Nellcor ULTRA CAP N-6000 User manual

SERVICE
MANUAL
NELLCOR®
ULTRA
CAP™
PULSE
OXIMETER
AND
CAPNOGRAPH,
MODEL
N-6000
Caution:
Federal
law
(U.S.)
restricts
this
device
to
sale
by
or
on
the
order
of
a
physician.
Nellcor
Incorporated
25495
Whitesell
Street
Hayward,
California
94545
U.S.A.
510
887-5858
1-800-NELLCOR
©
1993
Nellcor
Incorporated
.
028065A-0193

Corporate
Headquarters
Nellcor
Incorporated
25495
Whitesell
Street
Hayward,
California
94545
USA
Tel
510-887-5858
ii
European
Office
Asia/Pacific
Headquarters
Nelleor
BV
Hambakenwetering
1
5231
DD
’s-Hertogenbosch
The
Netherlands
Tel.
+31.73.426565
Nellcor
Limited
Suite
1204C
Admiralty
Centre,
Tower
1
18
Harcourt
Road
Hong Kong
Tel.
+852.529.0363
European
Regional
Offices
Northwest
Europe
Nellcor
Europe
BV
Hambakenwetering
1
5231
DD
‘s-Hertogenbosch
The
Netherlands
Tel.
+31.73.426565
Middle
and
Central
Europe
Nelleor
GmbH
.
Black-&-Decker-Strasse
28
W-6270
Idstein
Germany
|
Tel.
+49
6126.5930
Southern
Europe
Nellcor
Sarl
3,
rue
du
Petit
Robinson
78353
Jouy-en-Josas
Cedex
France
Tel.
+33.1.39.46.96.58
European
Local
Offices
Nellcor
(UK)
Limited
University
of
Warwick
Science
Park
Coventry
CV4
7EZ
United
Kingdom
Tel.
+44.203.690220
Nelleor
(Belgium)
NV/SA
Interleuvenlann
62/8
.
B-3001
Leuven
Belgium
Tel.
+32.16.400467

11
12
13
2.1
2.2
2.3
2.4
2.5
3.1
3.2
3.3
3.4
TABLE
OF
CONTENTS
Symbols
AAA
We
Introduction...
나
Introduction.........................
…
1-1
Warnings,
Cautions,
and Notes
...
…
1-1
Description..................,..,............,...
..
1-2
1.8.1
Visible
and
Audible
Indicators.
„12
13.2
Audible
Indicator.......................
..
1-2
13.3
Visible
and
Audible
Alarms.
12
13.4
StatusMessages..............................
.
1-2
1.3.5
Automatic
Self-Test
and
Warm-Up
Time..
..
1-3
1.3.6
On-Screen
Menus..................,...,,,.......
…
13
1.3.7
N-6000
Default
and
Custom
Default
Set-Up
.
....
13
18.8
Trend
Memory......:...,...............,,,,,......
レレ
くく
レト
ニュ
レッ
くく
トー トト
ャ
ッッ て と
と ャ
て て
と
て て
とく
ここ
13
1.3.9
Automatic
Calibration..…..............,........,.,.......ss
ss
13
1.3.10
Battery
Operation
..................
13
1.3.11
Noninvasive
Oximetry
Sensors.........................
1-4
13.12
C-LOCK™
ECG
Synchronization
for
Pulse
Oximetry..
Principles
of
Operation.
.
2-1
Overview.........................
..
21
PulseOximetrySubsystem.......................
‚...
2-1
2.2.1
C-LOCK
ECG
Synchronization
„21
2.2.2
Automatic
Calibration..................
„22
2.2.8
Functional
versus
Fractional
Saturation...
2.2.4
-
Measured
versus
Calculated
Saturation
....
ーー
Capnography
Subsystem
............................
.
2-3
N20/02
Compensation
............
..
2-5
24.1
Standard
Gas
Conditions
...
..
25
2.4.2
Pressure
Broadening
Compensation...
2-5
2.4.3.
N20
Collision
Broadening...........
.
2-6
2.4.4
O2
Collision
Broadening.........................,.,...........ss
2-6
2.4.5
Water
Vapor
Effect...
issues
2-6
2.4.6
BTPS/ATPS
Compensation.
2-6
2.4.7
Summary:
Reported
CO2
Values...
..
2-6
.
Factory
Calibrated
Sensor
μμ
ΕΟΟ
ρου
..
2-7
Circuit
Analysis
Introduction.................
.
3-1
CO2
Module
Circuit
Details.
..
3-2
3.2.1
Sensor
Operation.........
..
3-2
3.2.2
CO2
Module
Hardware
....
...
..
3-2
3.2.3
bDigital.....................
...
..
3-2
8.2.4
Motor
Control........
3-3
3.2.5
Signal
Amplifier..
…
3-8
3.2.6
Heater
22220
…
..
3-8
3.2.7
Source....
3-3
8.2.8
Barometer...
..
3.2.9
StatusLEDs....................
…
3-3
3.2.10
Status
LED
Summary............
...
BA
Oximetry
Module
Block
Diagram
Analysis
…
3-4
©
8.8.1
OximetryModule........................
...
3-4
Detailed
Oximetry
Module
Circuit
Analysis
3-6
3.4.1
Oximetry
Module
sense
ας
3-6

3.5
3.6
4.1
4.2
4.3
4.4
5.1
5.2
5.3
6.1
6.2
7.1
7.2
7.3
8.1
8.2
8.3
9.1
iv
TABLE
OF
CONTENTS
(continued)
Main
Processor
PCB
Circuit
Details.
essences
8.5.1
Microprocessor
Section.......,...............,....,......,
8.5.2
Power.
Control.........
3.5.3
Reset
and
Watehdog......................
3.5.4
Battery
Backup
and
Memory
Retention............
8.5.5
Internal
and
External
Serial
Communications..
8.5.6
Graphical
Display
..................,....,,.....,..,.
3.5.7
BEPROM..............
8.5.8
FLASH
Memory.
8.5.9
Real
Time
Clock....
3.5.10
Audio
Generation
....
3.5.11
Charging
Indicator................
3.5.12
Patient
Isolation
Power
Supply...
3.5.13
Front
Panel
Display
Controller..
Power
Supply-Charger
PCB
Circuit
Details.........
3.6.1
Power
Supply
Charger
Theory
of
Operation..
3.6.2
AC
Rectifier
and
+12
volt
Battery
Charger...........
3.6.3
+5
Volt
Logic
and
Display
Power
Supply
....
3.6.4
C-LOCK
QRS
Sync
Input
Circuit...
…
Routine
Maintenance
......................
Introduction............
Cleaning
Instructions............
Changing
Mains
Voltage
Input
....
Replacing
or
Changing
the
Fuse...
Packing
and
Shipping......
Överview...........................
Repacking
In
Original
Carton..
Repacking
In
New
Carton
.......
Testing
and
Calibration
............
Description.....................
COs
Display
ConventionS
0
Disassembly
Guide........
Introduction............
..
”Disassempbly
Procedure
ees
enten
ennen
nernnse
6.2.1
RemovinginstrumentCover...........................
eker
6.2.2
Removing
了
attery.
<
scccrcre*
6.2.3
Removing
Oximetry
and
C02
Subassembly..
„
6.2.4
Removing
Main
Processor
PCB...........................
0040000400000
000000e0
en
6.2.5
Removing
Front Panel
Assembly
.............................,................,.....
6.2.6
Removing
Pull-Out
Card
Tray...
6.2.7
Removing
Speaker............................
6.2.8
Removing
Power
Supply-Charger
Assembly.
6.2.9
Removing
Transformer..........
7.2.1
Service
Screen
and
Display
Conventions...................
7.2.2
Breath
Detection
and
Display
of
End-Tidal
CO2
Values..
…
72
Calibration
and
Accuracy
Check
ui
iii
..
7-6
Troubleshooting........................
.
8-1
Introduction.......
.
8-1
Advisory
Messages.
ーー
..
8-6
Status
Messages.....
...
..
8-7
Spare
Parts........
…
..
9-1
In
го
ас
6101...
sise
9-1

TABLE
OF
CONTENTS
(continued)
10
¡ACA
essences
10.1
Overview.........
1
Specifications.............
11.1
Physical/Environmental.........
11.2
Components
and
User
Interface..
113
Controls
and
Indicators...........
114
Connectors................
185
Performance.
11.6.
Calibration.....
11.7.
Airway
Adapter
EP
O
KO KO
R
O
R
K
KK
O
RR
ek
6
11.8
再
jectrical.
0
0000
Koh
OK
S
KO
RS
O
O
KOK
KK
P
KKK
oP
P
een
εως
2
Schematic
Diagrams..
.
12.1
Introduction...................,......,.................
see
K R
PK
O
v
Kone
LIST
OF
FIGURES
2-1
Oxyhemoglobin
Dissociation
Curve..........................................
0.0000
öseeeeeea
2-2
Nondispersive
Infrared
CO2
Analyzer..
3-1
Overall
Block
Diagram..........................
3-2
Oximetry
Module
Overall
Block
Diagram...
3-3
LRD
Driver
Cireuit
に
し に
に に
に
レート
ーーー・
3-4
Input
Signal
Processing.
3-5
Input
Amplifier,
Synchronous
Detector,
and
Filter/Amplifier
8-6
A:D
Conversion
Circuits.......................,.......,,,...........
8-7
Support
Circuits...................
..
4-1
North
American
Fuse
Аттапветейв.............
линии
нии
ини
тни
низине
нение
4-2
European
Fuse
Атгапретейф...............
ss
наити
6-1
Disassembly
Diagram..............
7-1
Failed
LED
Digits
7-1
Service
Screen
IndicatiomS
レレ
ーー
トレ
ーー
トト
ーー
スー
スト
ドー
くに
トー
トト
トト
トス
スレ
(レレ
に
レレ
に
トス
レト
ストー
て
8-1
Troubleshooting
Guide
The
following
are
trademarks
of
Nellcor
Incorporated:
NELLCOR,
which
is
registered
in
the
United
States
and
other
countries;
C-LOCK,
which
is
registered
in
the
United
States;
OXISENSOR;;
and
ULTRA
CAP,
which
is a
commercial
trademark.
The
N-6000
is
covered
by
the
following
patents:
U.S.
Patent
No.
4,621,643;
4,653,498;
4,700,708;
4,770,179;
4,802,486;
4,869,254; 4,928,692;
4,934,372;
and
corresponding
patents
in
other
countries,

O/O
Sc
nx
el
κ
SYMBOLS
FRONT
PANEL
On/Standby
eke
Pulse
tone
volume
MENU
Audible
alarm
off
A
+
ni
C-LOCK
signal
lost
Low
battery
Battery
charging
220/240V~
INSTRUMENT
ONLY
FRONT
PANEL
A
vi
Attention:
Refer
to
operator’s
manual!
Freeze
Menu
Audible
alarm
on
C-LOCK
in
use
Battery
in
use
Type
BF
equipment,
(patient
electrically
isolated)

SYMBOLS
BACK
PANEL
=
Battery
fuse
Fuse
replacement
AS
ECG
input
O
Eguipotential
ground
MO
On/Off
<>
RS232
RS232inputoutput
port
Type
BF
equipment,
À
(patient
electrically
isolated)
110/120
VOLT
INSTRUMENTS
ONLY
BACK
PANEL
Attention:
Refer
to
operator’s
Instrument
not
anesthetic
VAN
manual!
proof.
vii

WARNINGS
The
N-6000 contains
no
user-serviceable
parts.
For
protection
against
electrical
hazard,
all
service
must
be
performed
by
qualified
personnel.
.
[For
protection
against
fire
hazard,
replace
fuses
only
with
the
same
type
and
rating.
The
N-6000
is
a
patient-connected
medical
device.
Isolated
patient
connectors
protect
the
patient
from
potentially
dangerous
electrical
potentials
or
ground
paths.
To
protect
the
integrity
of
this
connection,
the
procedures
and
part
specifications
contained
in
this
manual
must
be
adhered
to.
viii

SECTION
I
Introduction
1.1
INTRODUCTION
This
manual
contains
information
for
servicing
the
NELLCOR®
ULTRA
CAP™
pulse
oximeter
and
capnograph,
model
N-6000.
Service
of
this
product
must
be
done
by
qualified
service
personnel,
who
have
a
technical
background
in
analog
and
digital
electronics.
Note:
This
manual
is
written
for
an
N-6000
configured
for
English-language
displays
and
service
screens.
If
the
unit
is
configured
for
French
or
German,
the
technician
must
change
the
language
option
to
English.
To
reach
the
language
selection
menu
from
the
main
monitoring
screen,
press
the
Freeze
button
on
the
front
panel
while
the
main
monitoring
screen
is
displayed,
and
then
press
the
second
and
fourth
soft
keys
from
the
left
at
the
same
time.
The
legend
“LANGUAGE”
(LANGUE
in
French
and
SPRACHE
in
German)
appears
above
the
fourth
soft
key
from
the
left.
Press
this
key
to
gain
access
to
the
language
selection
menu
and
select
“ANGLAIS”
or
“ENGLISCH.”
ㆍ
The
N-6000
is
a
compact,
microprocessor-controlled
instrument
used
for
continuous
real-
time
monitoring
of
oxygen
saturation
(SpO2),
pulse
rate
(PR),
inspired
CO2
(ins.),
end-tidal
carbon
dioxide
(ETCO2),
and
respiratory
rate
(RR).
.
SpO2
is
noninvasively
and
continuously
measured
using
Nellcor's
reusable
or
adhesive
SpO2
sensors
which
are
available
for
a
variety
of
patient
sites.
SpO2
values
are
displayed
numerically,
while
the
plethysmograph
is
displayed
as
a
waveform,
A
bar
graph
indicator
shows
the
patient’s
relative
pulse
strength
as
measured
by
the
pulse
oximetry
circuits.
ㆍ
ETCOs
is
determined
by
positioning
an
infrared
mainstream
CO2
sensor
across
an
intubated
patient’s
airway
(15
mm
diameter).
The
amount
of
CO2
present
at
the
end
of
exhalation
is
displayed
numerically,
while
CO2
is
displayed
as
a
waveform.
This
manual
is
intended
for
use
by
authorized
service
personnel,
trained
in
servicing
analog
and
digital
patient
monitoring
equipment.
Service
personnel
must
have
read
ahd
understood
the
N-6000
operator’s
manual
and
be
familiar
with
instrument
operation
before
attempting
maintenance
or
repair.
12
WARNINGS,
CAUTIONS,
AND
NOTES
Before
you
read
the
N-6000
service
manual,
it
is
important
to
understand
the
following
terms.
These
terms
identify
information
that
pertains
to
technician
and
patient
safety,
and
indicates
proper
operation
of
this
instrument.
WARNING:
A
warning
describes
a
condition
that
may
result
in
injury
to the
patient,
operator
or
service
technician.
WARNINGS
are
always
in
boldface
type
and boxed.
Caution:
Cautions
describes
a
condition
that
may
result
in
damage
to
the
instrument.
Cautions
are
always
in
boldface
type.
Note:
A
note
gives
information
that
warrants
special
attention.
1-1

1.3
DESCRIPTION
The
following
paragraphs
describe
the
N-6000
and
list
important
features.
1.3.1
Visible
and
Audible
Indicators
The
N-6000
features
numeric
displays
of
SpO2,
PR,
RR,
and
ETCO2:
.
C02
and
plethysmographic
waveform
displays:
and
a
pulse
amplitude
bar
graph,
which
*
provides
a
qualitative
indication
of
pulse
strength
at
the
oximetry
sensor.
9
Trends
are
acquired
for
all
parameters,
and
the
following
five
trend
screens
can
be
displayed:
SpO2
trend
alone;
CO2
trend
alone;
SpO2
and
CO2
trends;
pulse
rate
and
respiratory
rate
trends;
SpOz,
CO2,
pulse
rate,
and
respiratory
rate
trends.
ㆍ
When
the
operator connects
the
monitor
to
AC
power
and turns
on
the
rear-panel
on/off
switch,
a
battery
charging
indicator
lights.
1.3.2
Audible
Indicator
Pulse
rate
and
oxygen
saturation
are
indicated
audibly
with
a
tone
that
signals
each
pulse.
The
pitch
of
this
tone
changes
with
variation
in
SpO2,
rising
as
saturation
increases
and
falling
as
it
decreases.
This. early
warning
system
encourages
prompt
corrective
action
since
the
clinician
can
watch
the
patient
and
listen
for
SpO2
changes
simultaneously.
1.3.3
Visible
and
Audible
Alarms
The
monitor
has
both
visible
and
audible
alarms.
These
alarms
are
activated
when
a
variable
moves
outside
an
adjustable
limit
(operator-defined),
when
the
monitor
detects
loss
of
pulse
or
apnea,
or
when
a
sensor
gets
disconnected.
The
tone
and
pattern
of
the
audible
alarm
depend
on
the
priority
of
the
alarm
state.
Pressing
the
ALARM
SILENCE
button
turns
off
the
audible
portion
of
the
alarm
temporarily.
A
flashing
ALARM
SILENCE
indicator
adjacent
to
the
ALARM
SILENCE
button
alerts
the
operator
that
the
alarm
tone
has
been
silenced
temporarily.
A
steady
ON
of
the
ALARM
SILENCE
indicator
warns
that
one
or
more
parameters
have
had
their
audible
alarm
function
disabled.
Visible
alarms
appear
on
the
monitor
sereen,
and
unlike
audible
alarms,
they
are
always
operational.
Note:
The
SpO2
and
ETCO2
numeric
indicators
change
from
green
to
red
when
an
alarm
occurs.
Note:
When
the
French
language
is
chosen
and
one
or
more
of
the
parameters
have
had
the
error
audible
alarm
function
disabled,
a
single tone
occurs
every
three
minutes.
1.3.4
Status
Messages
A
status
message
is
displayed
in
case
an
error
condition
occurs.
An
error
condition
will
be
identified
by
a
module
identifier
and
an
error
code
number
that
assists
service
personnel
in
troubleshooting
the
problem.
(See
section
VIII
for
more
information.)

1.3.5
Automatic
Self-Test
and
Warm-Up
Time
The
monitor
automatically
performs
a
series
of
diagnostic
tests
when
turned
on.
The
system
self-
test
takes
approximately
15
seconds
after
the
operator
turns
on
the
monitor.
These
tests
confirm
that
the
program
memory,
data
memory,
and
internal
circuitry
are
functioning
properly,
while
allowing
the
mainstream
COz
sensor
to
warm.
The
CO2
sensor
warm-up
time
is
approximately
45
seconds.
If
an
error
is
detected,
a
status
message
appears.
(Refer
to
Section
8,
“Troubleshooting,”
for
more
information.)
—
1.8.6
On-Screen
Menus
The
on-screen
menu
guides
the
operator
through
all
system
functions.
Menu
items
are
displayed
at
the
bottom
of
the
screen
just
above
the
four
function keys
that
are
used
to
select
an
item.
When
a
function
key
is
pressed,
the
screen
displays
a
new
menu
with
additional
functions.
Press
the
MENU
button
to
display
the
top level
menu,
or
to
return
to
the
main
monitoring
screen.
1.8.7
N-6000
Default
and
Custom
Default
Set-Up
The
N-6000
power-on
default
settings
can
be
customized
according
to
institutional
requirements,
Once
configured,
the
custom
alarm
limits
will
always
be
in
place,
even
after
power-down.
1.3.8
Trend
Memory
The
N-6000
stores
up
to
24
hours
of
trend
data
for
CO2,
SpO2,
pulse
rate,
and
respiratory
rate.
Trend
memory
can
be
viewed
in
30-minute,
2-hour, 4-hour,
8-hour,
12-hour,
or
24-hour
segments.
When
the
memory
is
full,
the
oldest
data
are
automatically
erased
as
new
data
are
stored.
Data
stored
in
memory
can
be
viewed
on
the
sereen
and
can
be
printed
with
a
printer.
Patient
monitoring
continues
while
the
trend
data
are
being
printed.
1.3.9
Automatic
Calibration
The
SpO2
subsystem
of
the
N-6000
is
fully
self-calibrating.
It
is
calibrated
automatically
whenever
the
system
is
turned
on
and
periodically
thereafter.
Additionally,
it
is
recalibrated
automatically
whenever
a
new
oximetry
sensor
is
connected.
The
capnography
subsystem
of
the
N-6000
is
factory
calibrated.
1.3.10
Battery
Operation
If
external
power
is
lost
or
transportable
operation
is
necessary,
the
N-6000
can
operate
up
to
90
minutes
on
its
rechargeable
internal
battery.
This
operating
time
can
be
extended
up
to
180
minutes
by
using
the
GRAPHICS
w/BATTERY
power
saver
option.
ㆍ
By
choosing
GRAPHICS
w/BATTERY
ON
in
the
SYSTEM
menu,
the
display
graphics
(including
all
waveforms,
messages,
etc.)
will
continue
to
operate.
By
choosing
GRAPHICS
w/BATTERY
OFF,
the
display
remains
blank
during
battery
operation.
The
N-6000
displays
only
the
numerical
values
of
SpO2
and
ETCO2,
This
allows
up
to
180
minutes
of
battery
life.
.
When
the
GRAPHICS
w/BATTERY
option
is
selected
to
be
OFF
and
the
unit
is
operating
on
AC
MAINS
power,
the
display
will
only
blank
when
the
unit
transfers
to
battery
power.
Press
the
MENU
button
to
turn
the
GRAPHIC
w/BATTERY
option
ON.
The
display
returns
when
the
operator
reapplies
AC power
to
the
monitor.

1.3.11
Noninvasive
Oximetry
Sensors
Noninvasive
NELLCOR
oximetry
sensors
obtain
measurements
by
optical
means
alone,
using
two
light-emitting
diodes
(LEDs)
as
light
sources.
The
N-6000
adjusts
automatically
for
differences
in
tissue
thickness
or
skin
pigmentation.
Specific
sensors
are
available
for
neonates,
infants,
children,
and
adults.
Refer
to
specific
sensor
directions
for
use
for
complete
information.
1.3.12
C-LOCK™
ECG
Synchronization
for
Pulse
Oximetry
If
a
patient
is
moving
or
has
poor
perfusion,
C-LOCK
ECG
synchronization
can
enhance
signal
quality
for
measurements
of
oxygen
saturation.
When
this
feature
is
used,
the
N-6000
receives
two
separate
signals
that
reflect
cardiac
activity:
an
optical
signal
from
the
sensor
and
an
electrical
signal
from
the
ECG.
The
N-6000
uses
the
ECG
QRS
complex
to
help
identify
the
pulse
and
synchronize
the
SpOz
measurements.
When
C-LOCK
ECG
synchronization
is
used,
the
C-LOCK
IN
USE
symbol
appears
on
the
display.
If
the
ECG
signal
is
lost
or
deteriorates
to
the
point
that
it
can
no
longer
be
used,
the
C-LOCK
SIGNAL
LOST
symbol
appears
on
the
display.
No
symbol
is
displayed
if
the
C-LOCK
feature
of
the
N-6000
is
not
being
used.
:

SECTION
II
Principles
of
Operation
2.1
OVERVIEW
This
section
describes,
in
general
terms,
operating
principles
for
the
N-6000.
2.2
PULSE
OXIMETRY
SUBSYSTEM
The
N-6000
oximetry
subsystem
is
based
on
the
principles
of
spectrophotometry
and
plethysmography.
It
includes
an
‘electro-optical
sensor
and
a
microprocessor-based
module. The
sensor
has two
low-voltage
light-emitting
diodes
(LEDs)
as
light
sources,
and
one
photodiode
as
a
photodetector.
One
LED
emits
red
light
(nominal
660
nm)
and
the
other
emits
infrared
(nominal
920
nm).
When
the light
from
the
LEDs
passes
through
the
sensor
site,
part
of
the
light
is
absorbed.
The
photodetector
measures
the
light
that
passes
through,
which
indicates
red
and
infrared
absorption.
With
each
heartbeat,
a
pulse
of
oxygenated
arterial.blood
flows
to
the
sensor
site.
Oxygenated
hemoglobin
differs
from
deoxygenated
hemoglobin
in
its
relative
red
and
infrared
absorption.
The
N-6000
measures
red
and
infrared
absorption
to
determine
the
percentage
of
functional
hemoglobin
that
is
saturated
with
oxygen.
In
principle,
a
pulse
oximeter
measures
the
light
absorption
by
tissues
and
nonpulsatile
blood.
Absorption
is
also
measured
when
pulsatile
arterial
blood
is
in
the
tissue.
The
ratio
of
absorption
at
both
wavelengths
results
in
a
value
for
the
arterial
oxygen
saturation
(SpO2).
2.2.1
C-LOCK
ECG
Synchronization
Through
C-LOCK
ECG
synchronization,
the
N-6000
uses
an
ECG
signal
as
a
reference
point
for
identifying
the
pulse
and
synchronizing
SpO2
measurements.
This
enhances
signal
quality
during
patient
movement
and
when
the
patient’s
perfusion
is
poor.
When
provided
with
an
ECG
signal,
the
N-6000
receives
two
signals
that
reflect
cardiac
activity:
an
optical
signal
from
the
sensor
and
an
electrical
signal
from
the
ECG.
A
short
time
after
a
QRS
complex
is
detected,
an
optical
pulse
is
detected
at
the
sensor
site.
The
length
of
this
delay
varies
with
the
patient’s
physiology
and
with
the
location
of
the
sensor.
However,
for
a
given
patient,
the
length
of
the
delay
is
relatively
stable.
Through
C-LOCK
ECG
synchronization,
that
time
-
relationship
is
used
to
identify
“good”
pulses
and
reject
nonsynchronized
artifacts
such
as
random
motion.
If
an
ECG
signal
is
not
provided,
or
if
that
signal
deteriorates
so
that
it
can
no
longer
be
used,
the
optical
pulse
alone
is
used
to
determine
the
pulse
rate
and
to
initiate
saturation
measurements.
C-LOCK
ECG
synchronization
resumes
when
an
adequate
ECG
signal
is
available.
2-1

2.2.2
Automatic
Calibration
The
oximetry
subsystem
incorporates
automatic
calibration
mechanisms.
It
is
automatically
calibrated
each time
it
is
turned
on,
at
periodic
intervals
thereafter,
and
whenever
a
new
sensor
is
connected.
Also,
the
intensity
of
the
sensors
LEDs
is
adjusted
automatically
to
compensate
for
differences
in
tissue
thickness
and
pigmentation.
Each
sensor
is
calibrated
when
manufactured:
the
effective
mean
wavelength
of
the
red
LED
is
determined
and
encoded
into
a
calibration
resistor.
The
instrument’s
software
reads
this
calibration
resistor
to
determine
the
appropriate
calibration
coefficients
for
the
measurements
obtained
by
that
sensor.
2.2.3
Functional
versus
Fractional
Saturation
Because
the
N-6000
measures
functional
SaO2,
it
may
produce
measurements
that
differ
from
those
of
instruments
that
measure
fractional
SaO2.
Functional
SaQ2
is
oxygenated
hemoglobin
expressed
as
a
percentage
of
the
hemoglobin
that
is
capable
of
transporting
oxygen.
Because
the
N-6000
uses
two
wavelengths,
it
measures
oxygenated
and
deoxygenated.
hemoglobin,
yielding
functional
SaO2.
It
does
not
detect
significant
amounts
of
dysfunctional
hemoglobin,
such
as
carboxyhemoglobin
or
methemoglobin.
In
contrast,
some
laboratory
instruments
such
as
the
Instrumentation
Laboratory
282
CO-
Oximeter
report
fractional
SaQ2-—~oxygenated
hemoglobin
expressed
as
a
percentage
of
all
measured
hemoglobin,
whether
or
not
that
hemoglobin
is
available
for
oxygen
transport.
Measured
dysfunctional
hemoglobins
are
included.
Consequently,
to
directly
compare
N-6000
measurements
with
those
of
another
instrument,
that
other
instrument
must
measure
functional
SaO2.
If
it
measures
fractional
Sa02,
those
measurements
can
be
converted
using
the
following
equation:
fractional
saturation
functional
saturation
=
~
x
100
—
(%carboxyhemoglobin
+
%methemogiobin
2.2.4
Measured
versus
Calculated
Saturation
When
saturation
is
calculated
from
a
blood
gas
measurement
of
the
partial
pressure
of
arterial
oxygen
(PaO2),
the
calculated
value
may
differ
from
the
N-6000
SpO2
measurement.
This
is
because
the
calculated
saturation
may
not
have
been corrected
for
the
effects
of
variables
that
shift
the
relationship
between
PO2
and
saturation
(Figure
2-1):
temperature,
pH,
the
partial
pressure
of
carbon
dioxide
(PaCO2),
the
concentrations
of
2,3-DPG
and
fetal
hemoglobin.
2-2

100
—
t
pH
„==
=.
+
Temperature
.”
est
+
PCO2
>
n°
+
2,3-DPG
“e
„>“
¿e
>.
ㆍ
΄
+
=
+ e
ο
7 è
я
; .
+
pH
5
504
+
e
+
Temperature
©
4
+
+
PCO2
.
の
e
e
+
2,3-DPG
a
ifs
4 +
+
+
+
+
nf.
+
+
7
の
VA
0
|
50
100
PO:
(mmHg)
Figure
2-1:
Oxyhemoglobin
Dissociation
Curve
2.3
CAPNOGRAPHY
SUBSYSTEM
The
N-6000
uses
nondispersive
infrared
spectroscopy
to
quantitatively
measure
the
amount
of
CO2
present
at
the
end
of
exhalation
(ETCO2).
It
features
a
small
lightweight
“mainstream”
CO2
sensor
that
attaches
to
a
disposable
airway
adapter.
The
airway
adapter
is
inserted
into
the
ventilator
circuit
either
between
the
endotracheal
tube
and
the
ventilator
circuit
or
between
the
elbow
and
the
patient
wye.
The
CO2
mainstream
sensor
fits
on
top
of
the
adapter
and
does
not
come
into
contact
with
the
respired
gases.
Infrared
spectroscopy
can
be
used
to
measure
the
concentration
of
any
molecule
that
absorbs
infrared
light.
Of
the
normally
respired
gases,
only
CO2,
N20,
and
water vapor
selectively
absorb
specific
wavelengths
of
infrared
light
(i.e.,
have
an
infrared
spectrum).
This
absorption
pattern,
the
infrared
spectrum
of
a
molecule
(usually
displayed
as
a
graph
of
light
absorbed
by
a
molecule
versus
the
wavelength
of
light),
is
unique
to.
that
molecule.
However,
different
molecules
may
have
similar
spectra
that
have
overlapping
absorption
peaks.
Because
the
absorption
of
light
is
proportional
to
the
concentration
of
the
absorbing
molecule,
an
unknown
concentration
can
be
determined
by
comparing
the
absorbance
to
that
of
a
known
standard.
The
CO2
detection
mechanism
used
in
the
N-6000
(commonly
referred
to
as
the
CO2
sensor
or
“optical
bench”),
has
an
infrared
source
that
is
optically
filtered
to
provide
a
narrow
band
of
wavelengths
corresponding
to
an
absorption
peak
of
the
CO2
spectrum
(see
Figure
2-2).
During
monitoring,
light
first
passes
through
the
respiratory
gas
in
the
airway
adapter.
Next
it
reaches
the
narrow-band
infrared
filter,
which
was
selected
because
it
passes
wavelengths
that
are
selectively
absorbed
by
0602.
It
then
encounters
the
chopper
wheel,
which
rotates
many
times
each
second.
Three
measurements
are
obtained
during
each
rotation:
ο
A
sample
measurement
is
made when
the
light
passes
through
the
open
area
of
the
chopper
wheel
and
then
reaches
the
detector
(i.e.,
light
passes
through
the
respiratory
gas).

.
A
reference+sample
measurement
is
obtained
when
light
passes
through
a
reference
gas
cell
containing
a
known
CO2
concentration
and
then
reaches
the
detector
(i.e.,
light
passes
through
respiratory
gas
and
gas
in
the
reference
cell).
ㆍ
A
dark
measurement
is
made when
light
strikes
a
solid
area
of
the
wheel
(1.9.,
no
light
reaches
the
detector).
The
ratio,
“reference+sample/sample”,
is
then
used
in
the
sensor-specific
calibration
equation
to
determine
the
CO2
concentration
in
respiratory
gas.
Respiratory
rate
is
determined
by
the
N-6000
from
the
analysis
of
the
CO2
waveform
(capnograph).
The
N-6000
measures
the
partial
airway
pressure
of
CO2
in
the
patient
airway
adapter
at
normal
assumed
conditions
of
33°
C
and
fully
saturated.
Barometric
pressure
is
measured
directly
by
the
N-6000.
When
units
of
%
(by
volume)
are
displayed
on
the
front
panel
LEDs,
the
measured
value
is
expressed
as
%
dry
gas.
By
convention,
all
readings
of
CO2
posted
on
the
front-panel
LED
display
are
assumed
to
be
measured
from
a
patient
and
corrected
to
body
temperature
(37°
C).
When
PCO2
is
being
displayed
on
the
front
panel
LED
in
units
of
mmHg
or
kPa,
the
values
are
also
converted
to
fully
saturated
conditions
(BTPS
)
before
being
displayed.
The
N-6000
detects
and
counts
breaths
when
the
COz2
level
crosses
a
threshold,
which
is
set
dynamically
from
maximum
and
minimum
CO2
values.
Hence,
when
a
ventilated
patient
has
several
spontaneous
breaths
with
low
ETCO2
values
followed
by
a
mechanical
breath
with
a
higher
ETCO?
value,
the
N-6000
accurately
detects
and
counts
all
breaths.
The
value
of
ETCO2
that best
estimates
the
true
alveolar
CO2
value
is
the
maximum
value
obtainable
from
the
patient
during
forced
exhalations.
Unlike
conventional
capnometers
that
simply
average
all
breaths
together
to-display
the
ETCO2
value,
the
N-6000
looks
for
the
maximum
value
of
ETCO2
seen
within
the
last
8
seconds
and
displays
that
maximum
value
for
the
true
ETCO2.
This
results
in
a
stable
ETCO2
value
that
best
approximates
the
arterial
PaCO
value
in
the
patient.
Reference
gas
cell
Open
area
!
Light
Airway
Infrared
Chopper
Photodetector
source
adapter
filter
wheel
Figure
2-2:
Nondispersive
Infrared
CO2
Analyzer
2-4

24
N20/O2
COMPENSATION
Unless
compensated
for,
elevated
levels
of
oxygen
(O2)
and
nitrous
oxide
(N20)
in
the
airway
affect
the
measurement
of
the
concentration
of
CO2
by
a
collision
broadening
effect.
To
ensure
accurate
measurements,
the
N-6000
provides
two
levels
of
compensation:
(1)
nitrous
oxide
with
oxygen; and
(2)
high
oxygen.
Compensation
levels
are
user
selectable
via
display
screen
menu
selections.
The
N-6000
is
calibrated
for
low O2
compensation,
which
assumes
the
reference
state
of
0%
nitrous
oxide
and
20%
oxygen.
This
is
the
default
mode
where
COMP.
=
OFF.
In
the
N-6000,
an
increase
in
O2
beyond
the
reference
state
decreases
the
displayed
CO2
value
by
1.5%
of
reading
for
every
20%
of
increase
in
O2.
An
increase
in
N20
increases
the
displayed
CO2
value
by
1.6%
of
reading
for
every
20%
increase
in
N20.
The
first
compensation
level
assumes
50%
nitrous
oxide
and
50%
oxygen.
Selecting
COMP.
=
N20
provides
the
necessary
correction
factor
for
high
levels
of
N20.
When
this
correction
is
applied,
there
is
zero
error
under
these
assumed
conditions,
The
second
compensation
level
assumes
0%
nitrous
oxide
and
60%
oxygen. Selecting
COMP.
=
O2
provides
the
correction
factor
for
high
levels
of
O2.
Once
again,
when
this
correction
is
applied,
there
is
zero
error
under
these
assumed
conditions.
Finally,
barometric
pressure
changes
also
affect
measured
CO2
values.
However,
changes
in
barometric
pressure
require
no
user
intervention;
the
N-6000
automatically
adjusts
for
these
changes.
The
following
paragraphs
contain
further
in-depth
discussion
of
compensation
and
standard
conditions,
as
pertains
to
N-6000
operation.
2.4.1
Standard
Gas
Conditions
The
accuracy
specifications
for
the
N-6000
refer
to
the
following
standard
conditions:
test
gas
is
CO2
in
balance
air
(21%
O2),
at
airway
conditions
of
33
°C,
fully
saturated
(water
vapor
pressure
of
38
mmHg),
and
a
barometric
pressure
of
760
mmHg.
Additional
corrections
and/or
residual
errors
are
required
for
barometric
pressure
(altitude),
N20,
02,
and
water
vapor.
2.4.2
Pressure
Broadening
Compensation
Ambient
barometric
pressure
affects
measured
CO2
by
a
“pressure
broadening”
effect,
in
which
energy
exchanged
in
molecular
collisions
alters
the
absorption
spectrum
of
CO2.
In
the
N-6000,
barometric
pressure
is
automatically
measured
and
a
compensation
is
automatically
applied
to
CO2
values
reported
on
both
the
main
monitoring
screen
and
the
service screen.
However,
additional
residual
errors
may
be
observed
when
testing
reference
gases
at
pressures
different
from
the
standard
reference
state
of
760
mmHg
(sea
level).
These
amount
to
approximately
+0.5
mmHg
for
every
5000-foot
change
in
altitude.
2-5

