MRC CLRM-200 User manual

PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
MRC.VER.01-10.13
O
Op
pe
er
ra
at
ti
io
on
n
M
Ma
an
nu
ua
al
l
C
CL
LR
RM
M-
-2
20
00
0
High-Quality Portable Colorimeter

1
Operating Instruction of Universal Test Components
Application: used to test liquid reagent, paste (such as ketchup and coating), powder
(such as coffee), etc.
I. Component Structure
The structure of universal test components is shown in Figure 1.
①Base; ②Baffle; ③Square Integrator; (31) Measuring Port of the Square Integrator
④Cuvette; ⑤Cover Plate; ⑥Push Plate; ⑦Screw
Figure 1 The Structure of Universal Test Components
II. Usage
The usage of universal test components is shown in Figure 2.

2
i. Place &/50310 colorimeter according to Figure 2. The measuring port of &/50310 is
butted up against that of square integrator, as shown in Figure 3. Then, use ⑥push plate
to hold &/50310 colorimeter and tighten ⑦the screw. Check whether &/50310 colorimeter
is loose or not. If it is loose, please re-tighten the screw.
ii. Put the ④Cuvette into the ③Square Integrator. There are two matte surfaces in the
cuvette. Please only touch these two matte surfaces when hold the cuvette. In addition,
there are two transparent surfaces which engraved with “MAX”and “MIN”. Please put
one surface against the measuring port of the square integrator, as shown in Figure 3.
iii. Put ④the Cuvette into ③the Square Integrator. Cover the ⑤cover plate. There is
an arrow icon in the cover plate. Meanwhile, square integrator also has an arrow icon.
Please make sure the location of these two arrows is consistent.
After finishing the above steps, you can start the measurement.

3
Figure 2 The Matching Between &/50310 Colorimeter and Universal Test Components
Figure 3 The Local Matching Between &/50310 Colorimeter and Universal Test Components
III. Notes
i. Keep ③the square integrator clean. After finishing the measurement, please cover
the ⑤cover plate and keep the square integrator closed.
ii. Protect those two transparent surfaces of the cuvette from scratching and soiling.
iii. The amount of the object in cuvette must be between “MAX”and “MIN”.

Contents
Product Description ....................................................................................................................1
Cautions .........................................................................................................................................1
I. Button Description ...................................................................................................................2
II. Interface Description ..............................................................................................................3
III. Battery Description and Installation..................................................................................3
IV. 200 Colorimeter Operating Instruction .......................................................................6
(I)Turning Power On ................................................................................................6
1. Preparations Before Power On ............................................................................6
2. Turning Power On ..................................................................................................6
3. White Calibration and Black Calibration..............................................................6
(II)Measurement.........................................................................................................7
1. Locating and Measuring Method..........................................................................7
2. Standard Measurement.........................................................................................7
3.Sample Measurement .............................................................................................8
(III)Save Data................................................................................................................9
1. Auto Save................................................................................................................9
2.Manual Save...........................................................................................................10
(IV)Connect 200 Colorimeter to the PC.........................................................10
(Ⅴ)print............................................................................................................................11
V. System Function Description.............................................................................................12
1. Record and Standard Entering...........................................................................12
2. Calibration..............................................................................................................13
3. Tolerance Setting..................................................................................................14
4. USB Communication............................................................................................15
5. Delete Records.....................................................................................................16
6. Language Selection .............................................................................................17
7. Display Mode ........................................................................................................18
8. Average Measurement ........................................................................................18
9. Function Setting....................................................................................................19
Ⅵ. Product Parameters ............................................................................................................25
1. Product Features..................................................................................................25
2. Product Specifications.........................................................................................26
Appendix......................................................................................................................................27
1. The Color of Objects............................................................................................27
2. Human Eyes to Distinguish Colors....................................................................28

