MGE UPS Systems SA 2 Operating instructions

SA 2 & SA 2 IU
Medical Suction Collection Containers
Operating and Maintenance Manual
M.G. Electric (Colchester) Ltd
Altbarn Close, Colchester, Essex, CO4 9HY, United Kingdom
Tel: +44 (0) 1206 842244, Fax: +44 (0) 1206 845849
e-mail: sales@mgelectric.co.uk www.mgeworldwide.com

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 2 of 16
Dear Customer,
We take this opportunity to thank you for purchasing a SA Medical Suction Unit. Please read the
operating instructions and listed precautions thoroughly before attempting to operate the unit. MG
Electric manufactures its range in accordance with the requirements of BS EN ISO 9001 and BS EN
ISO 13485
INDEX
1. SAFETY INSTRUCTIONS 4
2. GENERAL DESCRIPTION 5
3. INSTRUCTIONS FOR USE 6
3.1
Before Operating Unit
3.2
SA 2 Set Up
3.3
Operating Environment
3.4
Connection
3.5
SA 2 Overflow Protection Valve
3.6
Cleaning Procedure
3.7
Transport
3.8
Storage
3.9
Instructions by Medical Staff to Patients
4. MAINTENANCE 10
4.1
Daily Procedures
4.2
General Layout
5. RECOMMENDED SPARES 11
5.1
Recommended Spares
6. TECHNICAL SPECIFICATION 12
6.1
General Dimensions
6.2
End of Life
7. NOTES 13
8. OTHER PRODUCTS IN THE SA RANGE 16

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 3 of 16
Return of Medical Equipment
Should you wish to return any equipment to MG Electric (Colchester) Limited (MGE), or one of our
designated distributors, Health Service Guideline HSG (93) 26 Decontamination of equipment prior
to inspection, service or repair) must be adhered to. Failure to follow this guideline will invalidate
any warranty claims and result in the equipment being destroyed.
Definition of symbols used in these instructions:
The Instruction for use must be referred to!
Manufacturers’ details and date of
manufacture
Safety Warning
Disposal in accordance with directive
2012/19/EU
Temperature Limits
Humidity Limits

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 4 of 16
1. SAFETY INSTRUCTIONS
The safety of the patient and SAM suction unit operator are the first priorities. It is therefore vital that
the following precautions are strictly observed:
WARNING!!!
•No modification of this equipment is allowed.
•Only original and approved spare parts and collection container systems must be used with
MG Electric products – failure to use original or approved spares will invalidate the warranty
and may cause injury or damage the SA 2/SA IU.
•Other than for routine daily procedures, any maintenance or repairs to MG Electric products
must be carried out by fully trained and qualified Electro-Biomedical engineer/technician
(EBME) or an authorised MG Electric dealer. Such persons are to be familiar with the relevant
standards, rules, accident prevention regulations, and operating conditions as a result of their
training, experience, and instruction. They are qualified to carry out the required activities and
in doing so recognize and avoid potential hazards. All testing on SA suction units should be
in accordance with ISO 10079-1
•Contamination may be present on any components. When cleaning or replacing any part of
the SAM suction unit appropriate protective clothing and gloves MUST be worn to avoid
contamination. Disposal of contaminated parts must be according to local protocols.
•Store the manual in a safe place, so that it is available to the trained personnel at all times.
•All SA 2 and SA IU collection containers must be securely mounted when in use.
•SA 2 and SA IU collection containers are NOT suitable for use in an MRI environment.
•The overflow valve may not operate fully against frothing. To prevent frothing anti-foam agent
maybe used.
•When replacing a full SA 2, be aware of its weight and ensure handling the container is
comfortable to avoid the possibility of spillage.
•Transport of the suction unit with a full SA 2 or SA IU collection container attached is not
advised.

