PLANMED SOPHIE User manual

PLANMED SOPHIE
MAMMOGRAPHY X-RAY UNIT
TECHNICAL MANUAL
788405/11En
8
1
OBL
OBL
CC
CC
OBL
OBL
LAT
LAT
kV
kV
kV
kV
mAs
mAs
mAs
mAs
CTL
CTL
LAT
LAT
3 2
SIN
SIN
DEX
DEX


Sophie Mammography X-ray unit TOC-1
TABLE OF CONTENTS
Technical Manual
Chapter A GENERAL & TECHNICAL DATA
1WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
2MANUAL VERSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
3UNIT MODELS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
4SOFTWARE REVISIONS & COMPATIBILITY . . . . . . . . . . . . . . . . . . . . A-3
5HOW TO FIND INFORMATION IN THIS MANUAL . . . . . . . . . . . . . . . . . A-4
6TECHNICAL SPECIFICATIONS (TOSHIBA E7272 X-RAY TUBE) . . . . A-5
7USER’S STATEMENT FOR PLANMED SOPHIE
(TOSHIBA E7272 X-RAY TUBE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
7.1 Definition of measurement criteria. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
7.2 Toshiba E7272 x-ray tube ratings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9
7.3 Cooling curves for x-ray tube. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-10
8TECHNICAL SPECIFICATIONS (TOSHIBA E7290 X-RAY TUBE) . . . A-11
9USER’S STATEMENT FOR PLANMED SOPHIE
(TOSHIBA E7290 X-RAY TUBE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-13
9.1 Toshiba E7290 x-ray tube ratings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-15
9.2 Cooling curves for x-ray tube. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-16
Chapter B UNPACKING & INSTALLATION
1PRE-INSTALLATION REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . B-1
1.1 Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
1.2 Environmental requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
1.3 Space requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
1.4 Radiation protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
1.5 Power requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
2UNPACKING THE UNIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
2.1 General about unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
2.2 Step-by-step unpacking instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
2.3 Driving the unit out of the transportation position . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
2.4 Attaching the unit to the floor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
2.5 Unpacking the accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7
3ATTACHMENT OF ACCESSORIES . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
3.1 Compression paddles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
3.2 Remote control attachment (optional for the mobile unit) . . . . . . . . . . . . . . . . . . . . . . B-8
3.3 Mobile Sophie special accessories attachment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
3.4 Exposure warning light (option). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
3.5 Printer (option). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
3.6 ADMARK darkroom film-marking system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8
4SETUPS AND CHECKS PRIOR TO USE . . . . . . . . . . . . . . . . . . . . . . . B-9
5RE-PACKING & TRANSPORTATION OF THE SOPHIE UNIT . . . . . . . B-10
5.1 Driving the unit into the transportation position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-10
5.2 Transporting the unit on its own wheels (non-mobile unit). . . . . . . . . . . . . . . . . . . . . B-10
6MOBILE SOPHIE HANDLING & TRANSPORTATION . . . . . . . . . . . . . B-11
6.1 Preparing the mobile unit for use after transportation . . . . . . . . . . . . . . . . . . . . . . . . B-11
6.2 Preparing the mobile unit for transportation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-12
Chapter C KEYBOARD FUNCTIONS & MODES
1KEYBOARD OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
2USER MODE FUNCTIONS SHORT-FORM . . . . . . . . . . . . . . . . . . . . . . C-2
2.1 Normal user functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
2.2 Special user settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
3SERVICE MODE FUNCTIONS SHORT-FORM . . . . . . . . . . . . . . . . . . . C-4
3.1 How to enter / exit the service mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4

TABLE OF CONTENTS
TOC-2 Sophie Mammography X-ray unit Technical Manual
3.2 Meaning of displays in the service mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
3.3 Service mode settings & special displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
4SERVICE MODE FUNCTION DETAILS . . . . . . . . . . . . . . . . . . . . . . . . . C-7
4.1 Transport position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-7
4.2 Display of internal temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-7
4.3 Displaying the real density value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-8
4.4 Displaying the exposure counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-8
4.5 Displaying the recent 49 error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-8
4.6 Print a list of the recent 49 errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-9
4.7 Ignore errors temporarily . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-9
4.8 Diagnostic display of internal signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-10
Chapter D HELP AND ERROR MESSAGES
1GENERAL ABOUT ERROR DISPLAYS . . . . . . . . . . . . . . . . . . . . . . . . . D-1
1.1 Stop-message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
1.2 Help messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
1.3 Error messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
2HELP MESSAGES SHORTFORM TABLE . . . . . . . . . . . . . . . . . . . . . . . D-2
3ERROR MESSAGES SHORTFORM TABLE . . . . . . . . . . . . . . . . . . . . . D-4
4DETAILED ERROR MESSAGE EXPLANATIONS . . . . . . . . . . . . . . . . . D-8
4.1 ERRORS 00-09 (User related, not stored in internal error list) . . . . . . . . . . . . . . . . . . D-8
4.2 ERRORS 10-16 (X-ray safety and AEC-sensor related) . . . . . . . . . . . . . . . . . . . . . . . D-9
4.3 ERRORS 20-29 (motion and compression related) . . . . . . . . . . . . . . . . . . . . . . . . . D-10
4.4 ERRORS 30-38 (tube head and power supply related) . . . . . . . . . . . . . . . . . . . . . . D-12
4.5 ERRORS 40-49 (sensor & switch related) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-13
4.6 ERRORS 50-59 (temperature & special sensor related). . . . . . . . . . . . . . . . . . . . . . D-15
4.7 ERRORS 60-68 (serial communication & CPU voltage related) . . . . . . . . . . . . . . . . D-16
4.8 ERRORS 71-78 (CPU software related) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-16
4.9 ERRORS 80-87 (CPU hardware related) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-17
4.10 ERRORS 90-97 (Erroneous sensor signals) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-18
Chapter E PREVENTIVE MAINTENANCE
1SYSTEM MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
1.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
1.2 Operator’s maintenance schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
2PREVENTIVE MAINTENANCE CHECKS . . . . . . . . . . . . . . . . . . . . . . . E-2
2.1 Preventive maintenance schedule. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2
2.2 Calibration and verification of the instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2
2.3 Determining half value layer (HVL / beam quality) . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
2.4 Checking X-ray tube efficiency (radiation output) . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4
2.5 Focal spot size measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4
2.6 Image quality evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-5
2.7 Verifying AEC-system control limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-5
3OTHER PREVENTIVE CHECKS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6
3.1 Verifying compression force measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6
3.2 Verifying filament pre-heating voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6
3.3 Verifying the anode voltage (kV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6
3.4 Verifying the tube current (mA) and the exposure time (Toshiba E7272 x-ray tube) . . E-8
3.5 Verifying the tube current (mA) and the exposure time (Toshiba E7290 x-ray tube) . . E-9
3.6 Verifying x-ray beam alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-10
3.7 Verifying light-field beam illumination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-10
3.8 Verifying visually defined x-ray fields. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-11
3.9 Verifying electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-11
Chapter F ADJUSTMENT & CALIBRATION
1SPECIAL REGIONAL & INITIAL SETUPS . . . . . . . . . . . . . . . . . . . . . . F-1
1.1 Regional setups before using the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
1.2 Other setups before using the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1

