NOVACARE LAS 500 User manual

Operating Instructions
LAS 500
Low air Loss System

LAS_500_eng page 2
Thank you for purchasing the novacare
LAS 500 Low Air
Loss Therapy System, a system for the treatment and
prevention of pressure ulcers.
A good decision.
Your novacare® – Team
Please read the follow instructions carefully and observe
the warning notice before commissioning the system
Table of contents :
1. The system / Mode of action / areas of application page 3
2. Severity of pressure ulcers Page 3
3. Maximum recommended patient weight Page 3
4. Indications / Contraindications Page 3
5. Controls / functions Page 4
6. Setting chart Page 5
7. Pre-application system preparation Page 5 / 6
8. Lock – Function Page 6
9. Backup – Function Page 7
10. Static mode Page 7
11. Bariatric- / Heavy weight function Page 7
12. Care – Function Page 7
13. Alarm function Page 7
14. Practical Test Page 8
15. System operation with one cell extracted Page 8
16. CPR – Valve Page 8
17. Caution note Page 8
18. Safety Page 9
19. Testing for toxicological harmlessness and Bio-compatibility Page 9
20. EMC Page 9
21. Maintenance / Inspection Page 9 / 10
22. Hygiene Page 10
23. Care, cleaning, disinfection Page 10
24. Technical information Page 11
25. Waste Disposal Page 11
26. Warning notice for private, non-professional operators Page 12
27. Information for installation and instruction Page 12
28. Warranty Page 13
29. Scope of Supply Page 13
30. Product Identification / Item no. Page 13
31.
Notes Page 14

LAS_500_eng page 3
1. The System / Mode of action / area of applications
The novacare
LAS 500 Low Air Loss Therapy System is a true Low-air-Loss-
System“. It can be operated in static as well as in alternating mode.
Its technology is based on a high air stream, in connection with large air openings
of the cells. The result is an air cushion effect. It enables the patient to sink in
deeply, resulting in a reduced pressure Therefore the system is especially suited
for the application on pain-sensitive patients.
Contrary to ordinary alternating pressure system the alternating mode operates in
a 3 to 1 principle. The system alternately ventilates two cells, while it removes air
from the third one. This results in a very low partial bearing pressure in the
alternate mode as well.
The novacare
LAS 500 Low Air Loss Therapy System is a mattress
replacement system for clinical, stationary and ambulatory care-taking.
The system consists of a control unit with digital operating unit and a mattress
system with 18 arranged in a transverse manner.
2.
Severity of pressure ulcers
The novacare
LAS 500 Low Air Loss Therapy System is recommended for
pressure ulcer treatment up to decubitus degree 4 (EPUAP) / high risk to very high
risk.
3. Maximum recommended patient weight
In standard operation the novacare
LAS 500 Low Air Loss Therapy
System is recommended for a maximum patient weight of 180 kg.
An additional bariatric- / increased function permits applications for a patient
weight up to 320 kg. For extra comfort we recommend the use of an optional
120cm wide bariatric mattress.
4. Indications / Contraindications :
Indications Treatment and prevention of pressure ulcers and decubitus
especially for patients at high risk, as well as for heavy weight
patients.
Contraindications: Alternating pressure should be applied to pain or pain-
sensitive patients or in case of instable spinal cord injuries!

LAS_500_eng page 4
5. Controls / Functions:
Legend:
1. Setting of patient weight (observe chart)
2. Setting of alternating pressure interval / Cycle
3. Alternating mode
4. Static mode
5. Care – Mode
6. Start button (Operate) or Standby
7. display alternating pressure interval / Cycle (5 – 95)
in increments of 5
8. Bariatric mode
9. Lock function
1
2
3
4
5
6
8
9
7

