Panoramic PC-1000 User manual

© Copyright 2008 Panoramic Corporation
PC-1000/Laser 1000 Service Manual
Panoramic Corporation Dental Panoramic/Cephalometric X-ray Machine
SM6000 Rev C

2© Copyright 2008 Panoramic Corporation
Table of Contents
Table of Contents..........................................................................................................................2-4
Introduction...................................................................................................................................5-6
Purpose.....................................................................................................................................5
Statement of Compatibility........................................................................................................5
Limited Warranty.......................................................................................................................6
Voltage Regulator Warning.......................................................................................................6
X-Ray Shielding Requirements.................................................................................................6
Pre-Installation Check .....................................................................................................................7
Electrical Requirements
Control Panel Orientation
Cephalometric Arm Orientation
Remote Switch
Darkroom Requirements
PC-1000 Installation
PC-1000 Components ..............................................................................................................8
PC-1000 Installation.............................................................................................................9-17
Tools Required.......................................................................................................9
Verify Power ..........................................................................................................9
Remove Packaging and Shipping Restraints........................................................9
Control Panel Relocation.......................................................................................9
Install Side Covers...............................................................................................10
Install Film Drum Assembly................................................................................. 11
Install Tubehead Assembly..................................................................................12
Install Temple Supports.......................................................................................12
Level Machine .....................................................................................................12
Perform mA Calibration .......................................................................................13
Perform Pulse Count Calibration....................................................................14-15
Perform X-ray Beam Alignment...........................................................................16
Install Top and Rear Covers................................................................................17
Complete Installation...........................................................................................17
PC-1000/Laser 1000 Installation
PC-1000/Laser 1000 Components .........................................................................................18
PC-1000/Laser 1000 Installation........................................................................................19-39
Tools Required.....................................................................................................19
Verify Power ........................................................................................................19
Remove Packaging and Shipping Restraints......................................................19
Control Panel Relocation.....................................................................................19
Install Side Covers...............................................................................................20
Install Film Drum Assembly.................................................................................21
Install Tubehead Assembly..................................................................................22
Install Temple Supports.......................................................................................22
Level Machine .....................................................................................................22
Install Cephalometric Arm....................................................................................23
Install Cephalostat Head Positioner ....................................................................23
Perform mA Calibration ..................................................................................24-25

3
© Copyright 2008 Panoramic Corporation
Table of Contents
PC-1000/Laser 1000 Installation (continued)
Perform Pulse Count Calibration....................................................................26-27
Perform Panoramic X-ray Beam Alignment.........................................................28
Remove Soft Tissue Shields................................................................................29
Perform Cephalometric X-ray Beam Alignment..............................................29-33
Perform Ear Rod Alignment............................................................................34-36
Install Soft Tissue Shield .....................................................................................37
Install Suggested Settings Decal.........................................................................37
Install Top and Rear Covers................................................................................37
Complete Installation...........................................................................................38
Notes.............................................................................................................................................39
Service Procedures
Remote Switch Installation......................................................................................................40
Screw Motor Height Limit........................................................................................................41
Screw Motor Assembly Exchange .....................................................................................42-43
Panoramic Radiography..................................................................................................Appendix A
Darkroom Procedures .....................................................................................................Appendix B
Manual Processing
Automatic Processing
Darkroom Light Leak Test
Loading The Panoramic Cassette
Loading The Cephalometric Cassette
Maintenance Schedule................................................................................................... Appendix C
PC-1000/Laser 1000 Labeling........................................................................................ Appendix D
PC-1000/Laser 1000 Specifications................................................................................Appendix E
PC-1000/Laser 1000 Space Requirements.....................................................................Appendix F
PC-1000/Laser 1000 Parts List ......................................................................................Appendix G
Final Assembly
Base Assembly
Main Chassis Assembly
Chinrest Arm Assembly
Head Support Assembly
Film Drum Assembly
Upper Belt Drive Assembly
Belt Drive Assembly
Roller Orbit Drive Assembly
Rotating Arm Assembly
Screw Motor Assembly
Tubehead Assembly
Electronics Assembly
Control Panel Assembly

4© Copyright 2008 Panoramic Corporation
Table of Contents
PC-1000/Laser 1000 Parts List (continued) ...................................................................Appendix G
Laser Collimator Assembly
Electromagnet Assembly
PC-1000/Laser 1000 Schematics................................................................................... Appendix H
System Wiring
Control Panel
Timer Circuit Board
X-ray Control
Laser Power/Control