Sophie Mammography X-ray unit TOC-3
TABLE OF CONTENTS
Technical Manual
2KEYBOARD-CONTROLLED (in normal mode) . . . . . . . . . . . . . . . . . . F-3
2.1 Automatic kV mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-3
2.2 Automatic rhodium/molybdenum filter selection mode . . . . . . . . . . . . . . . . . . . . . . . . F-3
2.3 Default magnification factor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-4
2.4 Automatic release of compression after exposure. . . . . . . . . . . . . . . . . . . . . . . . . . . . F-4
2.5 Initial descent speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-4
2.6 Compression speed decrease rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-5
2.7 Intermediate compression stopping force . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-5
2.8 Twincomp compression on/off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-5
2.9 Lift motor crawling speed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-5
2.10 Setting correct time and date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-6
2.11 Density offset adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-6
2.12 AAEC contrast adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-6
2.13 Automatic film labeling (on/off). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-6
2.14 Disabling bucky grid movement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-7
2.15 Alarm sound frequency (loudness) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-7
2.16 Dimming the displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-7
2.17 mA limit (on/off) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-7
3KEYBOARD-CONTROLLED (in service mode) . . . . . . . . . . . . . . . . . . F-8
3.1 How to enter the service mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-8
3.2 Factory preset (recalling settings for all parameters). . . . . . . . . . . . . . . . . . . . . . . . . . F-9
3.3 C-arm upright position calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-12
3.4 Breast thickness measurement calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-12
3.5 Compression force measurement calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-13
3.6 X- ray tube filament preheating voltage calibration . . . . . . . . . . . . . . . . . . . . . . . . . . F-14
3.7 AEC-sensor calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-15
3.8 kV-value fine-adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-16
3.9 Special system parameters setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-17
4FINE TUNING SOPHIE AEC FOR NONSTANDARD FILM/SCREEN
COMBINATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-24
4.1 Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-24
4.2 Testing the AEC for proper tracking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-24
4.3 Adjusting the thickness, tube potential and time compensations . . . . . . . . . . . . . . . F-24
4.4 Adjusting the filter and imaging mode adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . F-26
5MECHANICAL ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-28
5.1 MAG/LOAD mechanism positional calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-28
5.2 UP/DOWN mechanism positional calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-28
5.3 X-ray beam adjustment (Toshiba E7272 X-ray tube). . . . . . . . . . . . . . . . . . . . . . . . . F-29
5.4 X-ray beam adjustment (Toshiba E7290 X-ray tube). . . . . . . . . . . . . . . . . . . . . . . . . F-35
5.5 Light field beam adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-42
5.6 Filter position adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-43
5.7 Twincomp paddle damper adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-43
5.8 Gas spring tension adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-43
6ADJUSTING SENSORS & SWITCHES . . . . . . . . . . . . . . . . . . . . . . . . F-44
6.1 Collimator mechanism limit sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-44
6.2 Compression mechanism limit sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-49
6.3 MAG/LOAD mechanism limit sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-52
6.4 Labeling head disk position sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-54
6.5 C-arm rotation sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-54
6.6 UP/DOWN mechanism sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-58
6.7 Cassette table size identification switches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-61
6.8 Lower paddle identification switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-64
Chapter G TROUBLESHOOTING & FAULTFINDING
1DISPLAYING INTERNAL SENSOR SIGNALS . . . . . . . . . . . . . . . . . . . . G-1
2ELECTRICAL PROBLEMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-1
2.1 Unit completely dead. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-1
3MOTOR & MOTION RELATED PROBLEMS . . . . . . . . . . . . . . . . . . . . . G-2
3.1 DC motors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-2
3.2 Stepper motors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-3