1
Product Description
200 colorimeter is researched and developed in accordance with CIE (International
Commission on Illumination) and CNS (China National Standards). 200
colorimeter is a high precise colorimeter with simple user interface and stable
performance. It can be powered by both Li-ion battery and external DC power
supply.
White and black calibration is not required for 200 colorimeter which greatly simplifies
the measuring steps.
200 colorimeter adopts illumination locating which brings out more precise and quicker
locating.
200 adopts complex advanced algorithms which make measurement performance
more stable, more accurate and more compatible.
Cautions
● This colorimeter is a precise measuring instrument. Please avoid dramatic changes of
external environment when measuring. These changes, including the flicker of
surrounding light, the rapid change of temperature, will affect the measuring accuracy.
● Keep the instrument balanceable; make sure the measuring aperture cling to the test
sample, and no shaking or shifting when measuring. Please prevent the colorimeter
from fierce collision or crash.
● This instrument is not waterproof. Do not use it in high humidity environment or in water.
● Keep the colorimeter clean. Avoid dust, powder or solid particles entering the measuring
aperture and the instrument.
● Replace the white calibration cover and put the colorimeter into instrument cabinet when
not in use.
● Please take out the battery to prevent the colorimeter from damage if you don’t use it for
a long time.
● Please keep the colorimeter in a cool dry place.
● Any unauthorized changes to the colorimeter are not permitted, or it will affect the
measuring accuracy, even cause irreversible damage.

2
I. Button Description
The following is a brief introduction of the buttons. We will give more detailed information
about its function separately in next chapters.
Figure 1 Button Function
Button Function Introduction
1. Testing
2. Menu
3. Up/Print
4. Down/Save
5. Enter
6. Back

3
II. Interface Description
Figure 2 Interface Description
Interface Description
1. Power Switch: Press the button to turn on the colorimeter. Press the button
again. The button will pop up. Then the colorimeter is turned off.
2. DC Interface: Connect with AC adapter. It is used to connect to external power
source. The specification of external power source is 5V=2A.
3. USB Interface / RS-232 Interface: This interface is a common interface. The
instrument automatically judges the connection status. USB interface is used to
transfer data to PC. Its baud rate is 115200bps. RS-232 interface is used to
connect to the printer; its baud rate is 19200bps.
Note: When connecting to external power source, please press the power
switch to start the instrument.
III. Battery Description and Installation

4
Please use original Li-ion battery. Do not use other batteries, or it will cause irreversible
damage.
Please take out the battery to prevent the colorimeter from the damage of battery leakage
if you don’t use it for a long time.
When using external power source or connecting USB interface to PC, if you press the
power switch, it will charge the Li-ion battery. If you don’t want to charge the battery,
please take out the battery.
When charging the battery, dynamic battery icon will display on the top right corner of
“Standard Measurement” and “Sample Measurement” interfaces. In figure 3, the charging
icon displays in “Standard Measurement” interface; if no charging, the dynamic icon will
not appear.
Figure 3 Charging Icon

5
Figure 4 Battery Installation
Installing Battery
When installing battery, be sure the power switch is popped out (The power source of
colorimeter is cut off). Then remove the battery cover.
Insert the Li-ion battery into battery compartment and push it gently in right direction.
Put the battery cover on Li-ion battery, and then push it up.
Battery Specification: Li-ion 3.7V==0.5A.

