MG F2DY User manual

EM-5426 Rev.7 P. 1 / 2
MG CO., LTD. www.mgco.jp
5-2-55 Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
INSTRUCTION MANUAL
BEFORE USE ....
Thank you for choosing us. Before use, please check con-
tents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact our sales office or representatives.
■PACKAGE INCLUDES:
Signal conditioner ...............................................................(1)
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
■CONFORMITY WITH EU DIRECTIVES
• The equipment must be mounted inside a panel.
• The actual installation environments such as panel con-
figurations, connected devices, connected wires, may af-
fect the protection level of this unit when it is integrated
in a panel system. The user may have to review the CE
requirements in regard to the whole system and employ
additional protective measures to ensure the CE conform-
ity.
• Install lightning surge protectors for those wires connect-
ed to remote locations.
■POWER INPUT RATING & OPERATIONAL RANGE
• Locate the power input rating marked on the product and
confirm its operational range as indicated below:
24V DC rating: 24V ±10%, approx. 80mA
■GENERAL PRECAUTIONS
• Before you remove the unit or mount it, turn off the power
supply and input signal for safety.
■ENVIRONMENT
• Indoor use.
• When heavy dust or metal particles are present in the
air, install the unit inside proper housing with sufficient
ventilation.
• Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
• Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 30 to 90% RH
in order to ensure adequate life span and operation.
• Be sure that the ventilation slits are not covered with ca-
bles, etc.
■WIRING
• Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
■AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 30 minutes is required for
satisfying complete performance described in the data
sheet.
COMPONENT IDENTIFICATION
Body
Zero Adj.
Span Adj.
Connection
Diagram
Specifications
Euro Type
Connector Terminal Block
INSTALLATION
Set the unit so that its DIN rail adaptor is at the bottom.
Hang the upper hook at the rear side of the unit on the DIN
rail and push in the lower. When removing the unit, push
down the DIN rail adaptor utilizing a minus screwdriver
and pull.
DIN Rail
35 mm wide
Spring Loaded
DIN Rail Adaptor
CURRENT LOOP SUPPLY MODEL F2DY
F2DY
EM-5426 Rev.7 P. 2 / 2
MG CO., LTD. www.mgco.jp
5-2-55 Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the connection diagram on the side of the unit.
■EXTERNAL DIMENSIONS unit: mm (inch)
1234
5678
107 (4.21) [1.5 (.06)]
DIN RAIL
35mm wide
22.5 (.89)
75 (2.95)
• When mounting, no extra space is needed between units.
■CONNECTION DIAGRAM
+
–
+
–
OUTPUT
3
5
6
POWER
7
8
2
4
1
+
–
+
–
–
+
MONITOR
(1 – 5V DC)
2-WIRE
TRANSMITTER
V
WIRING INSTRUCTIONS
■EURO TYPE CONNECTOR TERMINAL
Applicable wire size: 0.2 to 2.5 mm2
Stripped length: 7 mm
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal
7 – 8 with a multimeter.
3) Input: Check that the input signal is within 0 – 100% of
the full-scale.
4) Output: Check that the load resistance meets the de-
scribed specifications.
ADJUSTMENT PROCEDURE
This unit is calibrated at the factory to meet the ordered
specifications, therefore you usually do not need any cali-
bration.
For matching the signal to a receiving instrument or in case
of regular calibration, adjust the output as explained in the
following.
■HOW TO CALIBRATE THE OUTPUT SIGNAL
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 30 minutes.
1) ZERO: Apply 0% input and adjust output to 0%.
2) SPAN: Apply 100% input and adjust output to 100%.
3) Check ZERO adjustment again with 0% input.
4) When ZERO value is changed, repeat the above proce-
dure 1) – 3).
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 30 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
We offer a series of lightning surge protector for protec-
tion against induced lightning surges. Please contact us to
choose appropriate models.

Other MG Power Supply manuals

MG 10DU User manual

MG

MG 10DU User manual

MG R8-PS1 User manual

MG

MG R8-PS1 User manual

MG MDC6 User manual

MG

MG MDC6 User manual

MG Mini-M M2DY User manual

MG

MG Mini-M M2DY User manual

MG M80DY User manual

MG

MG M80DY User manual

MG 10DY User manual

MG

MG 10DY User manual

MG M6NPS2 User manual

MG

MG M6NPS2 User manual

MG HDY2 User manual

MG

MG HDY2 User manual

MG 18KDU User manual

MG

MG 18KDU User manual

MG Mini-M M2DYS User manual

MG

MG Mini-M M2DYS User manual

MG W5DY User manual

MG

MG W5DY User manual

MG 10DNY User manual

MG

MG 10DNY User manual

Popular Power Supply manuals by other brands

Mercury CB-R Series user manual

Mercury

Mercury CB-R Series user manual

Keysight Technologies N2779A user guide

Keysight Technologies

Keysight Technologies N2779A user guide

Recom RACM30-K/277/PMP Installation and operating instructions

Recom

Recom RACM30-K/277/PMP Installation and operating instructions

Kramer VA-100 user manual

Kramer

Kramer VA-100 user manual

Huawei TP48200A-DX08A1 user manual

Huawei

Huawei TP48200A-DX08A1 user manual

Arkuz BlueNet Power+ installation guide

Arkuz

Arkuz BlueNet Power+ installation guide

RELCO BIG JOLLY DIM DYNA/CLO manual

RELCO

RELCO BIG JOLLY DIM DYNA/CLO manual

Placid Industries PS-12 manual

Placid Industries

Placid Industries PS-12 manual

TDK-Lambda THR4 Installation, operation and maintenance manual

TDK-Lambda

TDK-Lambda THR4 Installation, operation and maintenance manual

Lumel ZS20 quick start guide

Lumel

Lumel ZS20 quick start guide

Utilitech SP-035 quick start guide

Utilitech

Utilitech SP-035 quick start guide

Dycon D1531 installation manual

Dycon

Dycon D1531 installation manual

TET ATLAS-1200 instruction manual

TET

TET ATLAS-1200 instruction manual

Akyga AK-ND-12 user manual

Akyga

Akyga AK-ND-12 user manual

Delton 301D operation instruction

Delton

Delton 301D operation instruction

Optimal Power OPR Power Series manual

Optimal Power

Optimal Power OPR Power Series manual

Zendure Satellite Monitor CT installation manual

Zendure

Zendure Satellite Monitor CT installation manual

Goobay 67953 user manual

Goobay

Goobay 67953 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.