Opti-Med JM-LED02 User manual

LED Shadowless Lamp
JM-LED02
USER MANUAL
Opti-Med ®
Hangzhou Jupiter Medical Equipment Co., Ltd
1116, Gudun Road, West Lake District, Hangzhou, China
Tel: +86-15021634329 Web: www.opti-med.net
Email: [email protected]

1
www.opti-med.net
1. Expected Use
This product is used in various surgical lighting in hospital operating room. It can
also be use in medical lighting situations in emergency room.
2. Conditions of Use
2.1 Normal working conditions
Temperature: 10℃~30℃;
Relative humidity: ≤70%;
Atmospheric pressure range: 860hPa~1060hPa;
This equipment can not be worked in flammable and explosive gas environment.
The type of device security: I type equipment;
Working system: continuous operation;
Liquid injection protection level: IPXO.
2.2 Power parameters
Power supply voltage: AC220(±10%)V;
Frequency: 50Hz;
Input power: 150VA;
700 main lamp power: 80W;
500 main lamp power: 70W.
2.3 suitable mounting height: 3200~3300 mm
Low-storey may bring inconvenience to users. Please refer to the product diagram.
in special circumstances, you should contact with the manufacturer when purchasing.
3.Technical Specifications
Type
700 Main lamp
500 Main lamp
Diameter of Light Head (mm)
700
500
Quantity of LEDS
80
48
Illumination (10 Steps Adjustable)
40~160klx
40~160klx
Color rendering index
Ra>95
Ra>95
Color temperature (5 Steps Adjustable)
3500-5000K
3500-5000K
Depth of illumination (mm)
1000
1000
Spot diameter
120-250mm
120-250mm
Endoscopy Mode
Available
Available
Endoscopy Illumination
40,000Lux
40,000Lux
LED Life
>50,000 Hours
>50,000 Hours
The model of LED
OSRAM
GWCS8PM1.CM
OSRAM
GWCS8PM1.CM
UPS
Available
Available
Control Panel
LCD
LCD
Opti-Med ®

2
www.opti-med.net
4. Simplified Diagram
NANTONG MEDICAL APPARATUS CO.,LTD
NANTONG MEDICAL APPARATUS CO.,LTD
3300
360°
360°
360°
φ750
φ550
360°
360°
300°
300°
300°
300°
±45°
±45°
915 950 800 940
balance arm
2000
balance arm 3000
decorative layer
m ain lam p
assistant
lam p
handle
handle
control box
control box
rotating arm
floor
AC220V
RT1-20 2A
RT1-20 2A
power switch of
operating room
transform er of
the m ain lam p
transform er of
the assistance
lam p
constant-
current
source
constant-
current
source
drive control
drive control
m ain
lam p
assistant
lam p
user interface
user interface
Opti-Med ®

3
www.opti-med.net
5.Installation
The equipment is fixedly mounted on the roof. The floor must bear the weight of
the device,but also bear the force moment because of partial load and the operation. In
order to ensure the safety, the floor should have enough bearing capacity and strength.
Users can make construction according to the following pictures.
Floor installation program
φ360
L
pressure plate
floor plate
on the roof
roof of
decorative layer
m ounting
bracket
m ounting base
picture 1: the roof w ith decorative layer
pressure plate
floor plate
on the roof
m ounting disk
m ounting base
picture 2: the roof w ithout decorative layer
m o un ting
d isk
m o untin g b ase
screw M 1 2
nut M 1 2
sp ring w asher 12
fla t w a sher 12
p ressure p la te
p icture 3 : enla rg ed view o f screw a ssem b ly
Opti-Med ®

4
www.opti-med.net
φ360
φ240
φ120
8-φ14
material: GB700: Q235
ASTM : A283GRC A306GR55
JIS : G3113-87 SAPH38
φ300
φ360
φ120
φ300
thickness: 8m m
sup erficial treatm ent:antirust lacquer
thickness: 12m m
sup erficial treatm ent:antirust lacquer
picture 4: p ressure p late picture 5: m o unting d isk
All fasteners of the installed components must be reliable and fastening to prevent
the harm due to loose. When mounting, the bottom of mounting base must be adjusted
to level condition, otherwise it would cause the drift of lamp.
Ground wire (color: yellow or green) must be grounded reliably, in order to ensure
the use of safety.
With installed problems, please contact with the manufacturer before construction.
Opti-Med ®

5
www.opti-med.net
Mounted diagram
m ounting disk
ro tating arm
balance arm 2000
assistant lam p m ain lam p
balance arm 3000
Opti-Med ®

