Nivus NFP User manual

FLOW permanent
Instruction Manual
Flow Measurement Transmitter
NFP – NIVUS Full Pipe
Firmware Revision: 4.00
Original Manual: German / 23.05.2018
Rev. 04 / 23.05.2018

Instruction Manual
NIVUS Full Pipe
page 2 NFP - rev. 04 / 23.05.2018
NIVUS AG, Switzerland
Burgstraße 28
8750 Glarus, Switzerland
Phone: +41 (0)55 6452066
Fax: +41 (0)55 6452014
swiss@nivus.com
www.nivus.de
NIVUS Austria
Mühlbergstraße 33B
3382 Loosdorf, Austria
Phone: +43 (0) 2754 567 63 21
Fax: +43 (0) 2754 567 63 20
www.nivus.de
NIVUS Sp. z o.o., Poland
ul. Hutnicza 3 / B-18
81-212 Gdynia, Poland
Phone: +48 (0) 58 7602015
Fax: +48 (0) 58 7602014
www.nivus.pl
NIVUS France
14, rue de la Paix
67770 Sessenheim, France
Phone: +33 (0)3 88071696
Fax: +33 (0)3 88071697
www.nivus.fr
NIVUS Ltd., United Kingdom
Wedgewood Rugby Road
Weston under Wetherley
Royal Leamington Spa
CV33 9BW, Warwickshire
Phone: +44 (0)8445 3328 83
www.nivus.com
NIVUS Middle East (FZE)
Building Q 1-1 ap. 055
P.O. Box: 9217
Sharjah Airport International
Free Zone
Phone: +971 6 55 78 224
Fax: +971 6 55 78 225
middle-[email protected]
www.nivus.com
NIVUS Korea Co. Ltd.
#2502, M Dong, Technopark IT Center,
32 Song-do-gwa-hak-ro,
Yeon-su-gu,
INCHEON, Korea 21984
Phone: +82 32 209 8588
Fax: +82 32 209 8590
www.nivus.com
NIVUS Vietnam
21 Pho Duc Chinh, Ba Dinh,
Hanoi, Vietnam
Phone: +84 12 0446 7724
www.nivus.com
NIVUS Chile
Viña Cordillera Oriente 4565
Puente Alto, Santiago
Phone: +562 2266 8119
www.nivus.com

Copyrights and property rights
NFP - rev. 04 / 23.05.2018 page 3
Copyrights and property rights
This document and its contents are proprietary to NIVUS GmbH and are not to be repro-
duced or copied without the express written permission of NIVUS GmbH.
Violations oblige to compensation.
Important Note
This manual may exclusively - even in parts - be copied or translated in any other way with
the express written consent of NIVUS GmbH.
Translation
If the device is sold to a country in the European Economic Area (EEA) this instruction manu-
al must be translated into the language of the country in which the device is to be used.
Should the translated text be unclear, the original instruction manual (German) must be con-
sulted or the manufacturer contacted for clarification.
Copyright
No part of this publication may be reproduced, transmitted, sold or disclosed without prior
permission. Damages will be claimed for violations. All rights reserved.
Names
The use of general descriptive names, trade names, trademarks and the like in this manual
does not entitle the reader to assume they may be used freely by everyone. They are often
protected registered trademarks even if not marked as such.

Instruction Manual
NIVUS Full Pipe
page 4 NFP - rev. 04 / 23.05.2018
Table of Contents
Copyrights and property rights 3
Table of Contents 4
General 7
1About this manual .................................................................................7
1.1 Applicable documentation.............................................................7
1.2 Signs and definitions used ............................................................8
1.3 Abbreviations used........................................................................8
1.3.1 Colour code for wires and single conductors..................................... 8
Safety Instructions 9
2Used symbols and signal words ...........................................................9
2.1 Valuation of the accident level.......................................................9
2.2 Warning notices on the product (option) .....................................10
3Safeguards and Precautions...............................................................10
4Liability disclaimer...............................................................................11
5Use in accordance with the requirements...........................................11
6User’s Responsibilities........................................................................13
7Personnel requirements......................................................................13
Delivery, Storage and Transport 14
8Delivery ...............................................................................................14
9Reception inspection...........................................................................14
10 Storing.................................................................................................14
11 Transport.............................................................................................14
12 Return..................................................................................................14
Product specification 15
13 Overview .............................................................................................15
14 Device identification............................................................................15
15 Specifications......................................................................................16
16 Configuration.......................................................................................18
16.1 Device Types...............................................................................18
Functional Principle 19
17 Operating Range.................................................................................19
18 Flow Velocity Measurement................................................................19

