NJORD Atle 180 User manual

Atle 180
INSTRUCTIONS FOR USE
Atle 180 (20,000) IFU EN Rev J - 10/2022
®

II
To avoid injury, always read these Instructions for Use and accompanied documents before
using the product.
Mandatory to read the Instructions for Use
Design policy and Copyright
® and TM are trademark of Njord Medtech.
© Njord Medtech 2021
As our policy is one of continuous product development and we therefore reserve the right
to change/modify designs/constructions and specications without prior notice. The content
of this publication/document may not be copied either whole or in part without permission/
consent from Njord Medtech.

III
Table of contents
Foreword ����������������������������������������������������������������������������������������������������������1
Read this manual carefully! ������������������������������������������������������������������������������������������ 1
Safety Instructions ������������������������������������������������������������������������������������������2
Symbol used ����������������������������������������������������������������������������������������������������4
Product description�����������������������������������������������������������������������������������������5
Use of Atle® 180����������������������������������������������������������������������������������������������6
General horizontal transfer of patient ������������������������������������������������������������������������ 6
Transfer at bowl-shaped CT-scanner �������������������������������������������������������������������������� 8
Attachment solution with hooks in a stretcher or trauma mattress�������������������������� 8
Sheet Attachment Solution ������������������������������������������������������������������������������������������ 9
Charging and Hand Control���������������������������������������������������������������������������������������� 10
Buttons������������������������������������������������������������������������������������������������������������������������� 11
Brakes and Emergency Stop �������������������������������������������������������������������������������������� 11
Cleaning and service�������������������������������������������������������������������������������������13
Cleaning of Atle ����������������������������������������������������������������������������������������������������������� 13
Replacement of the wire ��������������������������������������������������������������������������������������������� 13
Maintenance ��������������������������������������������������������������������������������������������������������������� 15
Atle® Labels ��������������������������������������������������������������������������������������������������18
Technical Specication ���������������������������������������������������������������������������������19
Product data ���������������������������������������������������������������������������������������������������������������� 19
Dimensions ������������������������������������������������������������������������������������������������������������������ 21
Electromagnetic Compatibility���������������������������������������������������������������������22
Electromagnetic Compliance ������������������������������������������������������������������������������������� 22
Electromagnetic Compatibility���������������������������������������������������������������������23
Electromagnetic Emissions ���������������������������������������������������������������������������������������� 23
Electromagnetic Immunity ����������������������������������������������������������������������������������������� 24
Electromagnetic Immunity ����������������������������������������������������������������������������������������� 25

1
Foreword
Read this manual carefully!
The information in this manual is crucial for
proper use and maintenance of Atle® 180.
It will help protect your product as well as
ensure that it performs to your satisfaction.
Transferring a person always constitutes a
potential risk. This document contains safety
related information that must be read and
understood to help prevent injuries.
Njord Medtech strongly recommends and
warns that to avoid injuries that can be
attributed to the use of inadequate parts, only
parts designated by Njord Medtech should
be used on products and other appliances
supplied by Njord Medtech.
Unauthorized modications of Atle® 180
may affect its operational and functional
performance. Njord Medtech will not be held
responsible for accidents, incidents or lack
of performance that occurs as a result of any
unauthorized modication to the product.
If a serious incident occurs during the use
of Atle® 180, affecting the user, or patient
then the user or the patient should report
the serious incident to the manufacturer or
distributor. In the European Union, the user
should also report the serious incident to the
Competent Authority in the member state
where they are located.
Manufacturer Information
This product is manufactured by:
Njord International AB
Medicinaregatan 8
413 90 Göteborg
Sweden
hello@njordmedtech.com
www.njordmedtech.com/
Denitions used in this manual
Means: Failure to understand and follow
instructions may result in injury to yourself
and others.
Means: Important information regarding
correct use of the product. Failure to follow
instructions may cause damage to the
product.
NOTE:
WARNING:

