NJORD Atle 180 User manual

Atle 180
INSTRUCTIONS FOR USE
Atle® 180 (20,000) IFU EN Rev K - 06/2023
®

II
To avoid injury, always read these Instructions for Use and accompanied documents before
using the product.
Mandatory to read the Instructions for Use
Design policy and Copyright
® and TM are trademark of Njord Medtech.
© Njord Medtech 2021
As our policy is one of continuous product development and we therefore reserve the right
to change/modify designs/constructions and specications without prior notice. The content
of this publication/document may not be copied either whole or in part without permission/
consent from Njord Medtech.

III
Table of contents
Foreword����������������������������������������������������������������������������������������������������������1
Read this manual carefully! ������������������������������������������������������������������������������������������ 1
Safety Instructions ������������������������������������������������������������������������������������������2
Symbols used ��������������������������������������������������������������������������������������������������4
Product description�����������������������������������������������������������������������������������������5
Use of Atle® 180����������������������������������������������������������������������������������������������6
General horizontal transfer of patient ������������������������������������������������������������������������ 6
Transfer at bowl-shaped CT-scanner ������������������������������������������������������������������������� 8
Attachment solution with hooks in a stretcher or trauma mattress�������������������������� 8
Attachment Solution ���������������������������������������������������������������������������������������������������� 9
Charging and Hand Control���������������������������������������������������������������������������������������� 10
Buttons������������������������������������������������������������������������������������������������������������������������� 11
Brakes and Emergency Stop�������������������������������������������������������������������������������������� 11
Cleaning and service�������������������������������������������������������������������������������������13
Cleaning of Atle® 180 ������������������������������������������������������������������������������������������������� 13
Replacement of the wire��������������������������������������������������������������������������������������������� 13
Maintenance ��������������������������������������������������������������������������������������������������������������� 15
Troubleshooting ���������������������������������������������������������������������������������������������������������� 17
Atle® Labels �������������������������������������������������������������������������������������������������19
Technical Specification���������������������������������������������������������������������������������20
Product data ��������������������������������������������������������������������������������������������������������������� 20
Dimensions ������������������������������������������������������������������������������������������������������������������ 22
Electromagnetic Compatibility���������������������������������������������������������������������23
Electromagnetic Compliance ������������������������������������������������������������������������������������� 23
Electromagnetic Emissions ���������������������������������������������������������������������������������������� 24
Electromagnetic Immunity ����������������������������������������������������������������������������������������� 25

1
Foreword
Read this manual carefully!
The information in this manual is crucial for
proper use and maintenance of Atle® 180.
It will help protect your product as well as
ensure that it performs to your satisfaction.
Transferring a person always constitutes a
potential risk. This document contains safety
related information that must be read and
understood to help prevent injuries.
Njord Medtech recommends and warns
that to avoid injuries attributed to the use of
inadequate parts, only parts designated by
Njord Medtech should be used on products
and other appliances supplied by Njord
Medtech.
Unauthorized modications of Atle® 180
may affect its operational and functional
performance. Njord Medtech will not be held
responsible for accidents, incidents or lack
of performance that occurs as a result of any
unauthorized modication to the product.
If a serious incident occurs during the use of
Atle® 180, affecting the user or patient, the
user or the patient should report the serious
incident to the manufacturer or distributor.
In the European Union, the user should also
report the serious incident to the Competent
Authority in the member state where they are
located.
Manufacturer Information
This product is manufactured by:
Njord International AB
Medicinaregatan 8
413 90 Göteborg
Sweden
hello@njordmedtech.com
www.njordmedtech.com/
Definitions used in this manual
Means: Failure to understand and follow
instructions may result in injury to yourself
and others.
Means: Important information regarding
correct use of the product. Failure to follow
instructions may cause damage to the
product.
NOTE:
WARNING:

2
Keep the manual together with the product
so that it can be used as reference or
provide access to online documentation
at njordmedtech.com. Store the Quick
Reference Guide that comes with the product
well visible to all staff.
Atle® 180 is a medical device for horizontal
patient transfer and used in healthcare in
which the transfers are made from one
care bed to another examination table or
equivalent.
Intended Use
Atle® 180 has been designed to assist
caregivers in hospitals and should only be
used by healthcare professionals in care
environment.
The product is not intended for wet areas
and should not be used inside MRI (magnetic
resonance) examination rooms, i.e. not in
proximity to a magnetic eld.
WARNING: The product is not intended to
be used on inside MRI examination rooms
Atle® 180 should be used in an
indoor environment.
There is no need for medical device specic
training for professional healthcare users to
use Atle® 180.
Operational Life
Atle® 180 has been designed and tested to
achieve up to 40 000 cycles with a load of
180 kg. Corresponding to 4-6 years of use.
Maintenance according to the specications
in the section “Cleaning and Service” in this
manual applies.
Important Safety Instructions
The batteries that power Atle® 180 are
rechargeable and may not be replaced.
The batteries are charged via ordinary main
plugs. Atle® 180 should not and cannot be
used during charging. To avoid a fall accident
the caregiver should be attentive to the cord
and its placing during charging.
The wheels of Atle® 180 can be rotated 360
degrees to enable a smooth handling of the
device. The wheels are for safety reasons
possible to lock and should be locked while
the device is not intended to be in motion.
Use only accessories intended for Atle® 180.
Atle® 180 should only be used to transfer
patients horizontally between the beds. It is
not intended to be used for other transfers
e.g. move the patient to the side. Atle® 180
is not for moving any other type of object.
Safety Precautions
Do not position the system so that it is
difcult to operate the disconnection device.
The power cord to the control unit should
be positioned to avoid a tripping hazard and/
or damage to the cord.
Manufacturer will prove circuit diagrams,
component part list, and product description
to assist the service personnel in service
repair.
Uplug the power cord charger from the
control panel to disconnect the device from
supply mains.
Safety Instructions
NOTE: Atle® 180 should only be used for
two horizontal aligned surfaces
WARNING: Using Atle® 180 beyond its
life span may result in an incident causing
serious injuries

3
Before use
Before rst use, check that all parts
are included according to the product
description, clean Atle® 180 and test the
emergency stop button. Note that the battery
must be charged at least 24 hours before rst
use.
In case of deviation
In need of immediate shutdown during
transfer, the emergency stop switch on the
product is to be used.
Stop transfer immediately in case of irregular
sound from the engine or irregular speed of
transfer.
Policy on Number of Staff Members
Required for Patient Transfer
Atle® 180 is designed for safe usage with
one caregiver. There are circumstances
that may dictate the need for a two-person
transfer. An example of this is moving the
patient from a CT table to a bed. It is the
responsibility of the caregiver to determine
if a one or two-person transfer is more
appropriate.
Safe workload (SWL)
Maximum patient weight for Atle®
180 is 180 kg.
Product label
The main product label has a barcode
ID, serial number and date of manufacture for
the product.
Safety Instructions
1) Between Atle® 180 and bed unit
2) Between the wire and transfer bar
3) On the loops/hooks
4) When changing bobbin
5) At the wire guide
1
2
5
3
4
Fig 1
WARNING: Do not put ngers, hands or
other body parts where space is limited
(see Fig 1). This could pinch, cut or
seriously harm body parts
<180 Kg
WARNING: No modication of this
equipment is allowed
1

4
Atle® 180 unit has a traceability product label with the symbols below, the other symbols can
be found on Atle® 180 or in the Instructions for Use.
CE marking indicating
conformity with European
Community harmonies
legislation
Manufacture – This symbol
points out the manufacturer’s
name and address
Error indication
Battery indication
The Instruction for Use is
recommended to be read
Indicates the product is a
medical device according to the
EU Medical Device Regulation
2017/745
Intended for indoor use only
Points out a risk of pinching
!
The maximum load the device is
tared for safe usage / Maximum
patient weight
Points out the emergency stop
button
Date of manufacture
Class II equipment, double
insulation
Reset Points out the reset button for
the emergency stop
Charging output
Hand
Control Hand control output
STOP
IP 20 Protection class against moisture
<180 Kg
Temperature limitations
Humidity limitations
Serial number
Reference number
Symbols used
Waste electrical and Electronic
Equipment (WEEE) – do not
dispose of this product in general
household or commercial waste
The Instruction for Use must be
read

