NuWave AirSentric WB55 User manual

WB55 Manual –revision 2.8a NuWave Sensors Product
Set-up & Operating Manual
AirSentric
Indoor Air Quality Monitor
Model: WB55

WB55 Manual –revision 2.8 1
Contents
1. Introduction ....................................................................................................................................2
1.1 Overview.................................................................................................................................2
1.2 Technical Specification............................................................................................................2
2. Hardware Configuration .................................................................................................................2
3. Hardware Installation .....................................................................................................................3
3.1 Requirements for Installing the WB55....................................................................................3
3.2 Coverage Area and Placement................................................................................................3
3.3 Mounting System....................................................................................................................5
4. Power Supply ..................................................................................................................................5
5. Internet Connection........................................................................................................................5
5.1 Wireless Ethernet Gateway Connection.................................................................................5
5.2 Network Configuration ...........................................................................................................6
6. Online Software Setup....................................................................................................................6
6.1 Online Account Setup .............................................................................................................6
6.2 Account Registration...............................................................................................................7
6.3 Adding a monitor to your account..........................................................................................7
7. Viewing and Managing Sensor Data ...............................................................................................8
7.1 Sensor Status...........................................................................................................................8
7.2 Graphing Options....................................................................................................................9
7.3 Indoor Air Quality Bands (RAG Status)..................................................................................10
7.4 Export CSV data.....................................................................................................................11
7.5 Sensor Management Settings...............................................................................................11
8. User Profile Settings......................................................................................................................13
9. API.................................................................................................................................................13
10. Gateway Network Configuration ..............................................................................................13
11. Calibration.................................................................................................................................16
12. Important Precautions..............................................................................................................17
13. Recycling and disposal ..............................................................................................................17
14. Appendix ...................................................................................................................................18
14.1 Product Warranty .................................................................................................................18
14.2 Troubleshooting....................................................................................................................19
14.3 WB55 Dimensions.................................................................................................................20
14.4 CH20 Sensor - Interfering Gases ...........................................................................................21

WB55 Manual –revision 2.8 2
1. Introduction
1.1 Overview
Before installing the AirSentric WB55 please give this manual a thorough read through. The manual
provides all of the important information to help you successfully install the monitor and set up your
online monitoring account. Please retain this manual for future reference.
1.2 Technical Specification
See airsentric.com/resources/
2. Hardware Configuration
The WB55 IAQ monitor comes equipped with a Zigbee wireless module, which enables multiple
monitors to be installed in a single environment/area. Each unit then communicates back to a single
internet enabled network point - the Data Hub - which connects to the online monitoring account.
Figure 1
→
WB55 wireless setup with Ethernet gateway unit

WB55 Manual –revision 2.8 3
3. Hardware Installation
3.1 Requirements for Installing the WB55
In order to install the system, there must be access to the following;
•Access to a power connection
•Access to a wired internet connection for the gateway
•Once power is supplied and the data gateway is connected the unit will need to be added to
your online monitoring account. Please note that you only need to add the monitor device ID
to the online software.
Important Notes:
•Do not place your WB55 monitor near any heat source such as a radiator or at a window or
direct sunlight.
•Place WB55 within range of your GY36 gateway.
•Place in an open space –avoid placing in areas where airflow may be obstructed
•Do not place where building occupants may be breathing directly at the monitor as this can
have a disproportionate effect on sensor readings.
3.2 Coverage Area and Placement
Monitor Installation Density:
Monitor installation density may vary depending on whether compliance with a particular standard
is being followed.
For example –The Well Building Standard guidelines recommend installing monitors within a
regularly occupied or common space in the building at the following installation density:
•Projects with occupiable space of < 3250 m2 : 1 monitor per 325 m2 [3,500 ft2 ] in occupiable
spaces (minimum 2)
•Projects with occupiable space of 3250-25,000 m2 : 1 monitor per 500 m2 [5,400 ft2 ] in
occupiable spaces (minimum 10)
•Projects with occupiable space of > 25,000 m2 : 1 monitor per 1000 m2 [10,800 ft2 ] in
occupiable spaces (minimum 50)
•Monitors must be distributed throughout the project and to the extent possible,
representative of all HVAC zones, faces of a building, and regularly occupied spaces (e.g., open
office areas, private offices, conference rooms and classrooms).

