Olympic Medical BILI-LITE 80 User manual

OLYMPIC MEDICAL
51480-IMA-EN0103Q
Instruction Manual
OLYMPIC BILI-LITETM
Model 80/81
CAUTION
Read and be familiar with this manual before installing,
operating, or servicing this device. To ensure operator,
technician, and patient safety, use only as specified in
this manual.


OLYMPIC MEDICAL
5900 First Ave S
Seattle, WA 98108 USA
Toll-free: 1-800-426-0353 (US/Canada)
Phone: 206-767-3500 (worldwide)
Fax: 206-762-4200
Web: www.OlympicMedical.com
© 2005 Olympic Medical Corporation. All rights reserved.


OLYMPIC BILI-LITE Model 80/81 Instruction Manual i
Contents
Section 1Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Irradiance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Section 2Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Assembling the Bili-Lite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Testing the Bili-Lite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Section 3Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Operating the Bili-Lite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Positioning the Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Measuring Irradiance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Measuring Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Section 4Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Cleaning the Bili-Lite. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Cleaning the Bulb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Cleaning the Bili-Timer (Model 81) . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Replacing the Bulb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Bulb Timer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Replacing the Battery (Bili-Timer) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
LOW BAT Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Section 5Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Section 6Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Repair Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Repair Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Resetting the Thermal Protector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Replacing the Thermal Protector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Replacing the Fan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Replacing the Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Replacing the Fuse Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8

ii OLYMPIC BILI-LITE Model 80/81 Instruction Manual
Replacing the Toroidal Transformer . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Replacing the Power Switch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Replacing the Ceramic Bulb Socket. . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Returning for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
Section 7Replacement Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Section 8Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Electromagnetic Specifications – Manufacturer’s Declaration . . . . . . . . 8-3
Recommended Separation Distances Between Portable and
Mobile RF Communications Equipment and the Bili-Lite . . . . . . . . 8-5
Appendix ADrawings and Schematics . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1

OLYMPIC BILI-LITE Model 80/81 Instruction Manual 1-1
Section
1
Overview
This manual provides the necessary information to assemble, maintain, and service
the Olympic Bili-LiteTM Model 80/81. The operating and maintenance instructions
are intended for medical personnel; the service instructions are intended for
qualified service technicians.
Conventions
The following conventions are used in this manual:
Table 1.1
Conventions
Convention Description
NOTE
#Notes provide additional information to clarify a point in the text.
CAUTION
Cautions indicate situations that, if not avoided, could result in minor
to moderate injury to the patient or operator, or damage to the equipment.
WARNING
Warnings indicate situations that, if not avoided, could result in
serious injury or death to the patient or operator.
BUTTON
This character style represents buttons and controls that the user
can touch or press.

1-2 OLYMPIC BILI-LITE Model 80/81 Instruction Manual
Symbols
Symbols
The following symbols are located on the Olympic Bili-Lite and its packaging:
Table 1.2
Product symbology
Symbol Definition Symbol Definition
Alternating current (ac)
voltage
Line (hot conductor)
Atmospheric pressure Mains power OFF
Battery Mains power ON
Caution, read instructions Neutral (neutral conductor)
Electrical shock hazard Nonionizing
electromagnetic radiation
Eye protection to be worn
by infants
Shipping
Fuse Storage
Protective earth (ground) Temperature
Humidity, condensing Type B equipment

OLYMPIC BILI-LITE Model 80/81 Instruction Manual 1-3
Intended Use
Intended Use
The Olympic Bili-LiteTM Model 80/81 is a floorstand phototherapy light intended to
provide therapeutic irradiance to treat neonatal hyperbilirubinemia.
Model 81: The Olympic Bili-TimerTM is intended to measure total phototherapy
treatment time.
Description
The Olympic Bili-Lite Model 80/81 is a floorstand halogen phototherapy light. The
lighthood is adjustable from 46.5 to 63.0 in. (118 to 160 cm) above the floor.
The Model 81 includes an Olympic Bili-TimerTM, with photodetector, wiring, and
shelf for attaching the timer to the Bili-Lite. The Bili-Timer is a digital timer that
automatically measures total phototherapy exposure time. The photodetector on the
Model 81 automatically starts the timer when the phototherapy light is turned on.
Figure 1.1
Olympic Bili-Lite Model 80/81
NOTE
#While the Model 80 does not include a photodetector and cannot be equipped with an
Olympic Bili-Timer, the Models 80 and 81 are assembled, operated, and serviced in a
similar manner.
Light housing
Support post
Base
Casters
Height-adjust knob
Model 80 Model 81
Olympic Bili-Timer

1-4 OLYMPIC BILI-LITE Model 80/81 Instruction Manual
Accessories
Accessories
Olympic Medical offers the following items that are compatible for use with the
Bili-Lite Model 80/81:
Olympic Bili-MeterTM, which measures the irradiance output of phototherapy.
Olympic Bili-MaskTM, which protects the patient’s eyes during phototherapy
treatment.
To order these items, see page 7-1.
Figure 1.2
Bili-Meter with Type B-22 sensor
Figure 1.3
Bili-Mask styles
“Standard” “Easy-On”“Quick-Stick”

