Omron HBP-1100 User manual

Professional Blood Pressure Monitor
HBP-1100
•InstructionManual
•Moded’emploi
•Gebrauchsanweisung
•Manualediistruzioni
•Manualdeinstrucciones
•Gebruiksaanwijzing
•
РУКОВОДСТВОПОЭКСПЛУАТАЦИИ
•KullanımKılavuzu
Thank you for purchasing this OMRON Professional Blood Pressure Monitor.
Please completely read this Instruction Manual before using the monitor for the rst time.
Read this manual to ensure the safe and accurate use of the monitor.
IM-HBP-1100-E-01-08/2013
2294509-7A
13E1306
EN
FR
DE
IT
ES
NL
RU
TR
AR

TableofContents
Introduction
Intended Use.........................................................................................1
Exemptions............................................................................................2
Notes on Safety.....................................................................................3
UsingtheUnit
Components of the Product ..................................................................9
Options..................................................................................................9
Features of the Product.......................................................................10
Features / Functions of Unit................................................................ 11
Installing the Batteries ........................................................................13
Connecting the AC Adapter ................................................................13
Cuff Selection and Connection...........................................................14
Applying the Cuff to the Patient ..........................................................15
Zero Indicator Function.......................................................................17
Non-InvasiveBloodPressure(NIBP)Measurement
Measurement in “Normal Mode”.........................................................18
Measurement in “Auscultation Mode”.................................................20
Stopping the Measurement.................................................................21
Non-Invasive Pressure Measurement Principles ...............................22
Maintenance
Maintenance Inspection and Safety Management.............................23
Cleaning of the Device........................................................................23
Accessory Care...................................................................................23
Check before Use ...............................................................................24
Checking Pressure Accuracy..............................................................25
Troubleshooting...................................................................................26
List of Error Codes ..............................................................................29
Disposal...............................................................................................31
Specications
Technical Specications: HBP-1100...................................................32
Manufacturer’s Declaration.................................................................35

1
EN
Introduction
IntendedUse
MedicalPurpose
The device is a digital monitor intended for use in measuring blood
pressure and pulse rate in adult and pediatric patient population with arm
circumference ranging from 12 cm to 50 cm (from 5 inches to 20 inches).
IntendedUser
This device should be used by a medical professional.
PatientPopulation
This device is intended for use on adults and children of age 3-years and older.
Environment
The instrument is designed for use in physicians’ offices, hospitals,
clinics and other medical facilities.
MeasurementParameter
■ Non-Invasive Blood Pressure
■ Pulse rate
PrecautionsforUse
Warnings and cautions described in the instruction manual should be
observed at all times.

2
Exemptions
OMRONwillnotbearanyresponsibilitiesonthefollowingmatters.
1.
When a problem or damage occurs caused by the maintenance and/or repair
conducted by a person other than OMRON or the dealer specified by OMRON
2. The problem or damage of OMRON product caused by the product
of other manufacturer not delivered by OMRON
3. The problem and damage caused by the maintenance and/or repair
using the repair parts not specified by OMRON
4.
The problem and damage caused by the results not observing the Notes
on Safety or the operational method mentioned in this Instruction Manual
5. Under the circumstances not within the operating conditions of this
unit including the power source or the setting environment mentioned
in this Instruction Manual
6. The problem and damage caused by the result(s) of remodeling or
improper repair of this product
7. The problem and damage caused by act of god such as fire,
earthquake, flood, or lightning
1. The contents of this Instruction Manual may be changed without prior
notice.
2. We have thoroughly reviewed the contents of this Instruction Manual.
However, if an inadequate description or error is found, please let us
know.
3. It is prohibited to copy a part of or the entire Instruction Manual
without getting OMRON’s permission. Unless this Instruction Manual
is used by an individual (company), it cannot be used without getting
OMRON’s permission from the standpoint of the Copyright Law.