2
Keep the manual together with the product
so that it can be used as reference. Store the
quick reference guide that comes with the
product well visible to all staff.
Atle is a medical device for horizontal patient
transfer and used in healthcare in which the
transfers are made from one care bed to
another examination table or equivalent.
Intended Use
Atle® 180 has been designed to assist
caregivers in hospitals and should only be
used by Professional healthcare
facility environment.
The product is not intended for wet areas
and should not be used inside MRI (magnetic
resonance) examination rooms, i.e. not in
proximity to a magnetic eld.
WARNING: The product is not intended to
be used on inside MRI examination rooms
Atle should be used in an indoor
environment.
There is no need for medical device specic
training for professional healthcare users to
use Atle® 180.
Operational Life
Atle® 180 has been designed and tested to
achieve up to 40 000 cycles with a load of
180 kg.
Maintenance according to the specications
in the section“Cleaning and Service” in this
manual applies.
Important Safety Instructions
The batteries that power Atle® 180 are
rechargeable and may not be replaced.
The batteries are charged via ordinary main
plugs. Atle® 180 should not and cannot be
used during charging. To avoid a fall accident
the caregiver should be attentive to the cord
and its placing during charging.
The wheels of Atle® 180 can be rotated 360
degrees to enable a smooth handling of the
device. The wheels are for safety reasons
possible to lock and should be locked while
the device is not intended to be in motion.
Use only accessories intended for Atle.
Atle should only be used to transfer patients
horizontally between the beds. It is not
intended to be used for other transfers e.g.
move the patient to the side. Atle is not for
moving any other type of object.
Safety Precautions
Do not position the system so that it is
difcult to operate the disconnection device.
The power cord to the control unit should
be positioned to avoid a tripping hazard and/
or damage to the cord.
Manufacturer will prove circuit diagrams,
component part list, descriptions to assist to
service personnel in parts repair.
Uplug the power cord charger from the
control panel to disconnect the device from
supply mains.
Safety Instructions
NOTE: Atle should only be used for two
horizontal aligned surfaces
WARNING: Using Atle® 180 beyond its
life span may result in an incident causing
serious injuries. WARNING: No modication of this
equipment is allowed

3
Before use
Before rst use, check that all parts
are included according to the product
description, clean Atle® 180 and test the
emergency stop button. Note that the battery
must be charged at least 24 hours before rst
use.
In case of deviation
In case of need for immediate shutdown
during transfer, the emergency stop switch
on the hand control is used.
Stop transfer immediately in case of irregular
sound from the engine or irregular speed of
transfer.
Policy on Number of Staff Members
Required for Patient Transfer
Atle® 180 are designed for safe usage with
one caregiver. There are circumstances
that may dictate the need for a two-person
transfer. An example of this is moving the
patient from a CT table to a bed. It is the
responsibility of the caregiver to determine
if a one or two person transfer is more
appropriate.
Safe workload (SWL)
Maximum patient weight for Atle®
180 is 180 kg.
Product label
The main product label has a barcode
ID, serial number and date of manufacture for
the product.
Safety Instructions
1) Between Atle and bed unit
2) Between the wire and pulling member
3) On the hooks
4) When changing bobbin
5) At the wire guide
1
2
5
3
4
Fig 1
WARNING: Do not put ngers, hands or
other body parts where space is limited
(see Fig 1). This could pinch, cut or
seriously harm body parts.
<180 Kg

4
Atle® 180 unit has a traceability product label with the symbols below, the other symbols can
be found on Atle® 180 or in the Instructions for Use.
CE marking indicating
conformity with European
Community harmonies
legislation
Manufacture – This symbol
points out the manufacturer’s
name and address.
Error indication
Battery indication
The Instruction for Use is
recommended to be read
Indicates the product is a
medical device according to the
EU Medical Device Regulation
2017/745
Intended for indoor use only
Points out a risk of pinching
!
The maximum load the device is
tared for safe usage / Maximum
patient weight
Points out the emergency stop
button
Date of manufacture
Class II equipment, double
insulation
Reset Points out the reset button for
the emergency stop
Charging output
Hand
Control Hand control output
STOP
IP 20 Protection class against moisture
<180 Kg
Temperature limitations
Humidity limitations
Serial number
Reference number
Symbol used
Waste electrical and Electronic
Equipment (WEEE) – do not
dispose of this product in general
household or commercial waste
The Instruction for Use must be
read

