Newall DPG2000 Instruction Manual

INSTALLATION/USER
MANUAL
DPG2000
DIGITAL READOUT
SYSTEM
Newall Measurement Systems

CONTENTS
1. INTRODUCTION
1.1 EMC and Low Voltage Compliance
1.2 Technical Specifications
2. INSTALLATION
2.1 Mounting
2.2 Power Supply
2.3 Transducer Connections
2.4 Switching On
3. SET-UP
3.1 Sleep/Datahold
3.2 Input Type
3.3 Direction
3.4 Radius/Diameter
3.5 Linear Error Compensation
3.5.1 Calculating Linear Error Compensation
4. USER INSTRUCTIONS
4.1 Using the Keypad
5. OPERATING FUNCTIONS
5.1 Using Incremental
5.2 Using Absolute
5.3 The Digifind Function
5.4 Inch/Millimeter Conversions
5.5 Center Find
5.6 Datahold Function
5.7 Selecting An Axis Resolution
6. TROUBLESHOOTING
7. CLEANING
Contents
Newall Measurement Systems
1

Introduction
Newall Measurement Systems 2
1.0 INTRODUCTION
1.1 EMC and Low Voltage Compliance
The DPG2000 Digital Readout conforms to the relevant European standards for electromagnetic compatibility
and low voltage directive as detailed below.
BS EN 50081-2: Electromagnetic compatibility.
Generic Emission Standard - Industrial Environment
BS EN 50082-2: Electromagnetic compatibility.
Generic Immunity Standard - Industrial Environment.
BS EN 61010-1: Safety requirements for electrical equipment for
measurement, control and laboratory use.
1.2 Technical Specifications
Dimensions:
Height: 140mm (5.5in)
Width: 240mm (9.5in)
Depth: 80mm (3.2in)
Weight: 2.5kg (5.5lbs)
Operating Voltage: 115 or 230V (switch selection)
Supply Voltage Fluctuation: Not to exceed +/-15% of the operating voltage
Supply Frequency: 50 to 60 Hz
Maximum Power Consumption: 18VA
Operating Temperature: 0 to 45°C
Storage Temperature: -20 to 70°C
Transducer Inputs: Dependant on model, one, two or three Spherosyn/Microsyn transducers
Selectable Resolutions:
Spherosyn/Microsyn 10 5μm (0.0002in) / 10μm (0.0005in) / 20μm (0.001in) / 50μm (0.002in)
Microsyn 5 1μm (0.00005in) / 2μm(0.0001in) / 5μm (0.0002in) / 10μm (0.0005in)
Environmental Conditions: Indoor Use, IP20 (IEC 529)
Relative humidity - maximum 80% for temperatures up to 31°C decreasing
linearly to 33% at 45°C
Transient overvoltage according to INSTALLATION CATEGORY II of IEC664
POLLUTION DEGREE 2 in accordance with IEC664
Newall Measurement Systems reserves the right to change specifications without notice.
Certificate No FM36096
ISO 9001

2.2 Power Supply
BEFORE CONNECTING THE ELECTRICAL SUPPLY TO THE CABINET, CHECK THAT THE VOLTAGE
SELECTOR SWITCH IS CORRECTLY SET. Figure 1.2 shows the position of this switch.
The mains supply is connected through a detachable supply cord. The cabinet is supplied with a cord fitted
with a right-angled connector. If another supply cord is to be used, it must have a IEC320, 10A, EARTHED
mains connector with a cord rated for at least 10A.
2.0 INSTALLATION
2.1 Mounting
The DPG2000 is supplied with a fixing kit consisting of a M10 stud, nut and washers. Figure 1.1 shows the
use of this kit in conjunction with an optional mounting arm (Part No: 294-35670).
Be sure that the mounting arrangements are secure as the operator will need to apply pressure to the front
panel when using the keypad.
An optional mounting assembly (Part no: 294-37740) is available which allows for tilt and rotation of the
cabinet.
Select the location of the DPG2000 with due regard to safety and ease of operation. Keep clear of moving
parts and coolant spray. Ensure that the natural ventilation around the cabinet is not restricted.
To ensure correct operation of the DPG2000, it is recommended that the case is grounded to the machine.
Use a wire or strap of a least 1.5mm2(16 AWG) from the cabinet equipotential terminal, (Figure 1.2), to a
suitable point on the machine body. The wire should be as short as possible. The machine must also be
properly grounded to a good earth point.
Installation
Newall Measurement Systems
3
Mounting Arm
Figure 1.1 Standard Fixing with Optional Mounting Arm

