Metal Samples MS2701E User manual

MS2701E
High Resolution ER RS-485 Transmitter
Operator’s
Manual
Metal Samples Company
ADivision of Alabama Specialty Products, Inc.
152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358-4202 Fax: (256) 358-4515

Table of
Contents
I.Introduction..............................................................................................................................................1
A.GeneralDescription...................................................................................................................1
B.PrinciplesofOperation..............................................................................................................1
C.TechnicalSpecifications.............................................................................................................3
II.InstallationandOperation.......................................................................................................................4
A.ReceivingtheMS2701Etransmitter..........................................................................................4
B.Installation.................................................................................................................................4
1.PhysicalMountingandProbeConnection...................................................................4
a.Direct‐ProbeMounting
......................................................................................4
b.RemoteMounting..............................................................................................5
2.PowerandDataConnections.......................................................................................6
a.MakingConnections.........................................................................................6
b.Grounding.........................................................................................................7
c.SafeAreaInstallation........................................................................................7
d.HazardousAreaInstallation...............................................................................7
3.SetupandOperation.......................................................................................................9
a.ProbeSettings.................................................................................................9
b.RS‐485CommunicationSettings...................................................................10
i.BaudRateandDefaultCommunicationParameters.............................11
ii.DeviceAddress......................................................................................11
iii.CommunicationProtocols......................................................................11
iv.TerminationResistor.............................................................................11
v.DataRegisters.......................................................................................11
c.CalibrationandTesting.................................................................................12
i.Calibration...............................................................................................12
ii.Testingoutputzeroandspan.................................................................12
iii.TestingtheMS2701EtransmitterwiththeMeterProver.......................12
d.InterpretingData...........................................................................................13
i.MetalLoss................................................................................................13
ii.CalculatingCorrosionRate.......................................................................14
e.Commissioning..............................................................................................15
C.Maintenance............................................................................................................................15
D.Troubleshooting........................................................................................................................16
III.ServiceandWarrantyInformation.......................................................................................................17
A.Warranty...................................................................................................................................17
B.ObtainingServiceandReturningtheInstrumentforRepair....................................................17
C.InstrumentRepairForm............................................................................................................18
AppendixA‐Drawings...............................................................................................................................19
ControlDrawing(HazardousAreaWiringDiagram)–StandAlone.............................................20
ControlDrawing(HazardousAreaWiringDiagram)–Multi‐Drop...............................................21
SafeAreaWiringDiagram.............................................................................................................22
AppendixA–RevisionHistory...................................................................................................................23

1
I.Introduction
A.GeneralDescription
ThemodelMS2701E transmitterisahighresolutioninstrumentdesignedtomeasureandtransmit
corrosion(metalloss)datafromanElectricalResistance(ER)probetoaplantcontrolsystemorother
recordingdevice.TheMS2701E transmitterutilizesRS‐485communicationwhichallowsmultiple
unitstobedaisy‐chained,simplifyinginstallationandreducingassociatedcosts.TheuseoftheRS‐485
protocolalsoallowsthetransmittertobeplacedgreatdistancesfromthecontrolsystemorrecorder
whilemaintaininggoodnoiserejection.Practicaldistancescanbeupto4000feet.Additionally,the
MS2701E transmitteroffersuser‐selectableRTUorASCIIcommunicationprotocolsmakingithighly
versatile.
TheMS2701E transmitteriscompatiblewithalltypesofMetalSamplesERprobes,aswellasany
standardERprobefromothermanufacturers.Unlikecompetitors’products,theMS2701E
transmitterdoesnotrequirefactorymodificationtoaccommodatedifferentprobetypes.Theprobe
typecanbeeasilychangedatanytimeusingtheon‐boardprobeselectionswitches(seepage9.)
TheMS2701E transmitterisavailableasdirect‐mountorremote‐mount.Thedirect‐mountversionis
mounteddirectlytotheERprobe.Thisoptionoffersthesimplestinstallationandminimizesnoise
problems.Theremote‐mountoptionallowstheMS2701E transmittertobemountedindependently
from(butincloseproximityto)theERprobe.Itisthenconnectedtotheprobeviaashortprobe
cable.(Seepages5and6formountingdiagramsandspecifications.)
B.PrinciplesofOperation
TheMS2701E transmitteroperatesontheElectricalResistance(ER)techniqueandisusedin
conjunctionwithanERprobe.TheERprobeutilizesaresistivesensingelementmanufacturedfrom
thematerialofinterest(oracloseapproximation)whichisexposedtoacorrodingenvironment.This
iscalledtheExposedorCorrodingElement.TheresistanceoftheExposedElementisdirectlyrelated
toitsthickness,soastheelementcorrodestheresultinglossofmetalcausesaproportionalincrease
intheelement’sresistance.TheprobealsocontainsaninternalReferenceElementwhichisusedto
compensatefortheinfluencesoftemperatureontheExposedElement.
TheMS2701E transmitterisdesignedtoworkwithanystandardERprobe,butitisrecommendedthat
CylindricalandLargeFlushtypeprobesbeusedtoensureoptimumperformance.Theirphysical
designplacestheReferenceElementincloserproximitytotheExposedElementcomparedtoother
probetypes,providingmoreeffectivetemperaturecompensationandthusreducingtheeffectsof
thermalnoise.
Becausetheyaredesignedtocorrode,ERprobesaresacrificialinnature.EachERprobewillhavea
finitelifethatisbasedontheelementthickness.ERprobesareavailableinanumberofgeometries
andthicknessesdesignedtosuitawidevarietyofapplications.Table1liststhecommonERelement
optionsavailablefromMetalSamplesandtheeffectivelifeofeach.

