PCB Piezotronics 350M88A User manual

Model 350M88A
Low Outgassing, Shock, ceramic shear, ICP® accel, 0.5 mV/g, 10kg, 10ft integral cable, filtered,
base model 350C23
Installation and Operating Manual
For assistance with the operation of this product,
contact the PCB Piezotronics, Inc.
Toll-free: 716-684-0001
24-hour SensorLine: 716-684-0001
Fax: 716-684-0987
E-mail: [email protected]
Web: www.pcb.com
Manual 21354 Rev E
ECN 50523
Repair and Maintenance
PCB guarantees Total Customer Satisfaction through its
“Lifetime Warranty Plus” on all Platinum Stock Products
sold by PCB and through its limited warranties on all other
PCB Stock, Standard and Special products. Due to the
sophisticated nature of our sensors and associated
instrumentation, field servicing and repair is not
recommended and, if attempted, will void the factory
warranty.
Beyond routine calibration and battery replacements
where applicable, our products require no user
maintenance. Clean electrical connectors, housings, and
mounting surfaces with solutions and techniques that will
not harm the material of construction. Observe caution
when using liquids near devices that are not hermetically
sealed. Such devices should only be wiped with a
dampened cloth—never saturated or submerged.
In the event that equipment becomes damaged or ceases
to operate, our Application Engineers are here to support
your troubleshooting efforts 24 hours a day, 7 days a
week. Call or email with model and serial number as well
as a brief description of the problem.
Calibration
Routine calibration of sensors and associated
instrumentation is necessary to maintain measurement
accuracy. We recommend calibrating on an annual basis,
after exposure to any extreme environmental influence,
or prior to any critical test.
PCB Piezotronics is an ISO-9001 certified company whose
calibration services are accredited by A2LA to ISO/IEC
17025, with full traceability to SI through N.I.S.T. In
addition to our standard calibration services, we also offer
specialized tests, including: sensitivity at elevated or
cryogenic temperatures, phase response, extended high
or low frequency response, extended range, leak testing,
hydrostatic pressure testing, and others. For more
information, contact your local PCB Piezotronics
distributor, sales representative, or factory customer
service representative.
Returning Equipment
If factory repair is required, our representatives will
provide you with a Return Material Authorization (RMA)
number, which we use to reference any information you
have already provided and expedite the repair process.
This number should be clearly marked on the outside of
all returned package(s) and on any packing list(s)
accompanying the shipment.
Contact Information
PCB Piezotronics, Inc.
3425 Walden Ave.
Depew, NY14043 USA
Toll-free: (800) 828-8840
24-hour SensorLine: (716) 684-0001
General inquiries: in[email protected]
Repair inquiries: rma@pcb.com
For a complete list of distributors, global offices and sales
representatives, visit our website, www.pcb.com.
Safety Considerations
This product is intended for use by qualified personnel
who recognize shock hazards and are familiar with the
precautions required to avoid injury. While our equipment
is designed with user safety in mind, the protection
provided by the equipment may be impaired if equipment
is used in a manner not specified by this manual.
Discontinue use and contact our 24-Hour Sensorline if:
Assistance is needed to safely operate equipment
Damage is visible or suspected
Equipment fails or malfunctions
For complete equipment ratings, refer to the enclosed
specification sheet for your product.
Definition of Terms and Symbols
The following symbols may be used in this manual:
DANGER
Indicates an immediate hazardous
situation, which, if not avoided, may
result in death or serious injury.
Manual 21354 Rev E
ECN 50523
CAUTION
Refers to hazards that could damage
the instrument.
NOTE
Indicates tips, recommendations and
important information. The notes
simplify processes and contain
additional information on particular
operating steps.
The following symbols may be found on the equipment
described in this manual:
This symbol on the unit indicates that
high voltage may be present. Use
standard safety precautions to avoid
personal contact with this voltage.
This symbol on the unit indicates that
the user should refer to the operating
instructions located in the manual.
This symbol indicates safety, earth
ground.
Manual 21354 Rev E
ECN 50523
PCB工业监视和测量设备 - 中国RoHS2公布表
PCB Industrial Monitoring and Measuring Equipment - China RoHS 2 Disclosure Table
部件名称
有害物质
铅(Pb)
汞
(Hg)
镉
(Cd)
六价铬(Cr(VI))
多溴联苯 (PBB)
多溴二苯醚(PBDE)
住房
O
O
O
O
O
O
PCB板
X
O
O
O
O
O
电气连接器
O
O
O
O
O
O
压电晶体
X
O
O
O
O
O
环氧
O
O
O
O
O
O
铁氟龙
O
O
O
O
O
O
电子
O
O
O
O
O
O
厚膜基板
O
O
X
O
O
O
电线
O
O
O
O
O
O
电缆
X
O
O
O
O
O
塑料
O
O
O
O
O
O
焊接
X
O
O
O
O
O
铜合金/黄铜
X
O
O
O
O
O
本表格依据 SJ/T 11364 的规定编制。
O:表示该有害物质在该部件所有均质材料中的含量均在 GB/T 26572 规定的限量要求以下。
X:表示该有害物质至少在该部件的某一均质材料中的含量超出 GB/T 26572 规定的限量要求。
铅是欧洲RoHS指令2011/65/ EU附件三和附件四目前由于允许的豁免。
CHINA RoHS COMPLIANCE
Manual 21354 Rev E
ECN 50523
Component Name
Hazardous Substances
Lead (Pb)
Mercury (Hg)
Cadmium (Cd)
Chromium VI
Compounds
(Cr(VI))
Polybrominated
Biphenyls (PBB)
Polybrominated
Diphenyl Ethers
(PBDE)
Housing
O
O
O
O
O
O
PCB Board
X
O
O
O
O
O
Electrical Connectors
O
O
O
O
O
O
Piezoelectric Crystals
X
O
O
O
O
O
Epoxy
O
O
O
O
O
O
Teflon
O
O
O
O
O
O
Electronics
O
O
O
O
O
O
Thick Film Substrate
O
O
X
O
O
O
Wires
O
O
O
O
O
O
Cables
X
O
O
O
O
O
Plastic
O
O
O
O
O
O
Solder
X
O
O
O
O
O
Copper Alloy/Brass
X
O
O
O
O
O
This table is prepared in accordance with the provisions of SJ/T 11364.
O: Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit
requirement of GB/T 26572.
X: Indicates that said hazardous substance contained in at least one of the homogeneous materials for this part is above the limit
requirement of GB/T 26572.
Lead is present due to allowed exemption in Annex III or Annex IV of the European RoHS Directive 2011/65/EU.
Model Number
350M88A
ICP® SHOCK ACCELEROMETER
Revision: NR
ECN #: 53614
Perfo rmance ENG LISH SI
Sensitivity(± 30 %) 0.5 mV/g .05 mV/(m/s²)
Measurement Range ± 10,000 g pk ± 98,000 m/s² pk
Frequency Range(± 1 dB) 0.4 to 10,000 Hz 0.4 to 10,000 Hz
Frequency Range(- 3 dB) 0.2 to 25,000 Hz 0.2 to 25,000 Hz [1]
Electrical Filter Corner Frequency(- 3 dB) 13 kHz 13 kHz [2][3]
Mechanical Filter Resonant Frequency 23 kHz 23 kHz [2][4]
Resonant Frequency ≥ 100 kHz ≥ 100 kHz
Broadband Resolution(1 to 10,000 Hz) 0.04 g rms 0.39 m/s² rms [2]
Non-Linearity(per 1000 g (9810 m/s2)) ≤ 2 % ≤ 2 %
Transverse Sensitivity ≤ 7 % ≤ 7 %
Environmental
Overload Limit(Shock) ± 50,000 g pk ± 490,000 m/s² pk
Temperature Range(Operating) -10 to +150 °F -23 to +66 °C
Temperature Range(Storage) -40 to +200 °F -40 to +93 °C
Temperature Response See Graph See Graph [2][5]
Base Strain Sensitivity 0.002 g/µε 0.02 (m/s²)/µε [2]
Electrical