Installation
Newall Measurement Systems 4
Figure 1.2 Connection Diagram
The PROTECTIVE EARTH CIRCUIT of the mains supply MUST BE CONNECTED to the protective earth
terminal of the cabinet through the supply cord.
The supply cord should be secured to the mounting arm or pillar with cable ties to ensure that it cannot drop
into a hazardous position when disconnected from the cabinet.
The power supply cord must be routed away from moving parts, swarf, coolant or sources of heat.
Input connectors 1,2 or 3
according to model
Cabinet equipotential terminal
for grounding to machine
Mains supply fuse
Not to be replaced by the operator
Disconnect mains supply before replacing
IEC inlet (mains supply)
Mains voltage selector switch
Mains supply switch

Installation
Newall Measurement Systems
5
If a mains plug is not already fitted to the supply cord, or is of the wrong type, then a suitable EARTHED plug
should be used which complies with the relevant specifications for plugs and socket-outlets.
The mains supply fuse is a 20x5mm, type T0.5A, 250V. It is not to be replaced by the operator. If the fuse
blows it is an indication of some significant problem with the power source. Check the supply and wiring
carefully. Before replacing the fuse, the cabinet must first be disconnected from the supply by the removal of
the IEC socket from the inlet. Do not position the equipment so that it is difficult to disconnect the mains
supply.
NOTE: If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
2.3 Transducer Connections
Figure 1.2 shows the connection sockets at the back of the DPG2000. The DPG2000 is designed for use
with Newall’s Spherosyn and Microsyn transducers only. The transducers are connected to the DPG2000
with Bleecon type connectors. These connectors have a sliding sleeve that locks the connectors into their
sockets.
Switch off the DPG2000 before connecting, or disconnecting the transducers. To fit the connectors into the
appropriate socket at the back of the cabinet, first align the connector and then push firmly in place. You
should hear a click confirming that the locking sleeve has engaged. To remove the connector, pull back on
the connector sleeve to disengage the locking mechanism.
The transducers and digital readouts are connected at a separated extra low voltage (SELV) level. Any
additional interconnections must also be at SELV level.
2.4 Switching On
The mains supply switch for the DPG2000 is mounted on the back of the cabinet as shown in Figure 1.2.
When you switch on the DPG2000 the cabinet will automatically go through a brief self diagnostic routine.
During this routine, the name DPG2000 will be shown, then the software version number will be displayed
and finally, all segments of the displays will be lit.
After this routine, the DPG2000 is ready for use.
CONVENTIONS USED IN THIS MANUAL
The direction of travel of an axis refers to the travel of the tool relative to the
workpiece. Keys on the keypad are signified in bold print,
such as [ ent ] for the enter key.

Set-Up
Newall Measurement Systems 6
3.0 SET-UP
The Set-Up procedure allows you to change the main settings for the DPG2000. For normal use, you will
find that you only need to perform the Set-Up procedure once, and it is possible that the factory defaults are
suitable for your needs without change. (See Table 1)
The Set-Up procedure can only be activated just after power
is switched on to the cabinet. After switching on, press the
concealed key which is located under the "ne" of the newall
logo on the keypad (See Figure 1.3). The key must be
pressed before the end of the initial self diagnostic
routine.
When you have entered the Set-Up routine, the letters
‘SET-UP’ appear in the display.
The Set-Up procedure makes use of a menu system. The
main menu consists of a list of options that can be cus-
tomised for your use (See Table 2). You simply press the
[ ∧∧] [ ∨∨
]keys to scroll through this list until you reach the
option you wish to change. To change the option when
selected, press the [ Xo ] [ Yo ] or [ Zo ] key. On the single
axis version, the options can be changed by pressing the
blue [ 0 ] key.
To exit the Set-Up routine press the concealed key ("ne" of newall).
Default Setting DPG2000
Sleep On
Input Type Spherosyn
Resolution 5μm
Direction 1
Radius/Diameter Rad
Linear Compensation 0
Table 1 - Default Settings
CONCEALED KEY
FOR ENTERING
SET-UP ROUTINE
Figure 1.3 DPG2000 keypad