2
ElementTypeCompatibilityThicknessProbeLife
(Span)
TubularLoopCompatible
42
84
WireLoopCompatible
4010
8020
Flush(Small)Compatible
4 2
8 4
20 10
CylindricalPreferred
10 5
20 10
50 25
Flush(Large)Preferred
52.5
10 5
20 10
40 20
Table1.StandardERProbeElements
TheMS2701E transmittermeasuresanERprobeutilizingahigh‐resolution,16‐bitmeasurement.
Thisallowstheunittodetectmuchsmalleramountsofmetalloss,thusrespondingfasterto
corrosioneventsandupsets.At16‐bitresolutiontheMS2701E transmittercanmeasuremetalloss
amountsassmallas0.0015%oftheprobelife.
MetallossreadingstakenbytheMS2701E transmitterareoutputasadigitalRS‐485signalwhich
canbefedintoaplantcontrolsystemorotherrecordingdevices.TheMetalLossisrepresentedby
a16‐bitdigitalvaluebetween0and65535,where:
0=0mils(0%MetalLoss)
65535=ProbeLife*inmils(or100%MetalLoss)
*TheProbeLife(orSpan)oftheprobebeingused(seetable1above.)
TheMetalLossdatacanbeusedtocalculatethecorrosionratebetweenanyperiodsofinterest.
Thisiscoveredinmoredetailonpages13and14.MoreinformationonERprobes,theirtheory,
selection,andusecanbefoundonourwebsiteat http://www.alspi.com/erintro.htm.

3
C. Technical Specifications
Model MS2701E High Resolution ER RS-485 Modbus Transmitter
Physical Data
InstrumentWeight: 3.70 lb. (1.68 Kg)
Total Weight w/ Accessories: 5.76 lb. (2.61 Kg)
InstrumentDimensions: 8.0"H x 4.82"W x 3.45”D
(20.32cm H x 12.24cm W x 8.76”D)
OperatingTemperature: -4° to 158°F (-20° to 70°C)
StorageTemperature: -40° to 176°F (-40° to 80°C)
Enclosure Material: Cast Aluminum (Copper-Free)
MountingSpecifications: Direct probe mount or
Remote mount
(Up to a 2.5” (6.35cm) Dia. pole)
Performance Data
Measurement Type: ER measurement
Range: 0-100% of probe life
Resolution: 0.0015% of Probe Life (16-bit)
CycleTime: 75 Seconds
Electrical Data
PowerRequirements: 10 to 28 VDC
Maximum Probe Cable Distance: 30 ft (9.1 m)
Output Specifications: RS-485 Modbus, RTU or ASCII Protocol (Switch Selectable)
2400 / 4800 / 9600 / 19.2K Selectable Baud
32 Maximum Units (Addresses 1 to 32)
Hazardous Location Certifications – Transmitter
USA / Canada Conforms to ANSI/UL Std. 60079-0, 60079-11, 61010-1
CAN/CSA Std. E66079-0, 60079-11 & CAN/CSA C22.2 No. 61010-1
For Stand-Alone Use For Multi-Drop Use
Class I, Zone 1, AEx ia d IIC T4 Gb Class I, Zone 1, AEx ia d IIB T4 Gb
Zone 20, AEx ia d IIIC T130° C Da Zone 20, AEx ia d IIIB T130° C Da
-20° C ≤Ta ≤+70° C -20° C ≤Ta ≤+70° C
Europe and Worldwide For Stand-Alone Use For Multi-Drop Use
(ATEX and IECEx) II 1/2 G Ex ia d IIC T4 Gb II 1/2 G Ex ia d IIB T4 Gb
II 1/2 D Ex ia d IIIC T130°C Da II 1/2 D Ex ia d IIIB T130°C Da
- 20°C ≤Ta ≤+ 70°C - 20°C ≤Ta ≤+ 70°C
ATEX Certificate No: ITS14ATEX28092X
IECEx Certificate No: IECEx ITS 14.0052X
X - see Special Conditions (page 7)
Hazardous Location Certifications – Enclosure
Case Specifications: Explosion Proof (FM, CSA, CENELEC, UL)
IP 66, NEMA 4X, 7BCD, 9EFG
Included Accessories (Direct-Mount Model)
Meter Prover, Operations Manual
Included Accessories (Remote-Mount Model)
Meter Prover, Operations Manual, Probe Cable, Mounting Hardware Kit
Recommended Accessories
RS-485 Isolating Power Supply (Metal Samples Part # ET1548)
RS-485 Isolating Signal Repeater (Metal Samples Part # ET1549)
4.82”W
(12.24cm)
8.0”H
(20.32cm)
Figure 1. Overall Dimensions
3.45”D
(8.76cm)

