Nucletron Simulix-Evolution Imaging User manual

Simulix-Evolution Imaging
Service Manual
REF 112.912ENG-04
21f375 rev 02


How to contact Nucletron Offices
NUC-T380-13
HOW TO CONTACT NUCLETRON OFFICES
Unless otherwise stated the service dispatch call number is the main telephone number.
Head Office and Manufacturer:
Nucletron B.V.
PO Box 930
3900 AX VEENENDAAL
Waardgelder 1
3905 TH VEENENDAAL
The Netherlands
Tel.: +31 318 557 133
Fax: +31 318 550 485
Service: +31 318 557 111
Web site: www.nucletron.com
BeNeLux
Nucletron B.V.
PO Box 930
3900 AX VEENENDAAL
Waardgelder 1
3905 TH VEENENDAAL
The Netherlands
Tel.: +31 318 557 236
Fax: +31 318 557 270
Service: +31 318 557 111
Web site: www.nucletron.com
Germany
Theranostic Medizintechnik GmbH
Obere Dammstrasse 8-10
D-42653 Solingen
Tel.: +49 212 5875 153
Fax: +49 212 5875 269
Service: +49 212 5875 333
Web site: www.theranostic.de
France
Nucletron S.A.R.L.
Parc d’Activités l. Bernard Vergnaud
15 rue Paul Langevin
93274 SEVRAN Cedex
Tel.: +33 1 4936 2060
Fax: +33 1 4383 3638
Italy
Nuclital srl
Viale Elvezia no. 2
20052 MONZA (Milan)
Tel.: +39 039 322848 / 323148
Fax: +39 039 323913
United Kingdom
Nucletron UK Ltd.
Nucletron House
Chowley Oak
Tattenhall
CHESTER CH3 9EX
Tel.: +44 1 829 771111
Fax: +44 1 829 770979
Sweden
Nucletron Scandinavia AB
Box 1704
Klostergatan 12
SE–751 47 Uppsala
Sweden
Tel: +46 18 565 000
Fax: +46 18 565 011
Spain
Nucletron S.A.
Avda. de Castilla, 2
Parque Empresarial San Fernando
Edificio Francia, Planta 2e, Esc. A
28830 San Fernando de Henares (Madrid)
Tel.: +34 918 250 067
Fax: +34 918 250 069
Service: +34 918 250 068
Europe / Middle East and Africa
Poland
Nucletron Poland Sp. z o. o.
Oddzial w Warszawie
Al. Krakowska 285
02-133 Warszawa
Poland
Tel.: +48 22 868 1782
Fax: +48 22 868 2216

How to contact Nucletron Offices
NUC-T380-13
HOW TO CONTACT NUCLETRON OFFICES
Unless otherwise stated the service dispatch call number is the main telephone number.
USA
Nucletron Corporation
8671 Robert Fulton Drive
COLUMBIA, MD 21046
Tel.: +1 410 312 4100
Fax: +1 410 312 4199
North and South
America
Canada
Nucletron Canada Inc.
411 Legget Drive, Suite 502
Kanata, Ontario
K2K 3C9
Canada
Tel: +1 800 826 2258 / 613 886 1100
Fax: +1 613 592 6559
Hong Kong
Nucletron Asia Pacific Ltd.
Room 1005, Tower 2, Silvercord,
30 Canton Road, Tsim Sha Tsui,
Kowloon, HONG KONG
Tel.: +852 2311 2683 / 2684 / 3025
Fax: +852 2311 3672
Australia
Nucletron Pty. Limited
Suite 4, 26 Sparkes Street
Camperdown. NSW. 2050
Sydney
Tel.: +61 2 9517 1300
Fax: +61 2 9517 1311
China
Nucletron Asia Pacific Ltd.
Beijing Representative Office
Room 21D CITIC Building,
19 Jian Guo Men Wai Da Jie
Beijing, China
Post Code: 100004
Tel.: +86 10 8526 2340 / 41
Fax: +86 10 8526 2339
Web site: www.nucletron-china.com
India
Nucletron India Pvt Ltd.
No. 3, D’Silva Road
Mylapore
Chennai 600 004
Tel.: +91 44 24992877 / 24991871
Fax: +91 44 24991967
Web site: www.nucletron-india.com
Nucletron Asia Pacific Ltd.
Shanghai Representative Office
Shanghai, China
Tel.: +86 21 6207 6747 / 48
Fax: +86 21 6207 6746
Web site: www.nucletron-china.com
Asia Pacific
Singapore
Nucletron Asia Pacific Ltd.
Singapore Representative Office
89 Science Park Drive
Rutherford # 03 - 06
Singapore 118261
Tel.: +65 6779 1248
Fax: +65 6836 7496
Your Distributor / Agent

