ParTech TurbiTechw2 LR User manual

INSTRUCTION MANUAL
TurbiTechw² LR Sensors

** This page is intentionally le t blank **

TurbiTechw² LR Instruction Manual
Table of Contents
1 Foreword........................................................................................................................................................ 6
2 Introduction.................................................................................................................................................... 7
2.1 Manual Conventions.............................................................................................................................. 7
2.2 WaterWatch² Trademark........................................................................................................................ 7
2.3 Scope o Manual.................................................................................................................................... 7
2.4 External Sensors................................................................................................................................... 7
3 Sa ety Precautions......................................................................................................................................... 8
3.1 General.................................................................................................................................................. 8
3.2 Electrical installation.............................................................................................................................. 8
3.3 Operating............................................................................................................................................... 8
3.4 Service and Maintenance...................................................................................................................... 8
3.5 End o Li e Disposal............................................................................................................................... 9
4 The Sensor and Installation........................................................................................................................ 10
4.1 TurbiTechw² LR Sensor....................................................................................................................... 10
5 Mechanical Installation................................................................................................................................. 11
5.1 Operation Limits................................................................................................................................... 11
5.1.1 Temperature................................................................................................................................ 11
5.1.2 Pressure...................................................................................................................................... 11
5.1.3 Maximum Levels.......................................................................................................................... 11
5.1.4 Material Compatibility..................................................................................................................11
5.2 Sensor Components............................................................................................................................ 11
5.2.1 Transmitter Module...................................................................................................................... 12
5.2.2 Receiver Module.......................................................................................................................... 12
5.2.3 Cleaning Module.......................................................................................................................... 12
5.3 Mounting & Installation......................................................................................................................... 12
5.3.1 Process Connections................................................................................................................... 12
5.3.2 Sample Flow Rate....................................................................................................................... 12
5.3.3 Sample Pressure......................................................................................................................... 13
5.4 Optional Accessories........................................................................................................................... 13
5.4.1 Optional Flow Veri ication Detector..............................................................................................13
6 Electrical Installation.................................................................................................................................... 14
6.1.1 Electrical Installation.................................................................................................................... 14
6.1.2 Sensor Connections.................................................................................................................... 14
6.1.3 Extending Sensor Cables............................................................................................................ 15
7 Sensor Con iguration................................................................................................................................... 16
7.1 Sensor Con ig...................................................................................................................................... 16
7.2 Sensor Status...................................................................................................................................... 16
7.3 Add Sensor.......................................................................................................................................... 16
7.4 S:0x TurbiTechw² LR (0-30)................................................................................................................. 17
7.4.1 S:0x Manual Clean...................................................................................................................... 17
7.4.2 S:0x In o....................................................................................................................................... 17
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 3 of 3

TurbiTechw² LR Instruction Manual
7.4.3 S:0x Remove............................................................................................................................... 17
7.4.4 S:0x Modbus Address.................................................................................................................. 17
7.4.5 S:0x Clean Con ig........................................................................................................................ 17
8 Measurement Con iguration......................................................................................................................... 19
8.1 Measurement Con ig............................................................................................................................ 19
8.1.1 Measurement Status.................................................................................................................... 19
8.1.2 Add Measurement.......................................................................................................................19
8.2 M:0x – Measurement Channel............................................................................................................. 20
8.2.1 M:0x In o...................................................................................................................................... 20
8.2.2 M:0x Title..................................................................................................................................... 20
8.2.3 M:0x Units................................................................................................................................... 20
8.2.4 M:0x Set Zero.............................................................................................................................. 20
8.2.5 M:0x Set Cal................................................................................................................................ 20
8.2.6 M:0x Averaging............................................................................................................................ 20
8.2.7 M:0x Remove............................................................................................................................... 20
8.2.8 M:0x Display Position..................................................................................................................20
9 Calibration.................................................................................................................................................... 21
9.1 Calibration Frequency..........................................................................................................................21
9.2 Equipment required or Formazine Calibrations...................................................................................21
9.3 Calibration Zero................................................................................................................................... 22
9.4 Calibration Span.................................................................................................................................. 23
9.5 Span Calibration using the Take Sample Routine...............................................................................23
9.5.1 Take Sample................................................................................................................................ 24
9.5.2 Sample Result.............................................................................................................................25
10 Maintenance............................................................................................................................................. 26
10.1 General cleaning................................................................................................................................ 26
10.2 Inspection.......................................................................................................................................... 26
10.3 Dismantling the Sensor......................................................................................................................26
10.3.1 Removing the Modules..............................................................................................................26
11 Spares....................................................................................................................................................... 27
11.1.1 Service Parts.............................................................................................................................27
12 Sensor Faults............................................................................................................................................. 28
12.1 Unstable reading on controller........................................................................................................... 28
12.2 Error Codes rom 7300w² Monitor...................................................................................................... 28
12.2.1 Input High Limit/Low Limit..........................................................................................................28
12.2.2 Over range/Under range............................................................................................................28
13 General Application Notes.........................................................................................................................29
13.1 Preparation o Formazin Turbidity Standard......................................................................................29
13.1.1 Health & Sa ety Precautions......................................................................................................29
13.1.2 Method o Preparation............................................................................................................... 29
13.1.3 Preparation o Dilution’s............................................................................................................. 29
13.1.4 Calibration.................................................................................................................................. 29
13.1.5 Storage Li e............................................................................................................................... 29
14 Technical Support...................................................................................................................................... 31
14.1 Returning Equipment or Repair........................................................................................................31
Page of 3 225769IM - Issue - 05 Issue Date 30/0 /2019