TABLE OF CONTENTS
TOC-4 Sophie Mammography X-ray unit Technical Manual
4MECHANICAL PROBLEMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-6
4.1 Paddle holder related . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-6
5BEFORE CALLING FOR ASSISTANCE . . . . . . . . . . . . . . . . . . . . . . . . G-8
5.1 General required information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-8
5.2 How to report image quality problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-8
5.3 How to contact Planmed head office. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . G-8
Chapter H PARTS REPLACEMENT & REPAIR
1REMOVING & REPLACING COVERS . . . . . . . . . . . . . . . . . . . . . . . . . . H-1
1.1 Base covers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-2
1.2 Telescopic column covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-2
1.3 Stationary column covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-3
1.4 C-arm covers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-4
1.5 Hood . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-6
1.6 Cassette table cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-7
1.7 Collimator cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-7
2SOFTWARE VERSION UPDATING . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-8
2.1 Replacing software chips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-8
3REPLACING FUSES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-10
3.1 Mains fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-10
3.2 Power supply fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-10
4LIGHTFIELD BULB REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . H-12
5REPLACING THE X-RAY TUBEHEAD . . . . . . . . . . . . . . . . . . . . . . . . . H-13
5.1 Preparations before changing the tubehead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-14
5.2 Removal of the tubehead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-17
6REPLACING THE POWER SUPPLY . . . . . . . . . . . . . . . . . . . . . . . . . . H-18
7REPLACING CIRCUIT BOARDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-19
7.1 Location of PC-boards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-19
7.2 Boards in the base & column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-20
7.3 C-arm pc-boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-20
7.4 Tubehead assembly boards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-22
7.5 Keyboards & displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-22
8REPLACING SENSORS & SWITCHES . . . . . . . . . . . . . . . . . . . . . . . . H-23
8.1 Placement of sensors & switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-23
8.2 Limit sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-24
8.3 Lift motor rotation sensor replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-24
8.4 C-arm motor rotation sensor replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-25
8.5 MAG-motor rotation sensor replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-25
9MOTOR REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-26
9.1 Lift motor replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-26
9.2 C-arm rotation motor replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-28
9.3 Compression motor replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-34
9.4 Magnification motor replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-35
10 REPLACING THE LABELING HEAD . . . . . . . . . . . . . . . . . . . . . . . . . . H-37
11 GAS SPRING REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-38
11.1 C-arm is functional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-38
11.2 C-arm is not rotating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-38
11.3 Replacing the gas-spring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-38
12 REPLACING PARTS IN COMPRESSION MECHANISM . . . . . . . . . . . H-39
12.1 Replacing the twin-comp damper. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H-39
Chapter I SCHEMATICS & DIAGRAMS
COPYRIGHT PLANMED 2000-10
PUBLICATION NUMBER 788405 revision 11

Sophie Mammography X-ray unit A-1
Chapter
Technical Manual
AGENERAL & TECHNICAL DATA
1 WARNINGS
WARNING
THE FOLLOWING WARNINGS, CAUTIONS AND NOTES MUST ALWAYS BE CONSIDERED
WHILE SERVICING, IN ORDER TO AVOID EITHER PERSONAL INJURY OR DAMAGE TO
THE UNIT.
CAUTION RADIATION SAFETY RULES
Some procedures described in this manual produces X-ray radiation. Always follow rules
for radiation safety.
Never touch or scratch the surface of the x-ray tube radiation window. The window is
made of beryllium that can be harmful if inhaled or swallowed.
Never attempt to open the TUBE-HEAD. It does not contain any serviceable parts and this
would make it inoperable and void the warranty.
Never make any exposures without the Mo/Al filter in the beam, or without the beam limit-
ing device (collimator) in place. Otherwise the radiation safety cannot be guaranteed.
CAUTION ELECTRICAL SAFETY RULES
The unit contains hazardous voltages. While servicing internal parts, always disconnect
the unit from the mains (if possible) by removing the plug from the wall outlet, and wait for
2 minutes before touching any electrical parts.
Always replace the fuses with ones of the same type and rating. Otherwise patient, opera-
tor or equipment safety cannot be guaranteed.
The circuit boards can be damaged due to static discharges and requires careful handling
and storage.
CAUTION GENERAL SAFETY RULES
The unit must be serviced only by qualified personnel, trained by PLANMED Repairs and
parts replaced by unqualified personnel carry no warranty.
Periodical maintenance as described in this manual must be performed on a regular basis,
to ensure the safety and image quality of the unit.
Some procedures described in the unit could jeopardize the function of the unit, if not fol-
lowed as stated.

Chapter A - GENERAL & TECHNICAL DATA
A-2 Sophie Mammography X-ray unit
MANUAL VERSIONS
Technical Manual
2 MANUAL VERSIONS
PLANMED pursues a policy of continuous product development. Although every effort is made to
produce up-to-date product documentation this manual should not be regarded as an infallible
guide to current production specifications. Planmed reserves the right to make changes to speci-
fications without prior notice. This manual is valid for all versions of the Sophie Mammogra-
phy X-ray units with SID 650mm. In case you have the unit with SID 600mm, you need a
technical manual revision 9.
3UNITMODELS
All Sophie Mammography X-ray unit models are electrically and mechanically identical, except
for some minor details. This manual is valid for all versions since the differences are small.
The darkroom versus daylight differs only in the size and dimensions of the cassette table and
film marking device.
The mobile versus the stationary units differs only in the base wheel system and the X-ray tube
locking system (for transportation).
Daylight units can be converted to darkroom units (and vice versa), please contact PLANMED for
details.
NOTE THIS TECHNICAL MANUAL IS VALID FOR SOFTWARE REVISIONS: TUBE CPU 2.41
(OR LATER) AND REAR CPU 4.41 (OR LATER).
If you find that your SOPHIE doesn’t have some keyboard functions, or its keyboard
functions differs from procedures described in this manual, then either the software
must be replaced with at least the above mentioned revision, or you have to relate to
an older version of this manual. The former procedure is recommended.

