OfiTE 130 User manual

Model 130 Benchtop Consistometer
#120-90 - Standard
#120-90-DAS - With Computer for Data Acquisition and Control
Instruction Manual
Updated 4/24/2018
Ver. 6.0
OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 · Fax: 713.880.9886 · www.ote.com
©Copyright OFITE 2015

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 1
Intro..................................................................................................2
Description......................................................................................2
Features...........................................................................................2
Requirements .................................................................................2
Specications .................................................................................3
Components ...................................................................................3
Setup................................................................................................5
Consistometer.............................................................................5
Chart Recorder ...........................................................................6
Eurotherm .................................................................................10
Potentiometer Indicator.............................................................12
Software ........................................................................................13
Setup.........................................................................................13
Calibration.................................................................................16
Operation ..................................................................................17
Test Builder ...............................................................................18
Test Data...................................................................................19
Operation.......................................................................................20
Filling the Slurry Cup.................................................................20
Loading the Test Cell ................................................................22
Completing the Test ..................................................................25
Maintenance..................................................................................26
Potentiometer ...........................................................................27
Appendix .......................................................................................29
Potentiometer Calibration .........................................................29
Potentiometer Adjustment.........................................................31
Diagrams...................................................................................32
Warranty and Return Policy ........................................................34
Table of
Contents

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 2
During cementing operations, the time required for a cement slurry to set is
of primary concern. Under an ideal situation, minimal time would be required
to successfully pump the slurry, which immediately upon placement, begins
to develop compressive strength. However, if insufcient time is allowed to
fully pump the cement, it will be necessary to drill the cement remaining in the
casing string. Remedial operations such as this are very costly. Conversely,
cements that are successfully placed, but require considerable time to
cure, consume valuable rig time, which is also quite costly. Laboratory tests
should be conducted under simulated reservoir conditions to examine the
actual thickening time of the slurry. The OFITE Benchtop Consistometer
was specically engineered to determine the thickening time of well cements
under simulated downhole pressures and temperatures.
A cement is mixed and poured into the slurry cup assembly. The slurry cup
is placed into the test vessel and pressure is increased via an air-driven
hydraulic pump. A PID temperature controller governs an internal heater,
which maintains the necessary temperature prole, while a magnetic drive
mechanism rotates the slurry cup assembly at 150 RPM. A potentiometer
controls an output voltage, which is directly proportional to the amount of
torque the cement exerts upon an API-approved paddle. A chart recorder
registers cement consistency and temperature as a function of time.
Temperature and consistency are digitally displayed via LED indicators.
- Maximum Pressure: 16,000 PSI (110.3 MPa)
- Maximum Temperature: 400°F (204.4°C)
- Pressure generated via an air-driven hydraulic pump
- Drive table is rotated with a magnetic drive
- External cooling jacket aids cooling of test cell
- Electronic timer with alarm, elapsed 0.1 minute resolution
- Deadweight calibration unit included
- Temperature and consistency alarms provide automatic shutdown
- Safety head with rupture disk are provided
- Unit is fully capable of testing cements in strict accordance to the
guidelines as stated in API Specication 10
- Air/Nitrogen Supply (100 - 150 PSI / 690 - 1,035 kPa)
- Water Supply for Cooling (40 PSI / 276 kPa)
- Water Drain
- 220 Volt, 50/60 Hz, 25 Amp electrical power supply
Intro
Description
Features
Requirements

