Page 4of 4
4
Basic Trouble Shooting
The load cells in the 69PB/6950 base have tested and corner
compensated at the factory and should not require adjustment for
installation. If the base is displaying inaccurate weight readings or
corner weight errors, check the following before making
potentiometer adjustments:
•Check the cable; is it wedged between the leveling feet and
floor or in between the load cell and base structure?
•Is the base fully and firmly supported by the leveling feet?
Does the base rock or tip if weight is placed on one corner?
•Are there any obstructions between the load cells and the
leveling feet?
•Is the CPC connector firmly attached? Are the pins inside the
connector bent or damaged?
Attaching A Ramp
Remove the leveling feet from the load cell on the side of the base
the ramp is to be installed on. Place the leveling foot stem through
the ramp bracket and screw the leveling foot into the load cell.
Attach the ramp to the ramp brackets.
Operation
With the wheels in the DOWN position, move the base to the
location where it will be used. Place the wheels in the UP position
so that the base is on the leveling feet. Adjust the leveling feet so
the base is level and does not rock when weight is placed on
one corner or end. Place the ramp in the down position. Power up
the indicator and allow several minutes for warm up of electronics
and load cells. The system is now ready for weighing. To move to
another location, reverse the above procedures.