ORTEC 467 Service manual

Model
467
Time
to
Pulse
Height
Converter
and
SCA
Operating
and
Service
Manual
This
manual
applies
to
instruments
marked
"Rev
12"
(on
rear
panel)
7V.
Printed
in
U.S.A.
2730
(N152Q)
02C
0976

CONTENTS
Page
WARRANTY
v
PHOTOGRAPHS
vi
1.
DESCRIPTION
1
1.1.
Purpose
and
Features
1
1.2.
Operation
1
1.3.
Logic
1
2.
SPECIFICATIONS
2
2.1.
Performance
2
2.2.
Controls
3
2.3.
Inputs
3
2.4.
Outputs
3
2.5.
Electrical
and
Mechanical
4
3.
INSTALLATION
4
3.1.
General
4
3.2.
Connection
to
Power
4
3.3.
Connection
into
a
System
4
3.4.
Linear
Output
Signal
Connections
and
Terminating
Impedance
5
3.5.
Logic
Signal
Connections
5
4.
OPERATING
INSTRUCTIONS
5
4.1.
Time
to
Pulse
Height
Conversion
5
4.2.
Single
Channel
Analysis
6
5.
CIRCUIT
DESCRIPTION
7
5.1.
General
7
5.2.
TPHC
Circuit
7
5.3.
Single
Channel
Analyzer
Circuit
8
5.4.
Auxiliary
Logic
8
6.
MAINTENANCE
9
6.1.
Testing
Performance
9
6.2.
Corrective
Maintenance
12
6.3.
Troubleshooting
14
6.4.
Typical
DC
Voltages
15
6.5.
Modifications
15
6.6.
Factory
Repair
15
Schematics
and
Block
Diagram
467-0201-S1
467-0301-S1
467-0101-B1

ILLUSTRATIONS
Fig.
6.1.
Test
System
for
Checking
Conversion
9
Fig.
6.2.
Test
System
for
Checking
Converter
Resolution
10
Fig.
6.3.
Test
System
for
Checking
Count
Rate
11
Fig.
6.4.
Test
System
for
Checking
Differential
Linearity
11
Fig.
6.5.
Differential
Linearity
for
the
Indicated
Ranges
11
Fig.
6.6.
Test
System
for
Checking
External
Strobing
Mode
12
Fig.
6.7.
Using
Internal
Strobe
with
Stop
After
End
of
Conversion
13
Fig.
6.8.
Using
Internal
Strobe
with
Stop
Timed
Normally
13
Fig.
6.9.
Using
External
Strobe
with
Output
Timed
Normally
14
Fig.
6.10.
Using
External
Strobe
with
Strobe
Signal
Arriving
Before
Conversion
is
Complete
14

STANDARD
WARRANTY
FOR
ORTEC
INSTRUMENTS
ORTEC
warrants
that
the
items
will
be
delivered
free
from
defects
in
material
or
workmanship.
ORTEC
makes
no
other
warranties,
express
or
implied,
and
specifically
NO
WARRANTY
OF
MERCHANTABILITY
OR
FITNESS
FOR
A
PARTICULAR
PURPOSE.
ORTEC's
exclusive
liability
is
limited
to
repairing
or
replacing
at
ORTEC's
option,
items
found
by
ORTEC
to
be
defective
in
workmanship
or
materials
within
one
year
from
the
date
of
delivery.
ORTEC's
liability
on
any
claim
of
any
kind,
including
negligence,
loss
or
damages
arising
out
of,
connected
with,
or
from
the
performance
or
breach
thereof,
or
from
the
manufacture,
sale,
delivery,
resale,
repair,
or
use
of
any
item
or
services
covered
by
this
agreement
or
purchase
order,
shall
in
no
case
exceed
the
price
allocable
to
the
item
or
service
furnished
or
any
part
thereof
that
gives
rise
to
the
claim.
In
the
event
ORTEC
fails
to
manufacture
or
deliver
items
called
for
in
this
agreement
or
purchase
order,
ORTEC's
exclusive
liability
and
buyer's
exclusive
remedy
shall
be
release
of
the
buyer
from
the
obligation
to
pay
the
purchase
price.
In
no
event
shall
ORTEC
be
liable
for
special
or
consequential
damages.
QUALITY
CONTROL
Before
being
approved
for
shipment,
each
ORTEC
instrument
must
pass
a
stringent
set
of
quality
control
tests
designed
to
expose
any
flaws
in
materials
or
workmanship.
Permanent
records
of
these
tests
are
maintained
for
use
in
warranty
repair
and
as
a
source
of
statistical
information
for
design
improvements.
REPAIR
SERVICE
If
it
becomes
necessary
to
return
this
instrument
for
repair,
it
is
essential
that
Customer
Services
be
contacted
in
advance
of
its
return
so
that
a
Return
Authorization
Number
can
be
assigned
to
the
unit.
Also,
ORTEC
must
be
informed,
either
in
writing
or
by
telephone
[(615)
482-4411]
,
of
the
nature
of
the
fault
of
the
instrument
being
returned
and
of
the
model,
serial,
and
revision
("Rev"
on
rear
panel)
numbers.
Failure
to
do
so
may
cause
unnecessary
delays
in
getting
the
unit
repaired.
The
ORTEC
standard
procedure
requires
that
instruments
returned
for
repair
pass
the
same
quality
control
tests
that
are
used
for
new-production
instruments.
Instruments
that
are
returned
should
be
packed
so
that
they
will
withstand
normal
transit
handling
and
must
be
shipped
PREPAID
via
Air
Parcel
Post
or
United
Parcel
Service
to
the
nearest
ORTEC
repair
center.
The
address
label
and
the
package
should
include
the
Return
Authorization
Number
assigned.
Instruments
being
returned
that
are
damaged
in
transit
due
to
inadequate
packing
wil
l
be
repaired
at
the
sender's
expense,
and
it
wi
ll
be
the
sender's
responsibility
to
make
claim
with
the
shipper.
Instruments
not
in
warranty
will
be
repaired
at
the
standard
charge
unless
they
have
been
grossly
misused
or
mishandled,
in
which
case
the
user
will
be
notified
prior
to
the
repair
being
done.
A
quotation
will
be
sent
with
the
notification.
DAMAGE
IN
TRANSIT
Shipments
should
be
examined
immediately
upon
receipt
for
evidence
of
external
or
concealed
damage.
The
carrier
making
delivery
should
be
notified
immediately
of
any
such
damage,
since
the
carrier
is
normally
liable
for
damage
in
shipment.
Packing
materials,
waybills,
and
other
such
documentation
should
be
preserved
in
order
to
establish
claims.
After
such
notification
to
the
carrier,
please
notify
ORTEC
of
the
circumstances
so
that
assistance
can
be
provided
in
making
damage
claims
and
in
providing
replacement
equipment
if
necessary.

SCA
INHIBIT
IN
OUTPUT
DELAY
INHIBIT/RESET
CO!
n<'o^
;
;
GOING
LOGIC
INPUTS
START
INPUTS
DC
ADJ
TRUE
START
OUTPUTS
TRUE
STOP
TPHC
Blicy
o
STROBE
SYNC
RESET
INT
START
EXT
EXT
-I
INT
STOP
SCA
MODE
SCA
WINDOW
OUTPUT
NORMAL
TPHC
OUTPUT
0
STOP
INHIBIT
MODE
DELAY
MONITOR
O.M.O
(iS
G)
our
••••<
mvjil
m.
y-r.vriv^-,
if
"W-
r
f
%
*■
■
•
V.
;
.
-.
-
w-
■■■■'■