6
IV. 200 Colorimeter Operating Instruction
(I)Turning Power On
1. Preparations Before Power On
a) Check whether there is battery powered or external power supply.
b) Make sure the white calibration cover is connected to the instrument and well
installed. If it is loose or the white calibration cover isn’t well installed, you must
put the cover on and be sure it is connected to the instrument tightly.
2. Turning Power On
Press the power switch on the back side of the instrument, the LCD screen will
display MRC logo. After a few seconds, it will enter Standard Measurement
interface automatically, and the default display is L*a*b*C*H.
3. White Calibration and Black Calibration
Note: White and black calibration is not required for 200 colorimeter.
Only when the measured data are not accurate enough, manual calibration
is needed.
After turning on 200 colorimeter, press “Menu” to enter the main menu, as
shown in Figure 5. Select “Calibrate” to enter White and Black Calibration
interface, as shown in Figure 6.
Be sure the white calibration cover is well installed. Selecting “White Calibration”
and press “Enter” button, the instrument will prompt you to place the white
calibration plate. Then press “Enter” or “Testing” key to start white calibration.
Be sure the white calibration cover is removed. Select “Black Calibration” and
press “Enter”. The instrument will prompt you to direct the measuring aperture to
the air. Press “Enter” again or press “Testing” key to start black calibration.
Note: When starting black calibration, direct the measuring port to the air.
Be sure the black calibration is performing in a dark, no bright light source
environment. Keep the measuring port more than 3m away from any
reflective items (hands, desks, walls etc.).
By then, manual white calibration and black calibration are completed.

7
Figure 5 Main Menu
Figure 6 Manual Calibration
(II)Measurement
1. Locating and Measuring Method
200 colorimeter is locating through the facula. The method: enter “Standard
Measurement” or “Sample Measurement” interface, press the “Testing” key and
hold it. The facula will appear at the moment. You can observe the matching
status between the facula and the measured sample. At the same time, hold the
measuring aperture close to the measured sample and adjust it. Then the
alignment is achieved.
After the locating, release “Testing” button. The instrument will finish sample
testing in approx. 1 sec and display color parameters of the measured sample.
2. Standard Measurement
There are two conditions about standard measurement. One is performing
standard measurement after power on; another is after completing sample

8
measurement or other operations, press “Back” button continuously to enter
standard measurement.
a) Standard Measurement after Power On
After power on, the colorimeter will display “Standard Measurement”, as shown in
Figure 7. Then, align measuring aperture to the standard, press “Testing” key, the
screen will display color parameters of this standard. Press “Enter” button, the
instrument will enter “Sample Measurement” interface.
Figure 7 Standard Measurement Interface
b) Standard Measurement after Completing Sample
Measurement or Other Operations
After completing sample measurement or other operations, the instrument may
display an interface. At this moment, press “Back” button repeatedly until the
interface back to “Standard Measurement” interface, as shown in Figure 7. Then,
perform standard measurement according to step a).
3.Sample Measurement
After completing standard measurement, press “Enter” button, the instrument will
enter “Sample Measurement” interface automatically, as shown in Figure 8. Align
the measuring aperture to the test sample to perform sample measurement.

9
Figure 8 Sample Measurement Interface
Note: During the measurement (Approx. 1 sec), all buttons are ineffective.
(III)Save Data
There are two methods to save data.
1. Auto Save
Press “MENU” button to enter the main menu, as shown in Figure 5. Select
“Settings” to enter an interface shown in Figure 9. Select “Auto Save” to enter
status setting interface, as shown in Figure 10. Select “On” and press “Enter”
button to save the setting. After completing this setting, the measurement data
will be saved automatically.
Figure 9 Settings Interface

10
Figure 10 Auto Save
2.Manual Save
In Figure 10, select “Off”, then the data will not be saved automatically. If you want
to save data when measuring, please press “Down/Save” button after each
measurement.
(IV)Connect 200 Colorimeter to the PC
Press “MENU” button to enter the main menu, as shown in Figure 5. Then, select
“Comm” to enter an interface shown in Figure 11. Follow the instruction
information mentioned in the interface to connect 200 colorimeter to the PC
with USB cable. When the communication is successful, the instrument will enter
“Communicating…” interface, as shown in Figure 12.
Figure 11 USB Communication

11
Figure 12 “Communicating” Interface
(Ⅴ)print
Only when the colorimeter is in “Standard Measurement”, “Sample Measurement”
or “Record” interfaces, you can print data through the printer.
Connect colorimeter to the printer. When the colorimeter is in one of the above
interfaces, press and hold (about 5 second) “Up /Print” button to start the printer.