4
II.InstallationandOperation
A.ReceivingtheMS2701ETransmitter
ChecktheMS2701E Transmitterforanyshippingdamagewhenitisfirstreceived.Whentheunitis
unpacked,verifythatthefollowingitemsareincluded:
Transmitter
MeterProver
User’sManual
ProbeCable(forremote‐mountonly)
MountingHardware(forremote‐mountonly)
Intheeventofshippingdamage,quantityshortage,ormissingitems,itisrecommendedthatthe
eventisdocumentedimmediatelyandthatdigitalphotographsaretaken.Anyshortagesormissing
itemsshouldbereportedtoMetalSamplesimmediately.Intheeventofshippingdamage,aclaim
shouldbeopenedwiththeresponsiblecarrier.
B.Installation
CAUTION:Usingthisproductinanywayotherthanthatspecifiedwithinthismanualmayimpairthe
intrinsicsafetyprotection.
InstallationoftheMS2701E transmitterinvolvesthefollowingsteps:
1. PhysicalMounting
2. ElectricalConnection
3. SetupandProgramming
1.PhysicalMountingandProbeConnection
WhenselectingalocationtomounttheMS2701E transmitteritisimportanttoconsiderthe
surroundingenvironment.Toensureproperoperation:
Donotmountthetransmitterinalocationthatexceedsitsoperatingtemperature.
Avoidmountingthetransmitternearsourcesofstrongelectricalnoise.
Ensurethatthereissufficientclearanceforinstallationandtoopenthetransmittercover
afterwards.
a.Direct‐ProbeMounting
TheMS2701E transmitterisdesignedfordirect‐probemountingwhicheliminatestheneedfor
additionalhardwareandtransmitter‐to‐probecabling.Thisgreatlysimplifiesinstallation,reduces
costs,andminimizeselectricalnoisethatcanbecoupledontoprobecablingfromnearbyelectrical
equipment.
BeforemountingtheMS2701E transmitter,firstensurethattheprobeisinstalledproperlyand
securely.Duringinstallationitisimportantthatyoudonotapplyexcessiveforceontheprobeor
seals,asdoingsocouldbreakthesealandresultinsystemleakage.
TomounttheMS2701E transmitter:

5
1. Alignthekeywaysofthetransmitterandprobeconnectors.
2. Insertthetransmitterconnectorplugfullyintotheprobeconnectorreceptacle.
3. Securethetransmittertothetopoftheprobebytighteningthecouplingnut.
NOTE:Hand‐tightissufficient.Donotover‐tightenthecouplingnut.
NOTE:Neverforcetheconnectorstomate.Ifthereisresistance,stopandcheckforbentpinson
theprobeandforforeignmaterialinthefemalesocketsofthetransmitterconnector.Gently
straightenanybentpinsandclearanyforeignmaterialthatmaybefound.
Figure2.DirectMountInstallation
b.RemoteMounting
Whenitisnotpracticaltodirect‐probemounttheMS2701E transmitter,theunitcanberemote
mountedinstead.Inthiscasetheinstrumentismountedtoaseparatemountingpoleusingthe
optionalRemoteMountingHardwareKit.ThetransmitteristhenconnectedtotheERprobeviathe
probeextensioncable.
Figure3.RemoteMountInstallation
AlignKey
Insert
Connector
Tighten
CouplingNut
123

6
2.PowerandDataConnections
a.MakingConnections
TheRS‐485fieldwiringentersthetransmitterthroughthe1/2"femaleNPTconduitportshownin
Figure4below.
CAUTION:Whenusedinahazardousarea,theconduitorcableconnectionsmustbemadeinsuch
awaythatallhazardousarearequirementsaremet.CableGlandsandblankingplugsmustbe
certifiedforprotectiontypes‘d’,’tb’anditshouldberatedforIP66.
Figure4.RS‐485Connector
ElectricalconnectionsaremadetothemainPCBviatheterminalblocksJ2andJ9asshowninFigure6.
NOTE:DonotconnectcableshieldingtotheTransmitter.Theshieldmustremainfloatingatthe
Transmitter.
TheRS‐485protocolallowsupto32transmitterstobedaisy‐chainedwhichsimplifiesinstallationand
reducescosts.Figure5showsanexampleofatypicalarrangement.Fullwiringdetailsareshownon
pages21and22.
CAUTION:Besuretoobservetheappropriategasgroupratinglistedonpage3forstand‐aloneor
multi‐dropoperation.
Figure5.Typicaldaisy‐chainarrangement,where‘n’isthenumberofthelasttransmitter.
1/2” FNPTConduitPortfor
FieldWiringEntry
ProbeConnectorStem
orProbeCable