ORTEC
467
TIME
TO
PULSE
HEIGHT
CONVERTER
AND
SCA
MANUAL
1.
DESCRIPTION
1.1.
PURPOSE
AND
FEATURES
The
ORTEC
467
Time
to
Pulse
Height
Converter
and
Single
Channel
Analyzer
(TPHC/SCA)
measures
the
time
Interval
between
the
leading
edge
of
logic
pulses
furnished
to
Its
start
and
stop
Inputs
and
generates
an
analog
output
pulse
that
Is
proportional
to
the
measured
time
through
the
TPHC
output.
The
TPHC
output
pulses
are
appropriate
for
multichannel
analysis
to
obtain
timing
spectra.
They
are
also
connected
Internally
to
the
single
channel
analyzer
to
generate
an
SCA
output
logic
pulse
for
each
TPHC
pulse
with
a
peak
amplitude
within
the
adjusted
single
channel
limits.
There
are
15
full-scale
time
ranges
that
can
be
switch-
selected
with
the
467,
from
50
ns
through
80
us.
Each
TPHC
output
pulse
has
a
peak
amplitude
that
Is
propor
tional
to
the
ratio
of
the
measured
time
Interval
to
the
selected
full-scale
Interval,
and
the
range
of
these
pulses
Is
0
through
+10
V.
The
467
Is
an
extremely
accurate
and
versatile
Instrument.
It
Is
composed
of
a
very
stable
gated
time
to
pulse
height
converter,
a
low-droop
stretcher,
a
strobed
TPHC
output,
and
a
single
channel
analyzer
that
can
be
operated
In
either
a
normal
or
window
mode.
The
Integrated
assembly.
In
a
NIM-standard
double-width
module,
combines
excel
lent
time
resolution
over
a
broad
dynamic
range
with
excel
lent
temperature
stability
and
linearity.
It
Is
dc-coupled
throughout
to
prevent
plleup
and
count-rate
distortion.
1.2.
OPERATION
Start-to-stop
time
conversion
Is
accomplished
only
after
a
valid
start
has
been
Identified
and
after
a
stop
pulse
has
arrived
within
the
selected
time
range.
The
start
Input
Is
disabled
during
the
busy
Interval
to
prohibit
plleup;
the
stop
Input
Is
disabled
after
the
first
accepted
stop
signal
.
Unwanted
stop
signals
that
occur
Immediately
after
a
start
Input,
such
as
those
In
linear
accelerator
applications,
for
example,
can
be
rejected
by
a
Stop
Inhibit
Mode
switch
and
a
circuit
that
Is
time-adjustable
from
0.1
to
1.0
jUS-
An
Inhibit/reset
circuit
also
permits
the
operator
to
abort
and
cancel
a
measurement
after
a
true
start
has
been
recognized.
The
Input
gate
for
the
start
circuit
can
be
operated
In
either
an
anticoincidence
or
a
coincidence
mode.
Time
ranges
may
be
switch-selected
for
full-scale
Intervals
from
50
ns
to
80
/ts.
Each
time
measurement
Is
analog-
stored
In
a
low-loss
stretcher
amplifier
until
a
linear
gate
Is
opened
by
either
an
Internal
or
an
external
strobe.
The
Internal
strobe
can
be
obtained
from
either
the
start
or
the
stop
Input
pulse,
and
In
either
case
occurs
automatical
ly
at
a
selected
delay
following
the
reference.
An
external
strobe
can
be
used
for
a
prompt
output
at
the
strobe
time
provided
that
a
time
measurement
has
been
completed
and
reset
has
not
occurred.
A
rear
panel
switch
can
select
either
5
or
120
us
after
stop
for
an
automatic
reset
If
no
strobe
has
been
furnished.
If
reset
occurs
before
a
strobe,
no
TPHC
output
signal
Is
available.
There
are
two
other
sources
for
reset:
one
occurs
If
the
start-to-stop
time
Interval
exceeds
the
range
that
Is
selected
and
the
other
occurs
as
a
result
of
an
Input
pulse
through
the
Inhibit/
Reset
Logic
Input
connector
on
the
front
panel.
The
normal
setting
for
the
rear
panel
switch
Is
120
ijls;
the
5-/is
setting
should
be
used
only
If
the
stop-strobe
mode
Is
used
and
the
delay
Is
adjusted
to
minimum,
or
If
the
external-
strobe
mode
Is
used
and
the
strobe
wi
ll
be
furnished
within
the
selected
Interval.
The
peak
amplitude
of
the
TPHC
signal
Is
sampled
by
the
SCA
at
the
time
of
a
true-stop
Input.
If
the
amplitude
Is
within
the
adjusted
acceptance
range
of
the
SCA,
an
SCA
logic
output
Is
generated.
The
width
of
the
SCA
output
Is
from
the
stop
Input
until
the
subsequent
reset.
Since
this
output
occurs
before
the
TPHC
signal
Is
used
to
generate
Its
analog
output,
the
SCA
output
can
be
used
to
Inhibit
a
TPHC
output,
unless
the
analog
signal
Is
within
the
SCA
window,
and
to
thus
limit
the
range
of
a
timing
spectrum
as
It
Is
stored
In
the
multichannel
analyzer.
The
single
channel
analyzer
has
a
lower
level
discriminator
that
can
be
adjusted
through
the
full
linear
range
of
the
TPHC
signals
from
0
through
10
V.
The
range
for
Its
upper-level
discriminator
Is
also
0
through
10
V,
but
the
zero
reference
point
for
the
ULD
must
be
selected
on
the
rear
panel
with
the
Window/Normal
switch.
When
the
switch
Is
set
at
Window,
the
zero
reference
for
the
ULD
Is
the
adjusted
setting
of
the
LLD
control.
When
the
switch
Is
set
at
Normal,
the
zero
reference
for
the
ULD
Is
ground
zero
and
Is
equal
to
the
LLD
zero
point.
1.3.
LOGIC
An
Input
can
be
accepted
through
the
Start
Input
connec
tor
on
the
front
panel
unless
the
467
Is
busy
processing
a
previous
set
of
Information
or
the
response
Is
Inhibited
by
a
gate
Input
condition.
The
acceptance
of
a
start
Input
Is
essential
In
order
to
Initiate
a
response
In
the
467.
When
a
start
Input
Is
accepted, a
positive
logic
signal
Is
available
through
the
rear
panel
True
Start
Output
connector
and
Is
continued
until
the
leading
edge
of
a
subsequent
reset.
The

reset
can
be
caused
by
a
TPHC
output,
by
the
sensing
of
an
overrange
condition,
or
by
an
inhibit/reset
signal
through
the
front
panel
BNC.
The
true-start
signal
permits
the
internal
circuits
to
start
measuring
a
time
interval
and
enables
the
stop
input
circuit.
The
Stop
Input
BNC
can
accept
an
input
signal
after
it
has
been
enabled
by
the
true-start
condition.
It
may
be
enabled
immediately
at
true
start,
or
the
rear
panel
Stop
Inhibit
Mode
switch
can
be
set
at
In
and
there
will
be
a
delay
from
true
start
before
the
stop
signal
can
be
accepted;
the
delay
range
is
0.1
through
1.0
fxs.
When
a
stop
input
signal
is
accepted,
this
indicates
that
an
interval
has
been
measured
and
its
analog
equivalent
is
stored
and
available.
A
signal
is
furnished
through
the
true-stop
output
that
continues
until
the
leading
edge
of
a
subsequent
reset.
If
no
stop
input
is
accepted
before
an
overrange
condition
is
sensed
or
before
an
inhibit/reset
input
is
furnished,
the
measurement
wil
l
be
aborted
and
no
output
signals
for
either
SCA
or
TPHC
will
be
generated.
At
the
true-stop
time
the
SCA
is
enabled
to
sample
the
peak
amplitude
of
the
stored
timing
signal
and
to
determine
whether
its
peak
amplitude
is
within
the
single-channel
acceptance
range.
If
the
SCA
responds,
it
generates
an
SCA
output
that
goes
high
about
600
ns
after
the
leading
edge
of
the
true
stop
and
this
signal
continues
until
the
trailing
edge
of
the
subsequent
reset.
If
the
SCA
does
not
respond
(because
the
amplitude
is
either
less
than
the
LLD
or
greater
than
the
ULD),
no
SCA
output
is
generated.
The
front
panel
SCA
Inhibit
switch
determines
whether
the
SCA
response
is
essential
in
order
to
generate
a
TPHC
output.
If
the
switch
is
set
at
In,
a
TPHC
output
is
generated
only
if
the
SCA
has
responded.
If
the
switch
is
set
at
Out,
the
generation
of
the
TPHC
output
is
independ
ent
of
the
SCA
response.
The
TPHC
output
must
be
strobed.
The
source
of
the
strobe
can
be
switch-selected
from
the
true-start
or
true-
stop
signal
or
from
an
external
signal
.
If
true
start
is
selected
as
the
reference,
the
strobe
occurs
after
a
fixed
delay
that
is
selected
by
the
Multiplier
switch
so
that
it
will
accommodate
the
maximum
range
time;
if
the
switch
is
set
at
XI,
the
delay
is
2
(js;
for
the
X10
setting,
the
delay
is
10
/is;
and
for
the
XI00
setting,
the
delay
is
100
/is.
If
true
stop
is
selected
as
the
reference,
the
strobe
occurs
after
a
delay
that
has
been
adjusted
with
the
front
panel
TPHC
Output
Delay
control,
1
to
10
/is
after
the
leading
edge
of
the
true-stop
signal
.
If
the
Strobe
Sync
switch
is
set
at
Ext,
a
signal
must
be
furnished
through
the
Strobe
Ext
BNC
connector
to
strobe
the
output
promptly.
The
reset
interval
is
5
/is
and
no
output
can
be
strobed
after
the
leading
edge
of
the
reset
pulse.
There
should
be
no
interference
if
the
Strobe
Reset
switch
is
set
at
120
/is
unless
external
strobe
is
being
used
and
the
strobe
input
pulse
does
not
arrive
within
the
interval
before
reset.
Reset
can
occur
as
the
result
of
the
completion
of
a
read
interval
in
which
the
TPHC
signal
is
furnished
as
an
output,
or
of
an
overrange
indication
where
no
significant
peak
amplitude
is
available,
or
of
an
inhibit/reset
input
that
cancels
the
cycle
at
its
leading
edge
and
inhibits
further
response
by
the
467.
The
principal
purpose
for
the
automatic
reset
is
to
furnish
this
function
if
external
strobe
is
being
used
and
the
input
pulse
is
not
furnished.
If
reset
occurs
for
any
reason
before
the
TPHC
output
is
completed,
the
TPHC
output
width
is
reduced
by
the
reset.
A
busy
output
starts
at
the
leading
edge
of
the
true-start
output
and
continues
until
the
trail
ing
edge
of
the
subsequent
reset.
This
can
be
used
to
control
external
equipment
by
indicating
each
interval
during
which
no
new
start
input
can
be
accepted.
2.
SPECIFICATIONS
2.1.
PERFORMANCE
Time
to
Pulse
Height
Converter
Time
Resolution
<10
ps
(10~^^
s)
FWHM
on
50-
and
100-ns
ranges;
<0.01%
FWHM
of
ful
l
range
for
all
other
ranges.
Temperature
Instability
<±10
<±0.015%/°C
for
higher
ranges.
ps/°C
for
50-ns
range;
Differential
Nonllnearlty
<±2%
from
10
ns
through
full
range
for
50-ns
range;
<±2%
from
5%
range
to
full
range
for
all
higher
ranges.
Integral
Nonllnearlty
<±0.1%
from
10
ns
through
full
range
for
50-ns
range;
<±0.1%
from
5%
range
to
full
range
for
all
higher
ranges.
Single
Channel
Analyzer
Temperature
Instability
ULD,
<±0.01
%/°C.
LLD,
<±0.01
%/°C.
Nonllnearlty
Effectively
determined
by
the
10-turn
poten
tiometers.
ULD,
<±0.5%
over
10-V
range.
LLD,
<±0.5%
over
10-V
range.