Set-Up
Newall Measurement Systems
7
LINEAR ERROR COMPENSATION
USER DEFINABLE
SLEEP
ON
OFF
Table 2 - Set-Up Menu Options
INPUT TYPE
SPHEROSYN
MICROSYN 5
MICROSYN 10
READINGS
RADIUS
DIAMETER
DIRECTION
0
1
3.1 Sleep/Datahold
In the Set-Up routine, press the [ ∧] key until "SLP ON" or "SLP OFF" is displayed. Toggle between
"SLP ON" or "SLP OFF" by pressing the [ Yo ] key.
Select "SLP ON" if you want to be able to use the Datahold facility. (See Section 5.6 for a description of
Datahold). Otherwise select "SLP OFF".
When the selection is made press [ ∧] to continue to the next option.
3.2 Input Type
Select between Spherosyn, Microsyn 5 or Microsyn 10 for each axis.
Pressing [ Xo ] [ Yo ] or [ Zo ] moves between transducer types for that axis. Pressing [ ent ] defines
that axis to be the selected type.
Warning: The axis selected for the input type must match the actual transducer (Spherosyn
or Microsyn) which is plugged into the corresponding axis. Erroneous readings
will occur if this warning is not followed.
When the selection is made press [ ∧] to continue to the next option.

Set-Up
Newall Measurement Systems 8
3.3 Direction
The direction option allows you to change the direction of travel of each axis. For example, if after installation
the X axis is measuring positive from right to left, you can use this option to change the direction of the X
axis so it measures positive from left to right.
In the Set-Up routine, press the [ ∧] until "DIR" (direction) is displayed.
Each axis display will show "0" or "1"
Pressing each axis key [ Xo ] [ Yo ] or [ Zo ] reverses the counting direction. When the unit is displaying
the choices you wish to use, press [ ∧] to continue to next option.
3.4 Radius/Diameter
This option allows you to select any axis to display measurements at a two times (x2) factor. This is used on
lathes and other turning applications to display part diameter rather than part radius.
In the Set-Up routine, press the [ ∧] key until "RAD" or "DIA" is displayed.
Pressing each axis key [ Xo ], [ Yo ] or [ Zo ], switches between radius and diameter readings for that axis.
When the unit is displaying the choices you wish to use, press [ ∧]to continue to the next option.
3.5 Linear Error Compensation
This option allows you to apply a constant correction factor to all measurements displayed. This factor is
expressed in parts per million (PPM).
In the Set-Up routine, press the [ ∧] key until LC (Linear Compensation) is displayed. The display will show
"LC 0" or the last correction factor entered.
To insert, or change, a correction factor, select the axis [ Xo ] [ Yo ] or [ Zo ] and enter the factor you wish to
use.
The value is entered by pressing the [ ∧] and [ ∨] keys.
For example, to apply a factor of 200PPM to the X axis, press the following keys: [ X ] [ ∧] until the displays
show 200 then press [ ent ].
If you make a mistake you can revert to the previous settings by pressing the [ ce ] key. This cancel
function will not work after the [ ent ] key has been pressed.
If the unit is displaying measurements less than the actual measurement, enter a positive factor to
compensate. A factor of 200PPM means displays are measurement x 1.000200. When each axis is
displaying the correction factor you wish to use, press [ ∧]to return to the first option.
3.5.1 Calculating Linear Error Compensation
To establish a correction factor, check the measurements displayed by the DPG2000 against a known
distance. The correction factor should be established while in the standard operating mode and not in the
Set-Up routine.