5
© Copyright 2008 Panoramic Corporation
Purpose
Panoramic Corporation provides this printed manual as a guide for the operation of the PC-1000 dental
panoramic X-ray machine and the PC-1000/Laser 1000 dental panoramic/cephalometric dental X-ray
machine.
The PC-1000 will enable the user to take panoramic X-ray images. The PC-1000/Laser 1000 will enable
the user to take panoramic X-ray images, as well as cephalometric X-ray images. A laser alignment
device is incorporated into the PC-1000/Laser 1000.
The information contained in this manual is not all inclusive and Panoramic Corporation should be
contacted for assistance and clarification when necessary.
It is imperative that this equipment be installed, serviced, and used by personnel familiar with the
precautions required to prevent excessive exposure to both primary and secondary radiation. This
equipment features protective designs for limiting both the primary and secondary radiation produced
by the X-ray beam. However, design features cannot prevent carelessness, negligence, or lack of
knowledge.
Only personnel authorized by Panoramic Corporation are qualified to install and service this equipment.
Any attempt to install or service this equipment by anyone not so authorized will void the warranty.
Statement of Compatibility January 1, 1988
Please address any comments/questions concerning this statement of compatibility to:
Panoramic Corporation • 4321 Goshen Road • Fort Wayne, IN 46818 USA • Attn: Director of Engineering
The only components compatible with the PC-1000 are those supplied with the machine.
Regardless of possible statements made by other manufacturers, no one is authorized or certified to
make additions or deletions to this machine. Only the combination of components delivered with the
machine is certified compatible by the manufacturer. As compatible accessories become available, Pan-
oramic Corporation will certify them as compatible and make them available to the user.
Statement of Compatibility Addendum October 1, 1988
The Laser 1000 Cephalometric Attachment is certified by Panoramic Corporation to be compatible with
the PC-1000 dental X-ray machine, provided installation is performed by an authorized representative
utilizing specific installation instructions furnished by Panoramic Corporation.
Statement of Compatibility Addendum October 1, 1995
Laser 1000 Cephalometric Attachments manufactured after October 1, 1995 are compatible only with
Introduction

6© Copyright 2008 Panoramic Corporation
Introduction
Voltage Regulator Warning
Do not plug this machine into ANY voltage regulating device. Contact Panoramic Corporation with any
questions regarding this.
X-ray Shielding Requirements
The requirements for panoramic and cephalometric shielding for building, operator, and patient, depend
on state and local regulations. Contact your state Department of Health for compliance information.
Compliance could involve a blueprint review, facility check, wall construction, film badge implementation,
remote switch installation, and/or a lead apron. It is beyond the scope of this manual to advise on these
regulations.

7
© Copyright 2008 Panoramic Corporation
Pre-Installation Check
Electrical Requirements
Optimally, the PC-1000 and the PC-1000/Laser 1000 should have a dedicated 105-125 VAC, 20Acircuit
with line regulation of 5% or better. If a dedicated circuit is not available, a regular 20 A circuit will work
as long as it is not taxed by other loads beyond 5 A. A standard 115 VAC, three-wire, grounded, electrical
outlet should be installed by an electrician behind the machine.
Control Panel Orientation
The control panel on the PC-1000 and the PC-1000/Laser 1000 is mounted directly on the side of the
machine and requires no separate installation, wiring, or mounting. Although the control panel is mounted
on the patient's left side from the factory, it can be easily relocated to the opposite side during installation.
Verify with the doctor on which side the control panel should be located.
Cephalometric Arm Orientation
The cephalometric arm for the PC-1000/Laser 1000 is shipped in a separate box and can be mounted on
either side of the machine. Verify with the doctor on which side the cephalometric arm should be located.
Remote Switch
Some states and local governments require that the exposure switch to be remotely installed. A remote
switch kit is available from Panoramic Corporation. Refer to Appendix G for detailed procedures.
Darkroom Requirements
Proper darkroom facilities must be present before installing the PC-1000 or PC-1000/Laser 1000. The
darkroom will be used to process test films at the time of installation to verify proper calibration. Refer to
Appendix A for darkroom requirements and processing procedures.