Table of Contents
NFP - rev. 04 / 23.05.2018 page 5
Installation and Connection 22
19 General Installation Instructions..........................................................22
20 Transmitter Installation and Connection .............................................22
20.1 General........................................................................................22
20.2 Enclosure Dimensions.................................................................23
20.3 Hints on how to avoid electrostatic discharge (ESD)..................24
20.4 Transmitter Installation................................................................24
20.5 Electrical Installation....................................................................25
20.5.1 Transmitter Connection ................................................................... 25
20.5.2 Sensor Connection.......................................................................... 28
20.6 NFP Power Supply......................................................................29
20.7 Overvoltage Protection Precautions............................................30
Initial start-up 34
21 Notes to the user.................................................................................34
22 General principles ...............................................................................34
23 Operator Panel....................................................................................35
24 Display.................................................................................................35
25 Operation Basics.................................................................................37
Parameter Setting 38
26 Basics of Parameter Setting ...............................................................38
27 Operation Mode (RUN).......................................................................40
28 Display Menu (EXTRA).......................................................................43
29 Parameter Menu (PAR).......................................................................46
29.1 Parameter Menu “Measurement Place”......................................47
29.2 Parameter Menu “Velocity”..........................................................49
29.3 Parameter menu “Digital Input”...................................................49
29.4 Parameter Menu “Analog Output”...............................................51
29.5 Parameter Menu “Relay”.............................................................53
29.6 Parameter Menu “Settings”.........................................................57
29.7 Parameter menu “Storage mode”................................................58
30 Signal Input/Output Menu (I/O)...........................................................59
30.1 I/O sub menu “Digital Inputs”.......................................................59
30.2 I/O sub menu “Analog Outputs”...................................................60
30.3 I/O sub menu “Relay” ..................................................................60
30.4 I/O sub menu “Data storage / USB” ............................................61
30.5 I/O sub menu “Measuredata” ......................................................62
30.6 I/O sub menu “v-Sensor”.............................................................63

Instruction Manual
NIVUS Full Pipe
page 6 NFP - rev. 04 / 23.05.2018
30.7 I/O sub menu “v-Sensor noise” ...................................................64
30.8 I/O sub menu “v-Gate”.................................................................64
31 Calibration and Calculation Menu (CAL).............................................65
31.1 CAL Menu “Velocity” ...................................................................65
31.2 CAL Menu “Analog Outputs”.......................................................66
31.2.1 Basic Information on Simulation...................................................... 66
31.3 CAL Menu “Relays” .....................................................................67
31.4 CAL Menu “Simulation” ...............................................................68
Parameter Tree / Menus available 69
Troubleshooting 74
Maintenance and Cleaning 77
32 Maintenance........................................................................................77
32.1 Maintenance interval ...................................................................77
32.2 Customer Service Information.....................................................77
33 Cleaning..............................................................................................78
33.1 Transmitter ..................................................................................78
33.2 Sensors .......................................................................................78
34 Dismantling /Disposal..........................................................................78
35 Installation of spare parts and parts subject to wear and tear............79
36 Accessories.........................................................................................79
Index 80
Approvals and certificates 82

General
NFP - rev. 04 / 23.05.2018 page 7
General
1 About this manual
Important Note
READ CAREFULLY BEFORE USE!
KEEP IN A SAFE PLACE FOR LATER REFERENCE!
This instruction manual is required for the Installation of the flow measurement transmitter
NIVUS Full Pipe and is for the intended use of the device. This manual is oriented exclusively
to qualified expert personnel.
Read this instruction manual carefully and completely prior to installation and connection
since it contains relevant information on this product. Observe the notes and particularly fol-
low the warning notes and safety instructions.
Keep this manual in a safe place and make sure it is available for the users of this product at
any time.
If you should have problems to understand information contained within this instruction man-
ual either contact the manufacturer or one of the distributors for further support. The manu-
facturer cannot be held responsible for damage to persons or material due to incorrectly un-
derstood information in this instruction.
In case of selling the instrument this instruction manual shall be provided to the purchaser
since it is a part of the standard delivery.
1.1 Applicable documentation
For the installation and operation of the complete system extra instruction manuals or tech-
nical descriptions may be required apart from this manual.
•Technical Instructions for Correlation Sensors and Electronic Box
•Installation Instruction for Correlation and Doppler Sensors
These manuals are provided with the auxiliary units or sensors and/or are available as down-
load on the NIVUS homepage.

