12
Using the Avant 9700
LCD Display Views
The Avant 9700 features three unique LCD display views: Plethysmographic
Waveform view, Trend Waveform view, and Trend Histogram view.
Plethysmographic Waveform View
Plethysmographic Waveform view is the Avant 9700’s default. Approximately 3.4
seconds of data are displayed on the device’s LCD screen. Upon reaching the right
edge of the LCD screen, new data overwrites old data starting at the left edge of the
screen. Alarm limits and vertical scaling are shown at the bottom of the LCD display.
(Vertical scaling is a value—shown with x1, x2, x4, x8, or x16—identifying the
vertical scaling of the waveform. The 9700 may scale the vertical amplitude of the
signal in order to allow small signals to be displayed with good resolution and to
prevent large signals from being clipped. For example, when x16 is displayed, the
amplitude of the displayed signal is 16 times the actual size.)
NOTE: The color of the plethysmographic waveform is the same as the
color of the Pulse Strength Bargraph.
Trend Waveform View
Trend Waveform view allows viewing of up to the most recent 24 hours of patient
SpO2and pulse rate data stored in nonvolatile patient memory. The SpO2and pulse
rate waveforms appear on the same LCD screen, with SpO2data shown in green
and pulse rate data shown in red. The Left and Right Control buttons can be used
to move the cursor to any data point on the trend waveform display, and the Plus
and Minus buttons can be used to zoom in and out on displayed data. The zoom
range is 15 minutes (zoomed in completely) to 24 hours (zoomed out completely).
The length of time displayed is shown on the top line of the LCD. Time, date, SpO2,
and pulse rate data for the current cursor position appear at the top of the LCD
screen. If multiple data are displayed at a single point when the LCD is zoomed out, only the
lowest SpO2and pulse rate values are displayed in text at the top of the LCD.
NOTE: A single point on the horizontal time axis can represent a range of
data stored in memory, depending on the current level of zooming.
To ensure that all data is viewable, use the Plus and Minus buttons
to zoom in and out.
When the oximeter does not have data for any given period, the values are shown as
SpO2 = --- or HR = ---. When this occurs, the waveform for that period is drawn
off scale at the top of the LCD display screen (i.e., a line is shown at the top of the
screen to represent all missing data).
NOTE: Record gaps are identified with a white bar at the top of the LCD
display screen; individual records can be differentiated by
observing the blank space between waveforms.