Excitation Voltage 20 to 30 VDC 20 to 30 VDC
Constant Current Excitation 2 to 20 mA 2 to 20 mA
Output Impedance ≤ 200 Ohm ≤ 200 Ohm
Output Bias Voltage 8 to 14 VDC 8 to 14 VDC
Discharge Time Constant 1 to 2 sec 1 to 2 sec [2]
Settling Time(within 10% of bias) ≤ 10 sec ≤ 10 sec
Electrical Isolation(Case) ≥ 1,000,000 Ohm ≥ 1,000,000 Ohm
Physical
Sensing Element Ceramic Ceramic
Sensing Geometry Shear Shear
Housing Material Titanium Titanium
Sealing Hermetic Hermetic
Size (Hex x Height) 0.375 in x 1.04 in 3/8 in x 26.5 mm
Weight(without cable) 0.19 oz 5.4 gm [2]
Electrical Connector Integral Cable Integral Cable
Cable Termination 10-32 Coaxial Plug 10-32 Coaxial Plug
Cable Length 10 ft 3.05 m
Cable Type 031 Twisted Pair 031 Twisted Pair
Mounting Thread 1/4-28 Male 1/4-28 Male
[6]
All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.
ICP® is a registered trademark of PCB Piezotronics, Inc.
OPTION AL VERSIONS
Optional versions have identical specifications and accessories as listed for the standard model except
where noted below. More than one option may be used.
NOTES:
[1]Typical corner frequency for coupled electrical and mechanical filters.
[2]Typical.
[3]Electrical filter is a second order filter.
[4]Amplitude at resonance is +9 dB.
[5]Test conducted on shaker at 100Hz.
[6]See PCB Declaration of Conformance PS135 for details.
Date: 03/06/2023 Date: 03/06/2023 Date: 03/06/2023 Date: 03/06/2023 75594
Entered: ND Engineer: MB Sales: KK Approved: BAM Spec Number:
Phone: 716-684-0001
Fax: 716-684-0987
E-M ail: info@ pcb.com
3425 Walden Avenue, Depew, NY 14043
SUPPLIED ACCESSORIES:
Model ACS-14 High G shock accelerometer calibration using Hopkinson bar. (1)
Model ACS-22 NIST Traceable frequency response (100Hz to ±1 dB point) (1)
1
1
2
2
3
3
4
4
A A
B B
CODE
IDENT. NO.
52681
DWG. NO.
SCALE: SHEET
DRAWN CHECKED ENGINEER
TITLE
UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:
DIMENSIONS IN MILLIMETERS
[ IN BRACKETS ]
XX ±0.13
ANGLES 2 DEGREES
DIMENSIONS IN INCHES
DECIMALS XX ±.01
ANGLES 2 DEGREES
FILLETS AND RADII
.003 - .005
DECIMALS X ± 0.3
FILLETS AND RADII
0.07 - 0.13
INSTALLATION DRAWING
75595
1 OF 1
4X
MECHANICALLY FILTERED
SHOCK ACCELEROMETER
XXX ±.005
NJF 03/03/23 JDM 03/03/23 MJB 03/03/23
3425 WALDEN AVE. DEPEW, NY 14043
(716) 684-0001 E-MAIL: sales@pcb.com
1"≤LENGTH < 1' = +1"/ - 0
1'≤LENGTH < 5' = +2"/ - 0
5'≤LENGTH < 100' = +6"/ - 0
100'≤LENGTH = +1'/ - 0
CABLE TOLERANCES IN ENGLISH CABLE TOLERANCES IN METRIC
2.54cm ≤LENGTH < 30.5cm = +2.54cm/ - 0
30.5cm ≤LENGTH < 1.5m = +5.1cm/ - 0
1.5m ≤LENGTH < 30.5m = +15.2cm/ - 0
30.5m ≤LENGTH = +30.5cm/ - 0
350M88A
75595
PCB Piezotronics Inc. claims proprietary rights in
the information disclosed hereon. Neither it nor any
reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.
1.08 .03 27.3 .8
[]
REVISIONS
REV DESCRIPTION DIN
NR RELEASED TO DRAFTING 53614
RECOMMENDED MOUNTING TORQUE ON .375 [9.52] HEX: 2.5 - 4 FT-LBS [3.4 - 5.4 N-m]
1
PIN OUTS
FUNCTION 031 EB
SIGNAL/POWER RED CENTER PIN
GROUND WHITE SHELL
CASE N/A N/A
MOUNTING HOLE PREPARATION
Ø.22±.03 [5.4±.8] 30±.03 [7.6±.8] MIN
1/4-28 UNF-3B 20±.03 [5.1±.8] MIN
(M) OPTION
Ø.21±.03 [5.9±.8] 30±.03 [7.6±.8] MIN
M6x0.75-6H 20±.03 [5.1±.8] MIN
.38 .03 9.7 .8
[]
MIN
A
.013 A
.001 [.03]
63 [1.6]
.37 .03 9.4 .8
[]
2X 
.26 .03 6.5 .8
[]
.20 .03 5.1 .8
[] .02 .03 0.5 .8
[]
1/4-28 UNF - 3A
(M) OPTION: M6x0.75-4g6g
.375 [9.52] HEX 1FOR CABLE AND ELECTRICAL CONNECTOR
SEE SPECIFICATION SHEET
CLEAR HEAT SHRINK

