
Page 6
JCF/F Operating Instructions
Fitting the Cell
THE UPPER SCREWS ON THE SIDE MOUNTING BLOCKS ARE COVERED AND MUST
UNDER NO CIRCUMSTANCES BE REMOVED. TO DO SO WILL RESULT IN BREAKAGE
OF THE CELL. TO FIT THE CELL:
1. Carefully unpack the cell onto a scratch proof surface.
2. The cell mounting blocks are fitted to the cell by two screws protected by covers.
UNDER NO CIRCUMSTANCES REMOVE THESE SCREWS.
3. Below the covered screws are holes which allow access to the cell mounting screws.
4. Offer the JCF cell to the cabinet front panel, so that the cell fixing screws which
protrude through the cell mounting blocks, locate into the cabinet front panel holes.
5. Place the Allen key supplied through the left hand 6mm hole and locate into the
captive nut giving the screw 3 complete turns.
6. Next place the Allen key through the right hand 6mm hole and locate into the captive
nut giving that screw 3 complete turns.
7. Continue to tighten the two screws alternatively until the cell is secure.
8. Do not tighten fully as the cell postion needs to be adjusted.
Fitting the Pump Tubing
1. Check that the Drain Tube on the right hand side of the Cell is retained vertically by
its clip.
2. Fit the tube from the pump outlet to the left hand side of the Cell and the tube from the
filter inlet to the right hand side of the Cell.
3. Fill the Cell with fresh, 0.9% saline to within 5mm of the top. There is a small recess
moulded into the top left rear wall of the Cell. Filling to the lower edge of this recess
will give the correct amount of saline.
4. Insert the power supply lead, plug into the electricity supply and switch on.
5. Switch on the pump using the Blue switch and top up the saline in the Cell as the
saline it fills the tubing and filter.
Cell Adjustment
1. Select the base curve measuring function and adjust the cell position by hand until the
measuring cylinder is equally positioned about the left hand edge of the illuminate
centre thickness scale, and is parallel to the screen fixing strip.
2. Tighten the cell mounting screws. DO NOT OVERTIGHTEN.
总代理:上海朗善光学仪器有限公司 www.lenser17.com