Chapter A - GENERAL & TECHNICAL DATA
Sophie Mammography X-ray unit A-3
SOFTWARE REVISIONS & COMPATIBILITY
Technical Manual
4 SOFTWARE REVISIONS & COMPATIBILITY
Background
Since this unit is fully microprocessor controlled, its operation is controlled by the currently
installed software. PLANMED reserves the right to improve functionality or to add new features to
the product by modifying the software and/or hardware. Manuals are updated accordingly and
technical bulletins published in order to inform about these changes or additions.
Hardware compatibility
As a general rule; new software releases are compatible with all existing hardware. New software
releases (or possible upgrading kits containing new software) are normally directly compatible
with existing hardware, but not necessarily without some re-calibration procedures.
Software numbering & compatibility
Since the unit contains two communicating microprocessors, these must be able to “understand”
each other.
Incompatible software versions should never be used together, the unit would either be totally
non-functional, produce continuous error codes, or some important features or functions could be
missing. However, no damage or safety hazard would result if two non-compatible software chips
are accidentally used together.
The following table contains information about compatible software versions. Use versions
together only as listed in this table.
Table 1: SOFTWARE COMPATIBILITY LIST
REAR CPU TUBE CPU REAR CPU TUBE CPU
1.20 - 1.28 1.20 - 1.28 1.80 1.80, 2.00
1.59 1.59 1.81 1.81, 2.01
1.60 1.60 4.00 2.01
1.63 1.63 4.01/2.01 2.04
1.64 1.64 4.02/2.02 2.04
1.65 1.64 4.06/2.06 2.04
1.66 1.64 4.20 2.20
1.67 1.64 4.21 2.22
1.68 1.64 4.22 2.23
1.69 1.69 4.23 2.25
1.70 1.69 4.30 2.31
1.71 1.69 4.40 2.40
1.72 1.69 4.41 2.41
1.73 1.69
1.74 1.69
1.75 1.69, 1.70
1.76 1.75, 1.76
1.77 1.77, 1.78,
1.79

Chapter A - GENERAL & TECHNICAL DATA
A-4 Sophie Mammography X-ray unit
HOW TO FIND INFORMATION IN THIS MANUAL
Technical Manual
5 HOW TO FIND INFORMATION IN THIS MANUAL
For normal operating instructions, please refer to the “SOPHIE USER’S MANUAL”. However,
this manual contains also full explanations of all user accessible adjustments that can be done
with the keyboards.
For navigation, please refer to the table of contents in the beginning of this manual. The follow-
ing is a brief description of the chapters in this manual.
“GENERAL & TECHNICAL DATA” on page A-1 (this chapter)
This chapter contains general information as well detailed technical specifications of the unit.
Here you can find answers to all questions related to the unit specifications. Full technical speci-
fication details can be found here.
“UNPACKING & INSTALLATION” on page B-1
This chapter contains complete instructions how to unpack and install the unit, with information
about installation requirements. Also information about transportation and handling is included.
“KEYBOARD FUNCTIONS & MODES” on page C-1
This chapter contains a complete short-form presentation of all available special keyboard func-
tions. For explanation of normal user operation. Please refer to the “SOPHIE USER’S MANUAL”.
The special display modes are also described here.
The detailed descriptions of all adjustments performed with the help of the keyboards can be
found in the chapter “ADJUSTMENT & CALIBRATION”on page F-1.
“HELP AND ERROR MESSAGES” on page D-1
This chapter contains general information as well as a complete short-form presentation of all
available special messages (STOP, HELP, ERROR) that can appear on the displays. A detailed
description of the error messages and their reasons is also included.
“PREVENTIVE MAINTENANCE” on page E-1
This chapter describes the operators maintenance. Complete step-by-step instructions how to
check the performance of the unit are also included. Some of these checks must be performed
annually and before using the unit.
“ADJUSTMENT & CALIBRATION” on page F-1
This chapter contains detailed instructions of how to perform keyboard-controlled adjustments
and calibrations, mechanical adjustments and sensor & switch adjustments.
“GENERAL & TECHNICAL DATA” on page A-1
For help and guidance with general troubleshooting and faultfinding, please refer to this chapter.
It contains information of how to proceed in possible problematic situations.
“PARTS REPLACEMENT & REPAIR” on page H-1
This chapter contains detailed instructions how to replace both mechanical and electrical parts. It
also reminds of what possible calibrations or adjustments must be performed after replacing par-
ticular parts.
“SCHEMATICS & DIAGRAMS” on page I-1
This chapter contains electrical wiring diagrams and schematics.