Preface
NUC-T374-02 v
Preface
Identification
Manuals are identified by a part number. The part number consists of three
parts:
Example:
The part number is printed on the right bottom side of the front page.
Numbers printed on other pages are for internal revision control and may
differ.
CE Classification
Manuals made for products that have been CE (Conformité Européenne)
marked have the CE mark on the front page. The number behind the CE
Mark defines the responsible notified body.
Nucletron systems are classified according to IEC 601-1 as Class I, type B.
Nucletron systems meet the relevant provisions of the European Medical
Device Directive 93/42/EEC or the European Active Implantable Medical
Device Directive 90/385/EEC; this is based on conformity of the products
and the quality system according to ISO 9001:2000 and ISO 13485:2003.
For this reason mark is used.

Preface
vi NUC-T374-02
Directive WEEE
Symbol Meaning Remarks
Location: Near or on
the type identification
plate.
Symbol for separated collection of
electrical and electronic equipment per
Directive 2002/96 of the European
Parliament and the Council of the
European Union (Directive on Waste of
Electrical and Electronic Equipment). If
applicable to the type of device, it
indicates legally imposed obligations
within EU member states, Iceland,
Norway and Switzerland when the
equipment is disposed of, at the end of
its life-time.
Owners of marked equipment should
contact the organization that imported
the equipment into their country, when
they want to dispose of the equipment,
at the end of its lifetime. The Directive
prioritizes re-use of equipment over re-
use of components over re-use of
materials over disposal as waste. Article
5 part 2d allows producers to decline the
return of any used equipment that is or
may be biologically or radiologically
contaminated.
Conventions
Throughout this manual certain conventions are used. These are Warnings,
Cautions and Notes. They provide a means of prioritizing information to be
brought to the attention of the user. They are given as follows:
Warning: Designates a possibly dangerous situation. Non-observance may lead
to death or the most severe injuries
Caution: Designates a possibly harmful situation. Non-observance may lead to
minor injuries or damage to the product.
Note: Designates a possibly harmful situation. Non-observance may lead to
damage to the product or the environment.
Precautions
The user is expected to use the product (soft and hardware) in accordance
with the instructions given in this manual, which must be read before the
system is used. Any unauthorized deviation from the procedures laid down
in this manual can affect the contractual obligations between purchaser and
vendor.
Warning: Do not modify any part of Nucletron supplied computer's hardware,
software or software configuration. Installing or running an appli-
cation that has not explicitly been approved by Nucletron may
compromise patient safety and will result in termination of all
warranty and service contracts.

Preface
NUC-T374-02 vii
Warning: The Nucletron products are designed for use and maintenance using
only parts available from Nucletron. No liability for failure can be
accepted by Nucletron through the use of parts obtained from other
suppliers unless written permission is obtained from Nucletron.
Copyright
All rights reserved. No part of this manual may be reproduced or transmitted
in any form whatsoever, including photocopying and recording without
written permission of the copyright holder: Nucletron B.V., PO Box 930,
3900 AX VEENENDAAL, The Netherlands.
Concerning this Publication
Information contained in this publication is subject to change without
notice. Whenever the equipment is delivered with a documentation medium
such as a CD or DVD, the user shall always check the medium for latest
information. This information is part of Release Notes and Document
Upgrade files.
Neither Nucletron B.V. and its subsidiaries nor anyone else who has been
involved with the development, production or delivery of the documentation
be liable for any special, incidental or consequential damages, whether
based upon breach of warranty, breach of contract, negligence, strict
liability in tort or any other legal theory.