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 3
Specications Size 25 × 16 × 20 inches (63.5 × 40.6 × 50.8 cm)
Weight 215 lb (94.6 kg)
Crated Size 30 × 20 × 24 inches (76.2 × 50.8 × 61 cm)
Crated Weight 255 lb (115.8 kg)
Temperature Controller Digital PID, 1° Resolution
Internal Heater 2,500 Watt (5,000 Watt Available)
Slurry Cup 150 RPM Rotational Speed; 316 Stainless Steel;
Expansion Chamber
Maximum Temperature 400°F (204.4°C)
Maximum Pressure 16,000 PSI (110.3 MPa)
#120-00-026-1 Timer
#120-00-028 Consistency Indicator, Eurotherm 2408i
#120-001 Mineral Oil, 1 Gallon, Qty: 3
#120-104 Rupture Disk, 17,500 PSI
#120-208-1 Slurry Cup Thermocouple
#120-519 Slurry Cup Assembly, No Expansion Cap
#120-502 Molded Diaphragm
#120-503 Paddle Pin
#120-506 Paddle
#120-90-5 Chart Recorder, Eurotherm 6100A
#120-90-033 Air Filter
#120-90-035 Filter
#120-90-63 Temperature Controller, Eurotherm 3504
#130-79-15 Serial Cable
#130-76-16 USB Cable
#152-38 Power Cable
#120-628 Potentiometer Assembly:
#120-602 Calibration Spring
#120-603 Potentiometer Body
#120-604 Potentiometer Resistor
#120-605 Contact Spring
#120-606 Potentiometer Contact Arm
#120-607 Contact Strip
#120-608 Grounding Cable Retaining Screw
#120-609 Grounding Contact Spring
#120-75-10 Slotted Weight Set
#120-75-9 Weight Hanger
Components

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 4
#120-90-00 Cell Assembly:
120-90-049 Cell Cap Backup Ring
120-146 Mag Drive O-ring
120-147 Mag Drive
120-148 Retaining Ring
120-149 Test Cell O-ring
120-206 2,500 Watt Heater
120-257 Drain Plug O-ring
503-258V90 O-Ring, Qty: 2
#120-91 Spare Parts for #120-90:
#120-001 Mineral Oil, 1 Gallon, Qty: 2
#120-104 Rupture Disk, 17,500 PSI, Qty: 3
#120-105 High-Pressure Check Valve
#120-146 Mag Drive O-ring, Qty: 12
#120-148 Cell Cap Backup Ring, Qty: 2
#120-149 Test Cell O-ring, Qty: 12
#120-204 Heater Gasket, Qty: 2
#120-208-1 Slurry Cup Thermocouple, Qty: 2
#120-257 Drain Plug O-ring, Qty: 12
#120-501 Slurry Cup Sleeve, Qty: 2
#120-502 Molded Diaphragm, Qty: 25
#120-503 Paddle Pin, Qty: 24
#120-504 Pivot Bearing, Qty: 10
#120-505 Pivot Bearing Gasket, Qty: 5
#120-506 Paddle, Qty: 6
#120-507 7¾" Paddle Shaft, Qty: 10
#120-508 Diaphragm Retaining Ring, Qty: 6
#120-509 Drive Disc
#120-510 Drive Bar
#120-511 Slurry Cup Shear Pin, Qty: 24
#120-512 Slurry Cup Drive Pin, Qty: 12
#120-513 Slurry Cup Gasket, Qty: 12
#120-519 Slurry Cup Assembly, No Expansion Cap
#120-602 Calibration Spring, Qty: 6
#120-604 Potentiometer Resistor, Qty: 6
#120-606 Potentiometer Contact Arm, Qty: 6
#120-607 Contact Strip, Qty: 6
#120-628 Potentiometer Assembly

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 5
Setup
Consistometer
1. Carefully remove the instrument from the wooden crate and place it on a
stable surface.
2. Connect an air or nitrogen (100 – 150 PSI / 690 – 1,035 kPa) supply to the
air supply on the back of the instrument.
This unit uses ¼" NPT female connectors for all supply lines.
3. Connect the drain and coolant supply lines, also on the back of the unit.
4. Make sure all electrical switches are turned off and the unit is grounded.
Make the necessary electrical connections in accordance with local
codes.
5. Plug the cable into the top port on the unit cabinet and connect the other
end to the port on the back of the Control Box.
6. If you are using a computer, connect the Control Box to the computer with
the supplied cable. Newer units connect via a USB cable (#130-79-16).
Older units use an RS-232 (serial) cable (#130-79-15).
7. To ll the oil reservoir, remove the oil reservoir cap and pour mineral oil
into the reservoir until it is full. Replace the cap. Make sure the seal is air
tight. Use the sight glass on the side of the reservoir to check the oil level.
The oil level should be about ½" from the top of the sight glass.
Data Port
Potentiometer
Calibration
Port
Thermocouple
Port
Oil Reservoir
Cap
Test Cell
Air To Cylinder
Oil Reservoir Valve
Pressure Release

