4
6. Release the push button while the desired calibration solution value is displayed. (Remember to calibrate the lower value
first). The LED will begin to flash rapidly.
7. While the LED flashes rapidly, dip the pen in the prepared calibration solution so that the sensor is completely submerged and
swirl it around to remove any bubbles that may have been caught in the sensor.
8. While the LED flashes slowly, the display will show “[value] ppm”. Hold the sensor steady during calibration.
9. If the first calibration point is successful, the display will indicate “SAVEd”, then “PUSHnHLD” will be displayed.
10. Rinse the sensor in clean water (preferably DI, RO, or distilled) and pat it dry with a clean soft lint-free cloth (you will have
up to 25 seconds for this step).
11. Push and HOLD the push button to continue. The display will alternate between the remaining options for calibration values.
NOTE: If you press and release the button at this point, or let the pen timeout, a 1-point calibration will be stored. You would
need to start over for a 2-point calibration.
12. Release the push button while the 2nd calibration point value is displayed. The LED will begin flashing rapidly.
13. While the LED is flashing rapidly, dip the pen in the 2nd prepared calibration solution so that the sensor is completely
submerged and swirl it around to remove any bubbles that may have been caught in the sensor.
14. While the LED flashes slowly, the display will alternate between “CAL” and “[value]”. Hold the sensor steady during
calibration.
15. If the second calibration point is successful, the display will indicate “SAVEd 2PT” and the PT6 will power OFF.
16. Verify calibration by retesting the prepared calibration solutions. The values should be within ±10% of the stated value of the
standard solutions.
17. Remove the sensor from the solution and rinse it with clean water (preferably DI, RO, or distilled) and pat it dry with a clean
soft lint-free cloth. Use caution not to touch the ISE membrane.
18. Fill the protective cap half-full with lower valued prepared calibration solution. Carefully put the cap back on the sensor for
conditioning (using caution as excess solution may squirt out).
NOTE: Individual calibration points can be verified by retesting the prepared calibration solutions. The PT6 will also check the
gain/slope of the calibration during a 2-point calibration. If, after the second calibration point the PT6 detects an error in the
gain/slope, it will display “Error”. If this happens, clean and condition the sensor, then verify you are using approved and properly
prepared calibration solutions, and recalibrate.
NOTE: If at any point during calibration you do not submerge the sensor in solution before the flashing slows, allow the pen to
power OFF and start over.
III. 1-POINT CALIBRATION
1. Remove the protective cap. Rinse the sensor for about 5 seconds with clean water (preferably DI, RO, or distilled) and pat it
dry with a clean soft lint-free cloth.
2. Press and release the push button to turn the PT6 ON.
3. Push and hold the push button to enter menu mode.
4. Release the button while “CAL” is displayed. The display will alternate between “PUSHnHLD” and “CAL”.
5. Push and hold the push button. The display will alternate between “1 ppm”, “10 ppm”, “100 ppm”, “1000 ppm”, and “ESC”.
6. Release the push button while the desired calibration solution value is displayed. The LED will begin to flash rapidly.
7. While the LED flashes rapidly, dip the pen in the calibration solution so that the sensor is completely submerged and swirl it
around to remove any bubbles that may have been caught in the sensor.
8. While the LED flashes slowly, the display will show “[value] ppm”. Hold the sensor steady during calibration.
9. If the calibration is successful, the display will indicate “SAVEd”, then “PUSHnHLD” will be displayed.
10. Press and release the button to exit or allow the PT6 to time out WITHOUT pushing the button.
11. Verify calibration by retesting the prepared calibration solution. The value should be within ±10% of the stated value of the
standard solution.
12. Fill the protective cap half full with prepared calibration solution. Carefully put the cap back on the sensor for conditioning
(using caution as excess solution may squirt out).
NOTE: If at any point during calibration you do not submerge the sensor in solution before the flashing slows, allow the pen to
power OFF and start over.
IV. FACTORY CALIBRATION
The PT6 can be returned to factory default calibration using the FAC CAL function. This will erase any stored wet calibration.
NOTE: Default factory calibration resets the electronics only and does NOT take the condition of the sensor into consideration.
To return your unit to factory calibration:
1. Press and release the push button to turn the PT6 ON.
2. Push and hold the push button to enter menu mode.
3. Release the button when “FAC CAL” displays. The display will alternate between “PUSHnHLD” and “FAC CAL”.
4. Push and hold the push button. “SAVEd FAC” will be displayed indicating the pen has been reset to its factory calibration.