TurbiTechw² LR Instruction Manual
15 Technical Speci ication.............................................................................................................................. 32
15.1 Physical............................................................................................................................................. 32
15.2 Electrical............................................................................................................................................ 32
15.3 Measurement..................................................................................................................................... 32
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 5 of 3

TurbiTechw² LR Instruction Manual
1 Foreword
The TurbiTechw² amily o sensors are the latest version o the TurbiTech range. The 'w²' in the product title
denotes compatibility with the Waterwatch² range. The sensors are designed or use with the 7300w²
Monitor. The term TurbiTech is used or any Partech sel -cleaning, Suspended Solids or Turbidity Sensor.
The TurbiTechw² LR Sensor has been speci ically designed to monitor inal treated, potable and ilter water.
The sensor is extremely sensitive and is capable o discriminating changes in Turbidity o less than 0.1 NTU.
The 90° light scatter monitoring technique makes the sensor sensitive to a wide range o particle sizes, rom
colloidal material through to larger particles and raw waters and the large mineral particles sometimes ound
in distribution samples.
The sensor requires very little maintenance, with no special tools or training required. Calibration can be
carried out against either a wet Turbidity standard or using the dry calibration re erence cell that is supplied
with every sensor. The re erence cell provides an extremely easy, repeatable method o checking the
per ormance o the sensor without the cost and inherent errors associated with chemical standards.
The TurbiTechw² LR Sensor complies with the internationally recognised ISO7027 standard or Turbidity
measurement. The unique sensor design captures stray light which ensures an ultra stable zero point, this is
combined with an automatic sel -cleaning mechanism that prevent the build up o ouling within the sample
chamber.
Page 6 of 3 225769IM - Issue - 05 Issue Date 30/0 /2019
TurbiTechw² LR Sensor

TurbiTechw² LR Instruction Manual
2 Introduction
2.1 Manual Conventions
All dimensions stated in this manual are in millimetres unless otherwise stated.
The manual has been written assuming the user has a knowledge o instrumentation and an understanding
o the type o measurement being made. Training in the use o the 7300w² Monitor and sensors can be
provided, please contact Partech or urther in ormation.
Icons have been used throughout this manual to draw your attention to precautions and use ul notes.
They are categorised in the ollowing way-
GENERAL NOTES – Speci ications and general notes o interest to the user.
GENERAL CAUTION – Used where caution is required to prevent injury, damage, corruption o
data, loss o calibration or invalidation o warranty etc.
INSTALLATION NOTES – General installation notes o interest to the installer.
ELECTRICAL CAUTION – Used where there is a danger o electric shock to the installer or end
user, or where caution is required to prevent damage to the instrument.
MAINTENANCE NOTES – Used to highlight recommended maintenance procedures and help with
ault inding.
ENVIRONMENTAL NOTES – General notes on environmental issues, waste and disposal.
2.2 WaterWatch² Trademark
WaterWatchw² is the amily name or the w² range o monitors and sensors. Sensors and instruments
designed or speci ic use with the 7300w² Monitor will be su ixed with the w² trademark.
2.3 cope of Manual
This manual describes the installation, con iguration, testing and operation o the TurbiTechw² Sensor.
Please re er to 7300w² Monitor manual or standard unctions o the 7300w² Monitor.
2.4 External ensors
External sensors re ers to any sensors, expansion modules or instruments connected externally to the
7300w² Monitor.
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 7 of 3