7
ThemaximumpermissiblelengthofthefieldwiringbetweentheMS2701E transmitterandthe
controlsystemisdeterminedbythecontrolsystemsupplyvoltage,theelectricalresistanceofthe
cableandtheloadofthecontrolsysteminput.IftheTransmitteristobeinstalledinasafearea,refer
tosectionc.SafeAreaInstallationfordetails.IftheTransmitteristobeinstalledinahazardousarea,
refertosectiond.HazardousAreaInstallation.
b.Grounding
TheMS2701E transmitterenclosureisgroundedinternallythroughthewiringharness,butan
additional,externalgroundingterminalisprovidedaswell.Theenclosureshouldbegrounded
properlyusingtheexternalgroundingterminaltoensuresafeoperation.
c.SafeAreaInstallation
TheMS2701E transmitterisapprovedforuseinhazardousareas,butcanbeusedinnon‐hazardous
areasaswell.Forbasicsafeareawiringinformationrefertothecircuitdiagramshownonpage21.
CAUTION:Whenusedinnon‐hazardousareas,equipmentmustbesuppliedwithapre‐approvedpower
supplyunitorapprovedequipmentwhichmeetstheentityparametersshownbelow.
Thepre‐approvedequipmentmustbecertifiedtotheelectricalsafetystandardsforequipmentin
ordinarylocations(forexample,EN/IEC/CSAUL61010‐1orEN/IEC/CSAUL60950‐1).Failuretomeet
theserequirementswillvoidtheintrinsicsafetycertificationoftheequipmentandprohibititsfuture
useinhazardousareas.
d.HazardousAreaInstallation
CAUTION:Thissectionprovidesgeneralguidelinesforhazardousareawiring.However,regardlessof
anythingstatedhere,theMS2701E transmittermustbeinstalledinfullcompliancewiththecontrol
drawinglocatedonpage20andallofthelocalarearequirements.
CAUTION:WhenusedinHazardousareas,equipmentmustbesuppliedwithapre‐approvedpower
supplyunitorapprovedequipmentviaacertifiedintrinsicallysafebarrieroragalvanicallyisolated
barrier)withthefollowingentityparameters.
Input supply terminal RS-485 channel CN1 and connector CN2 (Combined)
Connector CN1 pins A & B (J2 on PCB): Ci = 68uF
Ui = 23.6 Vdc Li = 0
Ii = 352.8 mA Co = 32uF
Pi = 2.3 W Lo = 1.2 mH
Ci = 0
Li = 0 Input supply terminal, RS-485 terminal
and connector CN2 (combined)
RS-485 channel at each line Ui = 23 Vdc
Connector CN1 pins D & E (J9 on PCB): Ii = 538 mA
Ui = 3.7 Vdc Pi = 0.7 W
Ii = 225 mA Ci = 68uF
Pi = 0.206 W Li = 0
Uo = 5 V Co = 32uF
Io = 51 mA Lo = 0.14 mH
Po = 64 mW

8
Probe connector
Connector CN2 (J1 on PCB):
Uo=5V
Io= 352.8 mA
Po= 0.44 W
SpecialConditions
1. Probedielectricrating<500Vr.m.s.Donotexceed.
2. Donotinstalltheprobe(exposedportionofthesensingelements)intheexplosiondust
atmosphere(GroupIII).
3. Networked(“daisy‐chained”)arrangementnotforuseinGroupIIC/IIICenvironment.
4. Upto32transmittersareallowedinthedaisy‐chainarrangement.
5. OnlyCertifiedcableglandsandblankingplugsofprotectiontypes‘d’and‘tb’shouldbeused.
ProtectionratingofIP66shouldbemaintained.
Wheneveranelectricallydrivensensorormeasuringdeviceisusedinapotentiallyexplosive
environmentthemeasuringsystemmustbeinstalledinsuchawaythatelectricalenergyiseither
effectivelyisolatedfromtheexplosiveenvironment(viaexplosive‐proofcontainers,cableconduits,
etc.)ortheamountofelectricalenergyproducedinthehazardousareamustbelimitedtoan
intrinsicallysafelevel.
LimitingelectricalenergyisthemostpracticalmethodofprotectingtheMS2701E transmitterwhenit
isinstalledinahazardousarea.IntheMS2701E transmitter,electricalenergylimitsaremaintainedby
theuseofappropriatesafetybarriersinstalledinthepoweranddatawiring.Thesafetybarriersmust
belocatedinthesafeareaneartheboundarybetweenthesafeandhazardousareas.Thesafety
barrierwillrepeatthesignalcurrentgeneratedbytheTransmitterandwillrelaythesignaltothedata
receivingstation.
Caution:WhenasafetybarrierisusedwiththeMS2701E transmitter,thetransmitterwiring
connectionsmustbeconnectedtothesafetybarrier’shazardousareaterminals.Allotherconnections
mustbemadetothebarrier’ssafeareaterminals.
ThetypeofrepeatersafetybarriersusedwiththeMS2701E transmitterdependsonthespecific
classificationofthehazardousenvironmentinquestion.However,theentityparametersmustmatch
thoseoftheMS2701E transmitter.
CAUTION:Forhazardousareainstallations,themaximuminductanceandcapacitanceofthefield
wiringbetweenthesafetybarrierandthetransmittercannotexceedtheentityparametersofthe
selectedbarrier.