Preface
viii NUC-T374-02

Table of Contents
112.912ENG-04 Simulix-Evolution Imaging –Service Manual ix
Table of Contents
1. Introduction 1-1
1.1 Intended Audience ............................................................................................................. 1-1
1.2 Related Documents............................................................................................................ 1-1
2. Safety Instructions 2-1
2.1 General Information............................................................................................................ 2-1
2.2 Safety Features .................................................................................................................. 2-1
2.2.1 Basic Safety Features for All Simulix Systems ................................................... 2-1
2.2.2 Oncentra Simulation ........................................................................................... 2-1
2.2.3 Anti-virus Software.............................................................................................. 2-2
2.2.4 Simulix Error Messages ...................................................................................... 2-3
2.2.5 Door Interlock Switch (X-ray Safety)................................................................... 2-3
2.2.6 Data Integrity and Security.................................................................................. 2-3
2.3 Electrical Safety.................................................................................................................. 2-3
2.3.1 Classification ....................................................................................................... 2-3
2.3.2 Power Supply ...................................................................................................... 2-4
2.4 Fire Safety .......................................................................................................................... 2-4
2.5 Fluid Safety......................................................................................................................... 2-4
2.6 Radiation Safety ................................................................................................................. 2-4
2.7 X-Ray Exposure Interlocks................................................................................................. 2-5
2.8 Electromagnetic Safety ...................................................................................................... 2-5
2.8.1 Electromagnetic Emissions................................................................................. 2-6
2.8.2 Electromagnetic Immunity................................................................................... 2-7
2.8.3 Separation Distance............................................................................................ 2-9
2.9 Emergency .......................................................................................................................2-10
2.9.1 Emergency Stop Buttons .................................................................................. 2-10
2.10 Collision Safety................................................................................................................. 2-10
2.10.1 Collision Avoidance System.............................................................................. 2-10
2.10.2 Collision Touch Guards..................................................................................... 2-11
3. Flat Panel Detector and Interface Box 3-1
4. Drawings 4-1
5. Service Procedures 5-1
5.1 Cleaning ............................................................................................................................. 5-1
5.1.1 Vacuum Cleaning................................................................................................ 5-1
5.1.2 Wiping ................................................................................................................. 5-1
5.2 Touch Guard Check ........................................................................................................... 5-1
5.3 Preventive Maintenance..................................................................................................... 5-2
5.4 Corrective Maintenance ..................................................................................................... 5-2
5.5 Fuse Replacement ............................................................................................................. 5-2
5.6 Backup and Restore........................................................................................................... 5-3
5.7 Cone Beam CT................................................................................................................... 5-3
5.7.1 Cone Beam CT Air-scan Calibration................................................................... 5-3

Table of Contents
x Simulix-Evolution Imaging – Service Manual 112.912ENG-04
5.7.2 Calibrations for Cone Beam CT Isocenter Calibration........................................ 5-5
6. Spare Parts List 6-1
7. Smoothbase Service Manual 7-1
8. Additional Information 8-1

Introduction
112.912ENG-04 Simulix-Evolution Imaging –Service Manual 1-1
1. Introduction
1.1 Intended Audience
This manual is intended for hospital service technicians, Nucletron service technicians and Nucletron
installation technicians. It provides technical service information about the system.
1.2 Related Documents
Publications related to this manual are listed in the following table:
Title REF
[1] Simulix-Evolution Reference Manual 112.910
[2] Simulix-Evolution System Log Book 112.923
[3] Simulix-HQ Service Manual 112.904
[4] Simulix-HP Service Manual 112.458
[5] Smoothbase Service Manual 190.057
[6] Compact-60 User Manual 112.470
[7] Epsilon 50RF User Manual 112.934


