Nivus NivuCont Plus User manual

Instruction Manual
NivuCont Plus
®
Instruction Manual for
NivuCont Plus Measurement Device
(Original Instruction Manual – German)
Software revision number 2.23
NIVUS GmbH
Im Täle 2
D – 75031 Eppingen
Tel. 0 72 62 / 91 91 - 0
Fax 0 72 62 / 91 91 - 999
Internet: www.nivus.com
NivuCont Plus - Rev. 02 as of 18.02.2009 page 1

®
Instruction Manual
NIVUS Representatives
NIVUS Representatives:
NIVUS AG
Hauptstrasse 49
CH – 8750 Glarus
Tel. +41 (0)55 / 645 20 66
Fax +41 (0)55 / 645 20 14
Internet: www.nivus.com
NIVUS Sp. z o. o
Ul. Hutnicza 3 / B-18
PL – 81-212 Gdynia
Tel. +48 (0)58 / 760 20 15
Fax +48 (0)58 / 760 20 14
Internet: www.nivus.pl
NIVUS France
14, rue de la Paix
F – 67770 Sessenheim
Tel. +33 (0)388071696
Fax +33 (0)388071697
Internet: www.nivus.com
NIVUS U.K.
P.O. Box 342
Egerton, Bolton
Lancs. BL7 9WD, U.K.
Tel: +44 (0)1204 591559
Fax: +44 (0)1204 592686
Internet: www.nivus.com
page 2 Rev. 02 as of 18.02.2009

Instruction Manual
NivuCont Plus
®
Translation
If the device is sold to a country in the European Economic Area (EEA)
this instruction handbook must be translated into the language of the
country in which the device is to be used.
Should the translated text be unclear, the original instruction handbook
(German) must be consulted or the manufacturer contacted for clarifica-
tion.
Copyright
No part of this publication may be reproduced, transmitted, sold or dis-
closed without prior permission. Damages will be claimed for violations.
All rights reserved.
Names
The use of general descriptive names, trade names, trademarks and the
like in this handbook does not entitle the reader to assume they may be
used freely by everyone. They are often protected registered trademarks
even if not marked as such.
NivuCont Plus - Rev. 02 as of 18.02.2009 page 3

®
Instruction Manual
NivuCont Plus
1 Contents
1.1 Table of Contents
1Contents...............................................................................4
1.1
Table of Contents.............................................................................4
1.2
Declaration of Conformity NivuCont Plus........................................6
1.3
Declaration of Conformity „Ex“ NivuCont Plus.................................7
1.4
Ex-Approval Transmitter..................................................................8
2Overview and use in accordance with the requirements10
2.1
Overview........................................................................................10
2.2
Use in accordance with the requirements......................................11
2.3
Specifications.................................................................................12
3General Notes on Safety and Danger...............................13
3.1
Danger Notes.................................................................................13
3.1.1
General Danger Signs...................................................................13
3.1.2
Special Danger Notes....................................................................13
3.2
Device Identification.......................................................................14
3.3
Installation of Spare Parts and Parts subject to Wear and Tear ...14
3.4
Turn-off procedure.........................................................................14
3.5
User’s Responsibilities...................................................................15
4Functional Principle ..........................................................16
4.1
General..........................................................................................16
4.2
Unit Versions..................................................................................16
5Storing, Delivery and Transport.......................................17
5.1
Receipt...........................................................................................17
5.1.1
Delivery..........................................................................................17
5.2
Storing............................................................................................17
5.3
Transport........................................................................................17
5.4
Return............................................................................................17
6Installation..........................................................................18
6.1
General..........................................................................................18
6.2
Transmitter Dimensions, Installation and Connection...................18
6.2.1
General..........................................................................................18
6.2.2
Enclosure Dimensions...................................................................19
6.2.3
Transmitter Installation...................................................................20
6.2.4
Transmitter Connection..................................................................21
6.2.5
Connection Variations....................................................................22
6.3
Input Configuration.........................................................................23
6.3.1
Connection Examples....................................................................25
7Overvoltage Protection .....................................................27
page 4 Rev. 02 as of 18.02.2009