TurbiTechw² LR Instruction Manual
3 afety Precautions
3.1 General
Read the sa ety precautions care ully.
Check the delivery o your WaterWatch² sensor or damage. Any damage should be reported to your supplier
as soon as possible.
Use care when unpacking the sensor. NEVER use sharp instruments to open the packaging, as this can
cause damage to the sensor or cable.
Only use accessories speci ically manu actured by Partech or use with this sensor.
Read the operating instructions care ully be ore installing and operating this sensor.
Keep the cable connections dry and ree rom contamination during installation.
Keep the sensor away rom high voltage cables.
3.2 Electrical installation
Only suitably quali ied personnel or a competent person may install, operate or repair this equipment. The
installer must ensure all electrical installations comply with local wiring regulations and standards (re er to
BS7671 or UK installations).
Please check the sensor has been wired correctly. Incorrect wiring may causes damage to the sensor or
monitor.
The WaterWatch² amily o sensors are designed exclusively or use with the 7300w² Monitor. DO NOT
connect to other monitors.
Sensors may need to be correctly addressed to the monitor be ore use. Please read the Sensor
Configuration section o this manual or ull details.
3.3 Operating
Because these sensors have a wide range o applications, users must acquire the appropriate knowledge to
use these sensors in their speci ic application.
Partech are always available to provide advice and assistance in your application. Please contact Partech or
urther in ormation.
These sensors may need to be calibrated be ore use. Please read the Calibration section o this manual or
ull details o calibration procedures.
3.4 ervice and Maintenance
Be ore maintenance, this equipment must be isolated or disconnected rom HAZARDOUS LIVE voltages
be ore access.
Maintenance instructions or the TurbiTechw² sensor should be carried out as speci ied in this instruction
manual. Failure to carry out regular maintenance could invalidate the Warranty.
Services and repairs must be carried out by a Partech engineer. Partech can provide a service contract or
your system. Please ask or details.
Page 8 of 3 225769IM - Issue - 05 Issue Date 30/0 /2019

TurbiTechw² LR Instruction Manual
3.5 End of Life Disposal
Equipment should be recycled according to local regulations.
Any calibration solutions should be disposed o as described in the Manu acture Sa ety Data Sheet
accompanied with the calibration solution.
Partech can provide recycling and disposal o your old Partech equipment, and may also provide the same
service or other manu actures equipment when replaced with Partech equipment.
Partech may provide a trade-in or old Partech equipment when upgrading your system. Please contact us
or urther in ormation.
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 9 of 3