8© Copyright 2008 Panoramic Corporation
PC-1000 Components
PC-1000
Top Cover
Temple Supports
Forehead Support
Mirror
Tubehead
Chinrest
Handles
Base
Rotating Arm
Cassette Sleeve
Film Drum
kVp Setting
Knob
Control Panel

9
© Copyright 2008 Panoramic Corporation
PC-1000 Installation
If there is any question concerning
installation or calibration, please
contact Panoramic Corporation
immediately.
Tools Required
Multimeter, regular and phillips screwdrivers, allen
wrenches, small fluorescent screen, knife, level,
and a pulse counter or stopwatch.
Verify Power
Verify that the outlet is a 110 VAC grounded outlet.
Remove Packaging and Shipping Restraints
1. Carefully remove all shipping packaging from
the PC-1000, including the plastic wrap,
cardboard, and wooden cover. Unpack the
side and top covers shipped in the separate
box.
2. Remove the 3/4" hex head shipping bolt from
the top rear of the overhead chassis. The
shipping bolt is located between the aluminum
chassis and the rear of the machine. There
may be 1 or 2 shipping bolts. Leave these
bolts with the office for future relocation.
Control Panel Relocation
Verify with the doctor, on which side the control
panel needs to be located. To relocate it the
opposite side:
1. Remove the 4 7/16" mounting bolts from
the control panel.
2. Move the control panel to the opposite side
of the machine, while carefully routing the
wiring harness.
3. Reinstall the 4 7/16" bolts in the control
panel.

10 © Copyright 2008 Panoramic Corporation
Install Side Covers
1. Install 3 metal standoffs in the pre-threaded
holes on the side of the column opposite the
control panel.
2. Place the side cover without the power cord
over the standoffs.
3. Insert 3 allen head bolts into the holes in the
side cover and tighten into the metal standoffs.
4. Install 3 metal standoffs in the pre-threaded
holes on the control panel side of the column.
5. Locate the wire with the 15-pin molex
connector J7 in the overhead and connect it
to the 15-pin P7 connector on the top of the
control panel.
6. Connect the 6-pin molex connector P15 on the
main power cord in the side cover to the 6-pin
connector J15 on the top of the control panel.
7. Route the power cord through the notches on
the side of the control panel while placing the
side cover over the standoffs.
8. Insert 3 allen head bolts into the holes in the
side cover and tighten into the metal standoffs.
PC-1000 Installation

11
© Copyright 2008 Panoramic Corporation
PC-1000 Installation
Install Film Drum Assembly
1. Slide the black plastic cover off of the end of
the rotating arm.
2. Remove the cardboard packing material from
the rotating arm.
3. Remove the set screw, crown nut, and washer
from the film drum spindle.
4. Insert the film drum spindle down through the
rotating arm.
5. Install the film drum with the knob closest to the
floor. Slide the film drum onto the spindle up
as far as possible and tighten the film drum
knob to temporarily hold the film drum in place.
6. Slide the flat washer up on the film drum
spindle.
7. Install the crown nut on the film drum spindle
with the slots in the crown nut facing up.
8. Thread the crown nut until the lower edge of
the threaded hole in the film drum spindle is
aligned to the lower edge of a slot in the crown
nut.
9. Insert the set screw in the threaded hole in the
film drum spindle and tighten the screw.
10. While supporting the film drum, loosen the film
drum knob to allow the film drum to rest on the
washer and crown nut.
11. Slide the black plastic cover back on the end
of the rotating arm.
Note: The film drum will rotate with increased
friction after the film drum knob is tightened.
Manual rotation, while the knob is tight, should
be avoided to prevent abnormal wear.

12 © Copyright 2008 Panoramic Corporation
PC-1000 Installation
Install Tubehead Assembly
1. Connect the 6-pin molex connector from the
end of the rotating arm to the 6-pin connector
on the tubehead.
2. Carefully raise the tubehead into position facing
the film drum, ensuring that the wires are not
stressed or pinched.
3. Tighten the three set screws, then back each
out slightly to allow alignment of the tubehead.
Install Temple Supports
1. Install the 2 temple supports using the existing
screws and washers in the head support
assembly.
2. Ensure that the temple supports are angled
away from the machine and that the horizontal
red lines on the temple supports are facing
out.
Level Machine
With a level on the underside of the main
overhead chassis, level the machine in all
directions using the 4 four adjustable feet under
the base.

