Mindray P8-2TU User manual

P7-3T/P7-3Ts/P7-3TE/P7-3TU/P8-2Ts/P8-2TU/P8-3Ts
Ultrasonic Transducer
Operator’s Manual


Table of Contents
Operator's Manual i
Table of Contents
Introduction.............................................................................................................. I
Intellectual Property Statement.............................................................................. I
Responsibility on the Manufacturer Party ............................................................ I
Note........................................................................................................................... I
Warning .................................................................................................................. II
Warranty .................................................................................................................. II
Exemptions ............................................................................................................. II
Customer Service Department .............................................................................. II
Important Information: Customer Responsibility............................................... III
Meaning of Signal Words...................................................................................... III
1 Safety Information .................................................................................................. 1
1.1 SafetyClassication...........................................................................................................1
1.2 SafetyPrecautions .............................................................................................................1
1.3 Ambientconditions.............................................................................................................3
1.4 Symbols .............................................................................................................................4
2 Overview.................................................................................................................. 5
2.1 Applications........................................................................................................................5
2.2 PrecautionsForUse ..........................................................................................................5
2.3 Out-of-BoxInspection ........................................................................................................5
2.3.1 AccessoriesInspection........................................................................................5
2.3.2 TransducerInspection .........................................................................................6
2.4 Structure.............................................................................................................................6
2.4.1 P7-3T/P7-3TETransducerFunctionsbyPart......................................................7
2.4.2 P7-3Ts/P7-3TUTransducerFunctionsbyPart ....................................................8
2.4.3 P8-2Ts/P8-2TUTransducerFunctionsbyPart ....................................................9
2.4.4 P8-3TsTransducerFunctionsbyPart ...............................................................10
3 Transducer Controls and Inspection .................................................................. 11
3.1 InspectionbeforeUse ......................................................................................................11
3.1.1 VisualandTactileInspection ............................................................................. 11
3.1.2 TipDeectionControl ........................................................................................12

Table of Contents
ii Operator's Manual
3.1.3 Brake Operation ................................................................................................14
3.1.4 ScanplaneRotationOperation ..........................................................................16
3.2 ConnectingandDisconnectingtheTransducer ...............................................................18
3.2.1 TEETransducerconnectorwithLockHandle ...................................................18
3.2.2 TEETransducerConnectorwithoutLockHandle..............................................19
3.3 ElectricalSafety ...............................................................................................................20
3.3.1 ElectricalLeakageCurrentTest.........................................................................20
3.3.2 Bite-holeInspectionTest ...................................................................................21
3.4 ThermalSafety.................................................................................................................22
3.4.1 ThermalLimits ...................................................................................................22
3.4.2 ControlSettings,Temperature...........................................................................23
3.4.3 Temperature Calibration Test.............................................................................23
3.5 CleaningandDisinfectingtheTransducer ......................................................................24
3.6 CheckingafterTurningONtheSystem............................................................................24
4 Examination .......................................................................................................... 25
4.1 Pre-ExamInspection........................................................................................................25
4.2 CoupleGel,Sheath,andBite-guard ................................................................................25
4.2.1 Couple Gel.........................................................................................................25
4.2.2 Sheath ...............................................................................................................26
4.2.3 Bite-guard ..........................................................................................................26
4.2.4 OrderingSupplies..............................................................................................27
4.3 Examination .....................................................................................................................27
4.3.1 Cautions ............................................................................................................27
4.3.2 OperationofDeectionControls .......................................................................28
4.3.3 EmergencyRetraction .......................................................................................29
4.4 After Examinations ...........................................................................................................29
5 Cleaning and Disinfection ................................................................................... 31
5.1 BeforeProcessing............................................................................................................32
5.2 Cleaning ...........................................................................................................................32
5.2.1 CompatibleDetergents......................................................................................32
5.2.2 CleaningProcedureforTEEProbe ...................................................................32
5.3 Disinfections.....................................................................................................................33
5.3.1 CompatibleDisinfectants...................................................................................33
5.3.2 DisinfectionProcedure ......................................................................................33
6 Storage and Transportation ................................................................................ 35
6.1 Transducer .......................................................................................................................35
6.2 Tipcover...........................................................................................................................35

This page intentionally left blank.