5
Product description
1) Attachment solution (replaceable) 7) Bobbin (replaceable)
2) Pulling member 8) Wire guide
3) Wire (replaceable) 9) Emergency stop
4) Protective pads 10) Reset button for emergency stop
5) Wheel brake 11) Hand Control output
6) Hand Control with battery indicator 12) Charger output
13) Battery charger and power cord
To disconnect the device, pull out the charger from the appliance and/or pull out the plug
from main socket
Fig 2

6
Before transfer
1) Control that the patient is lying on its back
in a center position and the sheet sticks out
>20 cm from the patient’s shoulder and hips.
To facilitate the transfer sliding plates should
be used. Place optional sliding plates under
the patient, good position is under the feet,
the hips and the shoulders.
2) Prepare the transfer by laying the pulling
member on the sheet, at height with the
shoulder and hip on the patient. Then fold
over the remaining sheet. The pulling member
must lie close to the patient.
3) Place the bed in paralell and around 1/2
meter from the other bed unit.
Sheet attachment solution
4) Go to Atle, pull out both the wires at the
same time and place on the bed unit. Prepare
through open the loops of the attachment
solution.
Attach the loops over the sheet and around
the pulling member. Tighten the loop around
the pulling member so that the loop is
fastened tightly. Do the same on the other
side. (See “fastening device in sheets”, page
9).
NOTE: Sliding plates should be used for a
more secure and comfortable transfer.
Use of Atle® 180
General horizontal transfer of patient
The transfer technique described below can be used for lying patients from bed to bed,
stretcher to bed and bed to examination table.
NOTE: Maximum patients weight at transfer
with Atle is 180 kg.
1
2
3
4
Fig 3
Fig 4
Fig 5
Fig 6

7
During the transfer
5) Make sure that the bed and the
examination table are placed next to each
other and with the same height. Lock the
wheels on the bed. The remaining parts of
the sliding plates must lie over the bed unit.
Place Atle on the long side of the bed or
examination table. Set the height to Atle so
that the position of the wires is 2-3 cm over
the mattress on the examination table, see
gure below. Atle must stand against the
bed and the exits of the wires must be in line
with the ends of the pulling member. Finish
positioning of Atle by locking all wheels.
6) Initiate motorized transfer via hand control.
Press the pull button so that the wires are
stretched, then release the button for a
second, contact the patient and inform that
the transfer starts. Press the drag button
again and perform the transfer.
Make sure that the patient is following the
transfer, especially the feet, shoulders and
head. This is done with the help of one hand
by adjusting the sheets associated with the
transfer.
At the end of the transfer
7) End transfer when the patient is positioned
centered on the new bed unit. Remove the
fastener from the sheet and pulling member
by pulling out the wires slightly and widening
the loop forward through the “float”. Remove
the pulling member and the sliding plates.
Use of Atle® 180
WARNING: Both beds should be positioned
close to each other to avoid the risk of
patient falling between these. For additional
safety measures, sliding plates should be
used.
NOTE: Position Atle against the bed edge
with the position of the wires 2-3 cm over
the mattress on the examination table, to
avoid risk of tipping
WARNING: Avoid moving your arm / hand
between the pulling member and engine
part when the pulling member approaches
engine part at end of the patient transfer to
avoid the squeeze risk
WARNING: To eliminate the squeeze risk,
make sure no body parts of the patients is
between the bed and the examination table
or between bed and Atle
WARNING: To eliminate the squeeze risk,
ensure that the patient does not get stuck
with some body parts in relation to the
pulling member and or the wire during the
transfer
6
Fig 7
Fig 8
5