LAS_500_eng page 5
6. Setting chart:
Please consider that these pre-settings are recommendations only. Please conduct the
„Practical Test“ in any case and make individual adjustments until you determine the
correct setting for the patient in question.
7. Pre-application system preparation :
1. First, remove the existing mattress from the patient bed. Now, take the
complete mattress system from the carton and place it onto the patient bed
frame.
2. Remove the control unit from the box and attach it at the foot end of the bed,
using the fixation hooks.
! The automatically retractable hooks adjust to the foot end of the bed and
provide a secure hold. Upon removal of the control unit the hooks
automatically retract to the original position..
3. Make sure that the connection tube is located at the foot end of the bed.
4. Using the belt, affix the complete mattress to the bed frame.
5. Now connect the trio-plug as well as the CPC ventilation plug of the mattress
connection tubes to the connection outlet located on the side of the control
unit. Make sure that all plugs connect properly.
6. Insert the power plug into a 220 V power outlet.
7. Now activate the main power switch (ON / OFF) located on the
right side of the control unit.
ON
OFF

LAS_500_eng page 6
8. „STANDBY“ signals operational readiness. Activate the control
unit by pushing „I“. The LED-display „OPERATE“ comes on.
Return to „STANDBY“ by re-pushing of „I“.
9. Now push the mode button to select alternating (LED
– display „ALTER“ lights up) or static mode (LED –
display „STATIC“ lights up) depending on the desired
operating mode,
10. Duration of inflation is 2-5 min.
11. The „Cycle“- button sets the alternating intervals. This
function if only active in the “Alter” mode. Setting is achieved
in 5-minute increments (0-95 Min) by pushing the button.
Push the “Alter” Button to return to „5“.
12. Set patient weight by pushing the -/+ key.
Use setting chart for support.
13. The patient can now be placed onto the
mattress.
14. Now verify or correct the settings by conducting the manual check.
15. Please note that the system pressure might require adjustment when raising
the head rest section.
8. Lock – Function :
In order to prevent unauthorized changes in the default setting, the LAS 500
control unit is equipped with an automatic key lock function. It is automatically
activated after 3 minutes.
Release the key lock function by pushing the button with the key symbol. Keep it
pushed for 5 seconds, until the „Lock“ LED Display goes out.

LAS_500_eng page 7
9. Backup – Function :
In case of power loss, the settings used last are saved. New adjustments after
power loss are not necessary.
10. Static mode:
Set the mode button to „Static” (LED – display „STATIC“ lights up).
The unit can be operated in static mode for special applications (i.e. pain
patients)
11. Bariatric- / Heavy weight function :
The novacare
LAS 500 Heavy weight therapy systems features a „Bariatric-
Function accommodating especially large patients. We recommend the use of
the bariatric mattress (optional) in combination with this function.
The mode is activated by pushing the BARIATRIC – Button. The “Bariatric” LED
lights up.
12. Use of CARE Function:
The LAS 500 control unit features a CARE Function. By
activating this function all chambers of the mattress system are
inflated with maximum system pressure for 30 min. After 30
min., for the safety of the patient, the system automatically
switches to the last selected mode.
This special function facilitates re-positioning of the patient during
care-taking.
13. Alarm function :
The unit features an acoustic power loss alarm. Power loss is signaled by intermittent
beeping.
This can be eliminated by turning the unit off (ON/OFF – Switch).

LAS_500_eng page 8
14. Practical Test
An optimal correct pressure setting permits the patient to sink into the mattress by
about 20 to 30% of the original cell height. In alternating mode a hand or lower arm
should be inserted effortlessly between patient and basic mattress into the respective
ventilated cell region. Under no circumstances may the patient hit the mattress
bottom.
15. System operation with one cell extracted:
The LAS 500 mattress system permits continued operation with one cell extracted.
Release the CPC coupling of the affected cell, remove the cell und lock the coupling
with a CPC sealing plug (optional, item no.: 991016). Ensure that it locks in place
properly.
Now the system can be used once again without difficulty.
The extracted cell can in re-inserted in the opposite sequence.
16. CPR - Valve :
The novacare
LAS 500 Low Air Loss Therapy System is equipped with a CPR
emergency valve. It permits a spontaneous deflation for example to conduct
reanimation. In this case set the valve to switch position „open“.
17. Caution note :
• Protect control unit from humidity and direct wetness
• Keep tubes free of bends
• Do not use system in the immediate proximity of a source of heat.
• When moving the bed pull out the power plug.
• Never pull the cord.
• Sharp objects must be kept away. Do not secure overlay with inappropriate
means.
• Do not operate system in the immediate proximity of flammable gases (risk of
explosion)
• Use only original spare parts and consumables.
• If side rails are used, ensure the adherence to the mandatory minimum
distance. Side rails need to be boosted if the minimum distance cannot be met.