Instruction Manual
NivuCont Plus
®
8................................................................................Initial start-up
.............................................................................................29
8.1
General..........................................................................................29
8.2
Display and Control Elements .......................................................30
8.2.1
Display...........................................................................................30
8.2.2
Navigation Keys.............................................................................31
9Parameter Setting..............................................................32
9.1
Brief Description of the single Basic Menus..................................32
9.1.1
Description of PAR Submenus......................................................33
9.1.2
Download and store NivuCont Plus parameters............................35
9.1.3
Upload Parameters from PC to NivuCont Plus..............................42
9.2
Parameter Setting Principles.........................................................44
9.3
Parameter Menu (PAR).................................................................46
9.3.1
Parameter Menu „Settings“............................................................47
9.3.2
Parameter Menu „Analog Input“ ....................................................48
9.3.3
Parameter Menu „Measurement Values 1-3“................................50
9.3.4
Parameter Menu „Measurement Value 4“ .....................................57
9.3.5
Parameter Menu „Display“.............................................................58
9.3.6
Parameter Menu „Relay“................................................................59
9.3.7
Parameter Menu „Analog Output“..................................................65
9.3.8
Parameter Menu „Volume Calculation“.........................................67
9.3.9
Parameter Menu "Interface“...........................................................67
9.4
Signal Input / Output Menu (I/O)....................................................67
9.5
Calibration Menu (CAL).................................................................68
9.6
Information Menu (EXTRA) ...........................................................68
9.7
Block Diagram of NivuCont Plus....................................................71
10 Troubleshooting ................................................................77
11 Programming Example......................................................78
11.1
Level/Height...................................................................................78
11.2
Rack Difference Measurement ......................................................82
11.2.1
Application Example Rack Control ................................................82
11.2.2
Detailed Parameter setting of Rack Difference Measurement: .....84
11.3
Overflow Measurement..................................................................91
11.3.1
Application Example Overflow Measurement................................91
11.3.2
Detailed Parameter setting of an Overflow Measurement:............92
11.4
Basin Purification Control ............................................................100
11.4.1
Application Example Basin Purification Control...........................100
11.4.2
Detailed Parameter setting of a Basin Purif. Control via 1 Relay:101
12 Maintenance and Cleaning..............................................109
13 Emergency .......................................................................110
14 Dismantling/Disposal ......................................................110
15 Table of Pictures..............................................................110
NivuCont Plus - Rev. 02 as of 18.02.2009 page 5

Q:\Formulare\CE…
NIVUS GmbH
Im Täle 2
75031 Eppingen
Telefon: +49 07262 9191-0
Telefax: +49 07262 9191-999
Internet: www.nivus.de
EU Konformitätserklärung
EU Declaration of Conformity
Déclaration de conformité UE
Für das folgend bezeichnete Erzeugnis:
For the following product:
Le produit désigné ci-dessous:
Bezeichnung:
Description:
Désignation:
Multifunktionaler Messumformer NivuCont Plus
Multi-functional measurement transmitter
Convertisseur de mesure multifonctionnel
Typ / Type: NCP-…
erklären wir in alleiniger Verantwortung, dass die auf dem Unionsmarkt ab dem Zeitpunkt der Unterzeichnung
bereitgestellten Geräte die folgenden einschlägigen Harmonisierungsvorschriften der Union erfüllen:
we declare under our sole responsibility that the equipment made available on the Union market as of the date of signature of
this document meets the standards of the following applicable Union harmonisation legislation:
nous déclarons, sous notre seule responsabilité, à la date de la présente signature, la conformité du produit pour le marché de
l’Union, aux directives d'harmonisation de la législation au sein de l’Union:
• 2014/30/EU • 2014/35/EU • 2011/65/EU
Bei der Bewertung wurden folgende einschlägige harmonisierte Normen zugrunde gelegt bzw. wird die Konformität
erklärt in Bezug die nachfolgend genannten anderen technischen Spezifikationen:
The evaluation assessed the following applicable harmonised standards or the conformity is declared in relation to other
technical specifications listed below:
L’évaluation est effectuée à partir des normes harmonisées applicable ou la conformité est déclarée en relation aux autres
spécifications techniques désignées ci-dessous:
• EN 61000-6-2:2005 • EN 61000-6-4:2007/A1:2011 • EN 61010-1:2010
Diese Erklärung wird verantwortlich für den Hersteller:
This declaration is submitted on behalf of the manufacturer:
Le fabricant assume la responsabilité de cette déclaration:
NIVUS GmbH
Im Taele 2
75031 Eppingen
Allemagne
abgegeben durch / represented by / faite par:
Marcus Fischer (Geschäftsführer / Managing Director / Directeur général)
Eppingen, den 20.04.2016
Gez. Marcus Fischer
DE / EN / FR