6
www.opti-med.net
Mounted steps
1. according to picture 1 and picture 2 in the operation instruction, drill four Ф16mm
holes on the floor plate with impact drill in determined place. pressure plate and
mounting disk must be tightened with screw M12 in the appropriate length, according
to picture 3.
2. According to mounted diagram, connect the rotating arm to the mounting plate,
regulate it to horizontal place, and tighten it according to picture 3.
3. Connect two balance arms to rotating arms respectively, and connect the
conductive plug. then tighten the two parts with the distributive screw.
4. According to the instruction for the balance arm of Ondal, link the conductive plug,
connect two lamps to the balance arms respectively, and make sure that it is reliable.
5. After careful examination and confirmation, debug the equipment with energizing
according to the operation instruction.
6. Use
NANTONG MEDICAL APPARATUS CO.,LTD
handhold
push button
handle
picture 6:lamp head
Opti-Med ®

7
www.opti-med.net
Position adjustment
Pull the handhold on the lamp in order to the position and the point which meet
operators’ requirement.
Illumination adjustment
Press the power switch, the lamp lights. each time you press the button
“ increasing brightness”, the illumination will rise a gear. When up to the maximum
10 gear, the maximum illumination symbols will lights; each time you press the
button “weakening brightness”, the illumination will decline a gear, up to the lowest
illumination. The electronic control system of the equipment has memory function. So
it can adjust to the last used gear illumination when rebooting.
Spot adjustment
It is not required usually adjustment,because that the product has been set on
appropriate position when it is out of the factory. When using,place the lamp
Head at 1m away from the surgical site. If it needs to adjust slightly,rotate the
handle in order to the site which meet the surgical needs.
Handle’s remove
When disinfecting, hold the lower part of the handle with hand, and press the
button with your thumb, pull down. then you can remove it. When installation, keep
the holes on the direction of the button, and push it gently until the button pops out.
Handle’s disinfection
Do not put the handle in high pressure steam sterilizer, in order to avoid
deforming. It can use peroxyacetic acid solution, glutaruc dialdehyde solution and
medical alcohol to disinfect in normal temperature.
7. Maintain
picture 8
Opti-Med ®

8
www.opti-med.net
Damping adjustable rotating arm
The center seat of 2 rotating arms has 2 damping adjustment screws. If it drifts after
using for some time, you can tighten the damp with wrench according to picture 8.
picture 9
Lamp head lamp adjustment
Lamp head and two brackets should have appropriate damping, to avoid
transposition. If appearing the phenomenon, tighten the screws properly according to
picture 9.
Balance arm adjustment
Please operate it according to the instruction of Ondal.
Replace the fuse
If the lamp can not light ,in the case of normal electricity network, unscrew the
fuse holder , check whether the fuse is damaged and identify the reasons, according to
the following picture. After confirming, replace the fuse.
Opti-Med ®

9
www.opti-med.net
Precautions when maintaining
(A) the product is not anti-inlet device. It can not be cleaned with water when
cleaning. You can wipe it with wet cloth containing neutral detergent solution, then
wipe it with wet cloth containing clear water for several times. It also can be wiped
with wet cloth containing disinfectant.
(B) After surgery, cut off power supply of shadowless operating lamps, rise the lamp
head to the highest place to be convenient to clean the operating room and to avoid
that the cleaning equipments collide the shadowless operating lamps.
(C) Non-professionals do not dismount the device at option. In addition to the LED
module inside the device, there is no device what users can replace.
8.Warranty
The warranty period of our shadowless lamp is one year. We carry out the
national 3 guarantees policy. Please save the relevant receipt of purchasing and
maintenance card.
Do not let person without the authorization by the manufacturer dismantle the
equipment. If there is any needs on the fault or use of the equipment, please contact
with customer service of the manufacturer.
9.Tips for Environmental Protection
The equipment does not contain mercury vapor and explosive devices. Deal with
the packaging materials and the equipment after stopping working according to local
environmental regulations. Metallic article may become renewable resources, and
e-waste should be treated with harmless.
10. Packing List
(1) Rotating arm 1
(2) Balance arm Ondal for 500 head 1
(3) Balance arm Ondal for 700 head 1
(4) Main lamp (700) 1
(5) Assistant lamp (500) 1
(6) Handle 4
(7) Product certificate 1
(8) Manual 1
(9) Warranty sheet 1
(10) Instruction for balance arm Ondal for 500 head 1
(11) Instruction for balance arm Ondal for 700 head 1
(12) Inner hexagon spanner (L,M,S) 1
(13) Fuse(F2AL250V) 4
Opti-Med ®
Table of contents
Other Opti-Med Work Light manuals
Popular Work Light manuals by other brands

Parkside
Parkside PAAL 6000 B2 Operation and safety notes

ARRI
ARRI M-Series ARRIMAX 18/12 Service manual

GYS
GYS toPARC Pro operating instructions

Ozito
Ozito Power Xchange PXWLS-800U Original instructions

Envirolite
Envirolite EVT1042D3-05 manual

Deluce Lighting
Deluce Lighting LEDSTRIPKIT/WH Fitting instructions