I
©2023ShenzhenMindrayBio-MedicalElectronicsCo.,Ltd.AllRightsReserved.
ForthisOperator’sManual,theissuedateis2023-12.
Introduction
Thisoperator’smanualdescribestheoperatingprocedureforthetransducer.Toensure
safeandcorrectoperationofthetransducer,readtheoperator’smanualcarefullyand
understandthetransducersclearlybeforeoperation.
Fortheoperatingproceduresfortheultrasonicdiagnosticsystemandotherdevices,please
refertotherelevantmanuals.
Intellectual Property Statement
SHENZHENMINDRAYBIO-MEDICALELECTRONICSCO.,LTD.(hereinaftercalled
Mindray)ownstheintellectualpropertyrightstothisMindrayproductandthismanual.This
manualmayrefertoinformationprotectedbycopyrightorpatentsanddoesnotconveyany
licenseunderthepatentrightsorcopyrightofMindray,orofothers.
Mindrayintendstomaintainthecontentsofthismanualasconfidentialinformation.
Disclosureoftheinformationinthismanualinanymannerwhatsoeverwithoutthewritten
permissionofMindrayisstrictlyforbidden.
Release,amendment,reproduction,distribution,rental,adaptation,translationoranyother
derivativeworkofthismanualinanymannerwhatsoeverwithoutthewrittenpermissionof
Mindrayisstrictlyforbidden.
, and arethetrademarks,registeredorotherwise,ofMindrayin
Chinaandothercountries.Allothertrademarksthatappearinthismanualareusedonly
forinformationaloreditorialpurposes.Theyarethepropertyoftheirrespectiveowners.
Thispostingservesasnoticeunder35U.S.C.§287(a)forMindraypatents:http://www.
mindrayna.com/patents.
Responsibility on the Manufacturer Party
Contentsofthismanualaresubjecttochangewithoutpriornotice.
Allinformationcontainedinthismanualisbelievedtobecorrect.Mindrayshallnotbeliable
forerrorscontainedhereinorforincidentalorconsequentialdamagesinconnectionwith
thefurnishing,performance,oruseofthismanual.
Mindrayisresponsiblefortheeffectsonsafety,reliabilityandperformanceofthisproduct,
onlyif:
• allinstallationoperations,expansions,changes,modifications andrepairsofthis
productareconductedbyMindrayauthorizedpersonnel;
• theelectricalinstallationoftherelevantroomcomplieswiththeapplicablenationaland
localrequirements;and
• theproductisusedinaccordancewiththeinstructionsforuse.
Note
Thisequipmentmustbeoperatedbyskilled/trainedclinicalprofessionals.

II
Warning
Itisimportantforthehospitalororganizationthatemploysthisequipmenttocarryouta
reasonableservice/maintenanceplan.Neglectofthismayresultinmachinebreakdownor
personalinjury.
Warranty
THISWARRANTYISEXCLUSIVEANDISINLIEUOFALLOTHERWARRANTIES,
EXPRESSEDORIMPLIED,INCLUDINGWARRANTIESOFMERCHANTABILITYOR
FITNESSFORANYPARTICULARPURPOSE.
Exemptions
Mindray’sobligationorliabilityunderthiswarrantydoesnotincludeanytransportationor
otherchargesorliabilityfordirect,indirectorconsequentialdamagesordelayresulting
fromtheimproperuseorapplicationoftheproductortheuseofpartsoraccessoriesnot
approvedbyMindrayorrepairsbypeopleotherthanMindrayauthorizedpersonnel.
This warranty shall not extend to:
• Malfunctionordamagecausedbyimproperuseorman-madefailure.
• Malfunctionordamagecausedbyunstableorout-of-rangepowerinput.
• Malfunctionordamagecausedbyforcemajeuresuchasreandearthquake.
• Malfunctionordamagecausedbyimproperoperationorrepairbyunauthorized
servicepeople.
• Malfunctionoftheinstrumentorpartwhoseserialnumberisnotlegible.
• Othersnotcausedbyinstrumentorpartitself.
Customer Service Department
Manufacturer: ShenzhenMindrayBio-MedicalElectronicsCo.,Ltd.
Address: MindrayBuilding,Keji12thRoadSouth,High-techindustrialpark,
Nanshan,Shenzhen518057,P.R.China
Website: www.mindray.com
E-mailAddress: service@mindray.com
Tel: +86 755 81888998
Fax: +86 755 26582680
EC-Representative: ShanghaiInternationalHoldingCorp.GmbH(Europe)
Address: Eiffestraβe80,Hamburg20537,Germany
Tel: 0049-40-2513175
Fax: 0049-40-255726
Manufacturer: MindrayDSUSA,Inc.
Address: 800MacArthurBlvd.
Mahwah,NJ07430-0619USA
Tel: +1(201)995-8000