Chapter A - GENERAL & TECHNICAL DATA
Sophie Mammography X-ray unit A-5
TECHNICAL SPECIFICATIONS (TOSHIBA E7272 X-RAY
Technical Manual
6 TECHNICAL SPECIFICATIONS (TOSHIBA E7272 X-RAY TUBE)
ORIGINAL MANUFACTURER
PLANMED Oy, Asentajankatu 6, 00810 Helsinki, FINLAND
phone: +358-9-7590 5300, fax: +358-9-7590 5555
MODEL AND APPROVALS
MODELS: PLANMED SOPHIE, PLANMED SOPHIE MOBILE
APPROVALS: IEC safety class: I, IEC degree of protection: B, IEC enclosure class: IP20
Complies with IEC 601-1 safety and IEC 601-2-7 radiation regulations
X-RAY TUBE ASSEMBLY
X-ray tube Toshiba E7272
Anode type rotating anode (3000 rpm at 50 Hz)
Anode braking Automatic, electrical
Anode thermal capacity 300 000 HU
Anode target material Molybdenum
Anode target angle 16 degrees
Tube port material Beryllium
Focal spot sizes 0.3mm / 0.1 x 0.07 mm
Filtration 30µm Mo and 0.5mm Al or
30µm Mo and 25µm Rh
GENERATOR ASSEMBLY (for both tube types if not otherwise specified)
Generator constant potential, high frequency
Anode voltage range 20 - 35kV ± 2kV, virtually DC
Anode current range/0.3mm focus 70 - 110mA (70 - 120 mA @ 60Hz) ± 5mA
Anode current range/0.1mm focus 10 - 22mA (12 - 22 mA @ 60Hz) ± 1mA
mAs range/0.3mm focus 10 - 500mAs (optional 600mAs in GB) ± 4mAs
or ±10% (whichever is larger)
mAs range/ 0.1mm focus 10 - 200 mAs
Exposure time /0.3mm focus 0.1 - 5 seconds (optional 6 seconds) ±5%
Exposure time / 0.1mm focus 0.1 - 9.9 seconds ±5%
Cooling automatically controlled (internal fan)
Line voltage 208 - 240V~, 50 or 60Hz - single phase
Line voltage compensation automatic ±10%
Line voltage regulation ±10% (absolute minimum/maximum 187V~/265V~)
Fuses 2 x 15AT / 250V
Total power consumption max. 100VA idle, 4000VA max. for 5 seconds
Power factor > 0.95
Mode of operation intermittent
Maximum mains resistance Ω1 ¾
Maximum continuous
heat dissipation < 250W
BUCKY GRID SPECIFICATION
Grid manufacturer Gilardoni
Grid type GRIDGIL HT; Fiber interspaced vertical
Grid movement Variable amplitude oscillation
Grid ratio r5
Grid lines per cm N36
Focusing distance f065
Application limits 18x24: f150 f291
24x30: f153 f282
Material of absorbing strips Pb
Contrast improvement ratio K1.47
Selectivity Σ6.04
Grid exposure factor B1.69
Transmission of primary radiation Tp0.75
Max. deviation between the central
line indication and the true
central line ∆2

Chapter A - GENERAL & TECHNICAL DATA
A-6 Sophie Mammography X-ray unit
TECHNICAL SPECIFICATIONS (TOSHIBA E7272 X-RAY
Technical Manual
AUTOMATIC EXPOSURE CONTROL (AEC)
Sensors three selectable large area solid-state sensors
Density adjustment 15 density steps (13% change in OD per step)
plus fine adjustment ± 5 steps (0,05 OD per step)
Function modes Normal AEC, Advanced AEC (AAEC)
C-ARM
Rotation motorized, range -135°to +180°
Vertical movement motorized, range 795mm to 1350mm (31.3in to 53.1in)
SID 650mm
Compression fully motorized, variable speed, variable angle (twincomp)
Cassette sizes 18 x 24cm and 24 x 30cm
Magnification motorized, variable range from 1.3 to 1.8
MECHANICAL DATA
Dimensions (H x D x W) 1000 x 900 x 755 (mm) minimum
(36.6 x 35.0 x 29.7 inch minimum)
Weight 160kg (352lbs), unpacked
Color RAL 9016

Chapter A - GENERAL & TECHNICAL DATA
Sophie Mammography X-ray unit A-7
USER’S STATEMENT FOR PLANMED SOPHIE (TOSHIBA
Technical Manual
7 USER’S STATEMENT FOR PLANMED SOPHIE (TOSHIBA E7272 X-RAY
TUBE)
Radiation leakage technique factors
The maximum-rated peak tube potential is 35 kVp and the maximum rated continuous tube cur-
rent is 3.5mA for the maximum-rated peak tube voltage.
Minimum filtration
The Beam-limiting device contains three different filtrations 30µm molybdenum, 25µm rhodium
and 0.50mm aluminum. The measured half-value layers are 0.343mmAl, 0.373mmAl and
0.377mmAl at 30kVp. These measured values correspond to aluminum equivalents of 0.55 -
0.65mmAl.
Rate line voltage
208 - 240V~±10%. Line voltage regulation 10%.
Maximum line current
Maximum 22 Amperes at absolute minimum stated mains voltage (187V~)
Technique factors that constitute the maximum line current condition
27kV / 110mA
Generator rating and duty cycle
3kW. The duty cycle and wait period can be calculated using the following formulas:
•Wait period = Tw = kV*mA*exp. Time / 500W
•Duty Cycle = 1 / (1 + kV*mA / 500W)
Maximum deviation of peak tube potential from indicated value
±2kV
Maximum deviation of tube current from indicated value
±5mA large focal spot, ±1mA small focal spot
Maximum deviation of exposure time tube current product from indicated value
±4mAs or ±10%, whichever is larger
Maximum deviation of exposure time from indicated value
±5%
7.1 Definition of measurement criteria
Exposure time
The beginning and end points of the exposure time are defined at 70% of the peak radiation
waveform measured with a calibrated x-ray monitor.
Peak tube potential
Is defined as the measured high voltage mean value measured with a calibrated non-invasive
kVp meter.
Tube current
Is defined using the resistance and voltage over the feedback resistor measured with a calibrated
multimeter. The mA value is then the voltage divided by the resistance.
mAs product
Is defined as the product of the exposure time and the tube current measurements.
The nominal x-ray voltage together with the highest x-ray tube current obtainable
from the high-voltage generator when operated at its nominal x-ray tube voltage
35kV 80mA / 60Hz - large focal spot, 35kV 70mA / 50Hz - large focal spot
35kV 16mA / 60Hz - small focal spot, 35kV 14mA / 50Hz - small focal spot

