manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Newport
  6. •
  7. Measuring Instrument
  8. •
  9. Newport INFCP User manual

Newport INFCP User manual

INFCP and INFCP-xxxB
Process Meter
Using This Quick Start Manual
Use this Quick Start Manual to set up your Process Meter and begin
operation. Information is provided on how to:
Connect ac power Set basic options for operation
Connect the sensor Scale the meter.
Features with are for the “B” version which has three-color
programmable “Big” LED display - All segment characters shown are
for the “B” version.
IMPORTANT: For complete information on all setup options,
please refer to the Operator’s Manual.
Mount the Unit
1. Cut a panel opening using the
dimensions shown to the right.
2. Position the unit in the opening,
making sure the front bezel is
flush with the panel.
3. Install retaining clips on both
sides of the meter and tighten
against the panel.
Wiring
1. Remove the panel at the back of the unit.
2. Locate the TB1 connector.
3. Insert the correct wire in each terminal as shown in the following
figure and tighten the lockdown screws.
4. Tug gently on the wires to verify the connections.
Connect the Sensor
1. Locate the TB2 connector on the rear of the unit.
2. Attach the sensor wires and tighten the lockdown screws.
The diagram below shows the wiring for 4–20 mA sensors
with internal excitation.
3. Tug gently on the wires to verify the connections.
4. Replace the panel at the back of the unit.
NOTE: Refer to the Operator's Manual for setup requirements
for other sensor types.
Current Input Connections (4–20 mA)
with Intern l Excit tion
Using the Configuration Menu
To configure the meter, you use the buttons on the front panel.
To Take This Action
Display the Press the MENU button. The first function
Configuration Menu on the menu, INPT, displays.
Select a submenu 1. Press MENU until the function you
function want is shown.
2. Press 䊳/TARE.
The information you can change flashes.
Select a value 1. Press 䊱/MAX to display the option
for that submenu you want.
function 2. Press MENU to store it.
STRD quickly flashes, indicating that
the selection has been stored in memory.
Then the next menu function displays.
Go back to previous Press RESET once.
menu function
Exit the Press RESET twice. The unit displays
Configuration RST as it reinitializes. When a numeric
Menu value displays, the unit is in run mode.
(Optionally, you can press MENU to
move through all the menu functions
until the unit reinitializes.)
NOTE: This Quick St rt M nu l includes specific configur tion
p r meters for tr nsducers with n output r nge of 4–20 mA nd
24 V excit tion. Other sensor types m y require different
p r meters or ddition l ones. When this is the c se, we refer you
to the Oper tor's M nu l for det iled instructions.
234
W rning: Do not connect AC power to your device until you
h ve completed ll input nd output connections. This
device must only be inst lled by speci lly tr ined
electrici n with corresponding qu lific tions. F ilure to
follow ll instructions nd w rnings m y result in injury!
AC Powered Unit Connections
DC Powered Unit Connections
To Set the Input Type
1. Press MENU until the unit displays:
2. Press 䊳/TARE. The unit displays:
3. For this application you want 0-20. If 0-20 is not
displayed, press 䊱/MAX until it appears. Other choices
are 100M, 50M, 10V and 5V.
NOTE: Refer to the Oper tor's M nu l for more inform tion
on ch nging r nges.
4. Press MENU to select the sensor shown. The meter
displays the next menu item. If you changed input type,
the meter displays:
To Set the Decimal Point
1. If it's not already shown, press MENU
until the unit displays:
2. Press 䊳/TARE. The unit displays:
3. Press 䊱/MAX to move the decimal
point to the desired location. The
choices are FFFF, F.FFF, FF.FF,
and FFF.F.
4. Press MENU to select the decimal point position shown.
The unit displays:
MENU SUBMENU 䊳/TARE DESCRIPTION
INPT 100M,±50M ,!1 0V ,!±5V ,0-20*Input
DEC.P FFFF*,F.FFF,FF.FF,FFF.F Decimal Point
RD.S.O IN!1, RD!1 ,IN!2 ,RD!2 Scale and Offset
RD.CF R.1=T*,R.1=N
Reading Configuration
R.2=0, R.2=1 , R.2=2, R.2=3, R.2=4
*
R.3=F*,R.3=U
COLR GRN,RED , AMBR Display Color
S1.CF S.1=A*,S.1=B
Setpoint 1 Configuration
S.2=U*,S.2=L
S2.CF S.1=A*,S.1=B
Setpoint 2 Configuration
S.2=U*,S.2=L
S1.DB 0003*
Setpoint 1, Deadband
S2.DB 0003*
Setpoint 2, Deadband
OT.CF O.1=E *, 0.1=D Analog Output
O.2=C *, 0.2=V
Configuration
O.3=A *, 0.3=P
O.4=D, 0.4=R
O.5=F, 0.5=H
P.BND 0000 shown if 0.3 = P Proportional Band
M.RST 0000 shown if 0.3 = P Manual Reset
OT.S.O RD!1,OUT1,RD!2 ,OUT2
Output Scale & Offset