Instruction Manual
NIVUS Full Pipe
page 8 NFP - rev. 04 / 23.05.2018
1.2 Signs and definitions used
Image
Meaning
Remark
(Action) Step Action to be performed by you.
Note the numbering of action steps. Observe the
order of the working steps!
Cross-reference Reference to further or detailed information.
>Text< Parameter or Menu Indicates a parameter or a menu that is selected
or described.
Reference to docu-
ment Refers to an accompanying documentation.
Table 1 Structural elements within the manual
1.3 Abbreviations used
1.3.1 Colour code for wires and single conductors
The abbreviations of colours, wire and components follow the international colour code ac-
cording IEC 757.
BK black RD red TR transparent
BU blue WH white GNYE green/yellow
GN green YE yellow BN brown
GY grey PK pink

Safety Instructions
NFP - rev. 04 / 23.05.2018 page 9
Safety Instructions
2 Used symbols and signal words
2.1 Valuation of the accident level
The general warning symbol indicates the risk of personal injuries or death. In the text sec-
tion the general warning symbol is used in conjunction with the signal words described
below.
DANGER Warnings in high degree of risk
Indicates a high-risk, imminently hazardous situation which will result in death or serious
injury if not avoided.
WARNING Warnings in medium degree of risk
Indicates a possible danger with medium risk which may result in a life-threatening situa-
tion or (severe) bodily injury if it is not avoided.
CAUTION Warnings in low-risk or property damages
Indicates a possible danger with moderate risk which may result in minor or moderate per-
sonal injury or material damage if not avoided.
WARNING Danger by electric voltage
Indicates a hazard with a high risk of electric shock which may result in a life-threatening
situation or (severe) bodily injury if it is not avoided.
Important Note
Contains information that should be highlighted.
Indicates a potentially damaging situation which can result in a damage of the product or an
object in its environment.

Instruction Manual
NIVUS Full Pipe
page 10 NFP - rev. 04 / 23.05.2018
Note
Contains information and facts.
2.2 Warning notices on the product (option)
General warning label
This symbol is for operators to refer to this instruction manual.
Observing the information contained therein is required in order to maintain protection
measured provided by the instrument during installation procedures and operation.
Protective conductor
This symbol refers to the protective conductor of the unit.
Depending on the mode of installation the instrument shall be operated solely connected to
an appropriate protective conductor according to applicable laws and regulations.
3 Safeguards and Precautions
Working with NIVUS instruments requires to observe and to follow the safety
measures and precautions below generally and at any time. These notes and warnings
will not be repeated for each description within the document.
WARNING Germ contamination
Please note that due to the operation in the waste water field the measurement system and
cables may be loaded with dangerous disease germs. Respective precautionary measures
must be taken to avoid damage to one’s health.
Wear protective clothing.
WARNING
Observe occupational safety regulations
Before starting installation work, observing the work safety regulations need to be checked.
Disregarding may lead in personal injury.
WARNING Do not disable safety devices
It is strictly prohibited to disable the safety devices or to change the way they work.
Disregarding may lead in personal injury.

Safety Instructions
NFP - rev. 04 / 23.05.2018 page 11
WARNING Disconnect the systems from mains
Maintenance, cleaning and/or repairs (by qualified personnel only) may only be performed
when de-energised.
Disregarding may lead to electric shocks.
Putting into operation by trained experts only
The entire measurement system shall be installed and put into operation by trained expert
personnel only.
Integrated buffer battery
The exchange of the integrated buffer battery shall be carried out by NIVUS staff or per-
sonnel authorised by NIVUS only. Otherwise the guarantee expires.
4 Liability disclaimer
The manufacturer reserves the right to change the contents of this document including this li-
ability disclaimer without prior notice and cannot be held responsible in any way for possible
consequences resulting from such changes.
For connection, initial start-up and operation as well as maintenance of the unit the following
information and higher legal regulations of the respective country (in Germany e. g. VDE reg-
ulations) such as applicable Ex regulations as well as safety requirements and regulations in
order to avoid accidents shall be observed.
All operations on the device which go beyond installation or connection measures in principle
shall be carried out by NIVUS staff or personnel authorised by NIVUS due to reasons of safe-
ty and guarantee.
Operate the transmitter only in technically perfect working order.
Improper Use
Not being operated in accordance with the requirements may impair the safety. The manufac-
turer is not responsible for failures resulting from improper use.
5 Use in accordance with the requirements
Note
The instrument is intended solely for the purpose described below.
Modifying or using the instruments for any other purposes without the manufacturer’s writ-
ten consent will not be considered as use in accordance with the requirements.
The manufacturer cannot be held responsible for any damage resulting from improper use.
The user alone bears any risk.
The permanent flow meter Type NFP - NIVUS Full Pipe including the accompanying sensors
is designed for continuous flow measurement of slight to heavily polluted media in permanent
full pipes.

