Once BioShift Single Door UV-C Germicidal Chamber User manual

BioShift Single Door UV-C Germicidal
Chamber - Instruction Manual
This manual provides user guidelines and additional operation
instructions for the Small BioShift Single Door UV-C Germicidal Chamber.
(Model 24-0256)
For service or parts call your
sales representative, visit our website:
www.once.lighting/uv-c
Or contact us at:
Once Inc.
United States
15255 23rd Ave N
Plymouth, MN 55447
+1 (763) 381-5621
Germany
Am Südfeld 7
49377 Vechta
+49 (0) 4441 8898-10
info[email protected]om
Poland
Wloclawska 167
87-100 Tourn
+48 (56) 699-5330

TABLE OF CONTENTS
Introduction ................................................................................................................................................ 2
Specications .............................................................................................................................................. 3
Safety ......................................................................................................................................................... 4-5
Installation ................................................................................................................................................... 6
Touchscreen Display ............................................................................................................................ 7-9
Home Screen Components ................................................................................................ 7
Maintenance Screen Components .................................................................................. 8
Settings Screen Components ............................................................................................ 9
Programming ............................................................................................................................................10
Cycle Time .............................................................................................................................. 10
Sleep Timer ............................................................................................................................ 10
Maintenance ...................................................................................................................................... 11-13
Alert Notications / Errors ............................................................................................... 11
Check Lamp Life ................................................................................................................... 12
Order Replacement Lamps and Ballasts ...................................................................... 12
Changing Lamps and Ballasts ........................................................................................ 12
Reset Lamp Life .................................................................................................................... 12
Cleaning .................................................................................................................................. 13
UV-CDoseVerication..........................................................................................................13
Operating .................................................................................................................................................. 14
Contact Information .............................................................................................................................. 15

2
INTRODUCTION
Health and wellbeing is a critical component to sustain an enjoyable life. It’s why the
quality of the air we breathe and cleanliness of the items we touch every day matters.
The germicidal properties featured in the BioShift Single Door UV-C Germicidal Chamber
oer a proven and eective way to inactivate harmful micro-organisms. The BioShift
chamber is designed and approved for industrial and commercial uses, and is not to be
used in residential or home settings.
The BioShift chamber is a sanitation tool made easy to use alongside additional cleaning
eorts. It helps meet a wide range of end-user needs by limiting cross contamination
through the import of pathogens on everyday items like cell phones, eye glasses, and
tools.
Veried to inactivate common pathogens by a nationally recognized testing laboratory
specializing in antimicrobial, biocidal and virucidal eectiveness, the BioShift chamber
uses UV-C radiation that has been extensively researched for more than 40 years.
Note: All bacteria and viruses tested to date respond to UV-C radiation, however it is mostit is most
eective on hard surface objects.eective on hard surface objects. It may not inactivate bacteria, viruses and pathogens, in
areas of objects that are not exposed to the proper intensity and duration of UV-C radiation.
Thus, the BioShiftSingle Door UV-C Germicidal Chamber is most eectively used in addition
and as a complement to disinfection products and procedures, and NOT as a replacement.

3
SPECIFICATIONS
Dimensions
Outside 29.5” wide x 23” deep x 23.6” high (750 x 584 x 600 mm)
Weight 110 lbs (50 kg)
Functional
Operating Power / Current 110V: 110W / 1 A, 240V: 115W / 550 mA
Standby Power / Current 110V: 6W / 130 mA, 240V: 9W / 100mA
Input Voltage 110-240VAC, 50/60 Hz (voltage uctuation ±10%)
Transient over voltages allowed up to category II
Temporary over voltages allowed to occur on mains supply
Power Connection Connect to branch circuit rated 20A max
Fuse Protection 5x20mm, Slow Blow, 10A, 250V
Note: Disconnect main power supply before replacing fuse.
Débranchez l’alimentation principale avant de remplacer le
fusible.
Environmental
Operating 41°F (5 °C) to 104 °F (40 °C) temperature
Maximum relative humidity 80% for temperatures up to 88°F
(31°C), decreasing linearly to 50% relative humidity at 104° F (40°C)
Note: Under 41°F (5°C) temperature must run warm-up cycle before
disinfecting. See Operating section.
Storage -20 °F (-28 °C) to 140 °F (60 °C) temperature, 10-95% humidity
Note: Allow unit to reach operating temperature before use.
Pollution Degree 2
UV-C Light Source
Germicidal Lamps 20W UV-C Lamps (4)
Output 254 nm UV-C Light
Chamber Components
Controller See Touchscreen Display section.
Emergency Switch Button turns OFF chamber in case of emergency (re, smoke,
etc.).
Tray / Reective Liner The tray and reective liner rest inside the bottom of the
chamber and are removable for easy cleaning.
Window The tempered glass window is opaque to UV-C.
Germicidal Lamps There are four (4) 20W germicidal UV-C lamps.
Magnet Securely holds door in closed position while running.
Door Gasket Seals chamber during operation to keep light from escaping.

