Mopec 0811-000-000 User manual

ENGLISH (EN)
0810MPC000-LBL-P2 Rev A
https://www.mopec.com

EN
0810MPC000-LBL-P2 Rev A
Introduction
This instructions for use manual contains information intended
to ensure the safe, effective, and compliant use of your product.
This manual is intended for pathologists, biomedical equipment
technicians, and housekeeping staff. Keep and consult this reference
manual during the life of the product.
The following conventions are used in this manual:
▪
A
WARNING
highlights a safety-related issue. ALWAYS comply
with this information to prevent healthcare staff injury.
▪
A
CAUTION
highlights a product reliability issue. ALWAYS comply
with this information to prevent product damage.
▪
A NOTE supplements and/or clarifies procedural information.
For additional information, including safety information, in-service
training, or current literature, contact your Mopec sales representative
.
Description
The 115V Autopsy Saw uses high-speed oscillating action to quickly
cut through bone without harming soft tissue. Because exposure to
infected tissue is reduced, the saw does not require a blade guard.
The saw may be used for removing the skull cap, making linear cuts,
or taking small bone specimens.
Safety Directives
WARNINGS:
▪ Before using this equipment, or any component compatible with
this equipment, read and understand the instructions for use. Pay
particular attention to safety information. Become familiar with the
equipment before use.
▪ Only healthcare professionals trained and experienced in the use
of this medical device should operate this equipment.
▪ Upon initial receipt and before each use, operate the equipment
and inspect each component for damage. DO NOT use any
equipment if damage is apparent.
▪ DO NOT modify any system component or accessory, including
the power cord and power cord plug.
▪ ALWAYS position the equipment so that the power cord plug may
be easily disconnected from the power supply.
▪
DO NOT use this equipment in areas in which flammable
anesthetics or flammable agents are mixed with air, oxygen or
nitrous oxide.
▪
ALWAYS follow the current local recommendations and/or
regulations governing environmental protection and the risks
associated with recycling or disposing of the equipment at the
end of its useful life.
Definitions
The symbols located on the equipment and/or labeling are defined
in this section or in the Symbol Definition Chart. See the Symbol
Definition Chart supplied with the equipment.
SYMBOL
DEFINITION
This symbol is intended to alert service personnel
of the presence of voltage that may cause injury or
fatal electric shock. Disconnect the device from the
electrical supply before servicing.
On
Off
Accessories
WARNING:
Use only CMM-approved system components
and accessories, unless otherwise specified.
NOTE:
For a complete list of accessories, contact your Mopec
sales representative.
2 https://www.mopec.com/

0810MPC000-LBL-P2 Rev A
EN
Instructions
Cleaning
WARNINGS:
▪ ALWAYS disconnect the power cord plug from the facility power
outlet before installing a blade. Failure to comply may cause fire,
electric shock, or injury.
▪ Before operating the equipment, ALWAYS make sure the blade
is securely installed and the blade retaining nut is tight. An
improperly tightened nut may come loose and cause the blade to
fracture.
NOTE:
The wide portion of the adjustable section blade can be used
for removing the skull cap or making long linear cuts. The narrow
portion of the adjustable section blade can be used to remove a
small bone specimen.
1.
Insert the blade and arbor to its full depth in the triangular shaft
hole.
2.
Use the wrench attached to the power cord to tighten the set
screw firmly against the arbor.
CAUTION:
ALWAYS place the power cord where it will not be stepped
on, tripped over, or otherwise subjected to damage or stress.
3.
Connect the power cord plug to a facility power outlet.
4.
Hold the instrument close to the saw blade with one hand, with
the index finger extending under the oscillating head. In this
position, the motor serves as a counterbalance.
5.
Flip the toggle switch to turn on the saw.
WARNING:
DO NOT apply excessive pressure, such as
bending or prying, with the blade. Excessive pressure may
cause the blade to fracture.
NOTES:
▪ Cutting action is achieved by high speed oscillation of the blade.
The blade cuts on both the backward and forward strokes of
the excursion. To engage more blade teeth and allow for more
efficient cutting, the blade should be moved slightly up and down
as the cut progresses.
▪
Soft tissue moves with the blade and is not cut by the blade;
therefore, no blade guard is necessary to protect the operator.
6.
Push the blade into the bone.
7.
Withdraw and advance the blade in the direction of the cut.
8.
Repeat this motion of alternating pressure along the direction of
the cut until the cut is complete.
9.
Upon completion, flip the toggle switch to turn off the saw.
CAUTION:
To reduce the risk of damage to the power cord, ALWAYS
grasp the power cord plug, not the power cord, when disconnecting
the saw.
10.
Disconnect the power cord plug from the facility power outlet.
WARNINGS:
▪ ALWAYS disconnect the power cord plug from the facility power
outlet before cleaning the equipment. Failure to comply may
cause fire, electric shock, or injury.
▪ Upon initial receipt and before each use, clean the equipment as
indicated.
▪
DO NOT sterilize the equipment.
▪ DO NOT immerse or soak the equipment in liquid. DO NOT allow
moisture or liquid to soak into electrical plugs, receptacles, or
openings. Moisture or liquid inside the equipment may create an
unsafe electrical condition, cause corrosion, and/or damage the
electrical and/or mechanical components.
▪ Before each use, make sure the equipment is completely dry. DO
NOT use any equipment if moisture is present.
▪ DO NOT use solvents, lubricants, or other chemicals, unless
otherwise specified.
1.
Disconnect the power cord plug from the facility power outlet.
2.
Wipe the external surfaces of the equipment with a soft, lint-free
cloth moistened with a non-abrasive, mild detergent and water.
3.
Thoroughly dry the equipment with a soft, lint-free, absorbent
cloth.
Inspection and Maintenance
WARNINGS:
▪ Only individuals trained and experienced in the maintenance of
reusable medical devices should inspect, test, and maintain this
equipment.
▪
ALWAYS disconnect the power cord plug from the facility
power outlet before performing maintenance procedures on the
equipment. Failure to comply may cause fire, electric shock, or
injury.
NOTES:
▪ The Autopsy Saw has a lifetime lubrication system. No additional
lubrication is required.
▪ For service, contact your Mopec sales representative
.
https://www.mopec.com/ 3
This manual suits for next models
1
Other Mopec Saw manuals
Popular Saw manuals by other brands

Makita
Makita 5820 Parts Breakdown

King Industrial
King Industrial KC-10GC Service manual

MK Diamond Products
MK Diamond Products TX-3 Operations & parts manual

Westfalia
Westfalia ASSG18 Original instructions

Panasonic
Panasonic EY3551 - 18V WOOD SAW operating instructions

P.Lindberg
P.Lindberg 65651 Original manual