Philips DL8765 User manual

Specifications are subject to change without notice.
© 2015 Koninklijke Philips N.V.
All rights reserved
Philips Consumer Lifestyle BV
Tussendiepen 4, 9206AD Drachten, Netherlands
Fax +31 (0)512594316
4222.100.4515.2 (10/2015)
1
2
2
3
4
DL8765
5
6
7
8
9
10
1 - 2 cm
0.4" - 0.8"
11
12
13
14
15
16
17
18
19
20
>75% recycled paper
>75% papier recyclé

English
Introduction
Congratulations on your purchase and welcome to
Philips! To fully benefit from the support that
Philips offers, register your product at
www.philips.com/welcome.
General
The Philips wrist blood pressure monitor with
Bluetooth® enables you to perform blood pressure
measurements, heart rate (pulse) measurements,
transmit data via Bluetooth® to your mobile device
and display your personal measurement results in
the Philips HealthSuite health app. The device can
also be used as a standalone device.
This user manual contains important safety infor
mation and provides step-by-step instructions for
using the blood pressure monitor.
Read this information carefully before you use the
blood pressure monitor and save it for future refer
ence.
Features:
- 46.8mm×29.3mm display with white backlight
- Measure-during-inflation technology
- Supports 2 users
Intended use
The Philips wrist blood pressure monitor is a digital
monitor intended for use in measuring blood pres
sure and heart rate in adult patient population,
with wrist circumference ranging from 13.5 cm to
21.5 cm (about 5-8.5 inches). The device is intend
ed to be used in a home environment.
The warning signs and symbols are essential to
ensure that you use this product safely and cor
rectly and to protect you and others from injury.
Below you find the meaning of the warning signs
and symbols on the label and in the user manual.
Symbol for 'follow instructions for use'.
This symbol means that the part of the
device that comes into physical contact
with the user (also known as the applied
part) is of type BF (Body Floating) accord
ing to IEC 60601-1. The applied part is the
cuff.
Symbol for 'the device complies with
European Medical Device Directive
93/42/EEC requirements'. 0344 refers to
the notified body.
Symbol for WEEE, waste electrical and
electronical equipment. Electrical waste
products should not be disposed of with
household waste. Please recycle where
facilities exist. Check with your local
authority or retailer for recycling advice
and see chapter 'Battery recycling'.
This symbol means that this product con
tains batteries which shall not be disposed
of with normal household waste
(2006/66/EC).
Indicates the manufacturer, as defined in
EU Directives 93/42/EEC.
Indicates manufacturing date.
Symbol for 'direct current'.
Symbol for the 'Bluetooth Combination
mark'. The device uses Bluetooth for com
munication.
Indicates the manufacturer's batch code.
Indicates the manufacturer's serial num
ber so that a specific medical device can
be identified.
Fuse T1A/250V Φ3.6*10CCC.
Symbol for 'Class II Equipment'. The
adapter is double insulated (Class II) and
complies with IEC 60601-1.
Symbol for indoor use only.
12
5
7
9
1011
6
8
1 2
3 4
13
14
15
16 17
1

