MolecuLight i:X User manual

MolecuLight i:XTM User Manual
MolecuLight i:XTM User Manual Revision 1.1 Page 2 of 43
Trademarks
MolecuLightTM, MolecuLight i:XTM, MolecuLight DarkDrapeTM, MolecuLight AdapterTM, Standard Imaging ModeTM,
Fluorescence Imaging ModeTM, ST-imageTM, ST-videoTM, FL-imageTM, FL-videoTM are trademarks of MolecuLight, Inc.
Patent Information
The MolecuLight i:XImaging Device (hereafter referred to as the device, MolecuLight i:X or MolecuLight i:X Imaging Device)
contains patented technology developed by MolecuLight Inc.:
Canadian Patent No. 2,724,973
U.S. Patent No. 12/992,040
Global patents and further U.S. patents are pending.
Publication Information
User Manual, Part Number 1294, Revision 1.1
Contact Information
The MolecuLight i:Xand its accessories are designed and manufactured by MolecuLight Inc.
MolecuLight Inc.
425 University Ave., Suite 700
Toronto, Ontario, Canada
M5G 1T6
Telephone: +1-647-362-4684, or 1-877-818-4360 (North America Only)
Fax: 647-362-4730
www.moleculight.com
The European authorized representative for MolecuLight Inc. is Emergo.
Emergo Europe
Prinsessegracht 20
2514 AP The Hague
The Netherlands

MolecuLight i:XTM User Manual Revision 1.1 Page 3 of 43
Contents
1INTRODUCTION ........................................................................................................................................................................ 5
2WARNINGS, CAUTIONS, AND NOTES ........................................................................................................................................ 5
2.1 WARNINGS,CAUTIONS,AND NOTES ..........................................................................................................................................................5
2.1.1 General Warning Messages...........................................................................................................................................................5
2.1.2 General Caution Messages ............................................................................................................................................................6
2.1.3 Acronym Table ..............................................................................................................................................................................7
2.2 SYMBOLS ..............................................................................................................................................................................................7
2.2.1 Symbols Used on the MolecuLight i:X Imaging Device Label ........................................................................................................ 7
2.2.2 Symbols Used on MolecuLight i:X .................................................................................................................................................8
2.2.3 Symbols Used on MolecuLight i:X Packaging ................................................................................................................................8
2.2.4 Symbols Used for MolecuLight DarkDrape Label ..........................................................................................................................9
2.2.5 Symbols Used for MolecuLight DarkDrape Package Label ............................................................................................................9
2.3 CERTIFICATIONS .....................................................................................................................................................................................9
2.3.1 Classifications ..............................................................................................................................................................................10
2.4 ELECTROMAGNETIC COMPATIBILITY .........................................................................................................................................................10
2.5 INFORMATION ON LASER RADIATION OUTPUT............................................................................................................................................10
3CONTENTS .............................................................................................................................................................................. 11
3.1 MOLECULIGHT I:XSYSTEM CONTENTS .....................................................................................................................................................11
3.2 MOLECULIGHT I:XACCESSORIES AND CONSUMABLES..................................................................................................................................11
4MOLECULIGHT I:X IMAGING DEVICE OVERVIEW..................................................................................................................... 12
4.1 MOLECULIGHT I:X IMAGING DEVICE DESCRIPTION AND USE .........................................................................................................................12
4.2 HOW MOLECULIGHT I:XCREATES IMAGES IN FL-MODE ..............................................................................................................................14
4.3 MOLECULIGHT DARKDRAPE AND MOLECULIGHT ADAPTER DEVICE DESCRIPTION AND USE.................................................................................15
5INTENDED USE AND INDICATION FOR USE ............................................................................................................................. 15
5.1 OFF LABEL USE ....................................................................................................................................................................................15
6DEVICE BASICS........................................................................................................................................................................ 15
6.1 POWER BUTTON ..................................................................................................................................................................................16
6.1.1 Status Indicator LEDs...................................................................................................................................................................17
6.1.1.1 System Status LED..........................................................................................................................................................................17
6.1.1.2 Range Finder LED ...........................................................................................................................................................................17
6.1.1.3 Ambient Light Status LED ...............................................................................................................................................................18
6.1.1.4 Battery Status LED..........................................................................................................................................................................19
6.1.2 Rocker Switch ..............................................................................................................................................................................19
7QUICK START GUIDE ............................................................................................................................................................... 19
8ENVIRONMENTAL CONDITIONS THAT AFFECT USE ................................................................................................................. 20
8.1 LIGHTING............................................................................................................................................................................................20
8.2 OPERATING TEMPERATURE ....................................................................................................................................................................20
8.3 STORAGE ............................................................................................................................................................................................20
9OPERATING INSTRUCTIONS.................................................................................................................................................... 20
9.1 PROPER IMAGING TECHNIQUE ................................................................................................................................................................20
9.2 HANDLING MOLECULIGHT I:X.................................................................................................................................................................21
9.3 IMAGING WITH MOLECULIGHT I:X(IXCAMERA APP) ..................................................................................................................................22
9.3.1 Capture an Image ........................................................................................................................................................................22
9.3.1.1 Capture an ST-image ........................................................................................................... ...........................................................22
9.3.1.2 Capture an FL-image ......................................................................................................................................................................23
9.3.1.3 Capture a video ..............................................................................................................................................................................23
9.3.1.4 Capture an ST-video .......................................................................................................................................................................23
9.3.1.5 Capture an FL-video .......................................................................................................................................................................24
9.4 REVIEW IMAGES AND/OR VIDEOS USING THE IMAGE LIBRARY .......................................................................................................................25
MolecuLight i:XTM User Manual Revision 1.1 Page 4 of 43
9.5 USING THE ALBUM FEATURE (OPTIONAL) .................................................................................................................................................25
9.5.1 Creating a New Album.................................................................................................................................................................26
9.5.2 Accessing and Using Previously Created Albums ........................................................................................................................27
9.5.3 Capture Image and Video Capture Screens.................................................................................................................................28
9.5.3.1 Capture Screens when the Active Album is the Camera Roll .........................................................................................................28
9.5.3.2 Capture Screens when the Active Album is Created ......................................................................................................................28
9.5.4 Deleting Albums ..........................................................................................................................................................................28
9.6 ZOOMING IN AND OUT AND PANNING ......................................................................................................................................................29
9.6.1 Zooming in and out .....................................................................................................................................................................29
9.6.2 Panning........................................................................................................................................................................................29
9.7 DELETING IMAGES AND VIDEOS ...............................................................................................................................................................29
9.8 CHARGING MOLECULIGHT I:X.................................................................................................................................................................29
9.8.1 Charge the MolecuLight i:X Imaging Device................................................................................................................................30
9.8.2 Charge the MolecuLight i:X Display Screen.................................................................................................................................30
9.9 UPLOADING IMAGES AND VIDEOS TO COMPUTER .......................................................................................................................................30
9.10 DISPLAY SCREEN FUNCTIONALITY.............................................................................................................................................................31
9.10.1 Basic Display Screen Functionality for MolecuLight i:X Use....................................................................................................31
9.10.1.1 Home Button..................................................................................................................................................................................31
9.10.2 Connect to a Wi-Fi Network....................................................................................................................................................32
9.10.3 Connect to the Internet ..........................................................................................................................................................32
9.10.4 Create a Passcode ...................................................................................................................................................................32
10 MEASURING WOUND AREA ................................................................................................................................................... 33
10.1 SETTINGS MENU ..................................................................................................................................................................................33
10.1.1 Wound Border Thickness ........................................................................................................................................................33
10.1.2 Display of Length & Width Dimensions...................................................................................................................................33
10.1.2.1 Length & Width ................................................................................................................ ..............................................................34
10.1.2.2 Vertical & Horizontal......................................................................................................................................................................34
10.2 AUTO MODE .......................................................................................................................................................................................34
10.3 MANUAL MODE...................................................................................................................................................................................35
10.3 SAVING A WOUND MEASUREMENT .........................................................................................................................................................35
11 INTERPRETATION OF FLUORESCENCE IMAGES ....................................................................................................................... 35
11.1 COLOR BLINDNESS................................................................................................................................................................................36
12 CLEANING AND DISINFECTING MOLECULIGHT I:X................................................................................................................... 37
12.1 PRE-CLEAN THE MOLECULIGHT I:X ..........................................................................................................................................................37
12.2 DISINFECT THE MOLECULIGHT I:X............................................................................................................................................................37
12.3 CLEAN THE MOLECULIGHT I:X.................................................................................................................................................................37
12.4 PRE-CLEAN THE MOLECULIGHT ADAPTER .................................................................................................................................................38
12.5 DISINFECT THE MOLECULIGHT ADAPTER ...................................................................................................................................................38
13 MAINTENANCE OF MOLECULIGHT I:X ..................................................................................................................................... 38
14 DISPOSAL OF MOLECULIGHT I:X ............................................................................................................................................. 38
15 TROUBLESHOOTING AND SUPPORT ....................................................................................................................................... 38
15.1 FREQUENTLY ASKED QUESTIONS .............................................................................................................................................................38
15.2 IXCAMERA APP TROUBLESHOOTING ........................................................................................................................................................41
16 WARRANTY ............................................................................................................................................................................ 42
APPENDIX A: SPECIFICATIONS ............................................................................................................................................................ 43
APPENDIX B: MOLECULIGHT I:X IMAGING DEVICE OVERVIEW
APPENDIX C: MOLECULIGHT I:X IMAGING DEVICE QUICK START GUIDE
APPENDIX D: MOLECULIGHT DARKDRAPE AND ADAPTER INSTRUCTIONS FOR USE
APPENDIX E: MOLECULIGHT I:X WOUND MEASUREMENT QUICK START GUIDE