5
Product description
1) Attachment solution (replaceable) 7) Wire guide
2) Transfer bar 8) Emergency stop
3) Wire (replaceable) 9) Reset button for emergency stop
4) Wheel brake 10) Charger output
5) Hand Control with battery indicator 11) Hand Control output
6) Bobbin (replaceable) 12) Battery charger and power cord
To disconnect the device, pull out the charger from the appliance and/or pull out the plug
from main socket
Fig 2

6
Before transfer
1) Ensure the patient is lying on hirs back in a
centered position with sheets extending >20
cm from the patient’s shoulder and hip.
To facilitate the transfer, sliding boards should
be used. Place optional sliding boards under
the patient. Good positions are under the feet,
the hips and the shoulders.
2) Prepare the transfer by placing the transfer
bar on the sheet, at height with the shoulder
and hip of the patient. Then fold the remaining
sheet over the transfer bar. The transfer bar
should be placed close to the patient.
3) Place the bed parallel to and approximately
0.5 meters from the other unit. Atle is placed
on the long side of the empty unit.
Attachment solution
4) Go to Atle® 180, pull out both wires
simultaneously and place them on the empty
bedside/examination table unit. Open the
loops of the attachment solution.
Attach the loops over the sheet and around
the transfer bar. Tighten the loop around
the transfer bar so that the loop is fastened
tightly. Do the same on the other side (See
“attachment solution”, page 9).
NOTE: Sliding boards should be used for a
more secure and comfortable transfer
Use of Atle® 180
General horizontal transfer of patient
The transfer technique described below can be used for patients lying in bed to transfer them
horizontally between beds, from stretcher to bed and bed to examination table and vice versa.
NOTE: Maximum patient weight at transfer
with Atle® 180 is 180 kg
1
2
3
Fig 3
Fig 4
Fig 5
4
Fig 6

7
During the transfer
5) Ensure that the bed and examination table
are placed at the same height right next to
each other. Lock the wheels on the bed. The
sliding boards must protrude between the bed
and examination table.
Atle® 180 is positioned completely close to
the long side of the empty unit. Adjust the
height of Atle® 180 so that the position of the
wires is 2-3 cm above the mattress; see gure
7. The wires must be positioned in a straight
line from Atle® 180´s output towards the
transfer bar. Finish positioning Atle® 180 by
locking all of the wheels on Atle® 180.
6) Initiate motorized transfer via hand control.
Press the pull button so that the wires are
stretched, then release the button for a
second, contact the patient and inform that
the transfer starts. Press the pull button again
and perform the transfer.
Make sure that the patient is following the
transfer, especially the feet, shoulders and
head. This is done with the help of one hand
by adjusting the sheets associated with the
transfer.
At the end of the transfer
7) End the transfer when the patient is
positioned centered on the new bed unit.
Remove the fastener from the sheet and
transfer bar by pulling out the wires slightly
and widening the loop forward through the
wire holder. Remove the transfer bar and the
sliding boards.
Use of Atle® 180
WARNING: Both beds should be positioned
right next to each other to avoid the risk
of the patient falling between these. For
additional safety measures, sliding boards
should be used
NOTE: Position Atle® 180 completely
close to the bed edge with the position of
the wires 2-3 cm over the mattress on the
examination table, to avoid risk of tipping
WARNING: Avoid moving your arm / hand
between the transfer bar and engine part
when the transfer bar approaches engine
part at end of the transfer to avoid the
squeeze risk
WARNING: To eliminate the squeeze risk,
make sure no body parts of the patients is
between the bed and the examination table
or between bed and Atle® 180
WARNING: To eliminate the squeeze risk,
ensure that the patient does not get stuck
with body parts in relation to the transfer
bar or the wire during the transfer
6
Fig 7
Fig 8
5

