
P/N 814358 [Rev.2] 01/2015
General Warnings and Cautions
1. The MiniOX®3000 Oxygen Monitor will perform
to specifications only if it is used and serviced in
accordance with the manufacturer’s instructions.
This instrument is to be used only by qualified,
trained personnel who have carefully read the
operating manual and labels and who have observed
the information set forth. If this instrument does not
perform as described in this manual, the instrument
must not be used until the condition is rectified.
2. The MiniOX®3000 Oxygen Monitor must be calibrated
prior to each use. A two point calibration check must
be performed weekly. See Section 3, Operation. If
the instrument cannot be calibrated, the sensor must
be replaced. If the instrument is still unable to be
calibrated, the instrument must be serviced.
3. The oxygen sensor has a minimal response to
certain gases other than oxygen. Be aware of these
gases and their interference levels. SeeAppendix C,
Interferent Gases and Vapors.
4. The oxygen sensor is affected by changes in
barometric pressure. See Appendix B, Effects of
Pressure, Humidity and Temperature.
5. The sensor is a sealed unit containing a potassium
hydroxide electrolyte, which is caustic. If the sensor
should develop a leak and the contents come
in contact with skin or clothing, rinse area with
large quantities of water. In case of eye contact,
immediately flush eyes for at least 15 minutes,
holding eyes open. Call a physician. Dispose of a
leaking sensor immediately; it must be disposed of
in accordance with all applicable federal, state, and
local regulations.
6. Ensure a tight fit exists between the sensor and the tee
adapter. Ohio Medical tee adapters are engineered
to fit securely with Ohio Medical sensors. However,
the sensor retaining strap must be used to prevent
accidental separation of sensor and tee adapter.
7. Never operate the MiniOX®3000 Oxygen Monitor if it
is suspected that water or other liquids have entered
into the case. If this occurs, immediately turn the unit
OFF and contact your nearest Ohio Medical Service
Center for additional information.
8. Use of devices generating or emitting electromagnetic
radiation near the MiniOX®3000 Oxygen Monitor may
interfere with the proper operation of the product,
causing it to fail to perform as designed. Particularly,
the electromagnetic radiation from the interfering
device may cause the product to display incorrect/
erratic values or to stop operating. Special attention
should be paid to the patient if this occurs.
9. Never use the MiniOX®3000 Oxygen Monitor in
combustible atmosphere, such as occurs with
flammable anesthetics. Such use could result in
ignition of atmosphere.
10. Never allow an excess length of cable near the
patient’s head or neck, as such could result in
strangulation. Secure excess cable to the bed rail or
other suitable object.
FAILURE TO COMPLY WITH THESE WARNINGS
COULD JEOPARDIZE THE WELL-BEING OF THE
PATIENT AND/OR HEALTH CARE PROFESSIONAL.
1. Remove internal battery if unit is to be non-operational
for extended periods of time.
2. Observe polarity when inserting a new battery.
Incorrect connection may cause damage to the
instrument.
3. The MiniOX®3000 Oxygen Monitor must never be
immersed in any cleaning solution, autoclaved, or
exposed to temperatures greater than 70oC.
4. Use only genuine Ohio Medical Corporation
accessories and replacement parts. Failure to do
so may seriously impair the monitor’s performance.
Repair or alteration of the MiniOX®3000 Oxygen
Monitor beyond the scope of the maintenance
instructions or by anyone other than an authorized
Ohio Medical Corporation service person could
cause the product to fail to perform as designed.
5. Improper mounting of the sensor in a breathing circuit
may result in inaccurate readings. The sensor MUST
be mounted with the deflector pointing downward to
prevent moisture collection on the sensor membrane
(see Figures 3-3 and 3-4 in Section 3, Operation).
Install the sensor upstream from the humidifier to
minimize its exposure to moisture.
6. Never use a MiniOX®3000 Oxygen Monitor with a
cable that appears worn, cracked, or has damaged
insulation.
FAILURE TO COMPLY WITH THESE CAUTIONS
COULD RESULT IN INSTRUMENT DAMAGE AND/OR
FAILUREOFUNITTOPERFORMTOSPECIFICATIONS.
WARNING
CAUTION
MiniOX®3000 Oxygen Monitor General WARNINGS and CAUTIONS
WC-1