Instruction Manual
NIVUS Full Pipe
page 12 NFP - rev. 04 / 23.05.2018
The flow meter is designed and manufactured in accordance with the current state of the art
and with the recognised safety rules and regulations applicable at the time this document is
issued. Danger to persons or material damage cannot be completely ruled out, however.
The maximum permissible limit values as specified in chapter “15 Specifications” shall be
necessarily observed. Any case varying from these conditions which is not approved by
NIVUS GmbH in written form is left at the owner’s risk.
Ex protection
The Ex-version of the transmitter is designed to be used in areas with explosive atmospheres
(zone 1).
Approval measurement transmitter: II (2) G [Ex ib Gb] IIB
WARNING Risk of personal injury due to explosion hazard
Install the transmitter out of Ex zones!
The Ex approval of the sensors is part of the “Technical Description” and/or “Instructions
Manual”.
The approval is valid only in conjunction with the according identification on the transmit-
ter’s or sensor’s nameplate.
The Ex approval of the sensors is part of the “Technical Instructions for Correlation Sensors
and Electronic Box”.
Conformity certificates and test certificates
For installation and commissioning the conformity certificates as well as the test certificates
issued by the respective authorities shall be followed.
The Ex version of the NFP is adjusted solely to NIVUS pipe sensors type POA regarding
the intrinsically safe system review according to EN60079-25.
In the event of using sensors by third-party manufacturers the operator shall carry out a
system review according to EN 60079-25!
The required specifications of the Ex version of the NFP can be found in the according EC
type examination certificate IBExU07ATEX1081.

Safety Instructions
NFP - rev. 04 / 23.05.2018 page 13
6 User’s Responsibilities
Important Note
In the EEA (European Economic Area) national implementation of the framework directive
89/391/EEC and corresponding individual directives, in particular the directive 2009/104/EC
concerning the minimum safety and health requirements for the use of work equipment by
workers at work, as amended, are to be observed and adhered to.
In Germany e. g. the Industrial Safety Ordinance must be observed.
Make sure to have a local operating permit available and observe the associated conditions.
In addition to this you must observe environmental requirements and local laws on the follow-
ing points:
•Personnel safety (accident prevention regulations)
•Safety of work materials and tools (safety equipment and maintenance)
•Disposal of products (laws on wastes)
•Disposal of materials (laws on wastes)
•Cleaning (cleansing agents and disposal)
Connections
Operators shall make sure prior to operating the instrument that during installation and initial
start-up the local regulations (such as regulations for electrical connection) are observed.
7 Personnel requirements
Installation, commissioning and maintenance shall be executed only by personnel meeting
the demands as follows:
•Expert personnel with relevant training an appropriate qualification
•Personnel authorised by the plant operator
Qualified personnel
within the context of this documentation or the safety notes on the product itself are per-
sons who are sufficiently familiar with installation, mounting, starting up and operation of
the product and who have the relevant qualifications for their work; for example:
I. Training, instruction or authorisation to activate/deactivate, isolate, ground, and
mark electric circuits and devices/systems according to the safety engineering
standards.
II. Education and instruction according to the standards of safety engineering regard-
ing the maintenance and use of adequate safety equipment.
III. First aid training