Chapter A - GENERAL & TECHNICAL DATA
A-8 Sophie Mammography X-ray unit
USER’S STATEMENT FOR PLANMED SOPHIE (TOSHIBA
Technical Manual
The highest x-ray tube current together with the highest x-ray tube voltage obtainable
from the high-voltage generator when operated at its highest x-ray tube current
120mA 24kV / 60Hz - large focal spot, 110mA 24kV / 50Hz - large focal spot
22mA 26kV / 60Hz - small focal spot, 22mA 24kV / 50Hz - small focal spot
The x-ray tube voltage and x-ray tube current which result in the highest electric output
power
27kV 110mA / 60Hz - large focal spot, 34kV 80mA / 50Hz - large focal spot
26kV 22mA / 60Hz - small focal spot, 24kV 22mA / 50Hz - small focal spot
The nominal electric power for a load time of 0.1s and at the nominal x-ray tube voltage
35kV 80mA / 60Hz - 2800W large focal spot, 35kV 70mA / 50Hz - 2450W large focal spot
35kV 16mA / 60Hz - 560W small focal spot, 35kV 14mA / 50Hz - 490W small focal spot
Reference current time product
large focal spot: for all kV values 40mA / 0.1s / 4mAs
small focal spot: for all kV values 10mA / 0.1s / 1mAs
The lowest current time product within the specified range of compliance
large focal spot: 30mA / 0.1s / 3mAs
small focal spot: 10mA / 0.1s / 1mAs
The nominal shortest irradiation time in the AEC-mode
20ms

Chapter A - GENERAL & TECHNICAL DATA
Sophie Mammography X-ray unit A-9
USER’S STATEMENT FOR PLANMED SOPHIE (TOSHIBA
Technical Manual
7.2 Toshiba E7272 x-ray tube ratings
Manual exposures
With manual exposures the tube kV, mAs and focal spot can be selected by the user. The mA
value is automatically determined by the selected kV value, and the exposure time is determined
by the selected mAs value. The mA value used is always the highest possible value, that
depends on the kV value, focal spot, exposure time and line frequency (see following table on the
next page). The maximum exposure time is 6.0 sec. for the large focal spot, and 9.9 sec. for the
small focal spot. The minimum mAs value is 10 mAs and it can be increased in approx. 20%
steps to the maximum (see values below). The maximum mAs value for the large focal spot is
either 300, 500, 550 or 600 mAs, depending on the value of the service mode parameter 6. The
maximum mAs value for the small focal spot is 200 mAs.
The possible mAs values are: 10, 12, 14, 16, 18, 20, 25, 30, 35, 40, 50, 60, 70, 85, 100, 120, 140,
160, 180, 200, 225, 260, 300, 350, 400, 450, 500, 550, 600.
Automatic exposures
Automatic exposures are terminated when the radiation detector indicates that the required dose
has been reached. In this mode the kV is selected by the operator and mA is chosen to be the
maximum allowed mA for the selected kV and Hz (line frequency).
If the exposure time is longer than 1 sec (5 sec for the small focal spot) then a falling load tech-
nique is used. The exposure is started at the mA as specified in the following tables. The accu-
mulated dose is checked at 0.45 sec (2.45 sec for the small focal spot) and if it has reached more
than half of the required dose, the exposure can be continued at the same mA level, and will be
terminated within 1 second (5 seconds for the small focal spot) - well before the tube is over-
loaded.
If the accumulated dose is detected to be less than half of the required, a longer exposure time
than the one specified for the tube current can be expected. Therefore the tube current is
reduced at this time (after 0.45 or 2.45 sec exposure) to a lower mA that, when added to the
already produced tube loading, will not exceed the maximum tube rating. The falling load current
levels are listed in the two following tables.
Thermal ratings
Two separate methods are used to protect the tube from overload. One is measuring the tube
head enclosure temperature and if it exceeds 60 °C no further exposures are allowed until it falls
below 60 °C.
The other method keeps the average input power into the tube below or equal to 500W. The
exposure parameters are limited as indicated by the following tables, but this feature forces a
waiting time between the exposures to ensure that the average power into the tube never can
exceed this limit. For example if an exposure of 24 kV, 100 mA, 1 sec is performed a wait period
of 5 seconds is calculated and no further exposures are allowed until this time has elapsed.
There is no way to bypass these safety feature under any circumstances, and it is thus virtually
impossible to overload the X-ray tube.