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 6
Setup
Chart Recorder
The OFITE Benchtop Consistometer includes a Eurotherm Chessell 6100A
chart recorder for displaying and recording test data. It features a removable
drive for easily transferring test data to a PC for processing. The unit is setup
to automatically record data onto the disk during the test. However, for this
feature to work, you must have the disk inserted into the drive when the unit
is powered on. Otherwise, the data must be manually archived at the end of
the test.
Recorder Disk
Eject Button
Root Menu
Button
It is strongly recommended that you carefully study the Eurotherm
Chessell 6100A instruction manual before using this equipment.
To manually archive test data:
1. Press the “Root Menu” button.
2. Press “Operator” from the root menu.
3. At the top of the screen, choose “Archive” and then choose “Disk”.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 7
The chart recorder is shipped to you with a pre-programmed conguration
le. It is highly recommended that you do not change any of these settings.
However, in case of emergency, it is possible to restore your conguration
from the disk that is shipped with the unit.
1. At the top of the screen, press “Save/
Restore” and choose “Restore”.
2. Make sure all four options are selected,
then touch the “File Name” eld.
3. Press the up arrow to access the root
directory.
4. Select “Removable” and press the down
arrow to access the disk.
5. Choose the le titled “Cong” and press
“Open”.
6. When the progress bar at the top of the
screen stops, the restore process is
complete.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 8
To transfer the test data to a PC:
1. Remove the disk from the chart recorder and insert it into the appropriate
drive on the PC.
2. From the Start Menu, select “Programs” then “Eurotherm” and then click
“Review”. This will open the Eurotherm Review software application.
3. From the “File” menu, select “Transfer” and then click “Files”.
4. Click the “Browse” button, then choose the appropriate drive for your
removable media. Open the “History” folder and select the les you wish
to transfer.
5. Type a name in the “Name” eld and click “OK”.
At this point you will receive a warning message. Click “OK” again.
6. When the le transfer is complete, go to the “File” menu and click “New
Chart”.
7. Click “Add Point”.
8. From the “Instrument” drop-down menu, choose the name you selected in
step 5.
9. Choose “Group 1” from the “Log Group” menu.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 9
12. To jump directly to a specic data point, click
the “Go To” button at the top of the screen.
Choose the data point you wish to view and
click “OK”.
10. Now, select the Point IDs you wish to display on the chart. Hold down the
“CTRL” button to select more than one. Click “OK”.
11. The software will now create the chart based on the data collected from
the test.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 10
Setup
Eurotherm
The Eurotherm Model 3504 controls the temperature and displays the
pressure. When running a test with the computer, the software automatically
programs the controller. However, when running a test without the computer,
the controller must be programmed manually. It is strongly recommended
that operators carefully study the Model 3504 instruction manual before
attempting to program the controller.
When running a test without the computer, always build the program prior to
mixing the slurry and loading the slurry cup.
The Eurotherm 3504 can store up to eight custom programs. Each program
consists of multiple segments. In a segment, the temperature setpoint can be
changed or held for a certain amount of time.
Each program also can consist of two channels; one for temperature and one
for pressure. Since the Benchtop Consistometer does not offer automatic
pressure control, only Channel 1 should be programmed.
Selector
Temperature
Pressure
Page Scroll
To edit a program:
1. Press the “Page” button until the display reads “Program Edit”.
2. Use the up and down arrows to select the program you want to edit.
3. Press the “Scroll” button to move to the Channel parameter.
4. Use the up and down arrows to select Channel 1.
Programming Channel 2 is not supported on the Benchtop
Consistometer.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 11
5. Set the parameters for the test:
Press the “Scroll” button to select a parameter to edit. Then use the up
and down arrows to modify the value of that parameter.
a. Set the following parameters for each test:
- Ramp Units: Seconds, Minutes, Hours
- Cycles: The number of times to repeat the test.
b. You must choose a Segment Type for each segment in the test.
The Segment Type will determine the parameters available for that
segment:
Rate: Increase the setpoint by a certain rate (Ramp Rate)
- Target SP: The end temperature for the segment
- Ramp Rate: The rate of increase for the setpoint. For example,
if Ramp Units is set to minutes, and Ramp Rate is set to 5, the
temperature setpoint will increase by 5° per minute.
Time: Increase the setpoint over a certain time (Ramp Time)
- Target SP: The end temperature for the segment
- Duration: The amount of time it will take to increase the setpoint to
the target.
Dwell: Hold the current setpoint.
- Duration: The amount of time to hold the current setpoint.
Step: Increase the setpoint immediately. No ramp is involved.
- Target SP: The end temperature for the segment.
End: The nal segment in every test must be End.
- End Type: Dwell is recommended. This will hold the current
setpoint until the test is stopped.
The display is split into two areas. The top area shows the temperature
channel while the bottom shows the pressure channel. Before starting a
manual test, the temperature channel must be in auto mode and the pressure
channel must be in manual mode:
1. Press the “Scroll” button to move the selector around the screen.
When the selector is above the dividing line, the temperature channel
is selected. When it is below the dividing line, the pressure channel is
selected.
2. Press the “A/MAN” button to toggle the selected channel between auto
and manual mode.
In the temperature channel, the word “MAN” will appear at the top of the
display to indicate manual mode. This word will disappear in auto mode.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 12
Setup
Potentiometer Indicator
The Benchtop Consistometer includes a built-in Eurotherm 2408i 15V
Potentiometer Indicator. As the slurry cup rotates within the unit, the
potentiometer creates a small voltage charge, which is displayed on the
indicator. This charge increases as the strength of the slurry increases.
The potentiometer indicator features a high-voltage alarm. When the charge
reaches the specied level, an alarm condition is triggered and the heater,
pump, and motor are automatically shut off.
To change the alarm setting, perform the following steps:
1. Push the “Page” key twice. The display will read “AL”.
2. Push the “Enter” key. The display will read “IFSH”.
3. Push the up arrow key once to read the current setting.
4. Push the up or down arrow keys to change the setting.
5. After choosing the appropriate setting, press the “Page” key three times
to return to the original display.
Enter Key
Page Key