8
Transfer at bowl-shaped CT-scanner
When moving from the CT table to another
bed it is recommended that two people carry
out the transfer with Atle® 180.
One person monitors and conducts the
transfer with the hand control in the usual
manner while the other person supports the
patient on the other side of the examination
table during the transfer, over the curved
edge of the CT table. The person who
supports the patient can choose to lift at the
edge of the sheet to create a smooth transfer.
Attachment solution with hooks in a
stretcher or trauma mattress
For all types of mattresses with handles along
the side, the hook is attached to the loop of
the wires and then connected directly to the
handles on the mattress.
Fig 11
Use of Atle® 180
NOTE: It is possible to move the entire
mattress to the examination table with the
help of the patient’s weight and with sliding
boards under the matress
7
Fig 9
Fig 10
Fig 12
WARNING: The pull should be straight

9
Attachment Solution
Before transfer
1) Prior to patient transfer, check that the
sheet is stretched under the patient and lay
the transfer bar on the sheet close to the
patient and with the ends of the transfer bar at
the level of the shoulders and hips. Fold the
sheet over the transfer bar.
2) Pull out the wires simultaneously and place
them on the empty unit, open the loops for
preparation of the attachment solution.
3) Attach the loop around one end of the
transfer bar. Tighten by advancing the wire
holder and then pull the wire. Do the same on
the other side.
After Transfer
4) When the patient transfer has been
completed, release the wires and disconnect
it from the transfer bar.
Fig 13
Fig 14
Fig 15
1
2
3
Use of Atle® 180
NOTE: Place the transfer bar close to the
patient to ensure the optimal position in the
middle of the bed/examination table after
the transfer Fig 16
4

10
Charging and Hand Control
Charging Batteries
The battery pack is the type of rechargeable
Lead. The batteries are charged via the main
power outlet. The Atle® 180 uses a 24 Volts
that can deliver approx. 50 pulls per charge
with a new battery. The battery life is variable
(2-3 years) and is influenced by frequency of
use and weight of patient.
The battery needs to be charged when it is
empty. Only the supplied charger (Mascot
2241P) should be used. For safety reasons,
Atle® 180 cannot be used while charging.
The charging of Atle® 180 is stopped by
unplugging the power cord from the outlet.
The hand control indicates that the battery is
being charged by traveling lights for charging
status.
The battery charges fully in approx. six hours.
Note that the battery must be charged for
at least 24 hours non-stop before rst use.
Battery life is not adversely affected by the
fact that the battery is charged frequently.
Replacement of the battery should be done
by trained staff or by the manufacturer service
staff.
Hand Control
If the light on the right turns red it indicates
that something is wrong (error). The LED is
permanently lit if there is an internal hardware
fault in the control box. On the other hand, if
an error is caused by the user, e.g. that you
try to run a function when the batteries are
charging, the alert LED only lights up while
you are trying to activate a function (i.e. when
you press a button).
Battery indicator
When the light on the left is red, Atle® 180
must be charged. When one light is green,
the battery charge level is less than 33%,
when two lights are green, the battery charge
level is 33% or better, and when all lights
are green, the battery charge level is 66% or
better.
1) On / off button
2) On / off button indicator
3) Reset button for the overcurrent protection
4) Lowering and raising buttons
5) Motorized pull button
6) Error indication
7) Battery level
8) Battery indicator
Only use the supplied hand control.
Use of Atle® 180
NOTE: The battery must be charged for at
least 24 hours non-stop before rst use
Fig 17
WARNING: Replacement of the battery by
inadequately trained personnel could result
in a hazard
Fig 18

11
Use of Atle® 180
Buttons
Pulling button
The button with double arrows is used to pull
the patient towards Atle® 180. The transfer
continues as long as the button is pressed or
until it reaches its limit.
Raising / lowering buttons
The up and down arrows are used to raise
and lower the motor housing. The motor
housing will be be raised or lowered as long
as the button is pressed or until it reaches its
limit.
Reset button
An overcurrent protection is triggered when
Atle® 180 is exposed to too high a load. To
reset the overcurrent protection, the user is
required to press the reset button. The user
can then press the pull button again.
On/off button
When the green light is on, the device is on
and ready to use.
Brakes and Emergency Stop
Brakes
Foot operated brakes are tted on all the four
wheels.
To lock the wheels, press down the back of
the pad.
To release the brakes, push the lower part of
the pad up.
Emergency stop
The user can shut off the power at any time
by pressing the red emergency button on
the control panel. Reset the emergency
stop function by turning the emergency stop
button and then pressing the reset button
next to the red emergency stop button.
While nding the battery level again the error
indication on the hand control will light red
and there might be some sounds but Atle®
180 can be used as normal.
Fig 19
Fig 20
Fig 21
Fig 23
NOTE: When transferering heavier patients,
it is important to lock the wheels, especially
the wheels in the front
WARNING: Without the wheels locked,
Atle® 180 is a risk of tipping
unlocked
locked
NOTE: Only the supplied charger and hand
control should be used
Fig 22
WARNING: Periodic checking or
replacement of the charger is neccessary