2.2.
CONTROLS
Range
fisec
Switch-selectable
1
5-range
choices
of
.05,
0.1,
0.2,
0.4
or
0.8
/Js
multiplied
by
XI,
X10,
or
X100;
the
XI
position
can
be
internally
modified
to
be
X1000
to
extend
time
range
capabil
ity
to
800
/is.
Multiplier
Front
panel
3-position
selector
switch;
settings
select
multiple
factors
for
the
selected
time
ranges
of
XI,
X10,
and
X100,
resulting
in
15
time
ranges
from
50
ns
to
80
/is.
TPHC
Output
Delay
Front
panel
10-turn
screwdriver
potentiometer
adjusts
the
output
delay
from
the
stop
input
to
the
internal
stop
strobe;
range,
<1
/is
to
>10
/is.
Anti
Coinc/Coinc
Front
panel
slide
switch
selects
either
coincidence
or
anticoincidence
logic
for
gating
the
start
input
circuit.
SCA
ULD
Front
panel
10-turn
potentiometer
determines
the
window
width
or
the
upper-level
discriminator
setting;
range,
0
to
10
V.
SCA
LLD
Front
panel
10-turn
potentiometer
adjustable
from
0
to
10
V.
SCA
Inhibit
Front
panel
slide
switch.
In
In
this
position
the
TPHC
output
pulse
is
avai
lable
only
if
the
output
level
fal
ls
within
the
SCA
window.
Out
In
this
position
the
switch
has
no
effect
on
the
TPHC
output.
DC
Adj
20-turn
potentiometer
to
adjust
the
dc
level
over
the
range
±0.5
V.
Strobe
Sync
Rear
panel
3-position
slide
switch
for
select
ing
one
of
three
modes:
Int
Start
In
this
position
the
information
is
strobed
out
~2
/IS
after
the
start
pulse
when
the
Multiplier
switch
is
in
the
XI
position,
~10
//s
in
the
XI0
position,
and
~100
fis
in
the
XI00
position.
Ext
In
this
position
a
positive
pulse
fed
into
the
Strobe
Ext
connector
wil
l
strobe
the
information
to
the
output
if
the
strobe
pulse
has
a
magnitude
of
+3
V
or
larger.
Int
Stop
In
this
position
the
information
is
strobed
out
1
to
10
/ts
(adjustable
by
the
TPHC
Output
Delay
control)
after
a
true-stop
pulse.
Strobe
Reset
Rear
panel
2-position
switch
that
allows
the
converter
to
be
reset
either
5
/ts
or
120
/is
after
a
true-stop
pulse
if
a
strobe
pulse
has
not
been
received.
SCA
Mode
Rear
panel
2-position
slide
switch:
Normal
Allows
independent
use
of
upper
level
discrimi
nator
and
lower
level
discriminator.
Window
ULD
setting
is
added
to
LLD
setting
when
switch
is
in
this
position.
Stop
Inhibit
Mode
Rear
panel
2-position
slide
switch:
In
Rejects
stop
pulses
that
occur
within
100
ns
to
within
1
JUS
(adjustable
by
the
Stop
Inhibit
Delay
control)
after
a
true-start
pulse.
Out
In
this
position
switch
does
not
affect
the
operation
of
the
instrument.
Stop
Inhibit
Delay
A
20-turn
trim
potentiometer
mounted
on
the
rear
panel
allows
the
stop
inhibit
period
to
be
adjusted
from
~100
ns
to
~1
jus
after
a
true-start
pulse.
2.3.
INPUTS
Start
Input
Front
panel
BNC
connector.
Amplitude
—250
mV
minimum;
protected
to
±100
V.
Zjj^
=
50^2,
dc-coupled.
Rise
Time
No
limit,
but
rise
time
should
be
as
short
as
possible
to
provide
maximum
accuracy.
Pulse
Width
3
ns
at
—250
mV;
maximum
l
imit,
~4
/is.
Stop
Input
Specifications
same
as
for
the
Start
Input.
Gate
Logic
Input
Front
panel
BNC
connector.
Logic
0,
<+2
V;
logic
1,
>±2
V;
input
protected
to
±100
V.
Gate
signal
must
occur
10
ns
before
the
start
and
must
overlap
the
start
input
pulse.
Impedance,
~1
kJ2,
dc-
coupled.
Inhibit/Reset
Logic
Input
Front
panel
BNC
connector.
Ampl
itude
of
>4
V
resets
circuit
at
any
point
in
the
cycle
and
inhibits
start
pulses
for
the
duration
of
the
pulse;
input
protected
to
+12
V.
Strobe
Ext
Rear
panel
BNC
connector.
Amplitude
>+2
V;
protected
to
>±25
V.
Rise
Time
No
limit.
Pulse
Width
10
ns
minimum;
~4
/is
maximum.
Impedance
1
kf2
dc-coupled.
Control
Outputs
Prompt
with
strobe
input.
2.4.
OUTPUTS
TPHC
Outputs
Front
and
rear
panel
BNC
connectors.
100%
protected
from
short
circuit
and
excessive
duty
cycle.
Unipolar
0
to
+10
V
linear;
<500
ns
rise
time.
Width
Internal
ly
adjustable
from
~1.0
to
2.5
/is.
Output
Timing
Prompt
with
either
internal
or
external
strobe.
Impedance
<1f2
on
front
panel
and
93^2
on
rear
panel,
dc-coupled.
Output
dc
Level
Adjustable
from
0
to
±0.5
V
dc
with
front
panel
DC
Adj
screwdriver
control.
SCA
Outputs
Front
and
rear
panel
BNC
connectors.
100%
protected
from
short
circuit
and
excessive
duty
cycle.
Amplitude
^4
V
on
positive
logic
if
TPHC
pulse
is
in
the
LLD-ULD
window
and
0
V
if
TPHC
is
not
in
the
window.
Output
Timing
Pulse
begins
~600
ns
after
a
valid
stop
pulse
and
continues
until
TPHC
resets.
Impedance
lOfl,
dc-coupled.

True
Start
Output
Rear
panel
BNC
connector
provides
a
positive
logic
timing
output
to
indicate
the
interval
from
an
accepted
start
input
signal
until
reset.
Rise
Time
<100
ns.
Output
Width
The
interval
from
the
start
input
until
reset
time,
which
can
occur
at
strobe
time,
overrange,
or
120
/is
after
stop
signal.
Impedance
<10J2,
dc-coupled.
True
Stop
Output
Rear
panel
BNC
connector
provides
a
positive
4-V
pulse
to
indicate
valid
stop
and
the
interval
from
an
accepted
stop
input
signal
until
reset
occurs.
Rise
Time
<100
ns.
Impedance
=10^2,
dc-coupled.
Output
Width
The
interval
from
the
stop
input
until
reset
time.
TPHC
Busy
Output
Via
rear
panel
BNC
connector
to
indicate
the
total
time
that
the
467
is
involved
in
a
conversion;
amplitude,
+4
V;
<100
ns;
Z^,
10J2,
dc-coupled.
Output
width
Is
equal
to
the
interval
from
the
start
input
to
5/is
after
reset.
Stop
Inhibit
Monitor
Rear
panel
BNC
connector
provides
a
positive
3.5-V
pulse
to
indicate
the
time
period
during
which
stop
signals
are
inhibited.
Rise
Time
<100
ns.
Output
Width
Variable
from
100
ns
to
>1.0/is
with
Stop
Inhibit
Delay
trim
potentiometer,
beginning
when
a
true-
start
pulse
is
received.
Stop
pulses
are
rejected
until
this
pulse
returns
to
the
baseline
if
the
Stop
Inhibit
Mode
switch
is
in
the
In
position.
Impedance
~10J2,
dc-coupled.
2.5.
ELECTRICAL
AND
MECHANICAL
Power
Required
+24
V,
165
mA;
-H2
V,
320
mA;
-24
V,
120
mA;
-12
V,
140
mA.
Dimensions
NIM-standard
double-width
module
(2.70
in.
wide
by
8.714
in.
high)
per
TID-20893.
3.
INSTALLATION
3.1.
GENERAL
An
ORTEC
401/402
Series
Bin
and
Power
Supply,
or
equal,
in
which
the
467
will
be
installed,
is
intended
for
rack
mounting.
If
vacuum
tube
equipment
is
operated
in
the
same
rack,
there
must
be
sufficient
cool
air
circulating
to
prevent
localized
heating
of
the
all-transistor
circuits
in
the
467
and
in
the
other
modules
in
the
Bin
and
Power
Supply.
Rack-mounted
equipment
subjected
to
the
temper
atures
in
vacuum
tube
equipment
can
exceed
the
maximum
for
which
the
transistorized
circuits
are
designed
unless
this
precaution
is
taken.
The
467
should
not
be
subjected
.to
temperatures
in
excess
of
120°F
(50°C).
3.2.
CONNECTION
TO
POWER
The
467
is
designed
per
TID-20893
and
accepts
its
operating
power
requirements
through
a
mating
power
connector
when
it
is
installed
in
an
ORTEC
401/402
Series
Bin
and
Power
Supply.
As
a
safety
precaution,
always
turn
off
the
power
for
the
Bin
before
inserting
or
removing
any
modules.
If
all
the
modules
installed
in
the
Bin
are
ORTEC
400
and/or
700
Series
instruments,
there
wil
l
be
no
overload
on
any
portion
of
the
Power
Supply.
However,
if
any
modules
not
designed
by
ORTEC
are
included
in
the
Bin,
this
protection
may
not
be
effective;
monitor
the
dc
voltages
at
the
test
points
on
the
control
panel
of
the
Bin
after
all
modules
have
been
installed
and
the
power
is
turned
on,
in
order
to
determine
that
none
of
the
four
power
levels
have
been
reduced
by
an
overload.
3.3.
CONNECTION
INTO
A
SYSTEM
The
467
can
accept
both
start
and
stop
pulses
from
discriminators
that
furnish
NIM-standard
fast
negative
logic
signals
or
from
the
timing
output
of
a
photomultiplier
tube
base.
Typical
ORTEC
instruments
that
provide
compatible
signals
are
the
416A,
403A,
473,
and
260
discriminators
and
the
265,
269,
270,
and
271
Photomultiplier
Tube
Bases.
The
start
and
stop
inputs
wi
ll
properly
terminate
5012
cable,
and
this
type
is
recommended
to
ensure
proper
termination
of
the
signals.
No
input
or
output
connectors
need
be
terminated
when
they
are
not
in
use.
In
any
experiment
in
which
it
is
reasonable
to
assume
that
the
count
rates
for
start
and
stop
will
be
equal
or
nearly
so,
use
the
signal
furnished
from
the
origin
of
events
into
the
start
input
and
the
signal
furnished
from
the
response
into
the
stop
input.
The
467
will
then
measure
the
time
difference
T
from
origin
to
response
and
furnish
an
output
amplitude
that
is
some
fraction
of
the
selected
ful
l-scale
amplitude,
proportional
to
the
ratio
of
T
to
the
selected
ful
l-scale
time
range.
In
any
experiment
in
which
the
two
count
rates
differ
noticeably,
such
as
one
in
which
fewer
responses
than
event
origins
can
be
expected,
use
the
lower
count
rate
as
the
start
input
to
the
467.
This
assures
that
the
467
dead
time
will
be
minimized,
because
it
analyzes
the
time
difference