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 5 of 16
2. GENERAL DESCRIPTION
SA 2
- Reusable collection container with a nominal capacity of 2 litres, used for the collection of
body fluids during suction therapy.
SA 2 IU
- A specialist version of the SA 2 collection container with a wide bore patient connection
for Intra-Uterine procedures.
All further references of SA 2 in this manual refer to both SA 2 and SA 2 IU connection containers.
The SA 2 includes a sealed top with integral shut-off valve, an integral handle and tube connections.
The shut-off valve provides prevent from contamination of the SA suction unit. The SA 2 is
connected to the SA suction unit via a moulded ‘V’ bracket and has graduations at intervals of 250
millilitres for volume indication.
The collection container does not come into contact with the patient under normal use – there are no
contra-indications.
The SA 2 collection container is fitted as standard to all SA suction units, and are connected to the
SA suction unit via a moulded vacuum connector of the same type as that fitted to the filters.
•Place the SA 2 in the bracket. Connect the tubing from the filter to the angle
connector in the centre of the lid.
•Connect the patient tube to the patient connection.
•Turn on the aspiration unit.
•Check the desired vacuum is established.
•After the suction procedure – disconnect the patient tube.
Patient
Connection
Filter
Angled
Connector
Angled
Connector

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 6 of 16
3. INSTRUCTIONS FOR USE
3.1 Before Operating Unit
BEFORE operating your new SA 2 collection container, please read the following instructions
carefully.
Become thoroughly familiar with the operation and maintenance of the SA 2 before use. Only
persons trained in its use should operate the suction unit.
3.2 SA 2 Set Up
Warning! When replacing a full SA 2 collection container, be aware of its weight and
ensure handling the container is comfortable to avoid the possibility of spillage.
Transport of the suction unit with a full SA 2 attached is not advised.
Warning! The SA 2 must be securely mounted when in use.
The SA 2 collection container is fitted as standard to all SA suction units, and are connected to the
SA suction unit via a moulded vacuum connector of the same type as that fitted to the filters.
•Place the SA 2 in the bracket. Connect the tubing from the filter to the angle
connector in the centre of the lid.
•Connect the patient tube to the patient connection.
•Turn on the aspiration unit.
•Check the desired vacuum is established.
•After the suction procedure – disconnect the patient tube.
3.3 Operating Environment
Operation of the SA 2 must be within the following ambient condition.
+40°C
-5°C
%
80%
30%
Temperature Humidity
Warning! Never operate a SA suction unit in the presence of flammable gas such as
anaesthetic agents. This is an Explosion hazard!
An anti-foam agent may be put into the SA 2, without disinfectant solution, before use to reduce the
possibility of frothing. It should not however, be placed into the SA 2 for extended storage periods.
3.4 Connection
The silicon service tube (∅6mm I/D with 3mm minimum wall thickness) must be connected between
the filter and the vacuum port on the SA 2 lid. The tube has an elbow connector on each end for
fitting to the filter and for fitting to the ‘VACUUM’ port of the SA 2. The silicon tubing used in
conjunction with this equipment is a replaceable item. It should be changed regularly according to the
level of usage and where it has become in any way contaminated or damaged.