MolecuLight i:XTM User Manual Revision 1.1 Page 5 of 43
1Introduction
Congratulations on your purchase of the MolecuLight i:XTM Imaging Device. The handheld MolecuLight i:XImaging Device
allows you to rapidly visualize potentially harmful bacteria in real-time during wound assessment and includes a wound
measurement application for quick wound area measurements. MolecuLight is confident that as you use the MolecuLight
i:XImaging Device you will find new and helpful information about the bacterial load in and around wounds during the
clinical examination process, helping you to better manage wound treatment for your patients.
The device is designed to aid and simplify clinical assessment, microbial sampling and treatment of wounds provided by
clinicians at the point-of-care. Its compact, handheld size and intuitive and simple user interface are designed for use across
hospitals and wound care clinical settings.
MolecuLight Inc. provides technical support, a User Manual, a Quick Start Guide and other educational references to aid you
with the use of the MolecuLight i:XImaging Device. Please review this User Manual thoroughly before operating your
device.
2Warnings, Cautions, and Notes
Please read the following safety information before using the MolecuLight i:XImaging Device.
The MolecuLight i:Xis an imaging device that does not require contact with the patient. It captures images/videos and
utilizes safe levels of violet wavelength light to illuminate a wound for fluorescence imaging. The MolecuLight i:XImaging
Device does not require the use of exogenous imaging contrast agents.
The MolecuLight i:Xis a Class II medical device per Health Canada Regulations.
The MolecuLight i:X is a Class IIa medical device per MDD 93/42/EEC.
MolecuLight i:Xis intended for use by trained health care professionals.
There are no known contraindications or side effects associated with the use of the MolecuLight i:X.
Some patients may experience some warmth from the device during exposure.
The Range Finder Sensor (see Figure 2, item 14) contains a fully-enclosed laser. No hazardous laser radiation is emitted
during use.
No eye or skin protection is required for the user or patient when operating the MolecuLight i:X.
Refer to Section 2.1.3 for the meaning of acronyms used throughout this User Manual.
2.1 Warnings, Cautions, and Notes
Warnings, cautions, and notes are used to describe serious and non-serious safety conditions of the device. The symbols
used to describe these safety conditions are:
Warning
Messages with this heading indicate serious adverse reactions, potential safety hazards, and
limitations in use imposed by a condition labeled with a warning. The warning identifi
es steps that
should be taken if the incident occurs.
Caution
Messages with this heading indicate information regarding any special care to be exercised by the user
and/or patient for the safe and effective use of the device. All caution statements should be followed
to ensure data and device integrity.
Note
Messages with this heading provide additional information that increase the user’s understanding of
the operation of the device.
2.1.1 General Warning Messages
Warning
The use of an accessory or cable with the MolecuLight i:XImaging Device other than those specified in
Section
3 may result in increased emissions or decreased immunity of the MolecuLight i:XImaging
Device.
MolecuLight i:XTM User Manual Revision 1.1 Page 6 of 43
Warning
The MolecuLight i:XImaging Device comes fully assembled and ready for use. No modification of this
equipment is allowed. Modification of the device will void the terms of the warranty.
Warning
Do not stack the
MolecuLight i:XImaging Device.
Warning
MolecuLight i:Xis intended to be used in a hospital/clinic by trained health care professionals. Avoid
exposure to magnetic fields, electrostatic discharge and thermal ignition sources
during use of the
device.
Warning
Do not charge or use the device in areas with potentially explosive atmospheres such as fueling areas
or in areas where the air contains chemicals or particles.
Warning
Do not point the violet wavelength LEDs directly into eyes (see
Figure 2, item 12).
Warning
MolecuLight i:Xis not suitable for use in the presence of flammable anesthetic mixtures with air,
oxygen or nitrous oxide.
Warning
Protect MolecuLight i:Xagainst dust and moisture by storing it in its original shipping package
overnight or when not in use for prolonged periods of time.
Warning
Avoid strong physical shocks and dropping.
Warning
Any damage to the device (crack or other visible deformity) may affect the functional, intended use
Warning
Do not soak or immerse MolecuLight i:Xin water or other liquids.
Warning
Be aware of surroundings when operating MolecuLight i:Xin a dark environment. The environment
should be safe to prevent tripping or bumping into any objects
during fluorescence imaging
procedures.
Warning
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the MolecuLight i:X,
including cables that may be specified by MolecuLight Inc. Otherwise, degradation of the performance
of MolecuLight i:Xcould result.
2.1.2 General Caution Messages
Caution
Do not attempt to use the device while it is charging. The device will not function.
Caution
Keep the Illumination Zone clean and avoid covering the Violet Wavelength LEDs and camera sensor
window with your fingers, as this may affect illumination and image quality (see Figure 2, item 15).
Caution
Prior to charging the device, ensure the MolecuLight Power Cable P/N 1142 (described in Section 3.1)
is undamaged before plugging into a wall outlet. Use of an extension cord or an electrical power bar is
discouraged.
Caution
When charging is complete, the MolecuLight Power Cable P/N 1142 should initially be disconnected
from the wall electrical outlet and then removed from the MolecuLight i:XImaging Device.
Caution
Use of controls, adjustments or procedures other than those specified in this User Manual may result
in hazardous laser light radiation exposure from the Range Finder Sensor (Section 2.5).
Caution
The MolecuLight i:XImaging Device is restricted to use by trained health care professionals and should
be protected from unauthorized use.
Caution
The Heat Sink (Figure 2, item 7) may get warm after prolonged use. The device will shut down if
temperature exceeds 460C or 1150F.
Caution
Improper cleaning/disinfection
of the device may result in distorted images.
Caution
Images/videos should be regularly downloaded and saved to avoid loss of patient information.
Caution
Non-MolecuLight Apps may appear on the MolecuLight i:Xfor download. Do not download any non-
MolecuLight Apps in order to maintain the device configuration intended for use.

MolecuLight i:XTM User Manual Revision 1.1 Page 7 of 43
2.1.3 Acronym Table
Acronym
Meaning
RF
Radio Frequency
ST-imageTM
An image captured in Standard Imaging Mode
ST-videoTM
A video captured in Standard Imaging Mode
FL-imageTM
An image captured in Fluorescence Imaging Mode
FL-videoTM
A video captured in Fluorescence Imaging Mode
LED
Light Emitting Diode
2.2 Symbols
2.2.1 Symbols Used on the MolecuLight i
i:
XImaging Device Label
The following table lists the symbols and statements used on the MolecuLight i:XImaging Device label (silver label).
Symbol
Meaning
Caution, consult accompanying documents
Read User Manual
Date of manufacture
Catalog number
Serial number
Certification of safety testing
Dispose of the MolecuLight i:Xin accordance with your country’s legal requirements for the disposal
of electrical and electronic waste
RF transmitter for wireless transmission
Device duty cycle (if device is
continuously on for 15 minutes in Fluorescence Imaging Mode, violet
LEDs should be turned off for 5 minutes before Fluorescence Imaging Mode is accessed again)
CE Marking indicates European Conformity to essential requirements of applicable European
product directives; 0123 represents MolecuLight Inc.’s Notified Body
Rx only
For prescription use
Statement
Meaning
CAN ICES-3(B)/NMB-3(B)
Self-Declaration of Compliance (SDoC) to Industry Canada ICES-003(B)
The MolecuLight i:X Imaging Device
complies with Part 15 of the FCC Rules
For further information related to this statement, refer to Section 2.3
for specific certification and Section 2.4
for electromagnetic
compatibility
For use only with Power Cable P/N:
1142. A utiliser uniquement avec Câble
N/P: 1142
For further information related to accessories and consumables, refer
to Section 3
Contains FCC ID:BCG-A1421 and IC:579C-
A1421
In reference to iPod touch (part of the MolecuLight i:X, referred to as
the Display Screen)
Table 1: Symbols and statements on MolecuLight i:XImaging Device label
MolecuLight i:XTM User Manual Revision 1.1 Page 8 of 43
2.2.2 Symbols Used on MolecuLight i
i:
X
The following table lists the symbols used in the labeling of MolecuLight i:X.
Symbol
Source & Meaning
Symbol
Source & Meaning
Source: IEC 60417-5010
Meaning: Power Button
Source: MolecuLight symbol
Meaning: Ambient Light Status
Source
: MolecuLight symbol
Meaning
: Display Screen Power Button
Source: MolecuLight symbol
Meaning: Port for Charging Display Screen and
Data Transfer
Source: MolecuLight symbol
Meaning: System Status
Source: MolecuLight symbol
Meaning: Fluorescence Imaging Mode
Source: IEC 60417-5546
Meaning: Battery Status
Source: MolecuLight symbol
Meaning: Standard Imaging Mode
Source: IEC 60417-5031
Meaning
: MolecuLight i:XCharging Port
(D.C)
Source
: MolecuLight symbol
Meaning
: Home Button (for Display Screen)
Source: MolecuLight symbol
Meaning: Range Finder
Table 2: MolecuLight i:Xsymbols
2.2.3 Symbols Used on MolecuLight
i:
XPackaging
The following table lists the symbols and statements used on the MolecuLight i:Xpackaging label and indicates storage and
transportation conditions.
Symbol
Source & Meaning
Symbol
Source & Meaning
Source: ISO 15223-1:2012
Meaning:
Maintain within -10o
C
to 50
oC (14oF to 122o
F) ambient
temperature
Source: WEEE symbol, EU Directive 2012/19/EU
Meaning:
Dispose of the MolecuLight i:Xin
accordance with your country’s legal requirements for
the disposal of electrical and electronic waste
Source:
ISO 15223-1:2012
Meaning:
Non sterile
Source:
ISO 15223-1:2012
Meaning:
Maintain within 10% to 70% humidity
Source:
ISO 15223-1:2012
Meaning:
Fragile
Source: ISO 15223-1:2012
Meaning:
Maintain within 89 kPa to 102 kPa for
atmospheric pressure
Source:
ISO 7000:2014
Meaning:
Face up
Source:
MolecuLight symbol
Meaning:
Do not stack or compress
Source:
ISO 7000:2014
Meaning
: Recyclable
Source:
ISO 15223-1:2012
Meaning
: Keep dry
Table 3: MolecuLight i:Xtransportation and storage conditions

MolecuLight i:XTM User Manual Revision 1.1 Page 9 of 43
2.2.4 Symbols Used for MolecuLight DarkDrape Label
The following table lists the symbols and statements used on the MolecuLight DarkDrape label.
Symbol
Source & Meaning
Symbol
Source & Meaning
Source:
ISO 15223-1:2012
Meaning:
Catalog number
Source:
ISO 15223-1:2012
Meaning:
Do not use if package is damaged
Source:
ISO 15223-1:2012
Meaning:
Batch code
Source:
ISO 15223-1:2012
Meaning:
For single use only
Source:
ISO 15223-1:2012
Meaning:
Date of manufacture
Source:
ISO 15223-1:2012
Meaning:
Non sterile
Source:
ISO 7010:2011
Meaning:
Read User Manual
Source:
IEC 60417 – 5840
Meaning:
Type B applied part
Source:
ISO 15223-1:2012
Meaning:
Caution, consult
accompanying documents
Source: European Conformity Marking
Meaning:
Manufacturer's declaration that the
product meets the requirements of the applicable EC
directives
Table 4: Symbols and statements on the MolecuLight DarkDrape label
2.2.5 Symbols Used for MolecuLight DarkDrape Package Label
The following table lists the symbols and statements used on the MolecuLight DarkDrape packaging label.
Symbol
Source & Meaning
Symbol
Source & Meaning
Source
: ISO 15223-1:2012
Meaning
: Catalog number
Source:
ISO 15223-1:2012
Meaning
: Keep dry
Source
: ISO 15223-1:2012
Meaning:
Batch code
Source:
ISO 15223-1:2012
Meaning
: Do not use if package is damaged
Source:
ISO 15223-1:2012
Meaning
: Date of manufacture
Source:
ISO 15223-1:2012
Meaning
: For single use only
Source:
ISO 7010:2011
Meaning
: Read User Manual
Source:
ISO 15223-1:2012
Meaning
: Non sterile
Source:
ISO 15223-1:2012
Meaning
: Fragile
Source:
IEC 60417 – 5840
Meaning
: Type B applied part
Source:
ISO 15223-1:2012
Meaning:
Caution, consult
accompanying documents
Source: European Conformity Marking
Meaning:
Manufacturer's declaration that
the product meets the requirements of the
applicable EC directives
Table 5: Symbols and statements on the MolecuLight DarkDrape package label (10-pack and 50-pack)
2.3 Certifications
The MolecuLight i:XImaging Device complies with the following standards:
xSafety Testing per IEC 60601-1:2005+A1:2012 / EN 60601-1:2006+A1:2013
including Canadian National Differences per CAN/CSA-C22.2 No. 60601-1:14 and
US National Differences per ANSI/AAMI ES 60601-1:2005/A1:2012
xSafety EMC Testing per IEC 60601-1-2:2014 4th Edition / EN 60601-1-2:2015
xSafety LED testing per IEC 60601-2-57:2011 and IEC 62471:2006 / EN 60601-2-57:2011 and EN 62471:2008
xUsability Testing per IEC 60601-1-6:2010+A1:2013 / EN 60601-1-6:2010 and EN 62366:2008
xRisk Management per ISO 14971:2007 / EN ISO 14971:2012
MolecuLight i:XTM User Manual Revision 1.1 Page 10 of 43
xQuality Management System Requirements per ISO 13485:2016 / EN ISO 13485:2016
The MolecuLight DarkDrape complies with ISO 10993-1:2009 / EN ISO 10993-1:2009: Biological evaluation of medical
devices.
2.3.1 Classifications
MolecuLight i:Xhas an operational duty cycle of 15 minutes on, 5 minutes off. The implication for this duty cycle is that if
the MolecuLight i:Xis continuously used in Fluorescence Imaging Mode for 15 minutes, the device will need to be switched
to Standard Imaging Mode for 5 minutes before selecting Fluorescence Imaging Mode again to allow for optimal
performance of the Violet Wavelength LEDs.
The medical device classification of the MolecuLight i:Xin Canada is Class II medical device and a Class IIa medical device in
Europe. Alone, the MolecuLight i:Xdoes not require patient contact for use. The MolecuLight DarkDrape (used in
combination with the MolecuLight Adapter and MolecuLight i:Xto achieve an optimal lighting environment for imaging
wounds) is a Type B patient applied part which complies with medical device biocompatibility requirements.
The MolecuLight i:Xis classified as “IPX0” for fluid ingress, therefore it has no inherent protection against damage from
exposure to liquids.
2.4 Electromagnetic Compatibility
The MolecuLight i:XImaging Device meets the requirements of Safety EMC Testing per IEC 60601-1-2:2014 4th Edition / EN
60601-1-2:2015. Medical electrical equipment requires special precautions regarding electromagnetic compatibility (EMC)
and must be used according to the instructions in this User Manual.
While unlikely, it is possible that high levels of emitted radio-frequency (RF) electromagnetic interference (EMI) from other
portable and mobile RF communications equipment or nearby radio-frequency sources could result in performance
disruption of the MolecuLight i:XImaging Device.
To avoid the risk of increased electromagnetic emissions or decreased immunity from such emissions, use only accessories
recommended by MolecuLight (i.e. MolecuLight Power Cable, see Section 3.1). Connection of accessories not recommended
by MolecuLight will void product warranty and could result in malfunctioning of the MolecuLight i:XImaging Device or other
devices located in the area.
This device complies with Part 15 of the US FCC Rules. This device complies with Canadian ICES-003 (B).
Note
The emissions characteristics of the MolecuLight i:Xmake it suitable for use in industrial areas and
hospitals (CISPR 11 class A).
MolecuLight i:Xis not intended for use in a residential environment
and does
not comply with
IEC 60601-1-11:2015 General requirements for basic safety and essential performance --
Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in
the home healthcare environment.
2.5 Information on Laser Radiation Output
The MolecuLight i:XImaging Device uses a miniature pulsed laser-based range finder to determine the optimal distance
between the device and the wound for superior image quality. The laser module in the range finder emits light at 850 nm
which is invisible to the human eye. The laser’s individual pulse duration is 3.33 ns and the pulse train is 52.3 ms at a 200 ms
repetition rate. The laser complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No.50.
The laser is a Class 1 certified laser
Caution
Do not point the laser (from the
Range Finder sensor) into eyes (Figure 2, item 14).