only
after
a
start
signal
is
accepted.
When
the
response
Is
used
as
a
start
signal,
furnish
the
signals
from
the
origin
of
events
through
a
delay
line
into
the
stop
input,
and
adjust
the
delay
to
match
the
selected
full-scale
time
of
the
467.
At
each
start
input
signal
the
467
wil
l
analyze
the
time
until
its
related
origin
signal
is
furnished
to
the
stop
input.
The
time
measured
is
then
delay
time
minus
T,
and
produces
a
so-called
inverted
time
spectrum.
The
purpose
for
this
type
of
system
connection
is
to
reduce
the
number
of
conversions
and
the
corresponding
dead
time
during
the
experiment.
For
each
signal
accepted
through
the
start
input
there
must
be
a
conversion,
but
for
each
signal
through
the
stop
input
there
need
not
be
a
conversion.
For
each
start
signal
that
is
not
followed
by
a
stop
signal
within
the
selected
time
full
range,
the
converter
measures
a
time
equal
to
the
total
range,
even
though
no
output
pulse
is
generated.
3.4.
LINEAR
OUTPUT
SIGNAL
CONNECTIONS
AND
TERMINATING
IMPEDANCE
The
source
impedance
of
the
standard
TPHC
output,
with
the
0-
to
10-V
linear
range,
is
about
1J2
through
the
connector
on
the
front
panel
and
93J2
through
the
connector
on
the
rear
panel.
For
the
front
panel
circuit
the
interconnection
to
other
modules
does
not
usually
require
any
special
considera
tions,
especially
if
the
interconnecting
cable
is
shorter
than
4
ft
in
length.
Paralleling
several
loads
on
a
single
output
will
still
not
reduce
the
0-
to
10-V
signal
span
significantly
unless
the
combined
load
is
less
than
lOOfi.
The
rear
panel
TPHC
output
circuit
is
designed
for
use
of
93i7
cable
to
transfer
the
signals
into
a
measuring
circuit
that
has
an
input
impedance
of
at
least
10000.
With
this
series
impedance-matched
circuit
connection,
there
will
usually
not
be
any
interference
with
the
signal.
If
oscilla
tions
should
occur,
it
will
be
necessary
to
provide
an
additional
shunt
termination
of
1000
in
paral
lel
with
the
input
circuit
of
the
receiving
instrument,
but
this
wi
ll
result
in
about
a
50%
loss
of
signal
amplitude.
As
with
any
analog
instrument,
osci
l
lations
may
be
ob
served
occasional
ly
when
unterminated
lengths
of
cable
are
used.
Short
cable
lengths
(up
to
4
ft)
need
not
be
terminated.
When
longer
cable
lengths
are
required
for
transfer
of
a
linear
signal,
the
cable
should
be
terminated
in
a
resistive
load
equal
to
the
cable
impedance
to
prevent
reflections
and
osci
l
lations
in
the
cable.
Osci
llation
suppres
sion
can
be
effected
by
either
a
series
termination
at
the
sending
end
of
the
cable
or
by
a
shunt
termination
at
the
receiving
end.
For
convenience
a
BNC
tee
can
usually
accommodate
both
the
cable
and
a
mating
terminator
at
the
input
of
the
receiving
instrument.
These
units
are
available
commercial
ly,
including
BNC
terminators
with
nominal
values
of
50,
100,
and
1000J2.
ORTEC
stocks
a
limited
quantity
of
all
but
the
lOOOfi
terminators
for
your
convenience,
as
listed
below:
BNC
Tee
Connector
50F2
Terminator
lOOfZ
Terminator
ORTEC
C-29
ORTEC
C-28
ORTEC
C-27
When
a
shunt
termination
at
the
receiving
end
of
the
cable
is
impractical,
consider
series
termination
at
the
sending
end.
For
a
series
termination
the
full
signal
amplitude
span
is
avai
lable
at
the
receiving
end
only
if
the
input
impedance
is
many
times
the
characteristic
impedance
of
the
cable.
For
series
termination
install
the
correct
resistance
between
the
actual
amplifier
output,
on
the
etched
circuit
board,
and
the
output
connector.
Effectively,
the
terminating
resistance
is
in
series
with
the
input
impedance
of
the
receiving
instrument,
and
may
result
in
some
loss
in
signal
amplitude.
For
example,
if
the
series
terminator
is
9312
and
the
driven
load
is
90012,
the
available
signal
span
wil
l
be
only
about
90%
of
the
maximum
signal
amplitude
for
each
pulse.
The
termination
of
a
9312
cable
in
a
9312
load
will
cause
a
50%
loss
for
the
signal
.
3.5.
LOGIC
SIGNAL
CONNECTIONS
The
start
and
stop
input
circuits
accept
NIM-standard
fast
negative
pulses.
Each
of
these
input
circuits
is
designed
with
a
5012
input
impedance
and
is
intended
as
the
proper
termination
for
the
signals,
furnished
through
5012
cables.
Impedance
considerations
for
each
of
the
remaining
logic
inputs
and
output
for
the
467
are
noncritical
and
9312
cable
is
usually
used.
They
can
be
terminated
with
10012
to
prevent
ringing
if
the
signal
is
used
to
drive
a
high-
impedance
load.
4.
OPERATING
INSTRUCTIONS
4.1.
TIME
TO
PULSE
HEIGHT
CONVERSION
There
are
seven
front
panel
controls
on
the
467.
Of
these,
four
are
directly
associated
with
the
conversion
of
a
start-to-stop
interval
into
an
analog
equivalent
TPHC
output
pulse.
These
controls
are
Range
;usec.
Multiplier,
TPHC
Output
Delay,
and
Anti
Coinc/Coinc.
If
the
SCA
Inhibit
switch
is
set
at
In,
this
also
affects
the
generation
of
a
TPHC
output.

The
Range
jusec
and
Multiplier
switches
determine
the
ful
l-scale
limit
for
time
conversion.
Any
of
15
combinations
may
be
selected
as
follows:
Switch
Settings
Full-Sea
le
Time
Limit
Range
Multiplier
0.05
XI
50
ns
0.1
XI
100
ns
0.2
XI
200
ns
0.4
XI
400
ns
0.05
X10
500
ns
0.8
XI
800
ns
0.1
X10
1
IIS
0.2
X10
2
IIS
0.4
X10
4
us
0.05
XI00
5
IIS
0.8
X10
8
IIS
0.1
XI00
10
jUs
0.2
XI00
20
IIS
0.4
X100
40
IIS
0.8
XI00
80
/is
For
example,
with
the
Range
switch
set
at
0.05
and
the
Multiplier
switch
at
XlOO,
the
ful
l-scale
time
range
is
5
iJ.s.
Any
stop
input
signal
that
occurs
within
5
/ts
after
a
true-start
signal
wil
l
initiate
the
gating
of
an
output
pulse
through
both
TPHC
Output
connectors.
The
output
pulse
wi
l l
not
be
furnished
through
these
connectors
unless
it
is
strobed.
The
strobe
condition
is
selected
by
a
rear
panel
switch
and
can
be
based
on
the
time
of
the
true-start
signal,
delayed
by
an
amount
of
time
greater
than
the
selected
full-scale
time,
or
by
the
stop
signal
,
delayed
by
an
amount
of
time
adjusted
with
the
front
panel
TPFIC
Output
Delay
control,
or
by
an
external
strobe
input
signal
.
When
the
output
does
occur,
its
peak
amplitude
wil
l
be
proportional
to
the
ratio
of
the
measured
start-to-stop
interval
to
the
selected
full-scale
time,
in
a
0-
to
10-V
range.
Internal
logic
eliminates
any
pulse
ambiguity.
No
output
pulse
is
furnished
unless
a
stop
signal
is
accepted
within
the
selected
full-range
time.
A
stop
signal
is
not
effective
unless
it
is
preceded
by
a
true-start
signal
.
For
further
logical
control
either
a
coincidence
or
anticoincidence
mode
can
be
selected
for
gating
control
of
the
start
input
circuit.
To
eliminate
gating
for
the
start
input,
set
the
Gate
switch
at
Anti
Coinc
and
leave
the
gate
input
circuit
without
any
connection.
When
the
same
switch
setting
is
used
and
an
input
signal
is
furnished,
start
signals
are
not
accepted
when
the
gate
signal
is
+2
V
or
more.
For
coincidence
gating
of
the
start
input
circuit
set
the
Gate
switch
at
Coinc
and
furnish
a
signal
of
+2
V
or
more
through
the
Gate
Logic
Input
connector
when
start
signals
are
to
be
accepted.
If
a
signal
is
furnished
through
the
Inhibit/Reset
connector
on
the
front
panel
,
any
time
measurement
that
may
be
in
process
will
be
aborted
and
no
new
measurement
can
begin
until
the
inhibit/reset
signal
is
removed.
To
be
effective,
the
inhibit/reset
signal
must
precede
an
output
strobe
time.
The
rear
panel
Strobe
Sync
switch
selects
the
source
for
the
strobe
signal
for
the
TPFIC
output.
When
the
switch
is
set
at
Int
Start,
the
strobe
is
generated
by
a
delayed
true-start
signal
and
the
delay
is
fixed
at
a
time
that
is
longer
than
the
selected
full-range
time.
When
the
switch
is
set
at
Int
Stop,
the
strobe
is
generated
by
a
delayed
true-stop
signal
and
the
delay
is
adjusted
by
the
TPHC
Output
Delay
control
on
the
front
panel
within
the
range
of
1
to
10
/js.
When
the
switch
is
set
at
Ext,
an
input
pulse
must
be
furnished
through
the
adjacent
BNC
connector
and
the
TPHC
output
signal
is
strobed
promptly
at
the
external
signal
time.
The
timing
of
an
external
strobe
input
signal
must
be
within
the
switch-selected
interval
,
5
or
120
iis,
after
the
true-stop
pulse
for
the
measurement.
If
the
strobe
is
furnished
prior
to
stop,
the
signal
is
not
accepted.
In
the
case
of
a
strobe
pulse
failing
to
arrive
within
the
selected
interval
after
stop,
the
467
wil
l
have
been
automatical
ly
reset
internally:
so
there
is
no
available
output
and
the
strobe
signal
is
ignored.
If
the
467
is
to
be
used
as
a
stretcher
or
buffer
storage
of
the
analog
output
for
more
than
120
ijls,
a
simple
modification
can
increase
this
time
limit;
see
Section
6.5.
By
the
addition
of
four
jumpers
in
the
467
circuit
the
XI
setting
of
the
Multipl
ier
switch
can
be
replaced
by
a
X1000
setting.
If
these
jumpers
are
added,
as
described
in
Section
6.4,
the
minimum
time
range
setting
is
500
ns
and
the
maximum
range
is
800
/is.
In
any
application
of
the
467
in
which
stop
signals
occur
immediately
following
a
true
start
but
are
to
be
ignored,
in
favor
of
time
measurements,
to
stop
signals
that
occur
later,
the
rear
panel
Stop
Inhibit
circuit
can
be
used.
With
the
Mode
switch
set
at
In,
the
adjacent
screwdriver
control
can
be
adjusted
for
a
delay
of
0.1
to
1.0
/is
and
al
l
stop
signals
are
ignored
that
occur
within
the
adjusted
delay
time
fol
lowing
a
true-start
signal
.
The
adjusted
time
delay
can
be
monitored
through
the
rear
panel
Monitor
connector
to
measure
the
duration
of
the
delay
signal
that
occurs
at
each
true-start-input
time.
4.2.
SINGLE
CHANNEL
ANALYSIS
The
single
channel
analyzer
portion
of
the
467
measures
the
peak
amplitude
of
each
analog
TPHC
pulse
as
soon
as
it
is
formed.
If
the
peak
amplitude
is
within
the
lower
and
upper
limits
that
are
set
with
the
LLD
and
ULD
controls
on
the
front
panel,
an
SCA
output
signal
is
generated
after
a
400-ns
propagation
time.
The
signal
can
be
used
to
control