3
EN
NotesonSafety
The warning signs and symbol examples indicated below are intended to
ensure safe use of the product and prevent damage and injury to you and
others. The signs and symbols are explained below.
SafetySymbolsusedinthisInstructionManual
Warning Indicates the matters in which death or severe bodily
damage may arise as a result of incorrect handling.
Caution Indicates the matters in which bodily harm or material
damage may arise as a result of incorrect handling.
GeneralInformation
Note:
Indicates general information to keep in mind when using the unit and
other useful information.
Setup
Warning
• Do not use the cuff or AC adapter to lift the unit, it can also cause the
unit to malfunction.
• If the unit has broken down, contact your OMRON retail dealer or
distributor.
• Do not use in combination with a hyperbaric oxygen therapy device, or
in an environment where combustible gas may be generated.
• Do not use in combination with magnetic resonance imaging (MRI)
equipment. If MRI is to be performed, remove cuff connected to the unit
from the patient.
• Do not use with a defibrillator.
• Do not install the unit in the following locations:
- Locations subject to vibration such as ambulances and emergency
helicopters.
- A location where there is gas or flame.
- A location where there is water or steam.
- A location where chemicals are stored.
• Do not use at extremely high temperature, high humidity, or high
altitude. Use only within the required ambient conditions.
• Do not subject the unit to intense shock.
• Do not place heavy objects on the AC adapter cable, or allow the unit to
sit on the cord.
• Clinical testing has not been conducted on newborn infants and
pregnant women. Do not use on newborn infants and pregnant women.
• Do not plug in or unplug the AC adapter with wet hands.

4
Caution
• Do not install the unit in the following locations:
- Locations with dust, salt, or sulfur.
- Locations directly exposed to sunlight for extended periods of time (in
particular, do not leave in direct sunlight or near a source of ultraviolet
light for extended periods, as ultraviolet light will cause deterioration
of the LCD).
- Locations subject to vibration or shock.
- Near heaters.
• Do not use the unit near large equipment that uses a switching relay for
power ON/OFF.
Beforeuse/duringuse
Warning
• The unit complies with the EMC (Electro Magnetic Compatibility)
standard (IEC60601-1-2). As such, it can be used simultaneously with
multiple medical instruments. However, if instruments that generate
noise such as an electric scalpel or a microwave therapy device are
near the unit, check the operation of the unit during and after use of
these instruments.
• If an error occurs or a measurement result is questionable, check the
vital signs of the patient by auscultation or palpation. Avoid relying
solely on the measurement results of the unit when judging the patient’s
condition.
• Only trained healthcare providers should use this device. Do not allow
patients to operate this device.
• Properly connect the connectors and AC adapter cable.
• Do not place objects or liquids on top of this unit.
• Check the following before using the unit:
- Make sure the AC adapter cable is not damaged (wires are not
exposed or broken), and the connections are firm.
•
For the AC adapter connected to the unit, supplies, and optional devices,
use only the standard accessories or OMRON-specified products.
• Do not use in a location with moisture, or a location where water may
splash on the unit.
• This unit is intended for use in physicians’ office, hospitals, clinics and
other medical facilities.
• Do not use the unit if it emits smoke, an abnormal odor, or abnormal noise.
• Do not bring cellular telephones or transceivers into the room where the
unit is installed or being used.
• Do not connect multiple monitors to the same patient.
• Do not connect the unit to a power outlet that is controlled by a wall
switch.

5
EN
Caution
• Before using the unit, verify that none of the following apply to the
patient:
- Poor peripheral circulation, noticeably low blood pressure, or low body
temperature (there will be low blood flow to the measurement position)
- The patient uses an artificial heart and lung (there will be no pulse)
- An SpO2sensor and the cuff are attached to the same arm
- The patient has an aneurysm
- The patient has frequent arrhythmia
- Body motions such as convulsions, arterial pulsations, or trembling
(cardiac massage in progress, minute continuous vibrations,
rheumatism, etc.)
• Before use, visually inspect the unit to make sure there are no
deformations due to falling, and that there is no dirt or moisture on the
unit.
• When the unit has not been used for an extended period of time, always
verify that it operates normally and safely before use.
• Do not use in a location where the unit may easily fall. In the event that
the unit falls, verify that it operates normally and safely.
Cleaning
Warning
• When cleaning the unit, turn off the power and disconnect the AC
adapter from the unit.
• After cleaning the unit, make sure it is completely dry before connecting
to a power outlet.
• Do not spray, pour, or spill liquids into or onto the unit, accessories,
connectors, buttons, or openings in the housing.
Caution
• Do not use thinner, benzene, or other solvents to clean the unit.
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde
gas, high-concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the
manufacturer.
• Clean the unit regularly.