MolecuLight i:XTM User Manual Revision 1.1 Page 11 of 43
3Contents
Figure 1: MolecuLight i:XImaging Device and its components
3.1 MolecuLight i
i:
XSystem Contents
The MolecuLight i:XImaging Device comes fully assembled and ready for use. The standard system configuration includes
items 1-6 illustrated in Figure 1 and detailed in Table 6 below. These items allow the MolecuLight i:XImaging Device to be
used in a patient care setting.
# in
Figure 1
Item Description
1
MolecuLight i:XImaging Device
Device operates on Apple iOS software
2
MolecuLight Power Cable
Cable to charge the MolecuLight i:XImaging Device and download
images/videos to a computer; Includes geography
-
specific plug adaptors (where
applicable)
3
Quick Start Guide
Step by step guide for MolecuLight i:Xuse
4
User Manual
Provides detailed instructions on use of the MolecuLight i:X
5
Optical
Lens Wipes
5 x Optical lens cleaning towelettes, individually packaged for cleaning display
screen and optical windows (ProWorks)
6
MolecuLight WoundStickers
1 box of 2,000 stickers for wound area measurement with the MolecuLight i:X
Table 6: List of items included with the MolecuLight i:XSystem (P/N 1153)
Note
The MolecuLight DarkDrape and MolecuLight Adapter (Figure 1, items 7-8) are not part of the
MolecuLight i:XSystem configuration. They are optional accessories that may be ordered separately.
3.2 MolecuLight
i:
XAccessories and Consumables
MolecuLight i:Xaccessories and consumables are listed below in Table 7 along with suggested vendors from which the
items may be purchased (if not available from MolecuLight).
MolecuLight i:XTM User Manual Revision 1.1 Page 12 of 43
# in
Figure 1
Item Description
MolecuLight
Part #
Reorder from MolecuLight
2
MolecuLight
Power Cable
Cable to charge the MolecuLight i:XImaging
Device
and download images/videos to a
computer; Includes geography
-specific plug
adaptors (where applicable)
1142 Yes
3
Quick Start Guide
Step by step guide for MolecuLight i:Xuse
1295
Yes
4
User Manual
Provides detailed instructions on use of the
MolecuLight i:X
1294 Yes
5
Optical Lens
Wipes
(Box of
100)
Box of 100 optical lens cleaning towelettes,
individually packaged for cleaning
display
screen
and optical windows (ProWorks)
1182 Yes
6
MolecuLight
WoundStickers
1 box of 2,000 stickers for wound area
measurement with the MolecuLight i:X
1266 Yes
7
MolecuLight
Adapter
Reusable Adapter to connect MolecuLight
DarkDrape to the MolecuLight
i:Xto create an
optimal lighting environment for imaging wounds
1215 Yes
8
MolecuLight
DarkDrape
(Case
of 50)
Case of 50 MolecuLight DarkDrapes, individually
packaged for one
-
time use to create an optimal
lighting environment for imaging wounds
1212 Yes
N/A
eo
TECHSPEC
Lens Cleaning
Solution
Lens cleaning solution t
o clean the device’s
optical components
(Edmund Optics) N/A
No; available from Edmund
Optics, Fisher Scientific,
VWR, Staples.com,
Amazon.com
N/A
Fisherbrand Lens
Paper (4’’x
6’’)
Len paper t
o clean the device device’s optical
components
(Fisher Scientific) N/A
No; available from Fisher
Scientific,
VWR,
Staples.com, Amazon.com
N/A
CaviWipes
Disinfecting wipes to clean the device between
uses (Metrex)
N/A
No; available from Metrex,
K
-Dental, Amazon.com,
Fisher Scientific
Table 7: List of recommended MolecuLight i:Xaccessories and consumables
Warning
Do not connect the MolecuLight i:XImaging Device to a power supply using cables that have not
been approved by MolecuLight.
Note
Lens Cleaning Solution and Lens Paper may be used instead of the Lens Cleaning
Towelette.
Detailed information about cleaning and disinfecting the MolecuLight i:Xis provided in Section 12.
4MolecuLight i
i:
XImaging Device Overview
4.1 MolecuLight
i
:X Imaging Device Description and Use
The MolecuLight i:XImaging Device allows clinicians to quickly, safely and easily visualize potentially harmful bacterial
presence and distribution on skin and in wounds, in real-time at the point-of-care. The device is non-contact and no imaging
contrast agents are required for fluorescence imaging.
The MolecuLight i:XImaging Device is a handheld medical device comprised of a high-resolution color LCD display and
touch-sensitive screen with integrated optical and microelectronic components. MolecuLight i:Xuses its patented
technology to enable real-time standard light and fluorescence imaging of bacteria and tissue components in wounds and
surrounding healthy skin of patients.
The MolecuLight i:XImaging Device captures real-time images (JPEG format) and videos (MOV format) in Standard Imaging
ModeTM and Fluorescence Imaging ModeTM. These modes are identified as follows:

MolecuLight i:XTM User Manual Revision 1.1 Page 13 of 43
1. Standard Imaging ModeTM (or ST-Mode): Real-time imaging as per standard photography.
In Standard Imaging Mode, MolecuLight i:Xis able to capture the following images:
a. ST-imageTM: An image captured in Standard Imaging Mode
b. ST-videoTM: A video captured in Standard Imaging Mode
Note
Wound measurement is
only possible on an ST-Image (see Section 10).
2. Fluorescence Imaging ModeTM (or FL-Mode): Real-time imaging in Fluorescence Imaging Mode as a result of violet light
illumination. In Fluorescence Imaging Mode, MolecuLight i:Xis able to capture the following images:
a. FL-imageTM: An image captured in Fluorescence Imaging Mode
b. FL-videoTM: A video captured in Fluorescence Imaging Mode
The device provides image-guidance during visual assessment of wounds.
The device can capture an ST-imageTM, ST-videoTM, FL-imageTM, and FL-videoTM of wounds to quickly identify the presence
and extent of harmful bacteria. This helps to guide the clinical team towards a more comprehensive wound examination
through guided or ‘targeted’ debridement and sampling of potentially harmful bacteria. Images can be appended to a
patient’s clinical care record by exporting the images in JPEG format or printing them (see Section 9.9).
Figure 2: Clinician side, side views and Patient side of MolecuLight i:XImaging Device
1. Power Button: pressing button turns the device ON/OFF
2. Display Screen Power Button: pressing button turns Display Screen ON/OFF
3. System Status LED: indicates overall device performance
4. Battery Status LED: indicates device battery charge
5. Range Finder LED: indicates optimal distance from wound
6. Ambient Light Status LED: indicates optimal lighting environment for Fluorescence Imaging Mode
7. Heat Sink: dissipates heat during use
Warning
Heat Sink
may get warm after prolonged use
8. Display Screen: provides touch functionality and displays images in real-time
9. Home Button: pressing button turns Display Screen ON
10. Port for Charging Display Screen and Data Transfer: to be used with the white connecting cable on the MolecuLight
Power Cable, boot should be lifted for access
11. Rocker Switch for Standard Imaging Mode and Fluorescence Imaging Mode: allows the user to switch between
Standard Imaging Mode and Fluorescence Imaging Mode
12. Violet Wavelength LEDs: provide illumination when in Fluorescence Imaging Mode
MolecuLight i:XTM User Manual Revision 1.1 Page 14 of 43
13. Camera Sensor Window: protects the camera sensor which allows image and video capture
14. Range Finder Sensor: detects optimal distance from wound, and
Ambient Light Sensor: detects optimal lighting conditions for Fluorescence Imaging Mode
15. Illumination Zone: area includes camera sensor window, Violet Wavelength LEDs, Ambient Light Sensor and Range
Finder Sensor
Note
Keep the Illumination Zone clean and avoid covering the Violet Wavelength LEDs and camera sensor
window with your fingers
16. Label: lists certifications, references User Manual, indicates date of manufacture, contains MolecuLight Inc.’s address,
provides device model and serial number
17. Holding Contour: allows the user to grip the device securely
18. MolecuLight i:XCharging Port: allows device charging with black power supply cable on the MolecuLight Power Cable
19. Buttons for future functionality
MolecuLight’s training program includes the following to ensure user knowledge of how to operate the device:
xReading this User Manual in its entirety
xFollowing the Quick Start Guide (Appendix C) and the MolecuLight i:XImaging Device Overview (Appendix B)
xFollowing the MolecuLight DarkDrape & MolecuLight Adapter Instructions for Use (Appendix D) and the MolecuLight i:X
Wound Measurement Quick Start Guide (Appendix E)
Caution
The MolecuLight i:XImaging Device is restricted to use by trained health care professionals and
should be protected from unauthorized use.
4.2 How MolecuLight i
i:
XCreates Images in FL-Mode
The MolecuLight i:Xincludes two high-efficiency violet wavelength emitting LEDs that illuminate wounds and surrounding
skin for high-resolution and real-time fluorescence imaging of bacteria and tissues, without the need for contrast agents.
When wounds are illuminated by violet light, endogenous collagens in the connective tissue matrix emit a green colored
fluorescent signal. Some bacteria emit a red colored fluorescence signal due to the production of endogenous porphyrins,
and others emit a cyan colored fluorescence signal due to the production of endogenous pyoverdine. The MolecuLight i:X
Imaging Device simultaneously captures fluorescence from both bacteria and tissues and creates a composite image on the
high-resolution color LCD Display Screen. The user can easily and instantly visualize the presence and location of bacteria
within and around a wound.
Figure 3 illustrates how MolecuLight i:Xuses violet light illumination to create real-time fluorescence images.
Caution
Keep the Illumination Zone clean and avoid covering the Violet Wavelength LEDs and camera sensor
window with your fingers, as this may affect illumination and image quality (see Figure 2, item 15).
Note
The red color may appear orange to some users
.
Figure 3: MolecuLight i:XReal-time Imaging
Violet Wavelength LEDs homogeneously illuminate the wound, resulting in the generation of a red or cyan fluorescence
signal when bacteria are present due to the presence of endogenous porphyrins or pyoverdine.