the
subsequent
generation
of
a
TPHC
output
if
the
front
panel
SCA
Inhibit
switch
is
set
at
In.
Under
this
condition
the
SCA
must
generate
an
output
in
order
to
permit
the
generation
of
a
TPHC
output
at
its
strobe
time,
and
this
can
limit
the
range
of
time
signals
that
are
furnished
to
a
multichannel
analyzer
to
generate
a
spectrum.
An
inde
pendent
SCA
output
signal
is
furnished
through
both
the
front
and
rear
panel
SCA
Output
BNC
connectors
for
any
external
applications
that
may
be
desired.
A
rear
panel
switch
selects
either
Window
or
Normal
mode
for
the
SCA.
In
the
Window
mode
the
acceptance
range
is
from
the
LLD
setting
to
the
sum
of
the
LLD
plus
ULD
settings.
In
the
Normal
mode
the
LLD
control
and
the
ULD
control
each
operates
independently
and
the
acceptance
range
is
from
the
LLD
setting
to
the
ULD
setting;
the
ULD
setting
must
always
be
greater
than
the
LLD
setting
to
generate
an
SCA
output.
5.
CIRCUIT
DESCRIPTION
5.1.
GENERAL
The
circuits
for
the
467
are
shown
in
schematics
467-0201-SI
and
467-0301-81
and
in
block
diagram
467-0101
-B1.
All
these
drawings
are
included
at
the
back
of
the
manual.
The
components
that
are
included
in
schematic
467-020TS1
all
have
reference
designation
numbers
that
are
less
than
300,
and
those
on
the
467-0301-SI
drawing
have
reference
designations
in
the
300
series.
The
division
of
circuits
between
the
two
boards
on
which
the
compo
nents
are
mounted
is
indicated
in
the
block
diagram.
The
nomenclature
used
to
identify
the
integrated
circuit
packages
referred
to
in
this
manual
is
defined
below
for
the
example
IC2(10),
where
IC
=
integrated
circuit,
2
=
component
number,
(10)
=
pin
number.
Any
portion
of
an
IC
package
can
be
designated
by
its
output
pin
number.
5.2.
TPHC
CIRCUIT
The
start
input
circuit
contains
D1,
Q1,
and
resistors
R1
through
R4,
which
form
a
voltage
limiter
to
protect
the
base
input
circuit
of
Q2.
A
negative
input
signal
of
250
mV
or
more,
limited
at
700
mV,
is
amplified
by
Q2
and
Q3
and
causes
tunnel
diode
D3
to
switch
from
its
low
state
(—50
mV)
to
its
high
state
(—450
mV).
The
tunnel
diode
remains
in
the
high
state
until
reset
is
furnished
from
Q51
to
trigger
D3
back
to
its
low
state.
While
tunnel
diode
D3
is
in
its
high
state,
the
current
that
normally
flows
through
Q5
is
switched
to
Q4
to
perform
two
functions:
It
drives
the
true-start-output
circuit
(Q5,
Q6,
IC1,
and
Q56)
for
an
external
timing
signal
keyed
to
an
accepted
and
valid
start
input,
and
it
switches
the
current
that
normally
flows
through
Q12
to
Q11.
While
current
flows
through
Q11,
Q10
is
held
at
cutoff
and
a
constant
current
from
Q46
charges
the
selected
timing
capacitor
linearly;
the
timing
capacitor
(C6and
possibly
C32
or
C33)
is
selected
by
the
Range
Multipl
ier
switch,
S2.
If
the
XI
selection
is
changed
to
a
X1000
selection,
C29
and
C31
can
also
be
selected
as
a
part
of
the
timing
capacitor.
The
voltage
to
which
the
selected
timing
capacitor
can
be
charged
is
limited
to
3.6
V
by
Q7.
The
voltage
developed
across
the
timing
capacitor
is
appl
ied
to
the
input
of
a
stretcher
amplifier
that
includes
Q41
through
Q44.
This
amplifier
has
a
gain
of
unity
and
its
output
voltage
is
obtained
from
the
gate
of
Q44B.
If
a
stop
signal
is
not
furnished
within
the
selected
time
range,
Schmitt
trigger
Q53
and
Q54
fires
at
a
timing
capacitor
level
of
about
3
V
from
a
10-V
signal
on
Q37E.
This
identifies
an
overrange
condition.
The
output
triggers
the
reset
circuit,
which
includes
Q48,
Q49,Q51,
and
Q52.
The
reset
circuit
causes
tunnel
diode
D3
to
return
to
its
low
state
without
generating
a
TPHC
output
pulse.
If
an
external
stop
signal
is
received
prior
to
the
overrange
identification,
Q46
is
switched
off
and
the
timing
capacitor
receives
no
further
charge.
The
charge
that
accumulated
on
the
timing
capacitor
between
the
start
and
stop
input
pulses
determines
the
capacitor's
voltage
and
therefore
the
output
voltage.
In
the
stop
input
circuit,
diode
D9
and
its
associated
resistors
form
an
amplitude-limiting
circuit
similar
to
the
start
input
circuit.
A
valid
start
pulse
enables
the
stop
circuit
through
Q17.
The
stop
pulse
is
amplified
by
Q18
and
Q19.
If
the
stop
pulse
arrives
while
D11
is
enabled,
this
diode
is
driven
to
its
high
state
to
switch
the
current
from
Q21
through
Q20.
When
Q20
conducts,
Q45
is
driven
into
conduction
and
the
constant
current
through
Q46
is
cut
off.
The
selected
timing
capacity
is
not
charged
any
further

and
has
no
discharge
path;
so
It
holds
the
voltage
to
which
it
has
been
charged.
Switching
pair
Q21
and
Q20
also
performs
two
additional
functions.
If
the
internal
stop
strobe
is
used,
a
trigger
is
furnished
to
the
read
timer,
Q24
and
Q27
through
Q29.
This
circuit
generates
a
delay
of
1
to
10
ms
and
then
operates
gate
circuit
Q35
and
Q36.
Switching
pair
Q21
and
Q20
also
furnishes
a
strobe
to
the
SCA
circuit
through
Q22
and
Q23.
Read
timer
Q27
and
Q28
performs
two
functions:
It
opens
linear
gate
Q36
and
triggers
reset
through
Q52.
Timing
for
Q27
and
Q28
is
adjusted
by
R86,
and
the
gate
width
is
normally
set
at
about
2
fis.
The
linear
gate
is
opened
during
this
period
and
reset
occurs
at
the
end
of
the
period.
At
reset
tunnel
diode
D3
is
switched
to
its
low
state
and
cannot
be
switched
back
to
high
again
for
about
5
/ts;
this
feature
prevents
pi
leup.
When
D3
is
reset,
it
forces
D11
to
its
low
state
through
Q4,
Q12,
and
Q17.
Linear
gate
Q36
conducts
a
current
with
an
amplitude
proportional
to
the
charge
stored
in
the
timing
capacity,
and
the
charge
is
proportional
to
the
time
between
start
and
stop
input
signals.
The
gating
pulse
from
Q33
causes
Q35
to
conduct
to,
in
turn,
cut
off
Q36.
The
current
that
was
flowing
through
Q36
is
rerouted
into
OA-1
during
the
period
of
the
gating
pulse.
OA-1
is
a
high-gain
operational
amplifier
that
has
its
gain
regulated
to
—1
for
positive
input
signals
and
to
0
for
negative
input
signals.
The
output
from
OA-1
drives
another
operational
ampl
ifier,
OA-2
on
the
0301
board,
that
has
a
gain
of
—1
and
generates
the
positive
TPHC
output
signals.
The
TPHC
information
may
be
strobed
out
in
three
different
ways:
The
strobe
pulse
can
be
referenced
inter
nal
ly
to
the
stop
signal,
internally
to
the
start
signal,
or
externally.
In
each
case
a
one-shot
multivibrator
is
triggered
to
generate
the
gate
pulse.
In
the
stop-strobe
mode
R73
and
R74
with
C13
determine
the
time
from
the
stop
pulse
until
Q24
conducts
and
the
one-shot
circuit
is
triggered.
The
delay
is
adjusted
from
1
to
10
/ts
by
R73
on
the
front
panel.
In
the
start-strobe
mode
the
true-start
signal
from
IC1(12)
triggers
a
one-shot
multivibrator
that
includes
IC3(6),
C55,
R189,
and
R191.
The
period
of
this
one-shot
circuit
is
variable
internally
and
determines
the
delay
between
the
start
pulse
arid
the
strobe
pulse.
The
start-
strobe
signal
drives
the
gate
one-shot
that
includes
IC4,
Q59,
and
Q26.
In
the
external
strobe
mode
a
positive
pulse
through
the
Strobe
Ext
input
connector
drives
the
gate
one-shot
through
Q26.
Logic
signals
corresponding
to
true
start,
true
stop,
and
TPHC
busy
are
generated
by
the
circuits
that
include
IC1,
IC2,
Q56, Q57,
and
Q58.
If
for
any
reason
a
strobe
signal
is
not
received,
the
TPHC
circuits
are
reset
after
5
or
120/ts,
selected
by
S5,
through
Q33
and
the
overrange
circuit,
by
R103,
C19,
and
C22.
For
this
condition
there
is
no
TPHC
output
signal.
5.3.
SINGLE
CHANNEL
ANALYZER
CIRCUIT
There
are
two
discriminator
circuits
in
the
SCA.
The
lower-level
discriminator
uses
IC304, IC305,
and
IC306.
The
upper-level
discriminator
uses
IC301,
IC302,
and
IC303.
Both
channels
obtain
their
reference
voltage
from
D301
and
the
maximum
voltage
at
TP1
and
TP2
is
adjusted
to
—5
V
with
R301
and
R316.
The
TPHC
ampl
itude
from
Q37
is
divided
by
R333
and
R334,
and
50%
of
the
TPHC
level
is
compared
in
IC303
and
IC306
to
their
reference
levels,
determined
by
the
settings
of
R302
and
R317
on
the
front
panel,
and
to
the
setting
of
rear
panel
SCA
Mode
switch
S6.
A
signal
is
obtained
from
the
stop
circuit
through
Q23
to
trigger
a
one-shot
that
includes
IC308,
R336,
and
C352.
The
output
of
the
one-shot
is
used
to
strobe
the
lower
and
upper
comparators.
If
the
TPHC
signal
exceeds
the
lower
reference
but
does
not
exceed
the
upper
reference,
a
positive
logic
output
of
4
V
is
obtained
at
Q305
through
IC307,
IC308,
and
IC309.
If
the
TPHC
signal
is
less
than
the
lower
reference
or
greater
than
the
upper
reference,
the
gate
circuits
in
IC309
are
not
activated
and
there
is
no
SCA
output.
When
switch
S6
is
set
at
Window,
the
lower-level
and
upper-level
reference
voltages
are
summed
at
the
input
of
IC302.
For
this
condition
the
window
is
the
difference
between
the
lower-level
adjustment
and
the
sum
of
both
the
lower-
and
upper-level
adjustments
within
the
linear
range
limits
of
the
SCA.
5.4.
AUXILIARY
LOGIC
The
467
includes
an
input
gate
circuit,
a
stop
inhibit
circuit,
an
inhibit-and-reset
circuit,
and
an
SCA
inhibit
circuit.
The
function
of
each
of
these
logic
circuits
is
defined
in
Sections
4.1
and
4.2.
The
input
gate
circuit
consists
of
D7,
Q13,
Q15,
andQ16.
When
Gate
switch
S8
is
set
at
Anti
Coinc,
Q13
is
saturated
and
a
negative
start
input
pulse
can
trigger
tunnel
diode
D3.
If
a
positive
gate
input
signal
is
furnished
through
CN2,
Q16
conducts
and
clamps
D3
to
prevent
it
from
respond
ing.
With
the
Gate
switch
set
at
Coinc
Q13
clamps
D3and
prevents
it
from
responding
except
when
a
positive
input
is
furnished
through
CN2.
Stop
signals
can
be
inhibited
from
triggering
D11
through
a
controlled
interval
following
each
true
start.
This
function
uses
IC3
and
IC4.
The
start
signal
triggers
a
one-shot
multivibrator
that
includes
IC3,
C56,
R196,
and
R197.
The
output
of
IC4
is
normally
high
and
is
driven
low
during
the
period
of
the
one-shot.
If
Stop
Inhibit
switch
S7
is
set
at
In,
the
positive
signal
from
IC4
clamps
Q17
through
Q16
to
prevent
response
in
D11.
After
IC4
returns
to
low,
Q17
conducts
and
enables
D11
to
accept
a
stop
input
pulse.
A
positive
monitor
output
is
produced
through
Q60
during
the
inhibit
interval.

