OrthAlign HipAlign User manual

HipAlign
®
Surgical Technique Manual
Total HipArthroplasty
PosteriorApproach

2
Table of Contents
General Operation …………………………………………………………………….. 3
Tips and Warnings ……………………………………………………………………. 3
Pre-Op …………………………………………………………………………………. 4
Patient Positioning …………………………………………………………….. 4
Opening and Powering On the Navigation Unit ……………………………... 4
Back-Table Set Up ……………………………………………………………………. 5
Calibration Steps ………………………………………………………………. 6
Prepare Instruments …………………………………………………………... 7
Attach Laser ……………………………………………………………………. 7
Pre-Op Selections ………………………………………………………………. 8
Securing Base to Bone …...…………………………………..……………………….. 9
Securing Pelvic Base………….………………………………………………... 9
Attaching Sensors and Registration of Landmarks ………………………………... 10
Attaching Sensors ……………………………………………………………... 10
Aligning Jig ……………………….……………………………………………. 11
Adjusting the Registration Jig …...……………………………………………. 11
How to Register Landmarks …………………………………………………. 11
Registrations ………………………………………………………………………….. 12
Options Menu …………………………………………………………………………. 13
Prepare Impactor and Shell …………………………………………………………. 13
Set / Check Cup Angle ……………………………………………………………….. 14
Check Leg Length ……………………………………………………………………. 15
Verify Homepoint …………………………………………………………………….. 15
Repeat Registrations…….…………………………………………………………….. 16
End Procedure……..…….…………………………………………………………….. 16
Shut Down and Disposal …………………………………………………………….... 16
Compilation of Instruments with Part Numbers …………………………………..... 17
Instrument Tray Configuration for Sterilization .………………………………….... 18
Troubleshooting ……………………………………………………………………….. 19-20

3
GeneralOperation
Types of Beeps:
Short beep: Acknowledgment that button was pressed.
Long beep: Input was not accepted, or there was a fault.
Double beep: Unit completed registration or calibration step successfully.
Laser aperture may be open or
closed with the shutter
Left
Returns user to previous screen
Center
Turns unit on and progresses user
to next screen
Warnings
NAVIGATION UNIT / REFERENCE SENSOR:
Up and Down
Toggles between different options or
adjusts numerical input
NOTE: Holding Left and Up buttons
simultaneously powers unit off
Risk of communication errors increases if multiple OrthAlign Plus® systems are used in close proximity.
If navigation unit is dropped on floor, it must be discarded. If reference sensor is dropped on floor, it must be returned
to manufacturer for verification of function and calibration.
Probe scale and lens of reference sensor must be kept clean for reliable operation. Note that lens of reference sensor is
on its underside.
Avoid impacting jig and electronics. Verify position of jig and electronics after impacting and before checking angles.
LASERMODULE/OTHER INSTRUMENTS:
Inspect laser module prior to use. Do not use laser module if it appears damaged or moisture is
present under the lens.
If laser module fails to function after initial registration, abandon leg length and offset measurement
feature.
The laser module is a Class 2 laser product, maximum output: <1mW, emitted wavelength: 635nm,
duration > 0.25 seconds (continuous), beam divergence: 10 degrees, per IEC 60825-1: 2014-
05. This information can be found etched onto the back surface of the laser module.
Do not stare into beam. Eye damage could result.
Take care when using laser to avoid possible dazzle, flash-blindedness or after-image effects.
Use of controls of adjustments or performance of procedures other than those specified herein may result in hazardous
radiationexposure.
Use the movable shutter on the front of the laser module to expose the lens for use, and cover the lens to stop projection
of the laser beam.
Once the battery is installed, the laser is always on and the shutter must be closed to stop projection of the laser beam.
The strong magnetic fields near a neodymium magnet can affect pacemakers, ICDs and other implanted medical devices.
Keep the laser module, vertical laser target, adapter socket driver and any other instruments with magnets at least six inches
away from any such heart device.
SURGICALTECHNIQUE:
Utilize standard surgical practice in joint reduction (for joint stabilityand soft tissue laxity).
Take care to avoid damage to nervous and vascular structures when using registration probe to register landmarks or
when mounting jig.
Do not impact or hammer system components except where components are designated as impactor devices.
Acetabular shell impactor may be impacted.
Abort the use of the navigation feature if the landmarks cannot be identified or registered.
Remove all pins, screws, and other instruments prior to closing wound. Do not implant any system components.