Safety Instructions
112.912ENG-04 Simulix-Evolution Imaging –Service Manual 2-1
2. Safety Instructions
2.1 General Information
Carefully read all the instructions before use. Observe all warnings and cautions throughout these
instructions.
Note:
Only qualified staff may use the system.
WARNING:
In case of equipment malfunctioning, bring the patient into safety and get the assistance of a
qualified technician.
WARNING:
Always take radiation safety precautions.
WARNING:
If any system components or cables are damaged, contact a Nucletron qualified service technician to
check and repair the damaged item(s).
WARNING:
Do not modify the system in any way. It is not allowed to install any software application program
that is not released by Nucletron.
WARNING:
If data or program parts are manipulated (e.g. by a computer virus), the user should contact
Nucletron.
2.2 Safety Features
2.2.1 Basic Safety Features for All Simulix Systems
Safety features at start up:
• Simulix EPROM program check
• Simulix memory check. (RAM)
• Simulix watchdog circuit
• See also related documents [3] and [4] (see Section 1.2)
Safety features during normal use:
• Enable bar to enable movements in conjunction with the specific actuators.
• Collision avoidance system (software)
• Collision guards on collimator and flat panel detector to protect a patient against a collision with the
Simulix parts.
2.2.2 Oncentra Simulation
The Oncentra Simulation workstation program code has been thoroughly checked for possible viruses and
malfunctioning, and installed according to strict installation descriptions described in the installation manual.

Safety Instructions
2-2 Simulix-Evolution Imaging – Service Manual 112.912ENG-04
WARNING:
The service engineer is advised only to use operating system utilities when instructed to do so –
either by this manual or Nucletron bulletins.
Do not change the installed software, either by changing files, moving files or changing text files.
Installation and calibration of the system shall be done by qualified service engineers only.
Do not add and use other software than the software provided by Nucletron, unless explicit
permission has been granted by Nucletron.
Re-calibration is required after any change of Simulix FAD, AFD, Imager-X, Imager-Y. Any change
includes: any change of parts, any re-adjustment or re-calibration and any mechanical damage and
repair to the system.
The Oncentra Simulation workstation as installed in a network system should be protected against viruses
and malicious code changes by the hospital network administrator.
The main safety provisions featured by the Simulix-Evolution workstation are:
• An automatic return to fluoro mode when the fluoro foot-switch is activated, so that the fluoro image
always appears whenever fluoro radiation is given to the patient.
• Automatic capturing of the Simulix positions with the image as soon as the fluoro foot-switch is released.
• Calibrated distance measurements in images are always done in the isocenter.
• Hardware detection: After switching on, system checks are performed in the X-ray generator, the Simulix
system, the Oncentra Simulation system and related communications channels.
2.2.3 Anti-virus Software
If the anti-virus software is installed without enabling automatic “live” updates of virus definitions from the
Internet (for example when “live” updates are forbidden by hospital policy), then it is recommended to do a
manual update of the virus definitions every week.
The Oncentra Simulation software has been tested with Norton AntiVirus 2006 anti-virus software.
The following programs are excluded from continuously being checked by Norton AntiVirus (for performance
reasons):
• C:\Program Files\Microsoft SQL Server\MSSQL$SMOBA\binn\sqlservr.exe
• C:\Program Files\Nucletron\Smoothbase\SmoothbaseStoreScp.exe
• C:\Program Files\Nucletron\Smoothbase\Smoothbase.exe
• C:\Program Files\Nucletron\OncentraSimulation\AcquisitionServer.exe
• C:\Program Files\Nucletron\OncentraSimulation\PositionServer.exe
• C:\Program Files\Nucletron\OncentraSimulation\OncentraSimulation.exe
• C:\Program Files\Danware Data\NetOp RemoteControl\Host\NHSTW32.exe
If the software is installed on a Cone Beam CT setup then the following programs are also excluded from
continuously being checked by Norton AntiVirus:
• C:\Program Files\COBRA_Exxim\BackPrj_COM_Server.exe
• C:\Program Files\Nucletron\OncentraSimulation\ReconstructionServer.exe