MolecuLight i:XTM User Manual Revision 1.1 Page 15 of 43
4.3 MolecuLight DarkDrape and MolecuLight Adapter Device Description and Use
Used together, the MolecuLight DarkDrape and the MolecuLight Adapter provide a convenient and optimal environment for
imaging wounds when room lights cannot be turned off or windows cannot be blocked.
The MolecuLight DarkDrape provides a light-blocking shield that may be draped over a wound. The MolecuLight Adapter
provides an attachment that allows the MolecuLight i:X to connect to the MolecuLight DarkDrape. After the MolecuLight
DarkDrape is securely attached to the MolecuLight Adapter, the MolecuLight Adapter slides onto the MolecuLight i:X and
clicks into place.
Please see Section 12.4 for MolecuLight Adapter cleaning instructions.
Please see Appendix D for the MolecuLight DarkDrape and MolecuLight Adapter Instructions for Use.
Warning
Do not reuse the MolecuLight Dar
kDrape. It is for single use only.
Caution
Do not over
tighten the drawstring when securing the MolecuLight DarkDrape to the patient.
Caution
Avoid
touching the wound with the MolecuLight DarkDrape.
Caution
Be aware of compromised visibility of the imaging target when using the MolecuLight DarkDrape to
avoid injuring the patient.
Note
Ensure the MolecuLight DarkDrape does not obstruct the imaging field of view.
Note
The MolecuLight DarkDrape is designed for
fluorescence imaging on the lower leg or foot.
5Intended Use and Indication for Use
The MolecuLight i:Xinstantly visualizes the presence of potentially harmful levels of bacteria through endogenous
autofluorescence without the need for contrast agents or contact with the patient. The intended use of the device is to
assist health care professionals during the management of patients with wounds by enabling real-time visualization of
potentially harmful bacteria. The device is intended to be used as part of routine clinical wound assessment processes
which may include examination for characteristic signs and symptoms of infection. Under violet light illumination, the
MolecuLight i:Xcan capture and document images or videos of wounds and surrounding areas where fluorescent bacteria
may be present. The bacterial fluorescence signals detected by the device provide health care professionals with a visual
indication of bacterial presence, load, and location within and around wounds. This information can be used to guide health
care professionals in their selection, application, and response monitoring of wound therapies.
The indication for use of the device is to instantly visualize the presence of potentially harmful bacteria commonly found
within or around wounds including S. aureus, P. aeruginosa, E. coli, Coagulase-negative staphylococci, Enterococcus spp,
Proteus spp, Klebsiella pneumoniae, Beta-hemolytic streptococci (Group B) and Enterobacter spp during clinical assessment,
treatment, and monitoring of treatment response of wounds.
5.1 Off Label Use
The MolecuLight i:Xis intended only to be used with the iX Camera App for purposes of image or video
documentation of wounds. Using the MolecuLight i:Xfor other purposes, or using Apps other than the iX Camera App which
may be available on the MolecuLight i:Xdevice is considered off label use.
6Device Basics
The device consists of two primary elements:
1. MolecuLight i:XImaging Device – contains the electronics required to support its imaging functionality, including optical
filters, camera sensor, range finder and ambient light sensors and the violet light source. It also contains several LEDs to
indicate the status of various operational components of the device.
MolecuLight i:XTM User Manual Revision 1.1 Page 16 of 43
The MolecuLight i:XImaging Device contains a rechargeable battery which needs to be kept charged in order for
MolecuLight i:Xto work properly. Refer to Section 9.8 for details about charging the
MolecuLight i:X.
Primary features on the MolecuLight i:XImaging Device include:
xRocker Switch
xTwo Power Buttons
xFour Status indicator LEDs (see Figure 4)
xTwo Charging Ports
2. Display Screen – displays both standard light and fluorescence images on a touch-sensitive color LCD screen, which can
be controlled by the user touching the screen in order to control the imaging functionality of the device.
The Display Screen leverages iPod technology and has the following characteristics:
x4-inch (diagonal) widescreen display with Multi-Touch technology
x1136-by-640-pixel resolution at 326 pixel per inch
x800:1 contrast ratio (typical)
x500 cd/m2max brightness (typical)
xFingerprint-resistant oleophobic coating
The Display Screen contains a rechargeable battery, which needs to be kept charged in order for MolecuLight i:Xto function
properly. Refer to Section 9.8 for details about charging the Display Screen.
For instructions about using the device’s Display Screen functionality, please refer to Section 9.10.
Note
The Display Screen and the MolecuLight i:XImaging Device each contain separate rechargeable
batteries. Both batteries need to be charged separately in order for MolecuLight i:Xto work
properly. Refer to Section 9.8 for details about charging the MolecuLight i:X.
Figure 4: Schematic rendering of MolecuLight i:XImaging Device showing 1- System Status LED, 2- Battery Status LED 3- Range Finder
LED and 4- Ambient Light Status LED
6.1 Power Button
The MolecuLight i:XImaging Device has a Power Button for controlling the ON/OFF state of the electronics in the device.
The Power Button can toggle the MolecuLight i:Xto one of the following states:
xOFF: the MolecuLight i:Xis powered off
xON: the MolecuLight i:Xis powered on (Violet Wavelength LEDs turn on after Fluorescence Imaging Mode is
selected)
Note
The ON/OFF state of the MolecuLight i:Xis independent of the ON/OFF/SLEEP state of the Display
Screen. Refer to Section 9.10 for details about the ON/OFF state of the Display Screen.

MolecuLight i:XTM User Manual Revision 1.1 Page 17 of 43
The Power Button functions as follows:
Action
Effect
If the MolecuLight i:X is OFF, press and
hold the Power Button for 3 seconds
Switches MolecuLight
i:Xfrom OFF to ON.
If the MolecuLight i:X is ON, press and
hold the Power Button for 3 seconds
Switches MolecuLight
i:X from ON to OFF.
If the MolecuLight
i:X is ON during
Fluorescence Imaging Mode for greater
than
90 seconds
Switches Violet Wavelength LEDs from ON to OFF automatically (to
conserve battery).
Note You will need to move the Rocker Switch to Standard Imaging
Mode and then back to Fluorescence Imaging Mode to activate the
Violet
Wavelength LEDs.
If the MolecuLig
ht i:X is ON and you
don’t touch MolecuLight
i:X for 30
min
utes
Switches the MolecuLight i:Xfrom ON to OFF automatically after 30
min
utes (timing out to conserve battery).
Note If the Rocker Switch is in the Fluorescence Imaging Mode
position when powering the MolecuLight
i:XON, you need to move the
Rocker
Switch to Standard Imaging Mode and then back to Fluorescence
Imaging Mode to turn on the Violet Wavelength LEDs.
Reset
function (to be used when the
device or any of its functions are not
functioning as intended)
To reset the MolecuLight i:X, switch the device off, then hold the Power
B
utton for 8 seconds. The System Status LED will blink red initially, then
yellow
and then red again twice. This sequence indicates a reset was
performed.
Remove finger from button once this sequence is observed.
Press the Power Button again for 2-3 seconds to turn on the device.
Table 8: Power Button functions
6.1.1 Status Indicator LEDs
The MolecuLight i:Xis designed with four sensors to guide you in capturing optimal images/videos. Each sensor has a
corresponding indicator light (LED) on the MolecuLight i:X.
6.1.1.1 System Status LED
MolecuLight i:Xperiodically executes a self-diagnosis to ensure it is operating properly. The System Status LED
indicates the overall operational state of the MolecuLight i:XImaging Device.
Indicator
Meaning
Solid Green Light
MolecuLight i:XImaging Device is functioning as expected.
Solid Yellow Light
You can continue imaging, however follow up with MolecuLight as soon as possible.
Solid
Red Light
The device has malfunctioned. Stop using the device immediately and contact
MolecuLight to troubleshoot the problem.
Red Light, Yellow Light, and 2
Red Light flashes
The device is going through a reset (after the Power Button has been pressed).
Table 9: System Status LED
Note
When the device is OFF, no System Status LED color is visible. When you power on the device, the
System Status LED
blinks green 5 times for 2 seconds, staying solid green on the 5th blink. Blinking
indicates that the device is performing its self-diagnostic routine.
6.1.1.2 Range Finder LED
MolecuLight i:XTM User Manual Revision 1.1 Page 18 of 43
MolecuLight i:Xis equipped with a pulsed laser-based Range Finder sensor which is used to determine the distance
between MolecuLight i:Xand the wound. The Range Finder LED indicates to the user when the device is held at the correct
distance for optimal imaging.
LED Indicator
Meaning
Green Light
The optimal distance (8-12 cm) between MolecuLight i:Xand the wound being imaged has been
reached. See Figure 8.
Amber Light
The distance between MolecuLight i:Xand the wound is out of its optimal range (i.e. closer than 8 cm
or further than 12 cm) and/or the Range Finder sensor is unable to determine the device-to-wound
distance. The user can adjust the device-to-wound distance to obtain a Green Light for imaging.
Table 10: Range Finder LED
Note
The Range Finder LED turns green when the distance between Range Finder and the wound is within 8-12
cm.
The Range Finder LED should be used when in FL-Mode. It is also strongly recommended to be used when in ST-
Mode to obtain images with similar sizes (for presentation or comparison purposes).
The Range Finder LED should always be used when measuring wound area per Section 10.
Caution
Capturing a FL-image or FL-video outside the recommended device-to-wound distance may result in
suboptimal fluorescence and/or image quality.
Caution
A defect in the Range Finder’s function may result in distorted/inaccurate images/videos. See Section
15 Troubleshooting & Support, question 18 on how to verify the Range Finder function.
6.1.1.3 Ambient Light Status LED
FL-images and FL-videos need to be captured in a dark environment for optimal fluorescence image quality. The Ambient
Light Status LED indicates if the level of ambient light in the examination room is optimal for Fluorescence Imaging to aid
users in capturing optimal fluorescence images and videos.
Note
The Ambient Light Status LED is only ON in Standard Imaging Mode. It turns off in Fluorescence Imaging
Mode.
For use with Fluorescence Imaging Mode, view the Ambient Light Status LED immediately after turning off room lights and
prior to toggling the Rocker Switch to Fluorescence Imaging Mode.
Indicator
Meaning
Green Light
The ambient light level in the imaging environment is adequate for fluorescence imaging.
Amber Light
The ambient light level is too high and fluorescence images should not be captured.
The user should reduce the ambient light settings by turning off the room lights, blocking windows, turning
off computer monitors or other means until the Ambient Light Status LED turns green.
Note If you are unable to obtain optimal ambient light for Fluorescence Imaging Mode, MolecuLight
recommends using MolecuLight DarkDrape (used in conjunction with MolecuLight Adapter). For more
information, refer to Section 4.3 or contact MolecuLight.
Table 11: Ambient Light Status LED
Note
The Ambient Light Status LED is not active or ON when the Violet Wavelength LEDs are turned on
for imaging in Fluorescence Imaging Mode. In Fluorescence Imaging Mode, the Ambient Light
Status LED is OFF.
Caution
A defect in the Ambient Light Status LED may result in inaccurate images.
Caution
Ignoring the Ambient Light
Status LED indicator may result in inaccurate images.