4
Pre-Op
PatientPositioning
VIEW FROM SIDE VIEW FROM ABOVE
Parallel to table edge
Table
Patient must be placed in the lateral decubitus position.
Table edge
When positioning patient prior to surgery, take care to align the anterior pelvic landmarks (both ASIS’s and
pubic tubercle) in a vertical plane parallel to the long edge of the operating table. Ensure that pelvis is
securely held byan appropriate positioning device such as a peg board or vise-type patient positioner.
WARNING: Correct patient positioning is essential for accurate navigation. The system calculates cup angles
and changes in leg length and joint offset
based on the assumption that the pelvis is accurately positioned
during table registration.
Opening and Powering on the Navigation Unit
1.
Open navigation unit
package.
2.
Pass sterile inner blister
package into operative field.
3.
Peel back cover to
expose navigation unit
and sensor battery.
4.
Remove navigation unit
and sensor and laser
batteries from
blister
pack.
5.
Press center button to
power on navigation unit.

5
Back-Table Set Up
Select Joint
Use center button to select Hip.
SelectApproach
Use center button to select approach.
STEP 1: Install Sensor Battery
Insert battery into reference sensor and close cover.
TIP: Reference sensor LED will slowly flash yellow to indicate
battery has been inserted correctly. LED will momentarily flash
green and navigation unit will move to next screen when reference
sensor is found.
STEP 2: ConfirmSensor ID
Press center button if displayed number matches the one marked on reference sensor being
used. Otherwise, press left button to let navigation unit find another reference sensor. Repeat
until correct reference sensor serial number is displayed.
STEP 3: Attach to Pelvic Jig
Slide the navigation unit onto jig toward lever. Navigation unit should be firmly secured in
place on the jig. Attach reference sensor to coupler at other end of the pelvic jig as shown.
NOTE: Navigation unit must be secure on the pelvic jig before proceeding.
S1
Select Joint
Knee
Hip
S2
SelectApproach
Posterior
Anterior
1
Install Sensor Battery
Seeking Sensor…
2
Confirm Sensor ID
Confirm
####
Use Count
22
No
Yes
3
Attach to Pelvic Jig
<
<
<
<

6
6
Place on left Side
Calibration Steps
TIPS FOR CALIBRATION STEPS:
Navigation unit must be maintained in steady position, or calibration will not be accepted. A red hand will flash
on screen if unit is not steady.
Level circle must be green, but ball does not need to be in exact center of circle.
Do not move to next position too quickly. Wait for double beep, then place in next position as shown on screen.
Displayed numbers are the percent errors from the nominal jig angles on each axis, and should be 4.0 or less.
If navigation unit does not proceed to next screen after several seconds, check that navigation unit and reference
sensor are correctly attached to the bracket and press left button three times to repeat from first calibration step
(Step 4).
STEP 4:
Angle unit away from you so that pelvic jig rests on table
with navigation unit screen facing up.
STEP 5:
Angle unit toward you so that reference sensor rests on
table with navigation unit screen facing you.
STEP 6:
Rest navigation unit on left side on table.
STEP 7:
Angle unit away from you so that pelvic jig rests on table
with navigation unit screen facing up.
STEP 8:
Attach reference sensor to coupler on probe. Verify that
numbers are displayed on the screen bysliding the probe to
differentpositions.
4
Angle Forward
Cal 1
Cal 2
Cal 3
5
Angle Back
✔
Cal 1
Cal 2
Cal 3
8
Attach Sensor
Verify probe values are
displayed
102

7
Prepare Instruments
Remove required instruments from tray, and prepare them for use.
Select appropriate shell adapter. See “Shell
Adapter Options” table below.
Note: The Probe Calibration Jig is not required for this procedure.
WARNING: Only shells with an apex hole matching one of the sizes listed can be attached to impactor.
Hex Driver - 3.5MM, Teardrop
403105
Probe Calibration Jig –Tripod
403258
Laser Module
403370
Registration Probe
403305
Pelvic Base, Anterior
403306
Pelvic Bracket Assembly, Short
403307
Straight Cup Impactor
403338
Reference Sensor 5 Hip (RS5H)
405001 or 403087-06
Threaded Fixation Pin, 4.0mm x 110mm
with 3.2mm Proximal End, 403317-110
Femur Registration Marker,
3.5mm Hex Head, 403402
Screw, 4.5mm dia x 20mm,
3.5mm Hex Head, 403108
Thigh Plate, Slotted,
403376
Vertical Target, Magnetic,
403377
Offset Cup Impactor (Optional, to be used
with Adapter Socket Driver)
403309
Adapter Socket Driver (Optional)
403347
Shell Adapter
Options
7/16-20
Depuy, Signature
403339
3/8-24
Biomet, Stryker
403342
3/8-16
Smith & Nephew
403341
5/16-24
Zimmer, Medacta
403343
1/4-28
DJO
403340
3/8-32
Microport
403344