III
TollFree: +1(800)288-2121
Fax: +1(800)926-4275
Important Information: Customer Responsibility
• Theresponsibilityformaintenanceandmanagementoftheproductafterdelivery
resideswiththecustomerwhohaspurchasedtheproduct.
• WhenthetransducerissenttoMINDRAYforwarrantyorrepair,youmustdisinfectit
andkeepitintheoriginalshipping/carryingcasetopreventinfection.
• Thewarrantydoesnotcover:
– Damagetothetransducerduetopatientbiting,
– Damagetothetransducercausedbydisinfectingincorrectlyorwithchemicalsnot
recommendedbyMindray
• ThewarrantywillbevoidifthetransducerisnotreturnedinitsoriginalMindray
carryingcase.
• Thisequipmentshallnotbeusedbypersonsotherthanfullyqualifiedandcertified
medicalpersonnel.
• Donotmakechangesormodicationstothesoftwareorhardwareofthisproduct.
• InnoeventshallMINDRAYbeliableforproblems,damage,orlosscausedby
relocation,modication,orrepairperformedbypersonnelotherthanthosedesignated
byMINDRAY.
• Thepurposeofthissystemistoprovidephysicianswithdataforclinicaldiagnosis.The
responsibilityfordiagnosticprocedureslieswiththephysiciansinvolved.MINDRAY
shallnotbeliablefortheresultsofdiagnosticprocedures.
• MINDRAYshallnotbeliableforlossofdatastoredinthememoryofthissystem
causedbyoperatorerrororaccidents.
• ThismanualcontainsWarningsregardingforeseeablepotentialdangers.Bealertatall
timestodangersotherthanthoseindicated.MINDRAYshallnotbeliablefordamage
orlossthatresultsfromnegligenceorfromignoringtheprecautionsandoperating
instructionscontainedinthisoperator’smanual.
• Ontheoccasionofchangeoftheadministratorormanagerforthissystem,besureto
handoverthisoperator’smanual.
• Whendisposingofthissystem,contactyourMINDRAYCustomerServiceDepartment
orsalesrepresentative.DonotdisposeofthissystemwithoutconsultingMINDRAY
CustomerServiceDepartmentorsalesrepresentativefirst.MINDRAYdoesnot
assumeanyresponsibilityfordamageresultingfromdisposalofthissystemwithout
consultingMINDRAY.Areplacementtransducerwillnotbesenttocustomeruntilthe
defectivetransducerisreceived.
Meaning of Signal Words
Inthisoperator’smanual, DANGER, WARNING, CAUTIONandNOTE are
signalwordsusedtoindicatesafetyandotherimportantinstructions.Thesignalwordsand
theirmeaningsaredefinedasfollows.Pleaseunderstandtheirmeaningsclearlybefore
readingthismanual.

IV
Signal Word Meaning
DANGER Indicatesanimminentlyhazardoussituationthat,ifnotavoided,
willresultindeathorseriousinjury.
WARNING Indicatesapotentiallyhazardoussituationthat,ifnotavoided,
couldresultindeathorseriousinjury.
CAUTION Indicatesapotentiallyhazardoussituationthat,ifnotavoided,
mayresultinminorormoderateinjury.
NOTE Indicatesapotentiallyhazardoussituationthat,ifnotavoided,
mayresultinpropertydamage.