8
Transfer at bowl-shaped CT-scanner
When moving from the CT table to another
bed it is recommended that two people carry
out the transfer with Atle.
One person monitors and conducts the
transfer with the hand control in the usual
manner while the other person supports the
patient on the other side of the examination
table during the transfer, over the curved
edge of the CT table. The person who
supports the patient can choose to lift at the
edge of the sheet to create a smooth transfer.
Attachment solution with hooks in a
stretcher or trauma mattress
For all types of mattresses with handles along
the side, the hook is attached to the loop of
the wires and then connected directly to the
handles on the mattress.
Fig 11
Use of Atle® 180
NOTE: It is possible to move the entire
mattress to the examination table with the
help of the patient’s weight by attaching as
usual to the sheet and having sliding plates
under the mattress
7
Fig 9
Fig 10
Fig 12
WARNING: The pull should be straight

9
Sheet Attachment Solution
Before moving
1) Prior to patient transfer, check that the
sheet is stretched under the patient and lay
the pulling member on the sheet close to the
patient as possible and with the ends of the
pulling member at the level of the shoulders
and hips. TFold the sheet over pulling
member.
2) Pull out both the wires at the same time
and place on the bed unit, open the loops for
preparation of the attachment solution.
3) Attach the loops around the pulling
member at the ends on the pulling member.
Tighten by advancing the “float” and then pull
on on the wire. Do the same on the other side.
After Transfer
4) When the patient transfer has been
completed, release the wires and disconnect
it from the pulling member.
Fig 13
Fig 14
Fig 15
1
2
3
Use of Atle® 180
4
Fig 16
NOTE: The pulling member must be placed
close to the patient for the patient to end up
in the middle of the bed/examination table

10
Charging and Hand Control
Charging Batteries
The battery pack is the type of rechargeable
Lead. The batteries are changed via the main
power outlet. The Atle® 180 used a 24 Volts
that can deliver around 50 pulls per charge
with a new battery. The battery life is variable
(1-2 year) and is influenced by: Frequency of
use and weight of patient.
The battery needs to be charged when it is
empty. Only the supplied charger (Mascot
2241P) should be used. For saftey reasons, ie
to avoid tripping over the cable, Atle cannot
be used underneath charge. Atle 180 is than
de-energized by unplugging the power cord
from the mains. The hand control indicates
that the battery is being charged by traveling
lights for charging status.
It takes around six hours to charge the
battery. Note that the battery must be charged
at least 24 hours before rst use. Battery life
is not adversely affected by the fact that the
battery is charged frequently. Replacement
by the battery should be done by medically
trained staff or by the manufacturer service
staff.
Hand Control
If the light on the right lights up red, it is a
indication that something is wrong (error).
The LED is permanently lit if there is an
internal hardware fault in the control box. On
the other hand, if an error is caused by the
user, e.g. that you try to run a function when
the batteries are charging, the alert LED only
lights up while you are trying to activate a
function (i.e. when you press a button).
Battery indicator
When the light on the left is red, Atle must be
charged (you can drag X number of times).
When one light is green, the battery charge
level is less than 33%, when two lights are
green, the battery charge level is 33% or
better, and when all lights are green, the
battery charge level is 66% or better.
1) On / off button
2) On / off button indicator
3) Reset button for the overcurrent protection
4) Lower and raising buttons
5) Motorized pull button
6) Error indication
7) Battery level
8) Battery indicator
Only use the supplied hand control.
Use of Atle® 180
NOTE: The battery must be charged at
least 24 hours before rst use.
Fig 17
WARNING: Replacement by the battery of
inadequately trained personnel could result
in a hazard
Fig 18