A
signal
through
the
Inhibit/Reset
connector
on
the
front
panel
activates
the
overrange
circuit,
Q53
and
Q54,
and
resets
the
TPHC
circuit
at
any
time
during
a
cycle.
A
positive
input
through
CN6
drives
the
base
of
Q53
negative
through
Q55.
The
reset
circuit
forces
D3
to
its
low
state
and
clamps
it
until
the
external
signal
drops
below
approximately
3.5
V.
During
this
period
a
start
signal
wi
ll
not
be
accepted.
If
SCA
Inhibit
switch
S3
is
set
at
In,
an
SCA
output
signal
is
necessary
in
order
to
enable
a
strobed
TPHC
output.
With
the
switch
in
this
position
Q30
is
normal
ly
saturated
and
blocks
the
gating
signal
at
Q31.
A
positive
SCA
output
signal
("SCA
OK")
cuts
off
Q30
and
al
lows
the
gating
signal
to
propagate
through
Q31
to
the
gating
circuit.
6.
MAINTENANCE
3
6.1.
TESTING
PERFORMANCE
The
following
test
procedures
are
furnished
as
a
guide
during
installation
and
checkout
of
the
467
TPHC/SCA.
Test
Equipment
The
following
test
equipment
is
recom
mended.
Each
test
procedure
refers
to
this
list
by
the
unit
identification
number
for
the
required
item(s)
of
test
equipment.
An
equivalent
unit
may
be
substituted
for
any
item
in
the
list,
providing
that
it
furnishes
the
function
required
for
each
specific
application.
1.
Hewlett-Packard
222A
Pulse
Generator
2.
ORTEC
436
100
MHz
Discriminator
3.
ORTEC
416A
Gate
and
Delay
Generator
4.
ORTEC
425
Nanosecond
Delay
5.
Photomultiplier
tube
with
scintillator
and
radiation
source
6.
ORTEC
403A
Time
Pickoff
Control
7.
ORTEC
449
Log/Lin
Ratemeter
8.
Tektronix
Type
585
Oscilloscope
9.
ORTEC
6240
Multichannel
Analyzer
10.
ORTEC
414A
Fast
Coincidence
11.
ORTEC
444
Gated
Biased
Amplifier
Preliminary
Procedures
Take
the
following
preliminary
steps
when
the
467
is
installed:
1.
Check
the
module
visually
for
possible
damage.
2.
With
the
power
turned
off,
instal
l
the
467
into
a
NIM-standard
Bin
and
Power
Supply
such
as
one
of
the
ORTEC
401/402
Series.
3.
Check
the
installation
for
proper
mechanical
alignment.
4.
Switch
on
ac
power
and
check
the
dc
power
voltage
levels
at
test
points
on
the
402
Power
Supply
control
panel.
Basic
Switch
Settings
Set
the
467
controls
as
follows:
Range
Multiplier
OLD
LED
SCA
Inhibit
Logic
Input
Strobe
Sync
SCA
Mode
Strobe
Reset
Stop
Inhibit
Mode
0.05
/us
1
10
(fully
clockwise)
0
(fully
counterclockwise)
Out
Anti
Coinc
Int
Stop
Normal
120
/us
Out
Conversion
Tests
Use
the
typical
test
setup
shown
in
Fig.
6.1
and
supply
a
start
and
stop
pair
of
input
signals
with
known
time
difference
into
the
467.
Observe
the
TPHC
output.
Then
use
the
following
procedures:
1.
Adjust
the
delay
for
the
stop
input
to
25
or
30
ns
more
than
the
basic
13
ns
required
for
a
minimum
response.
2.
Check
to
see
that
the
full-scale
time
range
is
0.05/us
X
1,
or
50
ns.
PULSE
FANOUT
Neg.
Output
(11
(2)
or
(3)
Start
Input
0.5
v
2
ns
-
LJ
DELAY
(3)
Neg.
ORTEC
OSCILLO
Output
SCOPE
467
(8)
Stop
Input
Fig.
6.1.
Test
System
for
Checking
Conversion.

10
3.
Measure
the
signal
through
the
TPHC
Output
connec
tor.
It
should
be
about
6
V
for
a
30-ns
delay
or
5
V
for
a
25-ns
delay.
4.
Turn
the
Range
fisec
switch
through
its
.1, .2, .4,
and
.8
settings
and
observe
the
pulse
amplitude
at
each
setting:
each
successive
switch
position
should
decrease
the
pulse
amplitude
to
about
1/2
of
the
amplitude
for
the
previous
setting.
5.
Return
the
Range
/isec
switch
to
.05
and
set
the
Multiplier
switch
at
10.
The
output
amplitude
should
be
reduced
to
about
1
/10
of
the
reading
for
step
3.
Resolution
Tests
See
Fig.
6,2
for
the
typical
test
setup
used
for
resolution
checks.
The
start
and
stop
pulses
used
for
this
test
must
have
fast
rise
time
and
be
jitter-free.
The
minimum
delay
recommended
for
the
stop
pulses
is
15
ns.
The
resolution
of
any
scale
can
be
measured
with
this
setup,
and
the
main
consideration
is
that
each
stop
signal
delay
be
within
the
linear
region
of
the
selected
time
range.
The
testing
procedure
consists
of
the
following:
1.
Adjust
the
delay
for
the
stop
input
to
a
basic
setting
of
30%
to
80%
of
the
selected
time
range.
2.
Operate
the
system
and
obtain
a
timing
spectrum.
Normalize
the
output
amplitude
ful
l
range
for
the
normally
digitized
full
range
of
the
ADC
in
the
analyzer.
3.
After
you
have
accumulated
an
adequate
spectrum
to
assure
statistical
accuracy
of
photopeak
measurements
(~1000
counts
in
the
peak
channel),
identify
the
peak
channel
number
and
measure
the
FWHM
channel
number
limits.
Log
for
reference.
4.
Increase
the
delay
for
the
stop
signal
by
a
fixed
and
known
amount.
This
may
be
done
by
switching
in
a
fixed
delay
line
cable
(ORTEC
425)
or
by
careful
adjustment
of
the
Delay
unit
controls.
The
total
delay
for
the
stop
signal
must
still
be
less
than
100%
of
the
selected
time
full
range.
5.
Accumulate
a
spectrum
for
this
measurement
of
in
creased
time
intervals.
Neg.
Output
Start
Input
Neg.
Output
0.5
V
<2
n»
Tr
<1
n»
Stop
Input
(2>
ORTEC
BIASED
AMPLIFIER
Fig.
6.2.
Test
System
for
Checking
Converter
Resolution.
6.
Observe
the
relocated
photopeak
in
the
timing
spec
trum
and
record
its
peak
channel
number
and
its
FWHM
channel
number
limits.
7.
Subtract
the
peak
channel
number
in
step
3
from
the
peak
channel
number
in
step
6.
This
is
the
number
of
channels
that
represents
the
time
variation
injected
at
step
4.
8.
Using
the
formula
listed
below,
calculate
time
resolu
tion
effective
for
the
established
system
calibration:
At
per
channel
=
stop
delay
increase
channel
shift.
9.
With
the
equation
listed
below,
calculate
the
converter
resolution,
using
the
FWHM
channel
width
from
either
step
3
or
step
7.
These
widths
should
be
the
same
at
either
peak
location.
Time
resolution
(FWHM)
=
FWHM
channel
width
X
At
per
channel.
This
resolution
is
affected
adversely
by
any
jitter
that
may
be
present
in
the
discriminator
and
by
the
resolution
of
the
amplifier.
Allowances
should
be
made
for
these
contribu
tions.
Count
Rate
Tests
In
many
applications
it
is
important
for
a
time
to
pulse
height
converter
to
handle
high
count
rates,
both
external
and
internal.
Since
the
start
input
is
gated
internally
and
the
conversion
circuits
are
all
direct-coupled,
the
limit
for
its
external
count
rate
capability
is
determined
solely
by
the
input
pulse
width,
and
there
are
no
pileup
effects.
The
limit
on
the
internal
count
rate
is
imposed
by
the
conversion
and
reset
process,
where
the
start
input
is
disabled
through
a
converter
busy
interval
following
each
accepted
start
signal,
A
converter
busy
interval
is
the
measured
time
plus
approximately
7
/ts
for
start-stop
intervals
within
the
selected
time
range
or
is
the
selected
time
range
plus
4
/is
if
no
stop
signal
is
furnished
within
the
time
range.
The
fol
lowing
test,
based
on
the
system
connection
shown
in
Fig.
6.3,
permits
accumulation
of
a
basic
timing
spectrum
for
the
start-stop
input
pulses
at
60
Hz.
As
the
external
count
rate
for
start
only
is
increased
by
regulating
the
random
pulse
generator,
the
internal
pulse
rate
in
the
467
is
increased,
and
a
ratemeter
will
monitor
the
resulting
rate
at
which
the
internal
capabil
ity
is
impaired.
1.
The
photomultiplier
may
be
used
as
a
random
pulse
generator,
triggered
by
a
radiation
source.
Use
an
initial
sensitivity
setting
above
the
energy
level
for
a
zero
output
pulse
rate.
2.
Adjust
the
delay
for
the
stop
input
to
about
0.4
/ts.
3.
Select
the
0.5-/is
time
range
with
the
467.

