
EPS320B Cardiac Stimulator
4
Lay out and connect Stimulus cabling, (under ground if required) between EPS320 Stimulus
Generator Unit (SGU) and EP Amplifier (LS Pro and CardioLab recorders shown):
Warning: To prevent patient micro-electrocution through capacitively coupled leakage currents,
MP3070 and MP3126 cable installation must be (i) appropriately insulated and separated from any nearby
mains wiring to prevent injection of mains voltage through broken insulation, (ii) must not be installed next
to mains voltage–energised cables and (iii) should not be installed within grounded conduits. Installation
should be acceptance tested by biomedical engineer for Patient Applied Part, Class CF device.
Note that when longer (>15m) length of Extension Stim Cable is required, it is preferable to place the
SGU in the operating room on top of the Amplifier under the patient table and use a long 80’/25m
serial cable MP3084-25 to interconnect PC with SGU. This will reduce noise and ground leakage
currents in the whole installation.
Note: for CardioLab II, ECG
output is found on the IEB
Labelled “Anlg Out - 1”.