Dear customer,
Waismed is doing its best to supply you with the top quality products at
the highest quality levels.
We hope you will nd the Reloading System to your satisfaction.
Should you encounter any problem or inquiry please contact us or your
local dealer.
NOTES AND PRECAUTIONS:
! The spring used in the reloadable devices is weaker than the sterile BIG, in
order to enable easier reloading of the device.
! The Reloading Device is intended for use only with the Demo BIG supplied
with it.
! The use of the Reloading Device is only for authorized personnel who were
trained to use and reload the BIG.
i For addional informaon and safety issues refer to the included CD-
ROM.
i The training BIG is a non sterile device, and therefore should only be used
with the aached Foam or the WMTLeg (training leg).
i The manufacturer recommends the Reloading Device to be used no more
than 50 mes.
i The manufacturer recommends the Demo Reloadable Device to be used
no more than 50 mes.

3
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
In order to reload the BIG you will need the following parts:
I. RearBody
II. Piston
III. Washer(Ring)
IV. Spring
V. Needle
VI. SafetyLatch
VII. FrontBody

4
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
Position the Reloading Device with
Waismed sticker facing you.
The handles should be in Position
2.
Remove the cap at the top of the
Reloading Device. Twisting it
counter clockwise and pull it up.

5
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
Move the Washer (Ring) (III)
on top of the Piston (II) to the
small groove on the Needle-
Gripping Prongs.
Match the Piston (II) to the small
Pin inside the Reloading Device.

6
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
Slide the spring (IV) over the
Piston (II).
Slide the Rear Body of the B.I.G.
(I) over the Piston (II) and Spring
(IV); make sure that the Bolder
Raised Arrow matches the groove
on the back of the device and the
Shallower Arrow is pointing at the
red line on the front of the device.
! Arrows should be pointed
downwards.

7
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
Move the handles to Position 1.
With your ngers, press the
top of the BIG and press it
downwards. You should feel the
spring’s resistance.
While holding the handles
in Position 1, close the cap
by pushing it downwards and
twisting it clockwise until the
two red string holders match.

8
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
Lower the handles to Position
2 and press it downwards until a
click sound is heard.
! The handles base and the device
base should touch one another.
! Do not lean on the far side of the
handles as they might break.
Raise the handles to Position 3.
! Caution!
The spring (IV) is now loaded and
great care should be taken while
handling the BIG.
! Do not try to remove the BIG
without inserting the safety latch
(VI) (step 12).

9
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
Lower the red clip on the cap’s
back with your nger until you
can insert the Safety Latch (IV).
Insert the safety latch (VI).
! Make sure the safety latch (VI)
is rmly in place and could be
seen through the small hole on
the other side of the cap.

10
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
Remove the BIG from the cap and
place the Needle (V) hub into the
Needle Gripping Prongs, until the
Needle (V) is rmly and securely
held in place. (A click sound will
be heard).
! Do not try to open the cap
without inserting the safety latch
(VI) rst.
! For safety issues; when
placing the needle (V) hub into
prongs, make sure that the sharp
end of the needle is not directed at
any person.
Remove the Cap by twisting it
counter-clockwise and pulling it up.

11
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Tel: +972 - 9 - 9517444 | Fax: +972-9-9517666 | [email protected] | www.waismed.com
Rev 3 28122010
Close the Front Body (VII) of
the BIG by screwing it onto the
Rear Body (I).
! Never take out the safety Latch
(VI) when the device is loaded.
! Never point the BIG and/or the
needle at a person.
For further information, instructional movies, presentations and
article please go to www.waismed.com
LOOKING FOR A TRAINING?
Just go to www.waismed.com and click the
‘Online Training’ icon.
One of our representatives will contact you within 48 hours
and coordinate an online training session free of charge!

12
11 Galgalei HaPlada St., Herzliya, P.O.B. 12135, Zip 46722, Israel
Rev 3 28122010
Table of contents