4
RISK OF UV
EXPOSURE CAUTION BIOHAZARD
SAFETY
Definition of Symbols
Warning
• It is possible that emissions from this product may interfere with some sensitive
appliances when placed nearby or in the same circuit as those appliances. The
user should be aware of this potential and take appropriate measures to avoid
interference.
• Not all materials are UV-C safe. UV-C rays can cause physical damage to products.
If products are excessively warm when they are removed after cleaning, they may
have been damaged by the UV-C.
• Do not put humans or live animals in the chamber.
• Do not allow children in or near the chamber.
• Do not place ammable liquid(s) in chamber.
Lamps
The lamps inside emit short wave UV-C radiation. Overexposure to direct or reected
UV-C light can cause severe damage to the eyes and skin. Never look into an illuminated
UV-C lamp without proper eye protection. The chamber windows provide a safe way to
view inside chamber without the risk of UV-C exposure.
Use caution when handling and changing lamps, as they are subject to fracturing. The
lamps contain a small amount of mercury, a chemical known to the state of California
to cause birth defects or other reproductive harm. The mercury is contained inside the
lamp and will not be released unless the lamp is damaged. Use appropriate PPE when
handling.
The lamps are sensitive to the oils in human skin, always handle with gloves.
Disposal and Recycling
All components must be disposed of in accordance with state and local laws. Refer to
lamprecycle.org for more information.
Visit www.once.lighting/bioshift-warnings or scan the QR code for
additional warning information.

5
Note: If the equipment is used in a manner not specied by the manufacturer, the protection
provided by the equipment may be impaired.
SAFETY
Features
Emergency Stop The safety switch can be used at any time to stop the operation.
Pressing EMERGENCY STOP disconnects all power.
Open Door BioShift chamber will not operate when the door is open. The door
cannot be opened during cycle time unless EMERGENCY STOP is
pressed.
Chamber The sealed chamber is meant to protect users from UV-C radiation.
Window The tempered glass window is opaque to UV-C.
Controller The controller has a countdown timer which allows the user to
monitor Lamp Life, replacement recommended after 1000 cycles,
3000 hours, or after a failed Dose Verication Test (see Maintenance
section).

INSTALLATION
The BioShift chamber comes with four (4) padded feet on each corner to allow for easy
placement on any sturdy and hard, non-moving surface. Note: The BioShift®chamber can
be installed in dry locations only (Pour emplacements secs seulement). Product is approved
for indoor use at an altitude up to 2000m.
The input power switch is customizable to the front or rear side of the chamber. If this is
desired, contact Once for further instructions.
1. Use two people to carefully lift the BioShift chamber into place.
2. Plug supplied power cord into standard electrical outlet. Note: Alternate power
cords not permitted.
For small object disinfection, position tray in the center slot for optimal cleaning.
For larger, heavier objects, position tray in the bottom slot.
Removal
Always de-energize/disconnect power before moving.
6

TOUCHSCREEN DISPLAY
The touchscreen display is located on the front of the unit.
Home Screen Components
(1) Cycle Time Display
(2) Chamber Start / Stop
(3) Chamber Status
(4) Settings
(5) Maintenance
1
3
4
2
5
7