Safety Information
Operator's Manual 1
1 Safety Information
1.1 SafetyClassication
Pleaserefertothesafetyclassicationinformationintherelatedoperator’smanualforthe
ultrasonicdiagnosticsystemthatmatcheswiththistransducer.
1.2 Safety Precautions
Pleasereadthefollowingprecautionscarefullytoensurethesafetyofthepatientandthe
operatorwhenusingthistransducer.
DANGER
DO NOT use flammable gases, such as anesthetic gas or hydrogen, or flammable
liquids such as ethanol, near this system, because there is danger of explosion.
WARNING
●This ultrasonic transducer is only for use with the specified diagnostic ultrasound
system. Please refer the diagnostic ultrasound system operation manual to select
the proper transducer.
●The multiplane TEE Transducer should be used only by a qualified physician who
has received appropriate training in proper operation of the transducer and in
endoscopic techniques as dictated by current relevant medical practices.
●The multiplane TEE transducer is a precision instrument, which must be handled
with care. It may be damaged when dropped or abused. In particular, do not allow
the ultrasonic window in the tip to come into contact with a sharp object. Do not
touch this window unnecessarily. Never exert force onto the window.
●To avoid injury to the patient, if any irregularity, substandard function or unsafe
condition is observed or suspected, the TEE transducer should not be used.
●The insertion part of the transducer meets the water-proof requirements for
soaking disinfection. However the control handle and transducer connector are
not water-proof. Keep the control handle and transducer connector out of any
cleaning or disinfection solutions
●To avoid injury to the patient and damage to the transducer, use a bite-guard
during all transesophageal exams.
●The use of a bite-guard is mandatory. Failure to use the bite-guard may result in
damage to the transducer, which could result in a safety hazard. Damage to the
transducer due to biting is not covered by the transducer’s warranty.

Safety Information
2 Operator's Manual
●To avoid injury to the patient, avoid forceful intubation pressure which can
cause lacerations of the gastrointestinal tract with attendant and subsequent
perforation.
●The deflection may after prolonged use develop an unwanted amount of free
play. In that case, contact the Mindray Customer Service Department or sales
representative to readjust the steering of the transducer. In this way, the risk of
“buckling” or “U-turning” of the transducer in the esophagus is minimized.
●Confirm that the transducer and cable are normal before and after each
examination. A defective transducer may cause electric shock to the patient.
●When using intra-cavity transducers, do not activate the transducer outside the
patient’s body.
●The connector is not watertight, and should always be kept dry. The control unit,
although spray-watertight, should not be immersed.
●Do not subject the transducer to shock. A defective transducer may cause electric
shock to the patient.
●This equipment contains no operator serviceable components. To prevent electric
shock, do not remove any covers or panels.
●DO NOT use the transesophageal probe with the defibrillator.
CAUTION
●When using this transducer, wearing medical gloves can help to prevent infection.
●To avoid inadvertent damage to the transducer, read this user guide before
handling and cleaning the TEE transducer.
●Be sure to use ultrasound gel. Please use the ultrasound gel compliant with the
relevant local regulations. Proper management and use of the ultrasound gel is
essential to ensure that it will not become a source of infection.
Only use water-based coupling gel.
●Under normal conditions at full acoustic power the temperature of the tip
does not exceed 43°C. Follow the instruction in this user manual to check this
regularly.
●If the laparoscopic probe is faulty, freeze the image and stop using the
transducer, and then slowly remove the transducer from the patient. Do not use
the laparoscopic transducer before maintenance.
●Before each use or after a change of viewing modes/settings, pay attention to the
status of the ultrasound image. Do not use the probe to perform image acquisition
when the image is frozen. Ensure that the orientation of the ultrasound image is
correct. Check the orientation prior to the examination
●Before each use, the outer surface of the portions of laparoscopic probe which
are intended to be inserted into a PATIENT should be checked to ensure there
are no unintended rough surfaces, sharp edges or protrusions which may cause
HARM.
●Do not use the carrying case for storing the transducer. If the carrying case is
used for storage, it may become a source of infection.