MolecuLight i:XTM User Manual Revision 1.1 Page 19 of 43
6.1.1.4 Battery Status LED
The Battery Status LED indicates the MolecuLight i:XImaging Device battery charge level.
Note
This LED indicates the battery status for the MolecuLight i:XImaging Device only. The Display
Screen’s battery status and charging status operates independently of this functionality and is
displayed on the screen.
Indicator
Meaning
Solid Green Light
The MolecuLight i:Xbattery is fully charged or at high power (over 80% power).
Solid
Yellow Light
The MolecuLight i:Xbattery power is between 20-80% power and MolecuLight i:Xcharging is
advised.
Solid Red Light
The battery power has dropped below 20% power and the MolecuLight i:Xrequires immediate
charging.
Table 12: Battery Status LED
When the MolecuLight i:XImaging Device is OFF, the Battery Status LED is OFF, except when the battery is critically low and
not charging, at which point the LED blinks red every 2 seconds.
When the MolecuLight i:XImaging Device is charging, the Battery Status LED blinks to indicate charging status of the device.
6.1.2 Rocker Switch
The Rocker Switch enables the user to toggle between Standard Imaging Mode and Fluorescence Imaging Mode.
When the Rocker Switch is in its “down” position, MolecuLight i:Xis in Fluorescence Imaging Mode.
When the Rocker Switch is in its “up” position, MolecuLight i:Xis in Standard Imaging Mode.
To enable Standard Imaging Mode:
1.
Turn on the device (if it is not already on).
2.
Toggle the Rocker Switch to its “up” position
(Figure 5).
xMolecuLight i:Xis in Standard Imaging Mode.
To enable Fluorescence Imaging Mode:
1.
Toggle the Rocker Switch to its “down” position
(Figure 6).
x
MolecuLight i:Xis in Fluorescence Imaging Mode.
xViolet Wavelength LEDs turn on.
Figure 5:
Rocker Switch in Standard Imaging Mode (in ‘up’ position)
Figure 6:
Rocker Switch in Fluorescence Imaging Mode (in ‘down’ position)
7Quick Start Guide
Please see Appendix C for the MolecuLight i:XQuick Start Guide.
MolecuLight i:XTM User Manual Revision 1.1 Page 20 of 43
8Environmental Conditions That Affect Use
MolecuLight i:Xis intended to be used in a hospital/clinical setting.
Caution
MolecuLight i:Xis not suitable for use in the presence of flammable anesthetic mixtures with air,
oxygen or nitrous oxide.
8.1 Lighting
The level of ambient light in the imaging environment dramatically affects the quality of both the Standard Imaging Mode
and Fluorescence Imaging Mode images acquired with the MolecuLight i:XImaging Device. A green Ambient Light Status
LED indicates when the lighting conditions are optimal for imaging in Fluorescence Imaging Mode. During Standard Imaging
Mode the Ambient Light Status LED will be amber. Check the Ambient Light Status LED prior to toggling the Rocker Switch to
FL-Mode.
If you are unable to obtain optimal ambient light for Fluorescence Imaging Mode, MolecuLight recommends using
MolecuLight DarkDrape (used in conjunction with MolecuLight Adapter). For more information refer to Section 4.3.
Caution
Be aware of surroundings when using the MolecuLight i:Xin dark environments during
Fluorescence Imaging Mode. Failure may result in bodily injury due to tripping.
8.2 Operating Temperature
The device is designed to operate in environment temperatures between 10-30oC (50-86oF).
8.3 Storage
Store MolecuLight i:Xat room temperature in a secure location between uses.
Store the MolecuLight DarkDrape at room temperature in a dry place.
9Operating Instructions
9.1 Proper Imaging Technique
A standardized wound imaging protocol is critical to the accuracy and objectivity of digital wound evidence and image
capture. Images should be acquired while holding the device stable in the hands of the user and while the patient is not
moving (to avoid movement artifacts in the images). See Figure 7.
To capture images with MolecuLight i:X(in either Standard Imaging Mode or Fluorescence Imaging Mode):
xHold the device 8-12 cm away from the wound. The Range Finder LED is green when the distance between
MolecuLight i:Xand the wound is optimal.
xEnsure that the wound is centered in the camera’s field of view.
xHold MolecuLight i:Xsuch that the plane of the wound is parallel to the plane of MolecuLight i:X. See Figure 8.
Note
The optical glass windows (see Figure 7) must be clean for optimal Standard Imaging Mode and
Fluorescence Imaging Mode image capture
. Refer to Section 12 for information about cleaning
MolecuLight i:X.
Caution
Patient movement while the device is in close proximity to the patient may result in distorted images.
Caution
Angling the device incorrectly may result in poor image quality. The Illumination Zone on the device
should point directly at wound and be held at the recommended distance.
Failure
to do so may result
in poor image quality. Refer to Figure 8.

MolecuLight i:XTM User Manual Revision 1.1 Page 21 of 43
Figure 7: How to hold the device with one hand (left) and two hands (right)
Note
Hands are positioned away from the Violet Wavelength LEDs and are outside of the Illumination
Zone demarcated by the raised ridge.
Figure 8: Overview of optimal distance and angling of the device relative to the wound
9.2 Handling MolecuLight i
i:
X
MolecuLight i:Xdoes not require direct contact with the patient or wound during operation. The optimal distance for
imaging is 8-12 cm from the wound.
MolecuLight i:Xdoes not require the administration of any exogenous imaging contrast agents for imaging in Fluorescence
Imaging Mode.
MolecuLight i:Xcan be easily cleaned between uses. Instructions for cleaning and disinfection can be found in Section 12.
MolecuLight recommends the use of standard powder-free latex or nitrile gloves when using MolecuLight i:Xto reduce the
risk of bacterial contamination.
Caution
Residue from (some) medical gloves on optical components may result in distorted images.
MolecuLight i:XTM User Manual Revision 1.1 Page 22 of 43
9.3 Imaging with MolecuLight i
i:
X(iX Camera App)
The MolecuLight iX Camera App is a simple and intuitive application that allows you to quickly and easily capture images
and videos on your MolecuLight i:XImaging Device.
Warning
Only use the
iX Camera App to capture wound images and/or videos.
Figure 9: Display Screen
Caution
Do not “swipe up” the camera icon circled in red above. This is not the iX Camera App that is
intended for use.
The iX Camera App is easily accessible by following two simple steps:
1. Push the Home Button on the MolecuLight i:X.
The MolecuLight i:XDisplay Screen turns on.
If the screen is locked, use ‘slide to unlock’ at the bottom of the display to unlock the screen.
Note
The MolecuLight i:XImaging Device and the Display Screen must be powered on. Refer to Section
6.1 for details.
2. Press the iX Camera App icon.
The iX Camera App opens and is ready for use.
The iX Camera App allows you to:
xCapture an image
xCapture a 30 second video
xReview previously acquired images/videos via Image Library
xDelete images/videos
9.3.1 Capture an Image
The iX Camera App allows you to capture both a ST-image and FL-image by selecting the Camera icon.
9.3.1.1 Capture an ST-image
To capture an image in Standard Imaging Mode:
1. Move the Rocker Switch to the Standard Imaging Mode position.
2. Select the Camera icon. The Camera icon turns blue when selected.
3. When using the iX Camera App wound measurement functionality, refer to instructions provided in section 10
and Appendix E.
4. View the Range Finder LED.

MolecuLight i:XTM User Manual Revision 1.1 Page 23 of 43
The Range Finder LED advises when the optimal distance between (8-12 cm from the wound) the target and
MolecuLight i:Xis reached. If you wish to capture images in FL-Mode and ST-Mode with similar fields of view, this
feature will be critical to observe.
5. Aim the MolecuLight i:X Imaging Device towards the wound (with the screen facing the user) and center the
wound in the camera’s field of view.
A square briefly appears in the center of the field of view to indicate that the camera focus is being set. If the
autofocus is not satisfactory, tap on the screen on the target of interest to manually focus on the wound.
6. To take an image, tap anywhere within the blue rectangle indicating ‘CAPTURE IMAGE’.
You will hear a click sound and the screen will flash indicating that the ST-image was captured and saved on the
MolecuLight i:X.
Caution
Hold MolecuLight
i:Xsteady while imaging; shaking hands may result in distorted images.
9.3.1.2 Capture an FL-image
To capture an image in Fluorescence Imaging Mode:
1. Turn off the lights in the examination room. A dark environment is required to capture a good quality image.
Note
If lights cannot be turned off, use (optional) MolecuLight DarkDrape and MolecuLight Adapter. Refer to
Section 4.3 for details.
Caution
Do not use MolecuLight WoundStickers when capturing images in Fluorescence Imaging Mode.
2. View the Ambient Light Status LED.
3. The Ambient Light Status LED provides feedback as to whether or not the ambient light level is optimal. The
green color indicates that conditions are optimal to capture an image. If the LED is amber, try turning off
additional light sources (e.g. computer monitors) and covering windows until the LED turns green or use a
MolecuLight DarkDrape.
4. Move the Rocker Switch to the Fluorescence Imaging Mode position.
5. Select the Camera icon and ensure you are already within the iX Camera App.
If the Display Screen is off, press the Home Button to access the iX Camera App icon.
6. View the Range Finder LED.
The Range Finder LED advises when the optimal distance (8-12 cm from the wound) between the target and
MolecuLight i:Xis reached. The Range Finder LED is green when the distance is optimal.
7. Aim MolecuLight i:Xtowards the wound (with the screen facing the user). A square briefly appears in the center
of the field of view to indicate that the camera focus is being set. If the autofocus is not satisfactory, tap on the
screen on the target of interest to manually focus on the wound.
8. To take an image, tap anywhere within the blue rectangle indicating ‘CAPTURE IMAGE’.
You will hear a click sound and the screen will flash indicating that the FL-image was captured and saved on the
MolecuLight i:X.
Caution
Covering LEDs with your fingers will result in poor image quality. The Illumination Zone on MolecuLight
i:X(see Figure 7) represents the intended keep-out zone for the LEDs.
9.3.1.3 Capture a video
The iX Camera App allows you to capture both an ST-video and FL-video by selecting the Video icon.
9.3.1.4 Capture an ST-video
To capture a video in Standard Imaging Mode:
1. Move the Rocker Switch to the Standard Imaging Mode position.
MolecuLight i:XTM User Manual Revision 1.1 Page 24 of 43
2. Select the Video icon. The Video icon turns blue when selected.
3. View the Range Finder LED.
4. The Range Finder LED indicates when the optimal distance between (8-12 cm from the wound) the target and
MolecuLight i:Xis reached. If you wish to capture videos in FL-Mode and ST-Mode with similar fields of view,
this feature will be important to use.
Note
Monitor the Range Finder LED throughout the whole video recording session to maintain an appropriate
imaging distance.
5. Aim MolecuLight i:Xtowards the wound.
A square briefly appears in the center of the field of view to indicate that the camera focus is being set. If the
autofocus is not satisfactory, tap on the screen on the target of interest to manually focus on the wound.
6. To take a video, tap anywhere within the blue rectangle indicating ‘START RECORDING’.
The ST-video begins recording and a time counter will appear on the screen.
7. To stop the recording, tap the screen within the blue rectangle indicating ‘STOP RECORDING’. The time counter
stops.
The ST-video stops recording and is saved on the MolecuLight i:X.
Note
30 seconds is the maximum video length that can be acquired
at one time.
Note
Images and videos are saved in
the Image Library.
9.3.1.5 Capture an FL-video
To capture a video in Fluorescence Imaging Mode:
1. Turn off the lights in the examination room. A dark environment is required to capture a good quality image.
Note
If lights cannot be turned off, use (optional) MolecuLight DarkDrape and MolecuLight Adapter. Refer to
Section 4.3 for details.
Caution
Do not use MolecuLight WoundStickers when capturing images in Fluorescence Imaging Mode.
2. View the Ambient Light Status LED.
The Ambient Light Status LED provides feedback as to whether or not the ambient light level is optimal. If the
LED is amber, try turning off additional light sources (e.g. computer monitors) and covering windows until the
LED turns green, or use the MolecuLight DarkDrape.
3. Move the Rocker Switch to the Fluorescence Imaging Mode position (see Figure 6).
4. Select the Video icon. The Video icon is blue when selected.
5. View the Range Finder LED.
The Range Finder LED advises when the optimal distance (8-12 cm from the wound) between the target and
MolecuLight i:Xis reached. The Range Finder LED is green when the distance is optimal.
Note
Monitor the Range Finder LED throughout the whole video recording session to maintain an
appropriate imaging distance.
6. Aim MolecuLight i:Xtowards the wound. A square briefly appears in the center of the field of view to indicate
that the camera focus is being set. If the autofocus is not satisfactory, tap on the screen on the target of interest
to manually focus on the wound.
7. To take a video, tap anywhere within the blue rectangle indicating ‘START RECORDING’.
The FL-video begins recording and a time counter appears on the screen.