8
Attach Laser to Pelvic Base
Step 9: Attach Laser
Close shutter on the laser module.
Insert battery and close cover.
Attach laser to the posterior side of the bracket at a selected
angle. The laser can be angled downward to project more
proximal or angled straight to project more distally.
Ensure the laser module is securely
attached and does not
rotate or pivot
on the bracket.
NOTE: The laser remains on as long as the battery is installed. The laser
beam
is shown/hidden through the movement of a shutter.
WARNING: Do not shine laser beam into eyes.
Pre-Op Selections
Step 10: Select Hip
Select operative hip.
Step 12: Input Abduction
Input target cup abduction angle. This is the Radiographic Inclination angle defined as the
coronal plane projection of the angle between the acetabular axis and the longitudinal axis
of the body.
Step 13: Input Anteversion
Input target cup anteversion angle. This is the Radiographic Anteversion angle defined as
the angle between the acetabular axis and the coronal plane.
Left
10
Select Hip
L
Right
12
Input Abduction
Input target cup
Abduction angle
(Radiographic Inclination)
L
40
Input:
13
Input Anteversion
Input target cup
Anteversionangle
(Radiographic Anteversion)
L
15
Input:
9
Attach Laser
Insertbattery
Screwonto
bracket
<
<
Closed Shutter
Angled Straight
Angled Down

9
Securing Base
Securing Pelvic Base
Placement of the fixation pins in the iliac crest should be vertical to optimize orientation of
theinstrumentation.
For the first pin, make a stab incision
over the ipsilateral iliac crest, directly
superior to the greater trochanter.
Insert a 4.0mm threaded
fixation pin
vertically using a wire
driver.
Make a second stab incision, and insert
the second fixation pin in the iliac crest
2cm anterior parallel to the first
pin. You may use the pelvic base as a
guide for spacing between the pins.
Slide the Pelvic Base over both pins to
the level of the skin, and use the 3.5mm
Hex Driver to tighten the two bottom
screws to secure Base.
WARNING:
The base must be attached prior to dislocating the hip to ensure correct patient positioning and navigation accuracy.
Take care to avoid plunging through the medial wall of the iliac wing.
Verify that pelvic base is securely mounted to the bone.
Take care to minimize divergence of pins, as excessive bending can damage pins.
Ensure pins are aligned in the tracks within the pelvic base and the base is clamped on the cylindrical part of the pins.
H1
Secure Base on Bone
Pelvic
Base

10
Attaching Sensors and Registration of Landmarks
Attaching Sensors
Attach pelvic bracket and probe to pelvic base.
TIP: Attach reference sensor to
probe by aligning arrows on sensor
and probe.
TIP: Attach pelvic bracket, with
navigation unit attached, to coupler
on pelvic base while pushing on
pelvic bracket lever.
TIP: Attach probe assembly to pelvic
base by aligning arrows on probe
and base.
H2
Attach Sensors
Attach probe to jig
Attach jig to
pelvic base

11
Aligning Jig
Place the leg on the operative side in a neutral
abduction, flexion and rotation using a roll if necessary.
Place the thigh plate on the lateral distal thigh and secure using
Coban™ by wrapping around thigh on either side of plate.
Mark the boundary of the plate on the skin.
Place the vertical target on the plate
Open shutter on laser.
Align pelvic bracket so the laser beam projects crosshairs
on the vertical target mounted on the distal thigh and
remove vertical target.
NOTE: Ensure laser lens is clear. Close laser shutter when not in use.
WARNING:
Discontinue use of laser if crosshair pattern is obscured.
Discontinue use if unable to verify thigh plate is aligned with original
marking on skin or if unable to return it to original position.
Adjusting the Registration Jig
TIP: To adjust the laser alignment on the leg, loosen pelvic base locking screw, and adjust
angle of pelvic bracket.
Locking the Registration Jig
Use the 3.5mm hex driver to lock screw tightly to secure
jig in this position for remainder of procedure.
Close the shutter on the laser module.
WARNING: Ensure all connections and locking mechanisms are secure before proceeding with registrations.
How to Register Landmarks
INCORRECT:
TIP: Hold probe close to the tip
to maximize accuracy of point. TIP: When pressing buttons, support back of navigation unit to avoid flexing jig. If
registration is not accepted, keep probe tip stationary and gently squeeze center button.
H2A
Align Jig
Aim Laser at
Distal Thigh
CORRECT:
H2B
Lock Base
Tightenscrew

