
Page <6> V1.004/02/22
Newark.com/multicomp-pro
Farnell.com/multicomp-pro
sg.element14.com/b/multicomp-pro
5. Operation:
5.1 Preparing the test
Prior to every test, please ensure that the instrument is in perfect condition:
• For example, keep an eye out for a broken housing or leaking batteries.
• Always carry out a function test before using the voltage tester, see below.
• Check that the instrument is functioning properly(for example at a known voltage source)before and after every test.
• If the safety of the user can not be guaranteed, switch o the instrument and secure it to prevent unintentional usage.
Carrying out a function test
Connect the voltage tester probe tips for 4 to10 seconds and then disconnect. With the exception of the LED for low imped-
ance test, all LEDs should light up. All segments are illuminated on the LCD display.
The functioning of the voltage detector is to be checked shor tly before andafter use by using the testing element. If the indi-
cation “not ready”appears or if theindication of one or more steps fails, or if no functioning is indicated, the voltage detectoris
no longer to be used.
5.2 Voltage Test
• Connect both test probes with power source.
• As from a voltage of >6V the voltage tester switches on automatically.
• The voltage is displayed via LEDs. The dierent indicating signals of the voltage detector (including the ELV limit
indication)are not to be used for measuring purposes.
• The voltage is also shown on the LCD display.
• For AC voltage, the “AC” is illuminated;
For positive voltage, the “+” is illuminated;
For negative voltage, the “-” is illuminated.
• In the case of DC voltage,the polarity of the indicated voltage relates to the
voltage tester probe tip
• Once the safety extra-low voltage (50V AC /120V DC)is reached or
exceeded, the “ ” is illuminated, in the event of no battery power or main circuit failure,
• and an acoustic signal is emitted.
• Once voltage is applied to the measuring instrument,press the HOLD buttom, the LCD and
LEDs display shows the
recoarded reading.
• To delete the recorded value,press the HOLD button once again.the LCD and LEDs display once again indicates the
voltage currently being applied to the probe tips.
5.3 Single-pole phase Test
• The single-pole phase test is only possible when batteries are installed and in good condition.
• The single-pole phase test starts at an AC voltage of approx.100V (pole>100V AC).
• When using single-pole phase tests to determine external conductors the display function may be impaired under cer tain
conditions (e.g. for insulating body protective equipment on insulation locations).
• The single-pole phase testing is not appropriate to determine whether a line is live or not. For this purpose, the
double-pole voltage test is always required.
• Connect both test probes with power source.
• A signal sound indicates the phase.
• The “ ” LED is illuminated in the display.