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 7 of 16
3.5 SA 2 Overflow Protection Valve
The SA 2 is fitted with an overflow valve designed to shut off the vacuum when the fluid level reaches
1750ml, this will stop suction from the suction unit. When the valve operates, the suction unit must be
switched off and the full SA 2 replaced by an empty one. For SA suction units with two collection
containers and where appropriate the full SA 2 collection container may be replaced by the second
SA 2 through operation of the changeover valve located on the top of the unit.
It should be noted that even after the overflow valve has shut off, fluid might continue to be drawn into
the SA 2 to an extent dependent upon the level of vacuum in the SA 2 at the time when the valve
closed.
Warning! The overflow valve may not operate fully against frothing. To reduce frothing
anti-foam agent may be used. Liquid sucked through to the pump will cause
damage - As a precaution, a hydrophobic filter should be fitted to prevent liquid
passing through.
3.6 Cleaning Procedure
Warning! Contamination may be present on any components. When cleaning or
replacing any part of the SA 2 appropriate protective clothing and gloves
MUST be worn to avoid contamination. Disposal of contaminated parts must
be according to local protocols.
Warning! Solvent-based cleaning agents or abrasive cleaners must not be used on any
SA 2. Minimise contact with cleaning solution and rinse well immediately with
warm water. Prolonged immersion of parts in water above 60°C causes loss of
material properties and must be avoided. Do not soak in dilute or neat
disinfectant, as this will cause damage.
Warning! The black ‘O’ seal is a re-usable item but must not be autoclaved.
The SA 2 collection container is made from Medical Grade Polycarbonate and is autoclavable up to
138°C. The SA 2 must be cleaned and sterilised between uses. The SA 2 is connected to the
aspiration unit via silicon tube and two vacuum connectors. The tubing and connectors are
replaceable items and should be changed regularly according to the level and type of usage and when
it has become in any way contaminated or damaged. The black ‘O’ seal is a re-usable item but must
not be autoclaved.
The SA 2 can be cleaned, disinfected and sterilised by most of the well-known methods employed in
practice. Under certain conditions, however, contact with cleaning, disinfecting and sterilising media
may cause some damage, which manifests itself in the form of stress cracking. This usually leads to a
reduction in mechanical strength. This manual contains no specific recommendations as to suitable
cleaning agents and disinfectants for the SA 2because the choice is many, and the composition of
such products can change.
In order to avoid mechanical stress to the SA 2, it is advisable not to clamp or stack them during
cleaning, disinfection or sterilization.

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 8 of 16
3.6.1 Cleaning of Jar and Lid
In many cases it is sufficient to clean with warm or hot water to which some weakly acidic, neutral or
weakly alkaline cleaning agent has been added.
•Cleaning is a pre-requisite for all types of decontamination. Sterilisation will not be
effective on a jar that is still soiled. In addition, proteinaceous matter may become
baked onto the surface.
•Gloves, apron and eye protection must be worn when cleaning by hand.
•The jar, lid and other parts must be separated before cleaning to ensure all
surfaces that may be contaminated are cleaned.
•Special attention should be paid to the overflow valve assembly and passageway
attached to the lid.
•Always pre-clean SA 2 parts under water and in a deep sink using detergent and
hot water to remove visible contamination - A soft brush may be used.
•Items must be examined after the cleaning process, to ensure that all visible soiling
is removed.
•Rinse well to remove any cleaning agent and avoid surface damage.
3.6.2 Cleaning of the ‘O’ Ring Seal
In many cases it is sufficient to clean the ‘O’ seal with warm or hot water to which some weakly acidic,
neutral or weakly alkaline cleaning agent has been added. The ‘O’ ring seal is a re-usable item but
must not be autoclaved. Thoroughly dry the ‘O’ seal after cleaning. Inspect the ‘O’ ring seal after
cleaning to ensure it is not cracked or damaged in any way. Replace if necessary.
3.6.3 Sterilisation
Sterilising by steam autoclaving -
•Ensure removal of all visible soiling.
•The SA 2 must be dried prior to sterilisation. This reduces the likelihood of lime
scale build up and ensures adequate steam penetration.
•The SA 2 jar should be inverted on the tray.
•Sterilise in a high temperature Autoclave at 138°C for a minimum of 3 minutes.
In contrast to standard autoclaving conditions, i.e., 121°C for 15 to 30 minutes, the SA 2 may be
subjected to temperatures up to 138°C, thus reducing the amount of time needed for sterilisation.
When sterilising with steam, germicides and detergents must be rinsed thoroughly from parts prior to
autoclaving. Failure to thoroughly remove germicides and detergents from the parts prior to
autoclaving may cause accelerated degradation. Permanent immersion of parts in steam causes loss
of material properties and must be avoided.
Care must be taken to ensure that the SA 2 parts are not damaged by substances added to the
boiler feed water, such as alkaline corrosion inhibitors, and that the SA 2 jar is positioned in such a
way that no condensation can accumulate inside it as this may also cause damage.
The SA 2 complies with ISO 10079-1 and will undergo 30 sterilisation cycles before deformation may
become visible. Stress cracking and fracture may occur after further autoclave cycles. As a rule, it is
possible to sterilize the SA 2 many times before gradual chemical decomposition reduces the
mechanical strength to a level where it is no longer adequate for its application. Sterilization tests
have shown that even after 100 cycles of 30 minutes each at 120 to 125°C, the material still retains
comparatively good impact strength.
Service UK Health and Safety Act – The unit must be decontaminated internally and externally and
certified as such before returning for service. Help Line 01932 355277