Chapter A - GENERAL & TECHNICAL DATA
A-10 Sophie Mammography X-ray unit
USER’S STATEMENT FOR PLANMED SOPHIE (TOSHIBA
Technical Manual
7.3 Cooling curves for x-ray tube
Table 2: Maximum mA for Tube E7272 vs. different modes and line frequencies
kV
MANUAL EXPOSURE MODE AUTOMATIC (AEC) EXPOSURE MODE
Large focal spot
(0.3 mm)
Small focal spot
(0.1 mm)
Large focal spot
(0.3 mm)
Small focal spot
(0.1 mm)
0 - 1
sec
1 - 5
sec
1 - 6
sec
0 - 5
sec
> 5
sec
0 - 1
sec
0.45 - 5
sec
0.45 - 6
sec
0 - 5
sec
2.45 -10
sec
50
Hz
60
Hz
50
Hz
60
Hz
50
Hz
60
Hz
50
Hz
60
Hz
50
Hz
60
Hz
50
Hz
60
Hz
50
Hz
60
Hz
50
Hz
60
Hz
50
Hz
60
Hz
50
Hz
60
Hz
20 70 70 70 70 80 80 20 20 20 20 70 70 70 70 80 85 20 20 18 20
21 80 80 80 80 90 90 20 20 18 20 80 80 80 80 90 95 20 20 18 18
2290909090100100222218189090909010010522221618
23 100 100 100 100 100 100 22 22 18 18 100 100 98 100 98 108 22 22 16 18
24 110 110 100 100 90 90 22 22 16 18 110 110 98 108 93 103 22 22 16 17
25 100 110 90 100 90 90 20 22 16 16 100 110 95 97 90 92 20 22 14 15
26 100 110 90 90 90 90 20 22 16 16 100 110 91 98 89 96 20 22 14 15
27 100 110 80 90 80 80 18 20 14 16 100 110 87 87 83 83 18 20 14 14
28 90 100 80 90 80 80 18 20 14 14 90 100 85 88 79 82 18 20 12 14
29 90 100 80 80 80 80 18 20 14 14 90 100 82 88 78 84 18 20 12 13
30 90 90 80 80 70 70 16 18 12 14 90 90 80 77 79 72 16 18 13 13
31 80 90 70 80 70 70 16 18 12 14 80 90 77 78 73 72 16 18 10 12
32 80 90 70 80 70 70 16 18 12 12 80 90 75 78 69 72 16 18 10 11
33 80 90 70 70 60 60 16 16 12 12 80 90 72 78 67 73 16 16 10 11
34 80 80 70 70 60 60 14 16 12 12 80 80 70 68 65 65 14 16 11 10
35 70 80 60 70 60 60 14 16 10 12 70 80 67 68 64 65 14 16 11 10
0
20
40
60
80
100
120
140
160
180
200
210
012345678910
HEAT STORAGE (kJ)
ANODE THERMAL CHARACTERISTICS
TIME (min)
707W
565W
424W
283W
141W
COOLING
0
100
200
300
400
500
0 50 100 150 200 250
HEAT STORAGE (kHU)
TIME (min)
600
700
TUBE HOUSING
HEAT DISSIPATION CAPACITY

Chapter A - GENERAL & TECHNICAL DATA
Sophie Mammography X-ray unit A-11
TECHNICAL SPECIFICATIONS (TOSHIBA E7290 X-RAY
Technical Manual
8 TECHNICAL SPECIFICATIONS (TOSHIBA E7290 X-RAY TUBE)
ORIGINAL MANUFACTURER
PLANMED Oy, Asentajankatu 6, 00810 Helsinki, FINLAND
phone: +358-9-7590 5300, fax: +358-9-7590 5555
MODEL AND APPROVALS
MODELS: PLANMED SOPHIE, PLANMED SOPHIE MOBILE
APPROVALS: IEC safety class: I, IEC degree of protection: B, IEC enclosure class: IP20
Complies with IEC 601-1 safety and IEC 601-2-7 radiation regulations
X-RAY TUBE ASSEMBLY
X-ray tube Toshiba E7290
Anode type rotating anode (9700 rpm at 180 Hz)
Anode braking Automatic, electrical
Anode thermal capacity 300 000 HU
Anode target material Molybdenum
Anode target angle 16 degrees (large focal spot) / 10 degrees (small focal spot)
Tube port material Beryllium
Focal spot sizes 0.3mm / 0.1 mm
Filtration 30µm Mo and 0.5mm Al or
30µm Mo and 25µm Rh
GENERATOR ASSEMBLY
Generator constant potential, high frequency
Anode voltage range 20 - 35kV ± 2kV, virtually DC
Anode current range/0.3mm focus 80 - 110mA ± 5mA
Anode current range/0.1mm focus 21 - 35mA ± 1mA
mAs range/0.3mm focus 10 - 500mAs (optional 600mAs in GB) ± 4mAs
or ±10% (whichever is larger)
mAs range/ 0.1mm focus 10 - 310 mAs
Exposure time /0.3mm focus 0.1 - 5 seconds (optional 6 seconds) ±5%
Exposure time / 0.1mm focus 0.1 - 9.9 seconds ±5%
Cooling automatically controlled (internal fan)
Line voltage 208 - 240V~, 50 or 60Hz - single phase
Line voltage compensation automatic ±10%
Line voltage regulation ±10% (absolute minimum/maximum 187V~/265V~)
Fuses 2 x 15AT / 250V
Total power consumption max. 100VA idle, 4000VA max. for 5 seconds
Power factor > 0.95
Mode of operation intermittent
Maximum mains resistance ≤ 1Ω
Maximum continuous
heat dissipation < 250W
BUCKY GRID SPECIFICATION
Grid manufacturer Gilardoni
Grid type GRIDGIL HT; Fiber interspaced vertical
Grid movement Variable amplitude oscillation
Grid ratio r5
Grid lines per cm N36
Focusing distance f065
Application limits 18x24: f150 f291
24x30: f153 f282
Material of absorbing strips Pb
Contrast improvement ratio K1.47
Selectivity Σ6.04
Grid exposure factor B1.69
Transmission of primary radiation Tp0.75
Max. deviation between the central
line indication and the true
central line ∆2