Safety Information
Operator's Manual 3
●It is required to practice ALARA when operating ultrasound system. Minimize the
acoustic power without compromising the quality of images.
●The transducer and accessories supplied with it are not delivered disinfected.
High-level disinfection is recommended before use.
●Disposable components are packaged sterile and are single-use only. Do not use
if integrity of packaging violated or if expiration date has passed. Please use the
disposable components compliant with the relevant local regulations.
●Please use the disinfection solution that recommended in this operator’s manual,
otherwise Mindray will not be liable for damage caused by other solutions. If you
have any questions, please contact Mindray Customer Service Department or
sales representative.
●Before introducing the transducer: do not rub or spray the tip of the transducer
with an anesthetic agent.
●The damage of the transducer may be caused by the contact of improper gel or
cleaner:
–DO NOT dip the transducer in the strong polar solutions of ethanol, chloride
of lime, ammonium chloride, acetone and formaldehyde.
–DO NOT contact the transducer with solution or ultrasound gel containing oily
medium such as mineral oil or lanoline.
NOTE:
●Read the following precautions to prevent the transducer from malfunction.
– Before connecting or disconnecting the transducer, freeze or turn off the
ultrasonic diagnostic system.
– Clean and disinfect the transducer before and after each examination.
●Repeated disinfection will eventually damage the transducer, please check the
transducer’s performance periodically.
1.3 Ambient conditions
Topreventthetransducerfrombeingdamaged,donotuseitwhereitwillbeexposedto:
• DirectsunlightorX-rays
• Suddenchangesintemperature
• Dust
• Excessivevibration
• Heatgenerators
Table 1-1 Environmental Conditions

Safety Information
4 Operator's Manual
Probe
model Conditions Ambient
temperature
Relative humidity
(no condensation)
Atmospheric
pressure
P7-3T
P7-3Ts
P7-3TE
P7-3TU
Operating 0 °C to 40 °C 10%to85%RH 700hPato1060hPa
Storageand
transportation
-10 °C to 45 °C 30%to90%RH 700hPato1060hPa
P8-3Ts Operating 0 °C to 45 °C 10%to90%RH 700hPato1060hPa
Storageand
transportation
-25 °C to 55 °C 5%to95%RH 700hPato1060hPa
P8-2Ts/
P8-2TU
Operating 17.5 °C to 42.7 °C 20%to85%RH 700hPato1060hPa
Storageand
transportation
-20 °C to 55 °C 20%to95%RH 700hPato1060hPa
1.4 Symbols
Thissystemusesthesymbolslistedinthefollowingtable.Theirmeaningsareexplainedas
follows.
Symbol Description
Generalwarningsign
Type-BFappliedpart
Serial number
Dateofmanufacture
Indicatesthedirectionofthelockhandle.Thetopsymbol
indicatesthepositionoflockedhandle,theleftsymbolindicates
thepositionoftheunlockedhandle.
Manufacturer
Protectionagainsttemporaryimmersioninwater
ComplywiththerequirementsoftheCouncilDirective93/42/EEC
(MedicalDeviceDirective).
Temperature limit
Humiditylimitation
Atmosphericpressurelimitation

Overview
Operator's Manual 5
2 Overview
2.1 Applications
TheTEEtransducerisusedtogetherwiththediagnosticultrasoundsystem,andisintended
forclinicalultrasounddiagnosisandexamination.Refertotheoperator’smanualofthe
diagnosticultrasoundsystem.
2.2 Precautions For Use
WARNING
The physician must take into account all possible factors before starting the
examination.
PrecautionsforusingtheTEEtransducerare:
• P7-3T/P7-3Ts/P7-3TE/P7-3TU/P8-2Ts/P8-2TUisintendedtobeusedonadultsonly,
fetalimagingandpediatricimagingarenotpermitted.
• P8-3Tsisnotintendedforfetalimaging.
• Imagingwhenthepatientexhibitsthefollowingorsimilarconditions:
– Esophagealstricture,spasms,lacerations,andtroubleswallowing(dysphagia)
– Esophagealdiverticula,esophagealvarices(swollenveins)
– Gastrointestinalbleeding
– Pepticulcers,hiatalhernia,esophagealwebsandrings
– Recentradiationtreatmenttotheesophagus
– Inabilityofthepatienttoswalloworaccommodatethetransducer
– Historyofgastroesophagealdiseases
– Othertherapiesthepatientmaybeundergoing
2.3 Out-of-Box Inspection
2.3.1 Accessories Inspection
Thefollowingitemsaresuppliedwitheachtransducer.
Item Quantity Description
Ultrasonictransducer 1 /
Operator’sManual 1 /
TEETransducerLeakage
CurrentSafetyTestingGuide
1 /