TurbiTechw² LR Instruction Manual
4 The ensor and Installation
Whilst every attempt has been made to ensure that these instructions are correct, common sense and good
engineering practice should always be used, as every installation can present a new set o challenges and
di iculties. I you are in any doubt please contact Partech or your local distributor or urther in ormation.
4.1 TurbiTechw² LR ensor
All the TurbiTechw² Sensors in the amily use an LED in ra red light source, which o ers long-term
stability, low maintenance and high reliability. All TurbiTechw² Sensors use either Light Scatter or Light
Absorption to per orm their measurement.
In simple terms, turbidity means cloudy, hazy or not pure. All drinking water, both raw and treated
contains some degree o turbidity due to dispersed suspended particles such as silt, clay, algae,
organic/inorganic matter or micro organisms.
Turbidity is a measurement quanti ying the degree to which light travelling through the water sample is
scattered by the suspended particles. The transmission o light is determined by physical parameters
such as particle concentration, size, shape and chemical properties. The scattering o light increases
with a greater concentration o suspended load. Turbidity is commonly measured in Nephelometric
Turbidity Units (NTU). The nephelometric method compares the light scattered by the sample and the
light scattered by a re erence solution.
The TurbiTechw² LR Sensor measures the turbidity o the sample water in NTU's in accordance with
established measurement techniques. The measurement cell has been designed to reduce unwanted
stray light to an absolute minimum resulting in a very accurate zero point. The 7300w² Monitor has the
acility or converting the units to mg/l under some circumstances, urther details are provided in the
7300w² Monitor user manual.
The beam rom the LED light source is passed through the sample o water contained in the measurement
cell and the amount o light scattered by particles in the sample is measured at 90 degrees to the light
source. The 7300w² Monitor uses an LED light source with an optical eedback loop to maintain constant
light levels. The sample lows through the lowcell continuously while the TurbiTechw² LR Sensor is in
measurement mode. The large bore lowcell reduces the danger o ouling or clogging in raw water
applications.
The TurbiTechw² LR sensor is a dedicated lowcell sensor and uses a wiper type cleaner. In the
intended clean water applications this wiper is more than capable o keeping the optical sur aces
clean. The wiper should be replaced every two years to ensure the cleaning unction remains e ective.
The TurbiTechw² LR sensor is available in two ranges; 0 to 30NTU & 0 to 500NTU. The range is
dictated by the hardware o the lowcell so cannot be changed by the user. There ore the range should
be speci ied when the instrument is ordered.
All the example screens in this manual relate to the 0 to 30NTU sensor. All the menu options and
unctionality are identical between the ranges except or the actual numeric value and number o
decimal places o NTUs.
Page 10 of 3 225769IM - Issue - 05 Issue Date 30/0 /2019

TurbiTechw² LR Instruction Manual
5 Mechanical Installation
5.1 Operation Limits
5.1.1 Temperature
The upper operational temperature limit is restricted to 60°C. At low temperatures the limit o operation
is 0°C with the practical limitation being ice ormation in the sample. The construction o the sensor is
such that condensation should not present a problem.
5.1.2 Pressure
The standard lowcell provided is not intended or high-pressure applications and should not be
subjected to internal pressures in excess o 1 Bar.
5.1.3 Maximum Levels
High turbidity levels can cause ouling problems that will require manual intervention. An inherent
eature o the principal o operation o the TurbiTechw² Sensor is that at very high turbidity levels the
sensor output will be seen to reduce, this can in some circumstances produce alse low readings.
5.1.4 Material Compatibility
Care should also be taken to ensure material compatibility between the media being monitored and
the wetted parts o the assembly. The principle wetted parts are black Acetal Co-Polymer,
Polypropylene, with Glass Lens. I you are in any doubt about chemical compatibility please contact
Partech.
Regulations governing the use o equipment in contact with potable water exist and these need to be
understood by the user o this product. It is Partech's belie that the low sur ace area in contact with
the potable water and the normal installation practise o eeding the sample stream to waste mean that
this product is suitable or use in potable water treatment processing.
5.2 ensor Components
The TurbiTechw² Sensor comprises o the transmitting module (light source), receive optics, cleaning
mechanism, drain valve, and inter ace unit.
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 11 of 3

