
9) Cleaning. After use, discard the Whatmann paper sheet, remove the sealing mask
and the porous plate and rinse them in running water. They can be re-used.
Draining and maintenance
Pull the apparatus forward until its front foot is no longer resting on the bench. In this
position, the reservoir is tipped forward. Place a vessel under the brainsick and unscrew
its cap. Recover the transfer solutions and rinse the inside of the reservoir using a wash
bottle. Rinse regularly to avoid crystallization of saline solution inside the buffer
collection tank. If crystals form, they are difficult to remove later.
The reservoir must be regularly drained and rinsed. About 3 liters of the the buffer
solution can be collected in the trough before the vacuum pump could be affected.
Advice to Users
Conditions usually recommended for a transfer according to Southern :
•Agarose : 0.8 to 1%
•Gel size : 20 cm x 20 cm
•Gel thickness : 4 to 5 mm
•Size of fragment to transfer : up to 50 kb
•Vacuum : 40 to 50 mbar
•Time required for transfer : 45 to 60 minutes
If the vacuum is too great, the gel may be compressed, thus impeding the transfer. We
recommend that the vacuum should not exceed 55-60 mbar. If the vacuum is too low,
the transfer time may be unnecessarily extended. In extreme cases, only partial transfer
may be achieved. We advise that the vacuum should not drop below 40 mbar.
In Case of Problems
If no vacuum can be created, please check the following sources of problems:
•The sealing mask is incorrectly positioned.
•If the plastic sheet is creased above the seal or torn even slightly, air leaks are
likely to occur. Remove the frame and mask and check its positioning, particularly
at the corners of the reservoir.
•Bad positioning of the seal in its channel.
•Insufficient overlapping of the sealing mask and gel.
•The brainsick is not tight. Fasten its cap slightly.
If any other problems occur, please contact Technical Support: