Neomedic P1000 User manual

POWERPRESS UNIT
User Manual
Pneumatic Sequential Compressor
Neomedic
9601 Owens mouth Ave. #8
Chatsworth, CA 91311
Toll Free: 866-990-1168
Tel: 818-998-1023
Fax: 818-998-0277
www.air1000.com

Contents
1. Introduction ---------------------------------------------------------------------- 4
2. Clinical Application-------------------------------------------------------------- 5
3. Clinical Treatment Guide ------------------------------------------------------ 6
4. Garment type & Size------------------------------------------------------------ 7
5. Pump Description --------------------------------------------------------------- 8
6. Warranty ------------------------------------------------------------------------- 10
7. General Safety-------------------------------------------------------------------- 10
8. Caution---------------------------------------------------------------------------- 10
9. Trouble Shooting----------------------------------------------------------------- 11
10. Technical Description----------------------------------------------------------- 11
3

2. Clinical Application
Intermittent Compression Therapy
A thorough assessment of the patient should be made considering
the general state of the patient and the specic characteristics of their
presenting condition. The choice of system, i.e. garment and treatment
regime, is to be recommended by the medical practitioner for the patient.
Training and education of the patient in the use and understanding of
the system will help to ensure compliance.
Indications • Lymphedema - primary
- secondary;
• Oedema - dependent
- traumatic
- secondary to stroke;
• Chronic Venous Insufciency
Cautions 1. Proper application and connection to the pump is
essential.
2. Garments should be removed if the patient experiences
tingling, numbness or pain.
Contraindications
POWERPRESS unit should NOT be used in the following conditions:
1. Severe arteriosclerosis or other ischemic vasular diseases.
2. Known or suspected acute deep vein thrombosis.
3. Severe congestive cardiac failure.
4. Any local condition in which garments would interfere,
for example:
• Gangrene • Dermatitis
• Untreated, infected leg wounds
• Recent skin graft
5
1. Introduction
The Powerpress Unit system consists of a pneumatic pump which is light
and simple to use and a 4 chamber garment. It is suitable for home use.
The pump supplies air via connecting tubes to an inatable garment
allowing the application of controlled pressure to gently compress the limb.
This action assists in increasing the return of blood, excess uids, improves
venous stasis and encourages the reabsorption of waste products.
Compression Cycle:
36sec ination (9sec each chamber) / 24 sec deation
Chamber 1 (Foot) = 36sec
Chamber 2 (Ankle) = 27sec
Chamber 3 (Calf) = 18sec
Chamber 4 (Thigh) = 9sce
Variable pressure output ranges from 30-160 mmHg. The segments within
the garments are designed to prevent ridging and ensure high patient
comfort and compliance.
** We would recommend that this manual is read throughly before
proceeding with treatment. Please also refer to clinical guidelines and
contraindications.
4

3. Clinical Treatment Guide
An initial pressure of 40 mmHg (30 mmHg for management of lymphedema)
is suggested at the commencement of treatment. It may be necessary to
start at a lower level of pressure, dependent on the patient’s tolerance.
The pressure can be gradually increased over time, until the required
pressure is reached. The upper treatment pressure range is generally
60-70 mmHg.
A single treatment session is usually 20-60 minutes.
The above settings and timings are guidelines, and should not be used as
a substitute for clinical judgement and experience.
6
zip
Full Leg Sleeve
Open
Arm Sleeve
Open
zip
zip
X Large Full Leg Sleeve
Open
Circumference
Open
zip
Half Leg Sleeve
Open
zip
X Large Half Leg Sleeve
Circumference
Available Extenders size
Half Leg Sleeve:4”, 6”
Arm Sleeve: 4”, 6”
Full Leg Sleeve: 4”, 6”
4. Garments type & size
18”(L)
22”
19.5”(L)
18”
9”
20”
28.5”(L)
13”
26”
28”(L) 30”(L)
16”
12”
25”
18”(L) 19.5”(L)
9”
14”
28”(L) 30”(L)
22”
14”
30”
7

5. Pump Description
30
40
50 60 70 80 90 100
110
120
130
140
mmHg
PRESSURE
150
MIN MAX
20
PRESSURE
1
2
3
4
CHAMBER
1 2 3 4
side view
1 Main Pressure Controller Knob (Push to Lock, Pull to release)
2 Pressure Gauge (0 ~ 160mmHg)
3 Air Tubing Hose Connector
4 Power Switch
Operation: The pump should be placed securely on a at surface.
Before starting the pump ensure that the garments are properly applied,
the zippers are secured and the garment connecting tubes are attached
to the pump outlet ports.
8 9
How to SET UP Pressure?
• Model# P1000
1. Turn on Power Switch
2. Wait until deation time
During Deation Time; Pressure Gauge does not move
Air does not inate sleeves
Air in the sleeves comes out
Air does not come out from connection port
3. Turn Pressure Knob quickly and set to desired pressure
4. Connect Tubing Hose and wear sleeves.
During usage of Pump : Wait until deation time to set desired pressure
To Remove the Garment: Switch off the pump, disconnect the tubing
from the pump by removing the tube connectors.

6. Warranty
All Powerpress Unit shall conform to applicable specications in effect
from the invoice date to the end of the Warranty Period, and Power-
press Unit reserves the right to change such specications at any time
with or withour notice. The Warranty Period for Powerpress Unit prod-
ucts shall be as follows, calcuated from the date of invoice:
• The Powerpress Unit pump is guaranteed for 12 months
• The powerpress Unit garments are guaranteed for 6 months
7. Safety Warnings
• Electrical equipment may be hazardous if misused. The pump’s rear
case should only be removed by authorized technical personnel.
• Do not use the pump in the presence of ammable gases such as
anesthetic agents.
• Make sure that the main power cable is positioned to avoid causing
a trip or other hazard, and is clear of moving bed mechanisms or other
possible entrapment areas.
8. Caution
Electromagnetic compatibility (EMC). This product complies with the
requirements of applicable EMC Standards. The use of accessories not
specied by the manufacturer may result in increased emissions by, or
decreased immunity of, the equipment, affecting its performance.
10
9. Troubleshooting
If you should encounter a problem, please follow the fault nding guide
below, if the fault cannot be rectied, please refer to Service.
10. Technical Description
Model No: P1000
Supply Voltage: 120 V
Supply Frequency: 60 Hz
Electrical Rating: 14VA
Fuse Rating: F500mAH250V
Size: Length: 11.8” (300mm)
Width: 8.3” (210mm)
Height: 5.1” (130mm)
Weight: 4.5lbs (2kg)
Case Material: ABS Plastic
Mode of Operation: Continuous
11
Fault
Pump does not operate.
Pump operates but
garments will not
inate.
None of the above.
Check
Is power switch on?
Is power cord plugged in
correctly?
Blockage in garment
supply tube.
Garment not correctly
tted to pump.
Pressure control set too
low.
Air leak in garment.
Remedy
Check switch.
Check connections.
Ensure that the tube airway
is clear.
Check connections.
Increase pressure control.
Check garemt. Replace if
defective.
Hint - Reduced pressure
level indicates garment or
connector leak
None of the above. Call Service
Table of contents
Popular Personal Care Product manuals by other brands

Wellcare
Wellcare MHB10 instruction manual

Vitek
Vitek VT-2509Y Manual instruction

Otto Bock
Otto Bock 17PA1-WR Series Instructions for use

RITM OKB ZAO
RITM OKB ZAO RITMSCENAR Super Pro v.2 operating manual

FACE VITAL
FACE VITAL Ultrasonic Skin Scrubber user manual

Withings
Withings WAS01 Installation and operating instructions