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 9 of 16
3.7 Transport
The SA 2 will be adequately boxed and protected to ensure no damage occurs during normal
transportation of goods, providing the ambient conditions are within the following parameters:
Temperature max +60°C min -20°C Humidity max +80% min +30%
-20°C
There are no restrictions for land, air, or sea transport.
3.8 Storage
SA 2 must be stored in a dry, dust-free, well-ventilated environment. The storage environment
should not exceed the temperature and humidity conditions stated below. Avoid direct sun or UV
exposure and shield nearby sources of heat. The equipment should be stored in its original
packaging providing no damage is evident. Protect against ground moisture by storing on a shelf or
wooden pallet.
Temperature max +60°C min -20°C Humidity max +80% min +30%
-20°C
3.9 Instructions by Medical Staff to Patients
When the equipment is required by a patient for home use, Medical Staff must fully instruct the patient
on the safe operation of the equipment. In the event of equipment contamination or failure, the patient
must be advised to switch off the suction unit and contact the authority who loaned the equipment.

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 10 of 16
4. MAINTENANCE
With the need for hospitals to be able to ensure SA 2 collection containers are compatible with their
Sterile Services Department prior to purchase - the following information is provided for guidance and
the procedure outlined may be adapted to suit current practise.
Warning! Contamination may be present on any components. When cleaning or
replacing any part of the SAM unit appropriate protective clothing and gloves
MUST be worn to avoid contamination. Disposal of contaminated parts must
be according to local protocols.
Warning! No modification of this equipment is allowed.
4.1 Daily Procedures
•An anti-foam agent may be put into the SA 2, without disinfectant solution, before
use to prevent the possibility of frothing. It should not however, be placed into the
container for extended storage periods.
•Examine the collection container for damage. Replace if necessary.
•Examine the external tubing for ageing, damage, or contamination and replace if
necessary, using equivalent tubing.
4.2 General Layout
SA 2 SA 2 IU

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 11 of 16
5. RECOMMENDED SPARES
5.1 Recommended Spares
Only original and approved spare parts must be used with all SA 2 collection containers – failure to
use original spares will invalidate the warranty and may cause injury and/or damage to suction unit.
All spares can be purchased by the user from MG Electric directly. Not all available spares are listed
below. Please contact the MG Electric sales team for a full list. (sales@mgelectric.co.uk)
It is recommended that only competent persons should undertake the replacement of spare parts.
5.1.1 Collection Container Components
SA 2 2 litre autoclavable collection container SAM 2
SA 2 IU 2 litre autoclavable Intra-Uterine collection container SAM 2 IU
SA 2 bottle top assembly MSP1047
SA 2 IU bottle top assembly MSP1071
SA 2 overflow valve – 10pk MSP1048
SA 2 top grommets – 10pk MSP1049
SA 2 jar (No bottle top included) – 4pk MSP1050
SA 2 ‘O’ ring – 10pk MSP1051
5.1.2 Associated Components
Disposable bacterial filters – 24pk MSP1002
Disposable hydrophobic filters – 24pk MSP1003
Elbow connector and ‘O’ rings – 10pk MSP1004
Silicone tube (OD ∅12mm) – 25M MSP1156
SA patient tubing – 2M MSP1351