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 13
Software
Setup
1. Double-click the “Multi Consistometer” icon on the PC desktop.
2. To add a new unit, click “Add New”.
3. Select “Consistometer” from the list.
4. Choose a Communication Type: Serial or USB. Newer units connect to
the computer via a USB cable. Older units use an RS-232 (serial) cable.
5. Enter a name. The name must be different from all other Consistometers
connected to this PC.
6. Enter a slave ID. The slave ID must be different from all other
Consistometers connected to this PC.
7. Choose the COM port the new Consistometer is connected to.
8. Chose a folder to save the test data. It is recommended that each
Consistometer have its own data archive folder.
9. Choose a logo le (.jpg) to display on the printed chart.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 14
10. Make sure “With Pressure” is selected and “With Pressure Control” is
not. Pressure monitoring is available on the Benchtop Consistometer, but
pressure control is not.
11. Click “Finish” then “Save”.
12. The main screen shows all units added to the PC. For each one it shows
the type of equipment, name, slave ID, COM port, and data archive path.
Units with “[R]” after their name are currently running a test.
Right-click on a unit and select “Launch” from the menu.
13. Select “Setup” from the “Options” menu.
14. Select units for Temperature (°F or °C), Consistency (Voltage or Bc), and
Pressure (PSI, KSI, or kPa).
15. Choose an “Archive Path”. This is the folder where all test data will be
stored.
16. Choose a “Chart Logo File”. This is the logo that will print on the report at
the end of a test. This le must be in .jpg format.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 15
17. Select “Print to Printer” if you want a chart to automatically print when
you stop a test. Select “Print to File” if you want a .jpg le to be create
automatically when you stop a test.
Both of these options can be enabled at the same time. It is
recommended that “Print to File” always be enabled.
18. Choose values for the “Consistency #1”, “Consistency #2” and
“Consistency #3” elds.
These values tell the software to record the test time when the
consistency reaches each point. A note will be made on the chart when
the test is nished.
Make sure that if you select “Bc” for consistency units that you also use
Bc for these two elds. Likewise, if you chose “V” for consistency units,
you must use voltage for these two elds.
19. Select “Automatic stop collecting data” if you want the Consistometer to
stop collecting data during an alarm condition.
During an alarm condition, the motor, heaters, and pump all stop. By
default, when the alarm condition is resolved, everything restarts.
However, if the “Automatic stop collecting data” box is checked data
collection is also stopped, and only the motor and pump restart. The
heaters remain off and no more data is collected.
20. When nished, click “OK”.
21. Select “Set Alarm” from the “Utilities” menu.
22. Enter a value in Bc. If the sample reaches this consistency, the unit will
enter alarm condition.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 16
Software
Calibration
A calibration should be performed periodically to make sure the unit is
still providing accurate information. Also, be sure to calibrate any time the
potentiometer is changed. Always calibrate with the same potentiometer
that will be used during the test.
1. Select “Calibration” from the “Operations” menu.
2. Prepare the potentiometer for calibration as described on page 29.
3. Place the weight hanger (with no weights) on the end of the string. The
weight hanger is 50 g.
4. Wait for the value on the screen to stabilize and then click “Accept”.