MolecuLight i:XTM User Manual Revision 1.1 Page 25 of 43
8. To stop the recording, tap the screen within the blue rectangle indicating ‘STOP RECORDING’ and the time
counter will stop.
The FL-video stops recording and is saved on the MolecuLight i:X.
Note
30 seconds is the maximum video length that can be acquired at one time.
Note
Images and videos are saved in
the Image Library.
9.4 Review Images and/or Videos Using the Image Library
The iX Camera App allows you to browse through the Image Library and view individual images/videos.
By default, the Image Library displays all images and videos that have been saved. The most recently saved image/video is
displayed at the bottom of the Image Library. When the Image Library icon is selected, all images/videos appear in
thumbnail format and are available for review.
Caution
There may be a minor delay once the Image Library icon is selected before the image thumbnails appear.
Swipe up or down to scroll through recent images/videos in the Image Library. Swiping down scrolls to older
image(s)/video(s) and swiping up scrolls to newer image(s)/video(s).
Select the Image Library icon from the Menu Bar to view previously acquired images
and videos.
1. Tap the thumbnail of an image/video of interest and the selected image/video is displayed.
A video is indicated by the overlaid play icon as depicted in Figure 11.
2. The date and time of acquisition is always saved with the image in the bottom left corner as depicted in Figure 10.
This is only applicable to saved images. Acquired videos are not saved with the date and time of acquisition.
Figure 10: Viewing a Saved Image
Figure 11: Viewing a Saved Video
9.5 Using the Album Feature (Optional)
All acquired images/videos can be browsed from the Image Library where images/videos are saved sequentially per
section 9.4.
The optional Album feature is for users who wish to save images and videos into particular folders. The Album feature
provides an easy and convenient way to save images and videos together by creating and designating an Active Album.
An Active Album simply means that newly acquired images and videos will be added to the Active Album selected.
To enable the Album feature, open the Settings Menu and tap the Album Feature switch:
When the Album feature is enabled, the default Active Album to view saved images/videos on the MolecuLight i:Xis the
Camera Roll as depicted in Figure 12.
MolecuLight i:XTM User Manual Revision 1.1 Page 26 of 43
The Album feature may be selected from the Settings Menu where previously created Albums may be browsed and
selected to become the “Active Album” at any time.
Figure 12: Enabling the Optional Album Feature
Note
The Active Album must be selected prior to taking images/videos. Images/videos cannot be moved
between Albums.
Warning
Due to iOS restrictions, when downloading images to a computer through explorer/finder, individual
albums will not appear and users may manually sort photos using the unique image ID text overlay
saved on images.
9.5.1 Creating a New Album
To create a new Album:
1. Enable the Album icon and select ‘+’
2. Select ‘Create’ when prompted with ‘Create New Album: iX-….?’
3. The new Album is designated as the Active Album where subsequent images and videos will be saved.
A description below the Album name provides information about the number of images and videos and the time and
date of the last acquired image within a particular album.
4. The iX Camera App automatically generates the Album name with prefix ‘iX-’ followed by the next 4-digit number in
the sequence. Note that Album names cannot be edited.
5. As depicted in Figure 13, a unique Image ID containing the Album name followed by “photo number” appears on the
bottom left side of the image along with the date and time of acquisition. This Image ID will appear at the bottom of
the saved photo and will automatically increment each time a new photo is taken.

MolecuLight i:XTM User Manual Revision 1.1 Page 27 of 43
Figure 13: Creating a New Album
9.5.2 Accessing and Using Previously Created Albums
To add new images to a previously created Album, enable the Album icon from the Settings Menu :
1. Select the desired Album from the Settings Menu drop down list (as depicted above).
The selected Album will be highlighted grey.
2. The selected Album becomes the Active Album and is labelled as the ‘Active Album’ across the Menu Bar:
Newly acquired images and videos will be saved to the Active Album.
When browsing the Image Library with an Active Album designated, only images and videos saved to the selected
Active Album are displayed in the Image Library. All acquired images and videos can always be found on the Camera Roll.
Note
Images and videos captured in a designated Active album cannot be
moved into another Album.
Note
At any time, the user may go back to the Settings icon to disable the Album feature. When disabled, all
acquired images and videos can be browsed from the Image Library .
Note
When the total number of created Albums reaches 9999, a new Album “iX-0001” shall be created even if
an old Album with the same name exists. This does not overwrite the previous Album.
Figure 14: Menu Bar
MolecuLight i:XTM User Manual Revision 1.1 Page 28 of 43
9.5.3 Capture Image and Video Capture Screens
9.5.3.1 Capture Screens when the Active Album is the Camera Roll
When the Album feature is enabled with the Camera Roll as the Active Album, the Capture Image and Video Capture
screens display the Active Album “Camera Roll” across the top of the image, and the date and time of acquisition
across the bottom left side of the image as depicted in Figure 15.
9.5.3.2 Capture Screens when the Active Album is Created
When created Albums are used as the Active Album instead of the Camera Roll, a unique Image ID correlating to the Album
name followed by “photo number” appears on the bottom left side of the image in addition to the date and time of
acquisition.
As depicted in Figure 16, this Image ID will appear at the bottom of the saved image and will automatically increment each
time a new image/video is taken. Although the Image ID and date and time of acquisition is displayed during Video Capture,
videos are not saved with an Image ID or date and time of acquisition.
9.5.4 Deleting Albums
With the Album feature enabled from the Settings Menu :
1. Delete an Album by selecting the ‘Edit’ function within the Settings Menu .
2. If user selects the ‘Edit’ button, a minus sign appears next to each Album:
To exit the ‘Edit’ function, tap the ‘Edit’ button again.
3. When the minus sign is selected, a ‘Delete’ button will appear:
4. If ‘Delete’ button is selected a confirmation message appears.
5. Once user confirms to delete the Album, they will be asked whether or not to also delete
the images within that Album.
If user chooses not to delete the images after the Album is deleted, the images will be
found in the Camera Roll.
Figure
15
: Capture Screens with
Camera Roll as Active Album
Figure
16
: Capture Screens with
Created Album as Active Album
Figure 17: Deleting Albums

MolecuLight i:XTM User Manual Revision 1.1 Page 29 of 43
9.6 Zooming in and out and Panning
9.6.1 Zooming in and out
To zoom in and out on a displayed image:
1. Select the image. The image will fill the full screen.
2. Place your thumb and pointer finger at opposing points of the image.
3. Keep both fingers on the screen and gently move your fingers away from each other.
This action “zooms in” on the part of the image between the two fingers, enlarging the portion of the image between
the two fingers.
Note
The speed at which your fingers spread influences the rate of zoom.
4. To zoom back out, reverse the motion and gently "pinch" your fingers towards each other while keeping contact with
the screen.
This action “zooms out” on the part of the image between the two fingers, reducing the display of the portion of the
image between the two fingers.
Note
It is not possible
to zoom in and out on video recordings.
9.6.2 Panning
To pan a displayed image:
On a zoomed in image, drag the screen with one finger to move around the image or hold your finger near the edge of the
screen to pan to that side. Move your finger closer to the edge to pan more quickly.
Note
It is only possible to pan a displayed ST-image or FL-image that has been zoomed. It is not possible to
pan a recorded video. It is not possible to pan during image capture.
9.7 Deleting Images and Videos
To delete an image or video:
1. Select the image or video on the Display Screen.
2. Tap the Trash icon.
You will be prompted with “Allow ‘iX Camera’ to delete this photo?” or “Allow ‘iX Camera’ to delete this video?”
3. Select either ‘Don’t Allow’ or ‘Delete’.
Selecting ‘Delete’ will delete the image or video.
Note
Images are not permanently deleted using this method. Images can still be recovered on the device
from the deleted images folder.
9.8 Charging MolecuLight
i:
X
The Display Screen and the MolecuLight i:XImaging Device each contain a rechargeable battery. Both batteries need to be
charged separately in order for MolecuLight i:Xto work properly. The cables for charging both batteries are provided as one
item (MolecuLight Power Cable) so that the MolecuLight i:XImaging Device and Display Screen may be charged
concurrently. Geography-specific MolecuLight Power Cable Adapters (supplied) may be required to connect the
MolecuLight Power Cable to the wall outlet.
MolecuLight i:XTM User Manual Revision 1.1 Page 30 of 43
9.8.1 Charge the MolecuLight i
i:
XImaging Device
To charge the MolecuLight i:XImaging Device:
Figure 18: Charging MolecuLight i:X Imaging Device and Display Screen with MolecuLight Power Cable
1. Connect the MolecuLight i:XImaging Device to the black power supply cable on the MolecuLight Power Cable (P/N
1142) as in Figure 18. The black power supply cable charges the MolecuLight i:XImaging Device.
The MolecuLight i:XImaging Device begins charging. The Battery Status LED indicates the battery power as described in
Section 6.1.1.4.
2. When the Battery Status LED is solid green and no longer blinking, disconnect the MolecuLight Power Cable from the
wall. Then remove the black power supply cable from the device and store the MolecuLight i:Xfor its next use.
Note
Connection to wall outlet should be easily and readily accessible and not blocked when MolecuLight i:X
requires charging.
Note
MolecuLight recommends charging the MolecuLight i:Xwhenever it is not in use so that it is fully charged
for its next use.
Note
It takes up to 5 hours to fully charge the MolecuLight i:Xonce it is fully depleted. It takes approximately 2
hours to charge to 80%.
9.8.2 Charge the MolecuLight
i:
XDisplay Screen
To charge the MolecuLight i:XDisplay Screen:
1. Connect the white connecting cable on the MolecuLight Power Cable (P/N 1142) to the port on the MolecuLight i:X
Display Screen as shown in Figure 18.
2. Connect the USB power adapter to the other end of the white connecting cable as shown in Figure 18.
3. Plug the USB power adapter into an electrical outlet.
4. When the battery status on the Display Screen indicates that it is fully charged, disconnect the MolecuLight Power Cable
from the electrical outlet and then remove the MolecuLight Power Cable from the device.
Note
MolecuLight recommends charging the MolecuLight i:XDisplay Screen whenever it is not in use so that it
is fully charged for its next use.
9.9 Uploading Images and Videos to Computer
You can upload images and videos to a computer using supported applications on your Mac or PC.
To upload images and videos to a computer:
1. Connect the white connecting cable on the MolecuLight Power Cable (P/N 1142) to the port on the MolecuLight i:X
Display Screen. Connect the USB connector end of the white connecting cable to your computer as shown in Figure 19.
When prompted to ‘allow’ or ‘trust’ the computer on the MolecuLight i:X Display Screen, you may do so.
Communication between MolecuLight i:Xand your computer will be established automatically.
Note
If the USB connector end of the white connecting cable is connected to the USB power adapter,
disconnect it.
2. Do not use iTunes to copy or move images from MolecuLight i:Xto your computer. If you have iTunes on your
computer, you will automatically be prompted to access the images/videos via iTunes.