OLYMPIC BILI-LITE Model 80/81 Instruction Manual 1-5
Irradiance
Irradiance
The therapeutic irradiance of phototherapy lights is measured with radiometers,
instruments with optical filters that pass only selected portions of the blue-green
action spectrum of bilirubin. Irradiance (radiant power) is expressed in units of
microwatts per square centimeter per nanometer (µW/cm²/nm).
The Olympic Bili-Meter1—the radiometer widely used in nurseries—measures
irradiance from 425 to 475 nanometers (nm). Standard 60601-2-50 of the
International Electrotechnical Commission (IEC) also requires that irradiance from
400 to 550 nm be measured. The following tables show this data.
The irradiance data for the Olympic Bili-Lite Model 80/81, below, uses two
different types of blue fluorescent bulbs at varying distances. Average irradiance is
measured over an effective surface area of 9.5 in. (24 cm) in diameter. Maximum
irradiance is measured at the center of the area.
1. Each Olympic Bili-Meter is certified to be calibrated traceable to the United States National
Institute of Standards and Technology (NIST).
NOTE
#The differences between the Olympic Bili-Meter and the IEC measurements do not
represent a difference in the total irradiance output of the same bulbs. The difference
results from averaging the irradiance of the same bulbs over wavelength bands of different
widths. The Olympic Bili-Meter measures from 425 to 475 nm—nearly all irradiance of the
blue bulbs occurs in this 50-nm band. The IEC measures irradiance from 400 to 550 nm—a
150-nm band; therefore, the average irradiance per nanometer of the IEC measurement
will differ from the Bili-Meter measurement.
Table 1.3
Average irradiance using the Olympic Bili-Meter (425–475 nm)
Distance from lens to top of light sensor
20 in. (51 cm) 25 in. (64 cm) 30 in. (76 cm)
Irradiance µW/cm2/nm 14.0 8.4 4.9
Effective area (diameter) 5.7 in. (14 cm) 7.5 in. (19 cm) 9.5 in. (24 cm)
Table 1.4
Maximum irradiance using the Olympic Bili-Meter (425–475 nm)
Distance from lens to top of light sensor
20 in. (51 cm) 25 in. (64 cm) 30 in. (76 cm)
Irradiance µW/cm2/nm 17.6 12.1 7.3
Effective area (diameter) 5.7 in. (14 cm) 7.5 in. (19 cm) 9.5 in. (24 cm)

1-6 OLYMPIC BILI-LITE Model 80/81 Instruction Manual
Irradiance
Figure 1.4
Response, Olympic Bili-Meter
Table 1.5
Average irradiance using the IEC Standard (400–550 nm)
Distance from lens to top of light sensor
20 in. (51 cm) 25 in. (64 cm) 30 in. (76 cm)
Irradiance µW/cm2/nm 18.5 11.0 6.4
Effective area (diameter) 5.7 in. (14 cm) 7.5 in. (19 cm) 9.5 in. (24 cm)
Table 1.6
Maximum irradiance using the IEC Standard (400–550 nm)
Distance from lens to top of light sensor
20 in. (51 cm) 25 in. (64 cm) 30 in. (76 cm)
Irradiance µW/cm2/nm 23.2 16.0 9.7
Effective area (diameter) 5.7 in. (14 cm) 7.5 in. (19 cm) 9.5 in. (24 cm)
Wavelength (nm)
Response (%)

OLYMPIC BILI-LITE Model 80/81 Instruction Manual 1-7
Irradiance
Figure 1.5
Spectral energy distribution, halogen bulb
Figure 1.6
Effective surface area, Olympic Bili-Lite Model 80/81
Wavelength (nm)
Relative irradiance (µW/cm²)
At 20 in. (50 cm) distance
5.7-in. (14-cm) diameter
At 25 in. (64 cm) distance
7.5-in. (19-cm) diameter
At 30 in. (76 cm) distance
9.5-in. (24-cm) diameter
NOTE
#Effective surface area shown directly overhead. If the lamp is angled toward the patient, the
effective area will elongate proportionately to the angle.

1-8 OLYMPIC BILI-LITE Model 80/81 Instruction Manual
Irradiance

OLYMPIC BILI-LITE Model 80/81 Instruction Manual 2-1
Section
2
Assembly
Assembling the Bili-Lite
The Bili-Lite is shipped with the following:
Arm assembly, with bulb installed
Support post
Base
Caster wheels with hardware packet
Required Items:
Assemblies listed above
Scissors
Phillips-head screwdriver
1/2-in. socket driver
To assemble the Bili-Lite:
1Install the caster wheels:
aPlace the base with the bottom facing up.
bScrew the casters into the base, then tighten.
2Fully tighten the height-adjust knob on the support post, then cut the cable ties
that secure the post to the arm (see Figure 2.1).
3Secure the arm assembly (see Figure 2.1):
aWithout twisting the power cord, place the stub of the arm assembly into the
upper support post with the holes aligned.
CAUTION
Only qualified technicians should assemble this device.
Read and be familiar with this instruction manual before assembling this device.
WARNING
The spring inside the support post is compressed. Releasing the height-adjust knob
will permit the support post to extend suddenly and forcefully a distance of 16 in.
(40.6 cm). Fully tighten the height-adjust knob before cutting the cable ties that
secure the post to the arm. To adjust the height, securely hold the support post while
loosening the height-adjust knob.

