
5.6.3. Data listing cursor
In data listing interface, cursor1, cursor2 and cursor measurement, are all the
same as the cursors in timing waveform interface. The difference is that cursors
in timing waveform display interface are vertical cursor lines, but in data listing
display interface cursors are level white display row. There is only address
display of two cursor rows about parameter display under data listing (data values
in the listing already).
5.7. Sampling setting
The logic analyzer uses the sampling mode for data obtaining, which is sampling
to input “digital”, notice that it is not to collect sample to input signal directly, but
to digital generated through comparison and distinguishing for input signal and
threshold, and stores sampling data in memory. So, the basic requirement for
logic analyzer use is setting sampling parameters correctly.
5.7.1. Sampling source
The purpose of using logic analyzer is to analyze the logic state of signal to be
measured, so the aim signal to be sampled is external signal source surely. But
take the following three points for consideration, set an educational pulse
generator in instrument inner. First, many users may be not familiar with the logic
analyzer, some may use it the first time, so first make operation practice to
deepen apprehension for the correlation between various parameters’ setting,
and familiarize operational control for display interface, and not need to connect
mass of input test nips, very conveniently no doubt. Moreover, external signals to
be measured are usually unknown and complicated, may be mixed with random
interference signal too, so it is necessary to set threshold voltage suitably and
take sample to external signal directly, getting expectant result may be more
difficult. Because the Inner Code Generator’s logic relation is simple, waveform is
pure and standard; one can quickly master logic analyzer’s use with it. Second, in
practical application, if the test result is different from expectant one, sometimes
it is hard to make clear that the problem comes from the tested circuit or logic
analyzer, here as long as use inner code generator to take samples and contrast,
it is easy to find reasons. Third, when instrument fails to work and need repairing
and debugging, the code generator is very convenient of course.
The inner code generator is composed of two types of signals, 00 ~ 15 channels
are count waveforms, count value are 0~65535, and each clock cycle adds 1.
16~29 channels are shift pulse waveforms and high level pulse moves one bit to
the right in turn from low to high channel. 30~31 channels are used to measure
external clock1 and 2 (6.7.2). The code generator in the equipment, if only power
is open, runs in its independent clock continuously till power closes. The inner
code generator directly generates standard digital signals for sampling, not
passing threshold voltage comparator. If choose inner code generator, then
external digital signal cuts automatically, vice versa, they don’t affect each other.
Pressing 【source 】key can switch into inner code generator or external signal
source in cycle, instrument default setting is inner code generator. The internal
code-generator has its own adjustable clock, can emulate measured signals with
different velocities, press 【system】to select parameter pattern-clk and input
the clock cycle value with decimalization numbers 0~9, its unit is ns, resolution is
10ns,the last number on the right has no use. The minimum cycle value is 20 ns;
the maximum value is 999999990ns, approximate to 1s. After code generator’s
clock changes, waveform display’s change is visible. The default setting of the
clock cycle is 20 ns.
-13-