NIVUS GmbH
Im Täle 2
75031 Eppingen
Telefon: +49 07262 9191-0
Telefax: +49 07262 9191-999
E-Mail: info@nivus.com
Internet: www.nivus.de
EU Konformitätserklärung
EU Declaration of Conformity
Déclaration de conformité UE
Für das folgend bezeichnete Erzeugnis:
For the following product:
Le produit désigné ci-dessous:
Bezeichnung:
Description:
Désignation:
“Ex“ Multifunktionaler Messumformer NivuCont Plus
“Ex” Multi-functional measurement transmitter
“Ex”Convertisseur de mesure multifonctionnel
Typ / Type: NCP-xxxxxxE…
erklären wir in alleiniger Verantwortung, dass die auf dem Unionsmarkt ab dem Zeitpunkt der Unterzeichnung
bereitgestellten Geräte die folgenden einschlägigen Harmonisierungsvorschriften der Union erfüllen:
we declare under our sole responsibility that the equipment made available on the Union market as of the date of signature of
this document meets the standards of the following applicable Union harmonisation legislation:
nous déclarons, sous notre seule responsabilité, à la date de la présente signature, la conformité du produit pour le marché de
l’Union, aux directives d'harmonisation de la législation au sein de l’Union:
• 2014/30/EU • 2014/34/EU • 2014/35/EU • 2011/65/EU
Bei der Bewertung wurden folgende einschlägige harmonisierte Normen zugrunde gelegt bzw. wird die Konformität
erklärt in Bezug die nachfolgend genannten anderen technischen Spezifikationen:
The evaluation assessed the following applicable harmonised standards or the conformity is declared in relation to other
technical specifications listed below:
L’évaluation est effectuée à partir des normes harmonisées applicable ou la conformité est déclarée en relation aux autres
spécifications techniques désignées ci-dessous:
• EN 61000-6-2:2005 • EN 61000-6-4:2007 /A1:2011 • EN 60079-0:2012 +A11:2013
• EN 60079-11:2012 • EN 61010-1:2010
Ex-Kennzeichnung / Ex-designation / Marquage Ex : II (1)GD [Ex ia] IIC
EG-Baumusterprüfbescheinigung / EC-Type Examination Certificate / Attestation d'examen «CE» de type:
IBExU 04 ATEX 1208
Notifizierte Stelle (Kennnummer) / Notified Body (Identif. No.) / Organisme notifié (№ d’identification)
IBExU Institut für Sicherheitstechnik GmbH, 09599 Freiberg, Allemagne (0637)
Qualitätssicherung ATEX / Quality assurance ATEX / Assurance qualité ATEX:
TÜV Nord CERT GmbH, Am TÜV 1, 30519 Hannover, Allemagne (0044)
Diese Erklärung wird verantwortlich für den Hersteller:
This declaration is submitted on behalf of the manufacturer:
Le fabricant assume la responsabilité de cette déclaration:
abgegeben durch / represented by / faite par:
Marcus Fischer (Geschäftsführer / Managing Director / Directeur général)
Eppingen, den 26.07.2017
Gez. Marcus Fischer
NIVUS GmbH
Im Taele 2
75031 Eppingen
Allemagne
DE / EN / FR

®
Instruction Manual
NivuCont Plus
1.4 Ex-Approval Transmitter
page 8 Rev. 02 as of 18.02.2009

