METREL MA 2115 User manual

Resistance
MA 2115
MA 2115S
Decade
Instruction Manual
Code No. 20 750 678
Measuring and Regulation Equipment
Manufacturer
Ljubljanska 77, SI 1354 Horjul
Tel.: +386 1 7558 200; Fax: +386 1 7549 226
http://www.metrel.si; E-mail: [email protected]
Single resistor application Desired value has to be set by rotating
the appropriate knobs of rotary switches. Each knob contain
multiplier dial of marked range.
Example: to set resistance of 2850 kW select 2 x 1M,
8 x 100k, 5 x 10k and rest to 0. Banana plugs must be connected
into sockets that cover at least all positions different to 0.
Split chains If e.g. knob x 1k selects position, the MA 2115 is split
into high resistance decade part in range (0 – 9990 k) W and low
resistance decade part in range (0 – 999) W. Use appropriate
sockets.
Voltage divider Simple voltage divider can also be configured for
test purpose. Example: voltage divider with 1000:1 ratio and input
resistance of 2 MW has to be realized. Set 2 x 1M, 2 x 1k and rest to
0. Input voltage is connected between most left sockets and socket
in bottom right corner and for output voltage use socket between x
10 k and x 1 k and the same socket on right corner. If voltage
connected to divider is higher than 50 V connect PE into green
socket.
MAINTENANCE
Inspection To maintain operator safety and ensure reliability of MA
2115 it is good practice to inspect it on a regular basis. Check that the
enclosure is without defects such as scratches or breaks. If any defect
is found please consult repair centre, distributor and/or manufacturer.
Cleaning Use a soft cloth slightly moistened with soapy water or
spirit to clean the surface of the instrument and leave the instrument
to dry totally before using it.
Notes!
• Do not use liquids based on petrol or hydrocarbons!
• Do not spill cleaning liquid over the instrument!
Calibration It is essential, that MA 2115 be regularly calibrated. We
recommend an annual calibration.
Service Repairs under or out of warranty time: Please contact your
distributor for further information.
Resistance
MA 2115
MA 2115S
Decade
Instruction Manual
Code No. 20 750 678
Measuring and Regulation Equipment
Manufacturer
Ljubljanska 77, SI 1354 Horjul
Tel.: +386 1 7558 200; Fax: +386 1 7549 226
http://www.metrel.si; E-mail: [email protected]
Single resistor application Desired value has to be set by rotating
the appropriate knobs of rotary switches. Each knob contain
multiplier dial of marked range.
Example: to set resistance of 2850 kW select 2 x 1M,
8 x 100k, 5 x 10k and rest to 0. Banana plugs must be connected
into sockets that cover at least all positions different to 0.
Split chains If e.g. knob x 1k selects position, the MA 2115 is split
into high resistance decade part in range (0 – 9990 k) W and low
resistance decade part in range (0 – 999) W. Use appropriate
sockets.
Voltage divider Simple voltage divider can also be configured for
test purpose. Example: voltage divider with 1000:1 ratio and input
resistance of 2 MW has to be realized. Set 2 x 1M, 2 x 1k and rest to
0. Input voltage is connected between most left sockets and socket
in bottom right corner and for output voltage use socket between x
10 k and x 1 k and the same socket on right corner. If voltage
connected to divider is higher than 50 V connect PE into green
socket.
MAINTENANCE
Inspection To maintain operator safety and ensure reliability of MA
2115 it is good practice to inspect it on a regular basis. Check that the
enclosure is without defects such as scratches or breaks. If any defect
is found please consult repair centre, distributor and/or manufacturer.
Cleaning Use a soft cloth slightly moistened with soapy water or
spirit to clean the surface of the instrument and leave the instrument
to dry totally before using it.
Notes!
• Do not use liquids based on petrol or hydrocarbons!
• Do not spill cleaning liquid over the instrument!
Calibration It is essential, that MA 2115 be regularly calibrated. We
recommend an annual calibration.
Service Repairs under or out of warranty time: Please contact your
distributor for further information.
23
Mark on your equipment certifies that this equipment meets
the requirements of the EU (European Union) concerning
safety and electromagnetic compatibility regulations.
Producer:
METREL d.d.
Ljubljanska cesta 77
SI-1354 Horjul
Tel.: +386 1 75 58 200
Fax: +386 1 75 49 226
E-mail: [email protected]