Set-Up/User Instructions
Newall Measurement Systems
9
Key Purpose
Set the current position for the axis to zero (reset)
Clear entry
Enter key to confirm data entry
Switches between absolute and incremental readings
(LED indicates choice of mode)
Centre Find function
Select the Digifind function
Setting Axes Resolutions
Switches between inch and millimetre display
(LED indicates choice of mode)
Scroll Up
Scroll Down
4.0 USER INSTRUCTIONS
4.1 Using the keypad
Figure 1.3 shows the layout of the keypad. The keys are used as follows:
For example, you might use a known (actual) distance of 500mm, against which the DPG2000 displays
499.8mm. The correction factor you would then apply is:
(0.2mm/500mm) x 1,000,000 = 400PPM
If the DPG2000 displays 500.2mm over the same distance, the correction factor would be:
(-0.2mm/500mm) x 1,000,000 = -400PPM
Warning: Once you have entered a multiplication factor for an axis, all measurements will be
adjusted accordingly. If you wish to disable this adjustment, you will have to enter a
compensation factor of zero.
You can re-edit any Set-Up function by locating it using the [ ∧] and [ ∨] direction keys.
To leave the Set-Up routine press the concealed key, which is located under "ne" of the newall logo.
XoYoZo
abs
inc
ce
12
ent
ref0
in
mm
∧
∨
res

abs
inc
Xo
Yo
Zo
abs
inc
Operating Functions
Newall Measurement Systems 10
5.2 Using Absolute
(a) Purpose and use
When the DPG2000 is set to absolute mode it will display the
position relative to an established datum point.
(b) Keystrokes
Operator Steps Keystrokes Axes Display
Set to incremental mode
Reset axes
Operator steps Keystrokes Axes Display
Set absolute mode
30 100
140
190
5.0 OPERATING FUNCTIONS
5.1 Using Incremental
(a) Purpose and use
When the DPG2000 is set to incremental mode, it can be
used to display each new position relative to the last position.
This is also known as point-to-point operation.
On setting to incremental, you can reset each axis by
pressing [ Xo ] [ Yo ] or [ Zo ].
Each time you switch to incremental mode, the DPG2000 will display the position relative to the last reset
position while in the incremental mode.
(b) Keystrokes
30 70 40 50
0.000
0.000
0.000

Xo
Yo
Zo
abs
inc
Operating Functions
Newall Measurement Systems
11
(c) Establishing the datum
When you reset the display in absolute mode, you are setting the current position of your machine as your
datum point. All absolute positions will be measured relative to this datum.
To set the datum position, move the machine to the point you intend to establish as the datum and then
reset any, or all, axes while in the absolute mode.
5.3 Digifind
Digifind is a reference function used to find the datum should it be lost due to power failure or an accidental
key stroke.
(a) Using Digifind to re-establish a lost datum
In the event that the datum positions are lost, align each axis to within ±6mm (±0.25”) for Spherosyn or
±2.5mm(±0.098”) for Microsyn to the original datum point.
By pressing the [ refo] followed by the corresponding axis key, the display will update to a distance equal to
the distance from the current position to the datum point.
Each time the DPG2000 is powered up the cabinet will automatically use Digifind to compensate for any axis
movement of up to ±6mm (±0.25”) for Spherosyn and ±2.5mm(±0.098”) for Microsyn.
5.4 Inch/Millimetre
To change between inch and millimetre readings, press [ in/mm ]. The displayed data will be converted
instantly. The LED beside the key reminds you which mode is being used.
When you switch on the DPG2000, it will display the same unit of measurement which was set prior to
power loss.
Operator steps Keystrokes Axes Display
Set to absolute mode
Move machine to the
datum location
Reset axes 0.000
0.000
0.000

Xo
Xo
12
Operating Functions
Newall Measurement Systems 12
5.5 Centre Find
Centre Find halves the dimension displayed for any or all axes selected. You can use Centre Find in either
absolute or incremental mode. The keystrokes are the same in either case.
In the following example, Centre Find is being used on the X axis to find the centre point of a workpiece that
is 100mm wide.
In either absolute mode or incremental mode, once you have used Centre Find you can locate the centre
point by moving until the display is at zero.
If you are in absolute mode, remember that using Centre Find will set the datum to the centre point.
5.6 Datahold (Sleep)
Datahold allows you to disable the DPG2000 but retain power to the measurement transducers and the
memory circuits. You can use Datahold to prevent unauthorised or accidental use of the DPG2000 whilst
unattended.
To select Datahold, press the concealed key located under the "ne" of the "newall" logo (See Figure 1.3).
While the DPG2000 has been set to Datahold, the keypad will not function and the displays will be blank.
If the machine axis is moved, "DISPLACD" (displaced) will appear in the display window. The DPG2000 will
record all transducer movement and will update the display once Datahold is cancelled. If a key is pressed,
"TOUCHED" (touched) will appear in the display to alert the operator that data entry has been attempted.
To cancel datahold, press the concealed key located under the "ne" of the "newall" logo.
5.7 Selecting An Axis Resolution
The resolution to which an axis position is displayed, can be selected by pressing the [ res ] key. This will
display the current selection for all axes.
To alter an axis resolution select that axis by pressing either [ Xo ] [ Yo ] or [ Zo ]. The resolution can then
be changed by pressing the [ ∧∧] and [ ∨∨] keys. Once the desired resolution is displayed press [ ent ].
If you make a mistake you can revert to the previous settings by pressing the [ ce ] key.
Operator steps Keystrokes Axes Display
Locate to your first position
(One edge of your workpiece)
and zero the axis
Locate to the second position
(The other edge of the workpiece)
Use Centre Find to locate the
centre point
0.000
100.000
50.000