13
© Copyright 2008 Panoramic Corporation
PC-1000 Installation
Perform mA Calibration
1. Disconnect the yellow wire, 1-pin molex
connector in the overhead chassis.
2. Connect a DC milliammeter in series with
the molex connectors, positive lead to the
male connector, negative lead to the female
connector.
3. Power the machine on.
4. Using the function switch on the control panel,
select PANORAMIC L or PANORAMIC R.
5. Using the kVp setting knob on the control
panel, set the kVp meter to 80 kVp on the
PANORAMIC scale.
6. Toggle the RUN/STOP switch on the printed
circuit board in the overhead chassis to STOP.
7. Locate the mA adjustment knob P1 in the
overhead chassis.
CAUTION: X-rays will be emitted when the
exposure switch is depressed.
8. Depress the exposure switch.
9. The milliammeter should show 6.0 mA ±.05
mA. If it does not:
A. Adjust P1 clockwise (CW) to increase the
mA until the milliammeter shows 6.0 mA ±
.05 mA.
OR B. Adjust P1 counterclockwise (CCW) to
decrease the mA until the milliammeter
shows 6.0 mA ± .05 mA.
Note: Contact Panoramic Corporation if the mA
cannot be set to 6.0 mA ± .05 mA.
10. Disconnect the milliammeter.
11. Reconnect the yellow wire, 1-pin molex
connectors.
12. Toggle the RUN/STOP switch on the printed
circuit board in the overhead chassis to RUN.

14 © Copyright 2008 Panoramic Corporation
Perform Pulse Count Calibration
If a pulse counter IS available:
1. Disconnect the blue wire, 1-pin molex
connector in the overhead chassis.
2. Connect the positive lead of a pulse counter
to the male connector, negative lead to the
aluminum chassis or neutral wire from the
plug.
3. Power the machine on.
4. Using the function switch on the control panel,
select PANORAMIC L.
5. Using the kVp setting knob on the control
panel, set the kVp meter to 80 kVp on the
PANORAMIC scale.
6. Ensure that the RUN/STOP switch on the
printed circuit board in the overhead chassis is
set to RUN.
7. Locate potentiometer R57 on the printed circuit
board in the overhead chassis.
CAUTION: X-rays will be emitted when the
exposure switch is depressed.
8. Depress the exposure switch for the entire 12
second exposure.
9. The pulse counter should show 720 pulses
±10 pulses. If it does not:
A. Adjust R57 counterclockwise (CCW) to
increase the pulses and repeat the test
until the pulse counter shows 720 pulses
±10 pulses.
OR B. Adjust R57 clockwise (CW) to decrease the
pulses and repeat the test until the pulse
counter shows 720 pulses ± 10
pulses.
10. Disconnect the pulse counter.
11. Reconnect the blue wire, 1-pin molex
PC-1000 Installation

15
© Copyright 2008 Panoramic Corporation
If a pulse counter IS NOT available:
1. Power the machine on.
2. Using the function switch on the control panel,
select PANORAMIC L.
3. Using the kVp setting knob on the control
panel, set the kVp meter to 0 kVp on the
PANORAMIC scale.
4. Ensure that the RUN/STOP switch on the
printed circuit board in the overhead chassis is
set to RUN.
5. Use a stopwatch to time the duration of an
exposure by timing the red exposure indicator
on the control panel.
6. Locate potentiometer R57 on the printed circuit
board in the overhead chassis.
CAUTION: X-rays will be emitted when the
exposure switch is depressed.
7. Depress the exposure switch for the entire 12
second exposure.
8. The stopwatch should show that the exposure
is 12 seconds ± .5 seconds. If it does not:
A. Adjust R57 counterclockwise (CCW) to
increase the time and repeat the test until
the stopwatch shows that the exposure is
12 seconds ± .5 seconds.
OR B. Adjust R57 clockwise (CW) to decrease
the time and repeat the test until the
stopwatch shows that the exposure is
12 seconds ± .5 seconds.
PC-1000 Installation

16 © Copyright 2008 Panoramic Corporation
Perform X-ray Beam Alignment
1. Power the machine on.
2. Using the function switch on the control panel,
select PANORAMIC L.
3. Using the kVp setting knob on the control
panel, set the kVp meter to 90 kVp on the
PANORAMIC scale.
4. Toggle the RUN/STOP switch on the printed
circuit board in the overhead chassis to STOP.
5. Temporarily affix a fluorescent screen behind
the film drum mask (slotted plate in front of the
film drum).
CAUTION: X-rays will be emitted when the
exposure switch is depressed.
6. Darken the room and depress the exposure
switch.
7. The fluorescent screen should glow, denoting
the presence of X-rays. Adjust the tubehead
so that the X-ray beam is centered horizontally
in the film mask slot, and that the top edge of
the X-ray beam is aligned to the top edge of
the film drum mask slot.
8. Tighten the 3 set screws locking the tubehead
assembly horizontally and the large allen bolt
on the right side of the tubehead locking it
vertically. Verify that the X-ray beam is still
centered.
9. Remove the fluorescent screen.
10. Toggle the RUN/STOP switch on the printed
circuit board in the overhead chassis to RUN.
11. Using the kVp setting knob on the control
panel, set the kVp meter to 70 kVp on the
PANORAMIC scale.
12. Place a loaded film cassette sleeve on the film
drum and align the L1 pointer.
13. Depress the exposure switch for the entire 12
second exposure.
14. Process the film to verify proper alignment.
The film should be black with clear edges.
PC-1000 Installation