Symbol for 'Including RF transmitter'. This
means that this device emits non-ionizing
radiation. All devices with RF transmitters
or that use RF electromagnetic energy
must have a label with this symbol.
Indicates caution.The user should consult
the instructions for use for important cau
tionary information such as warnings and
precautions that cannot, for a variety of
reasons, be presented on the medical
device itself.
Symbol for micro-USB connector.
Symbol for USB connector.
This symbol on the device means: pro
tected against access to hazardous parts
with a finger and against vertically falling
water drops when tilted up to 15 degrees.
–20ºC
60ºC
Indicates the storage and transportation
temperature limits to which the medical
device can be safely exposed: -20°C to
60°C.
Symbol for the 2 year Philips guarantee.
The Green Dot ('Der Grüne Punkt' in Ger
man) is the license symbol of a European
network of industry-funded systems for
recycling the packaging materials of con
sumer goods.
General description (Fig. 1)
1 Bluetooth® symbol
2 Battery symbol
3 Lock switch
4 Socket for micro USB plug
5 Blood pressure classification
6 Heart rate/irregular heart rate detector
7 Movement detector
8 User IDs
9 Cuff
10 On button
11 User ID button
12 Heart rate
13 Diastolic blood pressure
14 Inflation/deflation indicator
15 Systolic blood pressure
16 USB plug
17 Micro USB plug
Important safety information
Read this important information carefully before
you use the device and save it for future reference.
Warning
- Please keep the unit out of reach of infants,
children or pets, since inhalation or swallowing
of small parts can be dangerous or even fatal.
- The device is not suitable for measuring the
blood pressure of children.
- The device is not suitable for persons who have
electrical implants.
- Do not use this blood pressure monitor on any
arm where intravascular access or therapy
(such as an intravenous drip or a blood transfu
sion), or an arterio-venous shunt (A-V shunt) is
present. The temporary interference to blood
flow by the blood pressure measurement could
result in injury.
- If you had a mastectomy (breast amputation)
do not use this blood pressure monitor on the
arm on the side of the mastectomy. The inflat
ing cuff can lead to pain, trauma and further
injury in the arm on the side of the mastectomy.
- Consult your doctor if your suffer from illnesses
prior to using the device.
- No modifications of this equipment are allowed.
This may result in increased emissions or
decreased immunity of the device.
- Do not use the blood pressure monitor during
charging as it can cause injury.
- Do not touch the output of the adapter as it can
cause injury.
- Do not dispose of batteries in fire. Battery may
explode or leak.
- If you experience discomfort during a measure
ment, such as pain in the arm or other com
plaints, press the 'on' button to release the air
immediately from the cuff. Loosen the cuff and
remove it from your wrist.
- On the rare occasion of a fault causing the cuff
to remain fully inflated during measurement,
open the cuff immediately. Prolonged high
pressure (cuff pressure >300 mmHg or constant
pressure >15 mmHg for more than 3 minutes)
applied to the arm, may lead to bruises (ecchy
mosis).
- Too frequent and consecutive measurements
could cause disturbances in blood circulation
and injuries.
- Beware of strangulation particularly for children
and infants due to cables.
- Do not connect the tube to other medical
equipment, as this could lead to dangerous
injuries.
- This device is not intended for patients outside
a home environment.
- Never use any accessories or parts from other
manufacturers or that Philips does not specifi
cally recommend. Using such accessories or
parts could cause a hazardous situation for the
user or damage to the device.
- Only use the accessories and detachable parts
authorized by the manufacturer. The use of
unauthorized parts or accessories may cause
damage to the device or injury to the user.
Caution
- Always check the device before you use it. Do
not use the device if it is damaged, as this may
cause injury.

- The effectiveness of this blood pressure moni
tor has not been established in pregnant
(including pre-eclamptic) women.
- Common arrhythmias (such as atrial or ventric
ular premature beats or atrial fibrillation) and
peripheral artery disease / arteriosclerosis can
affect the performance (accuracy) of this blood
pressure monitor. Please consult your doctor
how to best use this blood pressure monitor if
you suffer from any of these conditions.
- Only use this device for its intended purpose as
described in this user manual.
- Do not confuse self-monitoring with self-diag
nosis. This device allows you to monitor your
blood pressure. Do not begin or end medical
treatment based on the measurement results.
Always consult your doctor for treatment
advice.
- Do not take any therapeutic measures on the
basis of a self-measurement. Never change
prescribed medication without consulting your
doctor. Consult your doctor if you have any
questions about your blood pressure.
- If you are taking medication, consult your
physician to determine the most appropriate
time to measure your blood pressure.
- This device is not intended for use on extremi
ties other than the wrist or for functions other
than obtaining a blood pressure measurement.
- If the cuff pressure exceeds 40kPa (300mmHg),
the unit will deflate automatically. If the cuff
does not deflate when pressures exceeds
40kPa (300mmHg), detach the cuff from the
wrist and press the 'on' button to stop inflation.
- Do not attach the cuff on the same arm on
which other monitoring medical electrical
equipment is attached simultaneously, because
this could cause temporary loss of function of
those simultaneously-used monitoring medical
electrical equipment.
- Never attach the cuff on injured skin, an injured
arm or an arm under medical treatment as this
can cause further injury.
- Do not use the device in case of existing poly
ester or nylon material allergies.
- This device is not washable. Never immerse the
device in water and do not rinse it under the
tap.
- This device is not suitable for continuous moni
toring during medical emergencies or opera
tions.
- This device cannot be used with HF (High Fre
quency) surgical equipment at the same time.
- Never use compressed air, scouring pads, abra
sive cleaning agents or aggressive liquids such
as petrol or acetone to clean the device.
- If the battery can no longer be recharged or
when the device does not function properly
anymore (see 'Specifications'), please contact
the Philips Consumer Care center in your coun
try.
- Keep the device away from fire and heat
sources, as the battery can overheat, causing
fire or bursting. The battery could explode,
causing injury or death.
- After charging, remove the micro USB plug from
the device and remove the USB plug from USB
portal.
- The equipment is not AP/APG equipment and
is not suitable for use in the presence of a flam
mable anesthetic mixture with air, with oxygen
or nitrous oxide.
- To avoid measurement errors, do not use the
device near strong electrical or magnetic fields,
for example magnets, radio transmitters,
microwave ovens.
- To avoid measurement errors, do not use the
device near a strong electromagnetic field radi
ated interference signal or electrical fast tran
sient/burst signal.
- Use this device under the right environmental
conditions as indicated in this user manual. If
not, this could affect the performance, lifetime
of the device and measurement results.
- Use a soft cloth to clean the whole unit. Do not
use any abrasive or volatile cleaners.
- Only use the micro USB cable supplied to
charge the device.
- The device does not need to be calibrated for
two years of normal use. After two years the
measurements may become less accurate.
- If you have any problems with this device, such
as setting up, malfunction, maintaining or using,
please contact Philips Consumer Care.
- Do not open, disassemble or repair the device
yourself. Modification of the device voids your
warranty.
- Report to Philips Consumer Care if any unex
pected operation or events occur.
- Dispose of accessories, detachable parts, and
the ME equipment according to the local guide
lines.
- Do not attempt to replace your blood pressure
monitors battery. It is built-in and not change
able.
- Avoid charging your blood pressure monitor in
extremely high or low temperatures (see 'Speci
fications').
- Do not clean the blood pressure monitor when
it is being charged. Always unplug the charger
first before cleaning the blood pressure moni
tor.
Compliance with standards
- The device meets the relevant standards for this
type of Class IIa electrical medical equipment
for home use.
- This Philips device complies with all applicable
standards and regulations regarding exposure
to electromagnetic fields and complies with EN
60601-1-2.
Display
Sym
bol
Description Explanation
Systolic
blood pres
sure
Maximum blood pressure.
Also see section Systolic
and diastolic pressure

