1.4 STANDARD FEATURES AND SPECIFICATIONS
• Wall-mounted NEMA 4X polycarbonate enclosure with backlit alphanumeric user interface/display
• One internally powered electrically isolated 4-20 mA analog output for ow rate or ow velocity.
• Two scalable electrically isolated dry contact pulse output for totalizing ow, indicating ow
direction or ow alarm.
GENERAL SPECIFICATIONS
ACCURACY
± 1.0% of reading from 1 to 20 ft/sec
± 0.01 ft/s for velocities below 1 ft/sec
OVERALL FLOW RANGE
0.1 to 20 ft/sec
SENSING METHOD
Clamp-on ultrasonic, differential transit time
method indirect or reect mode
PIPE SIZE RANGE
2” through 24” nominal diameter
PROGRAMMING
Factory programmed for specic application
MEMORY
Non-volatile memory retains all program
parameters and totalized values in the event of
power loss.
DISPLAY
White, backlit alphanumeric display shows:
5-digit ow rate with oating decimal, 14-digit
totalizer, pulse output status, operating status,
and provides eld conguration.
Displays total ow, ow rate, ow velocity,
signal strength, ow direction, alarm
conditions, relay status & logger status
Rate display range: 0 - 9,999,999
Totalizer display range: 0 – 99,999,999
Totalizers will roll over to zero when maximum
count is exceeded.
OUTPUT SIGNALS PROVIDED
Analog output: Isolated 4-20 mA/0-5 VDC
(Internally powered, 1000Ω max impedance
eld, selectable)
Two Programmable pulse outputs:
Optically isolated dry contacts
Contact rating: 30 VDC, 10 mA maximum
Pulse duration: 50 ms
Programmable for scaled pulse, ow direction, or
ow alarm
Serial communications: BACnet®MS/TP or
FLUID TEMPERATURE RANGE
Standard: -40° F to 250° F
AMBIENT OPERATING TEMPERATURE RANGE
-5° F to 140° F
MECHANICAL
MATERIALS OF CONSTRUCTION
Electronics enclosure: Wall-mount, polycarbonate
with clear, shatter proof enclosure
Transducer mounting hardware: Anodized aluminum and
stainless steel
Transducers: Stainless Steel and plastic
ENCLOSURE RATINGS
NEMA 4X, weather tight
ELECTRICAL
This equipment is intended for INSTALLATION
CATEGORY (OVERVOLTAGE CATEGORY) II applications.
Installations must comply with all local, state and federal
building codes.
INPUT POWER – Factory selectable
10 - 30 VDC/VAC, 10 VA maximum
(Fuse not eld replaceable)
100 - 240 VAC 50/60 Hz, 10 VA maximum
(Internal: 5x20mm, 250V/2.0A fuse)
ELECTRICAL CONNECTIONS
Enclosed terminal connections, cable access through
three standard ½” conduit openings & one standard
¾” conduit opening
Transducer signals: (2) BNC-connectors
Input power: 2 terminal removable connector
(12 – 28 AWG)
Signal inputs/outputs: Removable connectors, 2 or 3
terminal (12 – 28 AWG)
WIRING
Transducer signals: Only use ONICON provided triaxial cable.
24 VDC/VAC input power: Use PVC jacketed copper cable
with a wire gauge suitable for the length of run and
required maximum current carrying capacity.
Provide a separate protective earth wire.
120/230 VAC input power: Use a 3-wire service
with one wire a protective earth ground.
Signal inputs/outputs: Use PVC jacketed copper
shielded cable with a wire gauge suitable for the
length of run and required maximum current
carrying capacity.