Instruction Manual
NivuCont Plus
®
The approval is only valid in connection with the respective indication on the
transmitter’s nameplate.
NivuCont Plus - Rev. 02 as of 18.02.2009 page 9

®
Instruction Manual
NivuCont Plus
2 Overview and use in accordance with the requirements
2.1 Overview
1 Display
2 Keypad
3 Terminal Clamp Housing
4 RS232 Interface
Fig. 2-1 Overview
page 10 Rev. 02 as of 18.02.2009

Instruction Manual
NivuCont Plus
®
2.2 Use in accordance with the requirements
The measurement device type NivuCont Plus is capable of processing
4-20 mA input signals. Additionally, it is possible to supply up to two 2 or 3-wire
sensors with a defined voltage. Here the allowed maximum values, as specified
in chapter 2.3 must be strictly kept. All cases which vary from these conditions
and are not passed by NIVUS GmbH in writing are left at owner’s risk.
The device is exclusively intended to be used for purposes as described
above.
Modifying or using the devices for other purposes without the written consent
of the manufacturer will not be considered as use in accordance with the re-
quirements. Damages resulting from this are left at user’s risk.
The device is designed for a lifetime of approx. 10 years. After that period an
inspection in addition with a general overhaul has to be made.
Ex-Protection
The Ex-version of the NivuCont Plus is for power supply of intrinsically safe sen-
sors in areas with explosive atmospheres (zone 0/1).
The transmitter always has to be installed outside of Ex-zones!
Approval
Transmitter: II (1) GD [EEx ia] IIC
Electric Values
Sensor circuit (2L/3L) in EEx ia IIC/IIB (type of protection)
per channel U0= 27,3 V
I
0= 93,4 mA
P
0= 637 mW
linear characteristic
Max. values connectable concentrated capacities and/or inductivities n
EEx ia IIC EEx ia IIB
C080nF 400nF
L01mH 5mH
The approval is only valid in connection with the respective indication on the
transmitter nameplate.
For installation and initial start-up the conformity certificates and test certifi-
cates of the respective authorities must be followed.
NivuCont Plus - Rev. 02 as of 18.02.2009 page 11

®
Instruction Manual
NivuCont Plus
2.3 Specifications
Transmitter
Power supply 85 - 264 V AC, 50 - 60 Hz or 18 - 36 V DC (stabilized)
Power consumption max. 10 VA
Wall mount enclosure
-
Material: Polycarbonate / on-site enclosure,
wall mount and DIN rail mount,
-
Weight: - wall mount approx. 1620 g, Protection IP65
- panel mount approx. 1770 g, Protection IP30
- 19“-rack mount: approx. 750 g, Protection IP20
Ex-Approval (optional) II (1) GD [EEx ia] IIC
Operating temperature -20 °C to +60 °C (-4 °F to +140 °F)
Ex: -20 °C to +40 °C (-4 °F to +104 °F)
Storage temperature -30 °C to +70 °C (-22 °F to +158 °F)
Max. humidity 90 %, non-condensing
Display Back-lit graphic display, 128 x 64 pixel
Operation 6 keys, menu driven in German, English and French
Inputs
-
analog input 1 and 2 (option) sensor power supply: 24 V DC,
max. 63 mA for Ex, max. 130 mA for non-Ex, 4 - 20 mA
Outputs
-
analog output 1: 0/4 - 20 mA output range, load max. 500 Ω,
12 bit resolution, galvanically isolated from analog input
-
analog output 2 (option): values see analog output 1
Relays 1 or 5 free programmable relays
Function: SPDT (Single Pole Double Throat)
unblocking potential: min. 100 mV, max. 250 V AC
current on contact: max. 2 A (cos. ϕ0,9)
circuit-breaking capacity: max. 300 VA
Measurement uncertainty Linearity: ±0,1 % of final value
Temperature Drift: ±0,05 % / 10 K ambient temperature
Interface / Communication via RS232
page 12 Rev. 02 as of 18.02.2009