Xo
ent
∧∨
res
Operating Functions
Newall Measurement Systems
13
The available resolution options depend on the transducer type selected in Set-Up. See Sections 1.2 and
3.2
You can use the resolution function in either inch or millimetre mode. The keystrokes are the same in either
case. The resolutions will be displayed in units corresponding to the selected mode.
Operator steps Keystrokes Axes Display
To enter resolution mode
To select the axis
Scroll through resolution options
To confirm your selection
0.005
0.005
0.005
0.005

Troubleshooting/Cleaning
Newall Measurement Systems 14
6.0 TROUBLESHOOTING
Symptom Solutions
1 Nothing happens when the Check that the DPG2000 is correctly connected to a working power
unit is switched on. source.
Check the power lead is not damaged.
Check the selector switch on the DPG2000 is set to accept the
correct power supply voltage.
Check the fuse. Note that if the fuse has blown, this suggests a fault
with the power source which must be corrected before the fuse is
replaced. (See Section 2.2)
2 When the unit is switched on This suggests that voltage of the power is to low.
the displays are frozen. Check that the power source is within the limits accepted by the unit.
(See Section 2.2)
3 The displays work, but reset This suggests either that the voltage of the power source is too low,
from time to time without the or that the power source has an intermittent fault.
keys being pressed. Check the power source as above.
Check that all connections are sound.
4 The displays work, but give This suggests that the there is a poor earth (ground) connection.
erratic readings, the last digit Both the DPG2000 itself, and the machine on which it is installed,
jitters or the measurements jump must have proper earth (ground) connections. (See Section 2.1)
to new Figures unexpectedly.
5"NO SIG" or “SIG FAIL” appears in This means that the DPG2000 is not receiving a proper signal from
the display. the measurement transducer.
Check that the transducer connection is good.
Check that there is no damage to the connectors or to the transducer
lead. If only one axis is displaying this message, turn off the DPG2000
and connect the transducer from a working axis into the faulty axis.
Turn the DPG2000 back on, if the same message appears, the fault is
likely to be in the DPG2000 and you should contact your local dealer.
7.0 CLEANING
Disconnect the cabinet from the power supply before cleaning.
It is recommended that the cabinet be wiped over with a lint-free cloth with a non corrosive/abrasive cleaning fluid.
Do not use compressed air.

NEWALL MEASUREMENT SYSTEMS LTD
Technology Gateway .Cornwall Road
South Wigston .Leicester .LE18 4XH
Tel: (0116) 264 2730 . Fax: (0116) 264 2731
Email: [email protected]
NEWALL FRANCE SARL
63 Rue Victor Hugo .F-59200 .Tourcoing .France
Tél. 03 20 01 03 13 .Fax 03 20 26 13 41
NEWALL ELECTRONICS INC
1778 Dividend Drive . Columbus .Ohio .43228 .USA
Tel: (1) 614 - 771 0213 .Fax: (1) 614 - 771 0219
Website: www.newall.co.uk
Code: 023-12610 UK Issue Date: March 2001
Table of contents
Popular Keypad manuals by other brands

Yale
Yale 22131 quick start guide

SEW-Eurodrive
SEW-Eurodrive MOVITRAC B operating instructions

Lutron Electronics
Lutron Electronics seeTouch QS Application note

Bentel Security
Bentel Security ALISON/S manual

DMP Electronics
DMP Electronics Thinline Aqualite 7000 Series Installation and programming guide

Verkada
Verkada BK11 install guide