MOTUS GI PURE-VU SYSTEM INSTRUCTIONS FOR USE
1.0 WARNINGS AND PRECAUTIONS
Please read this manual and follow its instructions carefully.
Warnings:
• This device shall only be used by order of a licensed physician as governed by the FDA.
• The Pure-Vu System shall only be used according to the instructions and operating conditions
described in this manual. Failure to do so could impact the performance and safety profile of the
device, damage the device or the endoscope.
• The Pure-Vu System shall only be used under the supervision of a licensed physician. Do not use Pure
Vu system for any purpose other than the intended use.
• The Pure-Vu System shall be used only by trained medical personnel.
• The Pure-Vu Disposable Kits are provided clean and intended for single patient use. Do not sterilize,
clean or re-use the disposables. Discard all of the disposable device sections properly after use.
• Confirm that the other devices used with the Pure-Vu System function properly to their intended use
and that their operation or safety is not compromised and is compatible with the Pure-Vu System as
noted in these instructions. If any component of the endoscopic system is not functioning properly, the
procedure should not be performed.
• The Pure-Vu System should only be used with APPROVED medical devices, complying with IEC
60601-1.
• The Pure-Vu System shall be connected only to a supply main with protective earthing.
• Check and confirm that all cords or cables are connected correctly and securely.
• Continually monitor the Pure-Vu System and the patient for any signs of irregularity.
• The Pure-Vu System is compatible with EMC regulations, but may cause radio interference or may
disrupt the operation of nearby equipment.
• If the system does not function properly, turn the Workstation o, abort the procedure, and follow the
instructions noted in section “Operating the Pure-Vu Workstation”.
• The Pure-Vu System shall never be stored in areas where the unit could get wet or be exposed to high
temperature, humidity, direct sunlight, dust, or salt, which could adversely aect the equipment.
• The Pure-Vu System shall never be installed, used or transported in an inclined position nor should it
be subjected to impact.
• The Pure-Vu Workstation shall be maintained in a clean condition during storage and be ready for
subsequent use.
• The Pure-Vu disposables , as well as the Workstation (WS), shall be used only with the Pure-Vu System
recommended equipment and accessories.
• Avoid using the cleansing mode on the same area of tissue for an extended period of time.
• New, single use waste containers shall be used for each patient (Hospital approved canisters or Motus
GI Waste Bag).
• A New, single use irrigation bag (IB) or Irrigation bottle (water or saline) shall be used for each patient.
Precautions:
• Advance cautiously when performing retroflexion, entering the pylorus, duodenum, or ileocecal valve.
• Care shall be taken to avoid pinching the tubing during advancement through the GI tract and/or
when applying torque to the endoscope.
• Only Motus GI personnel or a trained Motus GI designee are allowed to service the Pure-Vu System.