To change the color palette, swipe the image right or left.
•
To set the temperature threshold to “hot-red” and “cold-blue” palettes, press the
temperature reading in the middle of the temperature scale.
•
To change the temperature from Celcius to Fahrenheit, click on the settings button
, click on
Theromography Settings
and then on
Units
.
•
Pinch the touch screen for digital zoom.
•
To use the automatic temperature range, press the current temperature and then
press the button on the dial.
•
For best performance, the ambient room temperature should be 25° C.
•
For optimal performance wait between 8-10 minutes from start-up, to allow the IR-
CAM-FEVER to reach an initial steady temperature state.
•
In order to establish a baseline temperature, have 3 healthy people with
regular temperatures pass in front of the camera. Calculate from the highest
temperature of the 3 people and rasie the temperature dial 1.5° C (see
Configuring the Alert
for instructions). This will be the threshold for the alert for
an individual with a fever.