Standard Calls
To trigger a standard call, the user presses the
orange button on a call point or operates the
remote button of a pear push lead. This causes the
call point’s re-assurance LED to pulse slowly
Reset
To reset a call point, a member of sta presses the
green reset button. The re-assurance LED will stop
pulsing.
Assistance Calls
If assistance is required by a member of sta (but it
is not an emergency), pressing the orange button
on the call point whilst it is in Standard Call will
initiate an Assistance Call. This causes the call
points re-assurance LED to pulse more urgently
than when a Standard Call is made. If the call point
is not active and assistance is required, pressing the
orange button twice will initiate an assistance call.
Emergency Calls
In an emergency press and hold the orange button
until the call point stops beeping and the red LED
stops being lit. This makes an Emergency Call
regardless of the previous state of the call point.
This causes the call point’s re-assurance LED to
flash rapidly, more than Standard and Assistance
Calls.
Pear Push Lead Pull Out Function &
Call Level
By pulling out a pear push lead or plugging one
into the call point, a Standard Call is automatically
initiated for safety. This active call needs to be
reset. The pear push lead can be plugged into one
of two sockets on the bottom of the call point. The
Le socket generates a Standard Call and the right
socket generates an Assistance Call.
Once the call point is activated a paging message is
sent to the pagers and displays.
Sensors - example Floor Sensor Mats
The right hand socket on the call point is used to
plug in Sensors such as pressure mats. When the
pressure mat is activated the call point will generate
an Assistance Call onto the system.
The right hand socket on the call point is used to plug
in Sensors such as pressure mats. When the pressure
mat is activated the call point will generate an
Assistance Call onto the system. Any accessory can
be plugged into the aux socket and it can be either
mono or stereo 6.35mm.
Repeat Call and Escalate
The call point will resend a message every minute
until it is reset. From the first repeat onwards the
message will contain how long the call point has
been active for example:
“R:005 CALL 01Mins” or “R:011 SENSOR 03Mins”.
If a ‘standard call’ remains unanswered it will
automatically escalate to an ‘assist call’ aer six
minutes.