Instruction Manual
NIVUS Full Pipe
page 14 NFP - rev. 04 / 23.05.2018
Delivery, Storage and Transport
8 Delivery
The standard delivery of the transmitter NFP contains:
•Measurement transmitter NFP
•The instruction manual including the certificate of conformity and approvals. It contains
any relevant information on how to operate the measurement system.
Check additional accessories such as sensor, welding nozzle, tapping saddle, ball valve etc.
depending on your order and according to the delivery note.
9 Reception inspection
Check the packaging for visible damage immediately after receipt. Any possible damage in
transit shall be instantly reported to the carrier. Furthermore a written report shall be sent to
NIVUS GmbH in Eppingen.
Incomplete deliveries shall be reported in writing either to your local representative or directly
to the NIVUS head office in Eppingen within two weeks.
Note
Mistakes cannot be rectified later.
10 Storing
Observe the minimum and maximum values on environmental conditions such as tempera-
ture and humidity according to chapter “15 Specifications”.
The NFP shall be protected from corrosive or organic solvent vapours, radioactive radiation
as well as strong electromagnetic radiation.
Always store the instrument in its original packaging.
11 Transport
Do not expose the system to heavy shocks or vibrations.
Use the original packaging for transport.
12 Return
In case of a required reshipment return the unit at customer cost to NIVUS GmbH in
Eppingen using the original packaging.
Insufficiently franked shipments will not be accepted.

Product specification
NFP - rev. 04 / 23.05.2018 page 15
Product specification
13 Overview
1 Clear view door
2 Graphic Display
3 Keypad
4 Preparation for Cable Glands
5 Terminal Clamp Housing
6 USB-A-Interface
Fig. 13-1 Overview
14 Device identification
The instructions contained within this manual are valid only for the type of device specified on
the title page. The name plate is fixed on top of the enclosure and contains the following:
•Name and address of the manufacturer
•CE label
•Information on type and series, serial no. if available
•Year of manufacture: the first four digits of the serial number represent the year and the
week number of manufacture (1821NFP….)
•Additional Ex identification for Ex-version devices (as mentioned in chapter “5 Use in
accordance with the requirements”)
In case of enquiries and ordering replacement parts it is important to specify article number
as well as the serial number of the respective transmitter or sensor. This ensures correct and
quick processing.
Check the nameplate
Check the delivered instrument for accordance with your order by identifying the name-
plate.

Instruction Manual
NIVUS Full Pipe
page 16 NFP - rev. 04 / 23.05.2018
Check the nameplate for correct specification of the power supply.
The EU-Type examination certificate (incl. appendix) and the declaration of
conformity are located at the end of the manual.
Fig. 14-1 Nameplates AC/DC (Ex versions)
15 Specifications
Power supply
100…240 V AC, +10 % / -15 %, 47…63 Hz
24 V DC, ±15 %, 5 % residual fluctuation
Power consumption
AC: max. 18 VA, typ. 7 VA
DC: max. 15 W, typ. 6 W
Wall mount enclosure
Material:Polycarbonate
Weight: approx. 1200 g
Protection
IP65
Operating conditions
Protection class I
Overvoltage category II
Pollution degree 2
Altitude
AC unit for use in altitudes up to 3000 m above MSL.
At relay voltages >150 V the use is restricted to an altitude
of max. 2000 m (AC and DC units)
Operating temperature
-20…+60 °C / for Ex: -20…+40 °C
Storage temperature
-30…+70 °C
Max. humidity
90 %, non-condensing
Display
Back-lit full graphic LCD, 128x64 pixel
Operation
6 keys, menu driven in German, English, French and Polish
Inputs
1x digital input
1x flow velocity pipe sensor type POA connectable
Outputs
1/3x 0/4…20 mA, load 500 Ohm, 12 bit resolution,
accuracy 0,1 % (after calibration)
2x relays SPDT, maximum load 230 V AC / 2 A (cos.ϕ 0,9)
Data memory
4 MB, 64512 data points, for programming and readings
memory; read-out via front-side USB stick
Storage cycle
1 minute to 1 hour
Ex approval (option)
II (2) G [Ex ib Gb] IIB

Product specification
NFP - rev. 04 / 23.05.2018 page 17
Rated voltage
264 V AC
Sensor circuits
Ignition protection type Ex ib IIB
Sensor power supply per
channel (z. B. flow velocity
sensor POA V2)
Terminal no. 50…52
Uo9,9 V
Io629 mA
Rectangular characteristics
Co5 μF
Lo0,15 mH
Sensor communication
interface per channel
Terminal no. 53…54
Uo9,9 V
Io130,3 mA
Po322 mW
Linear characteristics
Co9,7 μF
Lo0,15 mH
Ui10,1 V
Ii136 mA
The maximum values also apply to concentrated capacitance / inductors that can be
switched on.
Table 2 Specifications
The EU-Type examination certificate (incl. appendix) and the declaration of
conformity are located at the end of the manual.
Sensors
Observe the specifications of the according sensors as described in the respective instruction
manuals or technical descriptions.