12
Registrations
Register Home
Place tip of probe into hole of the pelvic base and press center button.
WARNING:
Register Table
Position operating table horizontally.
Verify the sagittal plane of pelvis is level.
Move probe, with reference sensor attached, to align parallel long axis of body
with the tip pointing towards the feet.
Press the center button to register.
Correct patient positioning is essential for accurate navigation. The system calculates cup angles based on the assumption
that the pelvis is accurately positioned during table registration.
Table registration must be completed prior to dislocating hip, as torque applied during dislocation can move pelvis away
from the correct initial alignment.
Pelvic Base cannot be moved after this point.
Pelvic Bracket angle to the base cannot be adjusted after this point.
If fixation pins become loose, navigation must be aborted, since table registration is no longer valid.
Registering Femur
Make the standard incision and identify a point along greater
trochanter to insert a registration marker for femoral
registration. Insert femur registration marker.
Open laser shutter, place the vertical target on the thigh plate and
align the leg so that the laser projects onto
the lower part of the
target.
Trace the crosshair pattern with a marking pen.
While laser projection is aligned with laser marking, hold the
probe tip in the center hole of the marker and press the
center
button to register the femur.
Close shutter on the laser and remove vertical laser target.
TIPS:
1.
When marking the vertical target and registering the femur
registration marker, position
the thigh in neutral flexion,
abduction and rotation.
2.
Stack legs vertically atop each other in a stable position which
will be easy to return to prior to registration.
WARNING:
If laser module moves, leg length must be aborted since leg alignment cannot be replicated.
Discontinue use if unable to verify thigh plate is aligned with original marking on skin or if unable to return it to
original position.
Prepare Acetabulum
NOTE: Once registration is complete, proceed to dislocate the hip, resect the femoral head, and prepare the
acetabulum as per the implant manufacturer’s technique.
H4
H3
Register Home
Place probe tip in hole
of pelvic base
Keep pelvic base locked
throughout procedure
Register Table
Verify sagittal plane of pelvis is level and
align probe with coronal plane

13
Options Menu
Set/check the angle of the cup by
navigatingimpactor.
NOTE: Title changes to “Check Cup
Angle” after cup angle has been
“Set”.
Measure change in leg length and offset
from its pre-operative state.
M1
Selection Option
Set Cup Angle
Check Leg Length
Verify Homepoint
Repeat Registrations
End Procedure
Re-register Homepoint. The change in position
of probe tip between original and current
position is displayed.
Repeat pelvic landmark registrations.
Use to select other hip for a bi-lateral procedure.
The unit can also be powered off at this point by
pressing and holding left and up arrow buttons
until unit powers off.
Prepare Impactor and Shell
Select the appropriate adapter as listed on page 7 of this surgical technique.
Straight Impactor:
Snap the adapter into the impactor.
Thread the adapter into the acetabular shell.
The release lever can be used to release the adapter from the impactor.
Offset Impactor:
Snap the adapter into the impactor.
Thread the adapter into the acetabular shell.
Pull on the finger grip shown to release the adapter from the
impactor.
Following full seating of the shell, remove the impactor from the
implant adapter.
Use the socket driver to remove the adapter from the implant.
NOTE:
Align the screw holes of shell as desired; Adapter can be assembled in multiple angles relative to the impactor.
Vertical orientation of the Reference Sensor post is not required for accuracy; it is recommended for ease of use when
transferring the Reference Sensor to the impactor for navigation.
If none of available adapters fits on selected implant, impactor cannot be navigated.
WARNING: Prior to impacting, verify shell is fully seated on shoulder of impactor and that threads do not protrude beyond
outer face of shell.
Release Lever
Pull to release adapter

