2|ni.com |NI PXIe-4339/TB-4339 Installation Guide and Terminal Block Specifications
Electrical ...........................................................................................................................15
Mechanical........................................................................................................................16
Physical .............................................................................................................................16
Environmental Specifications ........................................................................................... 16
Operating Environment............................................................................................. 16
Storage Environment ........................................................................................................ 17
Shock and Vibration ......................................................................................................... 17
Safety ................................................................................................................................17
Safety Standards ............................................................................................................... 17
Electromagnetic Compatibility ......................................................................................... 18
CE Compliance ................................................................................................................. 18
Online Product Certification............................................................................................. 18
Environmental Management............................................................................................. 19
World Wide Support and Services............................................................................................ 19
Electromagnetic Compatibility Guidelines
This product was tested and complies with the regulatory requirements and limits for
electromagnetic compatibility (EMC) stated in the product specifications. These requirements
and limits provide reasonable protection against harmful interference when the product is
operated in the intended operational electromagnetic environment.
This product is intended for use in industrial locations. However, harmful interference may
occur in some installations, when the product is connected to a peripheral device or test object,
or if the product is used in residential or commercial areas. To minimize interference with radio
and television reception and prevent unacceptable performance degradation, install and use this
product in strict accordance with the instructions in the product documentation.
Furthermore, any modifications to the product not expressly approved by National Instruments
could void your authority to operate it under your local regulatory rules.
Caution To ensure the specified EMC performance, operate this product only with
shielded cables and accessories. Cables connected to the PXIe-4339 using the
TB-4339/B/C terminal block must use only twisted, shielded pair cables. Use one
twisted, shielded pair for each EX+/EX-, RS+/RS-, AI+/AI-, AO+/AOGND,
PFI0/DGND, and T+/DGND that is cabled. Cable shields must be terminated to
chassis ground at both ends.
What You Need to Get Started
To set up and use a NI PXIe-4339 module with a TB-4339/B/C terminal block, you need the
following items:
Hardware
– NI PXIe-4339 module
– TB-4339/B/C terminal block