Chapter A - GENERAL & TECHNICAL DATA
A-12 Sophie Mammography X-ray unit
TECHNICAL SPECIFICATIONS (TOSHIBA E7290 X-RAY
Technical Manual
AUTOMATIC EXPOSURE CONTROL (AEC)
Sensors three selectable large area solid-state sensors
Density adjustment 15 density steps (13% change in OD per step)
plus fine adjustment ± 5 steps (0,05 OD per step)
Function modes Normal AEC, Advanced AEC (AAEC)
C-ARM
Rotation motorized, range -135°to +180°
Vertical movement motorized, range 795mm to 1350mm (31.3in to 53.1in)
SID 650mm
Compression fully motorized, variable speed, variable angle (twincomp)
Cassette sizes 18 x 24cm and 24 x 30cm
Magnification motorized, variable range from 1.3 to 1.8
MECHANICAL DATA
Dimensions (H x D x W) 1000 x 900 x 755 (mm) minimum
(36.6 x 35.0 x 29.7 inch minimum)
Weight 160kg (352lbs), unpacked
Color RAL 9016

Chapter A - GENERAL & TECHNICAL DATA
Sophie Mammography X-ray unit A-13
USER’S STATEMENT FOR PLANMED SOPHIE (TOSHIBA
Technical Manual
9 USER’S STATEMENT FOR PLANMED SOPHIE (TOSHIBA E7290 X-RAY
TUBE)
Radiation leakage technique factors
The maximum-rated peak tube potential is 35 kVp and the maximum rated continuous tube cur-
rent is 3.5mA for the maximum-rated peak tube voltage.
Minimum filtration
The Beam-limiting device contains three different filtrations 30µm molybdenum, 25µm rhodium
and 0.50mm aluminum. The measured half-value layers are 0.343mmAl, 0.373mmAl and
0.377mmAl at 30kVp. These measured values correspond to aluminum equivalents of 0.55 -
0.65mmAl.
Rate line voltage
208 - 240V~±10%. Line voltage regulation 10%.
Maximum line current
Maximum 22 Amperes at absolute minimum stated mains voltage (187V~)
Technique factors that constitute the maximum line current condition
33kV / 97mA
Generator rating and duty cycle
3.2kW. The duty cycle and wait period can be calculated using the following formulas:
•Wait period = Tw = kV*mA*exp. Time / 500W
•Duty Cycle = 1 / (1 + kV*mA / 500W)
Maximum deviation of peak tube potential from indicated value
±2kV
Maximum deviation of tube current from indicated value
±5mA large focal spot, ±1mA small focal spot
Maximum deviation of exposure time tube current product from indicated value
±4mAs or ±10%, whichever is larger
Maximum deviation of exposure time from indicated value
±5%
Definition of measurement criteria
Exposure time
The beginning and end points of the exposure time are defined at 70% of the peak radiation
waveform measured with a calibrated x-ray monitor.
Peak tube potential
Is defined as the measured high voltage mean value measured with a calibrated non-invasive
kVp meter.
Tube current
Is defined using the resistance and voltage over the feedback resistor measured with a calibrated
multimeter. The mA value is then the voltage divided by the resistance.
mAs product
Is defined as the product of the exposure time and the tube current measurements.

Chapter A - GENERAL & TECHNICAL DATA
A-14 Sophie Mammography X-ray unit
USER’S STATEMENT FOR PLANMED SOPHIE (TOSHIBA
Technical Manual
The nominal x-ray voltage together with the highest x-ray tube current obtainable
from the high-voltage generator when operated at its nominal x-ray tube voltage
35kV 91mA - large focal spot
35kV 28mA - small focal spot
The highest x-ray tube current together with the highest x-ray tube voltage obtainable
from the high-voltage generator when operated at its highest x-ray tube current
110mA 29kV - large focal spot
35mA 28kV - small focal spot
The x-ray tube voltage and x-ray tube current which result in the highest electric
output power
33kV 97mA - large focal spot
31kV 32mA - small focal spot
The nominal electric power for a load time of 0.1s and at the nominal x-ray tube
voltage
35kV 91mA - 3185W large focal spot
35kV 28mA 980W small focal spot
Reference current time product
large focal spot: for all kV values 40mA / 0.1s / 4mAs
small focal spot: for all kV values 10mA / 0.1s / 1mAs
The lowest current time product within the specified range of compliance
large focal spot: 30mA / 0.1s / 3mAs
small focal spot: 10mA / 0.1s / 1mAs
The nominal shortest irradiation time in the AEC-mode
20ms
Table of contents
Popular Medical Equipment manuals by other brands

Leica BIOSYSTEMS
Leica BIOSYSTEMS CM3600 XP Instructions for use

NHS
NHS Cough Assist manual

Coaguchek
Coaguchek XS System Getting started

Tektronix
Tektronix Trimode TDP7700 Series Technical reference

Cobi Rehab
Cobi Rehab XXL-Rehab Comfort cushion user manual

Coopers of Stortford
Coopers of Stortford F890 Instructions for use

Ultra Ankle
Ultra Ankle Ultra Zoom Fitting instructions

Interacoustics
Interacoustics VRA201 Instructions for use

Schiller
Schiller Cardiovit AT-10 Service handbook

Sonoscape
Sonoscape Flagship S9 Quick manual

Yuwell
Yuwell 7E-A user manual

Sunrise Medical
Sunrise Medical TRU-TRAC TT-92B Series operating instructions