MolecuLight i:XTM User Manual Revision 1.1 Page 31 of 43
3. Close the iTunes window, select Finder (Mac) or Windows Explorer (PC) and select the MolecuLight Device under
Portable Devices. You will be able to access your images within the MolecuLight Device folder.
Note
If you delete photos or videos during data
transfer, they are removed from the Image Library.
Note
Images are not permanently deleted using this method. Images can still be recovered on the device from
the deleted images folder.
Note
You will require an Apple ID (consult with your Institution’s IT department), for accessing future
MolecuLight Apps.
Figure 19: Connecting MolecuLight i:Xto a computer to transfer images/videos
9.10 Display Screen Functionality
Figure 20: MolecuLight i:XDisplay Screen
9.10.1 Basic Display Screen Functionality for MolecuLight i
i:
XUse
Most of the buttons required to use the Display Screen are located on the touchscreen user interface.
The Home Button is the only physical button that controls the Display Screen.
9.10.1.1 Home Button
Pressing the Home Button takes you to the Home Screen and provides access to the iX Camera App.
You are able to lock the Display Screen while not using the MolecuLight i:X. Locking the Display Screen saves battery life and
prevents the device from responding if you inadvertently touch the screen.
MolecuLight i:XTM User Manual Revision 1.1 Page 32 of 43
To lock the Display Screen:
1. Press the Display Screen Power Button. The MolecuLight i:Xscreen becomes dark and the device is locked.
Note
The Display Screen locks automatically if you don’t touch the screen for 5 minutes.
To unlock the Display Screen:
1. Press the Home Button.
2. Drag the ‘slide to unlock’ slider to the right. The MolecuLight i:Xunlocks.
To turn on the Display Screen:
1. If the Display Screen has been fully powered off, press and hold the Display Screen Power Button.
2. When the Apple logo appears, wait for the MolecuLight i:XDisplay Screen to power on.
3. When the ‘slide to unlock’ slider appears, drag the ‘slide to unlock’ slider to the right which unlocks the MolecuLight i:X,
and access the iX Camera App.
To turn off the Display Screen:
1. Press and hold the Display Screen Power Button.
2. When the ‘slide to power off’ slider appears, then drag the slider to the right.
3. The MolecuLight i:XDisplay Screen powers off.
9.10.2 Connect to a Wi-Fi Network
MolecuLight i:X contains a 2.4GHz Wi-Fi module. Wi-Fi is a technology that allows you to wirelessly connect to a local-area
network.
To connect to a Wi-Fi network:
1. From the MolecuLight i:XHome Screen, swipe right to the center page and tap Settings.
The Settings screen is displayed.
2. Tap Wi-Fi.
The Wi-Fi screen is displayed.
3. Slide the Wi-Fi ON/OFF switch to ON.
MolecuLight i:Xautomatically searches for available Wi-Fi networks.
4. Tap the name of your desired Wi-Fi network.
5. If the Wi-Fi network is password-protected, a prompt appears to enter a password.
6. Enter the password for the Wi-Fi network.
7. Tap Join.
When your device is connected to a Wi-Fi network, a checkmark appears to the left of the network name. The Wi-Fi
symbol also appears in the status bar at the top left of your display.
Note
If you leave proximity to a connected Wi-Fi network and lose the connection to it, MolecuLight i:X
reconnects when you return to the same location.
9.10.3 Connect to the Internet
MolecuLight i:Xconnects to the Internet by joining Wi-Fi networks. When joined to a Wi-Fi network that is connected to the
Internet, MolecuLight i:Xconnects to the Internet automatically.
9.10.4 Create a Passcode
Note
It is advised to follow your IT department protocol before setting up a
passcode.
Default factory settings for the MolecuLight i:XImaging Device do not require the use of a passcode.
To prevent unauthorized access to data acquired on the MolecuLight i:Xyou may want to create a passcode.
To create a passcode for MolecuLight i:X, go to Settings > Passcode.
Note
If a cybersecurity breach is suspected contact MolecuLight at info@moleculight.com or call +1-647-362-
4684, or 1-877-818-4360 (North America Only).

MolecuLight i:XTM User Manual Revision 1.1 Page 33 of 43
Figure 21: Settings Menu
10 Measuring Wound Area
The iX Camera App wound measurement functionality features wound border detection, wound area measurement, and
wound length & width measurement capabilities. The wound border may be traced automatically (Auto Mode) or manually
(Manual Mode). Wound area measurement should only be performed on images captured in Standard Imaging Mode with
the use of MolecuLight WoundStickers. Refer to Appendix E for additional details on the iX Camera App wound
measurement functionality.
Note
The wound measurement functionality should only be used on images captured in Standard Imaging
Mode and should not be used on images captured in Fluorescence imaging mode.
Note
The use of 2 MolecuLight WoundStickers may produce an average wound measurement error (Area
and Dimension)
of 5%, when the image is taken with MolecuLight WoundStickers in the plane of the
wound.
Note
In the absence of MolecuLight WoundStickers, no
area measurement is provided.
Caution
The use of only 1 MolecuLight WoundSticker may produce an average wound measurement error (Area
and Dimension) of 15%, when the image is taken with MolecuLight WoundStickers in the plane of the
wound.
Caution
Only MolecuLight WoundStickers should be used for the wound measurement functionality.
Caution
Do not place MolecuLight WoundStickers on the wound.
10.1 Settings Menu
The Settings Menu accessed from the Menu Bar enables customization of the wound border thickness, and display of
reported length and width dimensions. The Album feature may also be enabled through the Settings Menu as
described in Section 9.5.
10.1.1 Wound Border Thickness
The displayed wound border thickness can be modified from the Settings Menu
accessed from the Menu Bar.
By default, the medium border thickness is selected.
Use the slider to modify the border thickness to the desired setting.
The 3 available settings are
1) Thin
2) Medium
3) Thick
10.1.2 Display of Length & Width Dimensions
The display of the Length and Width dimensions can be disabled or enabled using the switch from the Settings Menu
accessed from the Menu Bar.
When enabled, a selection between “Length & Width” or “Vertical & Horizontal” can be made. By default, the “Length
& Width” setting is enabled where the maximum length and perpendicular width of the wound is displayed on screen
whenever a wound measurement is taken.
When display of dimensions is disabled using the switch from the Settings Menu , length/width or vertical/horizontal
dimensions will no longer be reported on the wound image when a wound border is drawn and saved.
MolecuLight i:XTM User Manual Revision 1.1 Page 34 of 43
10.1.2.1 Length & Width
By default, the Length & Width setting is enabled:
Once the desired wound border is drawn, Length & Width dimensions will appear as blue and
green lines on the wound and their measurements will be reported in cm.
Wound length (blue line, L) represents the longest dimension of the wound in any direction.
Wound width (green line, W) is drawn perpendicular to wound length.
As wounds change in size over time the orientation of the maximum length of the wound may
also shift and therefore the maximum length may not always be reported from the same part
of the wound using this length and width option.
10.1.2.2 Vertical & Horizontal
Wound dimensions are also available in vertical and horizontal format:
Vertical length (blue line, V) represents the longest vertical distance (top to bottom of the
screen) within the wound boundary. Horizontal length (green line, H) represents the longest
horizontal distance within the wound boundary.
Selecting this method of measuring wounds will allow you to consistently measure the same
axes of the wound for tracking changes in vertical and horizontal wound distances over time.
To measure the wound consistently, wound images must be taken with consistent orientation.
This method of measurement can be combined with the well documented clock method to
accurately detect changes in the vertical and horizontal distances of the wounds.
To take measurements according to the clock method, capture wound images with the
MolecuLight i:Xin a portrait orientation such that the top of the device is aligned with the head
of the patient and the bottom with the feet of the patient. The measured vertical distance will
be a measurement of the wound from 12-6 o’clock and the horizontal distance will be a
measurement of the wound from 9-3 o’clock.
10.2 Auto Mode
The iX Camera App can automatically trace and measure the area of a wound. The following steps are required for
automated wound area measurements:
1. Ensure the Standard Imaging Mode wound image is captured with 2 MolecuLight WoundStickers (P/N 1266) placed
close to and within the plane of the wound. Measurement accuracy will be degraded if the MolecuLight WoundStickers
are off plane, in which case stickers may appear elliptical. For optimal results, ensure the MolecuLight WoundSticker
placement is at the 1 and 7 o’clock position or at the 11 and 5 o’clock position when holding the MolecuLight i:Xin a
portrait (vertical) orientation, as depicted in Figure 22 and Figure 23.
2. When capturing an image, ensure the MolecuLight i:XCamera Sensor is parallel to the wound and the MolecuLight
WoundStickers are in field of view.
3. Open the image of interest from the Image Library . Select the Auto Trace icon.
4. The Auto Trace feature traces the border of the wound with a white line (the MolecuLight WoundStickers will be
highlighted with a blue circle during this auto detection process).
5. When selecting Auto Trace, additional icons are available to expand or contract the detected border. Pressing
these icons expands or contracts the wound border as desired.
6. Once the desired wound border is achieved, a Checkmark icon is available to ‘Accept and Save’ the wound border
and the wound area will appear in cm2.
At any point, the user can select the Auto Trace icon to exit Auto Mode.
Note
Once the user has pressed the Expand icon until it reaches the largest detection threshold range,
the Expand icon will disappear, leaving the Contract icon.
Figure 22: Length & Width
Figure 23: Vertical & Horizontal

MolecuLight i:XTM User Manual Revision 1.1 Page 35 of 43
Note
Once the user has pressed the Contract icon until it reaches the smallest detection threshold
range, the Contract icon will disappear, leaving the Expand icon.
Note
T
he wound area measurement is not displayed unless the user selects the Checkmark icon.
10.3 Manual Mode
If Auto Mode is not tracing an accurate wound boundary, the iX Camera App also allows for manual tracing of the wound
border to measure wound area. The following steps are required for manual wound area measurements:
1. Ensure the Standard Imaging Mode wound image is captured with 2 MolecuLight WoundStickers (P/N 1266) placed
close to and within the plane of the wound. Measurement accuracy will be degraded if the MolecuLight WoundStickers
are off plane, in which case stickers may appear elliptical. For optimal results, ensure the MolecuLight WoundSticker
placement is at the 1 and 7 o’clock position or at the 11 and 5 o’clock position when holding the MolecuLight i:Xin a
portrait (vertical) orientation, as depicted in Figure 22 and Figure 23.
2. When capturing an image, ensure the MolecuLight i:XCamera Sensor is parallel to the wound and the MolecuLight
WoundStickers are in field of view.
3. Open the image of interest from the Image Library . Select the Manual Trace icon.
4. Use your index finger or a stylus to draw the border of the wound. Maintain contact with the Display Screen until
closing the wound border trace.
5. Select the Checkmark icon if the trace is acceptable.
6. If the trace is not acceptable, press the Manual Trace icon once to exit Manual Mode and press it again to reenter
Manual Mode.
7. Once the desired wound border is drawn, a Checkmark icon is available to ‘Accept and Save’ the wound border and
the wound area will appear in cm2.
Note
T
he wound area measurement is not displayed unless the user selects the Checkmark icon.
10.3 Saving a Wound Measurement
When a wound border is traced (in either Auto or Manual Mode), a Checkmark icon is available and when selected, a
pop-up message will appear prompting the user to ‘Accept and Save’ the image. The user should select ‘OK’ when the
wound border trace accurately represents the border of the wound. Once ‘OK’ is selected, the image with the wound
border, reported area measurement, and any length/width or vertical/horizontal dimensions (when enabled) will be saved
as a new image to the same Album as the original image.
11 Interpretation of Fluorescence Images
MolecuLight i:Xallows real-time visualization of the presence and distribution of bacteria in wounds (for more details, see
Sections 4.2 and 5). When the device is used in FL-Mode as described in Step 5 of the Quick Start Guide in Appendix C,
fluorescence images of a wound may be obtained.
The fluorescence images obtained with the MolecuLight i:XImaging Device have distinct fluorescence colors which are
associated with bacteria and other different wound components. Users may see the following characteristic fluorescence
colors when imaging wounds:
Color
Indicator
Red*
Typically associated with the presence of potentially harmful levels of bacteria
Green
Typically associated with connective tissues in skin
Dark/black areas
Typically associated with blood, highly vascularized tissues, necrotic tissue, highly pigmented tissue
(freckles, moles, etc.) or poorly illuminated areas
White
Typically associated with materials in the field of view that are white (bed sheet, clothing, paper,
wound measuring ruler, etc.) or oversaturation of color signal
Cyan
Typically associated with the presence of Pseudomonas aeruginosa
Table 13: Color indicators for interpretation of fluorescence images
MolecuLight i:XTM User Manual Revision 1.1 Page 36 of 43
Note*
Red color of bacteria may appear orange to some users.
Figure 24 provides examples of the Standard Imaging Mode (ST-Image) and corresponding Fluorescence Imaging Mode (FL-
Image) from typical diabetic foot ulcers (DFUs) showing the characteristic red and green fluorescence colors typically seen
during use of the MolecuLight i:XImaging Device for assessing a DFU.
11.1 Color Blindness
Correct interpretation of the images acquired with the MolecuLight i:XImaging Device requires that the user can
differentiate between green and red colors. A typical test for red-green color blindness involves looking at color swatches
(see the examples in Figure 25). Individuals who do not suffer from red-green color blindness will be able to see the number
56 in the left-hand color swatch shown below, and the number 74 in the right-hand color swatch shown below.
Figure 25: Color figure swatches used in a typical clinical color-blindness test
Caution
MolecuLight
i:Xshould not be used by individuals confirmed to be color blind.
Figure 24: Examples of MolecuLight i:Ximages in Standard
Imaging Mode and Fluorescence Imaging Mode. Red color
indicates the presence of bacteria (shown by arrows).