TurbiTechw² LR Instruction Manual
5.2.1 Transmitter Module
The transmit module, positioned to the right o the low cell, contains the solid-state emitter and the
light intensity control components. The light intensity is controlled automatically preventing signal
degradation with time. The light beam passes through the sample water. Light is scattered and
ocused onto the target area. The transmitter contains precision optical components and must be
handled with care.
There is an O-ring seal on the ace o the transmitter module. The light source module is retained
using a screw collar.
5.2.2 Receiver Module
The receive module itted on the ront o the low cell is sealed and needs no routine service. The
module contains a photo-diode that receives the light rom the solid-state light source. The output rom
the receiver module passes to the microprocessor. The received signal is increased proportionally with
the turbidity level.
The receiver uses a combination o optical gain and electronic gain to view a very small area o the
low cell. The receiver contains precision optical components and must be handled with care.
The receiver module may be removed rom the low cell by removing the cable rom the inter ace and
the rubber grommets on the bracket. The module can then be unscrewed rom the housing. There is
an O-ring seal on the ace o the receiver module.
5.2.3 Cleaning Module
The cleaning module is located in the top o the measuring cell and locates with a screwed collar.
Access to the cell or manual cleaning is most readily obtained by removing the cleaning motor
assembly. The cleaning arm has a simple wiper blade attached to it. The module is controlled by the
7300w² Monitor. A location peg makes sure the module is correctly aligned during replacement.
Remove this cleaning module or access to the low cell or cleaning and calibration. Ensure the O-ring
is still in place at the bottom o the cleaning module when replacing back into the low cell.
5.3 Mounting & Installation
Whilst the TurbiTechw² Sensor is suitable or outdoor location we recommend that the 7300w² Monitor
is installed indoors or within an outer enclosure. Mount in a location ree rom excessive vibration.
5.3.1 Process Connections
Sample Inlet
½” BSP Female thread itted with 12mm Hose-tail, material Polyamide. The inlet is itted with a Nickel
plated brass ball valve.
Outlet
¼” BSP Female thread itted with 8mm Hose-tail, material Polyamide. This should be ree lowing to air
and not impose any siphoning e ect on the system. The smaller outlet will naturally create a small
positive pressure within the low cell and help minimise the e ect o micro-bubbles on the reading.
Manual Drain
There is a manually operated drain tap at the bottom o the low cell. This allows the cell to be emptied
prior to cleaning or calibration. Isolate the sample low be ore draining the cell.
½” BSP Female with 12mm Hose-tail itted, material Polyamide. The manual drain is itted with a
Nickel plated brass ball valve.
5.3.2 ample Flow Rate
The sample low rate should ideally be in the region o 1 litre/minute. The minimum recommended low
rate is 0.5 l/min, a lower low rate will not cause any damage but will result in a slow system response
Page 12 of 3 225769IM - Issue - 05 Issue Date 30/0 /2019

TurbiTechw² LR Instruction Manual
time. The maximum low rate is 5 l/min, low rates higher than this can cause problems with pressure
and turbulence.
5.3.3 ample Pressure
The sensor is not intended as a pressure containing vessel, however a small positive pressure across
the cell is desirable or maintaining a smooth low and to keep gases dissolved in solution. The
maximum operating pressure is 1 Bar.
5.4 Optional Accessories
Bubbles in the sample will cause alse high and/or unstable readings, micro-bubbles can be generated
i the sample pressure is reduced prior to passing through the lowcell. A De-Bubbler is available, but
will generally not improve situations where micro-bubbles are problematic. For evening out erratic low,
including occasional air intrusions, the De-Bubbler is an ideal solution. Please talk to Partech about
options or mitigating sample conditioning issues.
Note, when itting the De-Bubbler, please ensure the outlet is positioned minimum 300mm higher than
the inlet to the low cell. This will ensure the lowcell has a minimum o 300mm head o pressure.
Adjust the low through the De-Bubbler “Sample Inlet” high enough to allow a low out o the De-
Bubbler vent. Once a low is established, allow to settle or a ew minutes, then reduce the low until
the Vent low just stops (a small dribble rom the vent is pre erable to allow or small luctuations in the
low). The De-Bubbler is now correctly set.
5.4.1 Optional Flow Verification Detector
The low veri ication detector is an optional extra or the measuring system. The low veri ication
detector is itted in-line with the over low drain outlet. Once connected to the system, the low
veri ication detector will provide a signal when no low is detected.
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 13 of 3