L .CF RS=E *,RS=D
Lockout Configuration
SP=E *,SP=D
L3=0*,L3=1
BRIT M.BrT,L.BrT ,H.BrT
Display Brightness
External Fuse Required:
Time-delay, UL 248-14 listed Time-lag, IEC 127-3 recognized
175 mA (115 Vac line) 125 mA (115 Vac line)
80 mA (230 Vac line) 63 mA (230 Vac line)
10-32 Vdc, 7.5 W
DEVICE
* Factory Default Settings
When using DC power, do not use internal excitation or
Isolated Analog Output for high color brightness. For
low or medium brightness, internal excitation is limited
to 24 V @ 25 mA, 5 V, 10 V, 12 V @ 35 mA.
In order to maintain the same degree of protection as the
AC units, always use a Safety Agency Approval DC source
with the same Overvoltage Category and Pollution Degree.
For immediate technical or application assistance please call:
Newport Electronics, Inc.
2229 South Yale Street Santa Ana, CA 92704 U.S.A.
TEL: (714) 540-4914 FAX: (714) 546-3022
Toll Free: 1-800-639-7678 http://www.newportUS.com e-mail:[email protected]
ISO 9001 Certified
Newport Technologies, Inc.
976 Bergar Laval (Quebec) H7L 5A1 Canada
TEL: (514) 335-3183 FAX: (514) 856-6886
Toll Free: 1-800-639-7678 http://www.newport.ca e-mail:[email protected]
Newport Electronics, Ltd.
One Omega Drive River Bend Technology Centre
Northbank, Irlam Manchester M44 5BD United Kingdom
Tel: +44 161 777 6611 FAX: +44 161 777 6622
Toll Free: 0800 488 488 http://www.newportuk.co.uk e-mail:[email protected]
Newport Electronics B.V.
Postbus 8034 1180 LA Amstelveen The Netherlands
TEL: +31 20 3472121 FAX: +31 20 6434643
Toll Free: 0800 0993344 http://www.newport.nl e-mail: [email protected]
Newport Electronics spol s.r.o.
Rudé armády 1868, 733 01 Karviná 8 Czech Republic
TEL: +420 69 6311899 FAX: +420 69 6311114
Toll Free: 0800-1-66342 http://www.newport.cz e-mail: [email protected]
Newport Electronics GmbH
Daimlerstrasse 26 D-75392 Deckenpfronn Germany
TEL: 49 7056 9398-0 FAX: 49 7056 9398-29
Toll Free: 0800 / 6397678 http://www.newport.de e-mail: [email protected]
Newport Electronique S.A.R.L.
9, rue Denis Papin 78190 Trappes France
TEL: +33 130 621 400 FAX: +33 130 699 120
Toll Free: 0800-4-06342 http://www.newport.fr e-mail: [email protected]
Mexico and Latin America
TEL: 001-800-826-6342 FAX: 001 (203) 359-7807
En Español: 001 (203) 359-7803
®
Safety Consideration
This device is m rked with the intern tion l C ution symbol.
The instrument is a panel mount device protected in accordance with
Class I of EN61010-1 (AC units) and Class III (DC units). Remember
that the unit has no power-on switch. Building installation should
include a switch or circuit-breaker that must be compliant to IEC 947-1
and 947-3.
SAFETY:
Do not exceed voltage rating on the label located on the
top of the instrument housing.
Always disconnect power before changing signal and
power connections.
Do not use this instrument on a work bench without its
case for safety reasons.
Do not operate this instrument in flammable or explosive
atmospheres.
Do not expose this instrument to rain or moisture.
EMC:
Whenever EMC is an issue, always use shielded cables.
Never run signal and power wires in the same conduit.
Use signal wire connections with twisted-pair cables.
Install Ferrite Bead(s) on signal wire close to the instrument if
EMC problems persist.
Scaling Without Known Loads
(Continued)
To define the maximum load (IN!2 and RD!2 ):
1. Repe t steps 3–9 bove, entering the v lues for IN!2
nd RD!2 .
2. Once you've completed all steps, the unit displays:
To begin operation:
Reinitialize the unit (press RESET twice
or press MENU until RST flashes on the
display). When a numeric reading
appears, the unit is operational.
Determining Reading Offset
The run mode reading for meters scaled without known loads
may reflect an offset. For example, say you set
RD!1
to 0 and
RD!2
to 100, but when the minimum load is applied, a
negative value of –1.5 displays on the front panel.
To adjust the display reading:
1. With zero load applied, note the reading on the display.
2. Subtract that amount from the RD!1 and RD!2 values you
originally entered.
In the example, the offset would be –1.5. If RD!1 is to read
0 in run mode, it must be reentered as 1.5. RD!2 must
likewise be reentered as 101.5 if the meter is to read 100
when the maximum load is applied.
3.
Repeat the steps for "Scaling Without Known Loads," but
when the values for IN!1 and IN!2 display, do not change
them. Instead, press MENU to move to the prompts for
RD!1 and RD!2 and make the necessary changes.
4. Reinitialize the unit and resume operation.
Scaling Without Known Loads
For 4–20 mA sensors, the values for the minimum and
maximum input loads are always as follows:
Minimum load (IN!1 )— 2000
Maximum load (IN!2 )— 9999.
If your installation uses a different sensor type, you must
calculate the values for IN!1 and IN!2 before proceeding
with the steps below. Use the formula provided in the
Operator's Manual.
To define the minimum load
(IN!1 and RD!1 ):
1. If it's not already shown, press MENU
until the unit displays:
2. Press 䊳/TARE. The unit displays:
3. Press 䊳/TARE again. The unit
displays the last setting for IN!1 .
(The first digit flashes.)
4. Change IN!1 as necessary:
Press
䊱
/MAX to set or change the digit's current value.
Continue to press
䊱
/MAX until the meter displays the desired
value for the flashing digit. Values can range from 0 to 9. For
the first digit, you can also enter a minus sign (–) or –1.
Press
䊳
/TARE to scroll to the digit(s) you want to change.
5. Press MENU to store IN!1 .
The unit displays:
6. Press 䊳/TARE. The unit displays the
last setting for RD!1 .
(The first digit flashes.)
7. Change RD!1 as necessary:
Press
䊱
/MAX until the meter displays the desired value for
the flashing digit. Values can range from 0 to 9. For the
first digit, you can also enter a minus sign (–) or –1.
Press
䊳
/TARE to scroll to the digit(s) you want to change.
8. Press MENU to store the value
shown for RD!1 . The unit displays:
Scaling With Known Loads
(Continued)
6. Press MENU to store IN!1 . The unit
displays:
7. Press 䊳/TARE. The unit displays the
last setting for RD!1 .
8. Change RD!1 as necessary:
Press
䊳
/TARE to scroll to the digit(s) you want to change
(it flashes on the display).
Press
䊱
/MAX to change the value of the flashing digit.
Values can range from 0 to 9. For the first digit, you can
also enter a minus sign (–) or –1.
9. Press MENU to store the value shown
for RD!1 . The unit displays:
To identify the maximum known load (IN!2 and RD!2 ):
1. Apply the maximum known load (100%).
2. Repe t steps 4–9 for IN!2 nd RD!2 .
Once you've completed all steps, the unit displays:
To begin operation:
Reinitialize the unit (press RESET
twice or press MENU until RST
flashes on the display). When a
numeric reading appears, the unit is
operational.
Scaling With Known Loads
To identify the minimum known load (IN!1 and RD!1 ):
1. If it's not already shown, press MENU
until the unit displays:
2. Apply the minimum known load (0%).
3. Press 䊳/TARE. The unit displays:
4. Press 䊳/TARE again. The unit
displays the last setting for
IN!1
.
5. Press 䊳/TARE again. The unit
displays the actual reading being
received from the sending device.
5678
MQS3597/N/0801
Warranty/Disclaimer
NEWPORT Electronics, Inc. warrants this unit to be free of defects in
materials and workmanship for a period of one (1) ye r from the date of
purchase. In addition to NEWPORT’s standard warranty period, NEWPORT
Electronics will extend the warranty period for one (1) ddition l ye r if the
warranty card enclosed with each instrument is returned to NEWPORT.
If the unit should malfunction, it must be returned to the factory for evaluation.
NEWPORT’s Customer Service Department will issue an Authorized Return
(AR) number immediately upon phone or written request. Upon examination
by NEWPORT, if the unit is found to be defective it will be repaired or replaced
at no charge. NEWPORT’s WARRANTY does not apply to defects resulting
from any action of the purchaser, including but not limited to mishandling,
improper interfacing, operation outside of design limits, improper repair, or
unauthorized modification. This WARRANTY is VOID if the unit shows
evidence of having been tampered with or shows evidence of being damaged
as a result of excessive corrosion; or current, heat, moisture or vibration;
improper specification; misapplication; misuse or other operating conditions
outside of NEWPORT’s control. Components which wear are not warranted,
including but not limited to contact points, fuses, and triacs.
NEWPORT is ple sed to offer suggestions on the use of its v rious
products. However, NEWPORT neither ssumes responsibility for ny
omissions or errors nor ssumes li bility for ny d m ges th t result
from the use of its products in ccord nce with inform tion provided by
NEWPORT, either verb l or written. NEWPORT w rr nts only th t the
p rts m nuf ctured by it will be s specified nd free of defects.
NEWPORT MAKES NO OTHER WARRANTIES OR REPRESENTATIONS
OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT
OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY
WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The
remedies of purch ser set forth herein re exclusive nd the tot l
li bility of NEWPORT with respect to this order, whether b sed on