9
3.SetupandOperation
a.ProbeSettings
Theprobeselectionswitchesarelocatedonthetopleftcornerofthemainboardasillustratedin
Figure6below.TheseswitchesareusedtosettheERprobetype.Theyalsoallowtheinstrumentto
beplacedintoseveraltestmodeswhichoutputfixedvalues,allowingverificationoftheoutputsignal.
Table2givestheswitchsettingsforeachprobetypeandthevarioustestmodes.
Figure6.ConnectorandSwitchDetails
Switch1
Setting
Position
1 2 3 4
Operating
Positions
Wire Loop Off (↓)On (↑) On (↑) Off (↓)
Tube Loop/Flush On (↑) Off (↓)On (↑) Off (↓)
Cylindrical*Off (↓)Off (↓)On (↑) Off (↓)
Test
Positions
Zero Output (0) On (↑) On (↑) Off (↓)Off (↓)
Maximum Output
(65535) Off (↓)On (↑) Off (↓)Off (↓)
Full Scale Sweep On (↑) Off (↓)Off (↓)Off (↓)
Table2.ProbeSwitchSettings
*Cylindricalprobesettingisillustrated.
Switch1‐Probe
SelectionSwitch
(SeeTable2below)
PowerConnector
(J2)
ProbeCableConnector
(J1)
DataConnector
(J9)
Switch2‐ RS‐485
ConfigurationSwitches
(SeeTable3below)

10
b.RS‐485CommunicationSettings
TheRS‐485configurationswitchesarelocatedonthetoprightcornerofthemainboardasillustrated
inFigure6.Theseareusedtosettheinstrumentaddress,baudrate,communicationprotocol,andto
enabletheterminationresistor.Table3givestheswitchsettingsforeachoftheseparameters,and
theyareexplainedinmoredetailbelow.
Switch2
Device
Address
Position
34567
1Off (↓)Off (↓)Off (↓)Off (↓)Off (↓)
2On (↑)Off (↓)Off (↓)Off (↓)Off (↓)
3Off (↓)On (↑)Off (↓)Off (↓)Off (↓)
4On (↑)On (↑)Off (↓)Off (↓)Off (↓)
5Off (↓)Off (↓)On (↑)Off (↓)Off (↓)
BaudRatePosition6On (↑)Off (↓)On (↑)Off (↓)Off (↓)
127Off (↓)On (↑)On (↑)Off (↓)Off (↓)
2400On (↑)On (↑)8On (↑)On (↑)On (↑)Off (↓)Off (↓)
4800Off (↓)On (↑)9Off (↓)Off (↓)Off (↓)On (↑)Off (↓)
9600On (↑)Off (↓)10On (↑)Off (↓)Off (↓)On (↑)Off (↓)
19200Off (↓)Off (↓)11Off (↓)On (↑)Off (↓)On (↑)Off (↓)
12On (↑)On (↑)Off (↓)On (↑)Off (↓)
Communication
Protocol
Position13Off (↓)Off (↓)On (↑)On (↑)Off (↓)
814On (↑)Off (↓)On (↑)On (↑)Off (↓)
ASCIIOn (↑)15Off (↓)On (↑)On (↑)On (↑)Off (↓)
RTUOff (↓)16On (↑)On (↑)On (↑)On (↑)Off (↓)
17Off (↓)Off (↓)Off (↓)Off (↓)On (↑)
Termination
Resistor
Position18On (↑)Off (↓)Off (↓)Off (↓)On (↑)
1019Off (↓)On (↑)Off (↓)Off (↓)On (↑)
ActiveOn (↑)20On (↑)On (↑)Off (↓)Off (↓)On (↑)
InactiveOff (↓)21Off (↓)Off (↓)On (↑)Off (↓)On (↑)
22On (↑)Off (↓)On (↑)Off (↓)On (↑)
ReservedPosition23Off (↓)On (↑)On (↑)Off (↓)On (↑)
924On (↑)On (↑)On (↑)Off (↓)On (↑)
NotUsedN/A25Off (↓)Off (↓)Off (↓)On (↑)On (↑)
26On (↑)Off (↓)Off (↓)On (↑)On (↑)
27Off (↓)On (↑)Off (↓)On (↑)On (↑)
28On (↑)On (↑)Off (↓)On (↑)On (↑)
29Off (↓)Off (↓)On (↑)On (↑)On (↑)
30On (↑)Off (↓)On (↑)On (↑)On (↑)
31Off (↓)On (↑)On (↑)On (↑)On (↑)
32On (↑)On (↑)On (↑)On (↑)On (↑)
Table3.RS‐485ConfigurationSwitchSettings

11
i.BaudRateandDefaultCommunicationParameters
ThedefaultcommunicationparametersfortheMS2701E transmitterareprovidedinTable4below.If
necessary,thebaudratecanbechangedusingPositions1and2ofSwitch2,asillustratedinTable3.
ParameterValue
BaudRate19200
DataBits8
ParityEven
StopBits1
Table4.DefaultCommunicationParameters
ii.DeviceAddress
Upto32MS2701E transmittersmaybeconnectedinamulti‐dropnetwork(“daisy‐chained”)usinga
singlecablerun.However,eachtransmittermustbesettoauniquedeviceaddresstoavoidconflicts.
ThisisdoneusingPositions3through7ofSwitch2,asillustratedinTable3.
CAUTION:Besuretoobservetheappropriategasgroupratinglistedonpage3forstand‐aloneor
multi‐dropoperation.
iii.CommunicationProtocols
TheMS2701E transmitteroffersuser‐selectableRTUorASCIIcommunicationprotocols,givingtheunit
abroaderrangeofcompatibility.ThecommunicationprotocolissetusingPosition8ofSwitch2,as
illustratedinTable3.
iv.TerminationResistor
Toensureproperoperation,theRS‐485networkmustbeterminatedproperlywithatermination
resistor.Tofacilitatethis,eachMS2701E transmitterhasabuilt‐in120terminationresistorthatcan
beswitchedonoroffusingPosition10ofSwitch2.Onceallunitshavebeeninstalled,besureto
enabletheterminationresistorofthelastunitinthenetwork.Theterminationresistorsofallother
transmittersshouldbeturnedoff.
v.DataRegisters
Table5liststhedataregistersoftheMS2701E transmitter.
NoRegisterAddressDescription
140001DeviceStatus
240002DeviceAddress
340003BaudRate
440024DACValue
540026ProbeSwitchSetting
Table5.ModbusDataRegisters