TurbiTechw² LR Instruction Manual
6 Electrical Installation
6.1.1 Electrical Installation
Unscrew the two cover screws on the lower panel o the 7300w² Monitor to reveal the Terminals. Each
terminal strip is labelled as illustrated below. (This equipment must be isolated or disconnected rom
HAZARDOUS LIVE voltages be ore access). Re er to the 7300w² Monitor user manual or ull description o
all the terminals within the monitor.
The maximum size wire that can be terminated is 2.5mm² CSA. All the connections are via removable
Plug/Socket terminals. To disengage the terminal strip, simply pull down to release.
6.1.2 ensor Connections
When routing the sensor cables, please ensure the cable is separated rom any mains cables. Although the
Partech w² sensors have a high resistance to inter erence, separation o mains and data cables is good
practice and should always be ollowed where practical.
All sensors in the w² range communicate with the monitor using the ModTechw² Protocol. This protocol has
been speci ically designed to take advantage o the advanced eatures and diagnostics designed into the w²
range o sensors.
Note- These sensors can NOT be used with other monitors that are not included in the w² amily.
All sensors within the w² amily o instruments are connected to the 7300w² Monitor using the same 4 wire
con iguration.
•RED and BLACK wires provide the 12VDC supply to the sensor.
•WHITE and GREEN provide data communication.
A maximum o two sensors can be directly connected to the standard 7300w² Monitor, additional sensors
can be added using the optional expansion boxes available separately.
Remove the 4-way connector rom the 7300w² Monitor by pulling downwards to disconnect or easy access
to the connections. Connect the sensor cores as ollows-
(Terminals rom le t to right on the 4 way connector)
Term 1 (Le t) - RED (+12V)
Term 2 - Black (0V)
Term 3 - White (Data A)
Term 4 (Right) - Green (Data B)
Always connect the screen drain wire with the Black (Term 2). Illustration Le t shows
drain wire and Black wire connected together, and covered in Black Heat shrink.
Always use Bootlace errules when terminating the sensors to ensure a good connection to the terminals.
Page 1 of 3 225769IM - Issue - 05 Issue Date 30/0 /2019

TurbiTechw² LR Instruction Manual
6.1.3 Extending ensor Cables
Sensors are usually supplied with 10M cables (longer cables can be provided i requested). These cables
can be extended to a maximum length o 100M. To ensure optimum per ormance, only use Partech
ModTechw² cable or extensions. Partech can supply junction boxes to allow or cable extensions. These
should be used on all installations where the cable length rom the sensor to the monitor exceeds 20M
(Partech Junction boxes include on-board iltering or long cable lengths). Junction boxes are also use ul or
local connection o sensors close to the sample point. This allows or easy replacement o sensors without
the need to pull back cables to the monitor. The junction box has an on-board terminator switch that can be
activated to terminate the network i the sensor is to be removed or long periods.
When joining cables, ensure the connection is ully waterproo . Any moisture ingress can e ect the
communication between the sensor and monitor.
ModTechw² Cable speci ication-
•2 Twisted Pair - Red/Black (Power) and Green/White (Data) with Screen and Drain wire
•Cores 24AWG (0,22mm²) 7 x 0,20mm
•Outer Insulation – PUR Polyurethane Blue (RAL5003), Diameter - 5mmØ
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 15 of 3

TurbiTechw² LR Instruction Manual
7 ensor Configuration
Be ore attempting to con igure the sensor, please read the user manual
that came with your monitor. The monitor manual will introduce you to
the basic set-up o the monitor, and will amiliarise you with the monitor
menu structure and buttons.
The monitor leaves the actory with no sensors pre-installed.
Assuming the monitor has been physically connected to a sensor, the
next step is to register and con igure the sensor be ore any measurements can be made. A single
sensor may provide one or more measurements. We advise only connecting one sensor at a time.
Once the irst sensor has been registered, connect the second and register again. Repeat or any
additional sensors.
All sensors must be registered to the monitor in this way, even i they are di erent types.
Please note that live measurements are not available until the Sensor
Con iguration stage has been completed.
7.1 ensor Config
From the MAIN MENU screen, select SENSOR CONFIG by pressing ,
and press to accept.
7.2 ensor tatus
This option allows the user to review the current status o the 8 sensor channels, these will all be set to
disabled until a sensor is added.
Once a sensor has been installed the display will be updated to indicate the sensor type installed and
it's status.
7.3 Add ensor
1. From the MAIN MENU screen, select SENSOR CONFIG by
pressing and press .
2. The SENSOR MENU should be displayed. Press to
highlight ADD SENSOR, and press .
3. The Monitor will now search all possible addresses (0 to 240)
to ind any attached sensors. During the search, any sensors ound will be displayed momentary
be ore continuing with the search.
4. Once the search is complete, the Monitor will display a list o sensors ound. Each sensor will be
automatically allocated a new address rom S:01 to S:08.
5. Repeat the above process to install a second, third or more sensors. A total o 8 sensors are
possible (expansion box may be required to add additional sensors).
6. Sensor addition is now complete.
7. I a single or multiple sensors have been ound the can be pressed to escape rom
continuing the search.
Page 16 of 3 225769IM - Issue - 05 Issue Date 30/0 /2019
M:01 NTU
0.14
TURBIDITY
Monitor: OK 09:59:45 17/05/12
MAIN MENU
Monitor Config
Expansion Config
ensor Config
Measurement Config
Alarm Config
Output Config
Information
SENSOR CONFIG
ensor tatus
Add ensor
<No ensors Installed>