11
RADIATION
SOURCE
PHOTO-
MULTIPLIER
and
SCINTILLATOR
(5)
FANOUT
(21
SCALER
(71
FAN
IN
1101
OR
TEC
ANALYZER
467
(9)
nonlinearity,
and
the
percent
of
deviation
is
the
difference
between
this
count
level
and
the
average
divided
by
the
average
count
level.
1.
Select
the
467
time
range
to
be
tested.
2.
Calculate
the
maximum
stop
pulse
repetition
rate
for
the
selected
time
range.
This
should
be
sl
ightly
lower
than
the
reciprocal
of
the
time
range.
For
example,
for
the
l-^is
time
range
the
reciprocal
is
1
X
10®,
and
a
pulse
generator
60
Counts/s
PULSE
GENERATOR
111
FANOUT
12)
DELAY
(3)
or
(4)
Fig.
6.3.
Test
System
for
Checking
Count
Rate.
4.
Adjust
the
system
for
a
timing
spectrum,
accumulated
for
the
60-Hz
input
pulses.
5.
Decrease
the
threshold
of
the
discriminator
(2)
to
generate
random
start
signals
with
no
corresponding
stop
signals.
Monitor
the
random
rate
with
the
ratemeter.
6.
Observe
the
timing
spectrum
as
the
random
input
rate
is
gradually
increased.
Watch
for
interference
in
the
accumu
lated
spectrum.
Differential
Linearity
Measurements
A
system
for
testing
differential
linearity
of
the
467
is
shown
in
the
block
diagram
in
Fig.
6.4.
In
this
system
the
random
pulse
generator
is
used
as
the
source
for
start
signals,
and
a
pulse
generator
with
a
fixed
rate
is
used
for
stop
signals.
The
measurable
time
interval
between
a
start
and
stop
is
a
random
value,
with
equal
probabi
lity
that
it
wil
l
be
any
time
difference
up
to
the
periods
between
the
regular
stop
signals.
For
an
infinite
number
of
TPHC
outputs
the
count
levels
for
each
channel
of
the
analyzer
should
be
equal
After
the
test
has
been
run
long
enough
to
assure
statistical
accuracy
(e.g.,
>25,000
counts/channel),
the
spectrum
should
be
similar
to
those
illustrated
in
Fig.
6.5.
Any
deviation
from
a
straight
line
represents
a
differential
RADIATION
SOURCE
PHOTO-
MULTIPLIER
and
SCINTILLATOR
(5)
DISCRIMI
NATOR
PULSE
GENERATOR
(1)
ORTEC
467
ANALYZER
(91
Stop
'v
1
1
1
1
1
1
'
'
1
1
1
1
1
1
'
1
100
ns
Range
=
.1
Multiplrer
=
XI
(Vert):
Full
Scale
5
x
10"
counts
(Horiz):
Full
Scale
=
105%
range
-
-
■
-
-
lol
1
MS
Range
=
.1
Multiplier
=
X1C
(Vert):
Full
Scale
=
5
x
10"
counts
(Horiz)
:
Full
Scale
=
105%
range
Range
=
.1
Multiplier
=
X100
(Vert):
Full
Scale
=
(Horiz):
Full
Scale
[
10
ms
5x10"
counts
=
105%
range
Fig.
6.4.
Test
System
for
Checking
Differential
Linearity.
Fig.
6.5.
Differential
Linearity
for
the
Indicated
Ranges.

12
rate
of
4
to
5
times
10^
should
be
satisfactory.
A
lower
rate
increases
the
time
required
to
run
the
test,
while
a
faster
rate
wil
l
reduce
the
upper
channel
response
because
of
analyzer
dead
time.
3.
Operate
the
system
and
monitor
the
dead-time
meter
on
the
analyzer.
Regulate
the
random
start
rate
to
cause
the
analyzer
dead
time
to
be
approximately
10%.
7.
Vary
the
strobe
delay
and
observe
that
there
is
no
change
in
the
TPHC
output
amplitude
but
that
the
output
delay
fol
lows
the
strobe
delay.
Normally
when
the
467
is
in
external
strobe
mode,
it
is
internally
reset
at
either
5
or
120
MS,
selected
with
the
rear
panel
Strobe
Sync
switch.
Testing
the
Single
Channel
Analyzer
The
single
channel
analyzer
can
be
tested
by
the
following
procedure:
4.
Clear
the
analyzer
to
zero
and
operate
the
system
unti
l
the
average
count
level
stored
in
each
channel
is
sufficient
to
ensure
statistical
accuracy.
5.
Compare
any
nonlinearity
indications
to
the
specifica
tions
listed
in
Section
2.
Some
nonlinearity
can
be
expected
in
channels
in
the
lower
5%
of
the
analyzer
range
as
shown
in
Fig.
6.5
because
of
the
stop
pulse
width
and
the
TPHC
gating
time.
Checking
External
Strobing
Mode
The
system
for
check
ing
the
external
strobing
mode
is
shown
in
Fig.
6.6.
This
system
can
be
used
to
verify
the
principles
of
operation
of
the
467.
1.
Set
the
delay
for
the
stop
signal
at
about
0.4
/is.
2.
Set
the
467
time
range
for
0.5
fis.
3.
Use
the
internal
strobe
mode
for
the
467.
Adjust
the
osci
lloscope
sweep
as
required
to
identify
the
TPHC
output
pulses.
4.
Adjust
the
delay
for
the
strobe
signal
greater
than
0.5
MS
to
ensure
that
it
wil
l
occur
later
than
the
full
time
range.
5.
Switch
the
467
to
its
external
strobe
mode
and
observe
the
TPHC
output
pulse.
It
should
be
identical
to
the
pulse
observed
in
step
3
except
for
the
time
at
which
it
occurs.
6.
Vary
the
stop
delay
setting
and
observe
the
effect
on
the
TPHC
output
pulse.
It
should
occur
at
a
constant
time
but
vary
in
amplitude.
If
the
stop
delay
is
increased
beyond
0.5
MS,
there
wil
l
be
no
output
pulse.
FAST
DISCRIMI-
Start
TPHC
OSCILLO
SCOPE
(8)
PULSE
True
Start
GENERATOR
Stop
467
(1) (2)
"/
1
Strobe
TRIGGER
NANOSECOND
DELAY
MICROSECOND
DELAY
2002773
(3)
or
(4)
(3)
Fig.
6.6.
Test
System
for
Checking
External
Strobing
Mode.
1.
Set
the
ULD
control
ful
ly
clockwise
at
10.00
dial
divisions
and
the
LLD
control
ful
ly
counterclockwise
at
0.
2.
Set
the
SCA
Mode
switch
at
Normal.
3.
Use
the
circuit
shown
in
Fig.
6.1.
Select
a
convenient
time
range
and
adjust
the
stop
delay
to
obtain
an
output
of
about
5
V
through
the
TPHC
Output
connector.
Note
that
an
SCA
output
is
present.
4.
Increase
the
LLD
control
setting
and
note
that
the
SCA
output
disappears
as
the
LLD
dial
moves
through
about
5.00
dial
divisions
(considering
the
accuracy
of
the
TPHC
output
amplitude
adjustment).
Decrease
the
LLD
setting
to
about
4.00
so
that
the
SCA
output
reappears.
5.
Decrease
the
ULD
setting
and
note
that
the
SCA
output
pulse
disappears
as
the
dial
moves
down
through
about
5.00.
6.
Move
the
SCA
Inhibit
switch
setting
to
In
and
note
that
the
TPHC
output
pulse
is
present
only
when
the
SCA
output
pulse
is
also
present.
Return
the
SCA
Inhibit
switch
to
Out.
7.
Change
the
SCA
Mode
switch
from
Normal
to
Window.
Set
the
LLD
dial
at
4.0
and
the
ULD
dial
at
2.0.
Change
the
stop
delay
to
change
the
TPHC
output
pulse
amplitude
from
3
V
to
7
V.
Note
that
the
SCA
output
pulse
is
present
only
when
the
TPHC
output
amplitude
is
within
the
range
of
4
to
6
V.
6.2.
CORRECTIVE
MAINTENANCE
Clean
the
surfaces
of
the
printed
circuits,
the
connectors,
and
all
chassis
parts
periodically
to
prevent
accumulated
dust
from
forming
leakage
paths
between
the
circuit
components.
The
467
should
require
maintenance
only
rarely,
as
indicated
by
a
malfunctioning
in
its
performance
under
known
conditions.
If
the
instrument
is
suspected
of
malfunctioning,
use
the
performance
tests
of
Section
6.1
above
to
aid
verification.
When
incorrect
operation
is
identified,
disconnect
the
467
from
its
position
in
the
system
and
perform
routine
diagnostic
tests
with
a
pulse
generator
and
oscilloscope.
Use
the
timing
diagrams
in
Figs.
6.7—6.10
to
isolate
the
problem,
and
use
schematic
diagrams
467-0201-SI
and
467-0301-SI
at
the
end
of
this
manual
to
localize
the
malfunctioning.