Other PCB Piezotronics Measuring Instrument manuals

PCB Piezotronics IMI SENSORS 607A12 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 607A12 User manual

PCB Piezotronics 357B04 User manual

PCB Piezotronics

PCB Piezotronics 357B04 User manual

PCB Piezotronics M355B03 User manual

PCB Piezotronics

PCB Piezotronics M355B03 User manual

PCB Piezotronics 352C34 User manual

PCB Piezotronics

PCB Piezotronics 352C34 User manual

PCB Piezotronics J357B34 User manual

PCB Piezotronics

PCB Piezotronics J357B34 User manual

PCB Piezotronics IMI SENSORS 627A71 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 627A71 User manual

PCB Piezotronics IMI Sensors 604B32 User manual

PCB Piezotronics

PCB Piezotronics IMI Sensors 604B32 User manual

PCB Piezotronics IMI SENSORS EX607A01 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS EX607A01 User manual

PCB Piezotronics EX357E90 User manual

PCB Piezotronics

PCB Piezotronics EX357E90 User manual

PCB Piezotronics IMI SENSORS 602D10 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 602D10 User manual

PCB Piezotronics 3713B112G User manual

PCB Piezotronics

PCB Piezotronics 3713B112G User manual

PCB Piezotronics M353B77 User manual

PCB Piezotronics

PCB Piezotronics M353B77 User manual

PCB Piezotronics 356A45 User manual

PCB Piezotronics

PCB Piezotronics 356A45 User manual

PCB Piezotronics PM119B11 User manual

PCB Piezotronics

PCB Piezotronics PM119B11 User manual

PCB Piezotronics M350A14 User manual

PCB Piezotronics

PCB Piezotronics M350A14 User manual

PCB Piezotronics IMI Sensors 608A11 User manual

PCB Piezotronics

PCB Piezotronics IMI Sensors 608A11 User manual

PCB Piezotronics IMI SENSORS M621B41 User manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS M621B41 User manual

PCB Piezotronics 608A11/-0112 User manual

PCB Piezotronics

PCB Piezotronics 608A11/-0112 User manual

PCB Piezotronics 3743F11200G User manual

PCB Piezotronics

PCB Piezotronics 3743F11200G User manual

PCB Piezotronics TLD339A36 User manual

PCB Piezotronics

PCB Piezotronics TLD339A36 User manual

PCB Piezotronics ICP TLD356A34 User manual

PCB Piezotronics

PCB Piezotronics ICP TLD356A34 User manual

PCB Piezotronics 337A36 User manual

PCB Piezotronics

PCB Piezotronics 337A36 User manual

PCB Piezotronics 333B45 User manual

PCB Piezotronics

PCB Piezotronics 333B45 User manual

PCB Piezotronics J357B03 User manual

PCB Piezotronics

PCB Piezotronics J357B03 User manual

Popular Measuring Instrument manuals by other brands

Triplett WireMaster XR-2 instruction manual

Triplett

Triplett WireMaster XR-2 instruction manual

Panametrics HydroProII user manual

Panametrics

Panametrics HydroProII user manual

HT HT64 user manual

HT

HT HT64 user manual

BEKA BA358C manual

BEKA

BEKA BA358C manual

Garmin GI 275 Maintenance manual

Garmin

Garmin GI 275 Maintenance manual

JDiag FasCheck BT-200 user manual

JDiag

JDiag FasCheck BT-200 user manual

WIKA A-AS-1 operating instructions

WIKA

WIKA A-AS-1 operating instructions

Niigata seiki LSL-26 instruction manual

Niigata seiki

Niigata seiki LSL-26 instruction manual

Interscan 4000 series instruction manual

Interscan

Interscan 4000 series instruction manual

MYPIN DR Series user manual

MYPIN

MYPIN DR Series user manual

SCHUNK IFT manual

SCHUNK

SCHUNK IFT manual

GTE CoboSafe-Scan operating manual

GTE

GTE CoboSafe-Scan operating manual

Ofil daycor superb user guide

Ofil

Ofil daycor superb user guide

Vaisala GMP252 user guide

Vaisala

Vaisala GMP252 user guide

Hach Lange DR 3900 user manual

Hach

Hach Lange DR 3900 user manual

JUMO 202710/30 operating manual

JUMO

JUMO 202710/30 operating manual

SKF LAGM 1000E Instructions for use

SKF

SKF LAGM 1000E Instructions for use

Triplett RHT02 user manual

Triplett

Triplett RHT02 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.