WB55 Manual –revision 2.8 4
•For projects with multiple floors, monitors must be distributed across different floors,
including the lowest and highest regularly occupied floor (excluding floors with only leased
space in WELL Core and Core & Shell projects).
•Multifamily Residential projects may only utilize continuous monitoring pathways in non-
dwelling unit spaces—dwelling units must meet performance requirements through methods
described in the Performance Testing Protocols for WELL section. To determine the monitor
density for all other non-dwelling unit spaces, follow monitor density rate for the total area of
the non-dwelling unit spaces only.
•For projects containing large open spaces (e.g., gyms, ball rooms, etc.), one monitor is
sufficient for an area of up to 2,500 m2 [27,000 ft2 ] if there is evidence that the air is evenly
mixed and contaminant sources are uniform (e.g., testing and balancing report indicating
ventilation rate is even throughout the space, IAQ report indicating ventilation rate and
contaminant levels are even throughout a space).
Installation Location Options:
•Monitors are installed on a wall (vertically), at a height of 1.1-1.7 m [3.6-5.6 ft] above the
finished floor.
•Monitors are installed on a ceiling (horizontally). This placement option is only applicable in
spaces, 1) with ceilings are no greater than 3.7 m [12 ft] above the finished floor, 2) that do
not utilize displacement ventilation, and 3) there is evidence that the air is evenly mixed.
Evidence for evenly mixed air within a space is submitted in the sensor data package as a
Letter of Assurance from an Engineer confirming that a space meets requirements of a
ventilation guideline listed in A03: Ventilation Design Part 1: Ensure Adequate Ventilation.
•Monitors must be at least 1 m [3.3 ft] away from: doors, windows, air supply/exhaust outlets,
air purifiers, or other potential influences (e.g., humidifiers, cleaning supplies, printers and
photocopiers). To the extent possible, sampling points should be at least 5 m [16.4 ft] from
exterior doors.
•Additionally, monitors measuring temperature and relative humidity must be at least 1 m [3.3
ft] away from: direct sunlight, mechanical system supply outlets, fans, heaters or any other
significant source of heat or cold.
Source: WELL Performance Verification Guidebook Q4 2021

WB55 Manual –revision 2.8 5
3.3 Mounting System
The WB55 can be wall mounted, using the included mounting plate, with anchor points on the
baseplate. It is recommended that where possible all 4 mounting holes are used when securing the
unit to a wall.
Please note: The unit is designed to be wall mounted vertically - long term use of the unit in the
incorrect orientation will result in dust build up within the particle sensor (if a particle sensor is
included in your configuration).
Figure 2
→
Wall mounting points for the WB55
4. Power Supply
The WB55 Sensor includes a 12V power supply adapter. The adapter converter can operate at 100 –
240VAC (47-63Hz) and is compatible with the mains power network of most continents.
5. Internet Connection
5.1 Wireless Ethernet Gateway Connection
Your wireless monitor will need to be in range of the gateway which may be up to 30m depending on
the environment.

WB55 Manual –revision 2.8 6
•To set up the Gateway please connect the Ethernet cable provided to the DATA HUB and then
connect to an Ethernet point or a spare Ethernet port on your router.
•Power on the device after connecting the supplied power supply. The device will power on
automatically and establish a connection with the WB55 Sensor.
Please note: It can take up to 5 minutes for the gateway to pair with the monitor. You can
check its status by viewing the sensor’s dashboard for data arriving online.
5.2 Network Configuration
The DATA HUB will also by default automatically configure itself to your network settings using DHCP.
It is possible to configure the sensor to connect to a static IP address. Please see page 12 of this manual
to complete this step.
6. Online Software Setup
6.1 Online Account Setup
To set up your online account to remotely monitor your IAQ monitor please navigate to
https://hex2.nuwavesensors.com/ on your computer internet browser.
On the webpage you will be prompted to sign in or create an account. As this is your first time to
access the account please click ‘Create Account’ just under the sign in section.
Figure 3
→
Login Interface on Webpage

WB55 Manual –revision 2.8 7
6.2 Account Registration
Please complete the form in order to complete the sign-up process. If you have any issues, please
(found on the sticker on the back of both devices).
Figure 4
→
Account Registration Form
6.3 Adding a monitor to your account
To add your new sensor, click ‘Add Device’ from the Manage Devices Page and complete the form
based on your sensor details;
•Device ID:Please enter the 16-digit sensor ID (sticker on the back of your sensor)