LAS_500_eng page 9
18. Safety:
• DIN EN 60601-1:2001 + A1:2006
• DIN EN 60601-1-2:1990 + A1:1993 + A2:1995 + A13:1996 (EMV)
• 2006/95/EG
• California – Fire test Class 1 (for cover)
• Bend-proof supply tubes
• CPR – emergency valve
• Acoustic power failure alarm
19. Testing for toxicological harmlessness
and Bio-compatibility
• DIN EN ISO 10993-1:1998
• ASTM F 963-96a
20. EMC
Electric devices may interact due to electro-magnetic radiation. We recommend
a safety distance of at least 1 meter distance especially for sensitive equipment.
Upon request we will provide you with a table for more detailed information
21. Maintenance / Inspection
The novacare
LAS 500 low air loss therapy system is a Medical product in
accordance with RL 93/42/EWG. Use and applications are subject to these legal
foundations.
Systems are suitable for reuse. However, hygiene directives must be observed and
flawless operation must be ensured.
Maintenance at regular intervals (mandatory every 2 years) is necessary to
preserve the function of the novacare
LAS 500 low air loss therapy system.
Used filters must be exchanged within this period.
Maintenance / inspection is conducted by the producer or a firm authorized by the
producer at the client’s expense.
If maintenance / inspection is not conducted at all or in due time or not by an
authorized firm, then the guarantee or warranty claims are not longer valid.
Damages or any other loss of function arising from lack of maintenance or untimely
maintenance / inspection or interferences by an unauthorized party, likewise lead to
loss of guarantee and warranty claims (see warranty, paragraph 28).

LAS_500_eng page 10
Maintenance / inspection includes the following actions:
• Replace membrane unit
• Replace air filter
• Replace synchronous motor
• Conduct function test / function check
• Sealing of case and attaching an inspection sticker
• Protective conductor check in accordance with DIN VDE 0751
Maintenance / inspection carried out by the novacare service department is recorded
and documented according to ISO 13485.
22. Hygiene
Systems designed for re-use must be prepared on the basis of locally valid
directive, laws or regulations “for Hospital hygiene and Infection prevention“.
Already validated procedures and hygiene schedules, if any, must be applied for
this preparation.
Please act according to your locally valid hygiene directives and the respective list
of disinfectants for disinfecting the products and use only disinfectants listed in
them. Follow the instructions for use prescribed by the respective manufacturer.
Guarantee claims become null and void if inappropriate and unlisted preparations
are used or are used inappropriately.
We recommend our cleaning service. Please refer to the current service price list
for pricing.
23. Care, cleaning, disinfection
The novacare
LAS 500 Low Air Loss Therapy System all around protective
cover can be cleaned chemical-thermally by machine up to 60°. For disinfection
we recommend the use of an approved disinfection cleaning additive. To increase
the life span clean cells by spray/wipe disinfection. In special cases chemical-
thermal cleaning is possible as well.
Please ensure that the cell openings are shut when cleaned by machine.
For use, follow the recommendations of the respective manufacturer.
A spray-wipe desinfection is possible in special or urgent case, such as interim
cleaning of acute soiling during an application.
This measure might not replace hygiene regulations currently valid in your country
and should be applied only in special cases (i.e. interim cleaning of acute
contamination during applications).