12
V. System Function Description
Except “Standard Measurement” and “Sample Measurement” interfaces, you should enter
other function interfaces through the main menu. Main menu is shown in Figure 13.
Figure 13 Main Menu Interface
1. Record and Standard Entering
a) Record
Select “Record” in main menu to enter “Standard Record”, as shown in Figure 14.
The figure shows the standard parameters. You can check different standard data
through “Up” and “Down”. In figure 15, “T002” is a standard number. After
selecting a standard, you can press “Enter” to check sample parameters and
color difference parameters, as shown in Figure 15. You can check different data
through “Up” and “Down”. No.001 in Figure 15 is the serial number of sample
measurement.
Figure 14 Standard Record

13
Note: “↖Delete﹡” means press Back key “↖” and hold it for
3 seconds, then the record will be deleted.
Figure 15 Sample Parameters under the T002 Standard
b) Standard Entering
In some cases, it’s needed to measure color difference under a saved standard.
Then, you can select “Record” in main menu to enter standard sample records
interface. You can search the needed standard data through “Up” and “Down”.
After finding it, press “Menu” button, and then the standard record is entered to
the measurement interface, press “Enter”, you can perform sample measurement
under this standard.
c) Sample Record Entered to a Standard
In some cases, it’s needed to use a stored sample as a standard. Then, you can
select “Record” to enter sample records interface, as shown in Figure 15. You can
search the needed sample data through “Up” and “Down”. After finding it, press
“Menu” button, and then the sample record is entered to the measurement
interface as a standard, press “Enter”, you can perform the color measurement
under this standard.
2. Calibration
After turning on 200 colorimeter, press “MENU” to enter the main menu, as
shown in Figure 13. Select “Calibrate” to enter manual white calibration and black
calibration interface, as shown in Figure 16.
Be sure the white calibration cover is well installed on the instrument. Select
“White Calibration” and press “Enter” button, the interface will prompt you to put
away the white calibration cover, then press “enter” again or press “Testing” key
to perform white calibration. Before performing black calibration, be sure the
white calibration cover has been removed. Select “Black Calibration” and press
“Enter” button, the instrument will prompt you to direct the measuring aperture to
the air, press “Enter” again or press “Testing” key to perform black calibration.

14
Note: When performing black calibration, direct the measuring port to the
air. Be sure the black calibration is starting in a dark, no bright light
source environment. Keep the measuring port more than 3m away from any
reflective items (hands, desks, walls etc.).
Then, manual white calibration and black calibration are completed.
Figure 16 Calibration Interface
3. Tolerance Setting
Select “Tolerance” in main menu to enter tolerance setting interface, as shown in
Figure 17. You can add or subtract the number in which the cursor is positioned
through “Up” and “Down” button. After setting the number to the needed one,
press “Enter” button, the cursor will jump to the last number. When the cursor is in
the last number, press “Enter” button to save the settings and return to the main
menu.
If you don’t want to set or modify the tolerance, you can press “Back” button to
return to the main menu.

15
Figure 17 Tolerance Setting
4. USB Communication
Select “Comm” in the main menu to enter an interface shown in Figure 18.
According to the prompt, connect 200 colorimeter to the PC with USB cable.
When the communication is successful, the instrument will enter
“Communicating…” interface, as shown in Figure 19.
Figure 18 USB Communication Interface
Table of contents
Other MRC Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Ohmic instruments
Ohmic instruments DPSC-35XR manual

Badger Meter
Badger Meter Magnetoflow Mag Meter Installation and operation manual

Emerson
Emerson X-STREAM Series instruction manual

Spectra Precision
Spectra Precision Focus 30 Service manual

Tru-Test
Tru-Test 703 user manual

IUL
IUL iPeak Europium Instructions for use