14
C3
Cup Angle
Abduction Anteversion
40⁰
15⁰
Navigation ended
Select Check Cup Angle
from menu, if desired
L
Set / Check CupAngle
Step C1: Dock Sensor
Remove probe assembly from bracket.
Remove reference sensor from probe.
Dock reference sensor on the jig by attaching it to the coupler on the jig (as shown on the display).
Step C2: Set Cup Angle
Hold hip stable until green light
appears.
There are 30 seconds of live navigation
Remove reference sensor from jig by
pushing lever and attach to impactor.
If sensor is moved too quickly, an error
message will appear and Step C1 must
be repeated.
NOTE: The blue bar on right indicates remaining time.
Align impactor at desired cup angle. Aligning crosshairs in center of bull’s eye will set cup
angle at inclination and anteversion angles entered earlier.
Angles displayed are calculated according to their radiographic definitions.
NOTE: The navigation unit proceeds to the next step if:
-
30 second timer expires
-
center button is pressed
-
reference sensor detects an impact strong enough to disrupt navigation
Step C3: Cup Angle
Inclination and anteversion cup angles are now displayed statically.
NOTE:
Once reference sensor is attached to impactor, navigation unit displays abduction and anteversion angles based on the
Table reference frame.
The OrthAlign Plus® System does not navigate or control shell impaction depth or insertion point. Please refer to implant
manufacturer’s technique guide for guidance.
WARNING:
Check cup angle after impacting as cup angle will change during impaction. Not checking cup angle will reduce accuracy.
Verify position of reference sensor and navigation unit after impacting before checking angles.
Visually confirm shell angles before proceeding.
C1
Dock Sensor
Attach sensor
to jig
C2
Set Cup Angle
Abduction Anteversion
36⁰ 10⁰
C2
Set Cup Angle
Abduction Anteversion
40⁰ 15⁰

15
Check Leg Length
Step F5: Attach Sensors
Attach probe with reference sensor to jig
Step F5A: Position Femur
Position the magnetic vertical target on the distal thigh
Open shutter on laser
Re-align leg so that the laser projection aligns with the marking on the vertical target.
NOTE: Ensure laser lens is clear.
WARNING:
Discontinue use of laser if crosshair pattern is obscured.
Discontinue use if unable to verify thigh plate is aligned with original marking on skin
or if unable to return it to original position.
Step F6: Register Femur
Hold the probe tip in the center hole of the screw and press the center button to register the femur.
Remove vertical laser target.
Step F9: Register Femur
Leg Length and offset are displayed.
Close the shutter on the laser.
Verify Homepoint
Step V1: Attach Sensors
Attach probe and probe mount to bracket. Attach bracket to pelvic
base.
Step V2: Verify Homepoint
Place tip of probe into hole of Pelvic Base.
Press center button to register Homepoint.
Step V3: Homepoint Accuracy
View change in Homepoint which is change in probe tip position between
original Homepoint and current Homepoint registrations.
NOTE: The number displayed may not be zero due to mechanical play and reference sensor noise. If displayed number is
greater than 3mm, it is suggested to verify all reference sensor attachments and jig connections are correct and secure
before repeating. Repeat registrations if accuracy or previous registrations is in question.
V1 Attach Sensors
Attach probe to
probe mount
Attach probe mount
to jig
Attach jig to pelvic base
V2
Verify Homepoint
Placeprobe tip in
hole of pelvic base
Keep pelvic base
locked throughout
procedure
V3 Homepoint Accuracy
Homepoint : 0 mm

16
Repeat Registrations
Repeating registrations will result in a loss of all previous registrations.
Determine if pelvic base has moved, and select ‘Yes’ or ‘No’ to continue repeating
registrations
End Procedure
Ending procedure will result in a loss of all previous registrations, and will return you
to Step 10 (Pre-Op Selections).
Remove pelvic base and pins prior to
closure.
Shut Down and Disposal
NAVIGATION UNIT MUST BE POWERED OFF AND
DISCARDED
Prior to disposal, power off navigation unit bypressing left
and up arrow buttons until unit is turned off. Additionally,
remove unit from pelvic bracket by pushing down on top
lever.
NOTE: The unit will power off after 90 minutes without a
button being pressed.
DO NOT DISCARD THE REFERENCE
SENSOR or LASER MODULE
Batteries must be removed from reference sensor and laser
module and discarded. Do not reuse batteries.
Return reference sensor and laser module to instrument tray.
!
Warning
All previous registrations
will be lost
Cancel Proceed
Warning:
Has Femur Tack or
Pelvic Base moved?
Yes
No
!
Warning
All previous registrations
will be lost
Cancel Proceed
Warning:
Remove Femur Tack and Pelvic
Base

