- TYPE TABLE
The table must be stable: must not sag during lean of the instrument (eg. Laboratory
bench or marble / granite counter top).
Must be as anti-magnetic and anti-static as possible.
Must be reserved to the balance.
Place the balance as close as possible to the table legs because there are less
vibrations rathern than in the middle of the table.
2 Recommendations for a correct use of the instrument
2.1 Preface
To obtain accurate and repeatable weighing pay attention to the following.
2.2 First power on
THE FIRST TIME THAT YOU CONNECT THE BALANCE TO THE MAINS, BEFORE
USE, WAIT AT LEAST 12 HOURS TO ENSURE THE WARM-UP OF THE
INSTRUMENT.
For a correct use of the instrument never disconnect the balance from the mains. If you
want to turn it off, use the ON / OFF button to put it in Stand-by (in this way it is not
necessary to wait each time the warm up time).
2.3 Using the Balance
Level the balance. It is fundamental always to level properly the instrument: regularly
check that the air bubble is at the center of the level. Help with block-feet to always
ensure the correct leveling of the balance.
Load the sample to weigh always in the center of the plate to avoid possible errors.
Open the glass doors of the cabinet as less as possible and use the tweezers to
load/unload the samples to be weighed.
Pay attention to possible Static charges that may be generated due to containers with
materials of low electrical conductivity or due to dry air with less than 40% moisture.
Electrostatic charges can alter the results of the weighings. It is recommended to use the
Ionizer Mod. Ion-A15 to eliminate static charges present on any substances to be
weighed or that have accumulated on the instrument.
- Because of static charges the results of weighings are always different, the balance is
not repeatable.
Pay attention to dynamic push: a big temperature difference between the sample to be
weighed and the weighing chamber creates air drafts along the samplel. A colder object