WB55 Manual –revision 2.8 8
•Name: Example; Cleanroom 2A
•Label: Completing this field allows you to create groups of sensors based on your preference
–example; 1st Floor. You can also leave this blank if you do not wish to create a group.
Figure 5
→
‘Add Sensor’ Interface Page
Once you have competed the above form click ‘Add Sensor’ button at the end of the form and your
sensor will be added. To add another sensor at any time please repeat the steps above.
7. Viewing and Managing Sensor Data
After setting up the sensor and gateway correctly as well as adding your sensor after logging in you
should now start to see data arriving on the sensor monitoring page. If you do not see data arriving
after a period of 15 minutes, please refer to the gateway configurations setting in the troubleshooting
section of this manual (page 17).
7.1 Sensor Status
Here you can view the current VOC reading as well as the current temperature and humidity reading
from your sensor.

WB55 Manual –revision 2.8 9
Figure 6
→
Sensor Status Examples
7.2 Graphing Options
Data arriving on the website will be visible on the sensor graph. You can change between graph views
of 1 hour, 12 hour or 24 hours.
Figure 7
→
Example Graph

WB55 Manual –revision 2.8 10
7.3 Indoor Air Quality Bands (RAG Status)
Each parameter is represented by a widget on your dashboard. Each widget displays a current RAG
status band (Red, Amber or Green) that changes in response to live sensor readings and indicates if
the status is considered to be good, fair or poor.
Figure 8
→
Examples of Sensor Widgets
Guidance on Indoor Air Quality performance bands is constantly evolving and we aim to reflect current
best practice. The default settings are shown below and bespoke settings may be available, please
enquire with your NuWave representative.
Figure 9
→
IAQ Performance Bands

WB55 Manual –revision 2.8 11
7.4 Export CSV data
For a longer-term view of the data or for detailed analysis you can download a CSV file of your data at
any time.
Due to download file size limitations this is limited at 1 weeks’ worth of data per download. The link
to download your data will be e-mailed to the e-mail address registered with the account.
Figure 10
→
Export Feature
7.5 Sensor Management Settings
At the bottom of each sensor you will find the sensor management settings. From here you can
manage settings such as re-naming the sensor.
Figure 11
→
Management Settings
Note: To save and changes make sure and click ‘Submit Change’ at the bottom of the form.
Project Location Settings:
The project locationcan be manually set on the map by editing the location pin. Press the Edit Button
to access editing settings.

WB55 Manual –revision 2.8 12
Figure 12
→
Project Location Settings
Once the location has been selected, press ‘Submit Changes’ to save the project location.
Time and Date Settings:
The time and date setting change the time zone settings for a particular device according to its
location. Once this is changed the data on the graph appears in this time zone.
Please note that this does not affect the time zone data of any data downloaded by CSV which will
always be saved in UTC format.
Figure 13
→
Time and Date Settings

WB55 Manual –revision 2.8 13
8. User Profile Settings
On the settings page you can edit and manage your user account details including;
•Change Password
•Change e-mail address associated with the account
•Address Location
Once any changes have been made click the ‘Save Changes’ button.
Figure 14
→
Account Settings Page
9. API
NuWave Sensors offers a RESTful API to monitor devices connected to our HEX monitoring software.
Resources regarding the API for your installed monitors are available at the following link address:
https://hex2.nuwavesensors.com/api/docs/
10. Gateway Network Configuration
The DATA HUB gateway is configured to use DHCP by default. This automatically detects network
settings on most standard networks and the sensor will be able to send data online without changing
any settings.

WB55 Manual –revision 2.8 14
If you need to configure a port on your network for the gateway the port and server address details
are as follows:
Remote Port Number: 58122
Remote Server Address: gateway2.nuwavesensordata.com
You can also edit the network settings and assign a static IP using the gateway web interface of the
gateway which is accessible using an internet browser. To access the gateway you must know the IP
address which can be found using the MAC address of the gateway (located on the bottom of
gateway).
When prompted, enter the following username and password;
Username: ‘admin’
Password: ‘admin’ or ‘alpha’
If you have any issues, please contact info@nuwavesensors.com
Figure 15
→
Gateway Homepage
Figure 16
→
Gateway IP Configuration Page