6
Maintenanceandinspection
Warning
• To use the unit safely and correctly, always inspect the unit when
starting work.
• Unauthorized modification is prohibited by law. Do not attempt to
disassemble or modify the unit.
Drycellbattery
Warning
• If battery fluid comes in contact with the eye, immediately flush
with copious amounts of water. Do not rub. Seek medical attention
immediately.
• Do not throw into flame, disassemble, or heat.
• Always disconnect the AC adapter from the unit before removing or
installing a battery.
• If the unit will not be used for a month or longer, remove the battery
from the unit and store.
• Do not attempt to disassemble or modify the battery.
• Do not apply pressure to and deform the battery. Do not throw, pound,
drop, bend, or hit the battery.
• The battery has positive/negative polarity. Do not insert batteries with
their polarities reversed.
• Do not connect the positive and negative terminals of the battery with a
wire or other metal object.
• Do not use the AC adapter and battery at the same time.
• Use only the specified type of battery.
Caution
• If battery fluid comes into contact with the skin or clothes, immediately
rinse with water.
• Do not use new and old batteries together, or use different types of
batteries together.
Non-InvasiveBloodPressure(NIBP)measurement
Warning
• If a cuff is used on a patient with an infection, treat the cuff as medical
waste, or disinfect before reuse.
• If frequently performing NIBP measurement using a cuff over an
extended period of time, periodically check the patient’s circulation.
In addition, wrap the cuff as indicated in the cautionary points in this
manual.
• Do not connect the NIBP cuff or cuff joint to a luer lock adapter.
• Do not bend cuff tube during inflation and deflation, particularly after a

7
EN
change of body position.
• Do not wrap the cuff on the following parts:
- An upper arm on which intravenous drip or a blood transfusion is
being performed.
- An upper arm on which SpO2sensor, IBP catheter, or other instrument
is attached.
- An upper arm with a shunt for hemodialysis.
• If measuring blood pressure with the cuff wrapped on the arm on
the side of the body where a mastectomy was performed, check the
patient’s condition.
Caution
• NIBP measurement should be performed on the upper arm.
• During NIBP measurement, stop excessive body movement by the
patient and minimize trembling.
• If a doctor has indicated that the patient has hemorrhagic diathesis or
hypercoagulability, check the condition of the arm after measurement.
• Use the appropriate cuff size to ensure correct measurements. If too
large a cuff is used, the measured blood pressure value tends to be
lower than the actual blood pressure. If too small a cuff is used, the
measured blood pressure value tends to be higher.
• Before and during measurement, verify that none of the following apply
to the patient:
- The part where the cuff is wrapped is at a different height than
the heart. (A difference of 10 cm (4 inches) in height may cause a
variation in the blood pressure value of up to 7 or 8 mmHg.)
- Body movement or conversing during measurement.
- Cuff wrapped over thick clothing.
- Pressure on the arm due to a rolled up sleeve.
• In the case of a cuff for adults, the cuff should be wrapped to a
tightness that allows two fingers to be inserted in between the cuff and
the arm.
• The accuracy of a flashing measurement value that is out of the
measurement range cannot be guaranteed. Always check the patient’s
condition before deciding what steps to take.
• Do not use the cuff if it is damaged or has holes.
• Only an OMRON GS CUFF can be used with this device. The use of
any other cuff may result in incorrect measurement.

8
Note:
Setup
• Read and understand the manual for each optional accessory.
This manual does not contain cautionary information for optional
accessories.
• Exercise caution with the cables and arrange so that the patient does
not become entangled or bound.
Beforeuse/duringuse
• Check the following after turning on the power:
- No smoke, abnormal odor, or abnormal noise is emitted.
- Press each button and verify that it operates.
- For functions that cause icons to light or flash, verify that the icons
light or flash (page 12).
- Measurement can be performed normally, and measurement error is
within the tolerance value.
• If the screen is not displayed normally, do not use the unit.
• When recycling or disposing of parts (including batteries) of the unit,
follow local government rules and regulations.
Cleaning
• For cleaning, see page 23.
Non-InvasiveBloodPressure(NIBP)measurement
• If the patient has acute inflammation, a pyogenic ailment, or an external
wound at the location where the cuff is to be wrapped, follow the
instructions of the doctor.
• Non-Invasive Blood Pressure measurement (NIBP) is performed by
compressing the upper arm.Some people may experience intense
pain, or transient spotting caused by subcutaneous hemorrhaging
(bruising) may appear. The spotting will disappear with time; however,
it may be appropriate to inform patients for whom this may be a
concern that spotting sometimes occurs, and if necessary, refrain from
measurement.
• To measure correctly, it is recommended that the patient relax and not
talk during measurement.
• To measure correctly, it is recommended that the patient rest quietly for
5 minutes before measurement.