Safety Instructions
112.912ENG-04 Simulix-Evolution Imaging –Service Manual 2-3
The following specific network ports are permitted to be checked by Norton AntiVirus:
Port number Traffic type Description
104 TCP Smoothbase DICOM import/export
6502 TCP NetOp
8080 UDP NetOp
53311 TCP Acquisition Server
53312 TCP Position Server
53313 TCP Reconstruction Server
The following specific folders are not checked by Norton AntiVirus:
• F:\Database (including subfolders) (only applicable for Oncentra ConeBeam)
• H:\CT (including subfolders) (only applicable for Oncentra ConeBeam)
• I:\CT (including subfolders)
This means that the database and temporary CT data are excluded from virus checking.
2.2.4 Simulix Error Messages
When unexpected movements are detected, the power of these movements is switched off and an
appropriate error message will appear on the user screen of the Simulix.
2.2.5 Door Interlock Switch (X-ray Safety)
A door sensor ensures that the X-ray is switched off if the door is opened.
2.2.6 Data Integrity and Security
Simulix-Evolution has features for data integrity and security:
• Password protected: protection of the Simulix Oncentra Simulation software against unauthorized use.
• Levels of user authorisation.
• Backup and restore: making a regular backup of the database can prevent the loss of data. It is
recommended to run data back-ups routinely at regularly scheduled intervals to avoid data-loss problems
due to power system instability or interruption.
• An Un-interruptible Power Supply (UPS) prevents the loss of data in case of power failure.
• Logging system.
2.3 Electrical Safety
See also related documents [3] and [4] (see Section 1.2).
2.3.1 Classification
The Simulix-Evolution complies with the EN60601-1 international standard (see preface). CE marking.
Any other equipment used with this system is assumed to comply with the EN60601-1 international standard.
The system is suitable for continuous operation.

Safety Instructions
2-4 Simulix-Evolution Imaging – Service Manual 112.912ENG-04
2.3.2 Power Supply
For protection against main voltage interruptions, the Simulix viewing station contains an Un-interruptible
Power Supply (UPS) with a battery. The UPS performs a self-test automatically to check the battery. For
details, please refer to the UPS documentation.
The Flat Panel Detector and the interface box are permanently powered for stability reasons.
WARNING:
The Simulix viewing station contains an Un-interruptible Power Supply (UPS) with a battery. This
means that an electrical voltage (115VAC / 220VAC) may be present internally, even when the system
is not connected to a mains-socket.
WARNING:
The flat panel detector and the interface box are permanently powered for stability reasons. When
switching off the Simulix, the flat panel detector and the interface box are still powered.
Note:
When the flat panel detector and interface box have been switched off and the power is switched on again,
no service action is needed.
Caution:
Switch off the Mains for the system and switch off the power to the flat panel detector before
removing or replacing fuses. Accidentally contacting the electrical circuits inside the system could
result in serious injury.
Always use the correct size and type of fuse when replacing blown fuses.
2.4 Fire Safety
WARNING:
Do not use the system in oxygen-enriched atmospheres (more than 21% of oxygen) and/or in
atmospheres containing explosives or flammable gases.
2.5 Fluid Safety
If fluid is spilt over system components, then inspect and check the proper functioning of the system.
Caution:
The system is not protected against the ingress of fluids. The classification is IPX 0.
2.6 Radiation Safety
WARNING:
Always take X-ray radiation safety precautions.
See also related documents [6] and [7] (see Section 1.2).
X-ray radiation is only possible when the operator activates the foot-switch or manually activates the
exposure buttons. The software on the Oncentra Simulation station cannot switch on the X-ray radiation by
itself.
The Oncentra Simulation software can also inhibit radiation. When Oncentra Simulation detects a failure (for
example when the communication with the simulator is lost), Oncentra Simulation will inhibit radiation.