TOUCHSCREEN DISPLAY
Located under the Maintenance button.
Maintenance Screen Components
(1) View Lamp Life Remaining
(2) Reset Lamp Life
(3) View Lamp Replacement Info
(4) Run 100mJ/cm2Dose Test
(5) Run 250mJ/cm2Dose Test
(6) Run 1000mJ/cm2Dose Test
1
4
2
3
8
5
6

9
1
4
2
3
TOUCHSCREEN DISPLAY
Located under the Home screen Settings button.
Settings Screen Components
(1) View Set Lamp Cycle
(2) Set Lamp Cycle
(3) Set Sleep Time
(4) Expired Lamp Life
(5) Contact ONCE®
5

10
PROGRAMMING
Cycle Time
The standard and recommended default disinfection cycle is 5 minutes. However,
the cycle time can be customized to better meet facility protocols. Note: The amount
of inactivation by UV-C radiation is proportional to the amount of exposure time.
TO SET CYCLE TIME:
• From the Home screen, choose the Settings button (see Settings Screen
Components).
• Choose the Set Cycle button. Note: The time is set by arrowing up or down,
programmed in MM:SS.
• Once desired lamp life cycle has been chosen, select the check mark to set.
PROGRAMMING ITEMS TO NOTE:
• Controller cannot be programmed if EMERGENCY STOP button is pressed in.
Pressing EMERGENCY STOP button disconnects all power, even to the controller.
• If disconnected from power, the BioShift chamber does not have to be reprogrammed.
Sleep Timer
In order to ensure the longest life, the BioShift chamber will automatically enter sleep
mode after 30 minutes of inactivity. This setting can be changed by following this path:
• From the Home screen, choose the Settings button (see Settings Screen
Components).
• Choose the Sleep Time option and set to desired time.
• Select the check mark to set.

11
MAINTENANCE
Alert Notifications / Errors
An alert symbol will appear in the upper left corner of the Maintenance button when
one of the following occurs:
• UV-C lamps have reached maximum cycle / hours.
• Replacement lamps have been installed and Lamp Life needs
to be reset.
Simply follow the red exclamation point to the source of the issue.
Note: Lamp life remaining will never go negative. Once it reaches zero, it will remain until
Lamp Life is reset.

12
MAINTENANCE
Note: Chamber must be de-energized and included safety glasses or other UV protective
eyewear should be used at all times when performing maintenance. The chamber must be
energized when performing UV-C Dose Verication.
Check Lamp Life
It is important to monitor Lamp Life and replace when necessary. Even though the
lamps have a long default life, they will lose their power over the course of time and
need to be changed. The remaining lamp life is always prominently displayed at the top
of the Maintenance menu (see Maintenance Screen Components).
Order Replacement Lamps and Ballasts
The information required for ordering new lamps include the lamp part number and
ONCE®contact information. This information can be found from the Maintenance screen
under Lamp Replacement Info.
20W UV-C Replacement Lamps 26-0053 / 929002100324
110-240V Replacement Ballast 26-0052 / 929002100325
Changing Lamps and Ballasts
To replace lamps, de-energize the chamber, then simply twist the lamp and pull out
from its housing. Slide the new lamp in and twist to lock it in place. Do not use excessive
force. It is recommended that all lamps are replaced on 1000 cycle or 3000 hour intervals,
whichever comes rst, or after a failed UV-C Dose Verication. After lamps have been
replaced, press the “Reset Lamp”button in the Maintenance menu to reset the lamp life
counter.
Replace ballasts if replacement lamps fail to provide proper intensity, or if lamps fail to
ignite. Ballasts must be installed according to country, state, and local NEC to the same
standard as shipped from the manufacturer. After de-energizing the chamber, ballasts
are accessed via the removable side panel located inside the chamber and must be
securely xed to the bottom of the chamber.
Reset Lamp Life
• From the Home screen, choose the Maintenance button (see Maintenance Screen
Components).
• Choose the RESET LAMP button.
• Lamp Life will be reset to default, unless custom values have been entered in the
Lamp Life Settings menu.