MolecuLight i:XTM User Manual Revision 1.1 Page 37 of 43
12 Cleaning and Disinfecting MolecuLight i
i:
X
Clean and disinfect MolecuLight i:Xbefore and after it is used and between wound assessment procedures. Failure to do so
may result in cross-contamination, and patient or user exposure to bacterial contamination. MolecuLight recommends
using CaviWipes as disinfecting wipes for cleaning and disinfecting (See Table 7 for ordering information).
The MolecuLight DarkDrape is for single use only and must not be used more than once. The MolecuLight Adapter can be
reused after being appropriately cleaned and disinfected.
Warning
Do not reuse
the MolecuLight DarkDrape. It is a single use only accessory.
12.1 Pre-Clean the MolecuLight
i:
X
Wear impervious gloves such as latex gloves when handling disinfecting wipes. Use a disinfecting wipe (CaviWipe) to
remove all debris from the MolecuLight i:Xby gently wiping ALL SIDES and surfaces (see Figure 2), including the Display
Screen and optical components identified within the red perimeters in Figure 26.
12.2 Disinfect the MolecuLight
i:
X
Use another disinfecting wipe (CaviWipe) to thoroughly wet ALL SIDES and surfaces (see Figure 2) of the MolecuLight i:X,
including the Display Screen and optical components identified within the red perimeters in Figure 26. Let the MolecuLight
i:X remain wet for 3 minutes at room temperature. Allow the MolecuLight i:X to dry.
12.3 Clean the MolecuLight
i:
X
Use a Lens Cleaning Towelette to thoroughly clean the Display Screen and optical components identified within the red
perimeters in Figure 26. There should be no visible debris, fingerprints or material on the Display Screen or any of the
optical components after cleaning. Allow the MolecuLight i:X to dry.
The MolecuLight i:X is now clean, disinfected and ready for use.
Figure 26: Image showing optical components (red) on Clinician Side and Patient Side of MolecuLight i:X
On the Clinician Side, red perimeter indicates location of the Display Screen.
On the Patient Side, red perimeter includes: Optical window for Violet Wavelength LED, Camera Sensor Window, Range Finder &
Ambient Light Sensor (listed from top to bottom as depicted)
Warning
Do not soak or immerse MolecuLight
i:Xin water.
Note
Prior to cleaning optical
components, ensure surfaces are disinfected as described in Section 12.2.
Caution
Improper cleaning/disinfection may
result in distorted images.
Note
The user should dispose of the cleaning and disinfection materials as well as MolecuLight DarkDrape (if
used)
according to the biohazard waste control processes at their Institution.
MolecuLight i:XTM User Manual Revision 1.1 Page 38 of 43
12.4 Pre-Clean the MolecuLight Adapter
Use a disinfecting wipe (CaviWipe) to remove all debris from the MolecuLight Adapter by gently wiping all sides and
surfaces, including tabs.
12.5 Disinfect the MolecuLight Adapter
Use another disinfecting (CaviWipe) wipe to thoroughly wet ALL SIDES, surfaces and tabs of the MolecuLight Adapter. Let
the MolecuLight Adapter remain wet for 3 minutes at room temperature. Allow the MolecuLight Adapter to dry. The
MolecuLight Adapter is now clean, disinfected and ready to use.
13 Maintenance of MolecuLight i
i:
X
MolecuLight i:Xperiodically executes a self-diagnosis to ensure it is operating properly, without requiring user input. The
System Status LED indicates the overall operational state of the MolecuLight i:XImaging Device.
If the System Status LED turns to amber or red (green is the indication that the device is functioning as intended), you may
Otherwise, except for regular cleaning (Section 12), there are no maintenance or calibration activities that need to be
performed.
For more information and details on service pricing, contact MolecuLight at support@moleculight.com.
14 Disposal of MolecuLight
i:
X
Comply with your country’s legal requirements for the disposal of electrical and electronic waste to dispose of the
MolecuLight i:X. The two rechargeable lithium ion batteries that power the MolecuLight i:Xmust be recycled appropriately
by following your country’s disposal regulations. For more information about recycling MolecuLight products, please
contact your local recycling program or MolecuLight at info@moleculight.com.
15 Troubleshooting and Support
15.1 Frequently Asked Questions
About who to contact:
1. Where do I call/email with questions?
xContact MolecuLight at info@moleculight.com or call +1-647-362-4684 or 1-877-818-4360 (North America
Only).
2. How can I purchase a device?
xContact MolecuLight at info@moleculight.com or call +1-647-362-4684 or 1-877-818-4360 (North America
Only).
About imaging bacteria:
3. How does the device detect bacteria?
xThe device illuminates wounds with a narrow band of violet light that causes bacteria commonly found in and
around wounds to fluoresce based on the presence of endogenous molecules called fluorophores (e.g.
porphyrins, pyoverdine, etc.). Refer to Section 4.2 for additional information.
4. Which type of bacteria can be detected?
xEvidence from clinical studies indicates the MolecuLight i:Xcan reliably detect the following bacteria species
that are relevant for clinical wound care: Staphylocccus aureus, MRSA, Pseudomonas aeruginosa, Escherichia
coli, Coagulase-negative staphylococci, Enterococcus spp, Proteus spp, Klebsiella pneumonia, Beta-hemolytic
streptococci (Group B) and Enterobacter spp.
About acquiring images:
5. How long does it take to get an image?

MolecuLight i:XTM User Manual Revision 1.1 Page 39 of 43
xImages appear immediately. ST-images and FL-images are captured instantly from a wound and displayed in
real-time on the device’s Display Screen and when the ‘CAPTURE IMAGE’ button is pressed.
6. How can I access the images/videos?
xImages/Videos are located in the Camera Roll and Albums. Refer to Section 9.4 for additional information.
xYou can transfer these images/videos from your device to any computer using the white connecting cable on
the MolecuLight Power Cable (P/N 1142). Refer to Section 9.9 for instructions.
About operating the MolecuLight i:X:
7. Does the device require any consumables to obtain FL-Images?
xConsumables are not required for operation of the device in optimal lighting conditions.
xIf you are unable to obtain optimal ambient light for Fluorescence Imaging Mode, MolecuLight recommends
using MolecuLight DarkDrape (used in conjunction with MolecuLight Adapter). For more information, refer to
Section 4.3, or contact MolecuLight at info@moleculight.com or +1-647-362-4684 or 1-877-818-4360 (North
America Only).
8. Who is qualified to use the device?
xTrained health care professionals in wound care.
9. How do I clean the device?
xThe MolecuLight i:Xcan be wiped clean using disinfecting wipes. Ensure excess fluid from the wipes does not
enter any openings and that the device is thoroughly dry before use. A lens towelette to clean the optical
components and Display Screen should also be used. For further information, please see Sections 3 and 12.
10. How do I take care of the device?
xMolecuLight i:Xis an optical imaging device and must be handled with care at all times.
i. Do not drop, knock or shake the device. Ensure there is no dirt, moisture or scratches over the camera
sensor window, LEDs and Display Screen.
ii. Do not immerse the device in liquid.
iii. Place the MolecuLight i:Xwith its screen facing up on a clean, dry, flat surface when not in use.
11. Is the device safe with laser light exposure and violet light illumination?
xYes, the device is classified in Risk Group 1 and the laser is Class 1 meaning there is not sufficient energy
produced to damage skin or eyes in normal use. It is advised to not point the device towards the eyes.
12. How do I turn on the device?
xThere are 2 dark gray power buttons to turn the device on. Press and hold each for 3 seconds to turn on the
device and the Display Screen respectively.
13. Why is the Display Screen off?
xThe Display Screen is turned on via the Display Screen Power Button. Push this button to turn on the Display
Screen. If it does not turn on, the Display Screen battery may need to be charged via the charging port located
below the Home Button using the white connecting cable on the MolecuLight Power Cable (P/N 1142).
xThe Display Screen could also be locked. Press and release the Display Screen Power Button or the Home Button
to reveal the screen.
14. Why can’t I see red fluorescence on the screen?
xThere may not be any red fluorescence signal generated by the target you are imaging.
xEnsure that the System Status LED is green, which indicates that the device optics and electronics are
functioning properly. For further information, contact support@moleculight.com.
xThe Rocker Switch may be set to imaging in Standard Imaging Mode.
15. What is the System Status LED?
xThis LED informs you that the device electronics and optics are functioning as designed. This LED should be
green. If the color changes to amber or red, contact MolecuLight as soon as possible. Refer to Section 6.1.1.1.
16. What does the Battery Status LED indicate?
MolecuLight i:XTM User Manual Revision 1.1 Page 40 of 43
xThis LED indicates the battery power of the device (for the optics and electronics) and is green when the charge
is >80%. Note that there is also a battery status for the Display Screen itself in the top right corner of the screen
that is independent of the device.
17. What does the Range Finder LED indicate?
xThe Range Finder notifies you when the optimal distance between the device and the wound is reached, i.e. 8-
12 cm. This is critical to follow in FL-Mode and it is also advised to observe the Range Finder LED when in ST-
Mode so that the images obtained will have a similar field of view.
18. How do I check if the Range Finder is working properly?
xYou can use a ruler at various distances from the device to verify the Range Finder LED Status. At a 5 cm
distance, the LED should be amber, at 10 cm, the LED should turn green and at 15 cm the LED should turn
amber again. If the results are incorrect, do not use the device and contact MolecuLight at
support@moleculight.com or call +1-647-362-4684 or 1-877-818-4360 (North America Only).
19. What does the Ambient Light Status LED indicate?
xFor optimal image quality in FL-Mode, it is important that the lighting conditions are in a dark environment. The
Ambient Light Status LED will notify you when conditions are optimal (the LED will be green).
20. How do I check if the Ambient Light Status LED is working?
xThe Ambient Light Status LED should be amber in a bright environment and turn green in a dark environment.
Make sure all lighting is off in order for the LED to turn green. If you have any questions, contact MolecuLight at
support@moleculight.com or call +1-647-362-4684 or 1-877-818-4360 (North America Only).
21. The Ambient Light Status LED turns off when the device is in FL-Mode. Is this intended?
xYes, this is the normal operation. The ambient light sensor which is used to measure the ambient light
conditions is also sensitive to the light emitted from the device during FL-Mode operation, therefore the
indicator LED turns off in this mode to ensure it does not provide a false reading. The lighting conditions should
be adjusted such that the Ambient Light Status LED is green prior to using FL-Mode.
xIf you are unable to obtain optimal ambient light for FL-Mode, MolecuLight recommends using MolecuLight
DarkDrape (used in conjunction with MolecuLight Adapter). For more information, refer to Section 4.3, or
22. How do I know that my Device is working properly?
xThe System Status LED indicates the functionality of the electronics and optics of the device. This LED is green
when the device is turned on and operating properly. If it is amber or red, contact MolecuLight at
support@moleculight.com.
23. How long does it take to charge the device?
xThe device charges to approximately 80% of its total capacity after 2 hours of charging and will be fully charged
after approximately 5 hours.
About operating the wound measurement app:
24. Why do I need to use the MolecuLight WoundStickers?
xIn the absence of MolecuLight WoundStickers, no area measurement is provided. The use of 2 MolecuLight
WoundSticker may produce an average wound measurement (area and/or dimension) error of 5%, when the
image is taken with MolecuLight WoundStickers in the plane of the wound. The use of only 1 MolecuLight
WoundSticker may produce an average wound measurement (area and/or dimension) error of 15%.
Table of contents
Popular Medical Equipment manuals by other brands

Acare
Acare AH-TX Series instruction manual

sensiplast
sensiplast 300418 Instructions for use

mectron
mectron Piezosurgery touch Cleaning and sterilization manual

Datex-Ohmeda
Datex-Ohmeda Aespire 7900 User's reference manual

Orliman
Orliman NEURO-CONEX 94303D INSTRUCTIONS FOR USE AND PRESERVATION

Dräger
Dräger Babytherm 8004 Instructions for use