MUX-100 SERVICE MANUAL 1
11
1
-
--
-
2
22
2
1
11
1-
--
-2
22
2-
--
-3
33
3
Entire system
Entire systemEntire system
Entire system
MUX-100 MUX-100J MUX-100JE
Total width 580mm
Total length 1150mm
Column height 1930mm 1780mm
Weight 370kg
Maximum traveling speed 5±0.5km/h (May vary depending on floor status.)
Cassette accommodation Light cassettes 14×17 (half size) 10 pcs
1
11
1-
--
-2
22
2-
--
-4
44
4
High voltage generator
High voltage generatorHigh voltage generator
High voltage generator
MUX-100 MUX-100J MUX-100JE
Model name BIR-320-125 BIR-200-125
Nominal maximum
power
12.5 kW (100kV, 125mA, 0.1s)
(product of maximum tube voltage by maximumtube current which can flow for 0.1 sec
at tube voltage of 100 kV)
Nominal maximum tube voltage and maximum tube current which can flow at nominal
maximumvoltage 125 kV, 100mA
Maximum tube current and maximum tube voltage which can realize maximum tube
current 160 mA, 80 kV
Combination of tube voltage and tube current which can realize maximum electrical
output 125 kV, 100 mA 100 kV, 125 mA 80 kV, 160 mA
Minimum tube current time product 0.32 mAs
Maximumtube current time product 320 mAs 200 mAs
Rating
Nominal shortest exposure time (during photo-timer radiography) 10 ms
Tube current time product
which can be set by tube
voltage
40 to 90 kV 0.32 to 320 mAs 0.32 to 200 mAs
91 to 100 kV 0.32 to 280 mAs 0.32 to 180 mAs
101 to 110 kV 0.32 to 250 mAs 0.32 to 160 mAs
111 to 120 kV 0.32 to 220 mAs 0.32 to 140 mAs
121 to 125 kV 0.32 to 200 mAs 0.32 to 125 mAs
Setting and display of
radiography tube voltage Setting range: 40 to 125 kV, increment = 1 kV Display: Digital display
Setting and display of
radiography tube current
time product
Setting range (200 mAs maximum only in MUX-100JE)
0.32, 0.36, 0.40, 0.45, 0.50, 0.56, 0.63, 0.71, 0.80, 0.90, 1.0, 1.1, 1.2, 1.4, 1.6, 1.8, 2.0, 2.2,
2.5, 2.8, 3.2, 3.6, 4.0, 4.5, 5.0, 5.6, 6.3, 7.1, 8.0, 9.0, 10, 11, 12, 14, 16, 18, 20, 22, 25, 28,
32, 36, 40, 45, 50, 56, 63, 71, 80, 90, 100, 110, 125, 140, 160, 180, 200, 220, 250, 280, 320
mAs
Display: Digital display
Anatomical program
8 regions×3 directions×3 types of body thickness = 72 types of anatomical programs are
available.
Following contents are stored in memory:
(a) Scan condition (kV, mAs) (b) Scan method
(c) Photo-timer lighting field selection (d) Photo-timer film density setting
Direct photo-timer scan
(option)
Photo-timer scans are available when optional phototimer light receiving unit is combined.
(a) Manual scan or photo-timer scan can be selected.
(b) Film density adjuster can be set to 7 steps (-3, -2, -1, 0, +1, +2, +3).
(c) Two types of combination of lighting field can be selected.