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 12 of 16
6. TECHNICAL SPECIFICATION
6.1 General Dimensions
Free air-flow (litres/minute): 50 lt/min nominal*
Vacuum: 700mmHg*
Performance classification: High vacuum High flow (ISO 10079-1)
Nominal collection container capacity: 2 Litres
Patient Connection: ∅6mm I/D bore – SA 2 ∅9mm I/D bore – SA 2 IU
*Measurements are quoted at sea level
6.2 End of Life
IMPORTANT INFORMATION
Correct disposal of the product in accordance with EC directive 2012/19/EU
At the end of its life, the product must not be disposed of as urban waste.
It must be taken to a special local authority differentiated waste collection centre or to a dealer
providing this service.

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 13 of 16
7. NOTES

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 14 of 16

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 15 of 16

SA 2 & SA 2 IU
Operation and Maintenance Manual
Manual 160 Issue 06
Page 16 of 16
8. OTHER PRODUCTS IN THE
SA
RANGE
Mains Powered Suction
SA 12 - General ward high vacuum suction unit
SA 14 – Twin jar minor operating theatre high vacuum suction unit
SA 15 - Intra-uterine aspirator suction unit
SA 16 – Twin jar intra-uterine aspirator suction unit
SA 17T - Thoracic Theatre suction unit
SA 17W - Thoracic Ward suction unit
SA 18 - Intensive care low vacuum suction unit
SA 19 - Twin jar Intensive care low vacuum suction unit
SA 35 - Major operating theatre high vacuum suction unit
SA 36 - Twin jar intra-uterine aspirator unit
SA S – Micro suction unit
Portable Suction
SA HSPY - General high vacuum suction unit
SA EPS - Battery powered portable suction unit (Neonatal Option available)
SA AUVAC - Portable foot operated suction unit
SA TVAC - Disposable, hand operated suction unit
Oxygen Flowmeters
SA XYFW - Oxygen flowmeter
SA XYHU - Oxygen humidifier
Pipeline Regulators
SA 50 - High vacuum pipeline regulator with remote probe
SA 51 - High vacuum pipeline regulator with direct probe
SA 52 - Low vacuum pipeline regulator with remote probe
SA 53 - Low vacuum pipeline regulator with direct probe
SA 54 - High vacuum pipeline regulator - remote probe & mobile trolley
Research and Development
Since 1954, when MGE produced their first surgical suction units, the SA range has become
accepted as the industry standard, both in the U.K. and throughout the world.
In recent years, the SA range has been greatly extended, with models now available for portable,
electrical suction and central suction requirements. They have been completely re-designed, using
lightweight, robust materials, achieving greater efficiency, and making them easier to clean and
operate. All MGE equipment is manufactured and assembled to very high standards of quality at the
modern factory in Colchester in accordance with BS EN ISO 9001 Quality Management System, BS
EN ISO 13485 and the Medical Device Directive 93/42/EEC.
© Copyright M.G. Electric (Colchester) Limited
All Rights Reserved.
Reproduction, adaptation, or translation without prior written permission is prohibited, except as
allowed under the copyright laws.
Manual 160 Issue 6 May 2022
This manual suits for next models
1
Table of contents
Other MGE UPS Systems Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Hologic
Hologic Aquilex Instructions for Use and Operator's manual

Hellenbrand
Hellenbrand Hill-Rom Evolution LI156E Series user manual

FLAEM NUOVA SpA
FLAEM NUOVA SpA Baby Flaem manual

Ossur
Ossur Figure 8 Clavicle Splint Instructions for use

Covidien
Covidien Puritan Bennett 560 user manual

Topcon
Topcon MC-4S user manual