Instruction Manual
NivuCont Plus
®
3 General Notes on Safety and Danger
3.1 Danger Notes
3.1.1 General Danger Signs
Cautions
are framed and labelled with a warning triangle.
Notes
are framed and labelled with a “hand“.
Danger by electric voltage
is framed and labelled with the Symbol on the left.
Warnings
are framed and labelled with a “STOP“-sign.
For connection, initial start-up and operation of the NivuCont Plus the following
information and higher legal regulations (e.g. in Germany VDE), such as Ex-
regulations as well as safety requirements and regulations in order to avoid ac-
cidents, must be adhered to.
All operations, which go beyond steps regarding installation, connection or pro-
gramming the unit are allowed to be carried out by NIVUS staff only due to rea-
sons of safety and guarantee.
3.1.2 Special Danger Notes
Please note that due to the operation in the waste water field transmitter, sen-
sors and cables may be loaded with hazardous disease germs. Respective
precautionary measures must be taken to avoid damage to one’s health.
NivuCont Plus - Rev. 02 as of 18.02.2009 page 13

®
Instruction Manual
NivuCont Plus
3.2 Device Identification
The instructions in this manual apply only for the type of device indicated on the
title page. The nameplate is fixed on the on the bottom of the device and con-
tains the following:
- Name and address of manufacturer
- CE label
- Type and serial number
- Year of manufacture
- Ex-label (on Ex-version devices only) as mentioned in chapter 2.2
It is important for enquiries and replacement part orders to specify article num-
ber as well as serial number of the respective transmitter or sensor. This en-
sures correct and quick processing.
This instruction manual is a part of the device and must be available for the
user at any time.
The safety instructions contained within must be followed.
It is strictly prohibited to disable safety measures or to change the way
they work!
3.3 Installation of Spare Parts and Parts subject to Wear and Tear
We herewith particularly emphasize that replacement parts or accessories,
which are not supplied by us, are not certified by us, too. Hence, the installation
and/or the use of such products may possibly be detrimental to the device’s abil-
ity to work.
Damages caused by using non-original parts and non-original accessories are
left at user’s risk.
3.4 Turn-off procedure
For maintenance, cleaning and repairs (authorised staff personnel only) the
device shall be disconnected from mains and shall be prevented from being
turned on again unintentionally.
page 14 Rev. 02 as of 18.02.2009

Instruction Manual
NivuCont Plus
®
3.5 User’s Responsibilities
In the EEA (European Economic Area) national implementation of the frame-
work directive 89/391/EEC and corresponding individual directives, in particu-
lar the directive 89/655/EEC concerning the minimum safety and health re-
quirements for the use of work equipment by workers at work, as amended,
are to be observed and adhered to.
In Germany the Industrial Safety Ordinance must be observed.
The customer must (where necessary) obtain any local operating permits re-
quired and observe the provisions contained therein.
In addition to this, he must observe local laws and regulations on
- personnel safety (accident prevention regulations)
- safety of work materials and tools (safety equipment and maintenance)
- disposal of products (laws on wastes)
- disposal of materials (laws on wastes)
- cleaning (cleansing agents and disposal)
- environmental protection
Connections :
Before operating the device the user has to ensure, that the local regulations
(e.g. for electric supply) on installation and initial start-up are taken into account,
if this is both carried out by the user.
NivuCont Plus - Rev. 02 as of 18.02.2009 page 15