Overview
6 Operator's Manual
Item Quantity Description
Carryingcase 1 /
Tipcover 1 /
Transducercoversuite 1OnlysuppliedwithP8-2Ts/P8-2TU.
(IncludingNon-sterileTEEcover,gel-lled
syringeandbite-guard)
Tips:
Thetipcoverenclosesandprotectsthedistalend/scanheadoftheendoscopefrombeing
exposedtomechanicalstrainduringtransportationandstorage.
2.3.2 Transducer Inspection
ThefollowinginspectionsshouldbeperformedontheTEEtransducerafterunpackingthe
contents.
• Performvisualandtactileinspection,see"3.1.1 VisualandTactileInspection".
• Performtipdeectioninspection,see"3.1.2 TipDeectionControl".
• Performbrakeinspection,see "3.1.3 Brake Operation".
• Performscanplanerotationinspection,see"3.1.4 ScanplaneRotationOperation"..
• Performleakagetest,see"3.3 ElectricalSafety".
ContactMindrayoryourlocalrepresentativeimmediatelytoreportanydamageor
discrepancies.
WARNING
To avoid injury to the patient, if any irregularity, substandard function or unsafe
condition is observed or suspected, the TEE transducer should not be used.
2.4 Structure
TEEtransducerisaphasedarraytransducer,mountedinasealedtipattheendofa
gastroscope.Thetransducerisintendedforimagingoftheheartthroughtheesophagus
andthestomach.Thearraycanberotated0°–180°.

Overview
Operator's Manual 7
2.4.1 P7-3T/P7-3TE Transducer Functions by Part
12 3456 78
10
9
No. Name Function
1Distaltipwith
transducer
Convertstheelectricalsignalintoanultrasonic
signal,focusingthesoundbeamsinagivendirection;
meanwhile,itreceivesthereectedultrasonicsignaland
convertsitintoanelectricalsignalfortransmissionover
thecable.Thelensonthesurfaceistheacousticlens.
Applyultrasoundgelontheacousticlensforcorrect
operation.
2Deectionsection Deectstoobtainanoverallscanning.
3Flexibleshaft Insertsintothemouthcavityoresophagus.Depthis
markedontheexibleshaft,unit:cm.
4Controlhandle Operatesonthetransducerdeectionorrotation.
5Left/RightDeection
controlwheel
Usedforcontrollingthetransducerleft/righttipdeection.
6Up/DownDeection
controlwheel
Usedforcontrollingthetransducerup/downtipdeection.
7Deectionlock Restrainthedeectionwheeladjustment.
8 Cable Transmitselectricalsignalsbetweenthetransducerbody
andconnector.
9Motorcontrolpush-
buttons
Controlstoscanplanerotation.Range:0°–180°
10 Connector Connectsthetransducertotheultrasonicdiagnostic
system.

Overview
8 Operator's Manual
2.4.2 P7-3Ts/P7-3TU Transducer Functions by Part
1 2 3 4 5 6 7 8
10
9
No. Name Function
1Distaltipwith
transducer
Convertstheelectricalsignalintoanultrasonic
signal,focusingthesoundbeamsinagivendirection;
meanwhile,itreceivesthereectedultrasonicsignaland
convertsitintoanelectricalsignalfortransmissionover
thecable.Thelensonthesurfaceistheacousticlens.
Applyultrasoundgelontheacousticlensforcorrect
operation.
2Deectionsection Deectstoobtainanoverallscanning.
3Flexibleshaft Insertsintothemouthcavityoresophagus.Depthis
markedontheexibleshaft,unit:cm.
4Controlhandle Operatesonthetransducerdeectionorrotation.
5Up/DownDeection
controlwheel
Usedforcontrollingthetransducerup/downtipdeection.
6Left/RightDeection
controlwheel
Usedforcontrollingthetransducerleft/righttipdeection.
7Deectionlock Restrainthedeectionwheeladjustment.
8 Cable Transmitselectricalsignalsbetweenthetransducerbody
andconnector.
9Motorcontrolpush-
buttons
Controlstoscanplanerotation.Range:0°–180°
10 Connector Connectsthetransducertotheultrasonicdiagnostic
system.