2-2 OLYMPIC BILI-LITE Model 80/81 Instruction Manual
Assembling the Bili-Lite
bSecure with the two 8-32 x 3/8 in. screws.
Figure 2.1
Assembling the Bili-Lite (Model 81 shown)
4Attach the power cord to the support post:
aLoosen the fastener securing the cord clip to the support post (see
Figure 2.1).
bPlace the power cord in the cord clip. Allow as minimum amount of cord
slack as possible between the clip and the arm assembly (see Figure 2.2),
then retighten the fastener.
Arm assembly
Power switch
Cord clip
Height-adjust knob
Access cover
Base
Wire connectors
Support post
Bili-Timer shelf
(Model 81)
Bili-Timer
(Model 81)
Cord cleat
Caster wheels
Light housing
NOTE
#Wrap cord around post as shown in Figure 2.1.
If assembling a Model 81, skip this step. You will attach the power cord in step 6 on
page 2-3.

OLYMPIC BILI-LITE Model 80/81 Instruction Manual 2-3
Assembling the Bili-Lite
Figure 2.2
Positioning the power cord
5Assemble the base:
aSet the base on a level surface, then slide the access cover up the lower
support post about 6 in. (15.2 cm) (see Figure 2.3).
Figure 2.3
Assembling the post and base
bLoosen the two column-securing screws in the post socket on the base (see
Figure 2.3), then insert the support post into the post socket and tighten the
screws.
cInsert the hex bolt and washer through the bottom of the base and into the
support post (see Figure 2.3), then use the socket driver to securely tighten
the bolt.
dPlug the wire connectors from the bottom of the support post into the wires
in the base, male to female.
eAttach the access cover to the base cover with the four screws and lock
washers. Use care not to pinch the wiring between the access cover and
base cover.
6If assembling a Model 81, attach the Bili-Timer:
aRemove the cord clip fastener.
Minimum amount of
slack in power cord
Cord clip
Access cover
Post socket
Hex bolt and washer
Column-securing screws (x2)
Wire connectors
Screws (x4)
Bulb timer
CAUTION
Grounding integrity is dependent on proper assembly and tightening of all bolts
and screws.

2-4 OLYMPIC BILI-LITE Model 80/81 Instruction Manual
Assembling the Bili-Lite
bAttach the Bili-Timer shelf with two flathead screws (see Figure 2.1 on
page 2-2) using one of the screws to reattach the cord clip fastener.
cPlace the power cord in the cord clip. Allow as minimum amount of cord
slack as possible between the clip and the arm assembly (see Figure 2.2 on
page 2-3), then retighten the fastener.
dPlace the Bili-Timer on the shelf and plug it into the connector next to the
power switch.
7Test for proper operation (see page 2-5).

OLYMPIC BILI-LITE Model 80/81 Instruction Manual 2-5
Testing the Bili-Lite
Testing the Bili-Lite
Perform the test procedure to ensure the Bili-Lite is working properly after
performing assembly and service procedures.
To test the Bili-Lite:
1Plug in the power cord, then press the power switch to the
ON
(|) position.
2Confirm the following:
The bulbs illuminate.
The fan is on and air is flowing from the end of the bulb housing.
3Check the bulb timer:
aTurn off the Bili-Lite.
bCheck the number on the bulb timer (see Figure 2.3).
cTurn on the Bili-Lite, wait six minutes, then turn off the Bili-Lite.
dCheck the number on the bulb timer. Verify that the timer has progressed
0.1 hours.
4Test the Bili-Timer (Model 81):
aWith the phototherapy lamp off, press
POWER
to turn on the Bili-Timer.
Confirm that the display shows 000.0, and the colon remains steady (i.e., is
not blinking).
bTurn on the phototherapy light. Confirm that the colon blinks and, after
approximately six minutes, the display shows 000.1 (i.e., one-tenth of an
hour).
cPress
ZERO
. Confirm that the display shows 000.0.
dPress
POWER
to turn off the Bili-Timer.
5Move the light housing to several positions—vertical and rotational; verify that
it remains at each set position.
6Press the power switch to the
OFF
(O) position to turn off the Bili-Lite.
NOTE
#If the device does not operate as expected, see Troubleshooting on page 5-1.

2-6 OLYMPIC BILI-LITE Model 80/81 Instruction Manual
Testing the Bili-Lite
This manual suits for next models
1
Table of contents
Other Olympic Medical Medical Equipment manuals