INSTRUCTIONS FOR YOUR SAFETY
Please read carefully and retain for future reference. It is a requirement
of IEC/EN 60601-1 that these instructions are made available to the
operator and qualied maintenance staff at all times.
The Mopec 5000 System has been examined
and found to be in compliance with the
standard for Medical Electrical Equipment IEC/
EN60601-1
www.mopec.com
(800) 362-8491
/05&OWFSUJHIUFOJOHPGthe Captive Blade
Clamp XJMMDBVTFthe CMBEFUPOPUTFBUQSPQFSMZ
BOEthe TBXnot toPQFSBUFQSPQFSMZ

Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
CONTENTS
'PSFXPSE
&RVJQNFOU$MBTTJGJDBUJPO
4BGFUZBOE8BSOJOHT
(FOFSBM%FTDSJQUJPO
4FRVFODFPG0QFSBUJPO
.PQFD4BX6OJU
.PQFD1PXFS4VQQMZ1PSUBCMF
.PQFD7BDVVN6OJU
'JUUJOH"DDFTTPSJFTBOE%FUBDIBCMF1BSUT
1SFWFOUJWF.BJOUFOBODF"OE*OTQFDUJPO
.PQFD'VTF3FQMBDFNFOU
.PQFD1PXFS4VQQMZ
5SPVCMFTIPPUJOH
$VUUJOH)JOUT
%JTJOGFDUJPOBOE4UFSJMJ[BUJPO
"DDFTTPSJFTBOE%FUBDIBCMF1BSUT
4DIFNBUJD8JSJOH%JBHSBN
'JMUFS4FSWJDF3FDPSE
$POEJUJPOTPG8BSSBOUZ
&MFDUSJDBM5FTU$FSUJGJDBUF
&RVJQNFOU3BUJOHT

'03&803%
8IFOOPUJOVTFPSEVSJOHNBJOUFOBODFBMXBZTJTPMBUFUIFFRVJQNFOUGSPNUIFFMFDUSJDBM
TVQQMZ
13&1"3"5*0/'0364&
6OQBDLBOEDIFDLFRVJQNFOUBHBJOTUUIFEFMJWFSZOPUF
$IFDLUIFEFUBJMTPOUIFFRVJQNFOUSBUJOHMBCFMDPOGPSNUPZPVSFMFDUSJDBMTVQQMZ
'PSBTTFNCMZJOTUSVDUJPOTQMFBTFSFGFSUPQBHF
*.1035"/55)*4&26*1.&/5.645#&(306/%&%
5IFXJSFTJOUIFNBJOTMFBEPGUIJTFRVJQNFOUBSFDPMPSFEJOBDDPSEBODFXJUIUIF
GPMMPXJOHDPEF
(306/%
/&653"-
-*7&
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
WW W
(SFFO
8IJUF
#MBDL
(SFFO:FMMPX
#MVF
#SPXO
Type B Equipment - Protection against electrical shock
Saw Unit Watertight Equipment IP67
Power Supply Splash Proof Equipment IP65
EQUIPMENT CLASSIFICATION
Mode of Operation:-
Continuous with Intermittent Loading up to 3 mins. Duty Cycle 50%
&OWJSPONFOUBMDPOEJUJPOTGPSUSBOTQPSUBOETUPSBHF
5FNQFSBUVSF )VNJEJUZ "UNPTQIFSJD
1SFTTVSF
'
'
I1BI1B

3
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
5IJTFRVJQNFOUNVTUCFHSPVOEFECFGPSFVTF
5SBJOFEQFSTPOOFMNVTUPOMZVTFUIJTQSPEVDU
1MFBTFSFBEBMMUIFJOTUSVDUJPOTCFGPSFVTF
$IFDLBMMFRVJQNFOUCFGPSFVTF
6TFUIFBQQSPQSJBUFTBGFUZFRVJQNFOU
%POPUJNNFSTFUIF1PXFS4VQQMZBOE7BDVVNVOJU
JO-JRVJE
6TFHPPERVBMJUZEJTJOGFDUBOUTBTBQQSPWFECZUIF)PTQJUBM
*OGFDUJPO$POUSPM
%POPUVTFDPSSPTJWFDMFBOJOHNBUFSJBMTPS)ZQPDIMPSJUFBT
UIFTFXJMMEBNBHFUIFVOJUT
%POPUMFBWFUIF)BOE1JFDFTPBLJOHJODMFBOFSGPSQSPMPOHFE
QFSJPETPGUJNFBTXJMMDBVTFBOJOHSFTTPGGMVJE
4"'&5:"/%8"3/*/(4
/PUF
OWFSUJHIUFOJOHof the Captive Blade Clamp
XJMMDBVTFthe CMBEFUPOPUTFBUQSPQFSMZBOE
the TBXnot to PQFSBUFQSPQFSMZ