contr ct, w rr nty, negligence, indemnific tion, strict li bility or
otherwise, sh ll not exceed the purch se price of the component upon
which li bility is b sed. In no event sh ll NEWPORT be li ble for
consequenti l, incident l or speci l d m ges.
CONDITIONS: Equipment sold by NEWPORT is not intended to be used, nor
shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in
or with any nuclear installation or activity; or (2) in medical applications or
used on humans. Should any Product(s) be used in or with any nuclear
installation or activity, medical application, or used on humans, or misused in
any way, NEWPORT assumes no responsibility as set forth in our basic
WARRANTY / DISCLAIMER language, and additionally purchaser will
indemnify NEWPORT and hold NEWPORT harmless from any liability or
damage whatsoever arising out of the use of the Product(s) in such a manner.
Direct all warranty and repair requests/inquiries to the NEWPORT Customer
Service Department. BEFORE RETURNING ANY PRODUCT(S) TO NEWPORT,
PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER
FROM NEWPORT’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO
AVOID PROCESSING DELAYS). The assigned AR number should then be
marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper
packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please
have the following information available
BEFORE contacting NEWPORT:
1. P.O. number under which the product
was PURCHASED,
2. Model and serial number of the
product under warranty, and
3. Repair instructions and/or specific
problems relative to the product.
FOR NON-WARRANTY REPAIRS,
consult NEWPORT for current repair
charges. Have the following information
available BEFORE contacting NEWPORT:
1. P.O. number to cover the COST of the
repair,
2. Model and serial number of product,
and
3. Repair instructions and/or specific
problems relative to the product.
NEWPORT’s policy is to make running changes, not model changes, whenever
an improvement is possible. This affords our customers the latest in technology
and engineering.
NEWPORT is a registered trademark of NEWPORT Electronics, Inc.
© Copyright 2001 NEWPORT Electronics, Inc. All rights reserved.
This document may not be copied, photocopied, reproduced, translated, or
reduced to any electronic medium or machine-readable form, in whole or in part,
without prior written consent of NEWPORT Electronics, Inc.
PATENT NOTICE
The “Meter Case Bezel Design” is a trademark of
NEWPORT Electronics, Inc., registered in the U.S.. This product is
covered by one or more of the following patents: U.S. Pat. No. Des.
336,895; 5,274,577; 6,243,021 / CANADA 2052599; 2052600 /
ITALY
1249456; 1250938 / FRANCE BREVET No. 91 12756 / SPAIN 2039150;
2048066 / UK PATENT No. GB2 249 837; GB2 248 954 / GERMANY DE 41
34398 C2. OTHER INTERNATIONAL PATENTS PENDING.
Return Requests/Inquiries
This device is marked with the international caution symbol. It is important
to read the Setup Guide before installing or commissioning this device as
it contains important information relating to safety and EMC.
WARNING These products are not designed for use in, and should not be
used for, patient connected applications.
It is the policy of NEWPORT to comply with all worldwide safety and EMC/EMI
regulations that apply. NEWPORT is constantly pursuing certification of its
products to the European New Approach Directives. NEWPORT will add the CE
mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but
NEWPORT Electronics, Inc. accepts no liability for any errors it contains, and
reserves the right to alter specifications without notice.
TRADEMARK NOTICE:
a
®
, , , , ,
and the “Meter Case Bezel Design”
are trademarks of NEWPORT Electronics, Inc.
N
EWP
O
RT
®
NEWP
O
RT
®
To Scale the Meter
You can scale the meter in one of two ways:
1. With a known load — This method uses input (load)
information sent from another device such as a scale or a
simulator for voltage or current.
2. Without a known load — This involves calculating the load
based on transducer specifications and manually entering
it to the meter.
For both methods, you must first identify the minimum input
load IN!1 and the corresponding display reading you want
RD!1 . Then you identify the maximum input load IN!2 and
its corresponding display reading RD!2 .
The decimal point is for display purposes only — you set it
where you want it to display for your application. When
entering IN!1 and IN!2 values, ignore any decimal point on
the display. However, you must enter RD!1 and RD!2 values
with the decimal point in the desired position.