2.4.3
N20
Collision
Broadening
Nitrous
Oxide
(N2O)
can
affect
the
CO2
measurement
by
both
direct
absorption
of
infrared
and
by
a
“collision
broadening”
effect
in
which
energy
exchanged
in
molecular
collisions
alters
the
absorption
spectrum
of
CO2.
The
infrared
narrow
bandpass
filter
used
in
the
N-6000
is
chosen
to
eliminate
any
direct
absorption
of
infrared
energy
by
N20.
However,
N20
collision
broadening
causes
an
increase
in
measured
CO2
of
approximately
+0.8%
per
10%
increase
in
N20
from
standard
conditions.
A
software-selected
(user-selectable)
N20
compensation
option
is
provided
to
correct
for
high
N20,
assuming
a
gas
composition
of
50%
N20
and
50%
O2.
CO2
values
reported
on
both
the
N-6000
main
monitoring
screen
and
on
the
service
screen
are
compensated
for
this
gas
mixture
if
the
02
compensation
option
is
selected.
2.4.4
O2
Collision
Broadening
Oxygen
(02)
can
affect
the
CO2
measurement
by
a
“collision
broadening”
effect
in
which
energy
exchanged
in
molecular
collisions
alters
the
absorption
spectrum
of
CO2.
O2
collision
broadening
causes
a
decrease
in
measured
CO2
of
approximately
-0.75%
per
10%
increase
in
O2
from
standard
conditions.
A
software-selected
(user-selectable)
02
compensation
option
is
provided
to
correct
for
high
O2,
assuming
a
gas
composition
of
60%
O2
(0%
N20).
CO2
values
reported
on
both
the
N-6000
main
monitoring
sereen
and
the
service
sereen
are
compensated
for
this
gas
mixture
if
the
O2
compensation
option
is
selected.
2.4.5
Water
Vapor
Effect
Water
vapor
also
has
an
effect
on
CO2
measurements.
In
normal
use,
it is
assumed
that
respiratory
gas
is
at
standard
airway
conditions
of
33°C,
fully
saturated.
Under
these
conditions,
water
vapor
(with
a
vapor
pressure
of
38
mmHg)
causes
an
increase
in
measured
CO2
of
6%.
The
N-6000
is
calibrated
to
include
this
effect,
and
CO2
values
reported
on
the
main
monitoring
screen
are
automatically
compensated,
while those
on
the
service
screen
are
not.
If
applying
a
dry
test
gas
to
verify
calibration,
CO2
values
reported
on
the
service
screen
should
be
used
for
comparison.
2.4.6
BTPS/ATPS
Compensation
The
N-6000
assumes
that
measured
respiratory
gases
are
at
standard
airway
conditions
of
airway
temperature
and
pressure
fully
saturated
(ATPS).
That
is,
it
is
assumed
that
the
gas
in
the
airway
adapter
is
measured
at
33
°C,
fully
saturated
with
a
water
vapor
pressure
of
38
mmHg,
at
ambient
barometric
pressure.
By
convention,
CO2
values
are
reported
on
the
main
monitoring
screen
after
conversion
to
deep
lung conditions
of
body
temperature
and
pressure
fully
saturated
(BTPS):
37
°C,
fully
saturated
with
a
water vapor
pressure
of
47
mmHg,
at
ambient
barometric
pressure.
Service
screen values
do
not
have
these
corrections
applied.
2.4.7
Summary:
Reported
CO2
Values
on
the
Main
Monitoring
and
Service
Screens
CO2
values
reported
on
the
service
screen
are
uncorrected
for
BTPS/ATPS
conditions
(including
water
vapor
correction).
That
is,
CO2
values
reported
on
the
service
screen
are
accurate
assuming
a
dry
test
gas
is
present
in
the
airway
adapter
(at
25
°C).
Corrections
for
N20
and
Oz,
if
selected,
are
also
applied
to
the
service
screen
values.
CO2
values
reported
on
the
main
monitoring
screen
have
all
corrections
applied:
BTPS/ATPS,
water
vapor,
and,
if
selected,
N20
and
O2.

2.5
FACTORY
CALIBRATED
SENSOR
Each
CO2
sensor
is
individually
factory-calibrated
over
multiple
gas
concentrations
and
multiple
temperatures.
The
N-6000
sensor
is
temperature-regulated
to
42
°C,
which
keeps
the
sensor
optical
components
and
electronics
at
a
known
constant
temperature.
Additionally,
the
sensor
is
factory-
calibrated
over
multiple
temperatures,
so
that
any
deviation
from
the
42
°C
set-point
is
automatically
compensated
by
temperature
calibration
coefficients
stored
with each
sensor
unit.

Table of contents
Other Nellcor Measuring Instrument manuals