®
Instruction Manual
NivuCont Plus
4 Functional Principle
4.1 General
The NivuCont Plus is a multifunctional process measurement transmitter capa-
ble to supply several 2 or 3-wire sensors. Computer options as well as links be-
tween the inputs concerning comparison, average value, total, linearization, cal-
culation of exponential functions and volume are possible. Various switching
functions such as limit value, in bounds, out of bounds, trend, timer as well as
pump management and pump control are available.
4.2 Unit Versions
The transmitters primarily vary in terms of power supply, Ex-protection and en-
closure construction. The current type of device is indicated by the article num-
ber, which can be found on a weatherproof label on the bottom of the enclosure
or the 19“ unit.
From this article key the type of device can be specified.
NCP- Type
10 1Input4-20 mA; 1output0/4-20 mA,load500 Ξ
12
15
Construction / Enclosure
19 Plug-in unit in euro format
VA 240 x185 x 119 mm; (WxHxD)
FB Panel mountIP30
Power Supply
AC 85 - 264 V AC / 50 - 60 Hz
DC 18 - 36 V DC
Approvals
0none
EIntrinsically safe sensor supplyfor Ex zone 0
SerialInterface
RRS232
Extensions
0noExtension
1
Software Extensions
00 none(standard software)
0B
0M Volume and flow calculation
0K
NCP- R
Field enclosure IP65
also for DIN rail mounting
Trend and control
Basin purification
Combination of trend and control basin
purification and volume
as well as flow calculation
twoanalog inputs and twoanalogoutputs
for 2 separate measurements, total,
difference, average
61 x 128.7 x 172 mm; (WxHxD) [12HP/3RU]
72 x 144 x 215 mm; (WxHxD) [12HP/3RU]
1Input4-20 mA; 1output0/4-20 mA,load 500 Ξ;
2 potential-free limit-/error message contacts, switching capacity 5 A/230 V
(ohm resistive load)
1Input4-20 mA; 1output0/4-20 mA,load 500 Ξ;
5 potential-free limit/error message contacts, switching capacity 5 A/230 V
(ohm resistive load)
Fig. 4-1 Type key for NivuCont Plus measurement transmitter
page 16 Rev. 02 as of 18.02.2009

Instruction Manual
NivuCont Plus
®
5 Storing, Delivery and Transport
5.1 Receipt
Please check your delivery if it is complete and in working order according to
the delivery note immediately after receipt. Any damage resulting from transport
or transit shall be reported to the carrier instantly. An immediate, written report
must be sent to NIVUS GmbH Eppingen as well.
Please report any shortcoming due to delivery to your representative or directly
to NIVUS Eppingen within two weeks in writing.
Mistakes cannot be rectified later!
5.1.1 Delivery
The standard delivery of the NivuCont Plus measurement system contains:
- the instruction manual with the certificate of conformity. All required steps to
correctly install and to operate the measurement system are listed herein.
- a NivuCont Plus transmitter
Additional accessories such as sensors etc. depending on order. Please check
by using the delivery note.
5.2 Storing
The following storing conditions must be strictly adhered to:
Transmitter: min. temperature: - 30 °C (-22 °F)
max. temperature: +70 °C (158 °F)
max. humidity: 90 %, non-condensing
The devices must be protected from corrosive or organic solvent vapours, radio-
active radiation as well as strong electromagnetic radiation.
5.3 Transport
Sensor and Transmitter are conceived for harsh industrial conditions. Despite
this do not expose them to heavy shocks or vibrations.
Transportation must be carried out in the original packaging.
5.4 Return
The units must be returned at customer cost to NIVUS Eppingen in the
original packaging.
Otherwise the return cannot be accepted!
NivuCont Plus - Rev. 02 as of 18.02.2009 page 17

®
Instruction Manual
NivuCont Plus
6 Installation
6.1 General
For electric installation the local regulations in the respective countries
(e.g. VDE 0100 in Germany) must be referred to.
The NivuCont Plus power supply must be separately protected by a 6 A slow-
blow fuse and has to be isolated from other facility parts (separate turn-off,
e.g. by using an automatic cut-out with >B< characteristics).
Before feeding the rated voltage the transmitter and sensor installation must be
correctly completed. The installation should be carried out by qualified personnel
only. Further statutory standards, regulations and technical rulings have to be
taken into account.
All outer circuits, wires and lines connected to the device must have a minimum
isolation resistance of 250 V. If the voltage exceeds 42 V DC an isolation resis-
tance with 500 kOhm min. will be required.
The cross-sectional dimension of the power supply wires must be 0.75 mm2
(0.03 in2) and must be in accordance to IEC 227 or IEC 245. To observe the de-
gree of device protection please refer to chapter 2.3.
The maximum allowed switching voltage on the relay contacts must not exceed
250 V. According to Ex protection it must be checked if the devices power sup-
plies must be integrated into the facility’s emergency shutdown conception.
6.2 Transmitter Dimensions, Installation and Connection
6.2.1 General
The transmitter mounting place has to be selected according to certain criteria.
Please strictly avoid:
- direct sunlight (use weatherproof cover if necessary)
- heat emitting objects (max. ambient temperature: +40 °C (104 °F))
- objects with strong electromagnetic fields (e.g. frequency converters, electric
motors with high power consumption or similar)
- corrosive chemicals or gas
- mechanical shocks
- installation close to footpaths or travel ways
- vibrations
- radioactive radiation
The panel mount enclosure is fastened by using the 4 clamp screws which are
integrated in the enclosure sides.
The rack unit is installed in a standard rotating or fixed frame by using 4 machine
screws size M6 or M8 in suitable length with the necessary nuts and shims.
page 18 Rev. 02 as of 18.02.2009