12
c.CalibrationandTesting
i.Calibration
TheMS2701E transmitterisfullycalibratedwhenshippedfromthefactory.Thecalibrationsettings
arefixedtoavoidaccidentalchangewhichcouldresultinerroneousdata.Nofieldcalibrationis
required.However,itisimportanttotesttheunituponinstallation,andduringperiodicmaintenance
inspections,toensuretheunitisoperatingproperly.
ii.Testingoutputzeroandspan
TheMS2701E transmitterProbeSelectionSwitches(page9)offerthreetestsettingsthatallowthe
unittobeplacedintovariousdiagnosticmodesasfollows:
1) ZeroOutput–Forcesadigitaloutputof0.
2) MaximumOutput–Forcesadigitaloutputof65535.
3) FullScaleSweep–Causestheoutputtocontinuallycyclefrom0upto65535.
Thesetestmodescanbeusefulwhentroubleshootingcommunicationproblemsbetweenthe
transmitterandtheDCS/PLCsystem.
iii.TestingtheMS2701EtransmitterwiththeMeterProver
AMeterProverisprovidedtoallowroutinechecksoftheMS2701E transmitter.TheMeterProver
simulatesaWireLooptypeprobeatafixedvalue.TotesttheunitwiththeMeterProver:
1) Disconnectpower.
2) DisconnecttheMS2701E transmitterfromtheprobe(oriftheunitisremote‐mounted,
disconnecttheprobeextensioncablefromtheprobe.)
3) ConnecttheMeterProvertotheprobeconnectorstem(ortotheprobeextensioncableifthe
MS2701E transmitterisremote‐mounted.)
4) LoosentheEnclosureLockScrew.
5) Unthreadandremovethetransmittercover.
6) ChangetheProbeSelectionSwitchestotheWireLoopposition(seepage9.)
7) Reconnectpowerandallowtheinstrumenttomeasureforseveralminutestostabilize.
8) Afterseveralminutesobservethetransmitteroutput.Theoutputshouldcloselymatchthe
valueprintedontheMeterProverlabel.
IfthetransmitteroutputmatchestheMeterProvervalue,youmayreturntheProbeSelection
SwitchestotheirprevioussettingandreconnecttheMS2701E transmittertotheprobe.Ifthe
transmitteroutputshowsasignificantdifferencecomparedtotheMeterProvervalue,further
troubleshootingmayberequired.Refertopage16fortroubleshootingorcontactthefactoryfor
furtherassistance.
BesuretoreinstalltheenclosurecoverandtightentheLockScrewwhenputtingtheMS2701E
transmitterbackintoservice.

13
d.InterpretingData
i.MetalLoss
TheMS2701E transmittermeasurestheMetalLossofanERprobeandconvertsthatvaluetoadigital
16‐bitRS‐485Modbusoutput.ThisoutputisdirectlyproportionaltothecumulativeMetalLossofthe
ERprobeatanygiventime.Theoverallspanoftheoutputisproportionaltothelifeoftheprobein
use,soforDCS/PLCprogrammingtheRS‐485Modbusoutputcanbescaledasfollows:
0=0mils(0%MetalLoss)
65535=ProbeLifeinmils(100%MetalLoss)
TheProbeLifecanbefoundinTable1andisalsolistedontheMetalSamplesprobetag(asthe
“Multiplier”.)
Figure7.MetalSamplesProbeTag
Figure8illustratestherelationshipbetweenthedigitaloutputandMetalLoss.Inthisexamplea
probewitha10‐millifeisassumed.However,theXaxiscouldbechangedtorepresentanyProbeLife.
Figure8.OutputRelationship
Asseenhere,thetransmitteroutputbeginsat0foranewprobe(zeroMetalLoss)andincreasesas
theprobeelementcorrodes,eventuallyreachingamaximumof65535whentheprobesensing
elementhasbeencompletelyconsumedbycorrosion(inthiscase,10milsofMetalLoss.)Atthistime
theprobehasreacheditsend‐of‐lifeandmustbereplaced.
0
16384
32768
49151
65535
0.0 2.5 5.0 7.5 10.0
RS‐485ModbusOutput
MetalLoss(mils)
New
Probe
Probe
Consumed

