
OPERATION:
•Connect the supplied 12 VDC
power adapter to the meter and
to the mains line.
•Connect the pH and ORP elec-
trode to the BNC socket of the
meter.
•Always remove the electrode
protective caps before taking any
measurement. If the electrodes
have been left dry, soak the tip
(bottom 2.5 cm) in rinse solution
(M10000B) for a few minutes to reactivate them.
•Make sure that the meter has been calibrated before taking any
measurements (see Calibration Procedure).
•Immerse the tip (2.5 cm) of the electrodes into the samples.
•Turn the instrument on by pressing the ON/OFF key.
•Allow the reading to stabilize and the meter will start continuous
monitoring.
•A blinking alarm LED will indicate when the measured pH value is
higher or lower than the setpoint depending on user selection.
•When the alarm LED is blinking, there is power in the output power
socket.
NOTE: The output power contact has no protection fuse inside the
meter. It is recommended to protected it outside, against failure.
SETPOINT:
•The setpoint can be selected by adjusting the knob to the desired
value.
•The selectable range for MC125 is from
4 to 8 for pH and from -200 to 600 mV
for ORP.
•The nature of the setpoint can be se-
lected by setting the proper switch to the
desired position (ABOVE or BELOW).
CALIBRATION PROCEDURE:
•Remove the protective cap from the pH electrode.
•Immerse the electrode tip into a new sachet of pH 7
calibration solution and allow the reading to stabilize.
•Adjust the pH7 calibration trimmer (on the front) to
display "7.00 pH".
•Open a new sachet of pH 4 calibration solution and
use a small quantity to rinse the electrode.
•Immerse the electrode tip inside the pH 4 sachet and
allow the reading to stabilize.
•Adjust the pH4/10 calibration trimmer (on the front) to
display "4.00 pH".
•The calibration is now complete and the meter is ready
to take measurements.
•It is suggested to recalibrate the meter at least
once a month, after prolonged warehousing time
and after pH electrode replacement.
INSTALLATION PROCEDURE:
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com