Http://www.metrel.si
© 2003, 2005 Metrel
Distributor:
No part of this publication may be reproduced or utilised in any form
or by any means without permission in writing from METREL.
WARNINGS
• If the Resistance decade is used in a manner not speci- fied
in this Users Manual, the protection provided by the
equipment may be impaired!
• Do not use the Resistance decade if any damage is no-ticed!
• If the voltage of circuit connected to the Resistance de-cade
is higher than safety level its metallic enclosure must be
connected to PE (ground) terminal of circuit!
• Do not exceed ratings marked on front plate!
• MA 2115 does not contain any user replaceable compo-nents.
TECHNICAL DATA
Resistance range:
- standard: 0 to 9 999 999 W
- optional on request: 7 decades with possible
extension to 9 x 1 MW and/or 9 x 0.1 W
Accuracy:
- individual decade unit : ± 1% + 0.08 W
- 9 x 0.1 W option: ± 2 % + 0.08 W
- overall: ± 1% + 0.4 W
Maximum ratings:
- continuous power dissipation: 1 W / resistor
- voltage / current: marked on front plate
- working voltage to case: 500 V d.c. / a.c. rms
Dimensions (w × h × d): 267 × 89 × 97 mm
Weight: 0,8 kg
Protection classification: Class I
Over-voltage category: CAT III 300 V
Pollution degree: 2
Working temperature range: (0 to 40) °C
Storage temperature range: -10 °C to 80 °C
Max. humidity: 85 % RH (0 to 40) °C
APPLICATION HINTS
In general always predict and set required value before connection
into circuit and/or turning power on of it to avoid destructive short
circuit currents.
GENERAL
MA 2115 Resistance Decade is intended for all application areas
where variable resistances selected by hand are required. It is
passive electric device housed in metallic case. It consists of 7
decades with each of its setting rotary switch with range multiplier
from 0 to 9, and. Safety F 4 mm sockets are connected to each
resistance chain that it could be individually accessible. It is also
possible to split resistance chain into two or more independent
insulated groups by selection of rotary switch position.
23
Mark on your equipment certifies that this equipment meets
the requirements of the EU (European Union) concerning
safety and electromagnetic compatibility regulations.
Producer:
METREL d.d.
Ljubljanska cesta 77
SI-1354 Horjul
Tel.: +386 1 75 58 200
Fax: +386 1 75 49 226
E-mail: [email protected]
Http://www.metrel.si
© 2003, 2005 Metrel
Distributor:
No part of this publication may be reproduced or utilised in any form
or by any means without permission in writing from METREL.
WARNINGS
• If the Resistance decade is used in a manner not speci- fied
in this Users Manual, the protection provided by the
equipment may be impaired!
• Do not use the Resistance decade if any damage is no-ticed!
• If the voltage of circuit connected to the Resistance de-cade
is higher than safety level its metallic enclosure must be
connected to PE (ground) terminal of circuit!
• Do not exceed ratings marked on front plate!
• MA 2115 does not contain any user replaceable compo-nents.
TECHNICAL DATA
Resistance range:
- standard: 0 to 9 999 999 W
- optional on request: 7 decades with possible
extension to 9 x 1 MW and/or 9 x 0.1 W
Accuracy:
- individual decade unit : ± 1% + 0.08 W
- 9 x 0.1 W option: ± 2 % + 0.08 W
- overall: ± 1% + 0.4 W
Maximum ratings:
- continuous power dissipation: 1 W / resistor
- voltage / current: marked on front plate
- working voltage to case: 500 V d.c. / a.c. rms
Dimensions (w × h × d): 267 × 89 × 97 mm
Weight: 0,8 kg
Protection classification: Class I
Over-voltage category: CAT III 300 V
Pollution degree: 2
Working temperature range: (0 to 40) °C
Storage temperature range: -10 °C to 80 °C
Max. humidity: 85 % RH (0 to 40) °C
APPLICATION HINTS
In general always predict and set required value before connection
into circuit and/or turning power on of it to avoid destructive short
circuit currents.
GENERAL
MA 2115 Resistance Decade is intended for all application areas
where variable resistances selected by hand are required. It is
passive electric device housed in metallic case. It consists of 7
decades with each of its setting rotary switch with range multiplier
from 0 to 9, and . Safety F 4 mm sockets are connected to each
resistance chain that it could be individually accessible. It is also
possible to split resistance chain into two or more independent
insulated groups by selection of rotary switch position.