Overview
Operator's Manual 9
2.4.3 P8-2Ts/P8-2TU Transducer Functions by Part
12 39
11
45678
10
No. Name Function
1Distaltipwith
transducer
Convertstheelectricalsignalintoanultrasonic
signal,focusingthesoundbeamsinagivendirection;
meanwhile,itreceivesthereectedultrasonicsignaland
convertsitintoanelectricalsignalfortransmissionover
thecable.Thelensonthesurfaceistheacousticlens.
Applyultrasoundgelontheacousticlensforcorrect
operation.
2Deectionsection Deectstoobtainanoverallscanning.
3Flexibleshaft Insertsintothemouthcavityoresophagus.Depthis
markedontheexibleshaft,unit:cm.
4Controlhandle Operatesonthetransducerdeectionorrotation.
5Deectionlocklever Restrainthedeectionwheeladjustment.
6Left/RightDeection
controlwheel
Usedforcontrollingthetransducerleft/righttipdeection.
7Up/DownDeection
controlwheel
Usedforcontrollingthetransducerup/downtipdeection.
8User-denedkey SetbytheuserintheUltrasoundsystempreset
9 Cable Transmitselectricalsignalsbetweenthetransducerbody
andconnector.
10 Motorcontrolpush-
buttons
Controlstoscanplanerotation.Range:0°–180°
11 Connector Connectsthetransducertotheultrasonicdiagnostic
system.

Overview
10 Operator's Manual
2.4.4 P8-3Ts Transducer Functions by Part
1 2 3 4 5 6 7 8
9
No. Name Function
1Distaltipwith
transducer
Convertstheelectricalsignalintoanultrasonic
signal,focusingthesoundbeamsinagivendirection;
meanwhile,itreceivesthereectedultrasonicsignaland
convertsitintoanelectricalsignalfortransmissionover
thecable.Thelensonthesurfaceistheacousticlens.
Applyultrasoundgelontheacousticlensforcorrect
operation.
2Deectionsection Deectstoobtainanoverallscanning.
3Flexibleshaft Insertsintothemouthcavityoresophagus.Depthis
markedontheexibleshaft,unit:cm.
4Controlhandle Operatesonthetransducerdeectionorrotation.
5Deectioncontrol
wheel
Thelargerwheelonthehandleisforcontrollingtheprobe
tipdeection.
6Scanplanerotation
wheel
Thewheelontopisthescanplanerotationcontrolwheel.
Controlstoscanplanerotation.Range:0°-180°.
7Deectionlock Restrainthedeectionwheeladjustment.
8 Cable Transmitselectricalsignalsbetweenthetransducerbody
andconnector.
9Connector Connectsthetransducertotheultrasonicdiagnostic
system.
This manual suits for next models
6
Table of contents
Popular Transducer manuals by other brands

controlair
controlair 595XP Installation, operation and maintenance instructions

Setra Systems
Setra Systems 209 installation guide

Signaltec
Signaltec TPS II installation manual

turck
turck IMX12-FI01-1SF-1I1R Series quick start guide

Viatran
Viatran 28H Installation data manual

Thyracont
Thyracont VSP63PN operating instructions

S+S Regeltechnik
S+S Regeltechnik THERMASGARD HFTM-VA Operating Instructions, Mounting & Installation

Fisher
Fisher 646 instruction manual

Hydac
Hydac HLT 1100 Series user manual

Lumel
Lumel P18D User Manual & Quick Start

CTC Union
CTC Union IFC-CCF40 Quick installation guide

Dynisco
Dynisco SJ600H Series instruction manual