WB55 Manual –revision 2.8 15
Figure 17
→
Gateway TTL Configuration Page
Figure 18
→
Gateway Misc Configuration Page
gateway2.nuwavesensordata.com
/

WB55 Manual –revision 2.8 16
Static IP Settings - Reset to DHCP
After setting a static IP address on the gateway and it is no longer accessible on the network you can
reset the gateway back to using DHCP using the following instructions:
1. Unplug the device.
2. Plug the device back on while using a reset pin or paper clip to press and hold the reset button
3. The LEDs will flash rapidly before changing to a cyan colour to indicate the reset is underway
–the reset button can now be released.
4. After a few minutes the LED will revert back to its normal state of green before trying to
establish a connection with devices and the internet.
5. The gateway should be accessible on the network again. Please login to ensure that DNS
settings are set to ‘auto’ (the greyed out settings are ignored). Finally, reboot the gateway to
complete the process.
Figure 19
→
Reset to DHCP
11. Calibration
Our sensors come factory calibrated and in most cases your sensors can be used for up to 3 years
without calibration.
Calibration may be required in the following circumstances:
•You may require calibration for compliance purposes such as compliance with the Well
Building Standard. In this case we recommend an annual calibration service
•If the devices are used in very polluted environments where a lot of dust is present this can
have an impact on the particle sensor accuracy
•Exposure to high levels of chemicals can have an impact on the VOC or formaldehyde sensor
•The temperature and humidity sensor will have a small amount of calibration drift annually
which can be adjusted for with a calibration
Please contact us directly in order to arrange a calibration service at support@nuwavesensors.com

WB55 Manual –revision 2.8 17
12. Important Precautions
Caution! This device is recommended for use indoors and in a dry location only.
•Avoid using the WB55 in very dusty environments as dust build up may obstruct the air flow
within the device.
•Take care when using the WB55 to route the power cable in a way that reduces the risk of
injury to others, such as by tripping or choking.
•Do not cover or obstruct vents around the WB55 sensor.
•Only use the power adapter supplied with the WB55.
•Do not insert anything through the vents.
•Do not inject gas, dust or chemicals directly into the WB55 monitor.
•Do not use this device near water.
•Do not drop or subject the device to undue shock.
•Do not place in insect-infested areas. Insects can block vent openings to the sensors.
The WB55 is designed to be maintenance free, but you should keep it clean and avoid dust build up -
especially around the air vents of the sensor which can reduce performance.
To clean the WB55:
1. Turn off mains power and remove the power adapter plug from the WB55.
2. Wipe the outside with a clean, slightly damp cloth. Don’t use soap or solvents!
3. Very gently vacuum around the vents of the WB55 sensor to remove dust obstructing the
vent openings.
Note:
•Never use detergents or solvents on your WB55 sensor or spray air fresheners, hair spray or
other aerosols near it.
•Do not place the WB55 in water or let water get inside the WB55 sensor.
•Do not paint your WB55 sensor.
13. Recycling and disposal
The WEEE symbol on this product means that the WB55 should be disposed of separate from ordinary
household waste at the end of its life in accordance with local regulations. Please take the WB55 to a
collection point designated by your local authority to be recycled to help conserve natural resources.