9
EN
UsingtheUnit
ComponentsoftheProduct
Before using the unit, make sure that no accessories are missing and that
the unit and accessories are not damaged. If an accessory is missing or
there is damage, please contact your OMRON retail dealer or distributor.
Main unit
StandardMedicalAccessories
• ACadapter
• GSCUFFM(22-32cm)
Others
• InstructionManual(Thispaper)
• GuaranteeCard
Options
OptionalMedicalAccessories
(withinthescopeofECMedicalDeviceDirective93/42/EEC)
ACadapter
Adapter S
60240HW5SW
(9515336-9)
GSCUFFXL
HXA-GCUFF-XLLB
(9065802-0)
GSCUFFL
HXA-GCUFF-LLB
(9065798-9)
* UK plug type
AC Adapter UK 60240H7000SW (9983666-5)
GSCUFFM
HXA-GCUFF-MLB
(9065799-7)
GSCUFFS
HXA-GCUFF-SLB
(9065800-4)
GSCUFFSS
HXA-GCUFF-SSLB
(9065801-2)
Caution
• Only an OMRON GS CUFF can be used with this device. The use of
any other cuff may result in incorrect measurement.

10
FeaturesoftheProduct
The blood pressure accuracy of the HBP-1100 is clinically proven. Easy
to use, the HBP-1100 is intended for use by medical professionals.
■ Zero indicator function (page 17): Before each measurement, this
device indicates that “zero setting” was successful.
■ Auscultation Mode
■ 5 cuffs available - (XL: 42 to 50 cm, L: 32 to 42 cm, M: 22 to 32 cm,
S: 17 to 22 cm, SS: 12 to 18 cm)
■ This device and cuff can be cleaned with a soft cloth moistened with
alcohol.
■ Compact, can be stored in a drawer
■ Motion stop function: When body movement is detected, this device
stops deflation for 5 seconds.
■ Irregular pulse icon: Helps identify changes in heart rate, rhythm, and
pulse that may be caused by heart disease or other serious health
problems.

11
EN
Features/FunctionsofUnit
Frontandbottomofunit
1[Power ON/OFF]
button Turns the power ON/OFF.
2[Auscultation] button Press to enter “Auscultation Mode”
(page 20).
3[START/STOP] button
Press to start blood pressure
measurement.While the cuff is
inflating, hold down to continuously
inflate (page 18).
4NIBP connector Connects the cuff tube.
5 Power connector Connects the AC adapter.
6Battery cover Open to install or replace the
batteries.
Othersymbols
Symbol Description Symbol Description
This shows the Type BF
applied part. Caution
Class 2 (AC Adapter) Consult the instruction
manual.
34
2
16
5

12
LCDDisplay
3
2
1
7
5
4
6
8
1SYS Displays systolic blood pressure.
2DIA Displays diastolic blood pressure.
3PULSE Displays the pulse rate.
4Irregular pulse
wave icon
Lights up in the measurement result display
if the pulse wave interval was irregular
or there was body movement during
measurement.
5Auscultation icon Lights up when “Auscultation Mode” is ON.
6Pulse
synchronization
icon
Flashes in synchronization with the pulse
during measurement.
7Battery
replacement icon* When this icon lights up, an E40 error also
appears. Replace the batteries. (page 13)
8Zero indicator
icon
Lights up when “zero setting” is being
performed prior to blood pressure
measurement. When “zero setting”
finishes, appears.
* Only when the batteries are installed.

13
EN
InstallingtheBatteries
Warning
• If battery fluid comes in contact with the eye, immediately flush
with copious amounts of water. Do not rub. Seek medical attention
immediately.
• Do not throw into flame, disassemble, or heat.
• Do not attempt to disassemble or modify the battery.
• Do not use the AC adapter and battery at the same time.
Caution
• If battery fluid comes into contact with the skin or clothes, immediately
rinse with water.
1.MakesuretheACadapterhasbeen
disconnected.
2.Removethebatterycoverfromthe
bottomoftheunit.
3.Insertthebatteriesinthecorrect
orientation.
4.Replacethebatterycover.
Batteryreplacementicon
When this icon lights up, an E40 error also appears. Replace the
batteries.
ConnectingtheACAdapter
ACpower
Verify that the power outlet supplies the specified voltage and frequency
(100 - 240 V AC, 50/60 Hz).
ConnecttheACadaptertothepowerconnectorontheunitandthe
poweroutlet.
2
1

14
CuffSelectionandConnection
Warning
• If a cuff is used on a patient with an infection, treat the cuff as medical
waste, or disinfect before reuse.
Caution
• Do not use the cuff if it is damaged or has holes.
• Use the appropriate cuff size to ensure correct measurements. If a
cuff that is too large is used, the measured blood pressure value tends
to be lower than the actual blood pressure. If a cuff that is too small is
used, the measured blood pressure value tends to be higher.
Note:
• Make sure that the connectors are tightly connected.
Selectingthecuff
Measurethecircumferenceofthepatient’sarmandselectthecuff
sizethatisappropriateforthecircumference.
It is important to use the correct sized cuff for a patient in order to get
an accurate reading.
Select the cuff that is suitable for the patient from the cuffs below.
Cuffname Armcircumference
(cm) (inch)
GS CUFF XL* 42 - 50 17 - 20
GS CUFF L* 32 - 42 13 - 17
GS CUFF M 22 - 32 9 - 13
GS CUFF S* 17 - 22 7 - 9
GS CUFF SS* 12 - 18 5 - 7
*Available as an optional accessory.
Connectingthecuff
ConnectthecufftubetotheNIBPconnector
ontheunitandturnclockwisetolock.