14
ii.CalculatingCorrosionRate
Asexplainedpreviously,ERprobesandinstrumentsreportMetalLoss.However,thevaluethatisof
ultimateinterestisCorrosionRate.TheCorrosionRateisessentiallyMetalLossovertime,sothe
CorrosionRatecanbecalculatedusingthefollowingformula:
CorrosionRatempy ∆MetalLossmils
∆Timedays 365
where:
∆MetalLoss(mils)isthedifferencebetweentwoMetalLossreadings
Time(days)isthetimedifferencebetweenthosetworeadings
Itisacommonpracticetoprogramthisformulaintothecontrol(DCS/PLC)systemandhaveit
calculateCorrosionRateonacontinualbasisfromtheMetalLossdata.Thechallengeindoingthisis
selectinganappropriatetimeinterval.Usinganintervalthatistooshortmaygiveerraticresults,
whileselectinganintervalthatistoolongmaygiveresultsthatareinsensitivetosystemupsets.The
idealtimeperioddependsonmanyfactors,andwillvaryfromsystemtosystem.Itmaytakesome
trialanderrortosettleonthebesttimeperiodforyourinstallation.
InsomecasesitmaybenecessarytoreviewtherawMetalLossdataandmanuallyapplythe
CorrosionRateformulatoperiodsofinterest.Forexample,lookatthegraphbelowandseehow
muchthecalculatedCorrosionRatecanvarydependingonthetimeperiodused.
Figure9.CorrosionRatescalculatedfromMetalLossdata
Whileeachoftheresultsisvalidfortheselectedtimeperiod,theoneofmostinterestisthevalueof
11.81mpywhichrepresentssometypeofsystemupset.WhentheCorrosionRateiscalculated
automaticallyonapre‐selectedtimeperiod,thereisnoguaranteethattheselectedtimeperiodwill
alwayscoincidewithsystemupsetssuchasthis.Thatiswhymanualreviewandinterpretationof
MetalLossdataisalsohelpful.

15
e.Commissioning
OncetheMS2701E transmitterhasbeeninstalled,tested,andproperlyconfiguredfortheprobeinuse,it
canthenbeclosedandputintoservice.First,performonelastvisualinspectiontoensurethatall
electricalconnectionsaresecureandthattheenclosureo‐ringisinplaceandisingoodcondition.Then
threadtheenclosurelidontothebasefully.Oncethelidhasbeenthreadedintoplace,tightentheLock
Screwtopreventunauthorizedtampering.
Figure10.EnclosureLockScrew
CAUTION:TheLockScrewontheinstrumentbasemustbetightenedsecurelytopreventunauthorized
personnelfromopeningtheMS2701E
transmitter,andensurethattheintrinsicsafetyisnotviolated.Only
qualifiedpersonnelshouldbeallowedtoinstall,operate,andmaintenancetheMS2701E
transmitter.
C.Maintenance
Onceinstalled,theMS2701E transmitterrequireslittlemaintenance.However,itisimportanttoverifythe
followingitemsperiodicallytoensurecontinuedsafeoperation.
CAUTION:BeforeperforminganytestsormaintenanceontheMS2701E
transmitter,ensurethatall
hazardousarearequirementsaremet.
InspectionItemFrequency
Inspecttheenclosureo‐ringforanysignsofdamage.Replaceasnecessary.Annually
Inspecttheprobeconnectoro‐ringforanysignsofdamage.Replaceasnecessary.Annually
Inspectallelectricalconnectionsforsignsofcorrosion,mechanicaldamage,orforeign
matterthatcouldcauseimproperoperationorcauseanelectricalshort.Cleanasnecessary.
Annually
Ensurethatthelockingscrewisinplaceandissecure.Annually
Inspecttheenclosureforanysignsofcorrosionorpaintdamage.Annually
Checkforanysignsofmoistureingresswithintheenclosure.Annually
ContactMetalSamplesforreplacementpartsorifinstrumentrepairisnecessary.

16
D.Troubleshooting
IftheMS2701E transmitterdoesnotseemtoperformasexpected,checkthefollowingitems:
CAUTION:BeforeperforminganytestsormaintenanceontheMS2701E
transmitter,ensurethatall
hazardousarearequirementsaremet.
1. Ensurethattheprobeisoperationalandisnotcompletelycorroded.Thiscanbedoneintwoways.
a. TesttheprobewithaportableERmeterifavailable.
b. Testtheprobewithaportableresistanceorcontinuitymeterasfollows:
i. Connectonetestleadtopin‘A’oftheprobes6‐pinconnector.
ii. Measurecontinuitytoeachoftheotherpins.Thereshouldbecontinuity(low
resistance)toeachpin.
NOTE:Continuityoneachpindoesnotensurethattheprobeisgood.However,ifyou
findanopencircuitonanypinsthenitisalmostcertainthattheprobeisbadandshould
bereplaced.
2. EnsurethattheProbeSelectionSwitchesaresetcorrectlyfortheprobebeingused.Confirmthe
probetypeandrefertoTable2onpage9toverifytheappropriateswitchsettings.
3. Performavisualinspectionofthecircuitboardstolookforanysignsofmechanicalorelectrical
damage.
4. Ensurethatallelectricalcablesandwiringareingoodcondition.
5. Ensurethatallelectricalcontactsaresecureandfreeofcorrosion.
6. Ensurethatthereisadequatesupplyvoltageatthepowerconnector.
7. Verifythatthesupplyvoltagepolarityiscorrect.
8. Ifthereisinsufficientsupplyvoltageonthepowerconnector,checkthesafetybarrier(if
applicable)forablownfuseoranyotherfailure.
9. TesttheMS2701E transmitterusingthesuppliedMeterProver(seepage12.)
Thesebasicchecksshouldindicatethesourceofanyproblem(probe,powersupply,wiring,etc.).Ifitis
determinedthattheMS2701E transmitterismalfunctioning,orifyouneedfurtherassistancein
troubleshooting,contactMetalSamplesTechnicalSupport.
CAUTION:IftheMS2701E transmittershowsanysignsofdamage,removeitfromserviceimmediatelyand
consultthefactory.