WB55 Manual –revision 2.8 18
14. Appendix
14.1 Product Warranty
Limited Product Warranty
THIS LIMITED WARRANTY CONTAINS IMPORTANT INFORMATION ABOUT YOUR RIGHTS AND
OBLIGATIONS, AS WELL AS LIMITATIONS AND EXCLUSIONS THAT MAY APPLY TO YOU AS PART OF THE
TERMS AND CONDITIONS OF SALE IN EFFECT AT THE TIME THAT YOU PURCHASE A NUWAVE SENSOR
TECHNOLOGY LIMITED PRODUCT.
What this warranty covers?
NuWave Sensor Technology Limited (“NuWave”) warrants to the original purchaser of this WB55
sensor (the “Product”) shall be free of defects in design, assembly material, or workmanship under
normal use for a period of two (2) years from the date of purchase (the “Warranty Period”).
NuWave does not warrant that the operation of the Product will be uninterrupted or error-free.
NuWave is not responsible for damage arising from failure to follow instructions relating to the
Product’s use. This Limited Warranty does not cover software embedded in the Product and the
services provided by NuWave to owners of the Product. Refer to the licence agreement accompanying
the software for details of your rights with respect to their use.
Remedies
NuWave will repair or replace, at its option, any defective product free of charge (except for shipping
charges for the product). Any replacement hardware product will be warranted for the remainder of
the original warranty period or thirty (30) days, whichever is longer. In the event that NuWave is
unable to repair or replace the product (for example, because it has been discontinued), NuWave will
offer either a refund or a credit towards the purchase of another product from NuWave in an amount
equal to the purchase price of the product as evidenced on the original purchase invoice or receipt.
What is not covered by this warranty?
The warranty is null and void if the Product is not provided to NuWave for inspection upon NuWave’s
request, or if NuWave determines that the Product has been improperly installed, altered in any way,
or tampered with. The NuWave Product Warranty does not protect against floods, lightning,
earthquakes, war, vandalism, theft, normal-use wear and tear, erosion, depletion, obsolescence,
abuse, damage due to low voltage disturbances such as brownouts, non-authorised program or
system equipment modification, alternation or other external causes.
How to Obtain Warranty Service
Please review the online help resources at airsentric.com/resources/ prior to seeking warranty
service. To get service for your WB55 sensor you must take the following steps;
1. Contact NuWave Sensors customer support. Contact information can be found by visiting
airsentric.com
2. Provide the following to the customer support agent;
a. The serial number found on the back of your WB55 sensor
b. Where you purchased the product
c. When you purchased the product
d. Proof of payment
3. Your customer service representative will then instruct you on how to forward your receipt
and your WB55 as well as how to proceed with your claim.

WB55 Manual –revision 2.8 19
It is likely that any stored data relating to the product will be lost or reformatted during service and
NuWave will not be responsible for any such damage or loss.
NuWave reserve the right to review the damaged NuWave product. All costs of shipping the Product
to NuWave for inspection shall be borne by the purchaser. Damaged equipment must remain available
for inspection until the claim is finalised. Whenever claims are settled NuWave reserves the right to
be subrogated under any existing insurance policies the purchaser may have.
Implied Warranties
EXCEPT TO THE EXTENT PROHIBITED BY APPLICABLE LAW, ALL IMPLIED WARRANTIES (INCLUDING
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE) SHALL BE LIMITED
IN DURATION TO THE DURATION OF THIS WARRANTY.
Some jurisdictions do not allow limitations on the duration of an implied warranty, so the above
limitation may not apply to you.
Limitation of Damages
IN NO EVENT SHALL NUWAVE BE LIABLE FOR INCIDENTAL, SPECIAL, DIRECT, INDIRECT,
CONSEQUENTIAL OR MULTIPLE DAMAGES SUCH AS, BUT NOT LIMITED TO, LOST BUSINESS OR PROFITS
ARISING OUT OF THE SALE OR USE OF ANY NUWAVE PRODUCT, EVEN IF ADVISED OF THE POSSIBILITY
OF SUCH DAMAGES.
Statutory Rights
This warranty gives you specific legal rights, and you may have other rights, depending on your
jurisdiction. These rights are not affected by the warranties in this Limited Warranty.
14.2 Troubleshooting
Problem
Possible Issue
Solution
No data arriving online
after 5 minutes
Outside of wireless range
The wireless range of the sensor can vary
greatly depending on the environment. To
test this please plug the WB55 in close range
to the DATA HUB. Data should arrive online
after about 5 minutes. In this case the Data
Hub will need to be re-positioned closer to
the sensor location.
Ethernet Cable not firmly
connected
Power off both the DATA HUB and WB55
SENSOR by plugging out the power supplies.
Please ensure that the Ethernet cable is
firmly connected to both the DATA HUB
gateway and the port on your broadband
router. Apply power to both devices and
check to see if data arrives after 5 minutes.
Figure 20
→
Troubleshooting Table
Table of contents
Popular Accessories manuals by other brands

Axor
Axor Massaud Flower Vase 42274000 Dimensions

Palram
Palram Sunscape installation guide

S+S Regeltechnik
S+S Regeltechnik Thermasgard ALTF 1 Operating Instructions, Mounting & Installation

Nortek
Nortek 2GIG Series installation instructions

Thule
Thule Omnistor 8000 installation manual

Pil Sensoren
Pil Sensoren P41-350-I-CM12 manual