GENERAL DESCRIPTION
The MPQFD 5000 Oscillating Saw has been developed to help minimize the risk of
cross infection in sensitive areas, providing it is used in accordance with the
manufacturer’s instructions.
The complete system comprises three basic units.
THE SAW UNIT
This unit is fed from the power supply. It is hermetically sealed allowing disinfection and
appropriate sterilization without dismantling. The enclosed motor cannot become
contaminated , consequently there is no hazardous aerosol effect. All external materials
used in its construction are noncorrosive.
THE FILTRATION UNIT0QUJPOBM
This unit is fed from the power supply. It gives highly ecient three stage ltration by way
of a paper lter, cloth lter and highly ecient cartridge lter (Hepa lter) respectively. It is
quite capable of ltering dry and partially damp material but must never be used to extract
liquids. This unit is not watertight and not designed for liquid extraction.
THE POWER SUPPLY
This unit is fed from the QPXFS supply and in turn delivers the appropriate electrical cur-
rents to the saw unit and the vacuum filtration unit. The complete system is activated by
the momentary switch on the saw unit. An intelligent current limiting device (which resets
itself automatically) is incorporated in the power supply to prevent overload usage of the
saw unit. This unit is watertight.
User preference may require the equipment to be supplied in its basic form which
comprises the saw unit and power supply only. A ltration unit may be easily added by the
customer at a later stage.
The complete MPQFD 5000 system is capable of handling health endangering dusts,
making it suitable for use in hospitals and industrial environments where the collection
and disposal of such debris is subject to stringent control.
This equipment is not suitable for use in an environment where an explosion risk is present.
The equipment should only be moved by using the handle provided and never by
pulling on the saw unit orQPXFSDBCMF
SEQUENCE OF OPERATION
Warning: This equipment is designed for use by properly trained personnel only.
MPQFD accepts no liability whatsoever for inappSPQSJBUFPSVOUSBJOFEVTBHF
1Plug equipment into appropriate QPXFS sPVSDF. Check.BJOT0O indicatoronpower
supply is lit.
28JUITBXGJSNMZJOIBOE depress and release the on/off button.
The complete system will then start.
3Depress and release the on/obutton once again and the system will then stop.
Always disconnect the equipment from theQPXFSPVUMFU when not in use.
4
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
5IF.PQFDJTEFTJHOFEGPSIJHIMZFGGJDJFOUDVUUJOHFYDFTTJWFQSFTTVSF
XJMMTMPXDVUUJOHTQFFEBOENBZDBVTFUIFTZTUFNUPTIVUEPXOEVFUP
PWFSMPBEPSCMPXGVTF

Captive Blade Clamp
Square Drive
(ANDPIECE
On/Off (I/O)
Momentary Button
Watertight 3TRAIN2ELIEF
Saw Unit Suppy Cable
M01&$ 5000 SAW UNIT
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289

Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
Status Indicator
Filtration Unit Output
energized when lit.
Captive Retaining Screws
Enclosure Cover
Status Indicator
Saw Unit Output
energized when lit.
4UBUVT*OEJDBUPS
1PXFS4VQQMZ
FOFSHJ[FEXIFOMJU
Status Indicator
Saw Unit overload detected
when flashing. Complete
system will shut down.Then
Auto-reset to“System off”
status in approx.20 seconds.
-AINS)NPUT,EAD
7ATERPROOF#ONNECTOR
3CREW,OCKING
.ON)NTERCHANGEABLE
(AND0IECE,EAD
7ATERPROOF#ONNECTOR
3CREW,OCKING
.ON)NTERCHANGEABLE
.01&$"VUPQTZ4BX1PXFS4VQQMZ
1PSUBCMF
0OWER3UPPLY
(ANDLE
"LANKING0LUG
3CREW,OCKING
!LLOWSFOR6ACUUM5NITTO
BEADDEDINFUTUREOPTIONAL

'JMUSBUJPO6OJU$MBNQ3JOH
Saw Unit
ON/OFF Button
Full System
Supply Lead
Saw Unit
Vacuum
Hose
Hose/Filtration
Unit Connector
Filtration Unit
Top Cover Hose/Cowling
Connector
Toggle
Clamp
Vacuum
Cowelling
Filtration
Unit
Filtration Unit
Chassis
Supply Lead
Filtration Unit
Handle
Cable Clip
PoweR3UPPLY%NCLOSURE
Status Indicator
lights
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
.01&$"VUPQTZ4BX7BDVVN6OJU0QUJPOBM
1PXFS4VQQMZ*OQVU-FBE