Sym
bol
Description Explanation
Diastolic
blood pres
sure
Minimum blood pressure,
also see section Systolic
and diastolic pressure.
Heart rate Number of heartbeats per
minute (pulse is typically
equivalent to heart rate).
Battery sta
tus
Indicates status of battery
during charging.
Measure
ment unit
Measurement unit of blood
pressure.
Inflation
indicator
The cuff is inflating.
Deflation
indicator
The cuff is deflating.
Irregular
heart rate
detector
Irregular heart rate detec
tion during the measure
ment.
User IDs Start measurement for
selected user, and transmit
the measuring result.
Movement
detector
Moving during the mea
surement will result in an
inaccurate result.
Blood pres
sure classifi
cation
Classification of measured
blood pressure following
WHO system (see 'Blood
pressure classification').
Bluetooth®
Smart sym
bol
The device uses Bluetooth
for communication.
Heart rate
detection
Heart rate detection during
the measurement
Battery status indications
Battery
symbol
Battery status
The battery is almost empty.
+
The battery is empty.
When you measure 3 times a day starting with a
fully charged battery, the device can be used for
about 20 days until a recharge is needed. In case
of normal use, the battery can be charged around
300 times.
Note: Data will be lost when the battery is com
pletely empty.
Charging
The battery of this device is a built-in rechargeable
li-polymer battery with a capacity of 420 mAh.
Use the original USB cable supplied to charge the
battery.
When the battery is empty, it takes approx. 2 hours
to fully charge the battery of the device.
1 Put the micro USB plug in the socket of the
device (Fig. 2).
2 Put the USB cable in a USB port of a compatible
charger.
Battery charging indications
Battery
symbol
Battery charging indication
Battery charging: half full
Battery charging: almost full
Battery fully charged
Using the blood pressure
monitor
This tubeless device uses the oscillometric method
to measure blood pressure and heart rate.
Before every measurement, the unit establishes a
“zero point” equivalent to the atmospheric pres
sure. Then it starts inflating the cuff. During the
measurement, the device detects the pressure
oscillations in the blood vessels generated by the
heart pumping blood through the body. These
pressure oscillations are used to determine systolic
and diastolic blood pressure as well as heart rate.
While measuring heart rate, the device also deter
mines the small variations between the individual
heartbeats. If these variations exceed a pre-
defined threshold, the irregular heart rate detector
symbol lights up.
Systolic and diastolic pressure
The heart consists of two large chambers, the ven
tricles and two smaller chambers, the atria. The
ventricles collect blood from the atria and expel it
towards the peripheral beds of blood vessels with
in the body and the lungs. The atria collect blood
from these peripheral beds and prime the ventri
cles.
When the ventricles contract and pump blood out
of the heart, the blood pressure reaches its maxi
mum value in the cycle, which is called systolic
pressure (Fig. 3).
When the ventricles relax and are filled again with
blood, the blood pressure reaches its minimum
value in the cycle, which is called diastolic pressure
(Fig. 4).
Blood pressure classification
Consult a doctor in case of questions about your
blood pressure. Your doctor can inform you:
- About your normal blood pressure range.
- If your measuring result falls out of the range.
- Whether your blood pressure has reached a
dangerous level.
The following table shows the classification system
for the blood pressure measurements used in this
Table of contents
Other Philips Blood Pressure Monitor manuals