13
GATE
INPUT
(ANTI)
START
INPUT
STOP
INPUT
TRUE
START
«
CONVERSION
CAPACITOR
OVERRANGE
CONV.
BUSY
OUTPUT
T
range-
No
Output
Fig.
6.7.
Using
Internal
Strobe
with
Stop
After
End
of
Conversion.
Calibration
Three
critical
cal
ibrations
that
may
need
to
be
made
are
described
below:
Discriminator
Sensitivity
The
discriminators
for
the
start
input
and
stop
input
are
calibrated
to
trigger
at
a
level
of
approximately
—250
mV.
Control
R148
is
adjusted
for
the
start
input,
and
control
R49
is
for
the
stop
input.
A
response
cannot
be
triggered
for
the
stop
discriminator
unless
the
start
discriminator
has
been
triggered
at
least
12
ns,
but
less
than
the
selected
time
range,
before
the
stop
signal.
1.
Set
up
the
pulse
generator
(1)
for
negative
250
mV,
approximately
100
ns
wide.
Connect
this
output
to
the
start
input.
2.
Monitor
the
true-start
output
and
adjust
the
start
sensitivity
trim
potentiometer
RMS
until
a
true
start
is
barely
triggered.
3.
Set
fanout
2
or
3
(Fig.
6.1)
to
trigger
start;
delay
the
250-mV
signal
pulse
and
insert
it
into
the
stop
input
of
the
467.
GATE
INPUT
lANTII
START
INPUT
STOP
INPUT
LT
TRUE
START
CONVERSION
CAPACITOR
OUTPUT
AMPLIFIERS
POS.
OUTPUT
OVERRANGE
CONV.
BUSY
OUTPUT
^
I
/
T
delay
*-
T
range
-
4ms
Fig.
6.8.
Using
Internal
Strobe
with
Stop
Timed
Normally.
DC
Level
and
Overrange
To
calibrate
the
dc
level
and
overrange,
set
up
the
circuit
as
shown
in
Fig.
6.1.
1.
Put
the
front
panel
gate
Anti
Coinc/Coinc
switch
to
Coinc
so
that
the
467
wi
ll
not
accept
pulses.
2.
Monitor
TP3
with
a
sensitive
VTVM
and
adjust
dc
zero
trim
potentiometer
R133
so
that
the
dc
voltage
at
TPS
is
zero.
4.
Monitor
TPS
(0200
printed
circuit
board)
with
the
oscilloscope
and
adjust
the
stop
sensitivity
R49
so
that
the
output
barely
triggers.
3.
Return
the
Gate
switch
to
Anti
Coinc.
Monitor
TPS
with
a
calibrated
osci
lloscope
and
adjust
stop
delay
to
obtain
a
10.1
-V
pulse
at
TPS.

14
GATE
INPUT
lANTII
START
INPUT
STOP
INPUT
TRUE
START
OUTPUT
STROBE
INPUT
(EXT.
DELAYED
TRUE
START)
CONVERSION
CAPACITOR
OVERRANGE
CONV.
BUSY
OUTPUT
T
-
-T
range
GATE
INPUT
(ANTI)
START
INPUT
STOP
INPUT
TRUE
START
OUTPUT
STROBE
INPUT
(EXT.
DELAYED
TRUE
START)
CONVERSION
CAPACITOR
OVERRANGE
CONV
BUSY
OUTPUT
»
—T
strobe
T
range
<
b
or
120
MS
Fig.
6.9.
Using
External
Strobe
with
Output
Timed
IMormally.
Fig.
6.10.
Using
External
Strobe
with
Strobe
Signal
Arriving
Before
Conversion
Is
Complete.
4.
Adjust
the
overrange
trim
potentiometer
(R99)
so
that
a
signal
is
barely
triggered.
This
sets
the
overrange
threshold
to
10.1
V.
Pulse
Width
Adjustment
Use
the
same
signal
and
circuit
as
in
above.
1.
Monitor
the
467
output
and
adjust
the
stop
signal
delay
to
obtain
an
output
pulse
amplitude
of
5
to
8
V.
2.
Adjust
the
pulse
gate
width
trim
potentiometer
RB6
to
obtain
the
desired
pulse
width.
This
is
factory-set
at
approximately
2
fis.
6.3.
TROUBLESHOOTING
Use
the
tests
in
Section
6.1
and
the
circuit
analysis
in
Section
5,
and
refer
to
schematics
467-020TS1
and
467-0301-S1
and
functional
block
diagram
467-0101-B1
to
locate
the
faulty
part
or
parts.
A
table
of
typical
voltages
is
furnished
at
the
end
of
this
section
for
further
identifica
tion
of
faulty
circuit
components.
Dynamic
Signal
To
check
signals
under
dynamic
condi
tion,
set
up
as
for
"Resolution
Tests"
using
Fig.
6.1.
There
are
several
tests
for
quickly
finding
trouble
locations
with
an
osci
lloscope.
1.
Observe
input
start
signal
to
be
>—0.3
V
on
50S7.
-10.7
V
2.
Check
col
lector
04
for
-13
V
—1
I—
+
4
V
3.
Check
true-start
output
4.
Check
that
the
input
stop
pulse
is
>—0.3
V
on
5012
and
follows
the
start
pulse
within
the
time
range
selected.
-3
V
+
10
V
5.
Check
signal
at
TP1
(0200
board)
to
be
r>1
us
<6
US.
6.
Check
signal
at
TP2
to
be
7.
Check
pulse
at
TP3
to
be
q.
-10
V
^
h-
-,Tv_r
~L.
T
h-

15
6.4.
TYPICAL
DC
VOLTAGES
Table
6.1
gives
the
typical
do
voltages
that
were
measured
using
the
following
control
settings:
Range,
0.2;
Multiplier,
10;
SCA
Inhibit,
Out;
Logic
Input
switch,
Anti
Coinc;
Strobe
Sync,
Int
Stop;
Stop
Inhibit
Mode,
Out.
6.5.
MODIFICATIONS
The
467
time
range
can
be
extended
from
its
standard
l
imit
of
80
/US
so
that
it
wi
l l
measure
time
with
a
maximum
range
of
800
(US
full
scale.
The
467-0200
printed
circuit
board
includes
al
l
the
required
components,
and
they
can
be
used
by
installing
jumpers
at
the
appropriate
points
in
the
printed
circuit
board.
The
jumpers
that
can
be
added
are
Indicated
in
schematic
467-0201-SI
with
the
flagged
reference
to
note
6.
A
total
of
four
jumpers
are
involved,
and
the
resulting
Range
Multiplier
switch
selections
wil
l
be
X10,
X100,
and
X1000,
the
XI000
selection
uses
the
switch
setting
that
is
originally
assigned
to
the
XI
setting.
A
slight
loss
of
signal
amplitude
may
be
observed
due
to
stretcher
droop.
6.6.
FACTORY
REPAIR
This
instrument
can
be
returned
to
the
ORTEC
factory
for
service
and
repair
at
a
nominal
cost.
Our
standard
proce
dure
for
repair
ensures
the
same
quality
control
and
checkout
that
are
used
for
a
new
instrument.
Always
contact
the
Customer
Service
Department
at
ORTEC,
(615)
482-4411,
before
sending
in
an
instrument
for
repair
to
obtain
shipping
instructions
and
so
that
the
required
Return
Authorization
Number
can
be
assigned
to
the
unit.
Write
this
number
on
the
address
label
and
on
the
package
to
ensure
prompt
attention
when
it
reaches
the
ORTEC
factory.
Table
6.1.
Typical
dc
Voltages.
Checkpoint
Voltage
(V)
(or
IC
State)
Checkpoint
VOLTAGE
(V)
(or
IC
STATE)
Checkpoint
Voltage
(V)
(or
IC
State)
Checkpoint
Voltage
(V)
(or
IC
State)
Q2E
-0,75
Q23B
+0.73
Q45E
-12
IC1(12)
L
B
-0,011
C
+0.015
C
0
IC1(13)
H
QBE
-0.68
Q24E
-11.32
Q46B
-11.37
IC2(1)
L
C
+
1
.1
C
0
Q47B
-11.38
IC2(8)
L
Q4E
+0.5
Q25B
-0.68
Q48E
+
12.00
IC2(13)
H
B
-0.05
C
-0.02
B
+
11.36
IC3(1)
+
1.56
C
-13.12
Q26C
0
C
-11.38
IC3(2)
+
1.70
Q5B
-0.23
Q27B
0
Q49B
+
12.5
IC3(3)
H
C
-0.61
C
-5.73
C
0
IC3(5)
H
Q6C
+3,67
Q28B
-0.17
Q50B
0
IC3(6)
L
Q7E
+0.73
Q29B
-0.9
C
+5.23
IC3(9)
H
B
-2
75
C
-0.17
Q51E
-12.04
IC3(10)
L
QBE
-7.3
Q31B
+0.67
B
-11,4
IC3(14)
+
1.36
B
-6
62
C
-12.6
C
-1.97
IC3(15)
+
1.57
C
+0.69
Q32B
+0.01
Q52B
-11
.92
IC4(7)
H
QlOE
+0
02
C
0
C
0
104(14)
H
B
+0.72
Q33E
0
Q53E
+2.58
107(1)
L
C
+
1
36
C
+
12
B
+3.24
107(2)
L
QllE
-12.7
Q34B
+0.74
Q54B
0
107(3)
H
C
-7,32
C
0
C
+8.00
107(5)
H
Q12C
-0.23
Q35E
-12.64
Q55B
0
107(7)
Ground
Q13B
+0.72
B
-12.6
Q56E
0
107(8)
H
C
+0.14
a36B
-12
B
+0.23
107(10)
L
Q14E
+
1.15
C
-3,31
C
+
12
107(13)
H
B
+
1
.85
Q37E
0
Q57E
0
107(14)
Vcc
C
+4.3
B
+0.60
B
+0.16
108(1)
Open
Q15B
+0.097
Q38E
-6.8
C
+
12
108(2)
H
C
+
7,5
B
-6.1
Q58E
0
108(3)
L
Q16B
0
Q39E
+
19.8
B
+0.19
108(4)
Ground
C
+0.32
Q40E
-0.7
C
+
12
108(5)
L
Q17C
+2.2
B
-0.08
Q59E
-0.40
108(6)
L
Q18E
-0.73
C
+5.14
B
+0,21
108(7)
H
B
0
Q40B
-0.08
Q60E
0
108(8)
Ground
Q19E
-0.68
C
+
19.2
B
+0.15
108(9)
H
(Open)
C
+
1
.14
0418
+
11.9
C
+
12
108(11)
Vcc
Q20E
+0.47
B
+
11.2
D18
Anode
-21.2
108(12)
L
B
+0.02
Q42E
+7,04
D25
Cstliode
+5.63
108(14)
+
1.5
C
+
1.14
Q44A-S
+0.56
IC1(5)
H
108(15)
+1,5
0218
-0.233
G
+0.015
IC1(6)
L
108(16)
Vcc
C
-5.31
D
+6.42
IC1(8)
L
109(1)
H
Q22E
-4.93
Q44B-S
+0.541
IC1(9)
H
109(6)
L
B
-5.31
G
+0.02
ICl(lO)
L
109(7)
Ground
C
-24
D
+6.41
ICl(ll)
H
109(12)
109(14)
L
Vcc
Table of contents
Other ORTEC Media Converter manuals