TurbiTechw² LR Instruction Manual
7.4 :0x TurbiTechw² LR (0-30)
Once the sensor has been added and registered, the monitor will provide
a list o unctions speci ic to the sensor. Press or to select the
sensor and press . The CONFIG MENU will display a list o sensor
unctions.
7.4.1 :0x Manual Clean
This unction allows the user to initiate a manual clean cycle. During
commissioning it is advisable to per orm a manual clean to test the
operation o the sensor. This test is also use ul or checking the
e ectiveness o the cleaning action in heavily ouling applications or as
part o the sensor servicing routine.
7.4.2 :0x Info
This unction provides a range o diagnostic in ormation that may be requested by Partech or ault
inding
7.4.3 :0x Remove
This allows the sensor to be removed or re-con iguration o the monitor or i a sensor has been added
on error. I a sensor has been replaced with a new sensors, the old sensor must be removed, and the
new sensor installed.
You will be prompted with 'Are you sure?' be ore the sensor is removed. Press to accept and
remove.
7.4.4 :0x Modbus Address
This option allows manual adjustment o the ModTechw² address or the sensor, under normal
circumstances this should not be changed
7.4.5 :0x Clean Config
This gives access to a new sub-menu to con igure the cleaning regime.
S:0x Clean Interval
This allows this user to set the time between automatic cleans, the time is
set in minutes. The actory setting is 360 minutes (6 hours), the requency
can be adjusted between 2 and 1440 minutes (24 hours). Under normal
circumstances cleaning should be no more requent that every 60 minutes.
Please remember the more requently the sensor cleans the quicker the
blade will wear out.
S:0x Clean Mode
The sensor 'clean mode' can be set to On/O .
S:0x Service Life
The actory setting or the sensors is 3500 cleans between services, this
equates to around 2 years o normal operation. This service li e will need to be adjusted in some
applications due to the variable nature o the solids being monitored. Abrasive solids such as sand will
cause the seal to wear out more quickly.
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 17 of 3
SENSOR CONFIG
ensor tatus
Add ensor
:01 TurbiTechw2 LR (0-30)
S:01 CONFIG
:01 Remove
:01 Modbus Address
:01 Clean Config
:01 Manual Clean
:01 Info
S:01 CONFIG
:01 Remove
:01 Modbus Address
:01 Clean Config
:01 Manual Clean
:01 Info
:01 CLEAN CONFIG
:01 ervice Life
:01 Clean ervice
:01 Clean Info
:01 Clean Interval
:01 Clean Mode

TurbiTechw² LR Instruction Manual
S:0x Clean Service
This allows the user to reset the clean counter when a service is carried out.
S:0x Clean Info
This option reports the number o cleans carried out by the sensor since the last reset, the time o the
next clean and estimated service li e
Page 18 of 3 225769IM - Issue - 05 Issue Date 30/0 /2019