FITTING OF ACCESSORIES AND DETACHABLE PARTS
BLADE REPLACEMENT
Untighten the Captive Blade Clamp with Allen wrench provided and remove.
Replace or index blade as required, ensuring it is seated properly on the square drive
shaft. If not discard, or resulting damage or inefficiency of the saw could result.
3 'JSNMZSFUJHIUFOthe Captive Blade Clamp with Allen wrench QSPWJEFE
/PUF%POPU overtighten, as this will cause the blade to seat improperly and
prevent the saw from operating properly.
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
PREVENTIVE MAINTENANCE AND INSPECTION
THE EQUIPMENT SHOULD BE DISCONNECTED FROM THE ELECTRICAL SUPPLY
WHILE THE FOLLOWING TASKS ARE BEING PERFORMED.
THE SAW UNIT
Daily
Clean thoroughly with recommended disinfectant using soft nylon brush, not wire or hard
bristle type. Rinse down and dry thoroughly.
Weekly
Lubricate retaining screw, blade clamp and square drive with light engineering or silicone
based grease. Graphite based grease is not recommended. Inspect equipment for signs of
damage.
Annually
Full service of Saw Unit is strongly recommended. Note! *GUIFTBXVOJUJTTUJMMXJUIJOJUhT
POFZFBSXBSSBOUZQFSJPEit can only be serviced or repaired by .PQFD. or their
appointed agent. 4FSWJDFQFSGPSNFECZBOZPUIFSQBSUZ will invalidate thF
warranty.
THE POWER SUPPLY
Weekly
Clean thoroughly with recommended disinfectant using soft damp cloth. Dry thoroughly
and inspect for damage and ill fitting connections

ISOLATE FROM ELECTRICAL SUPPLY BEFORE REMOVING COVER
5IFUXPQPXFSTVQQMZGVTFTBSFMPDBUFEPOUIFVOEFSTJEFPGUIFCPBSEJO
UIFQPTJUJPOTIPXO
The third fuse is an anti surge fuse. Fuse type and ratings are marked adjacent
to the fuse holders or alternatively see page 14.
Undo the four captive retaining screws to remove the cover.
Prior to replacement ensure that the rubber seal
set into the underside cover has not
been disturbed.
FUSE REPLACEMENT
Retighten cover fully ensuring that
the cover is seated properly.
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
Status Indicator
Filtration Unit Output
energized when lit.
Captive Retaining Screws
Enclosure Cover
Status Indicator
Saw Unit Output
energized when lit.
4UBUVT*OEJDBUPS
1PXFS4VQQMZ
FOFSHJ[FEXIFOMJU
Status Indicator
Saw Unit overload detected
when flashing. Complete
system will shut down.Then
Auto-reset to“System off”
status in approx.20 seconds.
Waterproof Electrical
Connectors
Screw Locking-Non
Interchangeable
Waterproof Electrical
Connector
Screw Locking-Non
Interchangeable
.01&$"VUPQTZ4BX1PXFS4VQQMZ

TROUBLESHOOTING
The information contained in this section covers common problems and suggests solutions.
If your system’s problem is not covered here, or you still need help, please contact the
agent/manufacturer.
Note. Disconnect equipment from the QPXFSTVQQMZ electrical supply before
attempting to implement any solutions of an electrical nature
Possible Cause
System does not start.
Check main plug fuse.
Check power supply fuses.
Check all electrical connectors are
mated correctly.
Bad connection.
SolutionSymptom
1PXFS supply indicator
not lit.
Fuse blown.
Fuses blow repeatedly. Incorrect supply voltage. $IFDLFRVJQNFOUSBUFEWPMUBHFJT
DPNQUBCJMFXJUIUIFFMFDUSJDBMTVQQMZ
Disconnect ltration unit connector
from the power supply. If fuses do not
blow when saw unit is switched on
then the ltration unit is defective.
Filtration unit has been used
to extract liquids.
System shuts down
repeatedly.
Overload indicator is on.
Distorted blade.
Blunt blade.
Incorrect tted blade.
DiscardCMBEF immediately.
Index blade so new portion comes in
contact with the workplace.
Check blade is fitted in accordance
with the manufacturer’s instructions.
See page 8.
See cutting hints
System shuts down.
Overload indicator is on. Wait until yellow overload indicator is
o(approx 20 seconds) and retry.
Check connections for the saw unit
and ltration unit are engaged
correctly with the power supply.
Faulty connection.
Excessive pressure on
workpiece.