LAS_500_eng page 11
24. Technical information:
• Control unit : Weight 6.1 kg
dimensions : 42 x 27 x 18.5 cm
Length Power cable 3 m
Spannungswerte : AC 220-240 V / 50 Hz
At maximum up to 2.5 VA / 575 W
Fuses : TIAH 250 VAC 1A
Protection class : Type BF / class 1
• Mattress - Standard : dimensions: 200 x 90 x 24 cm
Weight : ca. 13 kg
• Mattress - Bariatric : Dimensions: 200 x 120 x 24 cm
(Optional) Weight : ca. 15 kg
Material PU –coated
Polyester texture
• Cells : Dimensions : 90 x 26 x 22 cm
Material PU
• Max. Flow < 1200 ltr. / min.
• Cycle Adjustable 5 - 95 Minutes
(in 5-Minute- intervals)
• Surrounding conditions:
Temperature Application/operation 5°~ 50°
Transport / storage -10°~ 60°
Rel. humidity Application/operation 20 % – 90 %
Transport / storage 20 % – 90 %
25. Waste Disposal
Mattress system
The materials used for the mattress does not pose any danger for the environment.
They may be disposed off with the household waste
Disposal via official waste incineration is preferred.
Control units
Control units are considered electric equipment waste. Please contact the proper
waste disposal authorities.
Novacare gmbh electric equipment is registered (see symbol)
WEEE – Reg.- No. DE – 89 403 200

LAS_500_eng page 12
26. Warning notice for private, non-professional operators:
In case of system failure check:
Is the unit turned on
Is the power cord connected properly and is the power plug inserted correctly?
Is the mattress connection tube connected properly?
If this measure fails to be successful contact the responsible supplier or user
(authority responsible for the use and maintenance of the equipment)
Never attempt to repair the system yourself. Fuses may only be exchanged by
qualified and authorized personnel.
In case of damage pull the power plug.
Cleaning
Remove dirt or stains with lukewarm water and a cloth. Water may be treated with mild
cleaning agents. Do not use aggressive cleaning agents or solvents.
Wipe disinfection
Disinfect surface regularly after every cleaning. Wipe disinfection is preferred.
Comply with information provided by cleaning agent’s manufacturer or request instructions
from the responsible care-taking personnel.
27. Information for installation and instruction :
Novacare LAS 500 low air loss Therapy Systems are medical care products. They
are sold by novacare gmbh primarily to medical experts (care-taking personnel,
specialized medical dealers, users, medical facilities, etc.) and put at the disposal of
professional users
Experts and professional users are defined as persons who, based on their medical
or comparable training, have sufficient knowledge about the condition intended to be
treated with the system or its prevention, in order to enable the patient and/or the
non-professional user and medical layman to sufficiently operate the alternating
pressure (pressure care) system.
Installation and system set up at the patient’s as well as putting into service and
instruction of the non-professional user or the medical layman respectively are
carried out by medical experts.
These are to be carried out within the scope of the usual instruction obligations, for
example the description of functions, explanation of operating features, as well as
cleaning and system effectiveness, and the indication of potential risks.
Operation by non-professional users or medical laymen is possible but should not be
implemented without sufficient instructions.
Special training concerning function and operation of the system are not necessary
but can be offered upon request.

LAS_500_eng page 13
28. Warranty :
• The warranty is based on statutory regulations.
• Furthermore a warranty is ruled out, if the goods supplied by us are processed,
handled or modified by other parties without our consent or if our operating
instructions are not followed. If the seals are damaged, it will be assumed that
the aforesaid activities have been carried out.
• For medical devices as understood by the EU-Directive 93/42/EEC, whose
usage (also operations) is subject to regular service and maintenance, a
warranty can be offered only if the service intervals prescribed by the
manufacturer are adhered to.
• As far as warranty claims are asserted and if during the subsequent inspection
it is found that the damages are of wear and tear or are not subject to
warranty, then we are empowered to claim the expenses (inspection, transport
costs etc.) from the client.
• The use of parts or individual components of other systems or products or
combinations thereof are not permissible. Warranty claims cease to exist for
any damage resulting from this, which must be borne by the user.
29. Scope of Supply
• Control unit LAS 500
• Mattress system (Standard)
Optional:
Mattress system (Bariatric)
• Carry bag
• Operating instructions
30. Product Identification / Item no :
• Name : novacare
®
LAS 500 low air loss system
• Item no.: 99 9001

LAS_500_eng page 14
31. Notes :

LAS_500_eng page 15

LAS_500_eng page 16
D-67098 Bad Dürkheim, Bruchstrasse 48 Tel. 06322 / 9565 - 0, Fax 06322 - 9565 - 65,
www.novacare.de