TurbiTechw² LR Instruction Manual
8 Measurement Configuration
The monitor leaves the actory without any measurements con igured. Measurements can only be
added a ter installing the relevant sensor(s).
Once the sensor(s) have been registered with the monitor and installed, the measurements will now be
available.
8.1 Measurement Config
From the MAIN MENU screen, select MEASUREMENT CONFIG by
pressing , press . The screen shot to the le t shows the de ault
con iguration a ter the installation o the TurbiTechw² sensor on a single
sensor con iguration:
M:01 = Turbidity (NTU)
8.1.1 Measurement tatus
This option allows the user to review the current status o the 16 measurement channels, these will all
be set to disabled until a sensor is added. In the example above the irst channel will be occupied.
Once a measurement has been con igured the display will be updated to indicate the measurement
and it's status.
8.1.2 Add Measurement
Unless a measurement has been removed there would be no condition
where the 'Add Measurement' option is required as the only measurement
type is Turbidity (NTU) which is selected by de ault when a sensor is
installed. The ollowing description is included or the purposes o
completeness. Note the screenshot on the left shows 'Measurement
Config' with the Turbidity sensor removed.
1. The MEASUREMENT MENU should be displayed. Press to highlight ADD MEASUREMENT,
and press .
2. All available measurements will be displayed in a list. Press or to select the irst
measurement. Note Colour (HU) is the only option.
3. Press to select the measurement. Repeat the process i more measurements are required.
4. Each measurement will be allocated a measurement number rom M:01 – M:16. A total o 16
measurements may be displayed.
5. Press to return back to the display screen. Your irst
measurement con igured should now be displayed. I two
measurements were con igured, press to cycle through the
display screens to show 2 measurements, 4 measurements etc.
The Measurement Menu will list all con igured measurements in order M:01
to M:16 the list will also indicate the sensor number that is delivering the
signal or the measurement.
225769IM - Issue - 05 Issue Date 30/0 /2019 Page 19 of 3
MEASUREMENT CONFIG
Measurement tatus
Add Measurement
M:01 Turbidity ( :01)
MEASUREMENT CONFIG
Measurement tatus
Add Measurement
MEASUREMENT CONFIG
Measurement tatus
Add Measurement
M:01 Turbidity ( :01)

TurbiTechw² LR Instruction Manual
8.2 M:0x – Measurement Channel
Selecting a measurement channel will reveal a new sub-menu associated
with that measurement. In MEA UREMENT CONFIG press to
highlight the required measurement and press .
The sub-menu is as ollows:
8.2.1 M:0x Info
This option provides additional in ormation on the measurement, this
in ormation will only be required i a problem exists with the instrument
per ormance.
8.2.2 M:0x Title
This allows the title o the measurement to be changed rom it's de ault,
the measurement title is used in measurement mode to identi y the
measured value. A selection o standard terms are available along with a 'User De ined' option that
can be adjusted to suit your requirements. For example this could be changed to 'Turbidity Line 1'. The
maximum number o characters is 20.
8.2.3 M:0x Units
This allows the units o the measurement to be changed rom it's de ault, the measurement units are
used in measurement mode. A selection o standard terms are available along with a 'User De ined'
option that can be adjusted to suit your requirements. The maximum number o characters is 4.
8.2.4 M:0x et Zero
The term relate to the calibration o the system and is covered in the section below.
8.2.5 M:0x et Cal
The term relate to the calibration o the system and is covered in the section below.
8.2.6 M:0x Averaging
This allows the user to impose averaging on the measured value, this is used to reduce the speed o
reaction to the process changes. The value can be adjusted between 0 and 6000, the higher the value
the slower the reaction time.
8.2.7 M:0x Remove
This allows the user to remove a measurement that has been selected in
error or to allow re-con iguration o the system. Please use this option with
care, all user settings will be lost i the measurement is removed in error.
8.2.8 M:0x Display Position
This option allows the position o the measurement to be moved. For
example the Colour measurement can be changed rom M:01 to M:02 so it will appear second on the
list in MEASUREMENT CONFIG menu.
Page 20 of 3 225769IM - Issue - 05 Issue Date 30/0 /2019
M:01 CONFIG
M:01 Info
M:01 Title
M:01 et Zero
M:01 Averaging
M:01 Remove
M:01 Units
M:01 et Cal
M:01 CONFIG
M:01 Info
M:01 Title
M:01 et Zero
M:01 Averaging
M:01 Remove
M:01 Units
M:01 et Cal
MEASUREMENT CONFIG
Measurement tatus
Add Measurement
M:01 Turbidity ( :01)
This manual suits for next models
1
Table of contents
Other ParTech Accessories manuals
Popular Accessories manuals by other brands

Techly
Techly IUPS-PCP-81U2URJ user manual

Panasonic
Panasonic UE-608040 installation instructions

ROSE DISPLAYS
ROSE DISPLAYS ANOFRAME ROUND ASSEMBLED-MOLLY ANCHORS instruction sheet

SICK
SICK NAV350 operating instructions

Lanaform
Lanaform Bali manual

Hydac
Hydac CS 1000 Series Operating and maintenance instructions