11
Use of Atle® 180
Buttons
Pulling button
The button with double arrows are used to
pull the patient towards Atle. The transfer will
keep pull as long as the button is pressed or
until it reaches its limit.
Raising / lowering buttons
The up and down arrows are used to raise
and lower the motor housing. The motor
housing will be be raised or lowering as long
as the button is pressed or until it reaches its
limit.
Reset button
When Atle is exposed to a too high load,
an overcurrent protection is triggered. To
reset the overcurrent protection, the user is
required to press the reset button. The user
can then press the pull button again to move
on.
On/off button
When the green is light up the device is on
and ready to use.
Brakes and Emergency Stop
Brakes
Foot operated brakes are tted on all the four
wheels.
To apply brakes, step on the back portion of
pad.
To release brakes, push the top portion of the
pad forwards.
Emergency stop
The user can shut off the power at any time
by pressing the red emergency button on
the control panel. Reset the emergency stop
function by turn the emergency stop button
and click on the reset button next to the red
emergency button.
First-time users should practice the
emergency stop manoeuvre before operating
the device with a patient.
Fig 19
Fig 20
Fig 21
Fig 23
NOTE: When moving heavier patients, it is
very important to lock the wheels, especially
the wheels in the front
WARNING: If the wheels are not locked,
there is a risk of Atle tipping over
unlocked
locked
NOTE: Only the supplied charger and hand
control should be used
Fig 22
WARNING: Periodic checking or
replacement of the charger is neccesarily

12
After Use
When the expiration date of the device is
reached, the product must be recycled
according to set standards for each country
and hospital.
Use of Atle® 180
Fig 24
Emergency stop Reset button Hand Control
Output
Charger
Output

13
Cleaning of Atle
Surfaces that the caregiver touches must
be cleaned manually after each transfer.
Clean the surfaces of Atle regularly with one
time use cloth with hot water or pH neutral
detergent. In case of disinfection use the
regular surface disinfectant.
Pay attention to parts that are most likely to
be touched:
• The hand control
• The control box
• The handle
• Attachment solution
• The wires
Replacement of the wire
The wire and the bobbin should be replaced
with a new one in case of suspicion or risk of
contamination, visible damage to the surface
layer on the wire or discoloured wire.
Replacement of the bobbin is done as follows.
1) Open the cover to the bobbin by inserting
a nger into the opening on the side of the
cover
2) Take out the wire from the wire guide, pull
the wire to the side and through the opening.
3) Remove the bobbin by unscrewing the
screw and pulling out the bobbin.
Fig 25
Fig 27
1a
1b
Cleaning and service
WARNING: if the bobbin and wire are
not changed when there is a suspicion of
contaminates there is a risk of spreading a
infection
3a
Fig 26
Fig 28
2

14
4) Take a new bobbin and attach it by pushing
it into the pin. Finish by tightening the screw
again.
5) Put the wire back into the wire guide.
6) Close the cover
Cleaning and service
5
Fig 31
4
6a
Fig 30
Fig 32
Fig 33
NOTE: If the bobbin is placed incorrectly,
the bobbin can fall off the pin and the
patient transfer cannot be performed
Fig 29
3b
6b
WARNING: Do not use Atle®180 when
changing bobbin to avoid squeezing risk

15
Maintenance
Replacement of components must be done by medically trained staff or by the manufacturer
service staff.
In the event of more severe contamination, the cover and other external parts may need to be
unscrewed and the internal parts need to be cleaned by medical staff.
Preventive service is recommended once per year. This is done either by the company’s
medical technical staff at the hospital or by the manufacturer’s service technicians. Key
areas are engine maintenance, pull test and condition of the wire. After the service has been
performed, a functional test is done without a patient.
Atle 180 is subject to wear and tear, and the following actions must be performed when
specied to ensure that the product remains within its orginal manufacturing specication.
Cleaning and service
3) Press the emergency stop button and make sure that all electrical power is cut off.
No action should occur when activation the buttons on the hand control
4) Check all the functions on the hand control. Ensure that the hand control membrane
is intact
9) Check all the functions on the control box
10) Press the reset button for the emergency stop button to see that all electrical
power is on again
5) Check that Atle can be switched on and off with the power on / off button on the
hand control
7) Check that Atle can be raised and lowered by pressing the hand control up and
down button
6) Check that Atle pulls by pressing the hand control pull button and that it stops when
you release the pull button, the pulling member moves towards Atle when it pulls
8) Control the reset button on the controller by pressing the reset, you can drag with
Atle by pressing the pull button
Points to be inspected by user/service technician
Frequency
Before every use
1) Ensure that the battery charge indicator is within the normal range
Annually
Initially
X
XX
X X
X X
X X
X
XX
X
2) Check that Atle is charging when a battery charger is connected to the Atle, the
charging status LEDs on the controllers start to wander. If the battery does not receive
a charge or a short operating time on the battery, a replacement of the battery is
needed, see points for service technician
X
XX
Note: The list continue on the other side
NOTE: Do not use Atle while serviced or maintained