17
© Copyright 2008 Panoramic Corporation
PC-1000 Installation
Install Top and Rear Covers
1. Ensure that the RUN/STOP switch on the
printed circuit board in the overhead is in the
RUN position.
2. Slide the top cover on the machine from the
front. Ensure that the wiring on the control
panel side is properly routed.
3. Install the 6 screws on the underside of the top
cover.
4. Place the cover on the rear of the machine and
install the 3 screws.
Complete Installation
1. Verify all steps have been completed.
2. Refer to the PC-1000/Laser 1000 User Manual
and train the staff thoroughly.
3. Complete the Panoramic Corporation
Installation Report and the FDA 2579
paperwork and distribute accordingly.

18 © Copyright 2008 Panoramic Corporation
PC-1000/Laser 1000 Components
Cephalometric
ExposureTime Chart
90kVP 10 mA
RareEarthScreens And Film
Seconds
PatientSize:
SmallChild
LargeChild
AverageAdult
LargeAdult
Lateral AP-PA
.5-1.0
.6-1.5
.8-2.0
1.0-2.5
.8-1.5
1.0-2.5
1.5-3.5
2.5-4.0
LASER 1000
PanoramicCorporation®
PC-1000
Mid-Sagittal to Film Scale
Cephalostat Locking Knob
Suggested Settings Decal
Cephalostat Arm
Ear Posts
Nasion Support
Ear Rings
Ear Rods
Film Cassette
Horizontal Locking Knob
Vertical Locking Knob
Laser Switch
Collimator Lever

19
© Copyright 2008 Panoramic Corporation
PC-1000/Laser 1000 Installation
If there is any question concerning
installation or calibration, please
contact Panoramic Corporation
immediately.
Tools Required
Multimeter, regular and phillips screwdrivers, allen
wrenches, small fluorescent screen, knife, level,
and a pulse counter.
Verify Power
Verify that the outlet is a 110 VAC grounded outlet.
Remove Packaging and Shipping Restraints
1. Carefully remove all shipping packaging from
the PC-1000/Laser 1000, including the plastic
wrap, cardboard, and wooden cover. Unpack
the side and top covers, and the cephalometric
arm shipped in the separate box.
2. Remove the 3/4" hex head shipping bolt from
the top rear of the overhead chassis. The
shipping bolt is located between the aluminum
chassis and the rear of the machine. There
may be 1 or 2 shipping bolts. Leave these
bolts with the office for future relocation.
Control Panel Relocation
Verify with the doctor, on which side the control
panel needs to be located. To relocate it the
opposite side:
1. Remove the 4 7/16" mounting bolts from
the control panel.
2. Move the control panel to the opposite side
of the machine, while carefully routing the
wiring harness.
3. Reinstall the 4 7/16" bolts in the control
panel.

20 © Copyright 2008 Panoramic Corporation
PC-1000/Laser 1000 Installation
Install Side Covers
1. Install 3 metal standoffs in the pre-threaded
holes on the side of the column opposite the
control panel.
2. Place the side cover without the power cord
over the standoffs.
3. Insert 3 allen head bolts into the holes in the
side cover and tighten into the metal standoffs.
4. Install 3 metal standoffs in the pre-threaded
holes on the control panel side of the column.
5. Locate the wire with the 15-pin molex
connector J7 in the overhead and connect it
to the 15-pin P7 connector on the top of the
control panel.
6. Connect the 6-pin molex connector P15 on the
main power cord in the side cover to the 6-pin
connector J15 on the top of the control panel.
7. Route the power cord through the notches on
the side of the control panel while placing the
side cover over the standoffs.
8. Insert 3 allen head bolts into the holes in the
side cover and tighten into the metal standoffs.
This manual suits for next models
1
Table of contents
Popular Medical Equipment manuals by other brands

ARJO HUNTLEIGH
ARJO HUNTLEIGH Maxi 500 Quick reference guide

Care Fusion
Care Fusion Alaris GP Technical & service manual

Care Fusion
Care Fusion IVAC PCAM Directions for use

New Era Pump Systems
New Era Pump Systems NE-9000-G installation manual

B. Braun
B. Braun Aesculap Spine Instructions for use/Technical description

Hillrom
Hillrom Centuris Pro Instructions for use