12
First-time users should practice the
emergency stop manoeuvre before operating.
After Use
When the expiration date of the device is
reached, the product must be recycled
according to set standards for each country
and hospital.
Use of Atle® 180
Fig 24
Emergency stop Reset button Hand Control
Output
Charger
Output

13
Cleaning of Atle® 180
Surfaces that the caregiver touches must be
cleaned manually after each transfer. Clean
the surfaces of Atle® 180 regularly with
single-use cloths with hot water or pH neutral
detergent. For disinfection use the regular
surface disinfectant.
Pay attention to parts that are most likely to
be touched:
• The hand control
• The control panel
• The handle
• Attachment solution
• The wires
Replacement of the wire
The wire and the bobbin should be replaced
with a new one in case of suspicion or risk of
contamination, visible damage to the surface
layer on the wire or discoloured wire.
The bobbin is replaced as follows.
1) Open the cover to the bobbin by inserting
a nger into the opening on the side of the
cover
2) Take out the wire from the wire guide by
pulling the wire to the side and through the
opening.
3) Remove the bobbin by unscrewing the
screw and pull out the bobbin.
Fig 25
Fig 27
1a
1b
Cleaning and service
WARNING: if the bobbin and wire are
not changed when there is a suspicion of
contamination there is a risk of spreading
an infection
3a
Fig 26
Fig 28
2

14
4) Take a new bobbin and attach it by pushing
it onto the pin. Finish by tightening the screw.
5) Put the wire back into the wire guide.
6) Close the cover
Cleaning and service
5
Fig 31
4
6a
Fig 30
Fig 32
Fig 33
NOTE: Incorrectly placed bobbins may fall
off the pin and the patient transfer cannot
be performed
Fig 29
3b
6b
WARNING: To avoid squeeze risk, do not
use Atle®180 when changing bobbins

15
Maintenance
Replacement of components must be done by medically trained staff or by the manufacturer
service staff.
More severe contamination may require the external parts to be unscrewed, and the internal
components to be cleaned by medical personnel.
Preventive and check-up service is recommended once per year. The hospital’s technician
or the manufacturer’s service technician performs this procedure. Key areas are engine
maintenance, pull test and condition of the wire. After the service has been performed, a
functional test is done without a patient.
Atle® 180 is subject to wear and tear, and the following actions must be performed when
specied to ensure that the product remains within its orginal manufacturing specication.
Cleaning and service
3) Press the emergency stop button to make sure that all electrical power is cut off. No
action should occur when activating the buttons on the hand control
4) Check all the functions on the hand control. Ensure that the hand control cover is
intact
9) Check all the functions on the control panel
5) Check that Atle® 180 can be switched on and off with the power on / off button on
the hand control
7) Check that Atle® 180 can be raised and lowered by pressing the hand control up
and down button
6) Check that Atle® 180 pulls when you press the hand control pull button and that it
stops when you release the button. The wires should moves towards Atle® 180 when it
pulls
8) Check the reset button on the control panel by pressing the appropriate button; you
can transfer with Atle® 180 by pressing the pull button
Points to be inspected by user/service technician
Frequency
Before every use
1) Ensure that the battery charge indicator is within the normal range
Annually
Initially
X
XX
X X
X X
X
XX
X
2) Check that Atle® 180 is charging when a battery charger is connected to Atle® 180;
the charging status LEDs on the hand control will wander. If the battery is not taking
charge or the battery has a short operation time, replacement of the battery is
needed, see points for service technician
X
XX
Note: The list continues on the other side
NOTE: Do not use Atle® 180 while serviced or maintained