Safety Instructions
112.912ENG-04 Simulix-Evolution Imaging –Service Manual 2-5
The foot-switch is used to make fluoroscopy images. The images are visible on the Oncentra Simulation
station. After releasing the foot-switch the X-ray radiation is switched off. On the Oncentra Simulation station
the Last Image Hold (LIH) stays visible.
The Oncentra Simulation station shows the state of the image: Live or LIH.
In case of malfunction of the Oncentra Simulation station, it is not possible to save images on the Oncentra
Simulation station.
The operator can make an X-ray exposure of the patient on the auxiliary film cassette. If the auxiliary film
cassette is not placed in a correct manner or not present, then the X-ray radiation will not switch on when the
operator activates the exposure button. The simulator shows a warning message on the data monitor. When
desired the operator can now overrule the warning message and can make an exposure. See also related
document [1] (see Section 1.2).
2.7 X-Ray Exposure Interlocks
For the Simulix the following applies: A system of interlock switches and monitoring devices detects
abnormalities in the system (e.g. open covers, open doors, shutters too wide which would result in X-ray
outside the detector) and will prevent the operator from making an exposure until the system is brought into a
safe state.
WARNING:
Never bring body parts in the beam of an X-ray system.
2.8 Electromagnetic Safety
A diagram of the system interconnection with a list of cables is shown in Chapter 4, drawing 112.912-2xx.
WARNING:
The use of accessories and cables other than specified, with the exception of cables sold by
Nucletron as replacement parts for internal components, may result in increased electromagnetic
emissions or decreased electromagnetic immunity of the Simulix-Evolution system.
WARNING:
If the Simulix-Evolution system is used adjacent to other equipment, the Simulix-Evolution system
should be observed to verify normal operation in the configuration in which it will be used.

Safety Instructions
2-6 Simulix-Evolution Imaging – Service Manual 112.912ENG-04
2.8.1 Electromagnetic Emissions
Guidance and manufacturer's declaration - electromagnetic emissions
The SIMULIX-EVOLUTION system is intended for use in the electromagnetic environment specified below.
The customer or the user of the SIMULIX-EVOLUTION system should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11 Group 1
The SIMULIX-EVOLUTION system uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11 Class A
Harmonic emissions
IEC 61000-3-2 Not applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable
The SIMULIX-EVOLUTION system is suitable for use in all
establishments other than domestic and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Table 2-1 Electromagnetic Emissions

Safety Instructions
112.912ENG-04 Simulix-Evolution Imaging –Service Manual 2-7
2.8.2 Electromagnetic Immunity
Guidance and manufacturer's declaration - electromagnetic immunity
The SIMULIX-EVOLUTION system is intended for use in the electromagnetic environment specified below.
The customer or the user of the SIMULIX-EVOLUTION system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±8 kV for input/
output lines
±2 kV for power
supply lines
±8 kV for input/
output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Not applicable Mains power quality should be that of a typical
commercial or hospital environment.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m Not applicable
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE UTis the a.c. mains voltage prior to application of the test level.
Table 2-2 Electromagnetic Immunity (1)

Safety Instructions
2-8 Simulix-Evolution Imaging – Service Manual 112.912ENG-04
Guidance and manufacturer's declaration - electromagnetic immunity
The SIMULIX-EVOLUTION system is intended for use in the electromagnetic environment specified below.
The customer or the user of the SIMULIX-EVOLUTION system should assure that it is used in such an environment.
Immunity test IEC 60601
test level Compliance level Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
10 Vrms
10 V/m
Portable and mobile RF communication
equipment should be used no closer to any part of
the SIMULIX-EVOLUTION system, including
cables, than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
Pd 35,0=
Pd 35,0=80 MHz to 800 MHz
Pd 70,0=800 MHz to 2,5 GHz
where
P
is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and dis the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the SIMULIX-EVOLUTION system is
used exceeds the applicable RF compliance level above, the SIMULIX-EVOLUTION system should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orientating or relocating the SIMULIX-EVOLUTION system.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m
Table 2-3 Electromagnetic Immunity (2)
Other manuals for Simulix-Evolution Imaging
1
Table of contents
Other Nucletron Medical Equipment manuals
Popular Medical Equipment manuals by other brands

PFM Medical
PFM Medical Rotary 3004 M operating manual

Atmos
Atmos A 161 Battery Series operating instructions

COOK Medical
COOK Medical NJFT-8 manual

Mac Medical
Mac Medical D Series Installation operation & maintenance

FIOR & GENTZ
FIOR & GENTZ NEURO HiTRONIC maintenance

Lowenstein Medical
Lowenstein Medical JOYCEeasy X Instructions for use