MAINTENANCE
Cleaning
• Do not use industrial degreasers or strong solvents such as brake cleaners.
• Do not override safety features during maintenance / cleaning.
• When cleaning all reective aluminum surfaces, do not use abrasives, they will
reduce the reectivity.
• Outside: BioShift chamber can be cleaned with a damp towel. Strong solvents
or detergents are not recommended. A typical household cleaning solvent is
recommended.
• Inside: Clean inside chamber walls with ethanol (reagent grade) and a soft towel.
Be careful not to spray or apply chemicals directly on the UV-C lamps. This may
result in total lamp destruction.
• Inspect the aluminum panels, if they appear to be degrading and reducing
reectivity, consider replacing.
UV-C Dose Verification
It is important to regularly verify the UV-C dose output of the BioShift chamber to
ensure proper disinfection. Every 50 cycles or once per month, whichever comes rst,
the operator shall verify that the UV-C dose exceeds a minimum of 100 mJ/cm2in a
short, programmed test cycle. Before proceeding with the Dose Verication, ensure the
chamberhascompletedafull cycle withinthe last 5minutes sothatthe lampsarewarmed
up. With an empty chamber, place a card (below) face-up at the center of the tray in the
bottom position. On the touchscreen display, select the “100 mJ/cm2Dose Test” button
to start a test cycle. Upon completion of the test cycle, the operator will evaluate the color
development of the center circle. If the center color matches the 100 mJ/cm2region, the
chamber has passed the test. If the color does not match 100 mJ/cm, all the lamps in
the chamber must be replaced. Note: Do not use dose verication as a disinfection cycle.
Replacement 100 mJ/cm2test cards (929002100327) are available from ONCE®
at www.once.lighting/uv-cdosimeter. Do not store cards in direct sunlight.
13

OPERATING
1. Ensure power switch is turned on and Emergency Stop button is released.
2. Place objects in chamber following guidelines below.
• Do not stack items / supplies in the chamber.
• Leave 6” between each product in the chamber with no more than 3 across.
• Place the tallest products in the center.
• It is not recommended to put food through chamber unless it is sealed.
3. Ensure door is latched closed and press Start. When a cycle is started the timer
(located on the front controller) will count down. The chamber will turn OFF
automatically after programmed time (recommended 5 minutes).
4. If the BioShift chamber is kept at a temperature under 41°F (5 °C), a warm-up cycle
must be completed before beginning sanitation.
ADDITIONAL OPERATING FUNCTIONS:
• BioShift chamber door shall not be opened during radiation cycle.
• In case of emergency, EMERGENCY STOP button can be pressed.
• If cycle has been interrupted, the cycle will restart at the beginning of the cycle
when pressing Start again.
MONITORING:
• Monitoring of inserted material and lamp functions can be done
through the window, made with glass that blocks UV-C radiation.
• Countdown timer shows how much time in radiation cycle is left.
CONTROLLER:
A. If the door is open, pressing the Start button will not start the machine.
B. Upon disinfection cycle completion, CLEAN will appear on the screen with the
option to Start another cycle.
14

15
United States
15255 23rd Ave N
Plymouth, MN 55447
+1 (763) 381-5621
Germany
Am Südfeld 7
49377 Vechta
+49 (0) 4441 8898-10
info[email protected]om
Poland
Wloclawska 167
87-100 Tourn
+48 (56) 699-5330
Conforms to cETLus + LVD + CE (ETL CDR, IEC/
EN/ANSI/UL/CUL 61010-1 & IEC/EN/ANSI/UL/
CUL 61010-2 040 & IEC/EN 61326-1: 2013)
5018010
THANK YOU
Thank you for purchasing the
BioShift Pass-Through UV-C
Germicidal Chamber
Manual valid for all 24-0256 models
Model number can be found on product nameplate
For more information and references
visit us online at www.once.lighting/uv-c
For warranty information visit www.once.lighting/warranties
For service or parts call your
sales representative, visit our website:
www.once.lighting/uv-c
Or contact us at:
Once Inc.
This manual suits for next models
1