17
Instruments with Part Numbers
Part Number
Part Description
Quantity
403105
Hex Driver - 3.5MM, Teardrop
1
403258
Probe Calibration Jig - Tripod
1
403370
Laser Module
1
403305
Registration Probe
1
403306
Pelvic Base, Anterior
1
403307
Pelvic Bracket Assembly, Short
1
403338
Straight Cup Impactor
1
405001
or 403087-06
Reference Sensor 5 Hip (RS5H)
1
403317-110
Threaded Fixation Pin, 4.0mm x 110mm with 3.2mm Proximal End
3
403376
Thigh Plate, Slotted
1
403377
Vertical Target, Magnet
1
Additional Instruments
403402
Femur Registration Marker, 3.5mm Hex Head
403108
Screw, 4.5mm dia x 20mm, 3.5mm Hex Head
403339
Impactor Adapter, DPY
403342
Impactor Adapter, S/B
403341
Impactor Adapter, S&N
403343
Impactor Adapter, ZIM
403340
Impactor Adapter, DJO
403344
Impactor Adapter, MPO
Optional Instruments
403309
Offset Cup Impactor
403347
Adapter Socket Driver

18
Instrument Tray Configuration for Sterilization
Position Instruments in Instrument Tray as shown below for sterilization. Refer to the Instructions for Use
for details on cleaning and sterilization parameters.
Vertical Laser
Target
Shell Adapters
Laser Module
Thigh Plate
Pelvic Base

19
Troubleshooting
Dropped data on last registration.
Press center button to repeat
registration.
Check that battery is correctly positioned in
reference sensor. Battery may need to be replaced.
Check that no concurrent OrthAlign or KneeAlign
cases are being conducted. There is a risk of
communication interference between electronic units
if multiple cases are conducted in close proximity.
No reference sensor motion is
detected for registration.
Complete registration sooner. Reference sensor does not report
that it has started navigation. Keep reference sensor in docked
position until traffic light turns green.
The reference sensor was moved too
quickly. Make sure reference sensor
is moved smoothly from one
position to the next, without rapid
rotations.
Place reference sensor in docked
position on jig. Refer to image on
navigation unit for guidance.
The probe point is not in correct
Home position. Repeat this step
with probe tip situated in notch at
the base of jig.
All previous registrations will be lost.
Press left button to return to Step M1.
Press center button to proceed with
Repeat Registrations or End
Procedure.
Move probe, with reference sensor
attached, to align parallel with sagittal
plane. Adjust registration jig to align
navigation unit closer to coronal plane.
Reposition pelvis back to starting
position.
!
Warning
All previous registrations
will be lost
Cancel Proceed

20
Troubleshooting (cont.)
Reference Sensor’s internal temperature
exceeds limit of 70°C. Wait for sensor to
cool down, then continue.
Repeat current step. If error persists,
reference sensor battery may need to
be replaced.
Replace reference sensor battery.
Confirm that all connections are secure.
Choose ‘Yes’ or ‘No’ depending on whether
pelvic base or femur tack has moved during
surgery. If ‘Yes’ is selected, leg length and
offset measurements will be disabled.
Remove reference sensor and clean probe
markings, clean reference sensor window
located on bottom side and re-attach
reference sensor. If problem persists,
restart reference sensor by removing and
replacing reference sensor battery.
There is a major system fault and
navigation unit must be turned off and
replaced. Please recover faulty unit and
return to manufacturer for analysis.
Remove femur tack and screws
from the wound prior to closure.
A software error has occurred. Power unit
off and on again only one time. If System
fault reoccurs after re-starting, power unit
off and replace it. Contact OrthAlign and
return unit for further analysis.
Software Error
Press and hold left
and up buttons to
turn unit off
Then restart
Warning:
Has Femur Tack or
Pelvic Base moved?
Yes
No
Warning:
Remove Femur Tack and Pelvic
Base
!
Warning
Reference Sensor
too hot 89.0º C
Press center button
to retry
Table of contents
Other OrthAlign Medical Equipment manuals
Popular Medical Equipment manuals by other brands

aidapt
aidapt Quiet Knight Mattress Raiser Fitting and Maintenance Instructions

Reliance
Reliance FX920 INSTALLATION AND OPERATING INSTRUCTIONS, OWNERS MANUAL

Squirrel Medical
Squirrel Medical Diamond MRS Instructions for use

Fresenius Medical Care
Fresenius Medical Care DIASAFE 2008 K troubleshooting guide

schwa-medico
schwa-medico PierenFemme operating manual

KLS Martin
KLS Martin Independant IDP 401 Mounting instructions