16
10) Press the reset button for the emergency stop button to see that all electrical
power is on again
11) Verify the proper functioning of the four wheels, check the brakes and that they
spin well
1) Check all bolts, nuts and locknuts to ensure they are tight
2) If the product does not work as intended, immediately contact Njord Medtech for
support
13) Check that the wires can be pulled out of Atle® 180 by hand without signicant
resistance
15) Check that the wires are not damaged and open the covers over the bobbins and
check that the bobbins are not damaged, if damage see “Replacement of the wire”
16) Check the home position sensors, by pulling the transfer bar all the way into Atle®
180 and Atle® 180 stops pulling automatically (the sensors do not work if engine noise
is heard when the transfer bar is in the home position)
4) Check that Atle® 180 is charging when a battery charger is connected to Atle® 180;
the charging status LEDs on the hand control will wander. If the battery is not taking
charge or the battery has a short operating time, replace the battery*
3) Connect a dynamometer on the opposite side of the bed to Atle® 180, connect the
transfer bar / hooks to the dynamometer and press pull on the hand control. Read that
Atle® 180 gives 1150 - 1300N before the overcurrent protection is triggered
14) Check that the wires are completely loose and can be pulled out
X X
X X
X X
XX
X
X
X
X
X X
12) Check front and rear wheels regularly for hair and debris; clean when necessary X
Points to be inspected by service technician
Frequency
Before every use
Annually
Initially
XX
Cleaning and service
*The battery (power pack RPP10) can be ordered from Njord Medtech or designated distributor
Points to be inspected by user/service technician
Frequency
Before every use
Annually
Initially

17
Troubleshooting
Cleaning and service
Type of error Possible cause Action
The wires cannot be pulled out - The wire is stuck or tangled on
the bobbin
- Open the cover to the bobbin
and release the tangled wire
either by removing/untangling
the wire from the bobbin one
turn at a time or by taking out
the bobbins and pulling the wire
out completely
The transfer bar cannot be
attached
- The attachment solution has
been put on incorrectly
- The sheet is two fold or too
thick
- The attachment solution is
worn out
- Adjust according to
instructions under “Sheet
attachment solution”
- The battery is discharged
- The main power switch is
rejected
- The emergency stop has been
activated
- Overload: the patient is too
heavy
- The patient weighs more than
180 kg
- The mattress texture gives a
very high friction
- The mattress structure and
material allows the patient to
create too much friction
The torque/engine stops
The torque/engine does not
start via the hand control
- Charge Atle® 180
- Turn on the main power switch
- Deactivate the emergency
stop by press the “reset” button
on the control panel
- Atle® 180 is intended for
patients that weigh a maximum
of 180 kg
- Perform manual transfer
- The stop measure stops the
machine over 1200N
Hand control does not respond - The emergency stop button is
activated
- The plug to the hand control is
unplugged
- Battery needs to be changed
- Deactivate the emergency stop
by clicking on the “reset” button
on the control panel
- Summon technical trained
personnel or Service technician
Every time the wires are pulled
in or out, they are at different
lengths
- The wires are wound
differently on the bobbins
- Open the cover to the bobbin
and release the wire one turn
on the bobbin on the side where
the wire pulls short
The error indication on the hand
control glows red and Atle® 180
makes a sound
- The emergency stop button
has been released after
activation
- After the emergency stop has
been restored, Atle® 180 needs
a few minutes to nd the battery
status and while doing this, the
error indication will light up and
Atle® 180 will make light noises,
but it is possible to use Atle as
usual
Other manuals for Atle 180
1
Table of contents
Other NJORD Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Dräger
Dräger Isolette 8000 plus Instructions for use

Weinmann
Weinmann OXYWAY Click Device Description and Instructions for Use

OxySmart
OxySmart PC-60F user manual

Valleylab
Valleylab Force 2 user guide

Thuasne
Thuasne TOWNSEND SpryStep Max AFO Instruction guide

Handicare
Handicare LinidoSolutions LI2618.0006-12 manual