This manual suits for next models

1

Other METREL Measuring Instrument manuals

METREL MI 3290 User manual

METREL

METREL MI 3290 User manual

METREL EurotestXC MI 3152 User manual

METREL

METREL EurotestXC MI 3152 User manual

METREL A 1636 User manual

METREL

METREL A 1636 User manual

METREL MI 3102 Manual

METREL

METREL MI 3102 Manual

METREL MI 3321 MultiservicerXA User manual

METREL

METREL MI 3321 MultiservicerXA User manual

METREL A 1018 User manual

METREL

METREL A 1018 User manual

METREL VoltScanner MI 2130 User manual

METREL

METREL VoltScanner MI 2130 User manual

METREL A 1378 User manual

METREL

METREL A 1378 User manual

METREL SmarTEC User manual

METREL

METREL SmarTEC User manual

METREL MD 9272 User manual

METREL

METREL MD 9272 User manual

METREL MicroOhm 2A MI 3242 User manual

METREL

METREL MicroOhm 2A MI 3242 User manual

METREL MI 3103 User manual

METREL

METREL MI 3103 User manual

METREL MD 9260 User manual

METREL

METREL MD 9260 User manual

METREL A1199 User manual

METREL

METREL A1199 User manual

METREL MD 9235 User manual

METREL

METREL MD 9235 User manual

METREL A 1445 User manual

METREL

METREL A 1445 User manual

METREL A 1179 User manual

METREL

METREL A 1179 User manual

METREL GammaGT MI 3311 User manual

METREL

METREL GammaGT MI 3311 User manual

METREL A 1446 User manual

METREL

METREL A 1446 User manual

METREL DeltaGT MI 3309 User manual

METREL

METREL DeltaGT MI 3309 User manual

METREL MI 2792 PowerQ4 Plus User manual

METREL

METREL MI 2792 PowerQ4 Plus User manual

METREL MD 126 User manual

METREL

METREL MD 126 User manual

METREL MD 9272 User manual

METREL

METREL MD 9272 User manual

METREL CE MultiTesterXA MI 3394 User manual

METREL

METREL CE MultiTesterXA MI 3394 User manual

Popular Measuring Instrument manuals by other brands

Kimo MP 200 P user manual

Kimo

Kimo MP 200 P user manual

ProMinent Flow Control 3 operating instructions

ProMinent

ProMinent Flow Control 3 operating instructions

Kodak REELZ Film Digitizer user manual

Kodak

Kodak REELZ Film Digitizer user manual

wirsbo proMIX 101 installation manual

wirsbo

wirsbo proMIX 101 installation manual

EXFO FTB-5240S user guide

EXFO

EXFO FTB-5240S user guide

Rohde & Schwarz ZNA user manual

Rohde & Schwarz

Rohde & Schwarz ZNA user manual

LOVATO ELECTRIC DMG900 instruction manual

LOVATO ELECTRIC

LOVATO ELECTRIC DMG900 instruction manual

CWP Accurate pH 8 operating instructions

CWP

CWP Accurate pH 8 operating instructions

Riverbed Cascade Shark series quick start guide

Riverbed

Riverbed Cascade Shark series quick start guide

MULTISPAN AVH-16 quick start guide

MULTISPAN

MULTISPAN AVH-16 quick start guide

National Instruments FD-11603 manual

National Instruments

National Instruments FD-11603 manual

Ashcroft t5500e Installation and maintenance instruction manual

Ashcroft

Ashcroft t5500e Installation and maintenance instruction manual

Gehaka G810 user guide

Gehaka

Gehaka G810 user guide

Maxwell 25605 user manual

Maxwell

Maxwell 25605 user manual

HEIDENHAIN MSE 1201 installation instructions

HEIDENHAIN

HEIDENHAIN MSE 1201 installation instructions

Tintometer Lovibond MD100 instruction manual

Tintometer

Tintometer Lovibond MD100 instruction manual

Metro DataVac MET200 user manual

Metro DataVac

Metro DataVac MET200 user manual

Chopin AlveoPC Maintenance manual

Chopin

Chopin AlveoPC Maintenance manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.