Instruction Manual
NivuCont Plus
®
The clear view door of the measurement transmitter is provided with a protection
foil for protection during transport and from scratches during assembly. This pro-
tection foil has to be removed immediately after the assembly.
If the view door protection foil has been exposed to direct solar radiation for a
long period, it cannot be removed easily.
Cleaning of the front foil can be undertaken with spirit or if necessary with
car polish. If this is not successful, a new front door can be ordered from
NIVUS GmbH or your local representative.
6.2.2 Enclosure Dimensions
The transmitter is available in 2 different enclosures -
on-site enclosure (wall/DIN rail mounting) or 19“-unit.
213 (8.39in)
239 (9.41in)
198 (7.8in)
5.3 (0.21in)
10 (0.39in) 137 (5.39in) 30
(1.18in)
58.5 (2.3in)
185 (7.28in)
5.3 (0.21in)
118.2 (4.65in)
8 (0.31in)
Fig. 6-1 Wall Mount Enclosure Type VB
NivuCont Plus - Rev. 02 as of 18.02.2009 page 19

®
Instruction Manual
NivuCont Plus
135 (3HE / 5.31in.)
63 (12TE / 2.48in.)
72 (2.83in.)
215 (8.46in.)
144 (5.67in.)
enclosure tubus
terminal clam
p
rail
Fig. 6-2 Panel Mount Enclosure Type FB
6.2.3 Transmitter Installation
General
Please note during installation that electronic components may be destroyed
caused by electrostatic bursts. Due to this reason avoid high electrostatic charge
by providing proper ground connection.
Please note the following points when selecting the measurement place:
- The installation surface must be free of vibrations.
- The ambient temperature must be between -20 ºC and +40 ºC
(-4 °F and +104 °F)
- Keep away from power lines or frequency converters
(approx. 100 m (300 ft.)).
Panel Mount Installation:
Panel mount enclosures are to be installed by fastening the tube in the prepared
cut-out.
The most simple way to install a wall mount enclosure is to fasten a DIN rail with
a length of 210 mm (8.3 in.) and then to snap-on the enclosure.
It is possible to install the enclosure by using 3 screws as well. Use a pan head
screw with a head diameter of 5.5 ... 8.0 mm (0.22 ... 0.32 in.) for this. This
screw must be screwed into the mounting plate protruding 4 mm (0.16 in.). Then
hang the enclosure on the screw and additionally fix it with 2 more screws from
the terminal clamp housing. Please observe to screw them at least 40 mm into
the wall or 50 mm into appropriate dowels (to be set).
Before the first connection it is necessary to have a slight pressure on the
screw of the clamping connection to ensure safe opening and a correct
connection.
page 20 Rev. 02 as of 18.02.2009
Table of contents
Other Nivus Measuring Instrument manuals

Nivus
Nivus NivuLink Compact NLC0CLOG User manual

Nivus
Nivus NFP User manual

Nivus
Nivus NFP User manual

Nivus
Nivus NivuFlow Stick User manual

Nivus
Nivus OCM PRO CF User manual

Nivus
Nivus OCM PRO CF User manual

Nivus
Nivus osm-f Technical manual

Nivus
Nivus NivuSonic Clamp On User manual

Nivus
Nivus NivuCompact User manual

Nivus
Nivus NivuFlow Mobile 600 User manual