16
11) Verify the proper functioning of the four wheels, check the brakes and that they
spin well
1) Check all bolts, nuts and locknuts to ensure they are tight
2) If the product does not work as intended, immediately contact Njord Medtech for
support
13) Check that the ropes can be pulled out of Atle by hand without signicant
resistance and that the end of the ropes does not come loose when the ropes are fully
extended
15) Check that the ropes are not damaged and open the covers over the bobbins and
check that the bobbins are not damaged, if damage see “Replacement of the wire”
16) Check the home position sensors indicates that the pulling member is in home
position by pulling the pulling member all the way into the Atle then Atle stops pulling
automatically (if engine noise is heard even when the pulling member is at home the
home position sensors do not work)
4) Check that the Atle is charging when a battery charger is connected to the Atle, the
charging status LEDs on the controllers start to wander. If the battery does not receive
a charge or a short operating time on the battery, replace the battery*
3) Connect a dynamometer on the opposite side of the bed to Atle, connect the pulling
member / hooks to the dynamometer press pull on the hand control read that Atle
gives 700N -50N + 100N before the overcurrent protection trips
14) Check that the ropes are completely loose and can be pulled out
X X
X X
XX
X
X
X
X
X X
12) Check front and rear wheels regularly for hair and debris; clean when necessary X
Points to be inspected by service technician
Frequency
Before every use
Annually
Initially
XX
Cleaning and service
*The battery (REAC power pack RPP10) can be order from the responsible seller

17
Trouble shooting
Cleaning and service
Type of error Possible cause Measure
The wire cannot be pulled out - The wire is stuck in the
bobbin
- Summon medical technically
trained personnel or Service
technician
The pulling member cannot be
attached
- The attachment solution has
been put on incorrectly
- The sheet is two fold or too
thick
- The attachment solution is
worn out
- Adjust according to
instructions under “Sheet
attachment solution”
- The battery is discharged
- The main power switch is
rejected
- The emergency stop has been
activated
- Overload: the patient is too
heavy
- The patient weighs more than
180 kg
- The mattress texture gives a
very high friction
- The mattress structure and
material allows the patient to
create too much friction
The torque/engine stops
Error indication on the hand
control glow red (indicates
charging, charging level and
error indication)
- Problem with power supply or
motor.
- Charge the battery (connect
the charger to the device and
the wall socket see that the
charge indicator lights up on the
control, it is a wandering light
The torque/engine does not
start via the hand control
- Charge Atle
- Turn on the main power switch
- Deactivate the emergency
stop by press the “reset” button
on the control box
- Atle is intended for patient
which weights a maximum of
180 kg
- Performs manual transfer
- The stop measure stops the
machine over 700N
- Try to turn off the main switch
and turn it on again.
If the error persists, summon
medical technology trained
personnel or service technician
- Reset the device with the
“reset” button on the control
box
Hand control does not respond - The emergency stop button is
activated
- The plug to the device is
unplugged
- Battery need to be changed
- Deactivate the emergency stop
by click on the “reset” button on
the control box
- Summon medical technically
trained personnel or Service
technician
Sound from the control box -
error indication
- Emergency stop button has
been activated
- The overcurrent protection
can be tripped due to the wire
getting stuck or tangled or too
high load for the pulling
- The height adjustment can
be stuck in the bed so the
overcurrent protection is
triggered
- Deactivate the emergency stop
by turn the emergency stop
button and than press the
“reset”button
Other manuals for Atle 180
1
Table of contents
Other NJORD Medical Equipment manuals