TROUBLESHOOTING- continued
Possible Cause
Saw unit status
indicator lit but not
running although
ltration unit does.
Check the power supply connection
is properly engaged.
Contact manufacturer.
Defective motor.
SolutionSymptom
Faulty connection.
Excessive noise from
saw unit.
Distorted blade.
Badly tting or loose blade.
RFQMBDFblade.
Check blade is fitted in accordance
with the manufacturers instructions.
Contact manufacturer.
Bearing service required.
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289

CUTTING HINTS
The MPQFD 5000 oscillating saw has been designed and constructed to provide effortless
and highly efficient cutting in its appropriate areas of usage. We suggest a trial cutting
period applying the least pressure possible to the workpiece until fully accustomed with
the efficiency. Excessive pressure is not required as this will reduce the speed of cut.
If you experience diculty then consult the Troubleshooting Causes and Solutions on pages
10 and 11.
Suitable safety protection must be worn at all times during cutting.
DISINFECTION AND STERILIZATION
The saw unit only is capable of withstanding temperatures between - degrees 'to
+ degrees '. This must be born in mind when considering a disinfection or
sterilization method.
Autoclaving of this equipment is not recommended.
Only use cleaning agents that are safe for use on metals. We do not recommend the use
of HyQPDIMPSJUF or corrosive cleaners as possible corrosion could result.
We advise gentle scrubbing with a soft non metallic brush, rinsing with clean water and
drying where necessary. Do not direct high pressure water jets at any part of the equipment,
this could invalidate the warranty.
All cleaning agents should be mixed, stored and applied according to the manufacturers
instructions.
8FTVHHFTUBOPODBVTUJDEJTJOGFDUBOUTVDIBT#&4BOJ1BUI%JTJOGFDUBOU'PBNJOH
4QSBZPS#&4BOJ1BUI%JTJOGFDUJOH4QSBZ$MFBOFSGPSDMFBOJOHQVSQPTFT8F
SFDPNNFOEZPV/0564&BCMFBDITPMVUJPOUPDMFBOZPVSVOJU
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289

ACCESSORIES AND DETACHABLE PARTS
#%
#%
#%
#%
#%
#%
#%
#%
#%
#%
#%
#%
#%
#%
#%
Order Code Description
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
#%
#%
#%
#%
774BX1PXFS4VQQMZ
771PXFS4VQQMZ
)BOE1JFDF
#POF%VTU$PMMFDUPS7
4BXBOE1PXFS4VQQMZ7#POF
Bone %VTU$PMMFDUPS240V
Dust Collector Filter Service Kit
Segmented#MBEF
Circular#MBEF
2.5"Segmented#MBEF
3" Circular Blade
Nut Driver
Depth Gauge 64mm
Depth Gauge 76mm
Vacuum Hoods (10Pack)
Vacuum Pipe with QR Fittings
1BQFSGJMUFSTQBDLPG
$MPUIGJMUFS
.JDSPOGJMUFS
#%5Captive Blade Clamp Kit Inc Allen Wrench

.01&$ 5000 SCHEMATIC WIRING
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
START/STOP
MOTOR
PRINTED
CIRCUIT
SAW
MOTOR
SAW UNIT
Connections
shown at
opposite end of
power supply
enclosure for
clarity
A6
A5
A4
A3
I/O I/O
M + M -
SAW
PL5
1
2
3
4
5
EARTH
1
2
3
4
5
6
LIVE!
LIVE!
L Fused
N Fused
L
N
Mains IN
Pump Power OUT
PL1
PL2
2
1
2
1
F2
F1
Varistor
250V
220/230
S1
110/115
-V
VR1
RL1
F3
PL4
Transformer secondaries
EARTH
1
2
3
4
5A QB D1
KBU4D
Fusible Resistor
D4
S1PB-E3/84A
F4
22R
IN OUT
GND
Pump Control
Saw Control
5V Regulator
U1
Mains Supply (Green)
+C1
4700uF
63V
40V Vacuum (Green)
+C3
22uF
35V
0V
1 3
2
5V 0V
2
2
C5
100uF
25V
C4
100uF
25V
0V
Saw 40V
0V
C6
100uF
25V
1
1
OVERLOAD (Yellow)
R11
10k
C2
100uF
100V
R12
10k
R1
10k
R4
3k9
R5
100R
R6
0R1
R10
1k
R9
1k
0.5W
5V
21
SAW
PL3
Transformer Primaries
Thermal Fuse Out
P1*
P1
Thermal Fuse In
P2*
P2
R8
10k
On/Off
Q3
BCX56-16
RL2
D8
S1PB-E3/84A
R4
3k9
0.5W
D5
m
m
D2
m
m
2
10k
R7
S1PB-E3/84A
D7
m
m
D3
D9
20V
Q2
BCX56-16
D6
m
m1
2
3
4
8
PIC1 2F675
Vcc
GP5
GP4
GP3
Vss
AN0
GP1
GP2
IC1
6
5
7
!
!
! !
!
!Safety Critical Parts