17
III.ServiceandWarrantyInformation
A.Warranty
MetalSampleswarrantsthatanypartofthemodelMS2701E transmitterandaccessorieswhichprovesto
bedefectiveinmaterialorworkmanshipwithinoneyearofthedateoforiginalshipmenttoPurchaserwill
berepairedorreplaced,atMetalSamplesoption,freeofcharge.Thiswarrantydoesnotcover(1)probe
assemblies,(2)itemsexpendableinnature,or(3)itemssubjecttodamagefromnormalwear,misuseor
abuse,orfailuretofollowuseandcareinstructions.
AlldamageditemsaretobeshippedatPurchaser’sexpensetoandfromMetalSampleswhichshallhave
therighttofinaldeterminationastotheexistenceandcauseofadefect.
TheforegoingshallconstitutethesoleandexclusiveremedyofanypurchaserofMetalSamplesproducts
forbreachofwarrantyandISEXCLUSIVEANDINLIEUOFALLOTHERWARRANTIES,EXPRESSED,IMPLIED
ORSTATUTORY,INCLUDINGTHEIMPLIEDWARRANTIESORMERCHANTABILITYANDFITNESS.INNOEVENT
SHALLMETALSAMPLESBELIABLEFORSPECIALORCONSEQUENTIALDAMAGES,ORFORANYDELAYIN
THEPERFORMANCEOFTHISWARRANTYDUETOCAUSESBEYONDITSCONTROL.
Thetechnicalinformationandsuggestionscontainedhereinarebelievedtobereliable,buttheyarenotto
beconstruedaswarrantiessinceconditionsofusearebeyondourcontrol.
B.ObtainingServiceandReturningtheInstrumentforRepair
Ifyouexperienceproblemswithyourinstrumentpleasecontactthefactoryat256‐358‐4202andaskfor
customersupportforinstrumentation.Ourcustomersupportdepartmentwillassistyouin
troubleshootingyourinstrument.
Mostissuescanberesolvedoverthephone,butinsomecasesitmaybenecessarytoreturnyour
instrumentforfurtherevaluationandrepair.Inthiscase,pleaseobtainaReturnMaterialsAuthorization
(RMA)numberfromthesalespersonorsupporttechnician.ThisRMAnumberwillensurethatyour
instrumentisroutedtothecorrectdepartmentwhenitisreceivedatthefactory.
AfterreceiptofanRMAnumberyoumaypackyourinstrumentforreturn.Besuretopackyour
instrumentinasturdyboxandtopaditsufficientlytoavoiddamageduringtransit.Alsobesureto
completethe“InstrumentRepairForm”onthenextpageandincludeacopywithyourrepair.Thiswill
ensurethattherepairdepartmenthassufficientinformationregardingtheproblemsyouareexperiencing
withyourinstrument,aswellasthebilling,contact,andreturnshippingdetailsfortherepair.
OnceyouhaveobtainedanRMAnumber,completedthe“InstrumentRepairForm”,andpackedyour
instrumentsecurely,pleaseshipitprepaidtothefollowingaddress:
MetalSamples
152MetalSamplesRoad
Munford,AL36268
ATTN:RMA#_____
NOTE:BesuretolistyourRMAnumberintheattentionline(shownasblanksintheexampleabove.)

18
C.InstrumentRepairForm
ThisformmaybephotocopiedforusewhenreturninganinstrumenttoMetalSamplesforrepair.
Pleasefillinallknowninformationandencloseacopyofthecompletedformwiththeinstrument.
GeneralInformation
Model
Number
Serial
Number
RMA
Number
Dateo
f
Purchase*
*Ifknown.
ContactInformationforRepair
Contact
Name
Company
Phone
Number
E‐mail
Address
ReturnShippingInformation
Recipient
Name*
Company*
Return
Address
*Ifdifferentthanabove.
ReasonforReturn.(Provideasmuchdetailaspossible.Attachadditionalpagesifrequired.)
InvoiceInstructions(Fornon‐warrantyrepairs)
Invoicemefortherepair
(RequiresanopenaccountwithMetalsamples.)
Reference
PO#
Contactmeforcreditcardinformation
(Forsecuritypurposes,donotlistcreditcardinformationonthisform..)
Table of contents
Other Metal Samples Transmitter manuals

Metal Samples
Metal Samples CORR VELOX MS2801E User manual

Metal Samples
Metal Samples MS2500L User manual

Metal Samples
Metal Samples CORR Velox MS2800E User manual

Metal Samples
Metal Samples MS2600E User manual

Metal Samples
Metal Samples MS2601E User manual

Metal Samples
Metal Samples MS2901E User manual

Metal Samples
Metal Samples MS2700E User manual

Metal Samples
Metal Samples MS2500E User manual