15
EN
ApplyingtheCufftothePatient
The device can be used on either
the right or left arm.
Wrap the cuff on a bare arm or
over thin clothing. Thick clothing
or a rolled up sleeve will cause
inaccurate blood pressure
measurements.
1.Makesurethecufftubeisnotbent.
The cuff tube should be on the peripheral side.
2.WrapthecuffsothattheINDEXARTERY“ ”isdirectlyover
thebrachialartery.
The brachial artery is on the inner side of the patient’s upper arm.
1 to 2 cm
Make sure that the INDEX ARTERY “ ” is within the range. If outside
the range, there will be greater error in the blood
pressure value. In this case, use the appropriate cuff size.
* Attach the cuff so that the bottom edge is 1 to 2 cm from the inner
side of the elbow joint.
* The cuff should be wrapped to a tightness that roughly allows two
fingers to be inserted under the cuff.
3.Duringmeasurement,keepthebrachialarteryonwhichthecuff
iswrappedatthesameheightastherightatriumoftheheart.

16
Caution
• Make sure the cuff is wrapped in the correct arm position and is at the
same height as the heart.
NOTE:
• If measurement is difficult due to arrhythmia, use a different blood
pressure measurement method.
• If the patient has acute inflammation, a pyogenic ailment, or an
external wound at the location where the cuff is to be wrapped, follow
the instructions of the doctor.
• Non-Invasive Blood Pressure (NIBP) measurement is performed by
compressing the upper arm. Some people may experience intense
pain, or transient spotting caused by subcutaneous hemorrhaging may
appear. The spotting will disappear over time, however, if it is possible
that this will disturb the patient, try the following technique:
- Wrap a thin towel or cloth (one layer) under the cuff.If the towel or
cloth is too thick, there will be insufficient cuff compression and the
blood pressure value will measure high.
• If the patient moves or the cuff is touched, this may be falsely detected
as a pulse and over-inflation will occur.
• Do not inflate the cuff when it is not wrapped on the upper arm. This
may damage the cuff.

17
EN
ZeroIndicatorFunction
Before each measurement, this device indicates that “zero setting” was
successful.
■ When the power is turned on, the entire indicator blinks and then “zero
setting” starts. When completed, appears.
■ When the power is already on, measurement is started, then “zero
setting” takes place from the ready screen (which shows “0”). When
completed, appears.

18
Non-InvasiveBloodPressure(NIBP)Measurement
Measurementin“NormalMode”
1.Pressthe[START/STOP]button.
Blood pressure measurement is performed once.
2.Themeasurementresultsaredisplayed.
If a measurement value is outside the corresponding range below, the
value will flash.
SYS: 59 mmHg or less, or 251 mmHg or higher.
DIA: 39 mmHg or less, or 201 mmHg or higher.
PULSE: 39 bpm or less, or 201 bpm or higher.
■ Normal measurement ■ Measurement error / failure
Manualinationin“NormalMode”
If the cuff is not sufficiently inflated, it can be inflated manually.
During inflation, hold down the [START/STOP] button to inflate
continuously.
“-” appears below the value to indicate that manual inflation is in
progress.
Other manuals for HBP-1100
3
Table of contents
Languages:
Other Omron Blood Pressure Monitor manuals

Omron
Omron BP5350 User manual

Omron
Omron HEM-18 User manual

Omron
Omron SpotArm i-Q132 User manual

Omron
Omron EVOLV BP7000 User manual

Omron
Omron M3 Intellisense User manual

Omron
Omron RS3 User manual

Omron
Omron RS8 User manual

Omron
Omron RS2 User manual

Omron
Omron HEM-18 User manual

Omron
Omron HEM-711 User manual

Omron
Omron HEM-7154-E User manual

Omron
Omron HEM-711DLX User manual

Omron
Omron HEM-7156-A User manual

Omron
Omron RS6 User manual

Omron
Omron 7 Series User manual

Omron
Omron INTELLISENSE HEM-705CP User manual

Omron
Omron M2 Compact User manual

Omron
Omron M6W Manual

Omron
Omron BP760 User manual

Omron
Omron M2 Compact Instruction manual