FILTER SERVICE RECORD
This page is provided to document your filter service history to meet the requirements
of the C.O.S.H.H. legislation. Although the complete filter service kit includes Vacuum
Cowellings and Paper Filters, provision for these are not included in the service record as
they are regarded more consistent with the day to day servicing of the equipment.
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
1
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
2
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
3
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
4
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
5
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
6
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
7
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
8
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
9
Date of Service
Signed
Vaccum Pipe replaced
Cloth Filter replaced
Micron Filter replaced
10

CONDITIONS OF WARRANTY
Should the equipment covered by this warranty become inoperative because of defective
material(s) or workmanship, .PQFDXJMM undertake at their own discretion to repair or
replace it at no charge to the purchaser, subject to the following conditions :-
1That the equipment has been used with care and shows no sign of abuse, neglect or
accidental damage.
2 That the equipment has only been used for the purpose(s) for which it was designed.
3 That the preventive maintenance has been performed in accordance with the
manufacturer’s instructions.
4That no attempt to repair or dismantle any part of the equipment has been made, except
by a MPQFD representative or an appointed agent of MPQFD.
5That any claim for repair or replacement be made within one calendar month of the
problem arising.
6The conditions of this warranty remain in force for a period of twelve months from the
date of purchase.
7This warranty does not aect your statutory rights.
The policy of MPQFD is one of continuous product development, and ensuring our
customers the best possible service. To achieve this, specifications and details may be
subject to change without prior notice.
Equipment owners outside the United 4UBUFT should consult the distributor/agent from
whom they purchased the equipment regarding spare parts and repair facilities.
The distributor/agent will advise on the best course of action to take.
CERTIFICATE OF ELECTRICAL TEST
The following tests have been performed with 65 portable appliance tester and
apply only to the combination of equipment whose serial numbers are identified as
being supplied against this warranty.
Earth bond test
Insulation test
-FBLBHF
7QBTTMFTTUIBOPIN
7EDHSFBUFSUIBONPIN
7SNTMFTTUIBON"
EQUIPMENT RATINGS
Equipment Serial No. Supply Voltage Auxiliary Socket
Output
5000 Power Supply V-50/60HZ 75VA V-50/60HZ 750 W
5000 Saw Unit 40V Not Applicable
5000 Filtration Unit V-50/60HZ 750 W Not Applicable
Tested by Dated
Purchased by Dated
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289

21750 Coolidge Hwy.
Oak Park, Michigan 48237
USA
T +1 248-291-2040
F +1 248-291-2050
W www.mopec.com
2002/289
Medezine Limited
Churchill Way Chapeltown
Sheffield England
S35 2PY
T
F+44(0)114 257 6222
E sales@medezinesaws.com
W www.medezinesaws.com
+44 (0)114 257 6002
2002/289
TECHNICAL INFORMATION
Power Supply
Transformer manufactured to IEC – EN60742
Output – Saw 40 VDC
Enclosure –to IEC529, DIN40050
Protected to IP65
Protection- Electronic Current Limiting
Device Double Pole Fused
Constructed –to IEC60601-1
Weight 1.7kg
Input voltage – 230Vac or 110 Vac
Cable 4.0m tted
Hand Piece
Motor –DC Permanent magnet
Speed –no load 13200 cycles per min
Current – no load 40Vdc 0.75amp
Typical load 1.5 amps
Construction – Anodised Aluminium body with
5 bearing sealed mechanism
Protected to – IP67
Audio Output – 37 dB (A) no blade
Vibration Value – 4.12 m/s2
Weight – 1.3kg
Technical Requirements
93/42/EEC
IEC/EN 60601-1
EMC 2004/108/EC
3FW;
Table of contents
Other Mopec Saw manuals
Popular Saw manuals by other brands

MSW
MSW C-SAW210N user manual

Zipper Mowers
Zipper Mowers ZI-FKS250 Operation manual

Jet
Jet HBS-812G Operating instructions and parts manual

Makita
Makita DPB181RME instruction manual

Delta
Delta 426-05-651-0001 instruction manual

MK Diamond Products
MK Diamond Products MK-101 Owner's manual & operating instructions