5. The software will prompt you to add weight to the hanger. After adding the
weight, wait for the value to stabilize, then click “Accept”.
The software will graph the calibration points as you enter them. At any
time you can click the “Back” button to re-enter a calibration point.
6. Click “Save” to save the calibration data.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 17
Software
Operation
1. Click “Start Test”.
2. Enter the necessary information. The following elds do not affect your
test. The information will print on the report when the test is complete:
- Customer Info
- Additives
- Cement Mfr.
- Test Name
- Cement Density
- Cement Class
- Density Units
- Job Type
The following elds do affect your test:
- DAQ Time (mm:ss): determines how often test data is saved to le
- Alarm: sets the consistency alarm. Be sure to use the same
consistency units set in the “Setup” screen.
3. Select a “Test Prole”. Test proles are built in the Test Builder. Refer to
page 18 for instructions.
4. Click “OK” to start the test.
If the computer is disconnected from the Consistometer or loses power,
the test will not stop. Be sure to stop it manually. Refer to page 25 for
instructions on stopping a test manually.
5. When the test is complete, click “Stop”.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 18
Software
Test Builder
The Consistometer Software can control the temperature during a test. Tests
are programmed in the Test Builder.
1. Select “Test Builder” from the “Utilities” menu.
2. Click the “New Test” button to start a new test. Or click the “Copy Test”
button to start with an existing test.
3. On the Temperature tab click the “Add” button to add a step.
4. Choose a Step Type:
- Ramp: The software will increase the setpoint up to the target over
a specied time period. You will be asked for a time period (minutes)
and a target temperature.
- Step: The software will increase the setpoint to the target immediately.
You will be asked for a target temperature.
- Hold: Maintain the current setpoint for the specied period of time.
You will be asked for a time period (minutes).
5. Once you add a step, you can then go back and edit, delete, or move it
up or down in the step list.
6. Click “OK” to save the test. The new test will now be available in the “Start
Test” dialog box.

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ote.com 19
Software
Test Data
At the end of each test, all test data is saved in the “Data Archive Folder” that
was selected in the “Setup” screen. There you can nd an image showing the
chart for each test as well as a data le that can be opened in MS Excel.
Archived tests and calibrations can also be viewed in the software:
1. Select “Open Data Archive” or “Open Calibration Archive” from the “File”
menu.
2. On the left-hand side of the screen, choose a test or calibration to view.
The graph will be displayed in the chart area.
3. To print the chart, click the “Print Chart” button.
You can also print a chart or export test data during a test. Simply right-
click on the table at the bottom of the main screen. From the context
menu, select either “Print Chart” or “Export”.
a. “Print Chart” - This will print the current chart. If you have selected “Print
to Printer” in the options screen, the chart will print to the default printer.
If you have selected “Print to File”, a .jpg le will be created. See page
13 for more information about setting options.
b. “Export” - This will export the current test data to a le in the archive
folder. Refer to item 1 above for information about where to nd this le.
Context Menu
This manual suits for next models
2
Table of contents
Other OfiTE Measuring Instrument manuals