Instruction Manual
NIVUS Full Pipe
page 18 NFP - rev. 04 / 23.05.2018
16 Configuration
16.1 Device Types
The transmitter NFP is available in different versions. The transmitters primarily vary in terms
of maximum pipe diameter, power supply, Ex-protection as well as the number of analog out-
puts.
The current type of device is indicated by the article number, which can be found on a weath-
erproof label on the bottom of the enclosure. From this article key the type of device can be
specified.
NFP
Type
25
Nominal diameter DN100…DN500
28
Nominal diameter DN100…DN800
Construction
W0
Wall mount enclosure IP65
Power supply
AC
85…265 V AC
DC
20…28 V DC
ATEX approval
0
none
E
Intrinsically safe sensor supply in Ex-Zone 1
Extensions
1
1 analog output
3
3 analog outputs
NFP
W0
Table 3 Type key for NFP transformer

Functional Principle
NFP - rev. 04 / 23.05.2018 page 19
Functional Principle
17 Operating Range
The NFP is a permanent measurement system for flow measurement of slightly to heavily
polluted media with various consistencies in constantly full filled pipes.
Important Note
The measurement method used for flow velocity detection is based on the ultrasound re-
flection principle.
Hence, it is indispensable for the system to work that there are particles in the water, which
are able to reflect the ultrasonic signal sent by the sensor (dirt particles, gas bubbles or
similar).
18 Flow Velocity Measurement
The piezo crystal has a slope to the flow direction and operates as a flow velocity sensor.
Here an ultrasonic burst with a defined angle is sent into the medium. All the particles in the
measurement path (air, dirt) reflect a small amount of the ultrasonic signal. The result is a
particular signal depending on shape and size of the particles.
Hence, the multitude of the reflected signals results in a reflection pattern (see Fig. 18-1). The
piezo crystal receives this pattern again, which then is converted into electric signals and will
be saved in a built-in digital signal processor (DSP). This DSP is located within the active
pipe sensor.
1 Pipe wall
2 Pipe sensor
E1…E7 Reflecting particles
1, 2, 3, 4, n Measurement window
Fig. 18-1 Situation at first signal detection

Instruction Manual
NIVUS Full Pipe
page 20 NFP - rev. 04 / 23.05.2018
After a certain period a second ultrasonic burst is sent into the medium (see Fig. 18-2). The
newly generated reflection signal is saved in the DSP too.
In various sections there are different flow velocities, where in case of full filled pipes and suf-
ficient calming sections the maximum velocity is in the centre of the pipe. Depending on me-
dium viscosity, flow velocity and wall roughness this velocity decreases approaching the pipe
wall (flow velocity profile, see Fig. 18-4).
The reflecting particles have moved away from the first measurement point by varying dis-
tances due to the flow velocity profile prevailing in the pipe. Hence, a distorted reflection pat-
tern results (see Fig. 18-3). At the same time slightly different reflections will emerge addi-
tionally resulting from the fact that some particles have been turning around and thus have
another shape of reflection; some particles are no longer within the measurement range and
others have now moved into the measurement range.
1 Pipe wall
2 Pipe sensor
E1…E7 Reflecting particles
1, 2, 3, 4, n Measurement window
Fig. 18-2 Situation at second signal detection
The DSP checks both the received reflection patterns for similarities using the cross correla-
tion method. All existing signal differences are rejected so that two similar but temporarily off-
set signal patterns are left for velocity evaluation.
Depending on the inner pipe diameter set an overlay consisting of up to 16 measurement
windows of different lengths will be added. Then, in each measurement window the lag Δt of
the signal pattern is investigated (see Fig. 18-3).
Other manuals for NFP
1
Table of contents
Other Nivus Measuring Instrument manuals

Nivus
Nivus NivuFlow Stick User manual

Nivus
Nivus osm-f Technical manual

Nivus
Nivus NivuLink Compact NLC0CLOG User manual

Nivus
Nivus OCM PRO CF User manual

Nivus
Nivus NivuFlow Mobile 600 User manual

Nivus
Nivus OCM PRO CF User manual

Nivus
Nivus NivuSonic Clamp On User manual

Nivus
Nivus NivuCont Plus User manual

Nivus
Nivus NFP User manual

Nivus
Nivus NivuCompact User manual