Other Newport Measuring Instrument manuals

Newport 843-R User manual

Newport

Newport 843-R User manual

Newport 1936-R series User guide

Newport

Newport 1936-R series User guide

Newport Cornerstone 130B User manual

Newport

Newport Cornerstone 130B User manual

Newport iServer MicroServer User manual

Newport

Newport iServer MicroServer User manual

Newport INFINITY INFCR User manual

Newport

Newport INFINITY INFCR User manual

Newport OPHIR Photonics Spiricon Pyrocam IV User manual

Newport

Newport OPHIR Photonics Spiricon Pyrocam IV User manual

Newport FPM-8220 User manual

Newport

Newport FPM-8220 User manual

Newport PScout2 Series User manual

Newport

Newport PScout2 Series User manual

Newport 835 User manual

Newport

Newport 835 User manual

Newport 1936-R series User manual

Newport

Newport 1936-R series User manual

Newport 1825-C User manual

Newport

Newport 1825-C User manual

Newport QuantX-300 User manual

Newport

Newport QuantX-300 User manual

Newport Oriel MS260i Series User manual

Newport

Newport Oriel MS260i Series User manual

Newport 2835-C User manual

Newport

Newport 2835-C User manual

Newport Hygrodynamics 8074A User manual

Newport

Newport Hygrodynamics 8074A User manual

Newport PHH-60 User manual

Newport

Newport PHH-60 User manual

Newport wi Series User manual

Newport

Newport wi Series User manual

Newport 290 Series User manual

Newport

Newport 290 Series User manual

Newport 843-R User manual

Newport

Newport 843-R User manual

Newport N-STP-910 User manual

Newport

Newport N-STP-910 User manual

Newport IDP User manual

Newport

Newport IDP User manual

Newport 2832-C User manual

Newport

Newport 2832-C User manual

Newport Cornerstone 260B User manual

Newport

Newport Cornerstone 260B User manual

Newport CONEX-LDS User manual

Newport

Newport CONEX-LDS User manual

Popular Measuring Instrument manuals by other brands

LaMotte DC1500 FLUORIDE manual

LaMotte

LaMotte DC1500 FLUORIDE manual

Keller ECO 2 manual

Keller

Keller ECO 2 manual

WIKA cpd8000-g user manual

WIKA

WIKA cpd8000-g user manual

Gossen MetraWatt METRACLIP 87 user manual

Gossen MetraWatt

Gossen MetraWatt METRACLIP 87 user manual

Guru Home Systems GU-HB-201-R1 installation guide

Guru Home Systems

Guru Home Systems GU-HB-201-R1 installation guide

Scaime GM 80 TG Operating	 instruction

Scaime

Scaime GM 80 TG Operating instruction

Inficon SQM-160 operating manual

Inficon

Inficon SQM-160 operating manual

PCB Piezotronics TLD356A32 Installation and operating manual

PCB Piezotronics

PCB Piezotronics TLD356A32 Installation and operating manual

ETG ETG6900 User instruction manual

ETG

ETG ETG6900 User instruction manual

HT HT9023 user manual

HT

HT HT9023 user manual

ADREL ALT-BMP manual

ADREL

ADREL ALT-BMP manual

Contrec 103D manual

Contrec

Contrec 103D manual

Endress+Hauser Liquiline CA80FE Brief operating instructions

Endress+Hauser

Endress+Hauser Liquiline CA80FE Brief operating instructions

Panasonic KW1M Eco-Power METER user manual

Panasonic

Panasonic KW1M Eco-Power METER user manual

Trumeter APM quick start guide

Trumeter

Trumeter APM quick start guide

horiba LAQUA WQ-300 Series quick guide

horiba

horiba LAQUA WQ-300 Series quick guide

Teledyne TIENet 301 Installation and operation guide

Teledyne

Teledyne TIENet 301 Installation and operation guide

socomec DIRIS B-10 instruction manual

socomec

socomec DIRIS B-10 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.